Certificate of Accreditation: Perry Johnson Laboratory Accreditation, Inc
Certificate of Accreditation: Perry Johnson Laboratory Accreditation, Inc
Certificate of Accreditation: Perry Johnson Laboratory Accreditation, Inc
ACCREDITATION, INC.
Certificate of Accreditation
Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:
(Hereinafter called the Organization) and hereby declares that Organization is accredited
in accordance with the recognized International Standard:
ISO/IEC 17025:2017
This accreditation demonstrates technical competence for a defined scope and the
operation of a laboratory quality management system
(as outlined by the joint ISO-ILAC-IAF Communiqué dated April 2017):
Accreditation claims for such testing and/or calibration services shall only be made from addresses referenced within this certificate.
This Accreditation is granted subject to the system rules governing the Accreditation referred to above, and the Organization hereby
covenants with the Accreditation body’s duty to observe and comply with the said rules.
For PJLA:
Perry Johnson Laboratory The validity of this certificate is maintained through ongoing assessments based
Accreditation, Inc. (PJLA) on a continuous accreditation cycle. The validity of this certificate should be
755 W. Big Beaver, Suite 1325 confirmed through the PJLA website: www.pjlabs.com
Troy, Michigan 48084
Page 1 of 20
Certificate of Accreditation: Supplement
Servicio Industrial de Calibración SIDEC México, S.A. de C.V.
Calle Fernando Montes de Oca #848, Colonia Latinoamericana
Saltillo, Coahuila, México. CP. 25270
Contact Name: Yesenia Gonzalez Treviño Phone: 868-127-1861
Dimensional
MEASURED RANGE OR NOMINAL CALIBRATION AND CALIBRATION
INSTRUMENT, QUANTITY DEVICE SIZE AS MEASUREMENT EQUIPMENT
OR GAUGE APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Outside MicrometerFO 0.05 in to 40 in (28.8 + 2.36L) µin Gage Blocks
0.5 mm to 1 016 mm (7.33 x 10-1 + 2.35 x 10-3L) µm JIS B 7502
Inside MicrometerFO 0.05 in to 20 in (57.72 + 7.98 x 10-1L) μin
0.05 mm to 508 mm (1.47 + 7.97 x 10-4L) µm
Bore GagesFO 0.7 in to 16 in 760 µm Dial Gage Tester
JIS B 7503
CalipersFO 0.05 in to 40 in (298.97 + 3.03L) μin Gage Blocks
0.5 mm to 1 016 mm (7.37 + 3.02 x 10-3L) µm JIS B 7507
Height GageFO 0.05 in to 40 in (5.16 x 10-1 + 3.09L) µin Gage Blocks
0.5 mm to 1 016 mm -2 -3
(4.01 x 10 + 2.57 x 10 L) µm JIS B 7517
O
Micrometers Heads 0.05 in to 1 in 58 μin Gage Blocks
JIS B 7502
Optical Comparator Glass Scale
LinealO JIS B 7184
X Axis Linearity X= 200 mm (1.43 + 5 x 10-3L) µm
Y Axis Linearity Y= 200 mm
Optical Comparator 5X 0.05 % of magnification
MagnificationO 10X 0.05 % of magnification
20X 0.05 % of magnification
Optical Comparator 0° to 180° 0.006 6°
AngularityO
Optical Comparator 4 in of Y axis travel or 310 μin
Axis SquarenessO maximum
Y axis travel if maximum
is less than 4 in
Measuring TapeF 0.1 m to 50 m 0.29 mm Master Rule, Optical
Reticle, JIS B 7512
RulesF 0.01 m to 2 m 0.29 mm Master Rule, Optical
Reticle, JIS B 7516
Digital IndicatorsF 0.000 5 in to 1 in 63 μin Dial Gage Tester
JIS B 7503
MicroscopesO 5X 1.5 µm Glass Scale
Magnification 10X 1.6 µm JIS B 7153
20X 1.7 µm
F
Feeler Gages 0.001 5 in to 0.2 in 29 μin Micrometer
JIS B 7524
Pin GagesF 0.01 in to 3 in 34 μin ASME B 89 1.5
Ring GageF 0.01 in to 3 in 34 μin Linear Height
Dimensional
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Thread Plug Gage 6-32 to 4-14 70 μin ANSI/ASME B1.2 EA-10/10
(Pitch Diameter)FO Wires Mitutoyo
Mod. 313-101 and
Digital Micrometer
Thread Plug Gage 6-32 to 4-14 38 μin ANSI/ASME B1.2 EA-10/10
(Major Diameter)FO Digital Micrometer
Coating Thickness 24 µm to 1 520 µm 8 µm Micrometer
StandardFO Manufacturer’s Manual
Surface GraniteFO 0.001 in to 0.1 in 0.002 in Repeat o Meter
Repeat Reading JISB 7513
Coating Thickness GageFO 24 µm to 1 520 µm 8 µm Coating Thickness Standard
ASTM E376-17
Acoustic
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Sound MeterFO 94 db to 114 db 1.1 dB Sound Level Calibrator
CEM Model SC-05 Class 2
IEC 61672-1
Mechanical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Roughness MeterFO Roughness Master:
Ra 117 μin 1.6 μin Specimen JIS B 0601
Ry 370 μin 4.2 μin
Dial Torque Wrench 1.13 N·m to 670 N·m 0.44 % of reading Torque Transducer, Torque's
Click Torque Wrench Analyzer, ASME B 107.300
Digital Torque Wrench 670 N•m to 980 N•m 2 % of reading Parallelepiped Weights
and Torque ScrewdriverFO ASME B 107.300
Pressure GagesFO 3 psi to 300 psi 0.58 psi Pressure Calibrator: Druck
Digital Calibrator: Heise
301 psi to 10 000 psi 0.59 psi Trasducer Ashcroft
NOM-013-SCFI
Pressure GagesFO 10 000 psi to 15 000 psi 0.32 % of reading Digital Pressure Gauge
NOM-013-SCFI
Mechanical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Vacuum MetersFO -10 psi to 3 psi 0.008 3 psi Pressure Calibrator Druck
NOM-013-SCFI
TransducersFO 3 psi to 300 psi 0.059 psi Pressure Calibrator Druck
301 psi to 10 000 psi 0.59 psi Digital Calibrator Heise
Transducer Ashcroft.
Pressure TransmitterFO 3 psi to 300 psi 0.059 psi NOM-013-SCFI
301 psi to 10 000 psi 0.59 psi
Pressure RecorderFO 3 psi to 300 psi 0.059 psi
301 psi to 10 000 psi 0.59 psi
Indirect Verification of 175 HBW to 463 HBW 2.1 HBW Test Blocks
Brinnell Hardness ISO 6506-2
Tester HBW 10/3000FO ASTM E10
Indirect Verification of 20 HRC to 25 HRC 0.19 HRC Test Blocks
Rockwell Hardness 25 HRC to 45 HRC 0.35 HRC ISO 6508-2
Testers HRCFO ASTM E18
45 HRC to 70 HRC 0.32 HRC
Indirect Verification of 35 HRB to 40 HRB 0.45 HRB
Rockwell Hardness 40 HRB to 70 HRB 0.33 HRB
Testers HRBFO
70 HRB to 100 HRB 0.33 HRB
Immersion 1 g/cm³ to 1.79 g/ cm³ 1.1 x 10-4 g/cm³ Doubly Distilled Water
DensimeterFO Weight Set Weights E2,
Precision Balance.
Cuckow Method-
CENAM Technical Publication
Flow MetersFO 1 L/min to 200 L/min 0.03 L/min Master Containers, Timer
Ultrasonic Flowmeter
NOM-012-SCFI and
CENAM Technical Publication
MicropipettesF 100 µL to 1 000 µL 1.5 µL Analytic Balance Doubly
PipettesF 1 mL 0.009 mL Distilled Water.
CENAM Technical Publication
2 mL 0.012 mL
5 mL 0.018 mL
10 mL 0.023 mL
20 mL 0.035 mL
25 mL 0.035 mL
50 mL 0.057 mL
BurettesF 10 mL 0.023 mL
25 mL 0.057 mL
50 mL 0.057 mL
100 mL 0.12 mL
Mechanical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
BeakerF 25 mL 0.046 mL Analytic Balance Doubly
50 mL 0.046 mL Distilled Water.
CENAM Technical Publication
100 mL 0.11 mL
150 mL 0.11 mL
250 mL 0.11 ml
400 mL 0.28 mL
600 mL 0.28 mL
800 mL 0.46 mL
1 000 mL 0.46 mL
2 000 mL 0.69 mL
FlaskF 1 mL 0.029 mL
2 mL 0.029 mL
5 mL 0.029 mL
10 mL 0.029 mL
20 mL 0.029 mL
25 mL 0.046 mL
50 mL 0.046 mL
100 mL 0.11 mL
500 mL 0.28 mL
1 000 mL 0.46 mL
2 000 mL 0.69 mL
Test TubeF 5 mL 0.005 7 mL
10 mL 0.16 mL
25 mL 0.29 mL
50 mL 0.58 mL
100 mL 0.58 mL
250 mL 1.2 mL
500 mL 2.9 mL
1 000 mL 5.8 mL
2 000 mL 12 mL
Differential Pressure 0.03 inH2O to 20 inH2O 9.7 x 103 inH2O Fluke 721
GageFO (6 Pa to 5 000 Pa) (2.4 Pa) NOM-013-SCFI
AnemometerFO 0.6 m/s to 6.3 m/s 0.2 m/s Electronic Anemometer Lutron
ISO 1771-1 ASTM D5096
Mechanical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Air Flow MeterFO 5 slpm to 500 slpm 1 slpm + 0.8 % of reading Alicat Mass Flow Meters
CEM ME-008
Thermodynamic
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
OvenFO 20 ºC to 1 000 °C 0.2 °C Thermocouple Type S:
MufflesFO 20 ºC to 1 000 °C 0.2 °C RTD Pt100 and
Multimeter Keithley
ASTM E220
Hot RoomsO 20 ºC to 100 °C 0.21 °C RTD Pt100 and
Cold RoomsO -20 ºC to 20 ºC 1.5 °C Multimeter Keithley
ASTM E220
Liquid in Glass -15 °C to 200 °C 0.21 °C RTD Pt100: and Multimeter
ThermometersFO Keithley, Temperature Baths,
Drywell, Oven
CENAM Technical Publication
Bimetallic -15 °C to 50 °C 0.2 °C Thermocouple Type S, RTD
ThermometersF 50 °C to 400 °C 0.23 °C Pt100 and Multimeter Keithley;
Temperature Baths, Drywell,
400 °C to 500 °C 0.44 °C Oven / ASTM E220
Thermocouple Type JF 0 °C to 50 °C 0.2 °C
50 °C to 400 °C 0.23 °C
400 °C to 1 000 °C 0.44 °C
Thermocouple Type KF 0 °C to 50 °C 0.2 °C
50 °C to 400 °C 0.23 °C
400 °C to 1 000 °C 0.44 °C
RTD - Resistance -15 °C to 500 °C 0.16 °C Thermocouple Type S, RTD
Temperature DetectorF Pt100 and Multimeter Keithley;
Temperature Baths Drywell,
Oven/ASTM E644-11
Infrared Thermometer -30 °C to 150 °C 1.2 °C Blackbody Fluke 9133
PyrometersF ASTM E 1256
35 °C to 500 °C 1.3 °C Blackbody Fluke 4181
ASTM E 1256
500 °C to 1 000 °C 1.8 °C Thermocouple Type S and
Multimeter Keithley; Oven
ASTM E 1256
Thermodynamic
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Thermohygrometers- 20 °C to 60 °C 0.19 °C Thermohygrometer
Temperature OnlyF Chamber Generator
HygrometersF 10 % RH to 90 % RH 1.6 % RH ASTM E104-02
F
Humidity Tester 10 % RH to 90 % RH 1.6 % RH
Humidity ChamberF 10 % RH to 90 % RH 1.6 % RH
Electrical
MEASURED RANGE OR NOMINAL DEVICE CALIBRATION AND CALIBRATION
INSTRUMENT, SIZE AS APPROPRIATE MEASUREMENT EQUIPMENT
QUANTITY OR GAUGE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Measure Fluke 5500A
AC Voltage Euramet-cg-15
At the listed frequenciesFO
10 Hz to 45 Hz 1 mV to 32.999 mV 200 μV
45 Hz to 10 kHz 1 mV to 32.999 mV 160 μV
10 kHz to 20 kHz 1 mV to 32.999 mV 170 μV
20 kHz to 50 kHz 1 mV to 32.999 mV 200 μV
50 kHz to 100 kHz 1 mV to 32.999 mV 230 μV
100 kHz to 500 kHz 1 mV to 32.999 mV 460 μV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
10 Hz to 45 Hz 33 mV to 329.999 mV 960 μV
45 Hz to 10 kHz 33 mV to 329.999 mV 260 μV
10 kHz to 20 kHz 33 mV to 329.999 mV 350 μV
20 kHz to 50 kHz 33 mV to 329.999 mV 680 μV
50 kHz to 100 kHz 33 mV to 329.999 mV 1 100 μV
100 kHz to 500 kHz 33 mV to 329.999 mV 2 700 μV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
10 Hz to 45 Hz 0.33 V to 3.299 99 V 10 mV
45 Hz to 10 kHz 0.33 V to 3.299 99 V 19 mV
Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE MEASUREMENT EQUIPMENT
CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Measure Fluke 5500A
AC Voltage Euramet-cg-15
At the listed frequenciesFO
10 kHz to 20 kHz 0.33 V to 3.299 99 V 3 mV
20 kHz to 50 kHz 0.33 V to 3.299 99 V 10 mV
50 kHz to 100 kHz 0.33 V to 3.299 99 V 10 mV
100 kHz to 500 kHz 0.33 V to 3.299 99 V 23 mV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
10 Hz to 45 Hz 3.3 V to 32.999 9 V 60 mV
45 Hz to 10 kHz 3.3 V to 32.999 9 V 20 mV
10 kHz to 20 kHz 3.3 V to 32.999 9 V 30 mV
20 kHz to 50 kHz 3.3 V to 32.999 9 V 80 mV
50 kHz to 100 kHz 3.3 V to 32.999 9 V 190 mV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
45 Hz to 1 kHz 33 V to 329.999 V 580 mV
1 kHz to 10 kHz 33 V to 329.999 V 300 mV
10 kHz to 20 kHz 33 V to 329.999 V 2 300 mV
45 Hz to 1 kHz 330 V to 1 000 V 2 200 mV
1 kHz to 10 kHz 330 V to 1 000 V 2 600 mV
Equipment to Measure
AC Current
At the listed frequenciesFO
10 Hz to 20 Hz 0.029 mA to 0.329 99 mA 1.5 μA
20 Hz to 45 Hz 0.029 mA to 0.329 99 mA 1.3 μA
45 Hz to 1 kHz 0.029 mA to 0.329 99 mA 1.3 μA
1 Hz to 5 kHz 0.029 mA to 0.329 99 mA 1.9 μA
5 Hz to 10 kHz 0.029 mA to 0.329 99 mA 4.4 μA
Equipment to Measure
AC Current
At the listed frequenciesFO
10 Hz to 20 Hz 3.3 mA to 32.999 mA 87 μA
Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND MEASUREMENT CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS CAPABILITY EXPRESSED EQUIPMENT
APPROPRIATE AS AN UNCERTAINTY (±) AND REFERENCE
STANDARDS USED
Equipment to Measure Fluke 5500A
AC Current Euramet-cg-15
At the listed frequenciesFO
20 Hz to 45 Hz 3.3 mA to 32.999 mA 37 μA
45 Hz to 1 kHz 3.3 mA to 32.999 mA 35 μA
1 Hz to 5 kHz 3.3 mA to 32.999 mA 71 μA
5 Hz to 10 kHz 3.3 mA to 32.999 mA 210 μA
Equipment to Measure
AC Current
At the listed frequenciesFO
10 Hz to 20 Hz 33 mA to 329.99 mA 700 μA
20 Hz to 45 Hz 33 mA to 329.99 mA 380 μA
45 Hz to 1 kHz 33 mA to 329.99 mA 350 μA
1 Hz to 5 kHz 33 mA to 329.99 mA 700 μA
5 Hz to 10 kHz 33 mA to 329.99 mA 2 100 μA
Equipment to Measure
AC Current
At the listed frequenciesFO
10 Hz to 45 Hz 0.33 to 2.199 99 A 4.8 mA
45 Hz to 1 kHz 0.33 to 2.199 99 A 2.6 mA
Equipment to Measure
AC Current
At the listed frequenciesFO
45 Hz to 65 Hz 2.2 A to 11 A 14 mA
65 Hz to 500 Hz 2.2 A to 11 A 17 mA
500 Hz to 1 kHz 2.2 A to 11 A 40 mA
Equipment to Measure 33 mV to 330 mV 0.006 % of reading + 3μV
DC VoltageFO 0.33 V to 3.3 V 0.005 % of reading + 5 μV
3.3 V to 33 V 0.005 % of reading + 50 mV
33 V to 330 V 0.005 5 % of reading + 500 mV
330 V to 1 020 V 0.005 5 % of reading + 1 500 mV
Equipment to Measure 0.33 mA to 3.3 mA 0.47 μA
DC CurrentFO 3.3 mA to 33 mA 4.1 μA
33 mA to 330 mA 43 μA
330 mA to 2.2 A 820 μA
2.2 A to 11 A 8 100 μA
1 A to 550 A 0.25 % of reading Fluke 5500A/COIL
Euramet-cg-15
Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND MEASUREMENT CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS CAPABILITY EXPRESSED EQUIPMENT
APPROPRIATE AS AN UNCERTAINTY (±) AND REFERENCE
STANDARDS USED
Resistance to MeasureFO 1 mΩ to 10.99 Ω 9.3 mΩ Fluke 5500A
11 Ω to 32.999 Ω 19 mΩ Euramet-cg-15
33 Ω to 109.999 Ω 25 mΩ
110 Ω to 329.999 Ω 46 mΩ
330 Ω to 1099.99 Ω 0.15 Ω
1.1 kΩ to 3.299 99 kΩ 0.36 Ω
3.3 kΩ to 10.999 9 kΩ 1.5 Ω
11 kΩ to 32.999 9 kΩ 3.6 Ω
33 kΩ to 109.999 kΩ 18 Ω
110 kΩ to 329.999 kΩ 46 Ω
330 kΩ to 1 099.99 kΩ 0.24 kΩ
330 kΩ to 1 099.99 kΩ 0.56 kΩ
1.1 MΩ to 3.299 99 MΩ 0.56 kΩ
11 MΩ to 32.999 MΩ 36 kΩ
33 MΩ to 109.99 MΩ 0.56 MΩ
110 MΩ to 330 MΩ 1.7 MΩ
Capacitance to 0.33 nF to 0.499 9 nF 0.019 ηF
MeasureFO 0.5 nF to 1.099 9 nF 0.022 ηF
1.1 nF to 3.299 9 nF 0.032 ηF
3.3 nF to 10.999 nF 0.075 ηF
11 ηF to 32.999 ηF 0.22 ηF
33 ηF to 109.99 ηF 0.44 ηF
110 ηF to 329.99 ηF 1.5 ηF
0.33 µF to 1.099 9 µF 4.4 ηF
1.1 µF to 3.299 9 µF 17 ηF
3.3 µF to 10.999 µF 56 ηF
11 µF to 32.999 µF 190 ηF
33 µF to 109.99 µF 760 ηF
110 µF to 329.99 µF 3.1 µF
0.33 mF to 1.1 mF 14 µF
Temperature Calibration, 600 °C to 800 °C 0.44 ºC Electrical Simulation of
Indication, and Control 800 °C to 1 000 °C 0.34 ºC Thermocouple Output
Equipment used with Fluke 5500A
Thermocouple Type BFO 1 000 °C to 1 550 °C 0.3 ºC Euramet-cg-11
1 550 °C to 1 820 °C 0.33 ºC
Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Temperature Calibration, 0 °C to 150 °C 0.3 ºC Electrical Simulation of
Indication, and Control 150 °C to 650 °C 0.26 ºC Thermocouple Output
Equipment used with Fluke 5500A
Thermocouple Type CFO 650 °C to 1 000 °C 0.31 ºC Euramet-cg-11
1 000 °C to 1 800 °C 0.5 ºC
1 800 °C to 2 316 °C 0.84 ºC
Temperature Calibration, -250 °C to -100 °C 0.5 ºC
Indication, and Control -100 °C to -25 °C 0.16 ºC
Equipment used with
Thermocouple Type EFO -25 °C to 350 °C 0.14 ºC
350 °C to 650 °C 0.16 ºC
650 °C to 1 000 °C 0.21 ºC
Temperature Calibration, -210 °C to -100 °C 0.27 ºC
Indication, and Control -100 °C to -30 °C 0.16 ºC
Equipment used with
Thermocouple Type JFO -30 °C to 150 °C 0.14 ºC
150 °C to 760 °C 0.17 ºC
760 °C to 1 200 °C 0.23 ºC
Temperature Calibration, -200 °C to -100°C 0.33 ºC
Indication, and Control -100 °C to -25°C 0.18 ºC
Equipment used with
Thermocouple Type KFO -25 °C to 120°C 0.16 ºC
120 °C to 1 000 °C 0.26 ºC
1 000 °C to 1 372 °C 0.4 ºC
Temperature Calibration, -200 °C to -100 °C 0.37 ºC
Indication, and Control -100 °C to 800 °C 0.26 ºC
Equipment used with
Thermocouple Type LFO 800 °C to 900 °C 0.17 ºC
Temperature Calibration, -200 °C to -100 °C 0.4 ºC
Indication, and Control -100 °C to -25 °C 0.22 ºC
Equipment used with
Thermocouple Type NFO -25 °C to 120 °C 0.19 ºC
120 °C to 410 °C 0.18 ºC
410 °C to 1 300 °C 0.27 ºC
Temperature Calibration, 0 °C to 250 °C 0.57 ºC
Indication, and Control 250 °C to 400 °C 0.35 ºC
Equipment used with
Thermocouple Type RFO 400 °C to 1 000 °C 0.33 ºC
1 000 °C to 1 767 °C 0.4 ºC
Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Temperature Calibration, 0 °C to 250 °C 0.47 ºC Electrical Simulation of
Indication, and Control 250 °C to 1 000 °C 0.36 ºC RTD Output
Equipment used with Fluke 5500A
Thermocouple Type SFO 1 000 °C to 1 400 °C 0.37 ºC Euramet-cg-11
1 400 °C to 1 767 °C 0.46 ºC
Temperature Calibration, -250 °C to -150 °C 0.63 ºC
Indication, and Control -150 °C to 0 °C 0.24 ºC
Equipment used with
Thermocouple Type TFO 0 °C to 120 °C 0.16 ºC
120 °C to 400 °C 0.14 ºC
Temperature Calibration, -200 °C to 0 °C 0.56 ºC
Indication, and Control 0 °C to 600 °C 0.57 ºC
Equipment used with
Thermocouple Type UFO
Temperature Calibration, -200 °C to -80 °C 0.05 ºC
Indication, and Control -80 °C to 0 °C 0.05 ºC
Equipment used with
RTD Pt 385, 100 ΩFO 0 °C to 100 °C 0.07 ºC
100 °C to 300 °C 0.09 ºC
300 °C to 400 °C 0.1 ºC
400 °C to 630 °C 0.12 ºC
630 °C to 800 °C 0.23 ºC
Temperature Calibration, -200 °C to -80 °C 0.05 ºC
Indication, and Control -80 °C to 0 °C 0.05 ºC
Equipment used with
RTD Pt 3926, 100 ΩFO 0 °C to 100 °C 0.07 ºC
100 °C to 300 °C 0.09 ºC
300 °C to 400 °C 0.1 ºC
400 °C to 630 °C 0.12 ºC
Temperature Calibration, -200 °C to -190 °C 0.25 ºC
Indication, and Control -190 °C to -80 °C 0.04 ºC
Equipment used with
RTD Pt 3916, 100 ΩFO -80 °C to 0 °C 0.05 ºC
0 °C to 100 °C 0.06 ºC
Temperature Calibration, 100 °C to 260 °C 0.07 ºC
Indication, and Control 260 °C to 300 °C 0.08 ºC
Equipment used with
RTD Pt 3916, 100 ΩFO 300 °C to 400 °C 0.09 ºC
400 °C to 600 °C 0.1 ºC
600 °C to 630 °C 0.23 ºC
Electrical
MEASURED RANGE OR NOMINAL DEVICE CALIBRATION AND CALIBRATION
INSTRUMENT, SIZE AS APPROPRIATE MEASUREMENT EQUIPMENT
QUANTITY OR GAUGE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Temperature -200 °C to -80 °C 0.04 ºC Electrical Simulation of
Calibration, -80°C to 0 °C 0.04 ºC RTD Output
Indication, and Control Fluke 5500A
Equipment used with 0 °C to 100 °C 0.04 ºC Euramet-cg-11
RTD Pt 385, 200 ΩFO 100 °C to 260 °C 0.05 ºC
260 °C to 300 °C 0.12 ºC
300 °C to 400 °C 0.13 ºC
400 °C to 600 °C 0.14 ºC
600 °C to 630 °C 0.16 ºC
Temperature -200°C to -80°C 0.04 ºC
Calibration, -80 °C to 0 °C 0.05 ºC
Indication, and Control
Equipment used with 0 °C to 100 °C 0.05 ºC
RTD Pt 385, 500 ΩFO 100 °C to 260 °C 0.06 ºC
260 °C to 300 °C 0.08 ºC
300 °C to 400 °C 0.08 ºC
400 °C to 600 °C 0.09 ºC
600 °C to 630 °C 0.11 ºC
Temperature -200 °C to -80 °C 0.03 ºC
Calibration, -80 °C to 0 °C 0.03 ºC
Indication, and Control
Equipment used with 0 °C to 100 °C 0.04 ºC
RTD Pt 385, 1 000 ΩFO 100 °C to 260 °C 0.05 ºC
260 °C to 300 °C 0.06 ºC
300 °C to 400 °C 0.07 ºC
400 °C to 600 °C 0.07 ºC
600 °C to 630 °C 0.23 ºC
Equipment to Output AC Fluke 8808A
Voltage CEM EL- 010
At the listed frequenciesFO
20 Hz to 45 Hz 200 mV to 750 V 0.9 % of reading
45 Hz to 20 kHz 200 mV to 750 V 0.9 % of reading
20 kHz to 50 kHz 200 mV to 750 V 0.9 % of reading
50 kHz to 100 kHz 200 mV to 750 V 0.9 % of reading
Equipment to Output AC Fluke 80K40
Voltage CEM EL- 010
At the listed frequenciesFO
20 Hz to 45 Hz 750 V to 40 kV 5 % of reading
45 Hz to 20 kHz 750 V to 40 kV 5 % of reading
Electrical
MEASURED RANGE OR NOMINAL DEVICE CALIBRATION AND CALIBRATION
INSTRUMENT, SIZE AS APPROPRIATE MEASUREMENT EQUIPMENT
QUANTITY OR GAUGE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Output AC Fluke 80K40
Voltage CEM EL- 010
At the listed frequenciesFO
20 kHz to 50 kHz 750 V to 40 kV 5 % of reading
50 kHz to 100 kHz 750 V to 40 kV 5 % of reading
Equipment to Output 200 mV to 1 000 V 0.015 % of reading + 10 mV Fluke 8808A
DC VoltageFO CEM EL- 010
10 V to 40 kV 5 % of reading Fluke 80K40
CEM EL- 010
Equipment to Output 20 mA to 10 A 1.3 % of reading Fluke 8808A
AC CurrentFO 1µA to 10 A 0.08 % of reading CEM EL- 010
1A to 50 A 2 % of reading Shunt 50A, Fluke 8808A
CEM EL- 010
Equipment to Output 20 Hz to 1 MHz 0.01 % of reading Fluke 8808A
FrequencyFO CEM EL- 010
Equipment to Output 1 µΩ to 199.999 kΩ 0.03 % of reading
ResistanceFO 199.999 kΩ to 19.999 9 MΩ 0.25 % of reading Resistance Decade:
19.999 9 MΩ to 100 MΩ 1.8 % of reading Burster: Model 1443
Burster: Model 1448
Burster, Model 1441
CEM EL- 010
Equipment to Measure 25 mV to 90 mV 0.009 7 mV Fluke 753
DC VoltageFO 1 V to 2.5 V 0.000 15 V Euramet-cg-15
10 V to 25 V 0.001 5 V
100 V to 250 V 0.02 V
Equipment to Measure 1 V to 2.5 V 0.001 6 V
AC Current 10 V to 25 V 0.018 V
60HzFO
100 V to 250 V 0.19 V
Equipment to Measure 10 mA to 25 mA 0.003 3 mA Fluke 753
DC CurrentFO 35 mA to 90 mA 0.018 mA Euramet-cg-15
Resistance to 2 Ω to 9 Ω 0.01 Ω
MeasureFO 20 Ω to 90 Ω 0.029 Ω
200 Ω to 900 Ω 0.19 Ω
2 kΩ to 9 kΩ 0.002 1 kΩ
Equipment to Measure 25 Hz to 100 Hz 0.012 Hz
FrequencyFO 250 Hz to 1000 Hz 0.11 Hz
2.5 kHz to 10 kHz 0.0011 kHz
20 kHz to 45 kHz 0.011 kHz
Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE MEASUREMENT EQUIPMENT
CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Output 20 mV to 90 mV 0.008 9 mV Fluke 753
DC VoltageFO 0.2 V to 0.9 V 0.000 05 V Euramet-cg-15
2.5 V to 13 V 0.000 78 V
Equipment to Output 1 mA to 20 mA 0.003 2 mA
DC CurrentFO
Equipment to Output 1 Ω to 9 Ω 0.001 8 Ω
ResistanceFO 20 Ω to 90 Ω 0.018 Ω
0.2 kΩ to 0.9 kΩ 0.000 16 Ω
3 Hz to 10 Hz 0.011 Hz
Equipment to Output 20 Hz to 100 Hz 0.012 Hz
FrequencyFO
200 Hz to 1000 Hz 0.11 Hz
2 kHz to 20 kHz 0.002 3 kHz
25 kHz to 45 kHz 0.005 9 kHz
Temperature Calibration, 650 °C to 1 750 °C 0.52 °C Fluke 753
Indication, and Control Electrical Simulation of
Equipment used with Thermocouple Output
Thermocouple Type BFO Euramet-cg-11
Temperature Calibration, 400 °C to 2 200 °C 0.97 °C
Indication, and Control
Equipment used with
Thermocouple Type CFO
Temperature Calibration, 100 °C to 900 °C 0.26 °C
Indication, and Control
Equipment used with
Thermocouple Type EFO
Temperature Calibration, 100 °C to 900 °C 0.28 °C
Indication, and Control
Equipment used with
Thermocouple Type JFO
Temperature Calibration, 100 °C to 1 200 °C 0.33 °C
Indication, and Control
Equipment used with
Thermocouple Type KFO
Temperature Calibration, 150 °C to 850 °C 0.32 °C
Indication, and Control
Equipment used with
Thermocouple Type LFO
Temperature Calibration, 100 °C to 1 200 °C 0.33 °C
Indication, and Control
Equipment used with
Thermocouple Type NFO
Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Temperature Calibration, 200 °C to 1 600 °C 0.66 °C Fluke 753
Indication, and Control Electrical Simulation of
Equipment used with Thermocouple Output
Thermocouple Type RFO Euramet-cg-11
Temperature Calibration, 100 °C to 1 650 °C 0.57 °C
Indication, and Control
Equipment used with
Thermocouple Type SFO
Temperature Calibration, 100 °C to 350 °C 0.21 °C
Indication, and Control
Equipment used with
Thermocouple Type TFO
Temperature Calibration, 100 °C to 550 °C 0.33 °C
Indication, and Control
Equipment used with
Thermocouple Type UFO
Temperature Calibration, 100 °C to 7 000 °C 0.28 °C Fluke 753
Indication, and Control Electrical Simulation
Equipment used with of RTD Output
RTD Pt 385, 100 ΩFO Euramet-cg-11
Temperature Calibration, 100 °C to 550 °C 0.19 °C
Indication, and Control
Equipment used with
RTD Pt 385, 200 ΩFO
Temperature Calibration, 100 °C to 550 °C 0.15 °C
Indication, and Control
Equipment used with
RTD Pt 385, 500 ΩFO
Temperature Calibration, 100 °C to 550 °C 0.14 °C
Indication, and Control
Equipment used with
RTD Pt 385, 1 000 ΩFO
Temperature Calibration, 100 °C to 550 °C 0.16 °C
Indication, and Control
Equipment used with
RTD Pt 3916, 100 ΩFO
Temperature Calibration, 100 °C to 550 °C 0.18 °C
Indication, and Control
Equipment used with
RTD Pt 3926, 100 ΩFO
Temperature Calibration, 0.0 °C to 200 °C 0.36 °C
Indication, and Control
Equipment used with
RTD Cu 427, 10 ΩFO
Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE MEASUREMENT EQUIPMENT
CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Temperature Calibration, 0 °C to 200 °C 0.16 °C Fluke 753
Indication, and Control Electrical Simulation
Equipment used with of RTD Output
RTD Ni 672, 120 ΩFO Euramet-cg-11
Optical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE MEASUREMENT EQUIPMENT
CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
LuxmeterFO 100 lux to 3 500 lux 12 lux Luxmeter Master
Light Intensity Meter
Calibration System using
Comparison Method
PB-OPT-01
Spectral Reflectance λ = 400 nm to 700 nm Ceramic Color Standards
Color Spectrometers, 0 % to 100 % reflectance (ρ) (Series II)
ReflectanceFO Color Values: CENAM Technical Guide
CIE L: 0 to 100 CIE L*:0.2
CIE a*: -80 to 80 CIE a*: 0.1
CIE b*: -80 to 80 CIE b*: 0.1
Spectrophotometer λ: 230 nm to 700 nm 0.002 8 nm Neutral Density Filters
Transmittance FO (τ: 1 % to 95 %) Homium Oxide Glass
CENAM Technical Guide
Chemical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION AND CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS MEASUREMENT EQUIPMENT
APPROPRIATE CAPABILITY EXPRESSED AND REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
pH MetersFO 4 pH 0.017 pH Reference Material for pH:
7.41 pH 0.015 pH (Buffer Solutions)
NMX-AA-008-SCFI
9.18 pH 0.015 pH
Conductivity Meters 146 µS/cm 2 µS/cm Reference Material for
Fixed PointsFO 1 413 µS/cm 5 µS/cm Electrolytic Conductivity
PROY-NMX-AA-093-SCFI
12 880 µS/cm 50 µS/cm
Viscometers – Dynamic 6 024 mPa·s 27 mPa·s Reference Material for
RotationalFO 16 500 mPa·s 73 mPa·s Viscosity
CENAM Technical Publication
33 680 mPa·s 150 mPa·s
1. The CMC (Calibration and Measurement Capability) stated for calibrations included on this scope of
accreditation represents the smallest measurement uncertainty attainable by the laboratory when performing
a more or less routine calibration of a nearly ideal device under nearly ideal conditions. It is typically
expressed at a confidence level of 95 % using a coverage factor k (usually equal to 2). The actual measurement
uncertainty associated with a specific calibration performed by the laboratory will typically be larger than
the CMC for the same calibration since capability and performance of the device being calibrated and the
conditions related to the calibration may reasonably be expected to deviate from ideal to some degree.
2. The laboratories range of calibration capability for all disciplines for which they are accredited is the interval
from the smallest calibrated standard to the largest calibrated standard used in performing the calibration.
The low end of this range must be an attainable value for which the laboratory has or has access to the
standard referenced. Verification of an indicated value of zero in the absence of a standard is common practice
in the procedure for many calibrations but by its definition it does not constitute calibration of zero capacity.
3. The presence of a superscript F means that the laboratory performs calibration of the indicated parameter at
its fixed location. Example: Outside MicrometerF would mean that the laboratory performs this calibration
at its fixed location.
4. The presence of a superscript O means that the laboratory performs calibration of the indicated parameter
onsite at customer locations. Example: Outside MicrometerO would mean that the laboratory performs this
calibration onsite at the customer’s location.
5. The presence of a superscript FO means that the laboratory performs calibration of the indicated parameter
both at its fixed location and onsite at customer locations. Example: Outside MicrometerFO would mean that
the laboratory performs this calibration at its fixed location and onsite at customer locations.
6. Measurement uncertainties obtained for calibrations performed at customer sites can be expected to be larger
than the measurement uncertainties obtained at the laboratories fixed location for similar calibrations. This
is due to the effects of transportation of the standards and equipment and upon environmental conditions at
the customer site which are typically not controlled as closely as at the laboratories fixed location.
7. The term L represents length in inches or millimeters as appropriate to the uncertainty statement.
8. The term P represents pressure in units appropriate to the uncertainty statement.
9. The term Wt represents weight in pounds or grams (including SI multiple and submultiple units) appropriate
to the uncertainty statement.
Issue 04/2020 This supplement is in conjunction with certificate #L20-239 Page 20 of 20