RM HC103 en
RM HC103 en
RM HC103 en
HC103
Table of Contents
1 Introduction 5
1.1 Further documents and information................................................................................ 5
1.2 Explanation of conventions and symbols used ................................................................ 5
1.3 Compliance information................................................................................................ 6
2 Safety Information 7
2.1 Definition of signal words and warnings symbols ............................................................ 7
2.2 Product-specific safety notes ......................................................................................... 7
6 Operation 20
6.1 Operation orientation .................................................................................................... 20
6.2 User interface .............................................................................................................. 20
6.2.1 Operation keys.............................................................................................. 20
6.2.2 Screen navigation ......................................................................................... 20
6.2.3 Basic elements on the touch screen ................................................................ 21
6.2.4 Icons and buttons ......................................................................................... 21
6.2.4.1 Title bar icons ......................................................................................... 21
6.2.4.2 Action bar buttons ................................................................................... 21
6.2.4.3 Status message icons.............................................................................. 21
6.2.5 Home screen ................................................................................................ 22
6.2.6 Work screen ................................................................................................. 23
6.2.7 Input dialogs ................................................................................................ 23
6.2.7.1 Entering characters and numbers .............................................................. 23
6.2.7.2 Entering numerical values ........................................................................ 24
6.2.7.3 Changing date and time........................................................................... 25
6.2.8 Lists and tables ............................................................................................ 25
6.3 Performing a simple measurement................................................................................. 26
9 Troubleshooting 79
9.1 Error messages ........................................................................................................... 79
9.2 Error symptoms ........................................................................................................... 79
10 Technical Data 82
10.1 General data ............................................................................................................... 82
10.2 Model-specific data...................................................................................................... 83
10.3 Dimensions................................................................................................................. 84
10.4 Interfaces specification ................................................................................................. 85
10.4.1 RS232C....................................................................................................... 85
10.4.2 USB host...................................................................................................... 85
10.4.3 USB device................................................................................................... 86
12 Disposal 92
Index 93
For further questions, please contact your authorized METTLER TOLEDO dealer or service representative.
u www.mt.com/contact
Elements of instructions
In this manual, step-by-step instructions are presented as follows. The action steps are numbered and can
contain prerequisites, intermediate results and results, as shown in the example. Sequences with less than two
steps are not numbered.
Prerequisites that must be fulfilled before the individual steps can be executed.
1 Step 1
Intermediate result
2 Step 2
Result
WARNING A hazardous situation with medium risk, possibly resulting in death or severe injury if
not avoided.
CAUTION A hazardous situation with low risk, resulting in minor or moderate injury if not avoided.
NOTICE A hazardous situation with low risk, resulting in damage to the instrument, other
material damage, malfunctions and erroneous results, or loss of data.
Warning symbols
General hazard: read the User Manual or the Reference Manual for information about the hazards
and the resulting measures.
Goggles
Safety notes
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to injury and death. If the instrument
cannot be shut down in an emergency situation, people can be injured and the instrument
can be damaged.
1 Check that the voltage printed on the instrument is the same as your local power supply
voltage. If this is not the case, under no circumstances connect the instrument to the
power supply, but contact a METTLER TOLEDO representative.
2 Only use the three-core power cable with equipment grounding conductor supplied by
METTLER TOLEDO to connect your instrument.
3 Only connect it to a three-pin power socket with earthing contact.
4 Only use standardized extension cables with equipment grounding conductor for
operating the instrument.
5 Make sure that the power plug is accessible at all times.
6 Arrange the cables so that they cannot be damaged or interfere with the operation.
7 Keep all electrical cables and connections away from liquids.
WARNING
Injury or death due to toxic or corrosive substances
Heating up toxic or corrosive substances, e.g., acids, can result in toxic or corrosive vapors
that can cause injuries if they come in contact with the skin or the eyes or if they are
inhaled.
1 When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
2 Set up the instrument in a well-ventilated location.
3 When using dry substances that form toxic gases, place the instrument in a fume hood.
WARNING
Death or serious injuries due to flammable solvents
Flammable solvents in the vicinity of the instrument can ignite and lead to fire and
explosions.
1 Keep flammable solvents away from the instrument.
2 When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
NOTICE
Damage to the instrument due to corrosive substances and vapors
Corrosive substances and corrosive vapors can damage the instrument.
1 When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
2 Ensure that the instrument parts touching your sample substance can not get altered by
it.
3 Wipe off any condensation of corrosive vapours after an operation.
4 Work with small samples.
NOTICE
Damage to the instrument or malfunction due to the use of unsuitable parts
Only use parts from METTLER TOLEDO that are intended to be used with your
instrument.
3.1.1 Overview
Legend
1 Sample chamber 2 Level indicator
3 Sample pan handler 4 Touch screen
5 Leveling screw 6 Sample pan holder
7 Draft shield element 8 Contacts for temperature adjustment kit
(optional)
9 Heating module (with thermal overload 10 Reflector ring
protection)
11 Temperature sensor 12 Protective glass
13 Handles for opening the sample chamber 14 USB Host 1
15 USB Host 2 16 USB Device
17 RS232C 18 Kensington slot for anti-theft purposes
19 RS232C (for production only) 20 Power line fuse
21 Power supply socket 22 Fan
1 2 3 4
6 7 5
1 Model designation 2 Readability
3 Maximum capacity 4 Manufacturer
5 Power supply 6 Year of manufacture
7 Serial number (SNR)
120 °C
With halogen heating technology, maximum heating power is quickly reached and allows use of high temper-
atures. Uniform heating of the sample material ensures repeatability of the drying results and makes it possible
to use small amounts of sample.
The METTLER TOLEDO Moisture portfolio comprises a range of moisture analyzers that differ from each other in
both hardware and software.
The following features are available for the HC103 model:
Robust and compact design, ideal for limited laboratory space.
Intuitive operation thanks to a large touchscreen and easy to navigate user interface. See [User interface
Page 20]
Quick and easy cleaning. See [Cleaning Page 74]
Quick access to methods and results with shortcuts on the home screen. See [Working with shortcuts
Page 44]
Easy reporting. See [Exporting results Page 48]
ID management. See [Identifications management Page 63]
Setup wizard and instrument tutorial. See [Setup wizard Page 13] and [Help and tutorials Page 73]
Avoid direct sunlight Avoid vibrations Avoid strong drafts Avoid temperature fluctu-
ations
Sufficient distance for Moisture Analyzers: > 15 cm next to the instrument, > 1 m above the heating module.
4.2 Unpacking
Note
Store all parts of the packaging. This packaging guarantees the best possible protection for the transport of your
instrument.
After receiving the instrument, proceed as follows:
1 Open the package and remove drying unit and the accessories.
2 Remove the packaging from the instrument.
3 Check the Moisture Analyzer for transport damage and complete scope of delivery.
4 In the event of damage or missing accessories, immediately inform a METTLER TOLEDO representative.
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to injury and death. If the instrument
cannot be shut down in an emergency situation, people can be injured and the instrument
can be damaged.
1 Check that the voltage printed on the instrument is the same as your local power supply
voltage. If this is not the case, under no circumstances connect the instrument to the
power supply, but contact a METTLER TOLEDO representative.
2 Only use the three-core power cable with equipment grounding conductor supplied by
METTLER TOLEDO to connect your instrument.
3 Only connect it to a three-pin power socket with earthing contact.
4 Only use standardized extension cables with equipment grounding conductor for
operating the instrument.
5 Make sure that the power plug is accessible at all times.
6 Arrange the cables so that they cannot be damaged or interfere with the operation.
7 Keep all electrical cables and connections away from liquids.
Two different versions of drying units with country-specific power cable are available:
110 V AC or 230 V AC
Instrument is at the final location.
1 Connect the power cable to the power supply socket on the instrument.
2 Connect the power cable to the power supply.
See also
Drying unit Page 10
See also
Connecting the instrument Page 13
Test/Adjust Page 50
Requirements
PC with one of the following Microsoft Windows® 32-bit/64-bit operating systems: Win 7 (SP1), Win 8 or
Win 10.
USB cable to connect the instrument to the PC.
Downloading and installing the USB driver
1 Open the download portal (see above) on your PC.
2 Download the driver software and extract the .zip file.
3 Right-click on the extracted .exe installation program and select Run as Administrator.
4 If a safety warning appears, allow windows to perform the installation.
5 Click Next and follow the installer's instructions.
Connecting the instrument to the PC
1 Connect the moisture analyzer to the preferred USB port on the PC.
2 Follow the instructions on the PC and install the software automatically (recommended).
3 Check that the instrument is listed as a device (COM Port) on your PC.
4 Set USB interface in the peripheral settings of the instrument.
ON/OFF Home
USB host
56
Strip printer
Workstation
56
USB device
076,&6
Note
The USB host connection can only be used with specific devices. The following devices are officially supported:
USB memory stick (for specific characteristics see list below)
A bar code reader can be connected to the USB port. With a connected bar code reader, bar codes can be
added in a keyboard dialog.
Note for using USB memory stick
USB memory sticks up to 32 GB are supported.
Recommended formatting: FAT32 (NTFS not supported).
Max. cluster size: 32 KB.
Max. file size: 32 MB.
NOTICE
Damage to the touch screen due to pointed or sharp objects
Operate the touch screen with your fingers.
Note
The sensitivity of the touch screen is set to a default value. The sensitivity of the touch screen can be defined
with the function Touch screen adjustment in the section Settings.
The surface navigation of the touch screen works in the same way as on most common touch screens:
Selecting a button or an icon
Tap on it.
Scrolling
Drag the button in the scroll bar up or down.
Or tap or
Name Explanation
1 Status message Show information about the instrument status. The quick access icon (for settings
icons of brightness and sound) is displayed continuously. See Status Message Icons
2 Status bar Shows the user name, status icons as well as date and time.
3 Title bar Shows title of the current screen, contains special functions. See [Title bar icons
Page 21].
4 Content area Main working area of the menus and the applications. Shows content depending
on application or performed action. May also show graphic illustrations such as
the curve diagram of a data set (e.g. drying curve).
5 Action bar Contains action buttons which are required and available in the current dialog.
See [Action bar buttons Page 21].
Name Explanation
1 Main Menu Measurement
Start measurement.
Prerequisite:
Method is defined and all necessary settings are made.
Results
Display, print and export results
Method Definition
Define, edit, test or delete a method
Test/Adjust
Adjust or test the integrated balance and the heating module and perform
SmartCal tests.
Settings
Define instrument-, user-, and data-management settings. Find help and tutorials
under this menu item.
2 User shortcuts Shows user-specific shortcuts for frequently used methods. Shortcuts are saved in
the user profile.
3 Instrument info Shows general info about the instrument and the software (e.g. serial number,
software version).
5 3
Name Explanation
1 Shortcut button Adding/editing a shortcut to the current method (to the home screen).
2 Value panel Current measured (or predicted) values of the work process.
3 Parameter panel Parameters of the current work process. A detailed overview of the method
parameters appears after tapping the parameter panel.
Identifications (IDs) appear after tapping the ID panel for entering or editing
values (comments). ID panel appears only if identification input is activated in
the menu.
4 Action buttons Action buttons which are required and available for the current dialog (e.g. Back,
->0/T<-, Print, Save, Delete, OK).
5 Graphics panel Graphical illustrations, e.g. of the drying curves, instructions for the user for
performing tasks, and weighing-in aid.
1
2
Name Explanation
1 Input field Shows the data that has been entered.
2 Explanation field Indicates the maximum number of characters allowed.
3 Shift Switches the keyboard between standard and special characters.
Cursor left
Cursor right
Note
It is also possible to place cursor directly in input field by tapping the respective position.
2
3
Name Explanation
1 Switch If a switch appears, located on the left side in the title bar; it must be switched on
for the entry of a value: O = OFF, I = ON
2 Input field Shows the data that has been entered.
3 Explanation field The valid input range is indicated.
Delete button Deletes last character.
1
2
Name Explanation
1 Pick button Increases the value
2 Explanation field Shows the current value
3 Pick button Reduces the value
Confirm with [Set date]. (To exit without changing, tap [Cancel].)
1 3
4
2 5
Name Explanation
1 Content title Shows the subject of the current content
2 Content area Shows the relevant content
3 Arrow icon [ ] Shows the list in ascending order.
[ ] Shows the list in descending order.
4 Scrollbar [ ] Scroll up
5 Scroller
6 [ ] Scroll down
Note
CAUTION
Burns due to hot surfaces!
Parts of the instrument can reach temperatures that cause injuries if touched.
Sample, sample pan and other parts inside the sample chamber may still be hot.
Do not touch the place marked with the warning symbol.
u www.mt.com/moisture-methods
Standard drying
Standard drying
This drying program is set in the factory and is suitable for most samples. The sample is
heated to the drying temperature.
Factory setting
Drying temperature = 105 °C, Switch-off criterion 3 = 1 mg / 50 s
Rapid drying
Rapid drying
This drying program is primarily suitable for samples with a moisture content over 30%.
Following the start, the selected temperature is exceeded by 40% for 3 minutes (the
maximum heating temperature is 230 °C) to compensate for the cooling due to vapor-
ization and to accelerate the drying process. The drying temperature is then lowered to the
set value and maintained.
Factory setting
Drying temperature = 105 °C, Switch-off criterion 3 = 1 mg / 50 s
The graph below exemplifies the mode of operation of the switch-off operation (not to scale).
1) 2)
∆g
= = ( )
∆t
∆g
∆t
(10 s)
∆g
∆t
(50 s) ∆g
∆t
(140 s)
1 (1 mg / 10 s) 3 (1 mg / 50 s) 5 (1 mg / 140 s) t
Detailed information:
%MC – Moisture Content
The moisture content of the sample is displayed (and printed out) as a
percentage of the wet weight (WW = initial weight = 100%). This is the
MC
factory setting. WW
DW
During the measurement the value is constantly displayed in percent and
as a graphical drying curve. The measured value is marked by Before drying After drying
"%MC" (Moisture Content, e.g. 11.35 %MC) also for the printed results.
MC = WW – DW · 100 %
WW
AM = WW – DW · 100 %
DW
AD = WW · 100 %
DW
MC = – WW – DW · 100 %
WW
Automatic
Your instrument is set in the factory to the automatic operating mode. This mode can be used for most kind of
samples. When closing the sample chamber, the weight of the sample is recorded and the measurement is
started.
Manual
We advise you to use the manual operating mode for samples which contain readily volatile substances. The
initial weight (wet weight) important for the determination of the moisture content is recorded. In the manual
operating mode you have time for further preparation of the sample (e.g. mixing with quartz sand or even
distribution of the sample) while weight losses due to evaporation during the preparation time are measured
from the outset. As soon as the sample is ready for drying, close the sample chamber. As soon as the sample
chamber is closed, drying starts. In the manual operating mode you can open the sample chamber during a
drying operation. In contrast to the automatic mode, drying will not be stopped but simply interrupted until the
sample chamber is closed again.
7.1.1.3.2 Accessories
Navigation: Home > Method Definition > Method name > Workflow handling > Accessories
For a method, certain accessories can be specified that must be used for the measurements. During the
measurement, these accessories are then requested on the work screen.
The following parameters can be defined:
Parameter Description Values
Accessories Defines the main accessory (pan) in which the sample is Sample pan* | HA-Cage
prepared.
Additional acces- Defines additional accessories to be used with the sample pan. (deactivated)* |
sories (only (activated)
available if the
sample pan is the
main accessory:
Glass fibre
filter
Second glass
fibre filter
Downholder
* Factory setting
7.2 Measurement
Navigation: Home > Measurement
Introduction
Before a measurement can be performed, a method first needs to be defined. To define a suitable method for
your sample, see [Method definition Page 30].
The instrument guides you through the steps during a measurement and displays results in real time. Never-
theless, please note the instructions in this manual, especially at the beginning of your work with the Moisture
Analyzer, as a correct execution of the work steps and the handling of the samples can have a great influence
on the results.
After taring, the display prompts you to add the sample to the sample pan.
1 Add the sample to the sample pan. If you have defined
a start weight, weigh the sample using the weighing-in
aid.
2 Close the sample chamber.
3 Tap Start drying.
The drying process starts.
Note
The maximal drying time is 480 minutes (8 hours).
Removing sample
CAUTION
Burns due to hot surfaces!
Parts of the instrument can reach temperatures that cause injuries if touched.
Sample, sample pan and other parts inside the sample chamber may still be hot.
Do not touch the place marked with the warning symbol.
Stop drying
When you stop the measuring process by tapping [Stop drying], you can choose between two options:
Abort measurement and save data
Already sampled data are stored and an entry will be made in the results. The result is marked as aborted.
Cancel
The process will be continued.
Adding a comment
At the end of a measurement you can add a comment to the result of the measurement. This comment is stored
in the measurement result and can be printed. A comment can only be entered before leaving the current
measurement.
1 To create a comment, tap Note.
The keyboard dialog appears.
Icon Function
Lower weight limit (tolerance range)
Target weight
7.3 Results
Navigation: Home > Results
The device can save a total of 100 measurement results. When there are more than 100 results, the oldest
result will automatically be overwritten. Therefore we recommend to safe the results frequently.
Method list
1 Method list
Tap the method you want to evaluate.
The journal view appears.
2 Journal view
The journal view allows you to start different graphical evaluations from a measurement series. You can
perform the following functions:
To start filter function, tap .
The filter menu appears.
To start chart view, tap Chart.
The chart view of the series of measurement appears (4).
To start graphic view, tap a measurement result.
The graphic view appears (5).
See also
Peripherals Page 61
7.4 Test/Adjust
Navigation: Home > Test/Adjust
Introduction
For optimum precision and repeatable results of the Moisture Analyzer, the function of the instrument should be
tested on a regular basis and, if necessary, adjusted. The frequency of testing and adjusting depends on the
use and location of the instrument.
Main menu Submenu Further information
Adjustments Weight adjustment - internal see Weight Adjustment - Internal
Weight adjustment - external see Weight Adjustment - External
Temperature adjustment see Temperature Adjustment
7.4.1 Tests
Navigation: Home > Test/Adjust > Tests
In this menu item you can check several settings, such as the adjustment of the balance or the temperature of
the heating module. For relevant settings see [Test / adjustment settings Page 66].
Introduction
Routine tests are performed to ensure accurate results. Therefore the tests should be performed in fixed, regular
intervals depending on your Quality Management system and the results should be documented in a traceable
way.
Based on your process requirements, METTLER TOLEDO can help you to define the routine tests to be
performed. For further information, contact your local METTLER TOLEDO representative.
CAUTION
Burns due to hot surfaces!
Parts of the instrument can reach temperatures that cause injuries if touched.
1 Let the instrument cool down before setting up.
2 Hold the temperature kit by its handle only.
1
3
Testing procedure
Test settings are defined, see Test Settings.
Temperature kit is installed.
1 Tap Temperature test.
Work screen Temperature test appears.
2 Close the sample chamber.
3 Tap [Start test].
The testing process starts: The instrument first heats up to target temperature 1 (lower) for 15 minutes,
then, for another 15 minutes, heats further up to target temperature 2 (higher). The screen shows the
target temperature values (Target), measured at the sensor, and the actual temperature (Actual),
measured at the temperature kit.
The instrument indicates whether the test has passed or failed according to defined tolerances.
4 If an adjustment is needed, decide if you want to perform it now, see Temperature adjustment.
5 • To return to the menu Tests, tap [Tests].
• To print the results, tap [ ].
• To display or print the results at any time, tap [History].
• To return to the home screen, press [ ].
6 Let the instrument cool down before removing the temperature kit.
See also
Accessories and Spare Parts Page 87
Test settings Page 67
Temperature adjustment Page 56
Accessories Page 87
Timeout
In case of a timeout of over 75 seconds, the error message appears Weight out of range. The test has
been aborted. The test was not executed.
1 Confirm with [OK].
2 Repeat the test.
See also
Accessories Page 87
Testing procedure
The test is carried out in the same way as a measurement.
Note
Every SmartCal test has a maximum number of 30 test results. If there are more than 30 results for one test,
the oldest results will automatically be deleted.
The instrument has been connected to the power supply for at least 60 minutes in order to reach operating
conditions.
If available, a RHT sensor is connected.
The SmartCal package is acclimatized.
Test temperature and control limits are set.
The sample chamber is open.
1 Tap [SmartCal x1 test].
The method list is shown on display.
2 Select a method from the list.
The work screen appears.
3 If the test is performed without RHT sensor, tap [Enter room temperature and humidity], enter current room
temperature and humidity and confirm with [OK].
4 Enter SmartCalTM lot number if needed. Tap [OK]
5 Place the sample pan handler with the aluminum sample pan in the instrument.
6 Tap [->0/T<-] to tare the balance.
7 Remove a SmartCalTM stick from the blister pack, open it and distribute the entire contents evenly over the
sample pan. If necessary, carefully rotate and tilt the sample pan until it is fully and uniformly covered with
granulate.
8 Start the test immediately by tapping [Start drying].
The SmartCal test procedure starts.
When the test is finished, the instrument indicates whether the test has passed or failed. The display
shows the normalized result. If the SmartCal result has failed, see [Measures after failed test
Page 55].
9 • To return to the menu Tests, tap [Tests].
• To print the results, tap [ ].
• To display or print the results at any time, tap [History].
• To return to the home screen, press [ ].
7.4.2 Adjustments
Navigation: Home > Test/Adjust > Adjustments
Introduction
To guarantee correct results and high performance of the instrument, it is necessary to adjust its functions:
After setting up the balance for the first time.
After a change of location.
If a test indicates that an adjustment is due.
Test the instrument exclusively under operating conditions.
METTLER TOLEDO offers an adjustment service – please contact your METTLER TOLEDO representative.
See also
Test / adjustment settings Page 66
Timeout
In case of a timeout of over 75 seconds, the error message appears Weight out of range. Adjustment
aborted. The test was not executed.
1 Confirm with [OK].
2 Repeat the adjustment.
See also
Accessories Page 87
CAUTION
Burns due to hot surfaces!
Parts of the instrument can reach temperatures that cause injuries if touched.
1 Let the instrument cool down before setting up.
2 Hold the temperature kit by its handle only.
1
3
Start adjustment
Draft shield element is installed.
Sample pan handler and sample pan holder is removed.
Temperature kit is inserted.
1 Close sample chamber.
2 Tap [Start adjustment].
The adjustment process takes 30 minutes.
The instrument starts the adjustment process. The screen shows the current temperature and the
remaining time.
The instrument heats up to the target temperature 1 for determining the predefined lower temperature.
After 15 minutes, the instrument adjusts the lower temperature and continues with the process.
The instrument heats up to the target temperature 2 for determining the predefined upper temperature.
After 15 minutes, the instrument adjusts the upper temperature and concludes the adjustment.
3 • To print out the results, tap .
• To return to the home screen, tap .
4 After cooling down, remove Temperature kit and set up drying unit for measurement.
See [History Page 58] and [Setting up the instrument Page 14].
Note
Your calibrated temperature kit can be re-calibrated. Please contact your local METTLER TOLEDO representative
for more information.
7.5 Settings
Navigation: Home > Settings
In this section you will find information how to define the general settings of your instrument.
Main menu Submenu Further information
User preferences Languages see [Languages Page 59]
Screen see [Screen Page 60]
Sound see [Sound Page 60]
Protection see [Protection Page 60]
Instrument settings Regional Settings see [Regional settings Page 61]
Peripherals see [Peripherals Page 61]
Touch screen adjustment see [Touch screen adjustment
Page 63]
Application settings Identifications management see Identifications management
Print and export management see [Print and export management
Page 64]
Quality management Test / Adjustment settings see [Test / adjustment settings
Page 66]
System and data management Export / Import see Export / import
Back up / Restore see [Back up / Restore Page 71]
Export event logs see [Export event logs Page 72]
Reset see Reset
Update see Update
Help and tutorials Instrument tutorial see Instrument tutorial
7.5.1.1 Languages
Navigation: Home > Settings > User preferences > Languages
7.5.1.2 Screen
Navigation: Home > Settings > User preferences > Screen
The following parameters can be defined:
7.5.1.3 Sound
Navigation: Home > Settings > User preferences > Sound
The following parameters can be defined:
7.5.1.4 Protection
Navigation: Home > Settings > User preferences > Protection
The following parameters can be defined:
7.5.2.2 Peripherals
Navigation: Home > Settings > Instrument settings > Peripherals
In this menu item you can configure one printer and one host connection.
The MT-SICS host connection (remote client) is enabled by default and only usable with the METTLER TOLEDO
specific protocol MT-SICS.
MT-SICS commands and functions can be transmitted via the RS232 interface (COM1 port) or the USB device
interface. The use of the USB device interface requires the installation of an USB device driver on the PC, see
[Installing USB device driver Page 17].
Commands and further information please refer to the Reference Manual MT-SICS Interface Commands for
Halogen Moisture Analyzers downloadable from the internet.
uwww.mt.com/moisture
If the alignment of the touch-sensitive areas of the screen correspond not exactly with the position of the buttons
e.g. after a software update was performed, then you can solve this error by executing this function.
NOTICE
Damage to the touch screen due to pointed or sharp objects
Operate the touch screen with your fingers.
Customized printouts
With the options Custom 1 and Custom 2 the information that appear on the printout can be selected
individually. The information that appear on the printout can be defined in the sub-option Printout content.
The following information for Printout content can be enabled/disabled or defined:
Signature 2 (deactivated)* |
(activated)
Note
The information Date / Time and the end result are always activated and appear at the end of the printout.
The settings for the option Intermediate results are still set but do only appear on short and standard printout.
Temperature kit
The following parameters can be defined:
Temperature test
The following parameters can be defined:
SmartCal test
Note
There are 4 fixed SmartCal methods and 2 customizable SmartCal methods. All 6 methods are activated as
default values but the methods can individually be enabled or disabled.
The following parameters can be defined:
* Factory setting
Temperature adjustment
The following parameters can be defined:
7.5.5.4 Reset
Navigation: Home > Settings > System and data management > Reset
This menu item allows you to reset the Instrument to factory settings. Date, time, weight and temperature
adjustments/history are not affected.
NOTICE
Data loss after system reset.
After resetting, the instrument is in factory default state. All data such as settings, methods
or results is lost.
Save the data on an external storage device before you reset the instrument, see
[Export / Import Page 70] and [Back up / Restore Page 71]
7.5.5.5 Update
Navigation: Home > Settings > System and data management > Update
The Update menu item is only used during the software update process, see [Updating the software
Page 77].
Note
This function is not available on approved instruments.
Replacing the power line If the fuse is blown see chapter "Replacing the
fuse power line fuse"
Updating the software If recommended by METTLER TOLEDO see chapter "Updating the
software"
8.2 Cleaning
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to injury and death. If the instrument
cannot be shut down in an emergency situation, people can be injured or the instrument
can be damaged.
Disconnect the instrument from the power supply before cleaning or performing other
maintenance work.
CAUTION
Burns due to hot surfaces
The interior parts of the heating module as well as the parts in the sample chamber can
reach temperatures that can cause injuries if touched.
Wait until the heating module has cooled down completely before performing any
maintenance task.
Protective equipment:
Gloves
Goggles
See also
Temperature adjustment Page 56
CAUTION
Burns due to hot surfaces
The round halogen lamp can reach temperatures that can cause injuries if touched.
1 Do not remove the halogen lamp.
2 Wait until the heating module has cooled down completely before cleaning.
3 Remove any splashes, deposits, or spots from the halogen lamp with a mild organic
solvent such as ethanol.
WARNING
Danger of death or serious injury due to electric shock!
Risk of electrical shock or damage of the instrument when changing fuses.
1 Do not use a fuse of a different type or rated value.
2 The short-circuiting (bridging) of fuses is not allowed.
The power line fuse is located on the back of the drying unit. To change the fuse, proceed as follows:
1 Pull out the power cord.
2 Slide out the fuse holder (1) with a suitable tool such as a screwdriver.
3 Remove the fuse (3) and check its condition.
4 If the fuse is blown, replace the fuse with one of the same type and the same rated value.
Note
A spare fuse is located in the holder (2).
Update procedure
Backup of the current data is created, see [Back up / Restore Page 71]
Recent software update is downloaded and stored on an external device, e.g., USB stick or SD card.
Administrator rights are active.
1 Insert device with software for update into its respective port.
2 Wait until the external device is recognized by the instrument.
3 Select Home > Settings > System and data management > Update. If Update is not visible check if
administrator rights are active. If you have an approved version, contact your support representative.
Update screen appears.
4 Select the device on which the software is stored.
5 Confirm with [Update] and wait until the software update is completely finished. Do not remove the device
before the instrument has completely restarted.
The instrument reboots.
A message appears: Operating system is up to date. Update instrument software?...
6 Confirm with [Update].
7 Wait until the system returns to the home screen.
Software downgrade
It is not recommended to downgrade the software. However, in case of a setup that requires a software
downgrade in order to work, please observe the following safety instructions.
NOTICE
Data loss during software downgrade
During a downgrade of the software all device data such as drying methods, results,
SmartCal history, and settings will be removed.
Only let a service technician perform a software downgrade.
Environmental conditions
Height above sea level up to 4000 m
Ambient temperature range Operation: +10 °C to 30 °C
(operability guaranteed 5 °C to 40 °C)
Relative air humidity max. 80% up to 31 °C, linearly decreasing to 50% at 40 °C
20% - 80% and non-condensing conditions
Warm-up time At least 60 minutes after connecting the instrument to the power
supply; when switched on from standby, the instrument is ready
for operation immediately.
Materials
Drying Unit
Housing PBT, PBX45A (UL 94 V-0)
Inspection window grill PPS, A504X90 (U L94 V-0)
Protective glass Quartz glass
Halogen lamp Quartz glass
Reflector bracket PPS A504X90 (UL 94 V-0)
Draft shield, interior bottom plate Stainless steel, X2CrNiMo17-12-2 (1.4404)
Balance
Maximum capacity 101 g
Readability 1 mg
Minimum sample weight 0.5 g
Weighing technology Strain gauge
Adjustment External weight (100 g, accessory)
Moisture Content
Readability 0.01%
Repeatability (sd) with 2 g sample 0.10%
Repeatability (sd) with 10 g sample 0.015%
Interfaces
1 × RS232C (for production only)
1 × RS232C (9-pin socket)
2 × USB Host (Type A socket), USB 1.1
Supported memory sticks up to 32 GB
1 × USB Device (Type B socket), USB 1.1
Data
Export data format of results CSV (Comma-separated values)
PDF
Hardware
Drying unit
Sample chamber opening/closing manual
Leveling 2 leveling screws
Sample pan ø 90 mm
Maximum height of sample 25 mm
Thermal overload protection Bimetallic-element switch in heating module
Dimensions (w × h × d) 192 × 169/370 × 385
Weight, ready to measure 5.1 kg
Display WVGA color display, resistive touch-sensitive
370
169
103
70 297
385
402
192
144
10.4.1 RS232C
Schematic Item Specification
Interface type Voltage interface according to EIA RS-232C/
DIN66020 CCITT V24/V.28)
DATA
Max. cable length 15 m
RxD Signal level Outputs:
IN
TxD
+5 V … +15 V (RL = 3–7 kΩ)
OUT –5 V … –15 V (RL = 3–7 kΩ)
GND
Inputs:
+3 V … +25 V
–3 V … -25 V
5 1 Connector Sub-D, 9-pole, female
Operating mode Full duplex
9 6
HAND Transmission mode Bit-serial, asynchronous
SHA KE
Transmission code ASCII
CTS
IN
RTS Baud rates see setting options
OUT
Bits/parity see setting options
Stop bits see setting options
Handshake None, XON/XOFF, RTS/CTS (software selectable)
Quality Management
Certified adjustment weight, 100 g (F1) 30406400
Printers
RS-P25/01 (EMEA) printer with RS232C connection to 11124300
instrument
Paper roll (length: 20 m), set of 5 pcs 00072456
Paper roll (length: 13 m), self-adhesive, set of 11600388
3 pcs
Ribbon cartridge, black, set of 2 pcs 00065975
Anti-theft devices
Anti-theft cable with lock 11600361
Barcode readers
Corded USB barcode reader 30417466
Miscellaneous
Dust filter, set of 50 pcs 11113883
Software
EasyDirect Moisture (simple data collection) 30522676
EasyDirect
Moisture
4
5
T
Tables 25
Temperatur
Adjustment settings 69
Temperature
Switch-off criterion 32
Test 52
Test settings 68
Test 22
Balance 51
Heating device 51
Temperature 52
Weight external 53, 55
Tests
General 51
Text 23
Time 25
Touch screen adjustment 63
Troubleshooting 72
U
Unpacking 12
USB Driver
Installation 17
USB Memory stick
see memory stick 18, 70
User
Home screen 22
Interface 20
Shortcuts 22
Mettler-Toledo GmbH
Im Langacher 44
8606 Greifensee, Switzerland
www.mt.com/contact