ISO 7886-2 1996 Syringes For Use With Power-Driven Syringe Pumps
ISO 7886-2 1996 Syringes For Use With Power-Driven Syringe Pumps
ISO 7886-2 1996 Syringes For Use With Power-Driven Syringe Pumps
STANDARD 7886-2
First edition
1996-05-15
Contents
Page
1 Scope 1
2 Normative references 1
3 Definitions 1
4 Nomencicture 1
5 Cleanliness 1
8 Lubricant 2
9 Tolerance on graduated capacity 2
10 Graduated scale 2
11 Syringe design 2
12 Piston/plunger assembly 2
13 Nozzle 3
14 Performance 3
15 Packaging 3
16 Labelling 4
Annexes
E Bibliography 14
© ISO 1996
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced
or utilized in any form or by any means, electronic or mechanical, including photocopying and
microfilm, without permission in writing from the publisher.
International Organization for Standardization
Case postale 56 • CH-1211 Geneve 20 • Switzerland
Printed in Switzerland
© ISO ISO 7866-2:1996(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide
federation of national standards bodies (ISO member bodies). The work of
preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which
a technical committee has been established has the right to be rep-
resented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, aisc take part in the work. ISO
collaborates closely with the International Electrotechnical Commission
(IEC) on all matters of electiotechnical standardization.
ON 01/02/2010
a vote.
SUPPLI EDBYBSBUN DERLI CENC EF ROMISOFORSU NP HARMACEUTI CAL INDU STRIESLIMITED - VADODARA
needles and intravascular catheters for single use.
ISO 7886 consists of the following parts, under the general title Sterile
hypodermic syringes for single use:
— Part 1: Syringes for manual use
Annexes A, B and C form an integral part cf this part of ISO 7886. Annexes
D and E are for information only.
iii
ISO 7836-2:1996(E) © ISO
fritroduction
1 General
2 Design criteria
The use of syringes which were initially designed and used as manually-
operated devices in syringe pumps now makes it desirable to achieve
much tighter tolerances on syringe dimensions than normally required for
manual use.
Typically the hard height of a syringe has never been regarded as a particu-
larly critical dimension. Its tolerances are ordinarily relatively loose. The
hard-height dimension is a function of not only the total length of plunger
red and the barrel, but also the thickness of the piston and finger grips. The
piston thickness, by virtue of its relatively unsophisticated manufacturing
process, can vary considerably. Because all these components are manu-
factured in multicavity moulds from many moulds around the world, the
cumulative extreme tolerance buildup from cavity to cavity and mould to
mould and location to location is such that these previously noncritical
dimensions cannot be instantly tightened.
3 Syringe identification
iv
© ISO ISO 7886-2:1996(E)
use, such as mechanical sensing of the syringe outside diameter, are not
deemed feasible in the long term. This is due to overlapping ranges of
diameter of syringes produced by different manufacturers, and the lack of
relationship between the outside and inside diameters of a syringe. It is
also recognized that standardization of syringe barrel diameters across the
industry is not a realistic option.
A means by which the pump could automatically identify the syringe model
and use this to programme such information as barrel inside diameter,
plunger force and occlusion alarm settings is seen as the next stage of this
part of ISO 7886. A possible method of recognition is to identify the syr-
inge and nominal capacity by means of a marking code on the barrel,
printed at the same time as the syringe scale, and to use this to pro-
gramme the pump automatically. It is recommended that development of
such a system be worked on as soon as possible.
010
SUPPLIE DBYBSBU N DERLI C EN CEF ROMISOFORSUNP HARMAC EUTICALIN DUSTRI ESLI MITED - VADODARAON 01/02/2
1111110•M
IN
1
T
St
CE
This page intentionally left blank ar
TI
ut
in
ar
fil
bi
2
T
tr
tr
e
tic
p
p
n
S
t.
t.
E
INTERNATIONAL STANDARD © ISO ISO 7886-2:1996(E)
-•■■
•=a1m110••■
ISO 594-2:1990, Conical fittings with a 6 % (Luer) 6 Limits for acidity or alkalinity
taper for syringes, needles and certain other medical
equipment — Part 2: Lock fittings. Clause 6 of ISO 7883-1:1993 shall apply.
1) To be published .
1
© ISO
!SO 7586-2:1996(E)
Dimensions in millimetres
/ Limits for extractable mPtals 1:
T
ti
8 Lubricant t(
10 Graduated scale
11 Syringe design G 1
0
itical dimensions for the fit of the syringe in a syr-
inge bump shall be designated as shown in figure 1
and shall be as given in table 1.
cz)
All other dimensional and design requirements shall
be as specified in ISO 7886-1.
12 Piston/plunger assembly B Distance of the projection of the finger grips from the
outside surface of the barrel.
2
IS0
c ISO ISO 7883-2:1996(E)
netres
12.2 Fit of the piston in the barrel Table 2
Table 3
13 Nozzle
Maximum displacement of fluid
ml
13.1 Conical fitting
Syringe
Test pressure
nominal
The male conical fitting of the syringe nozzle shall be capacity kPall
in accordance with ISO 594-i and shall have a iocking
fitting in accordance with ISO 594-2. ml 7 40 70 90 133
..- 5 and < 10 0,03 0,1 0,15 0,2 0,25
13.2 Nozzle lumen -••• 10 and < 20 0,1 0,2 0,3 0,35 0,4
..-- 20 and <30 0,1 0,4 0,6 0,8 0,9
The nozzle lumen shall have a diameter of not less ..-. 30 and < 50 0,1 0,4 0,6 0,8 0,9
than 1,2 mm. ---- 50 0,4 1,2 1,5 1,8 2,1
1) 1 kPa = 7,5 mmH,0 (approx.) = 0,145 lbfiin2 (p.s.i.) (approx.).
14 Performance
SUPPLIED BY BSB UNDER LICENCEFROM ISO FOR SUN PHARMACEUTICAL INDUSTRIES LIMITED - VADODARA ON 01/02/2010
14.5 Plunger movement forces
14.1 Dead space
When tested as described in annex C, the maximum
Subclause 14.1 of ISO 7886-1:1993 shall apply. force (inclusive of any variation) required to initiate
movement of the piston, the maximum sustaining
force and the maximum allowable variation of the
14.2 Freedom from air and liquid leakage past sustaining force shall be as given in table 4.
the piston
Table 4
Subclause 14.2 of ISO 7886-1:1993 shall apply.
Piston movement forces
3
ISO 7886-2:1996(E) © ISO
a) the maintenance of sterility of the contents under b) the model identification, if a manufacturer offers
dry, clean and adequately ventilated storage con- mare than one product of the same nominal ca-
ditions; pacity.
c) adequate protection of the contents during normal The secondary container shall be marked in accord-
handling, transit and storage; ance with ISO 7886-1 and with the information listed
in 16.1 a) and b).
d) that once opened, the container cannot be easily
resealed, and it should be obvious that the con-
tainer has been opened.
16.3 Storage container
16 Labelling
16.5 Syringe barrel
The primary container shall be marked in accordance a) the manufacturer's name or logo;
with ISO 7886-1 and with the following information:
b) the model identification, if a manufacturer offers
a) the statement "Suitable for use with power-driven more than one product of the same nominal Ca-
syringe pumps" or equivalent; nacity.
4
© ISO ISO 7886-2:19961E)
Annex A
(normative)
ON 01/02/2 010
a different constant temperature, correct the results
to 20 °C. A.4.1 Time lag to achieve steady fiowrate
5
ISO 7886-2:1996(E) © ISO
NOTES A.4.1.7 Switch on the drive and begin the test period
at this instant. Allow the test to continue for a period
1 The balance should be tared just prior to starting the
of 2 h.
equipment, because the tare cycle takes a finite time and
the balance has to settle prior to starting the test.
A.4.1.8 Compute the flowrate at 0,5-min successive
2 The electronic balance shown in figure A.1 should be intervals and plot the tlowrate in ml/h against time, in
placed on an antivibretion mount and be protected from minutes, with a solid line on the graph (see A.4.4.1).
shock, vibration and draughts. The temperature should be Mark the 1 ml/h set flowrate on the graph using a
controlled during the tests as specified in A.2.2. dotted line (see figure A.2).
A.4.1.6 Set the test driver to a rate equivalent to a A.4.1.9 Determine the time taken from the instant of
flowrate of 1 mlih in the syringe under test. Set the start-up to achieve a steady flowrate by recording the
sampling interval of the computer to 0,5 min. time to reach 95 % of the set flowrate (see 14.3.1).
1 3 7 4 6
1.5
0,5
- 0,5
120
Test period, t (min)
Figure A.2 — Example of data gathered during first two hours of test
6
© ISO ISO 7886-2:1996(E)
A.4.4 Calculation of results A.4.4.2 Calculate Ep(max) and Eomin) for the 2-min
and 5-min observation-time windows using equations
A.4.4.1 Calculate the actual flowrate Qi , in millilitres
(2) and (3) respectively, over the analysis period t1
per hour, for each sampling interval for the first
(minutes) of the second hour of the test period.
120 min of the test period using equation (1).
SU PPLIED BYBSB UNDER L IC ENC E FROMISO F ORSUN PHAR MAC EUTICALIN DUSTRI ESLIMITED - VADODA RAON01/02/20 10
6G (m, – mi _ 1 )
(1) ithm as follows:
rp
Overall error. A (V
-5
E p Irntr,)
- 10
[tilt
1 2 3 L 5 32
Figure A.3 — Example of plot of data gathered during second hour of test
7
ISO 7866-2:1996(E)
A.4.4.3 Calculate the overall mean percentage Plot percentage variation Ep(maxl and Ep(min) against
flowrate error A using the following expression where observation window duration t, (minutes) and the
A is measured over the analysis period t, (the second overall mean percentage error A [derived from
hour of the test period): equation (4)1 measured over the analysis period ti
(minutes) of the second hour of the test period (see
100(Q— qv) example in figure A.3).
A=
qv
Indicate Ep(m„) and Ep(min) and the overall mean
60 (m — mk percentage flowrate error A by means of a solid line.
Q= i (nl/h)
tip Indicate the zero error by means of a dotted line.
8
c ISO ISO 7886-2:1996(E)
9
ISO 7886-2:1996(E) © ISO
Annex B
(normative)
B.2.1 Test apparatus, as shown in figure B.1. For B.3.6 By means of the three-way valve, connect the
syringes under test of nominal capacity less than syringe under test to the pressurizing syringe and the
50 ml, use a 1-ml pressurizing syringe. For syringes pressure gauge.
under test of nominal capacity 50 ml and above, use a
5-ml pressurizing syringe. B.3.7 Repeat B.3.4 and B.3.5. Calculate the differ-
ence between the volumes recorded in B.3.5 and
B.2.2 Distilled water, complying with grade 3 of B.3 7 for each test pressure and record it as the
ISO 3696. compliance of the syringe under test at that pressure.
0%11 ///,
1 4- 3 5
1 Pressurizing syringe with fine graduations
2 Pressure gauge
3 Syringe under test
4 Three-way valve
5 Nominal capacity mark
6 Securing clamp
Figure B.1 — Apparatus for determination of compliance
10
© ISO ISO 7886-2:1996(E)
Annex C
(normative)
C.1 Principle C.3.11 Ploi the forces required tc move the piston
during the second half of the test period.
A mechanical testing machine is used to expel water
while the force to move the piston is being recorded. C.3.12 For large syringes (i.e. greater than 30 ml
nominal capacity) and at low flowrates (i.e. less than
5 ml/h), repeat C.3.2 to C.3.11 three times, with the
fiducial line of the piston set at three different points
C.2 Apparatus along the syringe barrel, once for each setting.
Record the position of the piston seal in the c) the minimum force (Fmi n) required to sustain
C.3.1
movement of the piston. Disregard that portion of
barrel.
the trace that relates to the passage of the piston
through the parking position;
C.3.2 Fill the syringe with water (C.2.2) to beyond
the nominal capacity. d) the range of forces (F,x — Fmi n);
C.3.3 Attach the administration set and needle as e) the maximum total variation. expressed as a
C.3.4 Expel water until the fiducial line of the piston (Fmax — Fmin) 100
is at the nominal capacity mark of the syringe.
Fmax
C.3.5 Mount the syringe in the test fixture and
clamp the push-button to the driving head of the
mechanical testing machine (C.2.1). C.5 Test report
C.3.6 Set the translation speed of the piston to be The test report shall include at least the following
equivalent to the desired volumetric flowrate. information:
C.3.9 Measure and note the force required to initiate c) the minimum and maximum forces required to
movement of the piston. sustain movement of the piston;
C.3.10 Expel water for a period of 2 h, or until the d) the maximum total percentage variation of the
syringe is empty, whichever is shorter. sustaining force.
11
ISO /886-2:1990(E)
© ISO
Ii
oll
12
© ISO ISO 7886-2:1996(E)
Annex D
(informative)
D.2 Short-term accuracy It is assumed that the reference syringe driver will be
calibrated periodically and a certificate of accuracy
In order to comply with this part of ISO 7886, the provided. Design of such a device is regarded as
SUPPL IEDBYBSBUN DERLI CENCEFROMI SOF ORSU NPHAR MACEUTICALI N DUST RIES LI MITED -
13
ISO 7886.-2:1396(E) © 'so
Annex E
(informative)
Bib!iography
(11 ISO 595-1:1986, Reusable all-glass or metal-and-glass syringes for medical use — Part 1: Dimensions.
121 iSO 595-2:1987, Reusable all-glass or metal-and-glass syringes for medical use — Part 2: Design, perform-
ance requirements and tests.
(31 ISO 8537:1991, Sterile single-use syringes, with or without needle, for insulin.
14
Auelq -01 Ancuopap ;Bed sNi
SUPPLIED BY BSB UNDER LICENCE FROM ISO FOR SUN PHARMACEUTICAL INDUSTRIES LIMITED - VADODARA ON 01/02/2010
This pane intentionally left blank
,atavocx.ven. -
513%. I "It
Ir•at
H
0- 3
Iv
CD
4al Aneuop
SLPPL!ED BY BSB UNDER LICENCE FROM ISO FOR SUN PHARMACEUTICAL INDUSTRIES LIMITED - VADODARA ON 01/02/2010
ISO 7836-2:1996(E) ©
ICS 11.040.20
Descriptors: medical equipment, sterile equipment, disposable equipment, syringes, specifications, performance, dimensions, tests,
packaging, labelling.