HIV Ag Ab Card 4th Generation
HIV Ag Ab Card 4th Generation
HIV Ag Ab Card 4th Generation
Rapid visual test for the qualitative detection of HIV-1 p24 Antigen and Antibodies (IgM, IgG & IgA)
to HIV-1& HIV-2 in Human Serum/ Plasma/ Whole Blood
5) Add 1 drop (20 μl) of the sample (Serum / Plasma / Whole Blood) IBO
DY
Expiry Date Catalogue Number HIV
ANT
using sample dropper / micropipette into the sample wells of both C
Ab
Do not use if package Contains biological Material the HIV Antigen and HIV Antibody device respectively. (fig. b)
is damaged of Animal Origin
negative test result does not eliminate the possibility of HIV infection, and should be confirmed by
ELISA and/or RT-PCR.
OR
5. Positive test results do not rule out co-infections with other pathogens.
Whole blood sample collection from finger prick
PERFORMANCE CHARACTERISTICS OF HIV Ag & Ab CARD (4th GENERATION)
5a) Wipe the complete fingertip with the alcohol swab. Wait (fig. 5a) Analytical Sensitivity: The sensitivity of the HIV Ag & Ab CARD (4th Generation) has been determined for
until the finger has completely dried (minimum 30
HIV-1 p24 Antigen using WHO international standard: HIV-1 p24 antigen NIBSC Code No. 90/636 and it is
seconds).
equal to 100 IU/ml.
5b) Take the lancet and prick the side of the pulp (ball of the (fig. 5b) Sensitivity and Specificity studies were carried out on fresh as well as frozen samples from low risk as well as
finger) with the lancet, perpendicular to the lines of the fingerprint. high risk groups. Performance of the test with reference to sensitivity and specificity has been determined by
5c) Make sure a well formed drop of blood is present on the tip of the using clinical known negative and positive antibodies serum/plasma and whole blood samples and HIV p24
finger. Antigen control.
5d) Take the sample dropper/micropipette and collect 20 μl blood by The performance of the test was evaluated and compared with a licensed commercially available Elisa test.
dipping the tip of the sample dropper/ micropipette into the blood The results obtained are as follows:
drop. Add 1 drop of blood immediately in the sample well of the (fig. 5c)
HIV Antigen device. Sample Type No. of Samples Licensed Test HIV Ag & Ab Card
Tested (4th Generation)
5e) Repeat the above step (5d) for HIV Antibody device.
Caution: Care should be taken that the blood sample does not clot & the transfer to Negative 1200 1200 1199
the sample pad is immediate. Sample dropper should be discarded
immediately considering it to be biohazardous. HIV Positive 125 125 125
N
IGE
ANT
6) In case of blood samples, wait for 30 seconds (till the blood is HIV
Ag
Sensitivity: 100% Specificity: 99.91%
C
completely absorbed). IBO
DY
ANT
HIV Cross Reactivity: The cross reactivity of the kit has been tested using 60 samples from other diseases and
7) Add 1 drop Assay Buffer into the sample well of both device. C
Ab
control region “C” of the HIV Antibody device indicates that the sample is C Ab 16 ASO 3 NEGATIVE NEGATIVE
“REACTIVE” for both HIV-1 p24 Antigen and HIV-1 and/or HIV-2 Antibodies. (fig. f) External Evaluation: The performance of HIV Ag & Ab Card has been evaluated by National Institute of
NON-REACTIVE HIV ANTIGEN Biologicals, India. The result obtained of 3 different lots are as follows:
As shown in Fig. (g), appearance of only one distinct red line in the control C Ag
Sensitivity: 100% Specificity: 100%
region “C” of both the HIV Antigen and HIV Antibody device indicates that the HIV ANTIBODY
sample is “NON REACTIVE” for both HIV-1 p24 Antigen and HIV-1 and/or C Ab
Precision: Within run (Intra assay) & between run (Interassay) precision have been determined by testing 10
HIV-2 Antibodies. replicates of ten samples - four negative, one HIV antigen (p24) positive, one HIV-2 positive and four HIV-1
(fig. g)
Positive. The C.V. (%) of all the ten samples were within 10%.
INVALID
The test is invalid, if no control line appears HIV ANTIGEN HIV ANTIGEN LIMITED EXPRESSED WARRANTY DISCLAIMER
(with or without the appearance of test lines) C Ag C Ag
The manufacturer limits the warranty to the test kit, as much as that the test kit will function as an in vitro
diagnostic assay within the limitations and specifications as described in the product instruction-manual,
after the completion of test, either with clear HIV ANTIBODY HIV ANTIBODY
when used strictly in accordance with the instructions contained therein. The manufacturer disclaims any
background or with complete pinkish/ C Ab C Ab
warranty expressed or implied including such expressed or implied warranty with respect to merchantability,
purplish background (fig. h to k). (fig. h) (fig. i) fitness for use or implied utility for any purpose. The manufacturer’s liability is limited to either replacement of
Repeat the test with a new card. Invalid test HIV ANTIGEN HIV ANTIGEN
the product or refund of the purchase price of the product and in no case liable to claim of any kind for an
may be because of following reasons: C Ag C Ag amount greater than the purchase price of the goods in respect of which damages are likely to be claimed. The
(a) Improper storage at temperature other HIV ANTIBODY HIV ANTIBODY manufacturer shall not be liable to the purchaser or third parties for any injury, damage or economic loss,
than the recommended temperature. C Ab C Ab
howsoever caused by the product in the use or in the application there of.
(b) Wrong Procedure (fig. j) (fig. k) BIBLIOGRAPHY OF SUGGESTED READING
(c) Long atmospheric exposure of the test 1. Ajaka Z.L. et.al. (1994). HIV-1/HIV-2 seronegativity in HIV-1 subtype O infected patients. The Lancet,
device after opening the pouch. 343, 1393-1394.
(d) Use of turbid/ lipemic/ haemolyzed sample. 2. Granade, T. C., Nguyen, S., Kuehl, D. S., & Parekh, B. S. (2013). Development of a novel rapid HIV test
In case of invalid result, serum/ plasma test samples should be centrifuged at 10,000 rpm for for simultaneous detection of recent or long-term HIV type 1 infection using a single testing device.
15 minutes and retest using new Card. AIDS research and human retroviruses, 29(1), 61-67.
3. Nakatsuma, A., Kaneda, M., Kodama, H., Morikawa, M., Watabe, S., Nakaishi, K., ... & Ito, E. (2015).
LIMITATIONS AND INTERFERENCES
Detection of HIV-1 p24 at attomole level by ultrasensitive ELISA with thio-NAD cycling. PLoS One,
1. Any deviation from test procedure may lead to invalid/ erratic results.
10(6), e0131319.
2. Some samples show cross reactivity for HIV antibodies and/or p24 antigen. Following factors are 4. Parpia, Z. A., Elghanian, R., Nabatiyan, A., Hardie, D. R., & Kelso, D. M. (2010). p24 antigen rapid test
found to cause false positive HIV antigen and/or antibody test results: Naturally occurring antibodies, for diagnosis of acute pediatric HIV infection. JAIDS Journal of Acquired Immune Deficiency
Passive immunization, Leprosy, Renal Disorders, Mycobacterium avium, Herpes simplex, Syndromes, 55(4), 413-419.
Hypergammaglobulinemia, Malignant neoplasms, Tetanus vaccination, Autoimmune diseases,
VER-01 R-00
Blood Transfusion, Multiple myeloma, Haemophelia, Heat treated specimens, Lipemic serum, Anti-
nuclear antibodies, T-cell leukocyte antigen antibodies, Epstein Barr virus, HLA antibodies and other
retroviruses. in vitro diagnostic reagent, not for medicinal use
3. This is only a screening test. All samples detected reactive must be confirmed by using HIV Western
J. MITRA & CO. PVT. LTD.
Rev. Date: Jan.-24
Blot and/or PCR. Therefore for a definitive diagnosis, the patient's clinical history, symptomatology as
well as serological data, should be considered. The results should be reported only after complying A 180-181, Okhla Indl. Area, Phase-1, New Delhi-110 020, INDIA
MN/AAI/117