Elucidating The Decomposition Kinetics of Xanthate Compounds in Mining Waters by Headspace Gas Chromatrography-Mass Spectrometry
Elucidating The Decomposition Kinetics of Xanthate Compounds in Mining Waters by Headspace Gas Chromatrography-Mass Spectrometry
Elucidating The Decomposition Kinetics of Xanthate Compounds in Mining Waters by Headspace Gas Chromatrography-Mass Spectrometry
uk
Provided by Welcome to TRUSpace | TRUSpace
Faculty of Science
In the
(CHEMICAL BIOLOGY)
This thesis has been accepted as conforming to the required standards by:
Dated this 1st day of May, 2018, in Kamloops, British Columbia, Canada
Xanthates are a widely used class of floatation reagents used for the recovery of valuable
sulphide minerals. Floatation utilizes large quantities of water that lead to rapid hydrolysis of
xanthates in circuit leading to a reduction in mineral recovery and an overall decrease in mill
efficiency. Currently the knowledge on xanthate degradation is sparse with only a few studies
reporting kinetic parameters for xanthate degradation into carbon disulphide (CS2). A headspace
GC-MS method was developed to directly measure the presence of CS2 in the gas phase. Kinetic
parameters were then established for potassium amyl xanthate (PAX), sodium isobutyl xanthate
(SIBX), potassium isopropyl xanthate (PIPX), and sodium ethyl xanthate (SEX) based on the
generation of CS2. The decomposition of all xanthate studied followed first order kinetics and the
rate constants were determined using a 7650A auto sampler and a PAL3 auto sampler. The rate
constants were found to be 7.05 x 10-4, 4.07 x 10-4, 5.11 x 10-4, and 1.48 x 10-4 h-1 for PAX,
SIBX, PIPX, and SEX respectively at 25 oC using the 7650A auto sampler. For the PAL3 auto
sample the rate constants were found to be 3.71 x 10-6, 4.39 x 10-6, 1.86 x 10-6, and 4.03 x 10-6 h-1
for PAX, SIBX, PIPX, and SEX respectively at 30 oC. The rate constants were found to increase
as the temperature was increased. These data were used to calculate the activation energies for
each decomposition reaction. The activation energies were found to be 19.83, 10.80, 34.44, and
22.64 kJ/mole for PAX, SIBX, PIPX, and SEX respectively. These data show that the methods
presented are viable options for further analytical studies on xanthate decomposition kinetics and
ii
ACKNOWLEDGEMENTS
I would like to thanks my supervisor Dr. Kingsley Donkor for all the support and
guidance provided throughout the duration of the project. I also would like to thank Trent
Hammer for training me to use the GC-MS and for providing expertise on instrumentation. John
Andre from New Afton Gold Mine in Kamloops also deserves a thank you for providing samples
and approaching myself and Dr. Donkor about this research opportunity. A special thanks you to
the Thompson Rivers University Chemistry department, honours coordinators, and research
office for making this research possible. Finally, I would like to extend thanks to Dr. Jon Van
Hamme and Dr. Heidi Huttunen-Hennelly for agreeing to be a part of my honours committee.
iii
TABLE OF CONTENTS
ABSTRACT…………………………………………………………………………………… II
ACKNOWLEDGEMENTS…………………………………………………………………. III
LIST OF FIGURES………………………………………………….……………………… V
LIST OF TABLES…………………………………………………………………………… VI
1. INTRODUCTION…………………………………………………………………………. 1
2.1 REAGENTS…………………………………………………………………………… 7
2.2 INSTRUMENTATION……………………………………………………………...... 7
2.3 PREPARATION OF CS2 STANDARDS…………………………………………….. 8
2.4 PREPARATION OF XANTHATE STANDARDS………………………………….. 8
4. CONCLUSIONS……………………………………………………………………….... 22
5. FUTURE WORK……………………………………………………………………..…. 23
6. LITERATURE CITED…………………………………………………………………. 24
7. APPENDIX………………………………………………………………………………. 25
iv
LIST OF FIGURES
Figure 1. General structure of a xanthate anion. The R-group will always be one of four 1
alkyl groups: ethyl, isopropyl, isobutyl, or an amyl group.
Figure 5. Calibration curve for CS2 using the 7650A auto sampler. An injection volume 11
of 5 µL was used with a 20:1 split. The headspace volume was 1 ml and the
temperature was 25 °C. The equation of the line and correlation coefficient are
shown within the graph area.
Figure 6. Calibration curve for CS2 using the PAL3 auto sampler. An injection volume 12
of 75 µL was used with a 100:1 split. The headspace volume was 10 ml and
the temperature was 30 °C. The equation of the line and correlation coefficient
are shown within the graph area.
Figure 8. Plot of CS2 evolved from a specific xanthate: PAX, SIBX, PIPX, or SEX, 15
against time using the 7650A auto sampler. Samples were run at 25 °C and 5
µL was sampled from 2 mL of headspace.
v
LIST OF TABLES
Table 1. Rate constants for PAX, SIBX, PIPX, and SEX determined using the 7650A 15-16
auto sampler. The constant was calculated using curve of CS2 generated over
time.
Table 2. Rate constants for PAX, SIBX, PIPX, and SEX determined using the PAL3 17
auto sampler. The constant was calculated using curve of CS2 generated over
time.
Table 3. Activation energy of PAX, SIBX, PIPX, and SEX determined using the 20
Arrhenius equation and the rate constants calculates using the PAL3 auto
sampler.
vi
1. INTRODUCTION
Alkyl dithiocarbonates, more commonly known as xanthates, are a class of widely used
mineral collectors in mine ore slurry. When combined with a technique know as froth
floatation, xanthates are able to selectively and efficiently concentrate sulphide minerals such as
chalcopyrite (CuFeS2) (Li, et al., 2015). This has propelled xanthates to the forefront of choice
reagents for the floatation of copper and gold sulphide minerals (Rezaei, Massinaei, and
Zeraatkar Moghaddam, 2018). Xanthate molecules, shown in Figure 1, all share a common
dithiocarbonate group (-OCS2) and have a variable alkyl chain attached through an ester
valuable mineral particles from the matrix. The coarse mill product is placed into a large
floatation vat containing water and xanthates are added to begin the floatation process (911
Metallurgist, 2017). The negative sulphur atom in the xanthate will bind and reduce metal
containing sulphides. The xanthate will selectively impart hydrophobicity on the mineral of
interest so that it can be separated from the aqueous matrix and the other hydrophilic gangue
(Kemppinen, Aaltonen, Sihvonen, Leppinen, and Siren, 2015) The result is a micelle like layer
of xanthates with the alkyl tails facing outwards surrounding the mineral particle. An agitator
then aerates the tank to create bubbles. The hydrophobic tail of the xanthate molecule will
1
interact with these bubbles as they are produced and float to the surface carrying the mineral
with it. This process may be repeated many times with grinding between each collection to
further concentrate the mineral of interest. Previous research has suggested that floatation
efficiency and selectivity increase as the alkyl chain length is increased (Taguta, O'Connor, and
McFadzean, 2017); however, aside from the additional hydrophobicity that is imparted by the
larger hydrophobic group, a consensus has not been reached on an exact mechanism.
catalytic nature of water on xanthates in solution (Shen, Nagaraj, Farinato, and Somasundaran,
2016). Xanthates are typically supplied to mills in a powder or pellet form (Mining World,
2017), which can then be prepared to a 5-10% solution to be used for floatation. The standard
practice is to consume these solutions within three weeks due to degradation in the aqueous
media. As the xanthates degrade in solution the concentration is decreased at an accelerated rate
and produces multiple by-products, which can severely impact the recovery of sulphide
Leppinen, and Siren, 2015). The most common by-product of the degradation process is carbon
disulphide (CS2), a toxic volatile species known to accumulate in floatation mills (Shen,
Nagaraj, Farinato, and Somasundaran, 2016). Since the degradation process is continually
occurring within the floatation reactions, the generation of toxic CS2 presents not only an issue
with decreasing flotation efficiency, but on the health and safety of workers and the surrounding
environment. It is estimated that about half of the total xanthates are consumed during the
floatation reaction while the remainder is discharged into tailings waste (Li, et al., 2015). This
tailings waste will continue to undergo degradation and produce toxic species, which can
contaminate surrounding areas. Often the tailings water is discharged into a natural body of
2
water without any treatment. This makes it important to understand the decomposition process
of xanthates not only for industrial applications but also for waste management.
ultraviolet/visible (UV/Vis) spectrophotometry (Hao, Silvester, and David, 2000); however this
method cannot distinguish between metal species and degradation products. The goal of these
studies was to track the decrease in xanthate concentration over time in order to elucidate its
temperature, time and pH on the stability of potassium ethyl xanthate (PEX) using a UV/Vis
method. It was found that after thirteen days at a pH of 9, PEX absorbance did not decrease
significantly (0.451%/day). When the pH was decreased to 7 the rate of degradation increased
to 0.9024% per day, and increased even further to 2.099% as the pH was decreased to 5.
Similarly, when the temperature was increased from 10 oC to 25 oC the rates of decomposition
were at least tripled at each pH. By observing the appearance and disappearance of peaks when
the pH was changed, two theoretical decomposition reacts were proposed (Figure 2). In acidic
solution the formation of a xanthic acid intermediate (C2H5OCS2H) dominated; however they
did not describe how this reaction leads to an increase in the rate of decomposition. In basic
solution a similar reaction occurs but instead of a xanthic acid intermediate, the dominant
degradation products are ethanol and CS2. Again, the link between the different reactions and
how they potentiate the differences in the rate of decomposition has not been clarified.
3
A similar study presented the decomposition rates by calculating the rate constant (k) for
each condition using SEX (Zhongxi and Forsling, 1997). At 40 oC the rate constants all had
orders of magnitude of 10-6. The order of magnitude decreased to 10-7 and then 10-8 as the
temperature was reduced to 20 oC and 5 oC respectively. It was also found that the difference
between the rate constants at pH 8.0 and 6.6 increased as the temperature was increased. This
suggests that temperature may be the dominant factor driving decomposition, and at high
enough temperatures pH may not significantly affect the rate compared to temperature.
The inability to distinguish between the different xanthates in solution has lead to the
Aaltonen, Sihvonen, Leppinen, and Siren, 2015). These methods have allowed for the separation
and monitoring of the individual xanthates. A novel gas chromatography method was developed
by Li and colleagues (2015) for the detection of isobutyl xanthate (IBX) in surface and drinking
water. The research was based on the decomposition of IBX into CS2 under acidic conditions.
Li argued that previous methods that aimed to prevent the decomposition of the xanthate during
analysis have used pH levels that are too acidic for use with a C18 column. Additionally, CE
methods have suffered from high detection limits (10-6 to 10-7 g/ml), which are not compatible
with residual xanthate concentrations typically seen in floatation and wastewater circuits (7 x
10-7 g/ml). By detecting the evolution of volatile CS2 using a headspace GC method with an
The use of a stable degradation product to quantify xanthates has allowed for the
CS2 directly. A study by Shen and colleagues (2016) used a headspace GC - mass spectrometry
(GC-MS) for the detection of volatile CS2 evolved from the decomposition of SIBX. The
4
derived rate law from these experiments showed SIBX followed first order kinetics with respect
to the generation of CS2. The work also supported previous research by noting the increase in
degradation associated with increased temperatures and acidic pH. These changes were
reflected by the rate constants determined under various conditions. At 25 oC, k was found to be
9.3 x 10-4 h-1. This value increased to 1.7 x 10-2 h-1 when the temperature was increased to 50 oC,
and to 1.3 x 10-1 h-1 at 70 oC. A maximum rate of decomposition was measured at pH 2.2, and it
was noted that above pH 8, no significant change in the rate of decomposition was observed.
GC-MS is a technique that combines the separative technique of gas chromatography with
the detection ability of a mass spectrometer. A general schematic for the instrument is shown in
Figure 3. In GC, an inert gas such as helium is used as a mobile phase to carry the sample
through the column (Bacher, 2016). The column contains a solid support coated with a high
boiling point liquid polymer. Columns may be capillary based, containing a coating directly
around the inner diameter of the capillary. Separation is achieved based on interactions between
the stationary phase and the components of the sample. If a non-polar polymer is used, the
column will retain non-polar components of the sample while the polar components will be left
to flow through freely. Components are also separated based on their boiling points.
Compounds with low boiling points have high vapour pressure and will spend more time in the
gaseous phase than interacting with the liquid stationary phase, yielding shorter retention times
5
Figure 3 Schematic diagram of a gas chromatograph single quadrupole mass spectrometer
(GC-MS) (Li et al., 2015).
As the components of the sample are separated and eluted at different times they are
detected by the mass spectrometer. MS ionizes molecules in a vacuum as they emerge from the
GC using a beam of high-energy electrons (Clinical Mass Spectrometry, 1995). This removes
an electron from the highest occupied molecular orbital of the molecule to produce a radical
with a charge of +1. The molecule may also be fragmented extensively by excess energy to
produce fragments that can help elucidate structure. The charged molecules are then subject to a
quadrupole mass analyzer. Four cylindrical rods are arranged parallel to one another and a
radiofrequency voltage is applied to one pair of rods while an opposing DC voltage is applied to
the other pair. Ions travel down the quadrupole and the voltages are adjusted so only ions of a
particular mass to charge (m/z) ratio will have stable trajectories and reach the detector. The
voltages are continually adjusted to detect ions with a mass range of up to 5000 m/z. The
detector collects and counts the ions to produce a signal proportional to the number of ions that
have impacted its surface. This yields not only a mass spectrum showing the molecular mass of
6
the compounds in the sample, but a chromatogram that can yield quantitative information base
on parameters such as peak height or peak area, both of which are proportional to concentration.
Although literature does exist examining the decomposition of xanthates through the
evolution of CS2, no studies have been found comparing the kinetics of each of the xanthates
under a common method. This study was set out with two goals in mind. First, was to develop a
suitable headspace GC-MS method for the detection of CS2 in aqueous solution. Once the
instrument was calibrated, the decomposition kinetics of PAX, SIBX, PIPX, and SEX were
2.1 Reagents
Pure CS2 solution was purchased from Sigma-Aldrich Canada Ltd., Oakville, Ontario, Canada
and was of analytical grade. Xanthate standards were supplied by New Afton Gold Mine,
Kamloops, British Columbia, Canada, as both dissolved solid and pure solid. All solutions were
prepared using filtered 18 MΩ water. Any solutions containing dissolved xanthate was filtered
2.2 Instrumentation
All experiments were performed using an Agilent 7890B-GC coupled 5977A-MS (Agilent
Technologies, Santa Clara, CA). Half of the experiments used a 7650A (Agilent) automatic liquid
sampler with a 10 –µL headspace syringe, and the second half utilized a PAL3 (Agilent) auto
sampler system equipped with a headspace tool. In both cases an HP-5MS 5% phenyl methyl
silox capillary column was used (29.94 M x 250 µM x 0.25 µM) (Agilent Technologies, Santa
Clara, CA). The temperature program used was developed from a previous method (Shen,
7
Nagaraj, Farinato, & Somasundaran, 2016). The temperature was held at 35 oC for 2 min before
increasing the temperature by 50 oC per minute to 220 oC where it was held for an additional 1.3
The optimized conditions for the 7650A auto sampler were a flow rate of 1 mL/min and a
split of 20:1. The injection volume was 5 µL a single ion monitor program was set up to observe
the presence of CS2 at 76.00 m/z with a gain factor of 10.00. For the PAL3 auto sampler the split
was increased to 100:1 and the gain was reduced to 1 to account for the increased sensitivity of
accommodate the larger headspace needle used in the PAL3 system. The single ion monitoring
µL of pure CS2 to 100.0 mL with 18 MΩ water. Samples were then diluted further to 1 ppm
before being loaded into clean and dry 2 mL glass sample vials that were capped immediately.
Sample vials were shaken rapidly for 15 min in a flask shaker as suggested by Li and colleagues
(2015). These samples were used to confirm detection of CS2 and to optimize detection
conditions. Calibration standards were prepared using the same methodology to various
concentrations. For the PAL3 auto sampler the use of 20 mL glass headspace vials allowed
increased the loading volume to 10 mL. Samples were agitated at 30 oC for 15 min using the built
0.0500g of solid PAX (97%), SIBX (95%), PIPX (98%), and SEX (95%) was dissolved in 18
8
MΩ water to a volume of 500 mL to prepare 1000 ppm stock solutions of each. Stock solutions
were stored at 4 oC in the dark to minimize aqueous degradation. Fresh stock solutions were
prepared weekly. Standards were prepared by first filtering the stock solution through a 0.45 µM
nylon filter prior to dilution. 1 mL of sample was loaded into a clean glass vial for the 7650A
sampler, and 10 mL was loaded for the PAL3 sampler. Samples were then shook or agitated for
at least 15 min prior to injection. All 7650A samples were equilibrated under ambient conditions
and a pH of about 8.50. PAL3 sample were equilibrated at varying temperatures and a pH of
about 8.50.
standard solution prepared in water. This standard was chosen as it falls within the concentration
range of CS2 typically observed in mining circuits. A reproducible peak was observed around
1.88 min and had an m/z of 76 (Figure 4), using both auto sampler systems. This peak was
confirmed as the CS2 peak and a single ion-monitoring program was set to detect ions of 76 m/z
for the 7650A auto sampler. No interfering peaks were noted in the chromatogram as the solvent
peak comes out before 1 min, which is not detected by the mass spectrometer due to a minute
long solvent delay where no ions are detected for the first minute of a run.
9
Figure 4 A chromatogram of a 1 ppm CS2 standard solution. Sample volume injected was 5 µL
with a split of 20:1.
Once a suitable method had been developed for the detection of CS2, a linear range was
established and calibrated for. The solubility of CS2 at 20 oC has been reported to be as low as 2
ppm (World Health Organization, 2000), and as high as 2000 ppm (Chemical Book, 2017). To be
safe the highest calibration standard was chosen to be 3 ppm to see if linear range of 0.3 – 3 ppm
could be established. A calibration curve was generated from these standards by plotting the area
under the peak in the chromatogram against the concentration of CS2 (Figure 5). A strong
correlation between the concentration of CS2 and peak area was observed (R2 = 0.977). The limit
of detection (LOD), or the minimum amount of analyte that can be distinguished from the
absence of that analyte, and the limit of quantification (LOQ), or the minimum amount of analyte
that can be detected with analytical certainty, was found using the standard deviation of the
response and the slope. LOD was determined by multiplying the ratio of the standard deviation of
the lowest concentration standard over the slope of the line by 3.3, and LOQ multiplied the same
ratio by 10. These are the standard ratio multiples used for LOD and LOQ calculation in
analytical studies. The LOD was found to be 0.03 ppm, and the LOQ was found to be 0.09 ppm.
10
These values are higher than previously reported LODs using headspace GC-MS (0.002 ppm)
(Li, et al., 2015); however, by increasing the number of calibration standards and replicates this
Figure 5 Calibration curve for CS2 using the 7650A auto sampler. An injection volume of 5 µL
was used with a 20:1 split. The headspace volume was 1 ml and the temperature was 25 °C.
The equation of the line and correlation coefficient are shown within the graph area.
A similar calibration protocol was carried out using the PAL3 auto sampler. Since the
same instrument, column, and temperature program was used for both auto samplers the retention
time of CS2 was unchanged using the PAL3 system. Since the PAL3 system is equipped with a
temperature-controlled agitator module and uses dedicated headspace equipment, the sensitivity
of the instrument increased, and the signals were too strong to continue using the same method as
before. The split ratio was increased 100:1 and the gain factor on the MS was reduced to 1. An
11
injection volume of 75 µL was also used to accommodate the larger headspace needle used by the
PAL3 system. Additionally, the single ion monitoring was removed from the MS, as the
increased sensitivity it provided was no longer needed. A new range was established from 0.1 to
2 ppm (R2 = 0.977) (Figure 6). The LOQ and LOD were also established for this curve using the
same method as described above. The LOD was found to be 0.04 ppm and the LOQ was found to
be 0.12 ppm. The lowest standard used is slightly below the LOQ, so the implications of
increasing the number of repeats to reduce the LOD and LOQ will be substantial for its
application in industry.
Figure 6 Calibration curve for CS2 using the PAL3 auto sampler. An injection volume of 75
µL was used with a 100:1 split. The headspace volume was 10 ml and the temperature was 30
°C. The equation of the line and correlation coefficient are shown within the graph area.
The direct use of CS2 for calibration allows for the use of ambient conditions and native
pH values. Previous methods that indirectly calibrate for CS2 by degrading xanthates must rely
12
on acidic pH and high temperatures to produce a detectable amount of CS2 in an analytical
timeframe (Li, et al., 2015). This extra sample preparation time limits industrial applicability. The
current method proposed is limited to a narrower range due to the low solubility of CS2 in water;
however, the window for industrial applicability still falls within this range making it more
generated against initial xanthate concentration was prepared. The curve was prepared for each of
the four xanthates to be studied: PAX, SIBX, PIPX, and SEX. All the curves showed linear
relationships between the initial xanthate concentration and the amount of CS2 generated (Figure
7). All the correlations showed R2 values greater than 0.99, except for SIBX at 0.88. The linear
relationship seen confirms that the decomposition follows first order kinetics. The rate of change
in the production of CS2 equals that of the change in initial xanthate concentration. The rate law
wherein rate equals the slope of the line of the curve of CS2 concentration generated during
decomposition against time, k is the rate constant, and [xanthate] is the initial concentration of
13
It should be noted that this experiment was carried out only on the 7650A sampling system,
rate constants of each xanthate. For the 7650A sampler the four xanthates all showed linear
relationships with R2 values greater than 0.92 (Figure 8). The rate constant for the degradation
reaction of each xanthate was calculated by dividing the slope of the line by the initial xanthate
concentration in the sample. In all cases the concentration of xanthate used was close to 100 ppm.
The rate constants at 25 oC were found to be 7.05 x 10-4, 4.07 x 10-4, 5.11 x 10-4, and 1.48 x 10-4
14
h-1 for PAX, SIBX, PIPX, and SEX respectively (Table 1). These values are comparable to those
found in the literature using similar methodology (Shen, Nagaraj, Farinato, & Somasundaran,
2016); however literature reported by Zhongxi and Forsling (1997) reported rate constants at 25
o
C to be on the magnitude of 10-6 and 10-7.
Figure 8 Plot of CS2 evolved from a specific xanthate: PAX, SIBX, PIPX, or SEX, against time
using the 7650A auto sampler. Samples were run at 25 °C and 5 µL was sampled from 2 mL of
headspace.
Table 1 Rate constants for PAX, SIBX, PIPX, and SEX at 25 °C determined using the 7650A
auto sampler. The constant was calculated using curve of CS2 generated over time.
Xanthate k (h-1) ± uncertainty
15
PIPX 5.11 x 10-4 ± 0.62 x 10-4
A similar methodology was repeated using the PAL3 sampler; however, the addition of
the incubated agitator allows for the determination of k at varying temperatures. Another
adaption involved sampling from a single vial over time rather than from multiple vials due to the
capacity of the incubator. The rate constants for each xanthate at varying temperatures are shown
in Table 2. An important feature to note is that all the values are smaller than those calculated
using the 7650A auto sampler (Table 1). At 30 oC the rate constants were 3.71 x 10-6, 4.39 x 10-6,
1.86 x 10-6, and 4.03 x 10-6 h-1 for PAX, SIBX, PIPX, and SEX respectively at 30 oC. These
values did not increase significantly when the temperature was increased to 50 oC for PAX but
did increase slightly for SIBX, PIPIX, and SEX. At 80 oC all the rate constants except for SIBX
increased by an order of magnitude. These results are comparable to those published by Zhongxi
and Forsling (1997) which reported rate constants to the order of 10-6 for ethyl xanthates at 40 oC
16
Table 2 Rate constants for PAX, SIBX, PIPX, and SEX determined using the PAL3 auto
sampler. The constant was calculated using curve of CS2 generated over time.
Xanthate Temperature (oC) k (h-1) ± uncertainty
Comparing the two methods, it appears that the 7650A auto sampler system appears
superior for the determination of the decomposition kinetics of xanthates; however, a few
considerations must be taken into account. Firstly, the 7650A method utilizes a different analysis
method where no vial is sampled from more than once. 8 samples are prepared into 8 separate
vials and each is analyzed once to generate the curve. The PAL3 system uses an incubator that
has the capacity for 6 vials. The method was modified to not only accompany the restrictions of
the incubator, but also to automate the method further for studies at higher temperatures. Rather
than separating each xanthate into a number of different vials, one vial was utilized for each
17
xanthate and was repeatedly sampled over time. This allowed for a much higher throughput of
samples and a level of automation that would be applicable to an industrial setting. The
consequence of this is that the septum of the sample vial is punctured multiple times which can
create leaks where volatile CS2 can escape. This might explain why the slopes generated from the
PAL3 curves are reduced, over the time course of the study some of the CS2 was likely escaping,
which would artificially reduce values for the slope of the line.
There may be a few reasons the results presented here differentiate from those in the
literature. In terms of developing a higher throughput method for use in industry all solutions
were prepared without pH adjustment. It was noted that the innate pH of xanthate solutions are all
around pH 8.5 and degradation is not severely impacted until the pH drops below 6 (Kemppinen,
Aaltonen, Sihvonen, Leppinen, & Siren, 2015). Values reported by Shen and colleagues (2016)
used a pH of 6.8 for analysis while Zhongxi and Forsling (1997) used a pH of 8 and 6.6 and
reported similar values for both. The methods used here all differ from Zhongxi and Forsling’s
(1997) study as the decrease in xanthate concentration was monitored rather than the evolution of
CS2. The method also utilized UV/Vis spectrophotometry and bands could not be assigned
quantification compared to headspace GC-MS methods that measure CS2 directly. Both of the
comparative studies also only analyzed a single xanthate (SIBX or ethyl xanthate) so no
each xanthate were also investigated. A plot of Ln (k) against the reciprocal of temperature was
generated according to the Arrhenius equation. The activation energy (Ea) for the degradation of
18
each xanthate into CS2 was found by multiplying the slope of the graph (Ea/R) by the ideal gas
constant R (8.314 j mol-1). The activation energies were found to be 19.83, 10.80, 34.44, and
22.64 kj/mole for PAX, SIBX, PIPX, and SEX respectively. Only literature values for the
activation energy of O-ethyl S-methyl xanthate could be found in the literature (Adejoro, Esan,
Adeboye, & Adeleke, 2017). The researchers reported an Ea of 166.200 kj/mol for the S-
methylated ethyl xanthate. The values reported in this research are significantly lower which is
likely due to the chemical difference between the species. The methylated sulphur reduces the
overall polarity of the xanthate making it less susceptible to degradation reactions in water. This
would increase the activation energy so that more energy is needed to initiate the degradation of
the methylated compound. Additionally, Adejoro and colleagues (2017) utilized an in silico
This work has shown that the PAL3 method may be suitable for investigating the thermodynamic
properties of the xanthates in tandem with their kinetic properties. This was a pilot study that
focused on simple data manipulation. Further work needs to be done to expand the Arrhenius
19
Figure 9. Arrhenius plot of ln(k) against the reciprocal of temperature. Temperature used were
30 °C, 50 °C, and 80 °C. All k values were calculated from data generated using the PAL3 auto
sampler.
Table 3 Activation energy of PAX, SIBX, PIPX, and SEX determined using the Arrhenius
equation and the rate constants calculates using the PAL3 auto sampler.
Xanthate Ea (kj/mole)
PAX 19.83
SIBX 10.80
PIPX 34.44
SEX 22.64
20
4. CONCLUSIONS
Headspace GC-MS was shown to be a suitable technique for studying the decomposition
kinetics of xanthates in aqueous matrices at ambient and elevated temperatures. Conditions were
optimized for the GC-MS using both a 7650A and PAL3 auto sampler system. Both systems
were successfully calibrated for the detection of volatile CS2 and xanthates were confirmed to
follow first order decomposition into CS2. Rate constants were successfully calculated at 25, 30,
50, and 80 oC; however, the two sampling systems did not produce comparable results to one
another but did reflect values presented in the literature for both headspace GC-MS methods, and
UV/Vis spectrophotometric methods. This work has developed methods for both non-headspace
specific (7650A) and headspace dedicated (PAL3) instrumentation. The methods automation and
potential for high throughput make it ideal for use in an industrial setting where sample turnover
is fast and sample prep should be minimized. This work also provides foundations for analytical
studies into the degradation kinetics of xanthates in solution and decomposition thermodynamics.
21
5. FUTURE WORK
Future work for this research should focus on increasing the analytical certainty of the values.
Increasing the repeats of each sample and using fresh headspace vials and lids for each analysis
will likely accomplish this. Since xanthate degradation is heavily pH dependant future studies
should also explore how k changes in response to changes in pH. Additional thermodynamic
studies can also be carried out to further elucidate the thermodynamics of the degradation
reaction. Finally, since the floatation tank is a concoction of chemical activity, it may be
beneficial to monitor for any potential destabilizing or stabilizing cosolutes in solution that may
behaviour of each xanthate can be developed to allow for mining operations to adjust current
floatation protocol to correct for any xanthate degradation that may be affecting mineral recovery
in circuit.
22
6. LITERATURE CITED
911 Metallurgist. (2017). 911 Metallurgist. Retrieved April 02, 2018 from 911 Metallurgist:
www.911metallurgist.com/blog/flotation_collectors
Adejoro, I., Esan, T., Adeboye, O., & Adeleke, B. (2017). Quantum mechanical studies of the
kinetics, mechanisms, and thermodynamics of gas-phase thermal decomposition of ethyl
dithiocarbonate (xanthate). Journal of Taibah University for Science , 11 (5), 700-709.
Bacher. (2016, April 01). Gas Chromatography Theory. Retrieved April 02, 2018 from UCLA
Chemistry: www.chem.ucla.edu/~bacher/General/30BL/gc/theory.html
Chemical Book. (2017). Carbon Disulphide . Retrieved April 14, 2018 from Chemical Book:
www.chemicalbook.com/ChemicalProductProperty_EN_CB6279761
Clinical Mass Spectrometry. (1995). Basic theory of mass spectrometry. Clinica Chimica Acta ,
241, 15-71.
Hao, F., Silvester, E., & David, G. (2000). Spectroscopic characterization of ethyl xanthate
oxidation products and analysis by ion interaction chromatography. Analytical Chemistry
, 72 (20), 4836-4845.
Kemppinen, J., Aaltonen, A., Sihvonen, T., Leppinen, J., & Siren, H. (2015). Xanthate
degradation occuring in floatation process waters of a gold concentrator plant. Minerals
Engineering , 80, 1-7.
Kim, I., Suh, S., Lee, I., & Wolfe, R. (2016). Applications of Stable, Nonradioactive Isotope
Tracers in In Vivo Human Metabolic Research. Experimental and Molecular Medicine ,
48 (1), 1-10.
Li, N., Chen, Y., Zhang, C., Chen, W., Fu, M. Y., Fu, M., et al. (2015). Highly Sensistive
Determination of Butyl Xanthate in Surface and Drinking Water by Headspace Gas
Chromatography. Chromatographia , 78, 1305-1310.
Mining World. (2017, June). Alternatives to xanthate collectors for sulphide flotation
applications. Mining World Processing , 14 (3), pp. 24-28.
Mustafa, S., Hamid, A., Naeem, A., & Sultana, Q. (2004). Effect of pH, Temperature and Time
on the Stability of Potassium Ethyl Xanthate. Journal of the Chemical Society of Pakistan
, 26 (4), 363-366.
Rezaei, R., Massinaei, M., & Zeraatkar Moghaddam, A. (2018). Removal of residual xanthate
from flotation plant tailings using modified bentonite. Minerals Engineering , 119, 1-10.
23
Shen, Y., Nagaraj, D., Farinato, R., & Somasundaran, P. (2016). Study of xanthate
decomposition in aqueous solutions. Minerals Engineering , 93, 10-15.
Taguta, J., O'Connor, C., & McFadzean, V. (2017). The effect of alkyl chain length and ligand
type of thiol collectors on the heat of adsorption and floatability of sulphide minerals.
Minerals Engineering , 110, 145-152.
Trudgett, M. (2005). The Ultra-Trace Level Analysis of Xanthates by HIgh Performance Liquid
Chromatography. University of Western Sydney, Physcial Sciences . Sydney: University
of Western Sydney.
World Health Organization. (2000, April). Carbon Disulfide. Retrieved April 14, 2018 from
www.inchem.org: www.inchem.org/documents/icsc/icsc/eics0022.htm
Zhongxi, S., & Forsling, W. (1997). The degradation kinetics of ethyl-xanthate as a function of
pH in aqueous solution. MInerals Engineering , 10 (4), 389-400.
24
7. APPENDIX
25
750 ppm PIPX
26
7.2 Mass Spectra of Xanthate Degradation Process
PAX
27
SIBX
28
PIPX
SEX
29