JUMONG General Installation Manual
JUMONG General Installation Manual
JUMONG General Installation Manual
Installation Manual
Table of Contents
Chapter i Safety Information
Introduction
Manual Conventions
General Safety
Procedural Safety
Manufacturer’s Responsibility
Revision History
Introduction
Description
Dimension
Specification
Abbreviation
Chapter iv Installation
Introduction
Unpacking
Overview
Shipping Components
label
Wiring cable
Chapter v Generator
Interfacing
Tube calibration
Safety Information
i
Introduction
We, SG healthcare CO, LTD, have the performance and reliability of our equipment as main
objectives. And we strictly follow the quality control system in order to find and prevent any
possible defects.
This Equipment is built to provide medical diagnostic images. X-ray images could be taken by
altering the angle and distance to meet the required uses. Using this equipment without
acknowledging the purpose, could harm users and patient.
Following contents must be studied by the user for the safety in use and maintenance of the
equipment.
Manual Conventions
In this manual, we used the following three icons to reduce the chance of possible accidents.
Please for the safety, study through this manual about every safety warning.
l WARNING
l CAUTION
l NOTE
Warning
This Icon is used to warn about the possible accidents that could lead to severe
injuries and death
Caution
This Icon is used to give cautions in use and maintenance that could damage the
machine and harm the user.
Note
This Icon is to provide additional information.
General Safety
This equipment contains number of mechanical and electrical safety features that can
prevents injuries of users and patients.
Note
For the operation and maintenance of the machine, the user must be fully
educated about this manual. For the safety, please read every Warning, Caution
and Note in this manual before using the machine.
Note
If there is a need of repair in either mechanical or electrical problem,
Service must be provided by the qualified engineer.
Note
Broken or Damaged materials/ equipment must be disposed in accordance of
local regulations.
Electrical and Mechanical Safety
Only qualified service engineer must repair mechanical and electrical problems.
A. Electrical safety
For the safety of users and patients, all the cable must be insulated.
All the covers must be in place for the mechanical safety.
B. Mechanical safety
l When the machine is in the movement, make sure your hand and fingers are not on
the way.
l When the machine is in the movement, keep observe patient position.
l When the machine is in the movement, keep the surrounding safe and always
keep the machine area clean.
l Please do not operate the machine uncovered, unless instructed by a qualified
engineer.
Procedural Safety
This manual is designed to provide procedures that can ensure safety of users and quality
of product. Disregarding the procedure, could lead to serious defect in equipment and
harm users and patients’ safety.
Using wrong chemicals for cleaning and sterilization could damage the machine and
cause fire. We, manufacturer, will not provide warranty for the damage caused by
chemicals.
Warning
Do not turn on the machine while cleaning
Please switch off the machine before cleaning.
Caution
Do not spray cleaning liquid directly on the machine. Use soft fabric or cloth
to clean. And Do not spray cleaning liquid on other person.
Caution
Do not soak any part of equipment in cleaning liquid.
Do not use high pressure sterilizer.
Hardware/Software Modifications
Software is fully installed in the machine. The followings are the problems that you might
encounter when you add or partially remove the software without permission.
Warning
Do not remove the installed safety features.
Warning
Do not relocate the equipment without permission.
Only qualified engineer can remove the communication cable.
Warning
Follow the instructions regarding safety which is written on the label.
Environmental Safety and Operating Environment
The equipment must be place in the area where machine can be well ventilated.
Manufacturer will not be responsible any damage caused by chemical reaction, gas,
fire in any circumstances.
The following contents are about the conditions where the machine can perform well. If the
operating environment is out of the condition provided, it can damage the machine.
Warning
Operating the machine that are out of the specified condition range, could cause
malfunction of the machine.
41°F to 86°F
Tolerance Humidity Range 10% to 75% (non-condensing)
Radiation and X-Ray General Safety Information
Operating X-Ray machine and users need to meet following requirements:
The user who operate or deal with the machine, must follow and beware of the laws and
regulations.
If the users are educated and aware of the danger and safe use of the X-ray radiation, it
could be used safety without harming. And users need to be acknowledgeable of the safety
requirements and warning about X-ray.
Manufacturer’s Responsibility
The machine must be operated by the user who are educated adequately.
The manufacturer, his representatives and his agents will not be responsible for any damage
caused by over-exposure of X-ray radiation. We will not be responsible for any damage
caused by repair and installation provided by an unauthorized service engineer.
Record of Revisions
ii
Revision History
th
1.0 19 of Aug 2016 Preparing the manual
Introduction
iii
3.1 Introduction
This manual contains the installation, usual manual and spare part.
3.2 Description
This equipment uses X-ray tube, collimator to provide X-ray medical images. This
equipment is design to maximize space utility.
3.2.1 JUMONG General Components
Following picture displays main components of the equipment.
Specifications
Compatibility The diagnostic equipment can provide diverse images in limited
spaces.
Classification Class 1
Weight 570 kg
This equipment is not suitable in condition where the room filled with combinations of
combustible goods, air, oxygen and nitrogen.
3.5 Abbreviations
% Percent
AWG American Wire Gauge
Btu British Thermal Unit
°C Degree Celsius
CE Communautés Européenes
cm Centimeter
C.R.S Cold Rolled Steel
°F Degree Fahrenheit
ga Gauge
hPa Hecto Pascal
inHg Inches Mercury
Kg Kilogram
kJ Kilojoule
Lb Pound
M Meter
max. Maximum
min. Minimum
mm Millimeter
Sq/Ft Square Foot
Sq/M Square Meter
UL Underwriters Laboratories
Installation
iv
4.1 Introduction
This chapter describes basic installation of the machine.
4.2 Unpacking
When you receive the product, you will find the product in the wooden packing. To unpack
it, you need to use required tools. Before unpacking, please move the products in the open,
wide and safe area to unpack.
To unpack the wooden packing, you will need the following tools:
1. If you unpack the wooden packing; it will look like the following picture in a
vacuumed vinyl bag.
2. If you remove a vacuumed vinyl pack, it will look like the following:
4.3 Input Power & Grounding
JUMONG
General
JUMONG
General
l Safety Equipment
l Personnel Required
l Required good for installation
l Preparation of the installation site
l Safety helmet
l Safety glasses
l Lumbar pad
l Safety footwear
4 2
Tube stand base cover
ea
5 OP cover 1
ea
6 Tube Support 1
Set
7 Cable Hose 2
ea
Caution
Without removing weight locking bolt could cause serious trouble,
It is in tube stand and wall stand.
4.7 Installation of tube stand
4.7.1 Preparation for tube stand
• Remove the rail cover.
• Remove stopper
•
• Use the anchor bolts to fix the rail to the ground.
4.7.2 Tube stand assembly
Connectthe cable andhigh voltage cable from table and wallstand to tube,OP,generator.
1. From Table
Number Name Pin Color To
Cable 1 Emergency SW Pin 1 Brown OP Emergency
SW
Pin 2 Red OP Emergency
SW
2. From Stand
l Before the connection, put high voltage cables and ROTOR cables on the external hose
4.9.2 Installation of Grid
Remove the wall stand and table’s bucky cover, then assemble the Grid.
Following pictures shows assembled and rightly positioned covers for the Jumong general.
1. Verify that the HT cable terminations are clean, in good condition, and coated with
vapor proof compound.
2. Remove the dust caps that cover the high voltage terminals on the HT tank.
3. Connect the anode and cathode cables to the HT tank. Ensure that the cables are
plugged into the proper connectors on the HT tank. Refer to figure 1-3.
4. Be sure that the HT cable connectors are tight and there is no play between the
connector insulator and the screw-down ring.
Figure 1-4: Tube / thermal switch connections on H.V. auxiliary board (low speed starter)
5.1.4 Tube Rota Cable Connections (Cont)
E7252X
Stator : XS-AL Type (Rotor boost voltage : 120VAC)
Please check stator in your Tube label, and change the jumper cable.
5.1.6 Control Console Connections
Console
J21: RS232 to Pc
J22: RS232 to Pc (Service
Port)
Mini Console
PC RS232 Port
Membrane Console
Use these step to access the GENERATOR SETUP menu (membrane console)
Step Action(membrane console)
E11
E8
E7
E5
Step 4 : Calibration is complete you see the message, Push the hand switch.
Step 5 : Calibrated the same way is a large focus.
SG Healthcare Co., Ltd.
TEL : +82-02-449-6908
FAX : +82-02-449-6909