Any Scan
Any Scan
Any Scan
IMAGING TABLE
SPECT
CT
PET
AnyScan
AnyScan SC
AnyScan PC
AnyScan S
AnyScan C
Upgrade path
Multiple AnyScan upgrade path are possible*
Add-on upgrade path for equipment that grows with the practice
-- Starts with a purchase of AnyScan S (SPECT)
-- With addition of a CT unit the system is upgraded to AnyScan SC (SPECT/CT)
-- Further addition of a PET unit upgrades the system to AnyScan SCP (SPECT/CT/PET) triple modality system and all NM
examinations can be performed in the same room
AnyScan SC
AnyScan S
AnyScan
AnyScan PC
AnyScan
AnyScan S
SPECT Components
Digital Detector
The two rectangular jumbo FOV high stability detectors are characterized with high optical performance and excellent
mechanical quality are available in two quality options:
-- The HP (High Performance) detector is designed to perform standard NM examinations
-- The UHP (Ultra High Performance) detector with Tera-Tomo 3D SPECT-Q quantitative reconstruction engine performs
absolute tracer uptake quantitation SPECT analysis
The thickness of the 585 x 470 mm NaI(TI) scintillation crystal is either 9.5 mm or 15.9 mm is also available in true rectangular option.
There are either 56/60 pcs of high quantum efficiency PMT characterized by improved energy resolution, magnetic shielding
and long-term stability.
The thickness of the lead shielding is 12-32 mm, covering the range of high energy isotopes.
Detector Electronics
The detector electronics are built on a compact, highly integrated, one-box, easily serviceable construction without tuning
potentiometers.
-------
SPECT Gantry
The gantry has a small footprint, it was designed with improved safety factor to ensure high stability.
-- 180, 101 or 90 degree variable angle head positions with high precision positioning
-- Industry standard CAN connected automation computer
-- Hand controller located on the gantry monitor, the optional
additional hand controller is located on the opposite
side of the gantry
-- Pre-programmed robotic gantry motions
-- Detector integrated infrared body contouring option
-- 90 degree detector positioning with online body
contouring for cardiac examination
-- Full automatic motion calibrations
-- Maintenance-free mechanical design
-- Patient Aperture: 75 cm
AnyScan S
NEMA Specifications
Detector
HP*
UHP**
UHP** Typical
Energy range:
40960 keV
9.7%
9.5%
8.9%
1.9%
1.4%
1.2%
2.4%
1.9%
1.2%
Integral CFOV
2.4%
1.9%
1.6%
Integral UFOV
2.9%
2.4%
1.6%
3.6 mm
3.1 mm
2.9 mm
CFOV (FWTM)
7.2 mm
6.1 mm
5.9 mm
UFOV (FWHM)
3.7 mm
3.2 mm
3.0 mm
UFOV (FWTM)
7.4 mm
6.2 mm
5.9 mm
Differential CFOV
0.18 mm
0.09 mm
0.03 mm
Differential UFOV
0.20 mm
0.10 mm
0.04 mm
Absolute CFOV
0.38 mm
0.28 mm
0.14 mm
Absolute UFOV
0.40 mm
0.30 mm
0.18 mm
7.4 / 7.6 mm
7.3 / 7.5 mm
6.9 / 7.3 mm
13.9 / 14.8 mm
13.4 / 14.5 mm
12.8 / 14.2 mm
180 cpm/Ci
9.9 mm
9.8 mm
9.7 mm
Radial
9.9 mm
9.8 mm
9.7 mm
Tangential
7.5 mm
7.4 mm
7.3 mm
SPECT Component
Collimators
A wide range of high quality collimators are available for the AnyScancamera. All collimators are made of high quality
micro-linear or micro-cast lead. The extremely precise level of manufacturing ensures excellent image quality.
Removable infrared auto body contour sensors, set on the collimator touch plates, ensure safety for the patient and improves
the reproduction quality of a repeated studies. Detector integrated infrared body contouring is available as option.
Hole
length
(mm)
Name
Hole
size
(mm)
Septal
Thickness
(mm)
Nominal
Energy
(keV)
Septal
Penetration
at Nominal
Energy
Sensitivity
(cpm/Ci) 7%
Geometric
System
Geometric
System
LEHS
(1)
26
2.5
0.30
140
2.1%
13.7
14.1
1000
840
(9)
LEGP
(2)
35
1.9
0.20
140
1.1%
8.2
8.9
340
290
(9)
LEHR
(3)
35
1.5
0.16
140
1.0%
6.5
7.4
210
180
(9)
LEUHR (4)
40
1.4
0.16
10
0.4%
5.4
6.5
140
110
(9)
MEGP
(5)
35
2.5
1.2
300
1.9%
11.1
11.6
330
220
(10)
HEGP
(6)
55
3.4
1.6
364
5.9%
11.6
12.2
340
90
(11)
LEFB
(7)
25
1.4
0.18
140
1.2%
6.4
7.3
340
300
(9)
7.2
7.3
70
(11)
10.2
10.2
140
(11)
13.2
13.2
230
(11)
4
HEPH
(8)
~200
6
8
Resolution at 10 cm
(FWHM mm)
364
Multislice CT Component
AnyScan system combines the power of nuclear medicine with the precision of 16/8 slices fast helical CT technology.
AnyScans CT module supplies precise data for the attenuation correction and provides excellent diagnostic imaging capability.
AnyScan family offers alternative choice of 16 or 8 slice diagnostic CT or 8 slice non diagnostic CT.
The CT module consists of a high resolution ceramic detector. The rotation speed is up to 0.4* sec with a 360-degree rotation.
CT-Gantry
-- Patient Aperture: 70 cm
-- Scan Field of View: 50 cm (70 cm for localization and corrections)
-- Acquisition Times (360): 0.4*, 0.5, 0.7, 1, 1.5 and 2 sec.
X-Ray Generator
-- Power: 60 kW for diagnostic or 10 kW for non-diagnostic CT
X-Ray Tube
-- Tube Voltage: 80 - 140 kVp
-- X-Ray Tube Current: 5 - 500 mA for diagnostic or 5 - 100 mA for non-diagnostic CT
-- Anode Heat storage capacity: 7.5 MHU
-- Automatic dose control (with ultra-low-dose & pediatric protocols)
X-Ray Detector
-- Ceramic Detectors
-- Number of Detector Elements: 24 x 896
-- Number of DAS Rows: 16 / 8 or 8
-- Detector Width (Z Dimension): 20 / 10 mm or 10 mm
Image Reconstruction
-- Reconstruction Matrix: 512 x 512 or 1024 x 1024
Scan Parameters
-- Acquired Slice Thickness: 16 rows of 0.625 mm / 1.25 mm or 8 rows of 1.25 mm
-- Displayed Slice Thickness: 0.625 (for diagnostic CT only), 1.25, 2.5, 5, 10 mm
*optional
AnyScan SC
PET Component
The PET module is carefully designed to integrate with the other modalities of AnyScan. The high sensitivity and excellent
resolution facilitate fast examination process and accurate clinical diagnosis.
-----------
10
Molecular PET imager module with high resolution LYSO detector technology
Independent gantry under common PET/CT gantry cover
70 cm gantry bore diameter
90.2 cm ring diameter
High precision digital imaging electronics
High spatial resolution
Excellent volume sensitivity
FBP-SSRB, 3DOSEM, 3DML and Tera-Tomo 3D PET reconstruction
Different FOV options for increased sensitivity
The selected FOV may be extended with a later purchase (Field upgradable)
Description
Unit
Basic field
Crystal
LYSO
Number of detectors
Crystal size
Extended Field
24
mm
3.9x3.9x20
Number of pixels
26448
39672
Number of PMT
288
432
mm
4.1
mm
4.9
mm
4.2
mm
5.1
Axial FOV
cm
Transaxial FOV
cm
System sensitivity
cps/kBq
15.2
23.0
50
5.8
9.1
Coincidence window
nsec
kcps
ps
100
Gantry aperture
cm
70
Ring diameter
cm
90.2
77
150
AnyScan PC
11
Shared Components
Triple Modality Imaging Table
The carbon fiber dual section imaging table can be extended extremely long, ensuring safe support for corpulent patients as well.
-----------
12
13
Software solutions
Nucline all modality acquisition software
Nucline is a fully configurable acquisition software framework, which offers standardized acquisition workflows for all
modalities in one common platform. Due to the common software framework Nucline provides fully integrated true hybrid
imaging solutions for all camera products of Mediso.
The new advanced 3D Geometric Planner of Nucline allows operators to use any kind of already acquired images as a
reference for the next acquisition in a smooth and intuitive visualization framework.
The human and pre-clinical acquisition line is fully covered by Nucline, which makes the whole system standardized and
intuitive. This way Nucline provides easy operation for users working in the field of both clinical and pre-clinical routine.
14
(b)
Unprocessed 15
min. scan
(c)
Tera-Tomo 2D Planar
7.5 min. scan
(d)
Unprocessed
15 min. scan
(a)
Tera-Tomo 3D SPECT Bone Enhancement
(512x512 volume rendering),
(b)
Standard OSEM
(128x128 volume rendering),
(a)
Tera-Tomo 3D SPECT Bone Enhancement
(512x512 volume rendering),
(b)
Standard OSEM
(128x128 volume rendering),
Comparison of Tera-Tomo 3D SPECT Bone Enhancement and Standard OSEM reconstruction (patient study, 608 MBq,
Images acquired by AnyScan SC Research Edition)
15
Software solutions
Tera-Tomo 3D CT Reconstruction Module
The Tera-Tomo 3D CT reconstruction engine implements a novel multi-material reconstruction method, which produces
optimal patterns for all specific tissues in the body. In addition, the engine also handles metal artefacts, since they are also
handled as separate materials during reconstruction. The engine contains an advanced streak reduction as well as a novel
edge-preserving noise compensation module. The Tera-Tomo 3D CT reconstruction is operated on multiple GPUs (Graphics
Processing Units) for optimal performance [2,3,4]. The reconstruction engine, supplied with all the above solutions, allows
ultra-low-dose CT acquisitions with excellent image quality without compromises, meaning radical dose reduction for patients.
16
17
Software solutions
The nuclear studies are right there. Anytime. From Anywhere.
The software tools that drive the solutions...
18
Left: 3D VR of a CT with surface cut plane. Middle: 3D VR of a 18F-FDG PET-CT with non-linear alpha functions
Right: 3D VR of the same PET-CT with different alpha functions and palettes.
Automated quadruple registration of a primary and interim 18F-FDG PET-CT image pairs
Tc MDP Whole Body bone SPECT AP - PA acquititions and their corresponding reconstructed SPECT/CT fusion with
spatially aligned synchronized cursors
99m
19
Cedars Sinai
Dedicated evaluation package to analyze quantitative blood-pool SPECT, cardiac perfusion and gated SPECT and PET studies and
compare to normal databases.
Cedars Sinai Quantitative Gated SPECT
-- Automated processing of myocardial perfusion SPECT and gated SPECT data
-- Left ventricular ejection calculation
-- 3D beating image displays
-- Volume curves and polar maps
Documentation
Automated bi-level macro-controlled printing and reporting. High quality inkjet colour and b/w hardcopy
-- On normal paper
-- On premium photo paper
-- 2400 dpi print quality
-- Special printing software for faithful printing
20
CT Software
InterView CT Expert
3D-solution for processing and analysis of medical images
-- Optimised patient CD management
-- Standard and cine mode display
-- Key images selection
-- DICOM and Windows printing
-- MPR
-- Axial / Coronal / Sagittal reconstructions
-- Oblique
-- Thick Slab creation (average)
-- Thick Slice (MIP/MinIP)
-- New MPR series creation
-- Parallel, curved or radial
-- Wide range of toolboxes
-- Dynamic color palette
-- Region of interest 2D-editing tool
-- Whole volume 3D
-- MIP
-- Volume rendering
-- Surface rendering
-- Conversion of digital images into DICOM format
-- Exam comparison
-- Multimedia report creation
-- Visual selections and automated bone masking
21
Dedicated CT Software
InterView CT Dedicated Analysis Suite
Includes the following modules:
InterView
InterView
InterView
InterView
InterView
CT Cardiac CT Analysis
CT Calcium Score Analysis
CT Cardiac Cath Analysis
CT 3D Viewer includes cardiovascular analysis capabilities
CT CT Colonoscopy Analysis
Automatic segmentation
Calcium Score calculations
Identifies the different arteries
Scores the calcified plaque areas
Agaston score and plaque volume indication
Total plaque burden calculation
Workflow to Approve findings
22
Auto Calcium Scoring
23
Automatic Colon Segmentation and Endoscopy view
Special Options
AnyScan S SINGLE single head SPECT camera
-----
AnyScan S SINGLE
AnyScan SC
-------
24
CT FIRST
AnyScan SC CT FIRST
The AnyScan SC SPECT/CT system combines SPECT and CT modalities in a single gantry placed in a common gantry
housing under the control of a single NuclineTM workstation, providing an optimal workflow solution for SPECT/CT
multimodal applications. Precise matching of the two modalities is ensured by an accurate alignment method.
An advanced reconstruction software is responsible for the accuracy of data, gained from the multimodal procedure.
25
SPECT Gantry
CT Gantry
PET Gantry
SPECT/CT/PET Imaging Table
Multimodality acquisition console
Power Distribution Unit
Dimensions, weights
Components
Length (mm)
Width (mm)
Height (mm)
Weight (kg)
SPECT gantry
1139
1714
1632
2100
CT gantry
778
2116
1916
1350
PET gantry
400
2116
1948
400
Imaging Table
2580
670
850
330
Collimators
685
562
46-68
70-125
Collimator Carts
750
560
1250
70
1000
600
1490
450
Optional Accessories
-----------------
26
AnyScan S
with SPECT-only imaging table
Room Layout
CT-component
-- Mains Voltage : 400/480 V (3 phase)
-- Mains frequency: 45-63 Hz
-- Max. power: 65 kVA (diagnostic CT)
10 kVA (non-diagnostic CT)
PET-component
-- Mains Voltage: 115/230 V (1 phase)
-- Mains frequency: 45-63 Hz
-- Max. power: 2.2 kVA
Environmental Requirements
Constant room temperature and humidity rate required for 7 days / 24 hours
------
27
References
1: PCT/HU2012/000066
2: Gbor Jakab, Tams Huszr, Balzs Csbfalvi: Iterative CT Reconstruction on the GPU. In: VI. GRAFGEO Conference. Budapest,
Hungary, 02.21. 2012-02.22. 2012. pp. 124-131.
3: Jakab Gbor, Rcz Attila, Bkki Tams, Nmeth Gbor: Fully GPU Based Real Time Corrections and Reconstruction for Cone Beam
Micro CT. In: IEEE - High Performance Medical Imaging (HPMI): Science Symposium Conference Record (NSS/MIC). Orlando, USA,
2009.10.24-2009.11.01. Orlando: IEEE, pp. 4068-4071.
4: Gabor Jakab, Laszlo Szirmay-Kalos: Hybrid Monte Carlo CT Simulation on GPU. In: Lecture Notes of Computer Science (LSSC13),
2013.
Conformance Statement
Quality management system operated by Mediso Medical Imaging Systems complies with Council Directive 93/42/EEC Annex II. The
multimodality molecular imaging system was approved by a Notified Body.
Product design, development, production and services comply with EN ISO 13485 and EN ISO 14971. Medical device design and safety
testing has been performed in accordance with EN IEC 60601-1 and EN IEC 60601-1-2 standards.
Safety labels are attached to appropriate places on equipment and appear in all operation manuals.
The supplied software complies with DICOM standard.
The technical information provided here is not a detailed specification.
For details and up to date information please contact your local distributor or Mediso Medical Imaging Systems.
Trademarks:
Emory Cardiac Toolbox is the trademark of Emory University, Atlanta, GA.
Cedars Quantitative Gated SPECT, Cedars Perfusion SPECT, Cedars
Quantitative Bloodpool SPECT are trademarks of Cedars Sinai University
Medical Center, Los Angeles, California.
InterView FUSION, InterView XP, InterView CT
AnyScanand Tera-Tomo are trademarks of MEDISO Medical Imaging Systems.
MEDISO USA
E-mail: sales@medisousa.com
Web: www.medisousa.com
MEDISO GmbH
E-mail: info@mediso.de
Web: www.mediso.de
MEDISO reserves the right to change data without notice MEDISO 2014.