Optima 520 - Datasheet New
Optima 520 - Datasheet New
Optima 520 - Datasheet New
Personalized CT solution
Optima* CT520
Product Data Sheet
Adaptive statistical iterative reconstruction (ASiR) is GE’s industry advanced technology. Leading the way
with image reconstruction, the ASiR technique enables reduction in image noise and improvement in image
quality, low contrast detectability and contrast resolution.
The Optima CT520 scanner with the ASiR reconstruction algorithm demonstrates the same image standard
deviation performance at 40% lower dose.
Utilizing ASiR, images obtained can have equivalent IQ to an acquisition with 1.67 times the mA and Kw
which means 70kW, 583mA.
1. In clinical practice, the use of ASiR may reduce CT patient dose depending on the clinical task, patient size, anatomical location and clinical
practice. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality
for the particular clinical task.
ASiR is licensed for use with a GE x-ray tube. Use of a third-party x-ray tube will require an additional license for these features.
Volumetric Image Space Reconstruction (VISR2) provides a 3D Neuro filter that reduces noise without
compromising resolution, for clear visualization of brain, tumor, and pediatric cases. With the VISR 3D filter,
the scanner delivers up to 20% image quality improvement at the same dose, or the same image quality
with up to 36% dose reduction.
2. In clinical practice, the use of VISR may reduce CT patient dose depending on the clinical task, patient size, anatomical location and clinical
practice. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality
for the particular clinical task.
When ASiR is installed, 3D Neuro filter will be disabled.
Opti-Dose
Dose Check Provides the user tools to guide dose given in clinical practice and is based on the standard
XR-25-2010 published by the Association of Electrical and Medical Imaging Equipment
Manufacturers (NEMA).
Dose Check provides the following
• Checking against a Notification Value if the estimated dose for the scan is above your
site typical dose value
• Checking against an Alert Value where the user needs specific authority to continue the
scan at the current estimated dose without changing the scan parameters.
• Defining Alert Values for Adult and Pediatric with age threshold
• Audit logging and review
• Protocol Change Control
DICOM Dose In conjunction with Prospective display of CTDIvol, DLP and Dose Efficiency, helping
reporting clinicians reach ALARA dose, and keep track of it.
Dose computation CTDIvol (CTDI volume), DLP (Dose Length Product), and Dose Efficiency computation and
and display display during scan prescription provide patient dose information to the operator.
Color coding for kids’ Providing pediatric scan protocols based on the Broselow-LutenTM Pediatric System,
protocols designed to facilitate pediatric emergency care and reduce medical errors.
3D Dose modulation Before the scan, clinicians can select the desired Noise/IQ: CT then tailored automatically
exposure parameters, patient to patient and real-time x-y-z during each scan, resulting in
average of 40% dose reduction
Beam tracking Providing real-time X-ray follow-up, guarantying that high spatial resolution is reached with
techniques no post-patient collimation and no dose penalty.
Short gantry In conjunction with Hyper generator and the Solarix X-Ray Tube affords to get up to 350 mA
geometry and seamless throughput.
IQ Enhancement (IQE) may reduce helical artifacts which are important for
image quality of thin-slice helical scans. The Optima CT520 series scanner
with this feature can accelerate its helical pitch up to 70% (e.g. 0.562 to 1.75,
@16slice) when acquiring the same helical artifact level compared with the
same scanner with IQ Enhance disabled. This coverage speed is equivalent
to that of a 50-slice wider detector CT scanner at same table speed.
Optima CT520 series has a dedicated User Interface (UIF) for emergency
cases to start examination quickly. Patient Name and Patient ID are
assigned automatically. Once a protocol is selected, scan setup interface
displays.
With its Energy Saving Mode activated, GE's Optima CT520 Series scanner is designed to reduce electricity
consumption for operation and ambient cooling by more than 12,085 kWh per machine annually, an energy
savings up to 40% compared to prior GE technology.
Direct MPR Direct MPR with Auto-Batch feature, affording automatic real-time
direct reconstruction and transfer of fully corrected multi-planar
images, also allows customer to move from routine 2D review to
prospective 3D image review of axial, sagittal, coronal, and oblique
planes while enabling automated protocol-driven batch reformats
to be created and networked to their desired reading location.
Series Split Series Split provides to split multi groups prescribed in a View Edit
series to separated series by groups. This capability allows the
workflow improvement to diagnose an examination to compare
groups taken with same location.
Enhanced filming workflow The new filming tool provides you a friendly and powerful user
interface. It allows you to open multiple sessions at the same time
for one or more patients.
SmartPrep with Auto Trigger SmartPrep allows intermittent monitoring of IV contrast
enhancement in an area of interest. The contrast flow is
monitored by Low-Dose scans until the contrast enhancement
reaches the preferred point and then auto trigger function will
automatically initiate the scan prescription.
3-click Scan start The pre-programmable protocol setting functions enable a
minimum of 3 clicks to start a scan as minimum.
Direct connect Direct Connect allows remote Advantage Workstation (AW)
access to the Xtream FX console’s thin-slice data. (AW4.3* and
later)
Graphic Retro Graphic Retro provides the capability to graphically prescribe your
reconstructions using an existing image.
Auto transfer Auto transfer by Series to distribute images where you need them
when you need them.
Gantry
Aperture 70 cm
Tilt ± 30°
Maximum SFOV 50 cm
Rotational Speeds 360° in 0.8, 1.0, 1.5, 2.0, 3.0, 4.0 sec
Laser Alignment Lights Define both internal and external scan planes to ± 1 mm accuracy.
Operate over full range of gantry tilt; activated any time during exam (with
tube stationary).
Coronal light remains perpendicular to axial light as gantry tilts.
Patient Table
Horizontal Scan range Axial: up to 1600 mm; Scout: up to 1460 mm; Helical: up to 1450 mm
ISO -25 mm
IV Pole* integrated at the foot-end of the table helps prevent IV lines from becoming
crossed and tangled, and helps ensure that the lines stay securely in place
on the patient
Table automatically re-centers on scan plane with changes in vertical position (after
setting internal landmark with alignment lights on)
X-Ray Tube
Use of ASiR allows scanning at a lower mA and less tube heat output,
resulting in the ability for longer duration helical scans similar to the
capability of a 6.3 MHU tube.
Generator
Max generator power Max Power with 120 kV on Optima CT520 system: 70kW equivalent with
ASiR
Imaging System
Host Computer
• Dual Intel Xeon 4215 CPU 2.5 GHz 8 Core
• 64 GB as 4x16 GB, DDR4 16 GB ECC 288-Pin Registered DIMM 2666 MHz or higher frequency
• 1TB SATA 7200 RPM HDD*1, 512 GB SATA SSD*1, 512 GB M.2 TLC SSD*1
Image Processor
• NVIDIA Quadro P620 Graphics Card
• 4 mDP Outputs
• 2 GB GDDR5, 2000 MHz, Memory Interface: 128-bit, Memory Bandwidth: 64 GB/s
• GP107 GPU
• 512 CUDA cores
• PCI Express 3.0 x16
• up to 4x 5120 x 2880 x 24 bpp @ 60Hz
Hard disk
- SAS interface
- Assigned to system and application software only
Software Architecture
Software architecture based on industry standards and client-server design
• Peripherals:
- DVD-R/CD-R (DICOM Interchange)*:
- 4.7 GB capacity (DVD)
- Up to 7,168 image storage (DVD)
- 20x DVD, 48x CD Max
- Supports CD-R, DVD-R
• USB:
- 64 GB
- Assigned to scan data file and protocol file storage/retrieval
• Color monitors:
- 19 inch diagonal width
- 1280 x 1024 dot resolution
- Non-interlaced flicker-free presentation
- 76 kHz Horizontal deflection frequency
- 72 Hz Vertical deflection frequency
• Scan control keyboard (English language) assembly with intercom speaker, microphone and volume controls
• 2-Button Scroll Mouse
*Option
Anti-Virus software
Standard McAFee Anti-Virus software
Image Networking
• Exam transfer 10 frames per second on dedicated 1 GB connection
• Standard auto-configuring 100 BaseT/10 BaseT Ethernet (UTP connection)
• Supports gigabit ethernet capability
• Direct network connection; multi-suite ethernet card not required for gateway out of suite
• Protocols supported:
- DICOM 3.0 network send (one IP address at a time) and receive, pull/query, and storage
commitment push
- AdvantageNet (GenesisNet) point-to-point send, receive and pull/query (no broadcast)
- InSite® point-to-point
- TCP/IP (for system administration)
Compatible options
• Options on CT Console
- A DICOM Print Interface is standard on the system - Euro Misc USB Keyboard
- English UIF & USB Keyboard - Finnish UIF
- Asian USB Keyboard - Japanese UIF
- Chinese UIF & USB Keyboard - ConnectPro HIS/RIS Interface with Performed
- French UIF & USB Keyboard Procedure Step (PPS) M4 compatibility
- German UIF& USB Keyboard - USB Bar Code Reader
- Scandinavian UIF & USB Keyboard - Exam Split
- Danish UIF & USB Keyboard - Xtream™ injector/Enhanced Xtream Injector
- Dutch UIF & USB Keyboard - SmartStep
- Italian UIF & USB Keyboard - Neuro 3D Filter
- Norwegian UIF & USB Keyboard - ASiR™
- Spanish UIF & USB Keyboard - LCD Monitor
- Swedish UIF & USB Keyboard - GE Color Printer
- Portuguese, Braz UIF & USB Keyboard - Table Tray
‡Option
With Optima CT520 being a sub-millimeter isotropic CT scanner, it is now possible to specify coronal and sagittal
image quality. Reformatted resolution (0.35 mm+/-0.05 mm) is visually demonstrated on the Catphan High
Contrast Resolution Insert Module CTP528.
50% 4.0
10% 6.5
0% 8.5
50% 8.5
10% 13.0
0% 15.4
Detail Kernel
50% 5.92
10% 10.98
4% 12.78
Visual resolution Z
0.31 mm+/-0.02 mm
17lp/ cm 16lp/ cm 15lp/ cm 14lp/ cm 13lp/ cm
Low-Contrast Detectability
Noise
=< 0.35% at 11.8 mGy CTDIvol with ASiR (ASiR 40% enabled)
Helical
=< 0.35% at 22.2 mGy CTDIvol when ASiR is disabled
=< 0.35% at 26.5 mGy CTDIvol with ASiR (ASiR 40% enabled)
Axial
=< 0.35% at 44.95 mGy CTDIvol when ASiR is disabled
CTDIw (Helical)
CTDIw (Axial)
Multiple Acquisition Maximum Scan Multiple scans can be acquired in one series to produce up to 3000
Time contiguous helical images. Up to 2000 rotation helical coverage are possible
in multiple series.
Scan Fields of View 25 cm for adult head; 25, 50 cm for body; 25 cm for pediatric head
AutoScan™ Fully automates longitudinal table movement and start of each scan.
Trauma Patient Allows patient scans and image display/analysis without entering patient
data before scanning.
Biopsy Simplified prescription for single or multiple scans around an arbitrary table
position aids biopsy studies.
10 sec 350
20 310-350
30 260-335
40 225-260/295
50 200-260/265
60 175-240
70 145-215
80 135-200
90 120-185
100 110-175
110 100-165
120 90-140
4 235-305
5 210-250/270
6 185-240
3 180-235
4 135-195
3 120-185
4 95-140
Helical scan: 16x0.625 mm, 1.375 pitch, 0.1 mm image interval, the image number (slice number) vs different
rotation cycles, shows in below table
2 13.5 136
3 27.2 273
4 41 411
5 54.7 548
6 68.5 686
Generating images at fine intervals, as small as 0.1 mm, enables reconstructed images that exceed 64 images per
gantry rotation. The number of images able to be generated per gantry rotation is a function of rotations and
coverage.
Orientation AP, RLAT, PA, LLAT (preset); or any angle from 0° - 359° with 1° inclement
(manually selected)
Scan Time 0.8, 1.0, 1.5, 2.0, 3.0, 4.0 sec full scans (360° acquisition)
Intergroup Delay (IGD) Minimum IGD is the same as minimum ISD, also user selectable.
Minimum scan-to-scan cycle of 1.8 sec possible for 0.8 sec scan speed with
Scan-to-Scan Cycle
minimum ISD’s.
Overlapped Reconstruction* The system needs to have the overlapped reconstruction feature, which
enables 32 slices per rotation in axial scanning modes and delivers
improved Z-axis visualization performance relative to non-overlapped
reconstruction.
Reconstruction Algorithms Soft, Standard, Detail, Bone, Bone Plus, Lung, Edge and chest
Queued Recon Requests will be processed continuously and simultaneously with other
processes on the system including scanning. Prospective recon will be
prioritized over retrospective recon.
Priority Recon Queuing One touch selection marks most recent rotation for next available recon.
Available during or after scanning.
*Option
4 commercial availability depends on each country's regulatory approval
Application
Volume Viewer Volume Viewer is an innovative and powerful suite of productivity enhancers (Volume
Rendering, Volume Analysis and Navigator) that includes:
• Dynamic Volume Review™ for Fast Screening
• Curved Volume of Interest
• Protocol Management and Loading
• Review Layout Presets
• Multiple VR Objects Merge
• Pseudo Surface Shading Mode
• Predefined Cut Planes
• Volume Rendered Navigator views
• VR Preset save/recall
• 3D Rendered Lumen View
• Automatic Path Tracking
• Path Bridging (in case of occlusions)
• SmartCursor™ for Easy Navigation
• Synchronized Reformatted Views
• Cut visualization mode
Advanced Vessel Advanced Vessel Analysis† is the ultimate tool to assess and quantify vascular
Analysis structures, including stenosis analysis, stent planning procedures, post stenting or
vascular surgery follow-up.
Protocol driven tools to perform quick, flexible and accurate quantitative analysis of
vascular anatomy.
Provides maximum, minimum and mean intraluminal diameter measurements
Provides cross-sectional areas of true orthogonal sections of the aortoiliac systems at
selected anatomical points
Clinical benefits include stenosis sizing, pre- and post- surgical assessment, stent
planning
Measurements in % stenosis or mm of stenosis, and measurement of length and
dimension of stenosis.
Interventional
SmartView Offers real-time 12 frames per second CT fluoroscopy with image latency of less than about
Fluoro 200 ms. Detailed targeting is supported by multiple acquisition modes, up to and including
high-resolution acquisition at 1.25 mm (16x0.625 mode only). High dose efficiency is made
available with continuous and quick check scan modes.
A simple and efficient user interface provides six user-selectable display layouts, in-room
image review and WW and WL control Flip, rotate, roam and zoom capabilities maintained
during acquisition. Feature ceiling-mounted in-room LCD monitor and full-featured handheld /
cradle-mounted controller.
Smart Step SmartStep enables a step and shoot-imaging mode for performing biopsies and other
interventional procedures. A Highly Functional Image Display presents a set of 3 interventional
images in 3 viewports.
CT Perfusion 4 CT Perfusion Neuro allows the user to process dynamic image data of the brain & generate
Neuro information with regard to changes in image intensity over time.
Components
Power supply
Main disconnect to be located within 32 feet (10 m) of the PDU. The facility must also
provide a protective disconnect device with low voltage, low energy
local and multi-point remote capability, in the line feeders to the
PDU.
Cooling Requirement
Recommended cooling values should not be used for calculating system input power requirements.
*** Recommended Allowance: 7,200/24,566
Growth Allowance: 9,825/33,520
Minimum Allowance
Suite Size
Room Environment
Warranty
The published Company warranty in effect on the date of shipment shall apply. The Company reserves the
right to make changes.
All specifications are subject to change.
Regulatory Compliance
This product is designed to comply with applicable standards under the
Radiation Control for Health and Safety Act of 1968.
Laser alignment devices contained within this product are appropriately
labeled according to the requirements of the Center for Devices and
Radiological Health.
This product is a certified device which satisfies regulations regarding Electro-Magnetic Compatibility (EMC)
and Electro-Magnetic Interference (EMI), pursuant to IEC-60601-1-2
*Option
Disclaimer:
The Optima CT520 system is not validated with any Flat Table Top (FTT) for use in providing image data for
RADIOTHERAPY TREATMENT PLANNING (RTP), additionally GEHC does not offer any FTT to be optionally sold for RTP.
About GE Healthcare
GE Healthcare provides transformational medical technologies and services that are
shaping a new age of patient care. Our broad expertise in medical imaging and
information technologies, medical diagnostics, patient monitoring systems, drug
discovery, biopharmaceutical manufacturing technologies, performance improvement
and performance solutions services helps our customers to deliver better care to more
people around the world at a lower cost. In addition, we partner with healthcare leaders,
striving to leverage the global policy change necessary to implement a successful shift to
sustainable healthcare systems.
Our “healthymagination” vision for the future invites the world to join us on our journey
as we continuously develop innovations focused on reducing costs, increasing access, and
improving quality around the world. Headquartered in the United Kingdom, GE
Healthcare is a unit of General Electric Company (NYSE: GE). Worldwide, GE Healthcare
employees are committed to serving healthcare professionals and their patients in more
than 100 countries. For more information about GE Healthcare, visit our website at
www.gehealthcare.com.
GE Healthcare
9900 Innovation Drive
Wauwatosa, WI 53226 U.S.A.
Chalfont St. Giles Buckinghamshire UK
www.gehealthcare.com