Radiographic Examination: Standard Practice For
Radiographic Examination: Standard Practice For
Radiographic Examination: Standard Practice For
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
A~u11 7
Designation: E1742/E1742M-23
INTERNATIONAL
This standard has been approved for use by agencies of the U.S. Department of Defense.
approved in 1995. Last previous edition approved in 2018 as El742/E1742M- 18. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
DOI: 10.1520/El 742_El 742M-23. Standards volume information, refer to the standard's Document Summary page on
2 the ASTM website.
This practice replaced MIL-STD-453.
5.4.8 Luminance/illuminance light meters are procured and tions need not be re-verified as long as those conditions are maintained.
6.1.1 A drawing, sketch, or photograph of the component 6.4 Radiographic Identification-Each radiograph shall
showing the location of the film and IQI with respect to the carry the identification or serial number of the component and
radiation source for each exposure. view number, when multiple views are taken. Each radiograph
6.1.2 The angle of the radiation beam in relation to the shall also carry the identification of the NDT facility examining
component, the source-to-film distance, and any blocking or the component and the date of the examination. Radiographs of
masking, if used. a repair area shall be identified with Rl, R2, R3, and so forth,
6.1.3 Part zones, if applicable, should be included (see 6.2). indicating the number of times that repairs were attempted.
This may be accomplished through drawings and tables or by Alternative schemes may be used for identification of repair
reference to documents where such information is found. radiographs so long as each film is clearly identified to relate to
6.1.4 The nominal exposure for X-ray machines, the a particular repair area. For explosives and propellants, the
voltage, milliamps, time (or rads as applicable), and effective conditioning temperature shall be identified on each X-ray film
focal spot size. For radioisotope sources, the isotope type, if the ordnance has been conditioned to a temperature other
source strength (curies), exposure time, and source size. than facility ambient for purposes of examination.
6.1.5 Film designation (for example, brand, type, and pro-
cessing parameters), intensifying screens (for example, type 6.5 Examination and Coverage-The number of parts
and thickness of screens), or filters (for example, filter material, examined, and the radiographic coverage of each part shall be
thickness, and location) if used, film loading instructions (for as specified by drawings, radiographic techniques, radio-
example, when using multiple film exposure techniques), and graphic manuals, handbooks for aircraft technical orders, or
the desired optical density range. other specifications, as applicable. Areas to be examined shall
6.1.6 Thickness and type of material. be identified on the drawing by using the symbols in accor-
6.1. 7 The IQI size and type, and the required radiographic dance with AWS A2.4 or other systems of designations that are
quality level. If alternate IQis are used (see 5.5.3), include easily identified on the drawing. If the number of parts to be
details of the design or reference to documents where such examined and the amount of coverage of each part is not
information is found. specified, all parts shall be examined and shall receive 100 %
6.1.8 Thickness and type of material for shims or blocks, or radiographic coverage.
both, if used. 6.6 Examination Sequence-The sequence for radiographic
6.1.9 Name and address of the NDT facility and the date, or examination in the production operation should be specified in
revision, of the procedure. the manufacturing or assembly process specification, contract,
6.1.10 Radiographic identification scheme used to correlate or purchase order. If not specified, radiographic examination
part-to-film. If the examination procedures are similar for shall be performed at a stage in the process of manufacturing
many components, a master written procedure may be used or assembly at which discontinuities can be detected. Radio-
that covers the details common to a variety of components. All graphic examination may be performed before heat treatment,
written procedures shall be approved by an individual qualified provided liquid penetrant or magnetic particle examinations are
and certified as a Level ill for radiography in accordance with performed after heat treatment.
5.1. 1.
6.7 Nonfilm Techniques-When permitted by purchase
6.2 Acceptance Requirements-When examination is per- order, contract, or specification, radioscopic/radiological ex-
formed in accordance with this practice, engineering drawings, aminations using nonfilm techniques shall be in accordance
specifications, or other applicable documents shall indicate the with Practices E 1255, E2033 or E2698 or a nonfilm specifica-
criteria by which the components are judged acceptable. tion approved by the cognizant engineering organization as
Complex components may be divided into zones and separate required. Prior approval shall be obtained from the Level ill
criteria assigned to each zone in accordance with its design radiographer of the cognizant engineering organization (see
requirements. When used, direct references to ASTM reference 5.1.1 ).
radiographic standards shall include the grade level for each
type of discontinuity permitted for each part or zone. 6.8 Multi-Film Techniques-Film techniques with two or
more films of the same or different speeds in the same film
Norn 3-lnformation on reference radiographs can be obtained from
the Annual Book of ASTM Standards, Vol 03.03 or from ASTM Head-
holder, to be used in either single or superimposed film
quarters. viewing, shall be permitted provided that the applicable radio-
graphic quality level, and optical density requirements (see 6.9
6.3 Surface Preparation-Components may be examined
and 6.10), are achieved for the area of interest.
without surface preparation or conditioning except as required
to remove surface conditions that may interfere with proper 6.9 Radiographic Quality Levels-The five quality levels
interpretation of radiographs. listed in Table 3 may be assigned on the basis of IQI thickness
6.3.1 Castings, forgings, and weldments may be radio- and the perceptibility of one, two, or three holes in the
graphed in the as-cast, as-forged, or as-welded conditions hole-type IQI image on the radiograph. If the quality level is
provided the following requirements are met. not specified on the drawing or other applicable documents, it
6.3.1.1 For castings and forgings, the surface condition shall shall be Level 2-2T. Unless otherwise specified by the cogni-
not interfere with evaluation. zant engineering organization, hole-type IQis used for exami-
6.3.1.2 Accessible surfaces of welds shall be prepared in nation of material 0.25 in. [6.35 mm] or less in thickness shall
accordance with A2. l . be 0.005 in. [0.127 mm] minimum thickness.
1000000
~.,,,,.
.
100000 i.,, ~ "
....:::s ...
~
- •
....:::sa. .
,,, ..
i.illl""' ~
.
0 , t....11111""'"
CD 10000
...
~
""' -
·s:I ~ -- i...iill"
.5 ,,,, ~" ....
,,. .
it
,,,,.~ ....
<II
1000 - . -
--
,,,,. ....
- .... ....
~
_, -
#
-
100 ... .....
1 1.5 2 2.5 3 3.5 4 4.5
Maximum allowable optical density
- - - candela/m 2
- footlamberts
NoTE 1-This figure is a depiction of the abscissa axis: Maximum Allowable Optical Density versus ordinate axis: Candela/m2 and footlamberts in
graphical format from tabular data derived from Specification E1390 and ISO 5580 (also known as EN 25580). Conversion from tabular data to a graph
accounts for the step in the line. For Film Viewer Output of 10 000 candela/m 2 (2919 Footlamberts), the Maximum Allowable Optical Density shall be
3.0. Regarding the ordinate axis: Candela/m 2 , the minimum luminance level required for the average human eye to achieve photopic eye response (that
is where maximum resolution and contrast discrimination occurs) is at 10 candela/m2 • At levels below this value the eye responds scotopically which
means lower contrast discrimination and resolution. While photopic vision typically occurs at a threshold of 10 candela/m2 for the average human eye,
this curve takes advantage of the fact that at lower optical densities most viewers can achieve an amount of light that guarantees that virtually all operators
will be viewing film in the photopic vision mode, that is 30 candela/m2 for optical densities <2.5. A theoretical advantage of this curve is that it
compensates for the reduced contrast sensitivity of radiographic film at lower optical densities.
NoTE 2-NDT film systems, classified corresponding to Test Method E 18 I 5 system classes "Special, I and II," with or without lead screens, are suitable
for the extended viewing range above an optical density of 4, due to their high gradient (Go.00 = 4 >6) at D 4 above fog and base. These double sided =
NOT film systems have a high silver content and do not saturate as early as medical and classes Ill, W-A, W-B and W-C film systems. The basic advantage
of increasing the optical density is the increase of contrast with optical density. Since the contrast/noise ratio also increases (with the square root of optical
density), the perception of indications of small flaws improves significantly with higher optical density. The operator should mask all film areas of lower
optical density to avoid blinding (dazzling). Blinding reduces the eye perception and requires longer eye adaptation time. High brightness viewing stations
also heat up films depending on the optical density and viewing time. The operator shall prevent overheating to protect the film integrity.
NOTE 4--Letter (b) applies to welds where the IQI is placed on the part; 6.21 Contrast-The contrast of the radiograph shall be
(b) does not apply to welds when blocks are used. determined by measuring the difference in optical density of
6.17.1 Radiograph Qualification Using Two IQ/s-When the film through the IQI and the adjacent material. The
the optical density varies by more than is specified in 6.17, two minimum optical density difference shown in Fig. 2 shall be
IQis used in the following manner are acceptable. If one IQI achieved between the IQI and the base metals for Radiographic
shows an acceptable sensitivity in the most optically dense Quality Levels 1 and 2.
portion of the radiograph, and the second IQI shows an 6.22 Back Scatter Radiation-During each exposure the
acceptable sensitivity in the least optically dense portion of the film shall be monitored for back scatter. Each film holder shall
radiograph, the two IQis shall serve to qualify the radiograph have a lead letter "B" a minimum of0.5 in. (12.7 mm) high and
within these optical density limits plus the +30 %/-15 % of a minimum of 0.0625 in. ( 1.6 mm) thick positioned behind the
6.17. Additional IQis may be used, as necessary in subsequent film and within the general area of the film to be viewed.
exposures, to cover the entire thickness range of the object. For Should the image of the lead letter "B" appear on the
components such as castings and forgings, where there are radiograph as a light image, the radiograph shall be considered
changes in wall thickness and wall alignment and the use of unacceptable and screens, lead backing, or other appropriate
two IQis is not possible, the use of one IQI is acceptable. The measures shall be implemented on subsequent exposures to
IQI thickness shall be based on the thinnest wall being reduce back scatter. The appearance of a dark image (higher
radiographed and shall be placed on the thickest wall section. optical density "B" image) should be disregarded unless the
The required +30 % to -15 % optical density tolerance (see dark image could be confused with, or interfere with, interpre-
6.17), need not be met; however, the optical density in the areas tation in the area of interest. When identical parts, or segments
of interest shall be between 1.5 and 4.0 (and up to 4.5 by of parts, are to be examined by the same technique, the letter
agreement of CEO; see Note I of Fig. 1) and the required "B" may be used to qualify the initial exposure only and may
radiographic quality level shall be obtained. be omitted for subsequent exposures as long as the proximity
6.18 Non-Requirement of /Qls-The IQis are not required and nature of backscattering objects and similar conditions are
when: maintained constant. A new qualification exposure shall be
6.18.1 Examining assemblies for debris, made whenever any of the following parameters are changed:
6.18.2 Conducting radiography for defect removal provided 6.22.1 X-ray machine voltage (kilovolts, Megavolts, or
final examination of the area includes an IQI, gamma source type),
6.18.3 Examining to show material details or contrast be- 6.22.2 Exposure time (milliampere-time),
tween two or more dissimilar materials in component parts or 6.22.3 Source-to-film distance,
assemblies including honeycomb areas for the detection of 6.22.4 Screens, collimation, masking, or filters,
fabrication irregularities or the presence or absence of material, 6.22.5 Film type,
or 6.22.6 Film processing parameters,
s
i,....-
i.,...-'
,,.,,.- ~
~
/'
v
/
/
0,01 0.02 0.03 0.04 0.05 0.06 0.08 0,1
Optical Density Difference
FIG. 2 Minimum Optical Density Difference Between IQI and Ad·
jacent Base Metal Material Images
6.22. 7 Type or amount of backing material, or NoTE 5-The unsharpness values specified in Practice EI 030/E 1030M
6.22.8 Radiation source change or relocation. (Casting) or Practice E 1032 (Welding) shall be imposed if Practices
E1030/E1030M or EI032 are specified in the contractual agreement.
6.23 Source-to-Film Distance-The minimum allowable NOTE 6--Unit of measurement for SFD, Ug, F, and t may be in either
source-to-film distance shall be calculated by the following English or SI units as long as they are consistent (not mixed).
equation, using the material thicknesses and geometric un-
sharpness values in Table 4, unless otherwise specified by the 6.25.1 As an alternative to markers, view numbers (see 6.4),
cognizant engineering organization. may be used provided orientation of the radiographs to the part
can demonstrate the required radiographic coverage and the
SFD = (Ft!Ug)+t (2)
location of defects can be accomplished.
where: 6.26 Crack Detection-When parts are radiographed to
SFD = source-to-film distance, detect in-service cracks in aircraft assemblies and the parts
Ug = geometric unsharpness, being examined are covered by aircraft skin, only the area of
F = size of radiation source, using the manufacturer's
the film that falls within a 10° cone of radiation ( 10° total solid
nominal size or the effective focal spot size in
angle, apex at the radiation source, central axis of cone equal
accordance with Test Method El 165, and
to central axis of radiation beam) shall be considered valid for
t = distance from the source side of the object to the film
(regardless of whether or not the object is in contact interpretation.
with the film). 6.27 Quality Assurance Provisions:
6.24 Identification-Marking of the radiographic film and 6.27 .1 Responsibility for Examination-The NDT facility is
component shall be provided so that the radiograph may be responsible for furnishing all supplies in conformance to
traced to the component being examined (see 6.28.6). contract or purchase order requirements and, unless otherwise
6.25 Location Markers-The images of location markers for specified in the contract or purchase order, the performance of
coordination of the part with the film(s) shall appear on the film all examination requirements contained herein. The examina-
without interfering with interpretation and with such an ar- tion provisions contained herein shall become a part of the
rangement that it is evident that complete coverage was NDT facility overall examination system or quality program.
obtained. The location marker positions shall be established on 6.27 .2 Examination Report-The results of all radiographic
the part, and the position of the markers shall be maintained on examinations shall be recorded and kept on file in accordance
the part during radiography. If the entire part can be radio- with the contract or purchase order. The examination reports
graphed with one film for each view, and the orientation of the shall reference this practice, or the applicable specification and
part with respect to the film is obvious, then location markers provide for traceability to the specific part or lot examined and
are not required. shall include the examiner's identification, the date of the
examination, the disposition of the component (accept or
reject), and the reason for rejection of any items.
TABLE 4 Geometric Unsharpness (Ug) (Maximum)
6.27 .3 Film Processing Control-Film processing shall be
Material Ug, in. Ug,mm
Thickness, in. [mm] controlled and monitored in accordance with Annex A4.
Under 2 (50.8) 0.020 0.5 6.27.4 Viewers-Maximum readable optical density shall be
2 [50.8) through 4 (101.6) 0.030 0.75
Over 4 (101.6) 0.040 1.0 determined as follows and measurements shall be in accor-
dance with Specification El 390:
ANNEXES
(Mandatory Information)
Al.I Alternate Hole-Type /Qls-The IQI design shall be as Al. I .4 Material identification shall be the chemical symbol
follows: of the predominant element. For example, if iron is the
A 1.1. l Image quality indicator dimensions shall be in ac- predominant element, "FE" shall be used. Identification of
cordance with Fig. Al.I. some common materials are listed in Fig. A 1.1 . When the
material is a composite or does not have a predominant
A 1.1.2 The IQis shall be fabricated from the same material
element, a controlled system for IQI identification shall be
group (see 3.2.4) or radiographically similar materials (see
established and referenced in the written procedure (see 6.1 ).
5.5.2) as the object to be radiographed.
Al.1.3 The IQls shall be identified as to material group (see Al.1.5 The IQI thickness identification shall consist of a
Fig. A 1.1 ) and thickness relative to the thickness of the two-digit number that expresses the material thickness in one
component to be radiographed. Identification shall consist of hundredths of an inch, .XX. For example, a specimen thickness
lead letters and numbers, or a material of similar radiographic of 0.75 in. requires an IQI identified by the two-digit number
opacity. .75.
NOTE!-
Minimum IQI thickness = 0.005 in. ± 10 %.
Minimum diameter for lT hole= 0.010 in. ± 10 %. (Hole is not IT for thicknesses less than 0.010 in.)
Minimum diameter for 2T hole= 0.020 in. ± 10 %. (Hole is not 2T for thicknesses less than 0.010 in.)
Minimum diameter for 4T hole = 0.040 in. ± 10 %. (Hole is not 4T for thicknesses less than 0.010 in.)
Design tor IQI thlcknes8e8 up to and Including 0.160 In.
TDia.
T ± 10%--.f (.-
Design for IQI thickness of 0.180 In. and over made In 0.020-ln. Increments
Symbol Material
SS Stainless steel
AL Aluminum
FE Steel
MG Magnesium
cu Copper
Tl Titanium
A2.l Fabrication Welds-Accessible surfaces of welds to The long axis of the IQI shall be parallel with and at least 0.125
be radiographed shall be prepared as necessary in accordance in. (3.2 mm], but not more than 1.25 in. (32 mm] from the weld
with the welding process specification if applicable, and the edge.
valleys between beads, weld ripples, or other surface irregu-
larities are blended to a degree such that the resulting radio- A2.2.5 Circumferential Welds-Except as provided in
graphic contrast due to surface condition cannot mask or be A2.2.5. l and A2.2.5.2, radiography shall be performed in
confused with that of any defect. which the radiation passes through only one wall. When
double-wall techniques are used, either superimposed or ellip-
A2.2 /QI Selection for Welds-For butt joints on simple tical (offset) projections may be used.
structures such as pipe or plate, the thickness on which the IQI A2.2.5. l Double-Wall Exposure, Double-Wall Image-
is based shall be the single-wall thickness plus any reinforce- Welds in pipe and tubes 3.5 in. [88.9 mm] and less in nominal
ment allowed. For T-joints, lap joints, comer joints, fillet size (diameter) may be radiographed using the double-wall
welds, weld surfacing, or for joints in complex assemblies method where the radiation passes through both walls and both
where both sides of the joint are not accessible, the IQI shall be walls are viewed together for acceptance. An IQI, based on the
based on the total thickness through which the X-ray beam will double-wall thickness plus twice the weld reinforcement, as
pass. If different size sections are joined, the IQI shall be based
applicable, shall be placed on the source side, on top of a shim
on the thinner wall section. Backing strips or rings are not
approximately equal to twice the weld reinforcement. As an
considered as a part of the weld, base material, or reinforce-
alternate, the same IQI and shim may be placed on the source
ment (bead) thickness upon which the IQI is based. If subse-
side of a like section that is placed adjacent to the weld being
quent machining or mechanical preparation reduces the weld
radiographed. When impractical to do the preceding, the IQI
thickness, IQI shall be based on final product thickness. If the
may be placed on top of a block that is approximately equal to
required sensitivity cannot be obtained, r~d~ographic ex~na
tion shall be required after final machirung or mechanical twice the wall thickness plus twice the weld reinforcement, as
preparation. applicable. The IQI and block shall be placed on_ a lo~
material-density material (such as polystyrene plastic or its
A2.2. l IQ/ Placement for Welds-The IQis shall be placed equivalent), adjacent to the weld being examined, so that the
on the source side parallel with, and at least 0.125 in. [3.2 mm], top of the block is level with the upper surface of the tube or
and not more than 1.25 in. [32 mm] from the weld being pipe. The minimum source-to-film distance shall be calculated
radiographed. When the weld is not radiographically similar or
using the outside diameter of the pipe or section as the material
where part geometry precludes placement, the IQI wire or hole
thickness "t" (see 6.23).
of the required sensitivity, as applicable, may be placed over
the weld but outside of the area of interest. When weld- A2.2.5.2 Double-Wall Exposure, Single-Wall Image-For
reinforcement (weld bead), backing rings or strips are not welds in pipe and tubes greater than 3.5 in. (88.9 mm] in
removed, a shim of the same material group or radiographi- nominal size (diameter), only the weld closest to the film shall
cally similar material (see 5.5.2) shall be placed between the be viewed for acceptance. An IQI, corresponding to the
IQI and base material to provide approximately .the same single-wall thickness plus any single-wall weld reinforcement,
thickness of material under the IQI as the average thickness of shall be placed adjacent to the weld on the inner wall on top of
the weld reinforcement plus the wall thickness and backing a shim equal to the weld reinforcement. When this is not
strip or ring. possible, the same IQI and shim may be placed on the inner
wall of a like section of a tube or pipe placed adjacent to the
A2.2.2 Examination Coverage for Welds-Unless otherwise
specified on the engineering drawing or other contrac11;1al weld being radiographed. When a like section of the tube or
documents, welded parts requiring radiography shall be rad10- pipe is unavailable, the IQI may be placed on a block adjacent
graphed for 100 % of the length of the weld zone, including the to the weld being radiographed. The block thickness shall be
weld heat affected zone. equivalent to twice the wall thickness plus any single-wall
weld reinforcement. The top of the block shall not be any lower
A2.2.3 Longitudinal Circumferential Weld Junctures- than the inner surface of the inner wall of the tube or pipe. The
Where portions of longitudinal welds adjoining a circumferen-
minimum source-to-film distance shall be calculated using the
tial weld are being examined simultaneously with the circum-
single-wall material thickness as ''t" (see 6.23).
ferential weld, additional IQis shall be placed on the
longitudi.nal weld at the outer edge of the cone of radiation A2.3 Electron Beam (EB) Welds-The central beam of
used to ~diograph the circumferential weld. radiation shall be in line with and parallel to the fusion
f
A2.2.4 Longitudinal Welds-For longitudinal welds, an IQI
shall be laced at the extreme end of the area to be interpreted.
interface. Cognizant engineering organization approval is re-
quired when the beam to weld angle exceeds 2°.
A3. PROPELLANTS/EXPLOSIVES
A3.1 Installed Explosives/Propellants-For the radiogra- ET = equivalent thickness of propellant/explosive for each
phy of explosives/propellants contained within a vessel, the vessel wall material traversed by the center of the
IQI shall be based on a chord length thickness of the explosive/ X-ray beam,
propellant that is traversed by the X-ray beam. When it is not Ma = mass attenuation coefficient of the predominant ele-
possible to obtain the required radiographic quality level using ment of the vessel section, and
an IQI that is based only on the explosive/propellant thickness, Mb = mass attenuation coefficient of the predominant chemi-
the chord lengths through the vessel wall material layers may cal element of the propellant/explosive.
be used in determining the total equivalent thickness of NoTE A3.l-These relationships were derived experimentally and have
explosive/propellant upon which the IQI thickness is to be been determined adequate for the evaluation of vessels containing
explosives/propellants.
determined. The equivalencies shall be determined in the The IQis may be fabricated from inert materials that have approxi-
following manner: mately the same physical material density as the propellant/explosive to be
radiographed. As an alternate, the IQI may be fabricated from, and based
ET= ACIB (A3.l) on, a material with higher material density with a thickness that has been
For energy levels of 320 keV and less, the following equated to the total equivalent propellant/explosive thickness as deter-
equation shall be used: mined by the method previously shown.
A4.1 Processing Equipment and Solutions-For both auto- NOTE A4. l-The test piece shall be a block of uniform thickness, or a
mated and manual processing, solutions and equipment shall stepwedge. Both the block and stepwedge shall be large enough to ensure
that no radiation due to undercutting has affected the area of interest.
be maintained in accordance with manufacturer's recommen-
dations to ensure a consistent and acceptable level of perfor- A4.2. l .2 Monitor Film-To monitor processor performance,
mance. Guide E999 should be consulted for guidance on film a Monitor Film shall be processed to the standardized tech-
processing. The following shall be performed and recorded: nique or a sensitometric film strip shall be used. The Monitor
Optical density shall be measured in the same areas as the
A4.2 Automated Processing: Control Film and compared to the Control Optical densities.
A4.2. l Processor Performance-Processor performance The Monitor Optical densities shall be within :!: 10 % of the
shall be determined by comparing a Monitor Film to a Control current Control Film. If the Monitor Optical density is out of
Film, as scheduled in A4.2. l. I and Table 2 in freshly mixed tolerance, necessary steps shall be taken to correct the problem,
seasoned developer. and another Monitor Film shall be processed to verify that the
problem has been corrected. Monitor Film for the past four
A4.2. l. l Control Film, shall be a sensitometric film strip, or
weeks shall be maintained for auditing purposes. Any produc-
pre-exposed film of a standard test piece (see Note A4. l ).
tion film processed from the last successful check need not be
When using the standard Test Piece Method to produce the
discarded provided the radiographic quality level and optical
Control Film, establish a standardized technique and dedicate
density requirements of 6.9 and 6.10 are achieved. Optical
one batch of film to reduce variables. The Control Film shall
density measurements shall be recorded as scheduled in Table
have at least one area with an optical density of 2.0 to 2.5 H&D
2.
inclusive, and can include an unexposed ar~a of the film for
monitoring base fog. (Sufficient blocking should be used to A4.2.2 Replenishment Rates-The rate at which developer
ensure that the area for monitoring base fog receives no and fixer are replenished shall be maintained within the
radiation exposure.) Control Optical densities shall be mea- manufacturer's recommended range and checked when solu-
sured and recorded, establishing nominal control values. tions are changed, during preventative maintenance, or main-
(Newly changed solutions should be seasoned in accordance tenance associated with the replenishment system. Replenish-
with Guide E999 prior to processing the Control Film.) ment rates shall be checked in accordance with the
APPENDIX
(Nonmandatory Information)
Xl.l The conversion to SI units for thicknesses and hole element, "FE" is used. Identification of some common mate-
diameters is for reference use only. rials is listed in Fig. X 1.1 .
Xl.1.1 The IQI is identified as to material group (see Fig. Xl.1.3 The IQI thickness identification consists of a two-
X 1.1 ) and thickness relative to the thickness of the component digit number. For example, a specimen thickness of 0.75 in.
to be radiographed. Identification consists of lead letters and (19.05 mm) requires an IQI identified by the two-digit number
numbers, or a material of similar radiographic opacity. .75 [19.05 mm].
X 1.1.2 Material identification is the chemical symbol of the Xl.1.4 Notes 1-5 represent direct conversion to SI units for
predominant element. For example, if iron is the predominant reference use only.
• E1742/E1742M - 23
Symbol Material
SS Stainless steel
AL Aluminum
FE Steel
MG Magnesium
cu Copper
Tl Titanium
Committee E07 has identified the location of selected changes to this standard since the last issue
(E1742/E1742M-18) that may impact the use of this standard. (Approved Dec. 15, 2023.)
(1) Made various editorial changes. (6) Deleted previous 3.2.1 as this is defined in Terminology
(2) Added sentence addressing weldments and brazing to 1.2. E 1316. Renumbered remaining subsections.
(3) Corrected titles of Guide E94/E94M and Practice E1030/ (7) Revised 5.1.1.
El030M. (8) Changed "film" to "radiograph" in 5.2.4, 5.4.3.1 , 5.4.4.1 ,
(4) Deleted previous 2.4 and 2.5 and renumbered subsequent 5.4.4.5, 6.11 , and 6.22.
subsections. (9) Revised Sections 5 and 6.
(5) Deleted ISO 9712 from 2.5. ( 10) Added Note 5.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Baff Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
PRINTED DATE:
No reproduction or networ1<ing permitted without license from Accuris Nol for Resale,2024-02-20 15:07:39 UTC