Qute Service Manual - Ver 1
Qute Service Manual - Ver 1
Qute Service Manual - Ver 1
Electrolyte Analyzer
Service Manual
Introduction
Section 1. Introduction
Section 3. Installation
Section 8. Troubleshooting
Introduction
Note: This manual is designed to support the product. Regional service centres are
the first level of service, where faulty assemblies can be most effectively repaired at a
central repair facility. The second level of service is performed at our factory
instrument service facility where additional service procedures requiring resources
not available at regional service centres are performed.
1.2 Safety
The following safety instructions must be observed at all times during the
operation of the Quantilyte System. The operating manual must be kept near the
instrument and must be accessible to the user at all times.
Follow all warnings and instructions marked on the instrument and in this
manual. The instrument must only be operated by personnel who have been trained
to use the system. It is strongly recommended that all first time personnel read this
manual prior to working with the instrument.
The instrument must only be used in accordance with its intended use. Use
only the consumables and accessories described herein (e.g. tubes). The
manufacturer assumes no liability for any damages, including those to third parties,
caused by improper use or handling of the system.
The operator may only perform the maintenance work described in this
manual. Use only the parts recommended in this manual for servicing. The tests and
maintenance work defined by the manufacturer should be performed to make sure
that the operator remains safe and that the instrument continues to function
correctly.
For safe electrical operation of the installed instrument the relevant regulatory
provisions have to be observed.
Check that the operating voltage is set correctly on all instrument components
before the system is connected to the main power supply. This product must be
operated from the type of power source indicated on the type label. If you are not
sure of the type of power available, consult an authorized sales person or your local
electric power company.
Introduction
Use a 3-wire grounding type plug to connect the instrument and all
peripherals to the main supply. Use only extension cables with a protective conductor
and grounded contact. Grounding of the instrument and its peripherals devices to
the same protective earth potential must be ensured. Never interrupt the grounding
contacts. There is a potential risk of electric shock if the protective conductor is
interrupted within or outside the device, or if it has been disconnected.
If you can see that the instrument has become unsafe to use, switch it off and
disconnect it from the main power supply. If liquid comes in contact with the
electronics of the instrument, switch it off and disconnect it from the main power
supply. Clean and dry the affected areas of the machine.
All system surfaces must be dry while operating the Quantilyte. Only bottles,
tubing, and components supplied and recommended for containing liquids on the
instrument are to be utilized in conjunction with this machine.
Spare fuses must match the values (nominal voltage, nominal current, and
type) specified by the manufacturer. Fuses that are non-functioning must be
replaced. Repair of fuses should not be attempted. Additionally, the fuse holder
should never be short circuited.
Never open screw-attached housing parts while the instrument is on. There is
a risk of injury due to moving parts, such as fans and motor drives.
Introduction
The protective cover should not be removed during a run. Additionally, the
user should not reach into the working area as serious injury could result. Improper
handling may cause serious damage to the instrument or result in injury to the user.
Avoid touching the moving parts while the system is in operation.
Slots and openings are provided for ventilation and are not meant as access
points into the system. To ensure reliable operation of the instrument and to protect
from overheating, these openings must not be blocked or covered.
The instrument may have sharp edges. Sheet metal parts and circuit boards
located behind protective covers could have sharp edges and contact may lead to
minor injuries. When handling, cut resistant gloves should be worn and caution
should be exercised to avoid contact with corners and edges of the machine.
There is a risk of infection when handling samples and test reagents. Avoid
contact between skin and mucous membrane and samples/test reagents or any parts
of the instrument which comes into contact with samples/test reagents.
Any parts of the Quantilyte that have come into contact with samples/test
reagents are to be treated as being potentially infectious. Reagents can lead to
irritation of the skin and mucous membranes. Use personal protective equipment
(PPE, such as appropriate gloves, a lab coat, and eye protection). Observe the
instructions in the package inserts for proper use of all reagents.
1.2.e Warning
Only professional doctors and nurses can use this analyzer for clinical under
certain conditions. Before using this analyzer, users must check the device and its
accessories to ensure its proper function and safety.
Introduction
Do not open the analyzer case. Otherwise it may cause electric shock.
Any services, upgrades should be conducted by authorized personnel.
Voltage : 220 V + 22 V
Amperage : TBD
Frequency : 50 Hz + 1 Hz
Power Consumption : TBD
Width : 27 cm
Depth : 38 cm
Height : 40 cm
Weight : 10 kg
Range
Resolution Precision
Item (mmol/L
(mmol/L) (%CV)
)
0.5 –
K+ 0.01 < 1.5%
20.0
Na+ 15 – 200 0.1 < 1.5%
Cl- 15 – 200 0.1 < 1.5%
Ca2+ 0.1 – 6.0 0.01 < 2.5%
Li+ 0.2 – 5.0 0.01 < 2.0%
pH 4 – 9 pH 0.01 pH < 3.0%
2.0 –
TCO2 0.1 < 4.0%
70.0
Serum Urine
Item
(mmol/L) (mmol/L)
K+ 3.5 – 5.5 50 – 100
Na+ 135 – 145 130 – 217
Cl- 98 – 106 170 – 250
Ca2+ 1.09 – 1.35 -
TCO2 20 – 33 (Adult) -
Introduction
Avoid spilling any liquids onto the apparatus. Use only a paper towel or a
cloth wet with either water or alcohol to clean the outside surfaces of the Quantilyte.
The symbols described here are used in the current manual, on the
instrument, and on its packaging. Warning messages are displayed using a
safety symbol.
Introduction
System Overview
The key component of electrode is the ion-selective membrane. The two sides
of membrane get contact with sample and internal Electrode Internal Solution
respectively. The side touching with sample is responding to the ion concentrations
change. The other side touch with internal Electrode Internal Solution, the conversion
from ion conduction to electron conduction is carried out by Ag/AgCl inner
electrode.
The instrument applies ISE to measure the ion concentrations in the sample.
1. Measured solution
2. Conjunctive point
3. Reference electrode
4. Ion selective electrode
5. Internal Electrode Internal Solution
6. Ion selective membrane
ISE is sensitive to the ion activity in the sample. When the ion concentration in
the sample is below 10-4M and activity coefficient is close to 1, the difference
System Overview
between ion activity and concentration can be ignored (when concentration goes
over 10-4M, the activity coefficient decreases, and the difference increases.)
When ISE gets contact with the measured solution, the measured ion in the
sample goes to ISE membrane due to the diffusion effects of the concentration
difference, which creates a potential between measure electrode and reference
electrode. The potential than an ideal ISE creates for the
These two types of liquid are used for calibration of the instrument. In
addition, there are several reagents used for washing and maintaining electrodes,
including:
Inside Liquid for Reference Electrode,
Inside Liquid for Test Electrode,
Wash Solution (C),
Activated Solution (D),
Electrode Wash Solution,
Excursion Calibration for Analyzer (A).
System Overview
Powering On
Before powering on the equipment, please make sure whether all the
reagents for calibrating are used out, whether the bottle of waste liquid is full
and whether all pipes are connected correctly. Upon powering on, the system
will conduct internal self-detection, diagnosing power supply, memory and
printer. Meanwhile, it will let the internal circuit enter heat-stable status. If the
result of self-detection is normal, the equipment will start washing the system.
System Washing
Activation
System Overview
System Calibration
“CO2 calibration Program” to calibrate the CO2 sensor. For normal operation,
it only requires one time of CO2 calibration in 2-3 days.
SampleMeasurement
centrifugal separated, some serum may still contain certain fibrinogen. It may
lead to obstruction when such serum is drawn to sample needle. Thus, user
should take care of it.
After testing all specimens, please use the Wash Solution (C)
accompanied to clean several times by selecting the cleaning procedure in the
menu. It shall thoroughly clean protein reside in pipes and on electrodes.
Cleaning is good for next measurement. For hospitals of little specimens, it is
necessary to thoroughly clean the system after testing a batch of specimens;
otherwise the residue protein shall get solid in pipes and on electrodes in the
following few days that may lead to obstruction. Power off the equipment
after the cleaning is done. If the equipment will not be used for a long time, it
is best to dismount pumping pipes to prevent aging and adhesion.
5. The bottles of all reagents should be fastened after using; avoid open
for long time.
During Maintaining
1. If the pipe and electrode is obstructed, please take off pumping pipes
and use the injector (put off need and put on soft pipe) to draw in
distilled water for reverse cleaning. Let water get out from sample
needle. It is prohibited to use needle to punch through the membrane
pipe in order to prevent from damaging electrode.
2. If the electrode is not going to be used for a long time, please use
distilled water to wash away the residue liquid inside the membrane
pipe of electrode, then seal and store it.
3 Whole blood samples should be analyzed within one hour since the
collection so as to avoid errors caused by sample transformation. Pipes should
be cleaned after the measurement.
1 Serum and plasma samples stored in refrigerator could be used for analysis.
However, before analyzing, their temperature should be returned to normal.
2 When preparing serum samples, do not add wrong substances (e.g. surface
active agent, anticoagulant), which may affect the measurement results or
damage sensors.
Urine Sample:
Before testing, use urine dilution to dilute the sample to 10 times (1:9). It is
recommended to use boric acid as the anticorrosive of urine sample in order
to prevent disturbances caused by other anticorrosive.
1. The calibration value of CO2 and the calibrate factors of K, Na, Cl, Ca will
still be stored in the equipment after powering off. But if the batteries
are exhausted, the data will be lost. Thus, when batteries are exhausted
or the CO2 test is abnormal, it needs to conduct CO2 calibration.
System Overview
a. Conduct CO2 calibration, if the testing results are still not right after
calibration, please check as follows;
e If the result still vary from standard CO2 quality control value, please
input calibrate factor in the CO2 calibrate program. For example, if input 1.2,
then the test result will be multiplied by 1.2. The factor will be automatically
stored in the equipment, no need to input every time.
3. EquipmentMaintenance
Daily maintenance
System Overview
Weekly maintenance
1. Check the electrode measuring pool every week; remove residual
solution and sample; keep the electrode measuring space clean.
4. Troubleshooting
Valve fault
1. Check if the fastening screws on the valve core as well as the motor
rotation shaft are fastened tightly and in right place.
2. Check if the valve core is too tight and cannot rotate due to being
unused for a long time
Plate fault
1. If the fastening screw in the middle of plate is loose, just fasten it.
3. If any pipe or electrode is obstructed, please take off pumping pipe and
use injector (take off needle and put on soft pipe) to draw in distilled
water for reverse cleaning. Let water out from sample drawing needle.
Solution:Take off electrodes and check one by one. After finding out the
obstructed
electrode, use an injector (with pipe, no needle) to draw in “protein-
removing liquid” and inject into obstructed electrode. After immersing for 3
minutes, use distilled water to clean it.
3)There is not enough inside liquid for electrode, the liquid level is below
Ag-rod and can not be sensed – add in inside liquid for electrode.
4)There is too much inside liquid for electrode, the liquid not be more than
2/3—3/4 of the chamber volume.
5)The coating on the Ag-rod of reference electrode falls off or gets white –
replace the Ag-rod or reference electrode.
11)If all calibration values are 240, then it is generally caused by bubbles in
the reference electrodes or testing liquid (sometimes bubbles are tiny and
hard to be identified). In this case, please take off the reference electrode and
flick the bottom of electrode (below the Ag-rod) with fingers repeatedly
.
12)After the electrode’s protein-removing process or when the equipment
has not been calibrated for a long time, the first self-detection may out of
control. Just activate it for 1-2 hours and then conduct calibration.
Low electrode slope may cause poor linearity, and sometime may affect the
repeatability of electrode. The major causes are:
3)The service life is expired. In the third case, user may need to replace
electrode; in the first case, user may treat it use protein-removing liquid.In the
second case, if the humidity is too large, use moisture-removing machine to
remove
moisture; if the temperature is too low, increase the indoor temperature. If the
above two methods
could not be provided, use a blow drier to heat the measuring pool and
remove moisture.
System Overview
adding reaction liquid). It may lead to the insufficiency of transferred CO2 and
low value of CO2.
But if the equipment is going to be powered on in 24 hours, then there is no
need to take them off.
(When putting pumping pipes back, make sure they are not installed
reversely, otherwise samples
could not be drawn in).
2)CO2 should be calibrated every 1-3 days so as to keep the result of CO2
reliable and
accurate.
CO2 value is low:Caused by short of reaction liquid due to that the pumping
pipe of reaction
liquid get sticky.
Solution:Take off the pumping pipe first and separate the stuck part;
then put the
pumping pipe back and manually turn the pumping pipe of reaction liquid
to draw in
“reaction liquid” until the “reaction liquid” drop in reaction pool.
CO2 value is high:It may be caused due to being calibrated for a long
time
Solution:Conduct CO2 calibration again.