OPTIFLUX 2000 Quick Start

Electromagnetic flow sensor

The documentation is only complete when used in combination with the relevant
documentation for the signal converter.

© KROHNE 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


CONTENTS OPTIFLUX 2000

1 Safety instructions 3

2 Installation 4

2.1 Scope of delivery............................................................................................................... 4


2.2 Device description ............................................................................................................ 5
2.3 Nameplate ........................................................................................................................ 6
2.4 Storage ............................................................................................................................. 7
2.5 Transport .......................................................................................................................... 7
2.6 Pre-installation requirements ......................................................................................... 7
2.7 General requirements ...................................................................................................... 8
2.7.1 Vibration .................................................................................................................................. 8
2.7.2 Magnetic field.......................................................................................................................... 8
2.8 Installation conditions ...................................................................................................... 9
2.8.1 Inlet and outlet ........................................................................................................................ 9
2.8.2 Bends in 2 or 3 dimensions..................................................................................................... 9
2.8.3 T-section ................................................................................................................................. 9
2.8.4 Bends .................................................................................................................................... 10
2.8.5 Open feed or discharge......................................................................................................... 10
2.8.6 Flange deviation .................................................................................................................... 11
2.8.7 Pump ..................................................................................................................................... 11
2.8.8 Control valve ......................................................................................................................... 11
2.8.9 Air venting and vacuum forces ............................................................................................. 12
2.8.10 Mounting position................................................................................................................ 13
2.9 Mounting ......................................................................................................................... 13
2.9.1 Torques and pressures......................................................................................................... 13
2.10 Temperatures ............................................................................................................... 16

3 Electrical connections 17

3.1 Safety instructions.......................................................................................................... 17


3.2 Grounding ....................................................................................................................... 17
3.3 Virtual reference for IFC 300 (C, W and F version) ........................................................ 19
3.4 Connection diagrams ..................................................................................................... 19

4 Technical data 20

4.1 Dimensions and weights ................................................................................................ 20


4.2 Vacuum load ................................................................................................................... 23

2 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 SAFETY INSTRUCTIONS 1

Warnings and symbols used


DANGER!
This information refers to the immediate danger when working with electricity.

DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.

WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.

CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.

INFORMATION!
These instructions contain important information for the handling of the device.

HANDLING
• This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
i RESULT
This symbol refers to all important consequences of the previous actions.

Safety instructions for the operator


CAUTION!
Installation, assembly, start-up and maintenance may only be performed by appropriately
trained personnel. The regional occupational health and safety directives must always be
observed.

LEGAL NOTICE!
The responsibility as to the suitability and intended use of this device rests solely with the user.
The supplier assumes no responsibility in the event of improper use by the customer. Improper
installation and operation may lead to loss of warranty. In addition, the "Terms and Conditions of
Sale" apply which form the basis of the purchase contract.

INFORMATION!
• Further information can be found on the supplied CD-ROM in the manual, on the data sheet,
in special manuals, certificates and on the manufacturer's website.
• If you need to return the device to the manufacturer or supplier, please fill out the form
contained on the CD-ROM and send it with the device. Unfortunately, the manufacturer
cannot repair or inspect the device without the completed form.

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 3


2 INSTALLATION OPTIFLUX 2000

2.1 Scope of delivery


INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.

INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.

INFORMATION!
The remote version will arrive in two cartons. One carton contains the converter and one carton
contains the sensor.

Figure 2-1: Scope of delivery


1 Ordered flowmeter
2 Product documentation
3 Factory calibration report
4 CD-ROM with product documentation in available languages
5 Grounding rings (optional)
6 Signal cable (remote versions only)

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.

4 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 INSTALLATION 2

2.2 Device description


Your measuring device is supplied ready for operation. The factory settings for the operating
data have been made in accordance with your order specifications.

The following versions are available:


• Compact version (the signal converter is mounted directly on the measuring sensor)
• Remote version (a measuring sensor with connection box and a separate signal converter)

Figure 2-2: Device versions


1 Remote version
2 Compact version with IFC 300 signal converter
3 Compact version with IFC 100 (0°) signal converter
4 Compact version with IFC 100 (45°) signal converter
5 Compact version with IFC 050 (10°) signal converter

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 5


2 INSTALLATION OPTIFLUX 2000

2.3 Nameplate
INFORMATION!
Check the device nameplate to ensure that the device is delivered according to your order.
Additional information (a.o correct supply voltage), can be found in the documentation of the
signal converter.

Figure 2-3: Example of nameplate


1 Name and address of the manufacturer
2 Type designation of the flowmeter and CE sign with number(s) of notified body / bodies
3 Calibration data
4 PED data

6 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 INSTALLATION 2

2.4 Storage
• Store the device in a dry and dust-free location.
• Avoid lasting direct exposure to the sun.
• Store the device in its original packaging.
• Storage temperature: -50 ...+70°C / -58...+158°F

2.5 Transport
Signal converter
• No special requirements.

Compact version
• Do not lift the device by the signal converter housing.
• Do not use lifting chains.
• To transport flange devices, use lifting straps. Wrap these around both process connections.

Figure 2-4: Transport

2.6 Pre-installation requirements


Make sure that you have all necessary tools available:
• Allen key (4 mm)
• Small screwdriver
• Wrench for cable glands
• Wrench for wall mounting bracket (remote version only)
• Torque wrench for installing flowmeter in pipeline

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 7


2 INSTALLATION OPTIFLUX 2000

2.7 General requirements


INFORMATION!
The following precautions must be taken to ensure reliable installation.
• Make sure that there is adequate space to the sides.
• Protect the signal converter from direct sunlight and install a sun shade if necessary.
• Signal converters installed in control cabinets require adequate cooling, e.g. by fan or heat
exchanger.
• Do not expose the signal converter to intense vibration. The flowmeters are tested for a
vibration level in accordance with IEC 68-2-64.

2.7.1 Vibration

Figure 2-5: Avoid vibrations

2.7.2 Magnetic field

Figure 2-6: Avoid magnetic fields

8 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 INSTALLATION 2

2.8 Installation conditions


2.8.1 Inlet and outlet
Use straight inlet and outlet pipe sections to prevent flow distortion or swirl, caused by bends
and T- sections.

Figure 2-7: Recommended inlet and outlet section


1 Refer to chapter "Bends in 2 or 3 dimensions"
2 ≥ 2 DN

2.8.2 Bends in 2 or 3 dimensions

Figure 2-8: 2 and 3 dimensional bends, in front of flowmeter


1 Bends in 2 dimensions: ≥ 5 DN; bends in 3 dimensions: ≥ 10 DN

2.8.3 T-section

Figure 2-9: Distance behind a T-section


1 ≥ 10 DN

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 9


2 INSTALLATION OPTIFLUX 2000

2.8.4 Bends

Figure 2-10: Installation in bending pipes

Figure 2-11: Installation in bending pipes

CAUTION!
Avoid draining or partial filling of the flow sensor

2.8.5 Open feed or discharge

Figure 2-12: Installation in front of an open discharge

10 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 INSTALLATION 2

2.8.6 Flange deviation


CAUTION!
Max. permissible deviation of pipe flange faces:
Lmax - Lmin ≤ 0.5 mm / 0.02"

Figure 2-13: Flange deviation


1 Lmax
2 Lmin

2.8.7 Pump

Figure 2-14: Installation behind a pump

2.8.8 Control valve

Figure 2-15: Installation in front of a control valve

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 11


2 INSTALLATION OPTIFLUX 2000

2.8.9 Air venting and vacuum forces

Figure 2-16: Air venting


1 ≥5m
2 Air ventilation point

Figure 2-17: Vacuum


1 ≥5m

12 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 INSTALLATION 2

2.8.10 Mounting position

Figure 2-18: Mounting position

2.9 Mounting
2.9.1 Torques and pressures
The maximum pressure and torques values for the flowmeter are theoretical and calculated for
optimum conditions and use with carbon steel flanges.

Figure 2-19: Tightening of bolts

Tightening of bolts
• Always tighten the bolts uniformely and in diagonally opposite sequence.
• Do not exceed the maximum torque value.
• Step 1: Apply approx. 50% of max. torque given in table.
• Step 2: Apply approx. 80% of max. torque given in table.
• Step 3: Apply 100% of max. torque given in table.

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 13


2 INSTALLATION OPTIFLUX 2000

Nominal size Pressure Bolts Max. torque [Nm] 1


DN [mm] rating
Polyolefin Polypropylene Hard rubber
25 PN 40 4 x M 12 - 22 11
32 PN 40 4 x M 16 - 37 19
40 PN 40 4 x M 16 - 43 25
50 PN 40 4 x M 16 - 55 31
65 PN 16 2 x M 16 - 51 42
65 PN 40 8 x M 16 - 38 21
80 PN 40 8 x M 16 - 47 25
100 PN 16 8 x M 16 - 39 30
125 PN 16 8 x M 16 - 53 40
150 PN 16 8 x M 20 - 68 47
200 PN 10 8 x M 20 68 - 68
200 PN 16 12 x M 20 45 - 45
250 PN 10 12 x M 20 65 - 65
250 PN 16 12 x M 24 78 - 78
300 PN 10 12 x M 20 76 - 76
300 PN 16 12 x M 24 105 - 105
350 PN 10 16 x M 20 75 - 75
400 PN 10 16 x M 24 104 - 104
450 PN 10 20 x M 24 93 - 93
500 PN 10 20 x M 24 107 - 107
600 PN 10 20 x M 27 138 - 138
700 PN 10 24 x M 27 163 - 163
800 PN 10 24 x M 30 219 - 219
900 PN 10 28 x M 30 205 - 205
1000 PN 10 28 x M 33 261 - 261
1 The specified torque values are dependent on variables (temperature, bolt material, gasket material,
lubricants, etc.) which are not within the control of the manufacturer. Therefore the values should be re-
garded as indicative only.
2 DN65 / PN16 is available with standard 8 bolt holes. On request 4 bolt holes is optional.

INFORMATION!
Other sizes / pressure ratings on request.

14 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 INSTALLATION 2

Nominal size Flange class Bolts Max. torque [lbf.ft] 1


[inch] [lb]
Polyolefin Polypropylene Hard rubber
1 150 4 x 1/2" - 6.7 3.2
1 1/2 150 4 x 1/2" - 13 9
2 150 4 x 5/8" - 24 17
3 150 4 x 5/8" - 43 29
4 150 8 x 5/8" - 34 23
6 150 8 x 3/4" - 61 38
8 150 8 x 3/4" 51 - 51
10 150 12 x 7/8" 58 - 58
12 150 12 x 7/8" 77 - 77
14 150 12 x 1" 69 - 69
16 150 16 x 1" 67 - 67
18 150 16 x 1 1/8" 105 - 105
20 150 20 x 1 1/8" 94 - 94
24 150 20 x 1 1/4" 133 - 133
28 150 28 x 1 1/4" 119 - 119
32 150 28 x 1 1/2" 191 - 191
36 150 32 x 1 1/2" 198 - 198
40 150 36 x 1 1/2" 198 - 198
1 The specified torque values are dependent on variables (temperature, bolt material, gasket material,
lubricants, etc.) which are not within the control of the manufacturer. Therefore the values should be re-
garded as indicative only.

INFORMATION!
Other sizes / pressure ratings on request.

CAUTION!
• Pressures are applicable at 20°C / 68°F.
• For higher temperatures, the pressure ratings are as per ASME B16.5.

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 15


2 INSTALLATION OPTIFLUX 2000

2.10 Temperatures
CAUTION!
Protect the device from direct sunlight.

Temperature range Process [°C] Ambient [°C] Process [°F] Ambient [°F]

min. max. min. max. min. max. min. max.


Hard rubber / Polyolefin 1
Separate flow sensor -5 80 -40 65 23 176 -40 149
Compact with IFC 300 -5 80 -40 65 23 176 -40 149
Compact with IFC 100 -5 80 -40 65 23 176 -40 149
Compact with IFC 050 -5 80 -40 65 23 176 -40 149
Polypropylene 2
Separate flow sensor -5 90 -40 65 23 194 -40 149
Compact with IFC 300 -5 90 -40 65 23 194 -40 149
Compact with IFC 100 -5 90 -40 65 23 194 -40 149
Compact with IFC 050 -5 90 -40 65 23 194 -40 149
1 Polyolefin is only available for DN200...1000
2 Polypropylene is only available for DN25...150

16 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 ELECTRICAL CONNECTIONS 3

3.1 Safety instructions


DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!

DANGER!
Observe the national regulations for electrical installations!

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.

WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.

INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.

3.2 Grounding
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel
against electric shocks.

Figure 3-1: Grounding


1 Metal pipelines, not internally coated. Grounding without grounding rings.
2 Metal pipelines with internal coating and non-conductive pipelines. Grounding with grounding rings.

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 17


3 ELECTRICAL CONNECTIONS OPTIFLUX 2000

Figure 3-2: Different types of grounding rings


1 Grounding ring number 1
2 Grounding ring number 2
3 Grounding ring number 3

Grounding ring number 1:


• 3 mm / 0.1" thick (tantalum: 0.5 mm / 0.02")

Grounding ring number 2:


• 3 mm / 0.1" thick
• Prevents damage to the flanges during transport and installation
• Especially for flow sensors with PTFE liner

Grounding ring number 3:


• 3 mm / 0.1" thick
• With cylindrical neck (length 30 mm / 1.25" for DN10...150 / 3/8...6")
• Prevents damage to the liner when abrasive liquids are used

18 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 ELECTRICAL CONNECTIONS 3

3.3 Virtual reference for IFC 300 (C, W and F version)


The virtual reference option on the IFC 300 flow converter provides complete isolation of the
measurement circuit.

Benefits of virtual reference:


• Grounding rings or grounding electrodes can be omitted.
• Safety increases by reducing the number of potential leakage points.
• The installation of the flowmeters is much easier.

Figure 3-3: Virtual reference

Minimum requirements:
• Size: ≥ DN10
• Electrical conductivity: ≥ 200 µS/cm
• Electrode cable: max. 50 m / 164 ft, type DS

3.4 Connection diagrams


INFORMATION!
For the connection diagrams please refer to the documentation of the applicable signal
converter.

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 19


4 TECHNICAL DATA OPTIFLUX 2000

4.1 Dimensions and weights

Remote version a = 88 mm / 3.5"


b = 139 mm / 5.5" 1
c = 106 mm / 4.2"
Total height = H + a

Compact version with a = 155 mm / 6,1"


IFC 300
b = 230 mm / 9.1" 1
c = 260 mm / 10.2"
Total height = H + a

Compact version with a = 82 mm / 3.2"


IFC 100 (0°)
(0 )
b = 161 mm / 6.3"
c = 257 mm / 10.1" 1
Total height = H + a

Compact version with a = 186 mm / 7.3"


IFC 100 (45°)
(45 )
b = 161 mm / 6.3"
c = 184 mm / 2.7" 1
Total height = H + a

Compact version with a = 101 mm / 3.98"


IFC 050 (10°)
(10 )
b = 157 mm / 6.18"
c = 260 mm / 10.24" 1
Total height = H + a

1 The value may vary depending on the used cable glands.

20 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 TECHNICAL DATA 4

INFORMATION!
• All data given in the following tables are based on standard versions of the flow sensor only.
• Especially for smaller nominal sizes of the flow sensor, the signal converter can be bigger
than the flow sensor.
• Note that for other pressure ratings than mentioned, the dimensions may be different.
• For full information on signal converter dimensions see relevant documentation.

EN 1092-1
Nominal Dimensions [mm] Approx.
size DN weight
[mm] Standard ISO Insertion H W [kg]
length length
25 150 200 140 115 5
32 150 200 157 140 6
40 150 200 166 150 7
50 200 200 186 165 11
65 200 200 200 185 9
80 200 200 209 200 14
100 250 250 237 220 15
125 250 250 266 250 19
150 300 300 300 285 27
200 350 350 361 340 34
250 400 450 408 395 48
300 500 500 458 445 58
350 500 550 510 505 78
400 600 600 568 565 101
450 600 - 618 615 111
500 600 - 671 670 130
600 600 - 781 780 165
700 700 - 898 895 248
800 800 - 1012 1015 331
900 900 - 1114 1115 430
1000 1000 - 1225 1230 507
1200 1200 - 1417 1405 555
1400 1400 - 1619 1630 765
1600 1600 - 1819 1830 1035
1800 1800 - 2027 2045 1470
2000 2000 - 2259 2265 1860

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 21


4 TECHNICAL DATA OPTIFLUX 2000

ASME B16.5 / 150 lb flanges


Nominal size Dimensions [inches] Approx. weight
[inches] [lb]
L H W
1" 5.91 5.39 4.25 7
1½" 5.91 6.10 5.00 11
2" 7.87 7.05 5.98 18
3" 7.87 8.03 7.50 26
4" 9.84 9.49 9.00 44
5" 9.84 10.55 10.00 49
6" 11.81 11.69 11.00 64
8" 13.78 14.25 13.50 95
10" 15.75 16.30 16.00 143
12" 19.69 18.78 19.00 207
14" 27.56 20.67 21.00 284
16" 31.50 22.95 23.50 364
18" 31.50 24.72 25.00 410
20" 31.50 26.97 27.50 492
24" 31.50 31.38 32.00 675

ASME B16.5 / 300 lb flanges


Nominal size Dimensions [inches] Approx. weight
[inches] [lb]
L H W
1" 5.91 5.71 4.87 11
1½" 7.87 6.65 6.13 13
2" 9.84 7.32 6.50 22
3" 9.84 8.43 8.25 31
4" 11.81 10.00 10.00 44
6" 12.60 12.44 12.50 73
8" 15.75 15.04 15.00 157
10" 19.69 17.05 17.50 247
12" 23.62 20.00 20.50 375
14" 27.56 21.65 23.00 474
16" 31.50 23.98 25.50 639
20" 31.50 28.46 30.50 937
24" 31.50 33.39 36.00 1345

22 www.krohne.com 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en


OPTIFLUX 2000 TECHNICAL DATA 4

4.2 Vacuum load

Diameter Vacuum load in mbar abs. at a process temperature of

[mm] 20ºC 40ºC 60ºC 80ºC


Hard rubber
DN200...300 250 250 400 400
DN350...1000 500 500 600 600
DN1200...3000 600 600 750 750
Polypropylene
DN25...150 250 250 400 400
Polyolefin
DN200...1000 0 0 0 0

Diameter Vacuum load in psia at process temperature of

[inches] 68ºF 104ºF 140ºF 176ºF


Hard rubber
8...12" 3.6 3.6 5.8 5.8
14...40" 7.3 7.3 8.7 8.7
48...120" 8.7 8.7 10.9 10.9
Polypropylene
1...6" 3.6 3.6 5.8 5.8
Polyolefin
8...40" 0 0 0 0

04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en www.krohne.com 23


KROHNE product overview
• Electromagnetic flowmeters
• Variable area flowmeters
• Ultrasonic flowmeters
• Mass flowmeters
© KROHNE 04/2014 - 4002298802 - QS OPTIFLUX 2000 R07 en - Subject to change without notice.

• Vortex flowmeters
• Flow controllers
• Level meters
• Temperature assemblies
• Pressure transmitters
• Analysis products
• Products and systems for the oil & gas industry
• Measuring systems for the marine industry

Head Office KROHNE Messtechnik GmbH


Ludwig-Krohne-Str. 5
47058 Duisburg (Germany)
Tel.:+49 203 301 0
Fax:+49 203 301 103 89
info@krohne.com
The current list of all KROHNE contacts and addresses can be found at:
www.krohne.com

You might also like