Fundamental Concepts of Bipolar and High-Density Surface EMG Understanding and Teaching for Clinical, Occupational, and Sport Applications: Origin, Detection, and Main Errors
<p>Maps of bioelectric signals: (<b>a</b>) example of an instantaneous surface EEG map obtained by a grid of electrodes on the scalp (reproduced from Figure 2 of [<a href="#B1-sensors-22-04150" class="html-bibr">1</a>]); (<b>b</b>) example of an instantaneous surface ECG map obtained by the simulation of a grid of electrodes on the torso (reproduced from Figure 4 of [<a href="#B2-sensors-22-04150" class="html-bibr">2</a>]).</p> "> Figure 2
<p>Example of motor unit action potential (MUAP) generation and propagation: electrode grids and maps of the right and left propagating potentials on the skin. Modified from (<a href="https://www.robertomerletti.it/en/emg/material/teaching/module5" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module5</a> accessed 29 April 2022, This date applies to all websites and URLs). (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup1</a>).</p> "> Figure 3
<p>Schematic representation of the fibers of two motor units (MU1 and MU2) that are innervated by two motoneurons (MN1 and MN2) in two innervation zones (IZ1 and IZ2): (<b>a</b>) the summation of the skin contributions of the propagating action potentials, as detected by a differential amplifier (bipolar or single differential detection), are depicted as MUAP1 and MUAP2 (observe that the contributions of deeper muscle fibers are smaller than those of the superficial fibers of each MU); (<b>b</b>) the interferential signal, which is the summation of the MUAP trains of many active MUs. Modified from Figure 5.1 of [<a href="#B22-sensors-22-04150" class="html-bibr">22</a>] with permission.</p> "> Figure 4
<p>Example of a bidirectional wave generated by dropping a stone in a narrow water channel (representing a single fiber). The floats represent a series of electrodes detecting the wave in subsequent times and in two directions. One pair of floats at equal distances from the stone detect identical waves and their difference is zero. Knowing the distance between floats and the delay between the waves they detect allows for the estimation of the wave propagation velocity. (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup2</a>). Courtesy of Dr. Alberto Botter.</p> "> Figure 5
<p>Schematic drawing of the right and left propagation of a motor unit action potential (MUAP) in time and space and the estimation of muscle fiber conduction velocity. When the distance between P1 and P3 is 20 mm and the delay is 5 ms, the muscle fiber conduction velocity is 20/5 = 4 mm/ms = 4 m/s. P1, P2, and P3 must be on the same side of the innervation zone. The signals detected by P1, P2, and P3 are called “monopolar”. The signals detected between P1–P2, P2–P3, and P1–P3 are called “bipolar” or “single differential (SD)”. The SD signal detected between P4 and P5 is near zero because of the symmetrical position of P4 and P5 with respect to the innervation zone. Modified from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module5" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module5</a>. (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup3</a>).</p> "> Figure 6
<p>Overlapping of the distributions of surface intensities: (<b>a</b>) distribution of intensity generated by a point source of light (or a source of electric potential) at depth h below the surface; (<b>b</b>) partially overlapping distributions of light intensities generated by two sources at depth h; (<b>c</b>) profiles of partially overlapping light intensities on the surface and along the line L. The same considerations apply to point sources (S<sub>1</sub> and S<sub>2</sub>) of electric potential whose surface maps extend along x and y, as indicated in (<b>b</b>). The intensity I<sub>P</sub> in point P on the surface is inversely proportional to the distance R<sub>1</sub> between S<sub>1</sub> and P and to the distance R<sub>2</sub> between S<sub>2</sub> and P, i.e., the voltage in P is V<sub>P</sub> = kS<sub>1</sub>/R<sub>1</sub> + kS<sub>2</sub>/R<sub>2</sub>, where k is a proportionality constant that accounts for the medium properties.</p> "> Figure 7
<p>Schematic drawing of the propagation of two motor unit action potentials (MUAPs) under a linear electrode array that is aligned with the fiber direction. MU1 discharges at time t<sub>1</sub> and MU2 discharges at time t<sub>2</sub>: example of the estimation of the innervation zones and the calculation of the muscle fiber conduction velocity (CV). (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup4</a>). Reproduced from (<a href="https://www.robertomerletti.it/en/emg/material/teaching/module6" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module6</a>).</p> "> Figure 8
<p>Example of experimental sEMG signals recorded with 15 single differential channels from a biceps brachii. We can see the “signatures” of at least 9 MUs, locate their innervation zones, and estimate their length and conduction velocity (CV). The slopes of the dashed lines are the CVs of the motor unit action potentials (MUAPs). This piece of information cannot be obtained by one pair of electrodes and requires detection by an array of electrodes placed along the muscle fiber direction, as in <a href="#sensors-22-04150-f007" class="html-fig">Figure 7</a>. IED = interelectrode distance. Modified from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module6" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module6</a>.</p> "> Figure 9
<p>Example of experimental sEMG signals recorded with a 12-electrode array providing 11 single differential (SD) channels from a biceps brachii. In the 45-ms window, we can see the “signatures” of three MUs and locate their innervation zones. In this case, global conduction velocity (CV) can only be estimated in the proximal and distal regions of the muscle. In the middle region (Channels 4 to 8), some motor unit action potentials (MUAPs) propagate upward and some downward. The concept of global, or mean, muscle fiber CV is meaningless in this region and can only be estimated for individual MUs after extracting their MUAPs by means of sEMG decomposition [<a href="#B23-sensors-22-04150" class="html-bibr">23</a>]. IED = interelectrode distance. Reproduced from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module6" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module6</a>.</p> "> Figure 10
<p>(<b>a</b>) Schematic representation of the transmembrane current as two opposite dipoles (a tripole) of current, which implies the representation of the action potential as a triangular waveform; (<b>b</b>) extinction of the tripole at the end of the fiber, where the dipole d<sub>1</sub> shrinks (at times t<sub>1</sub> and t<sub>2</sub>) and disappears (at time t<sub>3</sub>), leaving dipole d<sub>2</sub> whose field is no longer partially canceled by d<sub>1</sub>. As d<sub>2</sub> shrinks and disappears (at times t<sub>4</sub> and t<sub>5</sub>, respectively), the surface potential disappears; (<b>c</b>) example of a real monopolar motor unit action potential (MUAP) whose propagation phase is followed by a non-propagating transient due to the end-of-fiber (EOF) effect. This transient is partially removed by bipolar (SD) detection. In general, the sEMG signal includes propagating and non-propagating components, whose presence requires caution in selecting the signal segment to use for the estimation of muscle fiber CV. IED, interelectrode distance. Modified from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module5" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module5</a>.</p> "> Figure 11
<p>Slow-motion simulation of the generation, propagation, and extinction of an action potential produced by a single fiber whose neuromuscular junction is not in the middle of the fiber. Observe the two end-of-fiber effects taking place at different times at the fiber-tendon junctions T1 and T2. They generate two sharp non-propagating transients in the sEMG signal. For a motor unit, the end-of fiber effect(s) may be smaller and less sharp, depending on the scatter of the fiber terminations. (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup5</a>). Reproduced from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module9" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module9</a>.</p> "> Figure 12
<p>Effects of fiber depth within the muscle: skin thickness = 1 mm, subcutaneous (fat) tissue thickness = 3 mm, interelectrode distance = 5 mm. Observe how, in the monopolar recording, the end-of-fiber effect decreases slowly and becomes greater than the propagating component for depths near 8 mm within the muscle. This contributes to the crosstalk signal that is present on nearby muscles. (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup6</a>). Reproduced from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module9" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module9</a>.</p> "> Figure 13
<p>Effects of lateral displacement (transversal distance) between the electrode array and the fiber: same conditions as in <a href="#sensors-22-04150-f012" class="html-fig">Figure 12</a>. Observe that the end-of-fiber effect is visible for distances of greater than 10–15 mm, where it becomes larger than the propagating component. This contributes to crosstalk between muscles that is still detectable at lateral distances of 20–30 mm from the edge of a muscle, mostly as a non-propagating component. (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup7</a>). Reproduced from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module9" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module9</a>.</p> "> Figure 14
<p>Effects of the rotation of the fiber with respect to the array (or vice versa): the rotation center is the center of the fiber that coincides with that of the array and with the innervation zone (Channel 8). Observe how the detected propagating action potential decreases as it moves along the fiber. This is because the source moves away from the electrodes. The case of a fiber inclined and displaced laterally with respect to the electrode array (not presented) is frequent for muscles that have a fan-like or pinnate structure. More information is provided in [<a href="#B29-sensors-22-04150" class="html-bibr">29</a>]. (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup8</a>). Reproduced from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module9" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module9</a>.</p> "> Figure 15
<p>Effects of changing muscle fiber conduction velocity (CV). Observe that the width of the action potential and the end-of-fiber effect become narrower and the slope of the pattern increases as the CV increases. The observed changes affect both the single sEMG channels and the array of signals. When a single channel is recorded, the effects are not detectable. (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup9</a>). Reproduced from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module9" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module9</a>.</p> "> Figure 16
<p>Electrode configurations or “montages”: (<b>a</b>) monopolar (any combination or spatial filter can be obtained by a grid of monopolar signals); (<b>b</b>) bipolar or single differential (SD); (<b>c</b>) array of SD signals; (<b>d</b>) double differential (DD); (<b>e</b>) normal (perpendicular) DD (NDD) or Laplacian configuration. Other configurations are possible. The reference electrode R must be on a region with no sEMG activity and away from the muscle of interest. It should not be on the same muscle. The NDD system is widely used in EEG because of its selectivity. Some recently developed systems do not need a reference electrode. A is the amplification or gain of the differential amplifier. Modified from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module6" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module6</a>.</p> "> Figure 17
<p>Example of sEMG distribution on the erector spinae of a healthy subject: (<b>a</b>) grid of 8 × 4 electrodes (electrode diameter = 3 mm; center-to-center interelectrode distance IED = 10 mm). (<b>b</b>) two such grids are applied on each side of the spine to produce an array of 15 × 4 single differential signals whose root mean square (RMS) value (over a 1s epoch) is depicted in (<b>c</b>), according to the color scale; (<b>d</b>) interpolated image, obtained from (<b>c</b>), of the RMS distribution with an indication of the orientation of the fibers. The blue area in the middle corresponds to the innervation zone of the superficial motor units.</p> "> Figure 18
<p>Effects of electrode length on a signal: (<b>a</b>) the green signal propagates in space under the gray fixed electrode of length L, generating the instantaneous signal indicated by the red bar. The signal evolves in time, as depicted in panel (<b>b</b>). (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup10</a>). Courtesy of Dr. Alberto Botter.</p> "> Figure 19
<p>Electrode filtering of a monopolar experimental sEMG signal. (<b>a</b>) moving average performed by a round electrode over a propagating monodimensional signal, (<b>b</b>) a signal collected using a pin electrode is filtered by the transfer function of electrodes of different sizes (in one dimension only), producing a number of outputs that correspond to different electrode diameters. The signal is low-pass filtered and attenuated because only the filtering effect of the different electrode sizes are accounted for. In practice, a larger electrode covers more fibers, so the amplitude may not be reduced (it may actually increase) but the spectral alteration due to filtering remains and alters the amplitude and spectral features of the signal. Modified from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module9" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module9</a>.</p> "> Figure 20
<p>Concept of spatial filtering that is introduced by bipolar (SD) detection. Panels (<b>a</b>,<b>b</b>) show how the same triangular voltage profile propagating under an electrode pair can produce very different SD signals, depending on the interelectrode distance (IED). The same happens for the sEMG. To avoid this effect in sEMG detection, IEDs should be ≤5 mm (but values of up to 8–10 mm introduce signal alterations that are acceptable for most applications). This is particularly important for detection using electrode grids. It is less important for clinical bipolar detection, as long as the same electrode size, IED, and location are used in subsequent measurements for comparing sEMG features, for example, in pre- and post-treatment or intervention. (See animation in <a href="#app1-sensors-22-04150" class="html-app">Supplementary Materials, Sup11</a>). Reproduced from <a href="https://www.robertomerletti.it/en/emg/material/teaching/module9" target="_blank">https://www.robertomerletti.it/en/emg/material/teaching/module9</a>.</p> "> Figure 21
<p>Example of a biceps brachii muscle shortening under the skin during a dynamic contraction. When the forearm flexes from 150° to 70° (internal angle with respect to the arm), the biceps muscle shortens under the array and the map changes from Map 1 to Map 2. Considering the white and the black pair of electrodes as being positioned by operators A and B, the white pair would indicate that elbow flexion causes an increase in the sEMG amplitude whereas the black pair would indicate the opposite. Great caution must be used during the placement of single electrode pairs for the study of dynamic contractions because the muscle geometry changes taking place under the electrodes cause sEMG variations that may be erroneously attributed to changes in the neural muscle drive when electrodes are placed on or near to the innervation zones (IZs). Adapted with permission from Figure 6.5 of Ref. [<a href="#B22-sensors-22-04150" class="html-bibr">22</a>]. Copyright Springer Verlag Italia, 2012.</p> "> Figure 22
<p>A study of the anal sphincter pre- and post-episiotomy: (<b>a</b>) the anal probe with 16 silver electrodes equi-spaced along the circumference; the detection of the innervation zones (IZs) in the same woman before (<b>b</b>) and after (<b>c</b>) delivery with episiotomy. The black cylinders represent the neuromuscular junctions and the gray arcs depict the identified motor units (MUs). In this subject, no MUs were detected on the side where the episiotomy was performed. The results from 82 cases of episiotomy are shown in (<b>d</b>). Fewer MUs were detected on the right ventral (RV) quadrant, where the episiotomies were performed, suggesting damage to the innervation following surgery. Note: the information collected before the delivery was not used to plan the surgery. Quadrants: LV, left ventral; LD, left dorsal; RD, right dorsal; RV, right ventral. Adapted with permission from Figures 1, 2, and 4 from Ref. [<a href="#B45-sensors-22-04150" class="html-bibr">45</a>]. Copyright The International Urogynecological Association, 2014.</p> "> Figure 23
<p>Characteristic changes in the normal double differential spatially filtered (<a href="#sensors-22-04150-f016" class="html-fig">Figure 16</a>) high-density sEMG (HDsEMG) signal associated with different disorders: (<b>a</b>) myopathies lead to a change in the shape of the motor unit action potentials (MUAPs) while neuropathies lead to a decrease in the number of MUAPs contributing to the signal; (<b>b</b>) neuromuscular disorders involving the upper motor neuron, such as stroke, are characterized in the spatially filtered HDsEMG signal not only by fewer active motor units, but also by a lower variability in firing rate [<a href="#B49-sensors-22-04150" class="html-bibr">49</a>,<a href="#B54-sensors-22-04150" class="html-bibr">54</a>]. For a better representation of these effects, (<b>a</b>,<b>b</b>) have different temporal scales.</p> "> Figure 24
<p>Change in the mean conduction velocities of motor unit action potentials (MUAP CVs) with age in healthy subjects and in patients suffering from Duchenne muscular dystrophy or spinal muscle atrophy. The solid black line is the regression curve for the healthy subjects. The gray area contains 95% of all healthy subjects. Adapted with permission from Ref. [<a href="#B55-sensors-22-04150" class="html-bibr">55</a>]. Copyright Muscle and Nerve, 1997.</p> "> Figure 25
<p>Envelopes of bipolar sEMG (one plot per stride for 10 strides) from nine electrode pairs obtained by placing a 4 × 3 grid of electrodes, 15 × 18 mm in size, over the anterior aspect of the leg. The medial column of electrodes is placed on the edge between the tibialis anterior muscle and the tibia bone. The gray bands indicate the presence of activity in the mid and terminal stances due to signals coming from the peroneus longus. Adapted with permission from Ref. [<a href="#B66-sensors-22-04150" class="html-bibr">66</a>]. Copyright Journ. of Electromyography and kinesiology, 2007.</p> "> Figure 26
<p>Example of sEMG signals acquired during gait analysis of the leg of a healthy 10-year-old boy by means of pairs of circular electrodes: diameter of the conductive region = 10 mm; detection area = 78.5 mm<sup>2</sup>; center to center distance = 20 mm. Crosstalk on the tibialis anterior from the plantar flexors is outlined in purple. Reproduced from Ref. [<a href="#B72-sensors-22-04150" class="html-bibr">72</a>].</p> "> Figure 27
<p>Bipolar sEMG recordings from eight muscles of the left leg during the gait analysis of a subject: (<b>a</b>) “autoscale on”; (<b>b</b>) same scale for all channels. A superficial visual analysis of (<b>a</b>) would lead to incorrect conclusions.</p> ">
Abstract
:1. Introduction
- (a)
- (b)
- responding to the persistent demand for the simplification of teaching these concepts;
- (c)
- making users aware of the tutorials and teaching materials that are available.
2. Teaching Basic Physiological and Technical Concepts
2.1. The Concept of Motor Unit Action Potential (MUAP)
2.2. The Concept of Propagating Action Potentials and Their Velocity and Diffusion
2.2.1. Conduction Velocity
2.2.2. The Concepts of Volume Conductor and Crosstalk
2.3. Teaching Differential/Bipolar Detection, Innervation Zones, and the End-of-Fiber Effect
2.3.1. Teaching Action Potentials through Modeling
2.3.2. Spatial Filters
2.4. Two-Dimensional (2D) sEMG Representation and sEMG Images
3. The Role of Electrode Size and Interelectrode Distance in Bipolar sEMG
3.1. Electrode Size and Its Filtering Effect
3.2. Interelectrode Distance and Its Filtering Effect
4. Factors That Can Affect sEMG and Its Physiological Interpretation
4.1. Skin Treatment
4.2. Movement of the Muscle under the Electrodes
4.3. Identification of Innervation Zones
4.4. Examples of HDsEMG Applications in Neurophysiology
4.5. Issues of Bipolar sEMG Detection in Neurorehabilitation Applications
4.6. Issues in Ergonomics, Occupational Medicine, and Sport Sciences Applications
5. Discussion and Limitations of the Work
6. Conclusions
Supplementary Materials
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Acknowledgments
Conflicts of Interest
Acronyms
EEG | Electroencephalogram |
ECG | Electrocardiogram |
EMG | Electromyogram/electromyography |
sEMG | Surface electromyogram |
HDsEMG | High-density sEMG |
IZ | Innervation zone of a motor unit or muscle |
MU | Motor unit |
MUAP | Motor unit action potential |
NMJ | Neuromuscular junction |
IED | Interelectrode distance |
EOF | End-of-fiber |
CV | Conduction velocity of muscle fibers |
RMS | Root mean square |
SD | Single differential (bipolar) |
DD | Double differential |
NDD | Normal DD or Laplacian |
VL | Vastus lateralis |
VM | Vastus medialis |
References
- Lustenberger, C.; Huber, R. High density electroencephalography in sleep research: Potential, problems, future perspective. Front. Neurol. 2012, 3, 77. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Dössel, O.; Luongo, G.; Nagel, C.; Loewe, A. Computer Modeling of the Heart for ECG Interpretation—A Review. Hearts 2021, 2, 350–368. [Google Scholar] [CrossRef]
- Merletti, R.; Campanini, I.; Rymer, W.Z.; Disselhorst-Klug, C. Editorial: Surface Electromyography: Barriers Limiting Widespread Use of sEMG in Clinical Assessment and Neurorehabilitation. Front. Neurol. 2021, 12, 10–13. [Google Scholar] [CrossRef] [PubMed]
- Van Campenhout, A.; Bar-On, L.; Desloovere, K.; Huenaerts, C.; Molenaers, G. Motor endplate-targeted botulinum toxin injections of the gracilis muscle in children with cerebral palsy. Dev. Med. Child Neurol. 2015, 57, 476–483. [Google Scholar] [CrossRef] [PubMed]
- Chandra, S.; Afsharipour, B.; Rymer, W.Z.; Suresh, N.L. Precise quantification of the time course of voluntary activation capacity following Botulinum toxin injections in the biceps brachii muscles of chronic stroke survivors. J. Neuroeng. Rehabil. 2020, 17, 102. [Google Scholar] [CrossRef] [PubMed]
- Rojas-Martínez, M.; Alonso, J.F.; Jordanić, M.; Mañanas, M.Á.; Chaler, J. Analysis of muscle load-sharing in patients with lateral epicondylitis during endurance isokinetic contractions using non-linear prediction. Front. Physiol. 2019, 10, 1185. [Google Scholar] [CrossRef]
- Zwarts, M.J.; Stegeman, D.F. Multichannel surface EMG: Basic aspects and clinical utility. Muscle Nerve 2003, 28, 1–17. [Google Scholar] [CrossRef]
- Drost, G.; Verrips, A.; van Engelen, B.G.M.; Stegeman, D.F.; Zwarts, M.J. Involuntary painful muscle contractions in Satoyoshi syndrome: A surface electromyographic study. Mov. Disord. 2006, 21, 2015–2018. [Google Scholar] [CrossRef]
- Drost, G.; Stegeman, D.F.; van Engelen, B.G.M.; Zwarts, M.J. Clinical applications of high-density surface EMG: A systematic review. J. Electromyogr. Kinesiol. 2006, 16, 586–602. [Google Scholar] [CrossRef]
- Minetto, M.A.; Lanfranco, F.; Botter, A.; Motta, G.; Mengozzi, G.; Giordano, R.; Picu, A.; Ghigo, E.; Arvat, E. Do muscle fiber conduction slowing and decreased levels of circulating muscle proteins represent sensitive markers of steroid myopathy? A pilot study in Cushing’s disease. Eur. J. Endocrinol. 2011, 164, 985–993. [Google Scholar] [CrossRef]
- Mottram, C.J.; Heckman, C.J.; Powers, R.K.; Rymer, W.Z.; Suresh, N.L. Disturbances of motor unit rate modulation are prevalent in muscles of spastic-paretic stroke survivors. J. Neurophysiol. 2014, 111, 2017–2028. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Campanini, I.; Merlo, A.; Farina, D. Motor unit discharge pattern and conduction velocity in patients with upper motor neuron syndrome. J. Electromyogr. Kinesiol. 2009, 19, 22–29. [Google Scholar] [CrossRef] [PubMed]
- Merlo, A.; Montecchi, M.G.; Lombardi, F.; Vata, X.; Musi, A.; Lusuardi, M.; Merletti, R.; Campanini, I. Monitoring Involuntary Muscle Activity in Acute Patients with Upper Motor Neuron Lesion by Wearable Sensors: A Feasibility Study. Sensors 2021, 21, 3120. [Google Scholar] [CrossRef] [PubMed]
- Minetto, M.; Holobar, A.; Botter, A.; Ravenni, R.; Farina, D. Mechanisms of cramp contractions: Peripheral or central generation? J. Physiol. 2011, 589, 5759–5773. [Google Scholar] [CrossRef]
- Russo, A.; Aranceta-Garza, A.; D’Emanuele, S.; Serafino, F.; Merletti, R. HDSEMG activity of the lumbar erector spinae in violin players comparison of two chairs. Med. Probl. Perform. Art. 2019, 34, 205–214. [Google Scholar] [CrossRef]
- Sarcher, A.; Brochard, S.; Hug, F.; Letellier, G.; Raison, M.; Perrouin-Verbe, B.; Sangeux, M.; Gross, R. Patterns of upper limb muscle activation in children with unilateral spastic cerebral palsy: Variability and detection of deviations. Clin. Biomech. 2018, 59, 85–93. [Google Scholar] [CrossRef]
- Merletti, R.; Disselhorst-Klug, C.; Rymer, W.Z.; Campanini, I. Surface Electromyography: Barriers Limiting Widespread Use of sEMG in Clinical Assessment and Neurorehabilitation; Frontiers Research Topics; Frontiers Media SA: Lausanne, Switzerland, 2021; Volume 12, ISBN 9782889666164. [Google Scholar]
- Campanini, I.; Disselhorst-Klug, C.; Rymer, W.Z.; Merletti, R. Surface EMG in Clinical Assessment and Neurorehabilitation: Barriers Limiting Its Use. Front. Neurol. 2020, 11, 934. [Google Scholar] [CrossRef]
- Merletti, R.; Muceli, S. Tutorial. Surface EMG detection in space and time: Best practices. J. Electromyogr. Kinesiol. 2019, 49, 102363. [Google Scholar] [CrossRef]
- Merletti, R.; Cerone, G.L. Tutorial. Surface EMG detection, conditioning and pre-processing: Best practices. J. Electromyogr. Kinesiol. 2020, 54, 102440. [Google Scholar] [CrossRef]
- Heckman, C.J.; Enoka, R.M. Motor Unit. In Comprehensive Physiology; J. Wiley & Sons: Hoboken, NJ, USA, 2012; Volume 2, pp. 2629–2682. [Google Scholar]
- Barbero, M.; Merletti, R.; Rainoldi, A. Atlas of Muscle Innervation Zones; Springer: Milan, Italia, 2012; ISBN 978-88-470-2462-5. [Google Scholar]
- Holobar, A.; Zazula, D. Correlation-based decomposition of surface electromyograms at low contraction forces. Med. Biol. Eng. Comput. 2004, 42, 487–495. [Google Scholar] [CrossRef]
- Merletti, R.; Roy, S.H.; Kupa, E.; Roatta, S.; Granata, A. Modeling of surface myoelectric signals—Part II: Model-based signal interpretation. IEEE Trans. Biomed. Eng. 1999, 46, 821–829. [Google Scholar] [CrossRef] [PubMed]
- Merletti, R.; Lo Conte, L.; Avignone, E.; Guglielminotti, P. Modeling of surface myoelectric signals—Part I: Model implementation. IEEE Trans. Biomed. Eng. 1999, 46, 810–820. [Google Scholar] [CrossRef] [PubMed]
- Disselhorst-Klug, C.; Silny, J.; Rau, G. Estimation of the relationship between the noninvasively detected activity of single motor units and their characteristic pathological changes by modelling. J. Electromyogr. Kinesiol. 1998, 8, 323–335. [Google Scholar] [CrossRef]
- Farina, D.; Mesin, L.; Martina, S.; Merletti, R. A Surface EMG Generation Model with Multilayer Cylindrical Description of the Volume Conductor. IEEE Trans. Biomed. Eng. 2004, 51, 415–426. [Google Scholar] [CrossRef] [Green Version]
- Mesin, L.; Farina, D. Simulation of Surface EMG Signals Generated by Muscle Tissues with Inhomogeneity Due to Fiber Pinnation. IEEE Trans. Biomed. Eng. 2004, 51, 1521–1529. [Google Scholar] [CrossRef]
- Vieira, T.M.; Botter, A. The Accurate Assessment of Muscle Excitation Requires the Detection of Multiple Surface Electromyograms. Exerc. Sport Sci. Rev. 2021, 49, 23–34. [Google Scholar] [CrossRef]
- Minetto, M.A.; Botter, A.; Šprager, S.; Agosti, F.; Patrizi, A.; Lanfranco, F.; Sartorio, A. Feasibility study of detecting surface electromyograms in severely obese patients. J. Electromyogr. Kinesiol. 2013, 23, 285–295. [Google Scholar] [CrossRef]
- Piervirgili, G.; Petracca, F.; Merletti, R. A new method to assess skin treatments for lowering the impedance and noise of individual gelled Ag-AgCl electrodes. Physiol. Meas. 2014, 35, 2101–2118. [Google Scholar] [CrossRef]
- Hogrel, J.-Y. Clinical applications of surface electromyography in neuromuscular disorders. Neurophysiol. Clin. Neurophysiol. 2005, 35, 59–71. [Google Scholar] [CrossRef]
- Afsharipour, B.; Soedirdjo, S.; Merletti, R. Two-dimensional surface EMG: The effects of electrode size, interelectrode distance and image truncation. Biomed. Signal Process. Control 2019, 49, 298–307. [Google Scholar] [CrossRef]
- Schwartz, M.W. EMG Methods for Evaluating Muscle and Nerve Function; Schwartz, M., Ed.; InTech: London, UK, 2012; Volume 68, ISBN 978-953-307-793-2. [Google Scholar]
- Merletti, R.; Farina, D. Surface Electromyography: Physiology, Engineering, and Application; Merletti, R., Farina, D., Eds.; IEEE Press: Hoboken, NJ, USA; John Wiley & Sons, Inc.: Hoboken, NJ, USA, 2016; ISBN 9781119082934. [Google Scholar]
- Besomi, M.; Hodges, P.W.; Van Dieën, J.; Carson, R.G.; Clancy, E.A.; Disselhorst-Klug, C.; Holobar, A.; Hug, F.; Kiernan, M.C.; Lowery, M.; et al. Consensus for experimental design in electromyography (CEDE) project: Electrode selection matrix. J. Electromyogr. Kinesiol. 2019, 48, 128–144. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Farina, D.; Madeleine, P.; Graven-Nielsen, T.; Merletti, R.; Arendt-Nielsen, L. Standardising surface electromyogram recordings for assessment of activity and fatigue in the human upper trapezius muscle. Eur. J. Appl. Physiol. 2002, 86, 469–478. [Google Scholar] [CrossRef] [PubMed]
- Farina, D.; Merletti, R.; Indino, B.; Nazzaro, M.; Pozzo, M. Surface EMG crosstalk between knee extensor muscles: Experimental and model results. Muscle Nerve 2002, 26, 681–695. [Google Scholar] [CrossRef] [PubMed]
- Merlo, A.; Bò, M.C.; Campanini, I. Electrode Size and Placement for Surface EMG Bipolar Detection from the Brachioradialis Muscle: A Scoping Review. Sensors 2021, 21, 7322. [Google Scholar] [CrossRef] [PubMed]
- Nishihara, K.; Chiba, Y.; Suzuki, Y.; Moriyama, H.; Kanemura, N.; Ito, T.; Takayanagi, K.; Gomi, T. Effect of position of electrodes relative to the innervation zone on surface EMG. J. Med. Eng. Technol. 2010, 34, 141–147. [Google Scholar] [CrossRef]
- Rodrigues, F.B.; Duarte, G.S.; Marques, R.E.; Castelão, M.; Ferreira, J.; Sampaio, C.; Moore, A.P.; Costa, J. Botulinum toxin type A therapy for cervical dystonia. Cochrane Database Syst. Rev. 2020, 2020, CD003633. [Google Scholar] [CrossRef]
- Lapatki, B.G.; Van Dijk, J.P.; Van de Warrenburg, B.P.C.; Zwarts, M.J. Botulinum toxin has an increased effect when targeted toward the muscle’s endplate zone: A high-density surface EMG guided study. Clin. Neurophysiol. 2011, 122, 1611–1616. [Google Scholar] [CrossRef]
- Zoons, E.; Dijkgraaf, M.G.W.; Dijk, J.M.; van Schaik, I.N.; Tijssen, M.A. Botulinum toxin as treatment for focal dystonia: A systematic review of the pharmaco-therapeutic and pharmaco-economic value. J. Neurol. 2012, 259, 2519–2526. [Google Scholar] [CrossRef] [Green Version]
- Sultan, A.H.; Kamm, M.A.; Bartram, C.I.; Hudson, C.N. Anal sphincter trauma during instrumental delivery. Int. J. Gynecol. Obstet. 1993, 43, 263–270. [Google Scholar] [CrossRef]
- Cescon, C.; Riva, D.; Začesta, V.; Drusany-Starič, K.; Martsidis, K.; Protsepko, O.; Baessler, K.; Merletti, R. Effect of vaginal delivery on the external anal sphincter muscle innervation pattern evaluated by multichannel surface EMG: Results of the multicentre study TASI-2. Int. Urogynecol. J. 2014, 25, 1491–1499. [Google Scholar] [CrossRef]
- Enck, P.; Franz, H.; Davico, E.; Mastrangelo, F.; Mesin, L.; Merletti, R. Repeatability of innervation zone identification in the external anal sphincter muscle. Neurourol. Urodyn. 2010, 29, 449–457. [Google Scholar] [CrossRef] [PubMed]
- Peng, Y.; He, J.; Khavari, R.; Boone, T.B.; Zhang, Y. Functional mapping of the pelvic floor and sphincter muscles from high-density surface EMG recordings. Int. Urogynecol. J. 2016, 27, 1689–1696. [Google Scholar] [CrossRef] [PubMed] [Green Version]
- Ramaekers, V.; Disselhorst-Klug, C.; Schneider, J.; Silny, J.; Forst, J.; Forst, R.; Kotlarek, F.; Rau, G. Clinical Application of a Noninvasive Multi-Electrode Array EMG for the Recording of Single Motor Unit Activity. Neuropediatrics 1993, 24, 134–138. [Google Scholar] [CrossRef] [PubMed]
- Disselhorst-Klug, C.; Bahm, J.; Ramaekers, V.; Trachterna, A.; Rau, G. Non-invasive approach of motor unit recording during muscle contractions in humans. Eur. J. Appl. Physiol. 2000, 83, 144–150. [Google Scholar] [CrossRef] [PubMed]
- Buchthal, F.; Kamieniecka, Z. The diagnostic yield of quantified electromyography and quantified muscle biopsy in neuromuscular disorders. Muscle Nerve 1982, 5, 265–280. [Google Scholar] [CrossRef]
- Moloney, P.B.; Lefter, S.; Ryan, A.M.; Jansen, M.; Bermingham, N.; McNamara, B. The Diagnostic Yield of Electromyography at Detecting Abnormalities on Muscle Biopsy: A Single Center Experience. Neurodiagn. J. 2021, 61, 86–94. [Google Scholar] [CrossRef]
- Stegeman, D.F.; Kleine, B.U.; Lapatki, B.G.; Van Dijk, J.P. High-density Surface EMG: Techniques and Applications at a Motor Unit Level. Biocybern. Biomed. Eng. 2012, 32, 3–27. [Google Scholar] [CrossRef]
- Rasool, G.; Afsharipour, B.; Suresh, N.L.; Rymer, W.Z. Spatial Analysis of Multichannel Surface EMG in Hemiplegic Stroke. IEEE Trans. Neural Syst. Rehabil. Eng. 2017, 25, 1802–1811. [Google Scholar] [CrossRef]
- Williams, S.E.; Koch, K.C.; Disselhorst-Klug, C. Non-invasive assessment of motor unit activation in relation to motor neuron level and lesion location in stroke and spinal muscular atrophy. Clin. Biomech. 2020, 78, 105053. [Google Scholar] [CrossRef]
- Huppertz, H.J.; Disselhorst-Klug, C.; Silny, J.; Rau, G.; Heimann, G. Diagnostic yield of noninvasive high spatial resolution electromyography in neuromuscular diseases. Muscle Nerve 1997, 20, 1360–1370. [Google Scholar] [CrossRef]
- Rainoldi, A.; Gazzoni, M.; Casale, R. Surface EMG signal alterations in Carpal Tunnel syndrome: A pilot study. Eur. J. Appl. Physiol. 2008, 103, 233–242. [Google Scholar] [CrossRef] [PubMed]
- Wren, T.A.L.; Tucker, C.A.; Rethlefsen, S.A.; Gorton, G.E.; Õunpuu, S. Clinical efficacy of instrumented gait analysis: Systematic review 2020 update. Gait Posture 2020, 80, 274–279. [Google Scholar] [CrossRef] [PubMed]
- Campanini, I.; Cosma, M.; Manca, M.; Merlo, A. Added Value of Dynamic EMG in the Assessment of the Equinus and the Equinovarus Foot Deviation in Stroke Patients and Barriers Limiting Its Usage. Front. Neurol. 2020, 11, 583399. [Google Scholar] [CrossRef] [PubMed]
- Merlo, A.; Campanini, I. Impact of instrumental analysis of stiff knee gait on treatment appropriateness and associated costs in stroke patients. Gait Posture 2019, 72, 195–201. [Google Scholar] [CrossRef]
- Campanini, I.; Merlo, A.; Damiano, B. A method to differentiate the causes of stiff-knee gait in stroke patients. Gait Posture 2013, 38, 165–169. [Google Scholar] [CrossRef]
- Wren, T.A.L.; Chou, L.S.; Dreher, T. Gait and posture virtual special Issue “clinical impact of instrumented motion analysis”. Gait Posture 2020, 82, 108–109. [Google Scholar] [CrossRef]
- Wren, T.A.L.; Gorton, G.E.; Õunpuu, S.; Tucker, C.A. Efficacy of clinical gait analysis: A systematic review. Gait Posture 2011, 34, 149–153. [Google Scholar] [CrossRef]
- Ferrarin, M.; Rabuffetti, M.; Bacchini, M.; Casiraghi, A.; Castagna, A.; Pizzi, A.; Montesano, A. Does gait analysis change clinical decision-making in poststroke patients? Results from a pragmatic prospective observational study. Eur. J. Phys. Rehabil. Med. 2015, 51, 171–184. [Google Scholar]
- Mazzoli, D.; Giannotti, E.; Manca, M.; Longhi, M.; Prati, P.; Cosma, M.; Ferraresi, G.; Morelli, M.; Zerbinati, P.; Masiero, S.; et al. Electromyographic activity of the vastus intermedius muscle in patients with stiff-knee gait after stroke. A retrospective observational study. Gait Posture 2018, 60, 273–278. [Google Scholar] [CrossRef]
- Blanc, Y. EMG timing errors of pathologic gait. In Proceedings of the First General SENIAM (Surface EMG for Non- Invasive Assessment of Muscles) Workshop; Hermens, H.J., Ed.; RRD Publisher: Enschede, The Netherlands, 1996; pp. 183–185. [Google Scholar]
- Campanini, I.; Merlo, A.; Degola, P.; Merletti, R.; Vezzosi, G.; Farina, D. Effect of electrode location on EMG signal envelope in leg muscles during gait. J. Electromyogr. Kinesiol. 2007, 17, 515–526. [Google Scholar] [CrossRef]
- Benedetti, M.G.; Beghi, E.; De Tanti, A.; Cappozzo, A.; Basaglia, N.; Cutti, A.G.; Cereatti, A.; Stagni, R.; Verdini, F.; Manca, M.; et al. SIAMOC position paper on gait analysis in clinical practice: General requirements, methods and appropriateness. Results of an Italian consensus conference. Gait Posture 2017, 58, 252–260. [Google Scholar] [CrossRef] [PubMed]
- Hermens, H.; Freriks, B.; Merletti, R.; Stegeman, D.; Blok, J.; Rau, G.; Disselhorst-Klug, C.; Hagg, G. European Recommendations for Surface Electromyography; RRD Publisher: Enschede, The Netherlands, 1999; ISBN 90-75452-15-2. [Google Scholar]
- Blanc, Y. Electrode Placement for surface EMG. In AIM Project A2002 CAMARC-II (Computer Aided Movement Analysis in a Rehabilitation Context-II): Functional Evaluation Protocolos for Europe-Wide Network of Clinical Centres; RRD Publisher: Enschede, The Netherlands, 1994; pp. 25–30. [Google Scholar]
- Blanc, Y.; Dimanico, U. Electrode Placement in Surface Electromyography (sEMG) ”Minimal Crosstalk Area“ (MCA). Open Rehabil. J. 2014, 3, 110–126. [Google Scholar] [CrossRef] [Green Version]
- Blumenstein, R.; Basmajian, J. Electrode Placement in EMG Biofeedback; Williams & Wilkins: Baltimore, MA, USA, 1980. [Google Scholar]
- Merlo, A.; Campanini, I. Technical Aspects of Surface Electromyography for Clinicians. Open Rehabil. J. 2010, 3, 98–109. [Google Scholar] [CrossRef] [Green Version]
- Merlo, A.; Campanini, I. Applications in movement and gait analysis. In Surface Electromyography: Physiology, Engineering, and Applications; John Wiley & Sons, Inc.: Hoboken, NJ, USA, 2016; pp. 440–459. ISBN 978-1-118-98702-5. [Google Scholar]
- Shewman, T.; Konrad, P. Clinical Sequence Assessments and sEMG feedback. A Beginner’s Guide; Noraxon Inc.: Scottsdale, AZ, USA, 2011; ISBN 0977162249. [Google Scholar]
- Kofler, M.; Kreczy, A.; Gschwendtner, A. “Occupational backache”—Surface electromyography demonstrates the advantage of an ergonomic versus a standard microscope workstation. Eur. J. Appl. Physiol. 2002, 86, 492–497. [Google Scholar] [CrossRef] [PubMed]
- Felici, F. Neuromuscular responses to exercise investigated through surface EMG. J. Electromyogr. Kinesiol. 2006, 16, 578–585. [Google Scholar] [CrossRef] [PubMed]
- Felici, F.; Del Vecchio, A. Surface Electromyography: What Limits Its Use in Exercise and Sport Physiology? Front. Neurol. 2020, 11, 578504. [Google Scholar] [CrossRef]
- Martin, B.J.; Acosta-Sojo, Y. sEMG: A Window into Muscle Work, but Not Easy to Teach and Delicate to Practice—A Perspective on the Difficult Path to a Clinical Tool. Front. Neurol. 2021, 11, 588451. [Google Scholar] [CrossRef]
- Soderberg, G.L. Selected Topics in Surface Electromyography for Use in the Occupational Setting: Expert Perspectives; US Department of Health and Human Services: Washington, DC, USA, 1992; Volume 1.
- Kumar, S.; Mital, A. Electromyography in Ergonomics, 2nd ed.; Shrawan, K., Anil, M., Eds.; Routledge: England, UK, 2017; ISBN 9780203758670. [Google Scholar]
- Marras, W.S. Overview of Electromyography in Ergonomics. Proc. Hum. Factors Ergon. Soc. Annu. Meet. 2000, 44, 5-534–5-536. [Google Scholar] [CrossRef]
- Gazzoni, M.; Afsharipour, B.; Merletti, R. Surface EMG in Ergonomics and Occupational Medicine. In Surface Electromyography: Physiology, Engineering, and Applications; John Wiley & Sons, Inc.: Hoboken, NJ, USA, 2016; pp. 361–391. [Google Scholar]
- Varrecchia, T.; Ranavolo, A.; Conforto, S.; De Nunzio, A.M.; Arvanitidis, M.; Draicchio, F.; Falla, D. Bipolar versus high-density surface electromyography for evaluating risk in fatiguing frequency-dependent lifting activities. Appl. Ergon. 2021, 95, 103456. [Google Scholar] [CrossRef]
- Clarys, J.P.; Cabri, J. Electromyography and the study of sports movements: A review. J. Sports Sci. 1993, 11, 379–448. [Google Scholar] [CrossRef]
- Türker, H.; Sözen, H. Surface Electromyography in Sport and Exercise. Chapter 9; Turker, D.H., Ed.; InTech: London, UK, 2013; ISBN 978-953-51-1118-4. [Google Scholar]
- Rainoldi, A.; Gazzoni, M.; Melchiorri, G. Differences in myoelectric manifestations of fatigue in sprinters and long distance runners. Physiol. Meas. 2008, 29, 331–340. [Google Scholar] [CrossRef] [PubMed]
- Farago, E.; Macisaac, D.; Suk, M.; Chan, A.D.C. A Review of Techniques for Surface Electromyography Signal Quality Analysis. IEEE Rev. Biomed. Eng. 2022. in press. [Google Scholar] [CrossRef] [PubMed]
- Ivanenko, Y.P.; D’avella, A.; Lacquaniti, F. Muscle Coordination, Motor Synergies, and Primitives from Surface EMG. In Surface Electromyography: Physiology, Engineering, and Applications; John Wiley & Sons, Inc.: Hoboken, NJ, USA, 2016; pp. 158–179. ISBN 9781119082934. [Google Scholar]
- Besomi, M.; Hodges, P.W.; Clancy, E.A.; Van Dieën, J.; Hug, F.; Lowery, M.; Merletti, R.; Søgaard, K.; Wrigley, T.; Besier, T.; et al. Consensus for experimental design in electromyography (CEDE) project: Amplitude normalization matrix. J. Electromyogr. Kinesiol. 2020, 53, 102438. [Google Scholar] [CrossRef] [PubMed]
- McManus, L.; Lowery, M.; Merletti, R.; Søgaard, K.; Besomi, M.; Clancy, E.A.; van Dieën, J.H.; Hug, F.; Wrigley, T.; Besier, T.; et al. Consensus for experimental design in electromyography (CEDE) project: Terminology matrix. J. Electromyogr. Kinesiol. 2021, 59, 102565. [Google Scholar] [CrossRef]
- Rogers, E.M. Diffusion of Innovation; The Free Press: New York, NY, USA, 2003; ISBN 9780743258234. [Google Scholar]
- Jette, A.M. Overcoming Ignorance and Ineptitude in 21st Century Rehabilitation. Phys. Ther. 2017, 97, 497–498. [Google Scholar] [CrossRef]
- Hermens, H.J.; Freriks, B.; Disselhorst-Klug, C.; Rau, G. Development of recommendations for SEMG sensors and sensor placement procedures. J. Electromyogr. Kinesiol. 2000, 10, 361–374. [Google Scholar] [CrossRef]
- Manca, A.; Cereatti, A.; Bar-On, L.; Botter, A.; Croce, U.D.; Knaflitz, M.; Maffiuletti, N.A.; Mazzoli, D.; Merlo, A.; Roatta, S.; et al. A Survey on the Use and Barriers of Surface Electromyography in Neurorehabilitation. Front. Neurol. 2020, 11, 573616. [Google Scholar] [CrossRef]
- Jette, A.M. Moving research from the bedside into practice. Phys. Ther. 2016, 96, 594–596. [Google Scholar] [CrossRef]
- Gawande, A. Slow Ideas Some innovations spread fast. How do you speed the ones that don’t? New Yorker 2013, 29, 36–45. [Google Scholar]
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Campanini, I.; Merlo, A.; Disselhorst-Klug, C.; Mesin, L.; Muceli, S.; Merletti, R. Fundamental Concepts of Bipolar and High-Density Surface EMG Understanding and Teaching for Clinical, Occupational, and Sport Applications: Origin, Detection, and Main Errors. Sensors 2022, 22, 4150. https://doi.org/10.3390/s22114150
Campanini I, Merlo A, Disselhorst-Klug C, Mesin L, Muceli S, Merletti R. Fundamental Concepts of Bipolar and High-Density Surface EMG Understanding and Teaching for Clinical, Occupational, and Sport Applications: Origin, Detection, and Main Errors. Sensors. 2022; 22(11):4150. https://doi.org/10.3390/s22114150
Chicago/Turabian StyleCampanini, Isabella, Andrea Merlo, Catherine Disselhorst-Klug, Luca Mesin, Silvia Muceli, and Roberto Merletti. 2022. "Fundamental Concepts of Bipolar and High-Density Surface EMG Understanding and Teaching for Clinical, Occupational, and Sport Applications: Origin, Detection, and Main Errors" Sensors 22, no. 11: 4150. https://doi.org/10.3390/s22114150
APA StyleCampanini, I., Merlo, A., Disselhorst-Klug, C., Mesin, L., Muceli, S., & Merletti, R. (2022). Fundamental Concepts of Bipolar and High-Density Surface EMG Understanding and Teaching for Clinical, Occupational, and Sport Applications: Origin, Detection, and Main Errors. Sensors, 22(11), 4150. https://doi.org/10.3390/s22114150