Advanced Functions
l Four-Terminal AC Method l Self-Calibrating
The four-terminal, 1-kHz AC method uses four contact probes Minor drift and gain fluctuations within the internal measure-
to measure resistance independently of that of the measurement ment circuitry are automatically corrected to maintain high ac-
leads. curacy.
l Measurement Error Detection l Averaging Function
Detects test probe contact failure and broken leads, for 100% Stable readings can be consistently obtained by averaging two to
measurement reliability. 16 measurements.
3
l The 3 mΩ range (with 0.1 µΩ resolution) is ideal for testing ever l Provides high-speed measure-
lower-resistance large cells (BT3563 and BT3562). ment of high-voltage*3 battery
packs, for improving productiv-
l The 6 V range (with 10 µV resolution and 0.01% accuracy) is ideal
ity (BT3563).
for the high-precision voltage measurements required for cell test- *3
BT3563: up to 300 V
ing (BT3563 and BT3562). BT3562: up to 60 V
Measurement Parameters
and Applications BATTERY HiTESTER 3561
l For high-speed production line testing of small
battery packs for mobile and portable commu- 3561
nications devices
Up to
l For high-speed production line testing of small
cells
20V
l High-speed 10ms inspection in the 300 mΩ and 3 Ω ranges
l Improve inspection efficiency during mass production of Voltage measurement ranges: 20 V
compact cells
Resistance measurement ranges: 300 mΩ/3 Ω
Lithium-Ion and Secondary Batteries
Quick Response
with small cell measurement
Battery-Powered Devices
Handler Interface
Triggering, measurement configuration loading, and zero adjust-
Analog Output EXT I/O GP-IB RS-232C
ment can be externally controlled. Output signals provide com- (-01 suffix models only) (-01 suffix
parator results, end-of-measurement events, and measurement models only)
errors. (Because the BT3563/BT3652 are different from the 3561, 3561-01 Rear Panel
consult each model’s Instruction Manual for specific details or
designs.)
Comparator Functions
l Judges Resistance & Voltage Simultaneously l Composite Judgment Result Output
Resistance and voltage can be simultaneously judged Hi/IN/Lo by External I/O provides both separate and combined outputs of re-
independent comparators. Judgment results are sistance and voltage judgment results, so composite results can be
provided on the display, beeper, and external I/O. monitored.
The display allows confirming both results at a
glance. l Alternative Setting Methods
Set judgment thresholds by specifying high/low (Hi/Lo) values or
by specifying a standard value and deviation (%).
l Manual Comparator
Resistance comparator Comparator judgments can be executed only when required,
settings supporting flexible control by footswitch or PLC.
l Dual Beep Tones
Different beep tones distinguish IN and Hi/Lo judgments.
Both tones can be independently enabled or disabled.
Voltage comparator
settings
5
PC Application Program
Measurement data can be transferred to a PC for
importing to a spreadsheet program or storage as
CSV files. Interval and manual measurements can
be triggered by a keystroke or external trigger sig-
nal.
Data Printing
Measurement values, and those including judgment results and statistical
Printout Examples
calculation results can be printed using an RS-232C-compatible printer.
l Interval Printing
Elapsed time and measurement values can be printed over a speci-
fied interval. The interval can be set from 1 to 3,600 seconds.
Specifications
l BT3563,BT3562 and 3561 Specifications
Measurement types Resistance and voltage Zero-adjustment 1000 count range (both resistance and voltage)
Resistance measurement method Four-terminal AC (1 kHz) method Triggering Internal or external
Functions ΩV, Ω and V Delay time On/off, 0 to 9.999 seconds
Rated voltage [BT3563(-01)] Averaging samples On/off, 2 to 16 samples
±300 VDC rated input voltage
Statistical calculations Total data count; valid data count; maximum, minimum
±300 VDC maximum rated voltage to ground
and average values; standard deviation; population stan-
[BT3562(-01)]
dard deviation and process capability indices
±60 VDC rated input voltage
±70 VDC maximum rated voltage to ground Measurement value Measurement values are output via RS-232C upon trig-
[3561(-01)] output function ger input
±22 VDC rated input voltage Measurement value memory Up to 400 measurements
±60 VDC maximum rated voltage to ground Panel save/load Up to 126 configuration settings
Input resistance [BT3563(-01) and BT3562(-01)] Save Frequently Used Settings in Memory:
3 mΩ/30 mΩ/300 mΩ ranges: Approx. 90 kΩ Measurement function, resistance measurement range,
3 Ω/30 Ω/300 Ω/3000 Ω ranges: Approx. 1 MΩ auto-range setting, zero-adjust setting data, sampling
[3561(-01)] rate, trigger source, delay setting, averaging and com-
Approx.1MΩ parator settings, statistical calculation setting, display
Sampling rate Four steps – Extra Fast, Fast, Medium or Slow switching and key-lock.
Response time [BT3563(-01) and BT3562(-01)] Analog Output [BT3563-01 and BT3562-01 only]
Approx. 10 ms for measurements Measured resistance (displayed value, from 0 to 3.1 VDC)
Note: Response time depends on reference values and the mea- External interface External I/O, RS232C (9600, 19200 or 38400 bps),
surement object.
Printer RS-232C (all models), GP-IB (Model BT3563-01,
[3561(-01)]
BT3562-01 and 3561-01 only)
Approx. 3 ms for measurements
Note: Response time depends on reference values and the mea- Other functions Over-range display, measurement error detection, self-
surement object. calibration, dual comparators, key-lock
Total measurement time Sampling time + Response time
Items in the parentheses () indicate supply frequency settings; Tolerance: ±5 ms Items in the parentheses () indicate supply frequency settings; Tolerance: ±5 ms
for SLOW sampling, and ±1 ms for other settings. for SLOW sampling, and ±1 ms for other settings.
7
Measurement Ranges and Accuracy (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
year)
[Voltage Measurement]
Range 6V 60 V 300 V (only BT3563)
Maximum display Value ±6.00000 V ±60.0000 V ±300.000 V
Resolution 10 µV 100 µV 1 mV
Accuracy*3 ±0.01% rdg. ±3 dgt.
Temperature coefficient (±0.001% rdg. ±0.3 dgt.) / ºC
l 3561 l 3561
[Resistance Measurement] [Voltage Measurement]
Range 300 mΩ 3Ω Range 20V
Maximum display Value 310.00 mΩ 3.1000 Ω Maximum display Value ±19.9999 V
Resolution 10 µΩ 100 µΩ Resolution 0.1 mV
Measurement Current *4
10 mA 1 mA Accuracy *6
±0.01% rdg. ±3 dgt.
Measurement Current Frequency 1 kHz ±0.2 Hz Temperature coefficient (±0.001% rdg. ±0.3 dgt.) / ºC
Accuracy*5 ±0.5% rdg. ±5 dgt.
Temperature coefficient (±0.05% rdg. ±0.5 dgt.) / ºC
Open-Circuit Voltage 7 Vpeak
Measurement leads (for measuring batteries up to 60 V with BT3563, BT3562, or 3561) Measurement leads (3561 only)
φ5
Current
Voltage
23
φ2
Mainly for Small Secondary Batteries (with very small terminals) Measurement leads (for maximum precision, 3561 only)
1.8 mm dia. single-axis type for measuring 0.2 mm parallel pyramid-type pins for measuring PIN TYPE LEAD 9455
φ 0.8 A:260 mm (10.24 in), B:136 mm (5.35
small electrodes at thru holes and sub-millimeter objects
in), L:890 mm (2.92 ft), Not CE marked.
Current Note: The 9455 is a precision instru-
14mm