Nothing Special   »   [go: up one dir, main page]

GEIT 20032EN - Mic Probes PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

GE Inspection Technologies

UCI Probes
for MIC 10 and MIC 20 hardness testers.

Key Features
quick and easy hardness testing standardized according to ASTM A 1038 on-site and in any direction versatile probes for extensive application area

Typical Applications
heat-treated parts and surfaces, especially at difcult to access positions ready-assembled machines heat-affected zone of a weld thin coatings or layers

The UCI test method


The hardness testers of the MICRODUR family are characterized by their ease of use and extensive application possibilities. Hardness testing according to the UCI method enables quick and easy evaluation of on-site hardness indents without having to use a microscope for indentation evaluation: the probe is simply positioned and the reading is digitally displayed. Using these small and handy probes you can carry out your measurements anywhere and in any direction: whether it be vertical or overhead there are no limitations.

Standard Probes
These manual UCI probes are available with test loads from 10N to 98N and suitable for all standard applications, especially for hardness tests on heat-treated materials.

Probes equipped with a long oscillation rod


The probes MIC 201-L and MIC 205-L are not only suitable for all standard applications but, due to their long oscillation rod, also for measurements in deep cavities on test objects geometries.

Motorized Probes
Using a motor the test load for the 1N, 3N or 8N probe is applied automatically. Therefore, the motorized probes are particularity suited for hardness testing on finished precision parts, thin coatings or layers e.g. rotogravure cylinders.

Test method
With the standardized UCI method (ASTM A1038) the size of the hardness test indentation is detected electronically by determining the shift of ultrasonic frequency under load. This shift is proportional to the size of the indentation produced by the Vickers diamond in the material.

We have the right probe for you.


Your application determines the type of probe to be used. We have developed a whole range of probes with different test loads covering an extensive application area: standard probes, motor probes and probes equipped with a long oscillation rod. We can offer you the most suitable probe for various test object surfaces and geometries.

Hardness testing instruments


The UCI probes are suitable for operation with all UCI instruments: MIC 10, MIC 10DL, MIC 10R/RS MIC 20, MIC 20TFT

If you require additional information please visit our website www.ge.com/inspectiontechnologies

UCI probe models:


Probe model
MIC 201-A MIC 205-A MIC 2010-A MIC 201-AL MIC 205-AL MIC 2101-A MIC 2103-A MIC 211-A

Type
standard standard standard long rod long rod motorized motorized motorized

Load (N)
10 50 98 10 50 1 3 8

Diameter (mm)
19 19 19 19 19 40 40 40

Length (mm)
143 143 143 169 169 190 190 190

Tip diameter (mm)


2.4 3 3 3 3 3.7 3.7 3.7

www.ge.com/inspectiontechnologies

2006 General Electric Company. All Rights Reserved. We reserve the right to technical modications without prior notice. GEIT-20032EN (05/06)