Development and Validation of RP-HPLC Method For Simultaneous Estimation of Ivermectin and Clorsulon in Ivercam Injection
Development and Validation of RP-HPLC Method For Simultaneous Estimation of Ivermectin and Clorsulon in Ivercam Injection
Development and Validation of RP-HPLC Method For Simultaneous Estimation of Ivermectin and Clorsulon in Ivercam Injection
Corresponding author
Vegad Kunjal L.
Assistant Professor,
Department of Pharmcognosy
Sharda school of Pharmacy, Gandhinagar, Pethapur
kunjalvegad11@gmail.com
9825867380
523
Please cite this article in press as Vegad Kunjal L. et al. Development and Validation of RP-HPLC Method for Simultaneous
Estimation of Ivermectin and Clorsulon in Ivercam Injection. Indo American Journal of Pharmaceutical Research.2017:7(08).
Copy right © 2017 This is an Open Access article distributed under the terms of the Indo American journal of Pharmaceutical
Page
Research, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
www.iajpr.com
Vol 7, Issue 08, 2017. Vegad Kunjal L. et al. ISSN NO: 2231-6876
INTRODUCTION
Ivermectin (IVR) is chemically 22, 23-dihydroavermectin B1a+ 22, 23-dihydroavermectin B1b (Figure: 1); is an extremely
potent semisynthetic derivative of an antinematodal principle obtained from Streptomyces avermitilis. It binds to glutamate-gated
chloride ion channels which occur in invertebrate nerve and muscle cells. This leads to an increase in the permeability of the cell
membrane to chloride ions with hyper polarization of the nerve or muscle cell, resulting in paralysis and death of the parasite [1-5].
Clorsulon (CLOR) is 4-amino-6-(1,2,2-trichloroethenyl)-benzene-1,3-disulfonamide (Figure: 2); this drug has good efficacy
against mature rather than immature flukes when is given orally or via subcutaneous injection [3, 4]. Clorsulon is specific against
Fasciola spp. used for cattle at dose of 7 mg/kg to effect at great extent against fluke aged 8 weeks or older but the suggested (not
registered) dose rate against immature flukes for sheep is 15 mg/kg [6-7].
Objective:
Development and validation of High performance liquid chromatography method for estimation of Ivermectin and Clorsulon in
injection .
Pure Samples
Clorsulon and Ivermectin working standard was kindly gifted by NGL Fine Chem ,Mumbai.
www.iajpr.com
Vol 7, Issue 08, 2017. Vegad Kunjal L. et al. ISSN NO: 2231-6876
Wavelength determination
The detection was carried out in the UV region and wavelength selected for detection was 234 nm in mobile phase. 10µg/ml
of Clorsulon and Ivermectin working standard solutions were separately prepared in methanol. The zero order overlain spectrums of
the prepared solutions were recorded from 200 to 400 nm in double beam UV-visible spectrophotometer (Shimadzu, model 1800).
The isoabsorptive point was found at 234 nm and it was selected as wavelength of determination for both the drugs. The overlain
spectrum of Clorsulon and Ivermectin was shown in Figure: 3.
Chromatographic conditions
The HPLC system consisted of Shimadzu LC-20A system equipped with model LC-20AT pump, SPD- 20AT Shimadzu UV-
Vis (Diode array) detector, Hamilton syringe and DGU-20A5 online degasser, and a Rheodyne injection valve. Peak areas were
integrated using a Spinchrom Software program. Experimental conditions were optimized on a BDS Hypersil C 18 column (5μ, 250 x
4.6 mm), Thermo scientific at room temperature using 0.1 % KH2PO4: Methanol (60:40V/V) as mobile phase. Mobile phase was
flowed at 1 mL/min and all chromatographic experiments were performed at room temperature (25 ˚C ± 2˚C).
Preparation of KH2PO4:
2.72 gm of phosphate buffer was accurately weighed and dissolved in 1000ml of water, and then pH 5.5 was adjusted using
1% O-phosphoric acid.
Page
Mobile Phase: Prepare a filtered and degassed mixture of 60 volumes of buffer preparation (pH 5.5) and 40 volumes of Methanol.
www.iajpr.com
Vol 7, Issue 08, 2017. Vegad Kunjal L. et al. ISSN NO: 2231-6876
Precision
a) Intra-day precision
For Intraday precision, it was carried out by preparing 3 replicates of 3 different concentrations, within the linearity range and
measuring the peak area of each solution on the same day. % RSD (% relative standard deviation) was calculated.
b) Inter-day precision
For Interday precision, it was carried out by preparing 3 replicates of 3 different concentrations, within the linearity range and
measuring the peak area of each solution on 3 different days. % RSD (% relative standard deviation) was calculated.
Accuracy
To a fixed amount of pre-analyzed sample of CLOR 25μg/ml and IVER 2.5μg/ml, increasing amount working standard
solution of CLOR (20 ,25 and 30 ug/ml) and IVER (2, 2.5, and 3μg/ml) were added in 10 ml volumetric flask and made up to mark
with mobile phase Samples were injected into system and analyzed as described. Calculate the mean % recovery from peak areas
obtained.
www.iajpr.com
Vol 7, Issue 08, 2017. Vegad Kunjal L. et al. ISSN NO: 2231-6876
LOQ = 10 × (σ/S)
Where,
σ = Standard deviation of the Y-intercepts of the 5 calibration curves.
S = the mean slope of the 5 calibration curves.
Robustness
For robustness evaluation of HPLC method a few parameters like flow rate, mobile phase composition and pH were
deliberately changed. Robustness of the method was done at the concentration level 50 μg/ml and 5 μg/ml for CLOR and IVER
respectively.
www.iajpr.com
Vol 7, Issue 08, 2017. Vegad Kunjal L. et al. ISSN NO: 2231-6876
Precision
Intra-day precision
The data for intraday precision of CLOR and IVER are presented in Table 5 and Table 6 respectively. The % R.S.D. for
Intra-day precision was found to be 0.11-0.15 % for CLOR and 0.201-0.203% for IVER.
Inter-day precision
The data for inter-day precision of CLOR and IVER are summarized in Table 7 and Table 8 respectively. The % R.S.D. for
inter-day precision was found to be 0.100-0.106 % CLOR and 0.19-0.20 % for IVER.
www.iajpr.com
Vol 7, Issue 08, 2017. Vegad Kunjal L. et al. ISSN NO: 2231-6876
Accuracy
Accuracy of the method was confirmed by recovery study from marketed formulation at three level of standard
addition. Percentage recovery for CLOR was 98.9-101.3%, while for IVER, it was found to be in range of 99.3-101.5%, The
results are shown in Table. 9 and Table. 10. Recovery greater than 98 % justifies the accuracy of the method.
Amt of sample (µg/ml) Amt of std. spiked(µg/ml) Avg.Amt. recovery %recovery ± SD %RSD
80% 25 20 19.96 99.82 ± 1.3 1.32
100% 25 25 24.93 99.73 ± 0.8 0.82
120% 25 30 29.91 99.72 ± 0.6 0.64
Clorsulon Ivermectin
LOD 0.61 0.06
LOQ 1.86 0.20
Robustness
The data for robustness are presented in Table 12 and 13.
www.iajpr.com
Vol 7, Issue 08, 2017. Vegad Kunjal L. et al. ISSN NO: 2231-6876
Sr.no Label claim (mg) Result (w/w) %Assay Avg %Assay SD %RSD
1 100 99.91 99.91
Clorsulon
2 100 100.11 100.11 99.97 0.12 0.12
50µg/ml
3 100 99.88 99.88
Ivermectin 1 10 9.99 99.91
5µg/ml 2 10 10.01 100.11 100.11 0.19 0.19
3 10 10.03 100.31
Summary of Parameters
Table 15. Summary of parameters.
Result
Parameters
CLOR IVER
Linearity 0.999 0.999
Range 25-75 µg/ml 2.5-7.5 µg/ml
80% 99.82 ± 1.3 100.59 ± 1.12
Accuracy 100% 99.73 ± 0.8 100.24 ± 0.65
120% 99.72 ± 0.6 100.19 ± 0.49
Inter day %RSD=0.100-0.106 %RSD=0.19-0.20
Precision Intra day %RSD=0.11-0.15 %RSD=0.201-0.203
Inter day %RSD=0.1059 %RSD=1.6589
LOD 0.61 µg/ml 0.06 µg/ml
LOQ 1.86 µg/ml 0.20 µg/ml
CONCLUSION
The proposed RP-HPLC method was accurate and precise and applicable for the determination of Ivermectin and
Clorsulon without interference and with good sensitivity. Therefore proposed method can be used for routine analysis of
Clorsulon and Ivermectin in injection.
Abbreviations
HPLC : High Performance Liquid Chromatography
UV : Ultra violet
Mm : Millimetre
μm : Micrometre
LOD : Limit of Detection
LOQ : Limit of Quantitation
ICH : International Conference on Harmonisation
mL : Millilitre
μg/mL : Microgram/Millilitre
r2 : Correlation coefficient
mg : Milligram
kg : Kilogram
% : Percentage
% RSD : Percent relative standard deviation
CLOR : Clorsulon
IVER : Ivermectin
ACKNOWLEDGEMENT
It is very great pleasure and proud sense of reverence that I express my gratitude and thanks to Dr. Yogesh K. Patel,
Principal of the Sharda School of pharmacy.
530
Conflict of interest:
There is a no conflict of interest
Page
www.iajpr.com
Vol 7, Issue 08, 2017. Vegad Kunjal L. et al. ISSN NO: 2231-6876
REFERENCES
1. Tripathi KD. Essentials of Medical Pharmacology. 7 th ed. New Delhi: Jaypee Brothers Medical Publishers (P) LTD; 2011
2. British Pharmacopoeia (veterinary). Department of Health, Medicines and healthcare produces regulatory agency, London; 2005:
60-1.
3. Rehbein S, Visser M. Efficacy of an injectable ivermectin/clorsulon combination against Fasciola hepatica in sheep. Vet. Rec.
1999: 145: 468.
4. Oltean E. G, Nica A. Development and Validation of a RP- HPLC Method for the Quantitation Studies of Ivermectin in Solutions
Dosage Forms. Veterinary Drugs. 2011; 5(2): 68–70.
5. Corba J, Varady M, Praslicka J, Tomas ovica O. Efficacy of IVOMEC-F against naturally acquired gastrointestinal worms, lung
worms (Dictyocaulus filaria) and liver flukes (Fasciola hepatica) in sheep. Helminthologia. 1996; 33: 133–135.
6. Elham A. Mobarez, Hanan A. Azoz, Ashraf Sh. Darwish, Abd- Elazim M. El-Barawy. Modification And Validation of
Clorsulon And Ivermectin Analytical Method By High Performance Liquid Chromatography. 2016; 5(1): 136-148.
7. Limbani Rajen K, Modi Jignasa, Pasha T. Y. Method Development and Validation of Ivermectin And Clorsulon in their
Combined Dosage Form. International Bulletin of Drug Research. 2014; 4(6): 140-147.
8. Taylor W. G, Danielson T. J. Thin-Layer Chromatographic Detection of Ivermectin in Cattle Serum. J. Chroma. B. 1994; 661:
327-333.
9. British pharmacopoeia: 2010: volume I.1170.
10. Indian pharmacopoeia: 2010: volume III, sixth Edition: 159: 2662-2663.