The optimum light level is adjusted automatically to compensate for lamp degradation and flow cell staining. A thermal design that allows initial setup in less than two hours, typically; warm-up of the module is generally less than one hour for fast startup. A recycle valve that can be switched on to keep the system at operating conditions avoiding any warm-up time and saving solvent. Early maintenance feedback (EMF), with user-settable limits for continuous tracking of the reference flow cell purging. Built-in GLP/GMP compliance, such as operational qualification/performance verification (OQ/PV) maximize productivity. Automatic purging of the reference flow cell by applying user-settable purge and wait times. For unattended operation autozero can be programmed as part of the method.
Zero glass
Collimator lens
Countercurrent heat exchangers keep the optical unit and the flow cells at a constant temperature.
Flow cells
Mirror
Convenient front access for inspection and maintenance of tubings and valves. Self-diagnostics minimize downtime and ensure optimum system performance.
Enhance refractive index detector sensitivity while improving reproducibility and productivity
Specifications
nRIU Styreneacrylonitrile 20000 15000 10000 5000 0 -5000 -10000 0 5 10 15 20 25 30 Time [min] %RSD < 0.1% 20th day 10th day 5th day 1st day
Detection type: Short-term noise: Drift: Refractive index range: Flow cell: Temperature control: pH range: Control:
Deflection method 2.5 x 10-9 RIU* < 200 x 10-9 RIU/h* 1.001.75, calibrated 8 l cell volume, 5 bar (75 psi) pressure maximum Ambient +5 C to 55 C 2.39.5 Parameter entry, signal display, on-line help and diagnostics with the Agilent 1100 Series control module, optional PC (PCMCIA)-card for method, sequence and logbook storage and transfer. PC-based software for control and data evaluation with Agilent ChemStation Plus for HPLC Polarity, peak width Automatic zero Automatic purge and automatic solvent recycle Recorder/integrator: 100 mV or 1 V, with offset adjustment, output range selectable Controller-area network (CAN), GP-IB, RS-232, APG remote, optional local-area network (LAN). Remote: ready, start and shutdown signals
This overlay of 10 consecutive runs per day shows the remarkable intra- and interday retention times precision achieved with the Agilent 1100 GPC-SEC system.
Safety and maintenance: Extensive diagnostics, error detection and display leak detection, safe leak handling, leak output signal for shutdown of pumping system, low voltage in major maintenance areas
Carbohydrates
GLP features:
Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits (purge interval) and feedback messages. Electronic records of maintenance and errors, automated OQ/PV procedures. All materials recyclable 0 to 55 C constant temperature at < 95 % humidity (non-condensing) 180 x 345 x 435 mm (7 x 13.5 x 17 inches) 17 kg (38 lbs)
Housing: Environment:
5 6 7 8 1 2
Dimensions: (h x w x d) Weight:
10
12 14 Time [min]
Low limits of detection allow you to inject small amounts of sample. You save sample preparation time and extend column life.
* According to ASTM E-1303-95 "Practice for Refractive Index Detectors used in Liquid Chromatography." Reference conditions: response time 4 s, 35 C, 1 mL/min H2O (on-line degassing), restriction capillary.
Complete solution Agilent 1100 Series GPC-SEC system is a complete solution, based on the refractive index detector. The configuration comes with a startup kit containing: an application-specific column, reference standards, an application note with setup instructions and performance data
www.agilent.com/chem/1100