Standard Operating Procedure (Biochemistry - Diasys RESPONS - 910)
Standard Operating Procedure (Biochemistry - Diasys RESPONS - 910)
Standard Operating Procedure (Biochemistry - Diasys RESPONS - 910)
PROCEDURE
(BIOCHEMISTRY – DIASYS
RESPONS – 910)
PREPARED AS PER ISO 15189:2012
ISSUE NO: 04
AMENDMENT SHEET
Sl. Page Section/ Date of Amendment Reasons Signature Signature
No. No. Clause Amendment Made of
No. of QM Laboratory
Director
01 All All 09.03.2020 Issue No: 03 Change in
has been automated
withdrawn analyser
and issue no
04 is issued in
the system
TABLE OF CONTENTS
LAST
CHAPTER SL. REVISION
TITLE REVISION PAGE No.
No. No. NO
MADE
LAST
CHAPTER SL. REVISION
TITLE REVISION PAGE No.
No. No. NO
MADE
AMYLASE
18 00 00 2–4
I LDH
20 00 00 5-7
MACHINE OPERATION
II 00 00 1–1
PROCEDURE
SAFETY PRECAUTION
V 00 00 1–1
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 1 of 1
a) Purpose of
examination
Quantitative in vitro determination of lactate dehydrogenase (LDH) in serum
or plasma on DiaSys respons 910.
LDH
Method
Optimized UV-test according to IFCC and DGKC
c) Performance characteristics
Test has a linearity range of 5 - 1200 U/L
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 2 of 7
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
No interference was observed by ascorbic acid up to 30 mg/dL, bilirubin up
to 40 mg/dL and lipemia up to 2000 mg/dL triglycerides. Hemolysis
interferes because LDH is released by erythrocytes.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 3 of 7
t) Reference:
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 4 of 7
a) Purpose of examination
Quantitative in-vitro determination of α-Amylase in serum or plasma in
diasys response-910
Method
ENZYMATIC PHOTOMETRIC
c) Performance characteristics
Test has a linearity range of 3-2000 U/L.
e) Patient preparation:
QMSP 15
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 5 of 7
i) Calibration procedure
Periodic hardware calibration is done by the manufacturer. Parameter
specific calibration with standards is done in the following cases:
1. L-J Chart shows rejection Multi QC rule.
2. Lot change of reagent.
3. EQAS result shows outliers.
All auxiliary equipments are calibrated as per NABL 112.
j) Procedural steps
Assay procedure is performed automatically by the instrument.
Operation of instrument
Sample is collected, processed as per requirement and then the examination
is carried out following the operating procedure as described in chapter –II.
l) Interferences
Icterus: Interference less than 10% up to 20 mg/dl bilirubin.
Hemolysis: Interference less than 10% up to 2.5 g/l haemoglobin.
Lipemia: Interference less than 5% up to 1000 mg/dl intralipid.
t) Reference
1. Principle and procedure:
a. Kit Instruction.
b. John Bernard Henry, M.D.
2. Biological reference Interval :
a. Kit Literature.
b. Ref: Jacques Wallach, Md
Interpretation of diagnostic test 8th edition page
26-27. Lippincort.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 7 of 7
RUN CONTROL
Diasys Respons-910
Two level controls are run once daily for all of the parameters under the scope of accreditation.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 1 of 4
Receive the sample and confirm Lot No and Expiry Date with reference values specific to the
machine.
Make a note of the storage temperature.
Controls are stored at 2-8oC in specified Biochemistry Refrigerator
Vials are kept at room temperature for a maximum period of 1 hour while running tests.
Record sample validity period after opening or reconstitution.
Prepare aliquots (if applicable) and store in recommended temperature.
Internal Controls are run on a daily basis immediately after the specific machines are switched
on and in perfect running mode. Only authorized personnel in each department run controls.
Values observed are recorded and plotted in L-J charts.
Observe for any violation of Multi QC Deviation Rule and proceed accordingly.
QC Results are reviewed & if any of the parameters run out of 2SD or do not follow any of the MULTI
QC RULES described below, then a QC Level is run for the second time. If the results still do not fall
within the expected range, then the test parameter is calibrated, QC checked & samples run. All the
QC values are checked and filed and reviewed by the Consultant Pathologist.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 2 of 4
13S: One control observation exceeding the mean ± 3s; primarily sensitive to random error.
22S: Two consecutive control observation exceeding the same mean plus 2 s or mean minus 2 s
limit; sensitive to systematic error.
10X: 10 consecutive control observations falling on one side of the mean (above or below, with no
other requirement on size of the deviations); sensitive to systematic error.
41s: Four consecutive values on either side of mean are within 1SD to 2SD (41S).
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 3 of 4
D. Correlations of results:
Different Tests when results indicate the same pathology are used to confirm test results when
necessary. Both inter and intradepartmental correlation is done. Pathologist reviews the record
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 4 of 4
CALIBRATION PROCEDURE
Diasys Respons-910
Machine is calibrated by the manufacturer as and when required
Individual parameter calibration is done in-house by calibrators having
international traceability.
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 1 of 1
Issue No. 04 Issue Date: 09.03.2020 Prepared By: QM Copy No. Page 1 of 1
Ref: Wallsch. J interpretation of diagnostic test 7th edition page 29-31. Lippincort
Issue No. 04 Issue Date: 09.03.2020 Prepared By: Copy No. Page 1 of 1