Ev2xx HL
Ev2xx HL
Ev2xx HL
MANUAL
HANDLEIDING
MODE DEMPLOI
ANLEITUNG
Distributor:
Topac Inc
231 CJC Highway
Cohasset
MA 02025
Tel 781 740 8778
Fax 781 740 8779
www.topac.com
email sales@topac.com
EV202
EV222
EV215
EV231
EV232
EV233
EV243
EV261
EV262
EV265
June 2011
Table of contents
Manufacturer ..................................... 1
Warranty .......................................... 1
Introduction ...................................... 1
Servicing........................................... 1
Keyboard .......................................... 1
Very important ................................... 2
Set-up.............................................. 3
Method programming............................ 4
Manual programming ............................ 5
Temporarily changing parameters............. 5
Data-logging ...................................... 6
Alarms ............................................. 6
Inhoudstafel
Inleiding ........................................... 7
Toetsenbord....................................... 7
Fabrikant .......................................... 7
Waarborg .......................................... 7
Onderhoud ........................................ 7
Zeer belangrijk ................................... 8
Instellen ........................................... 9
Methode programmatie ......................... 10
Manuele programmatie ......................... 11
Tijdelijk parameters wijzigen ................. 11
Gegevens opslaan ................................ 12
Alarmen ........................................... 12
Table de matires
Prface ............................................ 13
Fabricant .......................................... 13
Clavier ............................................. 13
Garantie ........................................... 13
Service............................................. 13
Trs important.................................... 14
Configurer......................................... 15
Programmation mthode ....................... 16
Programmation manuelle ....................... 17
Changer temporairement les paramtres.... 17
Enregistrer des donnes ........................ 18
Alarmes............................................ 18
Inhaltsverzeichnis
Hersteller ......................................... 19
Einfhrung ........................................ 19
Garantie ........................................... 19
Service............................................. 19
Bedienungsfeld ................................... 19
Sehr wichtig ...................................... 20
Einstellen.......................................... 21
Methode Programmierung ...................... 22
Manuelle Programmierung...................... 23
Zeitlich Parameter ndern ..................... 23
Alarme ............................................. 24
Daten speichern.................................. 24
Introduction
This instrument is manufactured with the latest technology and needs no particular maintenance. Consort certifies that this instrument was thoroughly inspected
and tested at the factory prior to shipment and found to meet all requirements
defined by contract under which it is furnished. However, dimensions and other
physical characteristics may differ.
The normal operating temperature should be between 0 and 40C. Never use
the instrument in a room with high humidity (>95 %) or at very low temperatures
(condensation water!).
Connect the instrument only to an earthed power line. The required power source
is indicated on the label at the back of the instrument. Do not cut and splice the
power cord. When removing the power cord from the wall outlet, be sure to unplug
by holding the plug attachment and not by pulling the cord. Do not hold the plug
by wet hand.
Manufacturer
Consort bvba
Parklaan 36
B2300 Turnhout
Belgium
Tel (+32)(14)41 12 79
Fax (+32)(14)42 91 79
E-mail: info@Consort.be
Warranty
This instrument (excluding all accessories) is warranted against defective material
and workmanship for a period of thirty-six (36) months from the date of shipment
ex factory. Consort will repair all defective equipment returned to it during the
warranty period without charge, provided the equipment has been used under
normal laboratory conditions and in accordance with the operating limitations and
maintenance procedures in this instruction manual and when not having been
subject to accident, alteration, misuse or abuse. A return authorisation must be
obtained from Consort before returning any product for warranty repair on a
freight prepaid basis!
Consort is not liable for consequential damages arising out of the use or handling
of its products.
Servicing
In the event of this instrument being returned for servicing, the owner is requested
to remove the power supply lead and NOT to send the following items unless they
are suspect:
Manual
Cables
Accessories
If serious malfunctioning occurs, stop using the unit immediately and consult your
local Consort dealer.
Keyboard
EV202 EV222 EV215 EV231 EV232 EV233 EV243 EV261 EV262 EV265
Very important
This electrophoresis power supply is a high technology instrument available in
several versions. As it is capable of giving dangerous voltage levels by which high
power is involved, we suggest that you take a few moments to read this manual
thoroughly. Although this instrument is equipped with all necessary safety features
against abuse and other accidental failures, caution should be exercised when
working with high voltage equipment.
Avoid to touch the outlets with any conducting object and make sure there is
a second person present for your safety in case of any severe electric shock.
Never touch any part of the assembly (power supply, leads or tank) before having switched off.
Never make any other connections, such as e.g. putting several power supplies
in series or in parallel.
In order to prevent electric shock, never open the back plate nor remove the
cover.
Take care so that the power supply is not dropped to avoid damaging the cabinet which defeats safeguards or injuring yourself.
If the unit has been dropped or the cabinet has been damaged, unplug it
and have it checked by an authorised service technician to restore the safeguards.
The fact that the unit operates satisfactorily does not imply that the unit is
properly earthed or that it is completely safe. If in any doubt about the effective earthing of the unit, contact a qualified electrician.
Never block the ventilation holes or place the unit in any enclosure unless
proper ventilation is provided.
Never place the unit near or over a radiator, heat register or stove.
Avoid locations where the instrument is exposed directly to the sun light.
EV202 EV222 EV215 EV231 EV232 EV233 EV243 EV261 EV262 EV265
Set-up
----
STAND-BY
----
SET-UP
----
SET-UP
----
LANGUAGE
CHOOSE LANGUAGE:
ENGLISH
----
SET-UP
----
CONTRAST
----
SET-UP
1.
2.
3.
4.
5.
6.
7.
Select INFO and press SET to see general information about the electrophoresis
power supply.
----
ALARM
CURRENT<1mA
ALARM:
YES
----
----
SET-UP
POWER FAIL
POWER FAIL
DETECTION:
YES
----
----
SET-UP
STAND-BY
----
SET-UP
----
INFO
EV202 EV222 EV215 EV231 EV232 EV233 EV243 EV261 EV262 EV265
(not on EV222)
Method programming permits to store up to 9 different methods, each
with 9 steps, of frequently used parameters in the non-volatile memory
for future recall. Each step is able to recall a next one, providing a flexible multiple step function for special techniques. The method mode also
permits to program a linear voltage gradient for any step provided the
limiting current or power is not attained. Parameters can be changed
temporarily without interrupting the run.
1. Verify if the instrument is switched off.
2. Connect the electrophoresis tank to the instrument.
3. Switch the instrument on. If the timer is still running after switching
on, the power supply will automatically proceed with the experiment for the remaining time. Press STOP (within 10 seconds) if you
want to go in stand-by during the run. The parameters can then be
changed eventually. Press RUN to proceed with the experiment or
press SET, during 2 seconds, to go in stand-by again.
4. Select the METHOD mode and press SET.
5. Select the desired method and press SET.
6. Select the desired step and press SET.
7. Decide whether or not a voltage gradient has to be performed and
press SET.
8. Enter the desired voltage limit and press SET (= the voltage end
point in case of a gradient step).
9. Enter the desired current limit and press SET (this limit is automatically set to maximum and cannot be changed in case of a gradient
step).
10. Enter the desired power limit and press SET (this limit is automatically set to maximum and cannot be changed in case of a gradient
step).
11. Enter the desired time-units (h of kVh) and press SET.
12. Enter the desired time and press SET.
13. Decide if the experiment should automatically proceed with a next
step (resuming from point 6) or stop after this step. Enter the
desired procedure and press SET.
14. Press RUN to start the experiment.
At any time, you can return in the stand-by mode by holding SET pressed during 2
seconds.
At any time, you can avoid programming unnecessary parameters. To do so, simply
press RUN instead of SET after entering the last relevant value and the instrument
will immediately start the experiment keeping the other parameters to their previous values.
To ensure that the desired parameter is kept constant, we suggest to pre-set all
other parameters to maximum.
A blinking unit of the controlling parameter appears when the regulation is complete.
The timer will always count down.
step 2
step 3
step 4
Method programming
TIMER ACTIVE!
STOP NOW?
10s
----
----
STAND-BY
METHOD
ENTER
11
METHOD:
ENTER
1
11
STEP:
ENTER
11
GRADIENT:
ENTER
NO
11
VOLTAGE:
ENTER
400V
11
CURRENT:
ENTER
300mA
11
POWER:
ENTER
50W
11
TIME UNIT:
ENTER
h
11
TIME:
99:59h
ENTER
11
END METHOD:
YES
RUN
11
400V
300mA
99:59h
50W
step 5
time
EV202 EV222 EV215 EV231 EV232 EV233 EV243 EV261 EV262 EV265
Manual programming
Manual programming permits to set voltage, current, power and time
limits for a simple routine electrophoresis run. Parameters can be
changed temporarily without interrupting the run.
TIMER ACTIVE!
STOP NOW?
----
10s
STAND-BY
----
MANUAL
ENTER
VOLTAGE:
400V
ENTER
CURRENT:
300mA
ENTER
POWER:
50W
ENTER
TIME UNIT:
ENTER
TIME:
99:59h
RUN
99:59h
400V
300mA
At any time, you can return in the stand-by mode by holding SET pressed during 2
seconds.
At any time, you can avoid programming unnecessary parameters. To do so, simply
press RUN instead of SET after entering the last relevant value and the instrument
will immediately start the experiment keeping the other parameters to their previous values.
To ensure that the desired parameter is kept constant, we suggest to pre-set all
other parameters to maximum.
A blinking unit of the controlling parameter appears when the regulation is complete.
The timer will always count down. However, when no time limit has been programmed it will count up.
50W
SET
11
400
300mA
SET
11
400V
SET
400V
99:59h
50W
99:59h
300
11
50W
99:59h
300mA
SET
11
400V
300mA
50
99:59
Only the parameters of the running step can be changed. A gradient step
can never be changed.
1. Press SET to change temporarily the parameters without interrupting the run.
2. Change the actual voltage limit and press SET.
3. Change the actual current limit and press SET.
4. Change the actual power limit, and press SET.
5. Change the actual time and press SET.
6. The power supply will now continue the run.
50W
EV202 EV222 EV215 EV231 EV232 EV233 EV243 EV261 EV262 EV265
(not on EV222)
1. Select SET-UP and press SET.
2. Select DATA and press SET.
3. Select DURATION and press SET.
4. Select the desired time for the data-logging and press SET.
5. Select LOG and press SET.
6. Decide whether or not all data (voltage, current, power, time) during
the next experiments should be logged and press SET.
7. Select SET-UP and press SET to return.
Data-logging
----
STAND-BY
----
SET-UP
----
SET-UP
----
DATA
-----
DATA
-----
DURATION
LOG DATA
FOR:
10h
-----
DATA
-----
LOG
LOG
DATA:
-----
YES
DATA
-----
SET-UP
Alarms
GROUND LEAKAGE
SHORT-CIRCUIT
OVERLOAD
CURRENT<1mA
CALL SERVICE
***
ERROR
***
GROUND LEAKAGE
***
ERROR
***
SHORT-CIRCUIT
***
ERROR
***
OVERLOAD
***
ERROR
***
CURRENT<1mA
***
ERROR
***
CALL SERVICE
You could also try a general RESET: switch the instrument on while holding SET
pressed. Attention! All stored programs and data will be erased!
EV202 EV222 EV215 EV231 EV232 EV233 EV243 EV261 EV262 EV265
Specifications
VOLTAGE
CURRENT
POWER
PARAMETER RANGE
TIMER
VOLT-HOURS
DISPLAY
RESOLUTION
PROGRAMS
OUTPUTS
MIN. LOAD RESISTANCE
NO LOAD DETECTION
GROUND LEAKAGE DETECTION
OVERLOAD DETECTION
COMPUTER CONTROL
DATA-LOGGING
RS232
AMBIENT TEMPERATURE
RELATIVE HUMIDITY
POWER REQUIREMENTS
DIMENSIONS (WxDxH)
WEIGHT
EV243
0...400 V
0...300 mA
0...50 W
1...100% of full scale
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
3 in parallel, 4 mm sockets
30
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 75 W
24x20x13 cm
3 kg
EV231
0...300 V
0...1000 mA
0...150 W
1...100% of full scale
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
4 in parallel, 4 mm sockets
10
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 200 W
31x26x15 cm
6 kg
EV265
0...600 V
0...500 mA
0...150 W
1...100% of full scale
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
4 in parallel, 4 mm sockets
30
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 200 W
31x26x15 cm
6 kg
Specifications
VOLTAGE
CURRENT
POWER
PARAMETER RANGE
TIMER
VOLT-HOURS
DISPLAY
RESOLUTION
PROGRAMS
OUTPUTS
MIN. LOAD RESISTANCE
NO LOAD DETECTION
GROUND LEAKAGE DETECTION
OVERLOAD DETECTION
COMPUTER CONTROL
DATA-LOGGING
RS232
AMBIENT TEMPERATURE
RELATIVE HUMIDITY
POWER REQUIREMENTS
DIMENSIONS (WxDxH)
WEIGHT
EV202
0...300 V
0...2000 mA
0...300 W
1...100% of full scale
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
4 in parallel, 4 mm sockets
5
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 360 W
31x26x15 cm
10 kg
EV261
0...600 V
0...1000 mA
0...300 W
1...100% of full scale
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
4 in parallel, 4 mm sockets
15
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 360 W
31x26x15 cm
10 kg
EV215
0...1200 V
0...500 mA
0...300 W
1...100% of full scale
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
4 in parallel, 4 mm sockets
70
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 360 W
31x26x15 cm
10 kg
Specifications
VOLTAGE
CURRENT
POWER
EV232
0...3000 V
0...150 mA
0...150 W
EV233
0...3000 V
0...300 mA
0...300 W
EV262
0...6000 V
0...150 mA
0...300 W
PARAMETER RANGE
TIMER
VOLT-HOURS
DISPLAY
RESOLUTION
PROGRAMS
OUTPUTS
MIN. LOAD RESISTANCE
NO LOAD DETECTION
GROUND LEAKAGE DETECTION
OVERLOAD DETECTION
COMPUTER CONTROL
DATA-LOGGING
RS232
AMBIENT TEMPERATURE
RELATIVE HUMIDITY
POWER REQUIREMENTS
DIMENSIONS (WxDxH)
WEIGHT
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
4 in parallel, 4 mm sockets
600
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 200 W
31x26x15 cm
6 kg
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
4 in parallel, 4 mm sockets
300
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 360 W
31x26x15 cm
10 kg
0...99:59 h
0...99.99 kVh
LCD, 2x16 characters
1 V, 1 mA, 1 W
9x9 set of parameters
4 in parallel, 4 mm sockets
1200
on/off, programmable
3600 values
9600 b/s
0...40C
0...95%, non condensing
210...250 VAC, 50/60 Hz, 360 W
31x26x15 cm
10 kg
DECLARATION OF CONFORMITY
on behalf of Consort
WARRANTY CERTIFICATE
This instrument (excluding all accessories) is warranted against defective material and
workmanship for a period of thirty-six (36) months from the date of shipment ex factory.
Accessories and breakable items such as electrodes are not warranted unless proven to be
defective before shipment.
The original purchase order numbers, Consort invoice numbers and serial numbers of the
products must be provided.
CONSORT will repair all defective equipment returned to it during the warranty period
without charge (CIF Turnhout prepaid by sender), provided the equipment has been used
under normal laboratory conditions and in accordance with the operating limitations and
maintenance procedures described in the instruction manual and when not having been
subject to accident, alteration, misuse or abuse.
If the products have been used with or have come into contact with fluids, an MSDS (material safety data sheet) must be supplied prior to issuing a return authorisation.
A return authorisation must be obtained from CONSORT before returning any product for
warranty repair on a freight prepaid basis!