Troubleshooting
Troubleshooting
Troubleshooting
1 © Cepheid – Confidential.
Troubleshooting Approach
1. In case of problem, look for an error message on the screen (pop-up windows, message
box at the bottom of the screen or in the "Error" tab in the menu "View Results")
2. Look for the error code in the GeneXpert user manual and follow the recommended
corrective actions.
3. If the problem remains, contact your local service provider (if applicable) or Cepheid
assistance
Service Telephone Email address
4. Continue to use other modules in the meantime (if possible) and exclude faulty
module(s) from tests (see slides 8 & 9)
2 © Cepheid – Confidential.
Where to find Error messages?
3 © Cepheid – Confidential.
Where to find Error messages?
4 © Cepheid – Confidential.
Troubleshooting section in Operator Manual
5 © Cepheid – Confidential.
How to obtain System Log Report
6 © Cepheid – Confidential.
How to obtain System Log Report
7 © Cepheid – Confidential.
How to Exclude Module(s) from Tests
8 © Cepheid – Confidential.
How to Exclude Module(s) from Tests
4 Click on « OK »
9 © Cepheid – Confidential.
Modules not detected
Solutions:
- Does this error affect only one module? Always the same?
- Check the internal temperature of the modules (in the Maintenance menu)
- Electrical failure
Solutions:
1 – try to remove the cartridge from the software
- Select the module. Click “Open Door” to open the module door.
- If the door does not open, restart the system and repeat the above steps.
If above steps did not solve the issue, contact Cepheid in order to be guided on how to remove module manually (detailed
procedure and online support will be provided)
12 © Cepheid – Confidential.
Error Messages
Cause
Many different causes can lead to an
ERROR result. Click on Errors to know
more about the specific issue
Origin(s)
Most frequent issues, linked to sampling,
are detailed in the next slides. They should
be addressed by operators following the
advices contained in this document
Solution(s)
To easily reduce an unexpectedly high Error rate, it is essential that all operators identify errors linked to
sample preparation
13 © Cepheid – Confidential.
Results with Errors:
INVALID
Problem
SPC (Internal Control) failed
Origin
PCR was inhibited due to inhibitors
(pus, food particles, …)
Solution
Before mixing with Cepheid sample reagent (SR) for decontamination, check that
sample does not contain food particles, pus ...
Collect a new sample when necessary
14 © Cepheid – Confidential.
Results with Errors:
NO RESULT
Problem
Test could not be terminated and
Insufficient data were collected
Origin(s)
- Power failure during test
- “Stop Test” function was used.
(accidentally or deliberately)
- Computer freeze or crash during test
Solution
Secure the power supply.
Use “Stop Test” only when it’s necessary
Do not open other applications on the computer during tests
15 © Cepheid – Confidential.
Error Messages:
Probe check failed: 5006/5007
Problem
Probe Check control failed and test was
stopped before amplification
Origin(s)
- Incorrect processing of sample (viscosity)
- Incorrect sample volume
- Improper fluid transfer (bubbles)
- Incorrect storage of cartridges (Probe
integrity issues detected)
Solution(s)
- Make sure the sample is totally liquefied before transferring to the cartridges:
If after 15 minutes of incubation with Sample Reagent, sample is still too viscous do not load it into the
cartridge. Wait up to 10 more minutes.
- Add from 2 to 4 mL of the preparation in the cartridges
- Avoid making bubbles
- Store the kits between 2 to 28°C
16 © Cepheid – Confidential.
Error Messages:
Abnormal Pressure detected: 2008
Problem
Pressure reading exceeds the maximum
(2008)
Origin
- Filter of cartridge is clogged (due to too
viscous sample or particles)
- Pressure sensor failed
Solution(s)
- Make sure the sample is totally liquefied before transferring to the cartridges:
if after 15 minutes of incubation with Sample Reagent, sample is still too viscous do not load it into the
cartridge. Wait up to 10 more minutes.
- Make sure sample does not contain any solid particles
- Use a new cartridge and add only Sample Reagent, if problem persists it’s likely a module
problem: contact Cepheid
17 © Cepheid – Confidential.
Error Messages:
Module communication loss was detected: 2127
Problem
Communication is lost between modules and
software
Origin
- Ethernet connection between PC and GX is bad
- Power supply issue (main power or UPS fluctuations)
- Bad connection points between Gateway Board
and modules
- Room temperature is too high
Solution(s)
- Unplug and replug Ethernet cable between PC and instrument, restart the system
- Unplug and replug the communication cable between gateway board and GeneXpert. Restart system.
- Secure Power supply and use adapted UPS/Surge protector
- Check Room temperature (should be below 39°C)
18 © Cepheid – Confidential.
Error Messages:
Signal loss detected in the amplification curve: 5011
Cause
Signal loss detected in the amplification
curve.
Origin
Loss of tube pressure because the
cartridge tube is not airtight, or cartridge
valve is not working right
Solution(s)
- Make sure there is no bubbles in the reaction tube
- Use a new cartridge. If error repeats, this can be module related: modules fluidics are not
working well. Contact Cepheid support team.
19 © Cepheid – Confidential.
Cepheid Assistance & Support
When Contacting Cepheid, please prepare: The Serial number of the GeneXpert, the recorded error
messages, the description of the incident and when possible the archived runs concerned
Serial number of your GeneXpert (at the back side, near the power button)
21 © Cepheid – Confidential.