HORIBA YUMIZEN LINEARITY Linearity
HORIBA YUMIZEN LINEARITY Linearity
HORIBA YUMIZEN LINEARITY Linearity
35000.0
30000.0
25000.0
20000.0
15000.0
10000.0
5000.0
0.0
7.5
100
0.0
2.5
6.3
5.0
17.
20.
22.
27.
32.
42.
45.
50.
55.
57.
60.
65.
67.
72.
77.
82.
85.
87.
95.
12.
15.
25.
30.
35.
37.
40.
47.
52.
62.
70.
75.
80.
90.
92.
97.
AMR Graph HGB Linear (HGB)
25.0
20.0
15.0
10.0
5.0
0.0
0.0
5.0
100
2.5
6.3
7.5
12.
15.
20.
22.
27.
30.
35.
37.
42.
45.
52.
57.
60.
62.
67.
70.
75.
77.
82.
85.
90.
92.
97.
17.
25.
32.
40.
47.
50.
55.
65.
72.
80.
87.
95.
500.0
400.0
300.0
200.0
100.0
Remarks:
0.0
1
As per the graphs above it seems that the instrument is giving the value in the linear range as per the equipment specification.
The straight line drawn from '0' to the maximum value is more or less touching all points obtained by dilution of the specimen.
This proves that the linear range as specified by the equipment manufacturer is OK and acceptable.
Tech.sign Approved By