Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline by Gas Chromatography
Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline by Gas Chromatography
Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline by Gas Chromatography
Designation: D 3606 – 04
1
D 3606 – 04
TABLE 1 Instrument Parameters 5.13 Burets, automatic, with integral reservoir, 25-mL ca-
Detector thermal conductivity pacity.
Columns: two, stainless steel
Length, m (A) 0.8; (B) 4.6
Outside diameter, mm 3.2
6. Materials
Stationary phase (A) dimethylpolysiloxane, 10 mass % 6.1 Carrier Gas—Helium, 99.99 % pure. (Warning—
(B) TCEP, 20 mass %
Support (A) Chromosorb W, 60 to 80-mesh
Compressed gas under high pressure.)
(B) Chromosorb P, 80 to 100-mesh 6.2 Support—Crushed firebrick, acid-washed, 60 to 80-
Reference column Any column or restriction may be mesh and 80 to 100-mesh.
used.
Temperature:
6.3 Liquid Phases—1,2,3-Tris(2-cyanoethoxy) propane
Sample inlet system, °C 200 (TCEP) and methyl silicone.3
Detector, °C 200 6.4 Solvents:
Column, °C 145
Carrier Gas: helium
6.4.1 Methanol, reagent grade. (Warning—Flammable. Va-
Linear Gas Rate, cm/s 6 por harmful. Can be fatal or cause blindness if swallowed or
Volume flow rate, cm3/min approx 30 inhaled.)
Column head pressure, kPa (psi) approx 200 (30)
Recorder range, mV 0 to 1
6.4.2 Chloroform, reagent grade. (Warning—Can be fatal if
Chart speed, cm/min 1 swallowed. Harmful if inhaled.)
Sample size, µL 2 6.4.3 Methylene Chloride, for cleaning columns.
Total cycle time, min 8
Backflush, min approximately 0.75A
(Warning—Harmful if inhaled. High concentrations can cause
A
unconsciousness or death.)
This back flush time must be determined for each column system.
6.4.4 Acetone, for cleaning columns. (Warning—
Extremely flammable. Vapors can cause flash fires.)
5.2.1 Column A—One 0.8-m (2.5-ft) by 3.2-mm (1⁄8-in.) 6.5 Internal Standard:
outside diameter stainless steel column packed with 10 mass % 6.5.1 Methyl Ethyl Ketone (MEK), 99.5% minimum purity.
dimethylpolysiloxane (for example, OV-101) on Chromosorb (Warning—Flammable. Vapor can be harmful.)
W, 60 to 80 mesh. 6.6 Calibration Standards:
5.2.2 Column B—One 4.6-m (15-ft) by 3.2-mm outside 6.6.1 Benzene, 99+ mol %. (Warning—Poison. Carcinogen.
diameter stainless steel column packed with 20 mass % TCEP Harmful or fatal if swallowed. Extremely flammable. Vapors
on Chromosorb P, 80 to 100 mesh. can cause flash fires.)
5.3 Recorder, a strip chart recorder. An electronic integrat- 6.6.2 Isooctane (2,2,4–trimethyl pentane), 99+ mol %.
ing device or a computer capable of graphical presentation of (Warning—Extremely flammable. Harmful if inhaled.)
the chromatogram. The electronic integrating device or com- 6.6.3 Toluene, (Warning—Flammable. Vapor harmful.)
puter must be capable of measuring 0.1 volume % MEK with 6.6.4 n-Nonane, 99+ mol %. (Warning—Flammable. Vapor
satisfactory signal-to-noise. If a strip chart recorder is to be harmful.)
used, a 0 to 1–mV range recording potentiometer with a
7. Sampling
response time of 2 s or less and a maximum noise level of 60.3
% of full scale is recommended. The detector strip chart 7.1 Gasoline—(Warning—Extremely flammable. Vapors
recorder combination must produce a 4–mm deflection for a harmful if inhaled.) Samples to be analyzed by this test method
2–µL sample containing 0.1 volume % MEK when operated at shall be obtained using the procedures outlined in Practice
maximum sensitivity. D 4057.
5.4 Microsyringe, 5-µL capacity.
5.5 Volumetric Pipets, Class A, 0.5, 1, 5, 10, 15, and 20-mL 8. Preparation of Column Packings
capacities (see Specification E 694 and E 969). 8.1 Prepare two packing materials (one packing material
5.6 Measuring Pipets, 1 and 2–mL capacities calibrated in consists of 10 mass % dimethylpolysiloxane on Chromosorb
0.01 mL; 5–mL calibrated in 0.1–mL, for use in dispensing W; the other, 20 mass % TCEP on Chromosorb P) in accor-
volumes of benzene and toluene not covered by the volumetric dance with the following procedures:
pipets (see Specification E 1044 and E 1293) during prepara- 8.1.1 Dimethylpolysiloxane Packing—Weigh 45 g of the
tion of standard samples (see 11.1). Chromosorb W, 60 to 80 mesh and pour into the 500-mL flask
(5.10). Dissolve 5 g of the dimethylpolysiloxane in approxi-
NOTE 1—Other volume dispensing equipment capable of delivering the
mately 50 mL of chloroform. (Warning—Can be fatal if
specified volumes within the stated tolerance limits may be used as an
alternative to the requirements stated in 5.5 and 5.6. swallowed. Harmful if inhaled.) Pour the methyl silicone-
chloroform solution into the flask containing the Chromosorb
5.7 Flasks, volumetric, 25 and 100-mL capacity. W. Attach the flask to the evaporator (5.9), connect the
5.8 Vibrator, electric. vacuum, and start the motor. Turn on the infrared lamp and
5.9 Vacuum Source. allow the packing to mix thoroughly until dry.
5.10 Evaporator, vacuum, rotary.
5.11 Flask, boiling, round-bottom, short-neck, with 24⁄40
standard taper joint, 500-mL capacity. Suitable for use with 3
Packed column liquid phases such as OV 101 are considered to be of the
evaporator (5.10). dimethylpolysiloxane type. Other equivalent phases can also be used. Consult with
5.12 Lamp, infrared. the column manufacturer or phase supplier for information.
2
D 3606 – 04
3
D 3606 – 04
death) into the funnel and allow it to drain through the steel 4.6-m tubing (Column B) with the TCEP packing (8.1.2) using
tubing and into the drain beaker. Repeat the washing procedure the following procedure. Close one end of each tubing with a
with 50 mL of acetone. (Warning—Harmful if inhaled. High small, glass wool plug, and connect this end to a vacuum
concentrations can cause unconsciousness or death.) Remove source by means of a glass wool-packed tube. To the other end
the funnel and attach the steel tubing to an air line, using vinyl connect a small polyethylene funnel by means of a short length
tubing to make the connection. Remove all solvent from the of vinyl tubing. Start the vacuum and pour the appropriate
steel tubing by blowing filtered, oil-free air through or pulling packing into the funnel until the column is full. While filling
a vacuum. each column, vibrate the column with the electric vibrator to
9.2 Packing Columns—Preform Columns A and B sepa- settle the packing. Remove the funnel and shut off the vacuum
rately to fit the chromatograph. Pack the 0.8-m tubing (Column source. Remove the top 6 mm (1⁄4-in.) of packing and insert a
A) with the dimethylpolysiloxane packing (8.1.1) and the glass wool plug in this end of the column.
4
D 3606 – 04
10. Configuration of Apparatus and Establishment of 10.4 Determine Time to Backflush—The time to backflush
Conditions will vary for each column system and must be determined
10.1 Conditioning Column—Install Columns A and B as experimentally as follows. Prepare a mixture of 5 volume %
shown in Fig. 1 or Fig. 2 in accordance with the system isooctane in n-nonane. Using the injection technique described
preferred (5.1). Do not connect the exit end of Column B to the in 11.4 and with the preferred system (10.3) in the forward flow
detector until the columns have been conditioned. Pass helium mode, inject 1 µL of the isooctane – n-nonane mixture. Allow
gas through the column at approximately 40 cm3/min. Condi- the chromatogram to run until the n-nonane has eluted and the
tion the column at the listed temperatures for the specified time detector signal has returned to baseline. Measure the time in
periods. seconds, from the injection until the detector signal returns to
baseline between the isooctane and n-nonane peaks. At this
Temperature, °C Hours at Temperature
50 1⁄2 point all of the isooctane, but essentially none of the n-nonane,
100 1⁄2 should have eluted. One half of the time determined should
150 1 approximate the “time to backflush” and should be from 30 to
170 3
60 s. Repeat the run, including the injection, but switching the
10.2 Assembly—Connect the outlet of Column B to the system to the backflush mode at the predetermined “time to
detector port. Adjust the operating conditions to those listed in backflush.” This should result in a chromatogram of isooctane
Table 1, but do not turn on the detector circuits. Check the with little or no n-nonane visible. If necessary, make additional
systems for leaks. runs, adjusting the “time to backflush” until this condition of
10.3 Flow Rate Adjustment: all the isooctane and little or no n-nonane is attained. The “time
10.3.1 Column System Setup for Pressure Backflushing to backflush” so established, including the actual valve opera-
(Fig. 1): tions, must be used in all subsequent calibrations and analyses.
10.3.1.1 Open Tap A and B and close C; set the primary 11. Calibration and Standardization
pressure regulator to give the desired flow (Table 1) through 11.1 Standard Samples—Prepare seven standard samples
the column system (at an approximate gage pressure of 205 covering the range 0 to 5 volume % benzene and 0 to 20
kPa (30 psi)). Measure the flow rate at the detector vent, volume % toluene as follows: For each standard, measure the
sample side. Observe the pressure on gage GC. volume of benzene and of toluene listed below into a 100-mL
10.3.1.2 Close Tap A and open B and C. The pressure volumetric flask. Dilute to volume with isooctane (2,2,4–tri-
reading on gage GA should fall to zero immediately. If not, methyl pentane), with all components and glassware at ambient
open the needle valve until the pressure falls to zero. temperature.
10.3.1.3 Close Tap B. Adjust the secondary pressure regu- Benzene Toluene
lator until the reading of gage GC is 3.5 to 7 kPa (0.5 to 1 psi) Volume % mL Volume % mL
5 5.0 20 20.0
higher than observed in 10.3.1.1. 2.5 2.5 15 15.0
10.3.1.4 Open Tap B and adjust the backflush vent control 1.25 1.25 10 10.0
needle valve until the pressure recorded on GA approximates a 0.67 0.67 5 5.0
0.33 0.33 2.5 2.50
gage pressure of 14 to 28 kPa (2 to 4 psi). 0.12 0.12 1 1.0
10.3.1.5 Forward Flow—Open Taps A and C and close Tap 0.06 0.06 0.5 0.50
B (Fig. 1 B1). 11.2 Calibration Blends—Accurately measure 1.0 mL of
10.3.1.6 Backflush—Close Tap A and open Tap B. (There MEK into a 25-mL volumetric flask, and fill to the mark with
should be no baseline shift on switching from forward flow to the first standard sample (11.1). Continue doing this until all
backflush. If there is a baseline shift increase the secondary blends have been prepared.
pressure slightly.) (Fig. 1)
NOTE 2—Commercially prepared calibration standards may be used,
10.3.2 Column System Setup for Valve Backflushing (Fig. including those that are pre-mixed with the MEK internal standard.
2):
11.3 Chromatographic Analysis—Chromatograph each of
10.3.2.1 Set the valve in the forward flow mode (Fig. 2 B1), the calibration blends using the conditions established in 10.4
and adjust flow control A to give the desired flow (Table 1). using the following injection technique:
Measure the flow rate at the detector vent, sample side. 11.4 Injection of Sample:
10.3.2.2 Set the valve in the backflush position (Fig. 2 B2), 11.4.1 Use of an automatic liquid sample injection system is
measure the flow rate at the detector vent, sample side. If the highly recommended. If manual injections are to be made, the
flow has changed, adjust flow control B to obtain the correct injection technique in 11.4.2 is necessary so that sharp sym-
flow. (Flows should match to within 61 cm3/min). metrical peaks will be obtained.
10.3.2.3 Change the valve from forward flow to the back- 11.4.2 Flush the 5-µL microsyringe at least three times with
flush position several times and observe the baseline. There the sample mixture and then fill with about 3µ L of the sample.
should be no baseline shift or drift after the initial valve kick (Avoid including any air bubbles in the syringe.) Slowly eject
that results from the pressure surge. If there is a baseline shift, the sample until 2.0 µL remains in the syringe; wipe the needle
increase or decrease flow control B slightly to balance the with tissue and draw back the plunger to admit 1 to 2 µL of air
baseline. (A persistent drift could indicate leaks somewhere in into the syringe. Insert the needle of the syringe through the
the system.) septum cap of the chromatograph and push until the barrel of
5
D 3606 – 04
the syringe is resting against the septum cap; then push the NOTE 5—The order of elution will be nonaromatic hydrocarbons,
plunger to the hilt and remove the syringe immediately from benzene, MEK, and toluene using the prescribed dimethylpolysiloxane
the chromatograph. and TCEP. Fig. 4 is an example of a typical chromatogram.
11.5 Calibration—Measure the area of both aromatic peaks 12.4 Measurement of Area—Measure the areas under the
and of the internal standard peak as directed in 12.4. Calculate aromatic peaks and under the MEK peak by conventional
the ratio of the benzene peak area to the MEK peak area. Plot methods.
the concentration of benzene versus the ratio. Make the same NOTE 6—The precision statement in Section 15 was developed from
calculation and plot for toluene. See Fig. 3 for an example. This data obtained using electronic integrators or on-line computers. The
must be done to ensure that the entire chromatographic system precision statement may not apply if other methods of integration or peak
is operating properly and that the concentration of any one area measurement are used.
component has not exceeded the linear response range of any
part of the system: column, detector, integrator, and other 13. Calculation
components. The calibration should be linear. 13.1 Calculate the ratios of the peak areas of benzene and
toluene to the peak area of MEK. Determine from the appro-
NOTE 3—Calibrations using computer-based chromatography systems
are an acceptable alternative to the calibration procedure specified in 11.5.
priate calibration curve the liquid volume percent of benzene
NOTE 4—If the calibration has been shown to be linear, a least squares and toluene corresponding to the calculated peak ratios.
calculation may be performed to calculate a calibration factor. The 13.2 If the results are desired on a mass basis, convert to
precision statement in Section 15 was developed from data obtained from mass percent as follows:
calibration plots and may not apply if calibration factors are used.
Benzene, mass % 5 ~VB/D! 3 0.8844 (1)
12. Procedure where:
12.1 Preparation of Sample—Accurately measure 1.0 mL VB = volume percent benzene, and
of MEK into a 25-mL volumetric flask. Fill to the mark with D = relative density of sample at 15.6/15.6°C (60/60°F).
the sample to be tested and mix well.
12.2 Chromatographic Analysis—Chromatograph the
sample, using the conditions established in 10.4 “time to Toluene, mass % 5 ~V T/D! 3 0.8719 (2)
backflush” and the injection technique described in 11.4. The
valves must be turned to backflush mode at the time determined where:
in 10.4 so that undesirable components do not enter Column B. VT = volume percent toluene, and
12.3 Interpretation of Chromatogram— Identify on the D = relative density of sample at 15.6/15.6°C (60/60°F).
chromatogram the benzene, toluene, and the internal standard
MEK peaks from the retention times of the standards. 14. Report
14.1 Report the benzene and toluene contents in liquid
volume percent to the nearest 0.1 %.
6
D 3606 – 04
15.1.1 Repeatability—The difference between successive TABLE 3 Reproducibility
test results, obtained by the same operator with the same NOTE—X = the mean volume % of the component.
apparatus under constant operating conditions on identical test Component Range, volume % Reproducibility See Note
material would, in the long run, in the normal and correct
Benzene 0.1–1.5 0.13(X) + 0.05 15
operation of the test method, exceed the values in Table 2 only Benzene >1.5 0.28(X) 16
in 1 case in 20: Toluene 1.7–9 0.12(X) + 0.07 15
15.1.2 Reproducibility—The difference between two, single Toluene >9 1.15 16
SUMMARY OF CHANGES
Subcommittee D02.04 has identified the location of selected changes to this standard since the last issue
(D 3606–99) that may impact the use of this standard.
(1) Added additional Referenced Documents. (3) Added new Notes 1-3.
(2) Updated 5.5 and added new 5.6. (4) Updated 6.5.1, 6.6.2, and 11.1.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).