Evaluation of Performance and Uncertainty of Infrared Tympanic


Thermometers

Article  in  Sensors · April 2010


DOI: 10.3390/s100403073 · Source: PubMed

CITATIONS READS

11 485

2 authors, including:

Chiachung Chen
National Chung Hsing University
37 PUBLICATIONS   717 CITATIONS   

SEE PROFILE

All content following this page was uploaded by Chiachung Chen on 20 December 2013.

The user has requested enhancement of the downloaded file.


Sensors 2010, 10, 3073-3089; doi:10.3390/s100403073
OPEN ACCESS

sensors
ISSN 1424-8220
www.mdpi.com/journal/sensors

Article

Evaluation of Performance and Uncertainty of Infrared


Tympanic Thermometers
Wenbin Chung and Chiachung Chen *

Department of Bio-industrial Mechatronics Engineering, National ChungHsing University, 250


Kuokuang Rd., Taichung 40227, Taiwan; E-Mail: bse@dragon.nchu.edu.tw

* Author to whom correspondence should be addressed; E-Mail: ccchen@dragon.nchu.edu.tw;


Tel.: +886-4-2285-7562; Fax: +886-4-2285-7135.

Received: 21 February 2010; in revised form: 13 March 2010 / Accepted: 23 March 2010 /
Published: 31 March 2010

Abstract: Infrared tympanic thermometers (ITTs) are easy to use and have a quick response
time. They are widely used for temperature measurement of the human body. The accuracy
and uncertainty of measurement is the importance performance indicator for these meters.
The performance of two infrared tympanic thermometers, Braun THT-3020 and OMRON
MC-510, were evaluated in this study. The cell of a temperature calibrator was modified to
serve as the standard temperature of the blackbody. The errors of measurement for the two
meters were reduced by the calibration equation. The predictive values could meet the
requirements of the ASTM standard. The sources of uncertainty include the standard
deviations of replication at fixed temperature or the predicted values of calibration equation,
reference standard values and resolution. The uncertainty analysis shows that the
uncertainty of calibration equation is the main source for combined uncertainty. Ambient
temperature did not have the significant effects on the measured performance. The
calibration equations could improve the accuracy of ITTs. However, these equations did not
improve the uncertainty of ITTs.

Keywords: infrared tympanic thermometer; uncertainty; calibration equation


Sensors 2010, 10 3074

1. Introduction

Body temperature is an indication to express the health condition or pathological state.


Measurement of body temperature has become an essentially diagnostic method for medical treatment.
There are two traditional methods to measure the body temperature. The first type is the glass mercury
thermometer. This thermometer is inexpensive and easy to use. However, the response time is from 3
to 5 minutes. The glass material is extremely fragile and can be dangerous to the human body. The
second type is the electronic digital thermometer. Its sensing element is made of a thermistor or
resistance detector. This meter can measure the temperature within several seconds. However, the
electronic device is affected by aging problems. The sensing elements of digital thermometer still need
to have contact with the human body. Several problems exist in the clinical operation. The patient
reaction, such as children or infant, could affect the measurement of these contact thermometers.
The best method is to measure the core body temperature, such as the temperature of coronary
arteries. However, this is impossible except by using invasive surgical procedures. Recently, many
literatures reported that the core temperature can be measured by detecting the positions near the
membrane of the ear canal [1]. The infrared tympanic thermometer was developed to serve as a
detector for medical applications. The construction and operating principle of infrared tympanic
thermometer have been introduced in detail [2,3].
The reliability and accuracy of infrared tympanic thermometer have been discussed by many
researchers. Their results are inconsistent. Dodd et al. [4] compared the reading values of infrared ear
temperature for children aged between 0 and 18 with that of rectal thermometry. Their conclusions
indicated that the infrared ear thermometer would fail to detect fever in 75% of febrile children.
Craig et al. [5] found that the pooled mean temperature difference for rectal temperature minus
infrared ear temperature was 0.3 °C. The significant difference (significance was taken as p < 0.05)
was found between two sets of data. These authors suggested that the infrared ear thermometer did not
indicate the sufficient agreement with the body temperature measured by rectal temperature.
Kistemaker et al. [6] evaluated the performance of an infrared forehead thermometer. They concluded
that this Sensor Touch meter could work well in stable conditions. The average difference between the
infrared forehead thermometer and a rectal sensor ranged from 0.3 to 0.5 °C. Kocoglu et al. [7]
compared three body temperatures. The rectal and auxiliary temperatures were measured with glass
mercury thermometers. The aural temperature was measured by an infrared thermometer. They
concluded that the infrared tympanic thermometer could be applied in an emergency room setting.
Rosenthal and Leslie [8] compared the accuracy of an electronic and an infrared thermometer with
traditional glass mercury thermometry in the 95% confidence level. They found the average difference
between the reading values of infrared thermometer and glass thermometer was within 0.1 °C.
Stavem et al. [9] assessed the accuracy of infrared ear thermometry, by measuring the rectal and
esophageal temperature with thermistor thermometer and ear temperature by infrared thermometer.
The mean of two ear temperatures had better agreement with the rectal temperatures.
The inconsistency of the measurement may be explained by the performance, confidence level, and
uncertainty of the infrared thermometer. The factors affecting the performance of infrared tympanic
thermometer were discussed by Heusch et al. [10]. Their results indicated that the handedness, sex and
age were the significant factors conflicted the accuracy of the ear temperature measurement. Pusnik
Sensors 2010, 10 3075

and Drnovsek [11] found that several factors could affect the performance of infrared ear
thermometers; such as the position of thermometer related to the aperture of the blackbody, the drift of
a thermometer due to heating, the amount of times probe covers were used and the differences of
probe covers.
The calibration of the infrared thermometer is very important to ensure its performance. Because
the emissivity of canal is very close to unity, the temperature of a black body cavity is usually served
as the standard temperature for calibration. Cascetta [2] developed a blackbody cavity that consisted of
a copper cylinder. This cylinder contacted with the copper plate and maintained a constant temperature
by circulating water provided by a hot water bath, Pusnik et al. [12] compared the measurements of
infrared thermometer performed at several blackbodies. Three cavity shapes were found that could be
served as suitable standards for IR calibrations. Simpson et al. [13] described a commercial ear
thermometer calibrator that could be traceable to ITS-90.
Pusnik et al. [3] defined some important terms for the measurement of the infrared thermometer.
The performance of accuracy is closeness of the measurement result of thermometer and the true value
of a measurement. The difference between the average value of several measurements and the true
value was called systematic error. The diversity of individual measurements is presented as random
errors. However, the variability of measurements was not the only source that induced from the
difference between one measured values and average values of several measurements. Other
components could produce a variability source for measurement. The uncertainty of measurements
have been defined and explained by the guide to the expression of uncertainty [14]. The components of
uncertainty of an infrared ear thermometer were listed [3,12]. These components included the
repeatability of an IR thermometer, the reference thermometer, the blackbody radiator, the transducer
of instrument and the resolution of the IR thermometer.
Recently, uncertainty evaluation had been widely applied for physical and chemical sensors [15-17].
The effects of calibration equation on the measurement performance have been studied [18,19]. It is
very useful to study the influence of the factors affecting the performance of the IR thermometer.
As far as the authors know, there have been no reports of uncertainty evaluations of an infrared
tympanic thermometer.
The objectives of this study are to evaluate the accuracy and to calculate the uncertainty of two
types of infrared tympanic thermometer according to ISO GUM (Guide to the expression of
uncertainty in measurement). An adequate calibration equation is first established. Then the effect of
the calibration equation on the accuracy and uncertainty was compared.

2. Equipment and Methods

2.1. Infrared Tympanic Thermometer

Two types of infrared tympanic thermometers were adopted in this study. One is the OMRON
MC-510 Gentle Temp model (OMRON Co., Japan). The other is the BRAUN IRT-3020 Themoscan
model (Braun Co., Germany). The operating procedures for ear temperature measurement were
described in detail in their manual. The specifications of the two thermometers are listed in Table 1.
Sensors 2010, 10 3076

Table 1. Specification of the infrared clinical thermometers.


OMRON BRAUN
MC-510 IRT-3020
Sensing element thermopile thermopile
Measuring range 34–42.2 °C 34–42.2 °C
Resolution 0.1 °C 0.1 °C
Nonlinearity and 1. 36.0–39 °C, ±0.2 °C 1. 37.0–39 °C, ±0.1 °C
repeatability 2. (≤36.0, ≥39 °C), ±0.3 °C 2. (≤37 °C, ≥39 °C), ±0.2 °C

2.2. Standard Temperature

The standard temperature was prepared by a temperature calibrator and a blackbody cavity. This
standard temperature was simulated as the tympanic temperature to evaluate the performance of
two types of infrared thermometers. The standard temperature served as the standard for
performance testing.
The model of temperature calibrator is TC-2000 Scan Sense (Instrutek Co., Norway). The operating
temperature was ranged from –40 °C to 150 °C. The inner temperature of oil bath for calibration was
detected by a four-wire Pt-100 thermometer. The uncertainty of this calibrator is 0.03 °C according to
its specifications.
An aluminum cylinder was inserted into the oil bath of the calibrator. The length of the cylinder
was 14.5 cm and the diameter was 0.485 cm. The emissivity of this cylinder was assembled to be unity
calculated from the calculation equation of blackbody cavity [20].

2.3. Testing Procedures

The target temperature for calibration of infrared thermometer was maintained at 34.5, 36.0, 37.5,
39.0 and 40.5 °C. The performance test experiment was executed in an environmental chamber at three
levels of ambient temperature: 25 °C, 30 °C and 35 °C. The variation of the setting temperature was
controlled within ±1 °C.
Each infrared thermometer was randomly measured at five sets of standard temperatures at the same
ambient temperature. Five replicates were made for each standard temperature. As one measurement
was made, the infrared thermometer was taken out of the blackbody cavity for five minutes and then
put into the cavity again for further measurement.

2.4. Data Analysis

The performance of infrared thermometers was assessed by their accuracy and repeatability. The
definition of accuracy is the closeness with which a measurement approaches true value [11]. The error
is defined as:
Ei = Tr − Ts (1)
where Tr is reading value of the infrared thermometer in °C; and Ts is standard value of the infrared
thermometer in °C.
Sensors 2010, 10 3077

The standard deviation of data sets was calculated from replicates. These values were then further
analyzed by one and two way ANOVAs (Analysis of Variance) and other statistical procedure.
Significance was taken as P < 0.05. The P value is the smallest level of significance that would be to
reject the hypothesis.

2.5. Establish the Calibration

The regression analysis technique was applied to establish the calibration equation from the
relationship between the standard values and the reading values. The criteria for selecting of the best
equation are the coefficient of determination R2, t-tests of each parameters and residual plots [21,22].
To establish the calibration equation, the standard values of the standard temperature made by a
temperature calibrator were selected as the dependent value y. The reading values of the infrared
thermometer were viewed as an independent variable. The ambient temperature of the experiment
testing is assumed as the other independent variable. This calibration equation f(x) is called as the
inverse equation [23-25]:
y = f ( xi ) (2)

The form of y(x) can be expressed as follows:


y = b0 + b1 X 1 + b2 X 12 (3)

where X1 is the reading values of infrared thermometer, b0, b1, b2 are parameters.
Many texts dealing with the calibration equation reported that the reading values of the
thermometer were viewed as the independent variable and the standard values of temperature were
viewed as the dependent variable. This equation is called as the classical equation. However, the
inverse equation has shown the better predictive ability [23-25].

3. Sources of the Uncertainty for Infrared Tympanic Thermometer

According to the ISO GUM [14], the uncertainty of measurement is evaluated by “Type A” and
“Type B” methods. The Type A evaluation of standard uncertainty is the evaluation method by
statistical procedure. The Type B evaluation of standard uncertainty is the evaluation method by other
information of the measurement and instrument. There are several items of uncertainty source for
infrared tympanic thermometer.

3.1. The Standard Deviation

The standard uncertainty due to the estimate values of standard deviation of replicates ux is a Type
A uncertainty. It was easy to calculate from the replicates of the measurement made using the IR
thermometer at different conditions.

3.2. The Calibration Equation

The standard uncertainty due to the calibration of infrared thermometer is a Type A uncertainty.
The uncertainty in a predicted value uy is calculated by follows:
Sensors 2010, 10 3078

1 1 ( y − y)2
uy = s + + (4)
p n ∑ ( yi2 ) − (∑ yi ) 2 n

where s is standard deviations of calibration equation, p is the number of replicates in the calibration;
and n is the number of the sample data.

3.3. Uncertainty of the Reference Temperature

The uncertainty source of reference standard was provided by the manufacturer’s specifications.
The distribution of uncertainty was assumed as normal distribution. The estimate of uncertainty for the
standard temperature is:
U ref
uref = (5)
2
where Uref is the uncertainty source of TC-2000 temperature calibrator, and uref is the uncertainty due
to the reference temperature.

3.4. Uncertainty Due to Nonlinearity and Repeatability

The source Unon due to nonlinearity and repeatability is specified by manufacturers. The variation
response for this error source is assumed as a rectangular distribution. The uncertainty due to nonlinear
and repeatability unon is calculated as:
U non
unon = ± (6)
2 3
where Unon is the uncertainty source due to the nonlinear and repeatability that represents a
specification interval provided by the manufacturer; and unon is the uncertainty due to the nonlinear and
repeatability made using the IR thermometer.

3.5. Uncertainty Due to Resolution

The uncertainty measurement due to resolution is assumed to be a rectangular distribution. It is


considered as ±½ of the display scale value. The uncertainty value due to resolution ures is estimated as
the follows:
U res
ures = ± (7)
2 3
where Ures is the uncertainty source due to the resolution effect, and ures is the uncertainty due to the
resolution effect.

3.6. Uncertainty Due to Ambient Temperature Variation

The uncertainty source of ambient temperature was not mentioned by manufacturers. In this study,
the effect of the ambient temperature on the variation of measurements made using the IR thermometer
was evaluated by two-way ANOVA test. If the ambient temperature has significant effects on the
variation of measurement, the uncertainty due to the ambient temperature can then be further
Sensors 2010, 10 3079

estimated. The variation response from the temperature variation is assumed a rectangular distribution.
The uncertainty due to temperature variation is:
U tem
utem = ± (8)
3
where Utem is the uncertainty source due to the ambient temperature; and utem is the uncertainty due to
the ambient temperature.
The uncertainty due to the reference temperature, nonlinearity and repeatability, resolution, and
ambient temperature are classified as Type B uncertainty.
These uncertainty effects did not have significant correlations, and the combined standard
uncertainty uc then be calculated as follows:
a. Uncertainty estimation of a fixed measuring point.

uc1 = u x2 + uref
2
+ unon
2
+ ures
2
+ utem
2
(9)

b. Uncertainty estimation include the effect of calibration

uc2 = u y2 + uref
2
+ unon
2
+ ures
2
+ utem
2
(10)

4. Results and Discussions

4.1. Performance of OMRON MC-510 Thermometer

The relationship between reading values of MC-510 thermometer versus standard values is shown
in Figure 1. The reading values of this tympanic thermometer are close to each other. The effect of
ambient temperature on the reading values was inconstant. The error distribution of OMRON MC-510
thermometer at three ambient temperatures is shown in Figure 2. The error band is ±0.3 °C and the
error values moves from positive to negative with the increase of standard temperature. According to
the standard of the ASTM standard for allowable errors of clinical thermometers is 0.1 °C in the
temperature range 37–39 °C, 0.2 °C in the temperature range 36–37 °C and 39–41 °C. The allowable
error of clinical thermometers was within a band of ±0.3 °C in the range below 36 °C. From the data
distribution of Figure 2, the performance of reading values was corresponding with the requirement of
the ASTM standard in the temperature range below 37 °C. However, the reading values were diversity
in the temperature of 37.5 °C. Only the average values of several measurements could reduce
the measured errors. As the higher temperature ranged higher than 38 °C, the reading values
were underestimated.
Sensors 2010, 10 3080

Figure 1. The relationship between reading values of MC-510 thermometer versus


standard values.
Ta = 25 C Ta = 30 C Ta = 35 C

41
Read i n g v al u es o f i n frared t h ermo met er

40

39

38

37

36

35

34
34 35 36 37 38 39 40 41
St an d ard t emp erat u re, C

Figure 2. The error distribution of OMRON MC-510 thermometer at three


ambient temperatures.

Ta = 25 C Ta = 30 C Ta = 35 C

0.4

0.2
E rro rs s o f i n frared t h ermo met er, C

-0.2

-0.4

-0.6

-0.8

-1
34 35 36 37 38 39 40 41
St an d ard t emp erat u re, C

The relationship between standard temperature values and reading values of MC-510 thermometer
was established by regression analysis. The best equation is:
Tsta = 25.1634 − 0.4739 Tred + 2.1563Trea 2 (11)

R 2 = 0.992, s = 0.1965

where Tsta is the standard temperature and Trea is the reading temperature of IR thermometer.
Sensors 2010, 10 3081

The predicted errors of calibration Equation (11) are presented in Figure 3. All the predicted errors
in the range of standard temperature were below 37 °C and higher than 39 °C all within the ASTM
standard. However, the predicted errors at the standard temperature of 37.5 °C revealed high spreading
of the measurement. Above the standard temperature of 37 °C, the accuracy of the MC-510 infrared
tympanic thermometer had improved significantly.

Figure 3. The predicted errors of calibration equation of OMRON MC-510 thermometer.


0. 6

0. 4
res i d u al s o f p red i ct ed eq u at i o n s

0. 2

-0. 2

-0. 4

-0. 6
34 35 36 37 38 39 40 41
St an d ard t emp erat u re, C

4.2. Performance of BRAUN IRT-3020 Thermometer

The relationship between reading values of BRAUN IRT-3020 thermometer versus standard values
is shown in Figure 4. The closeness of the data distribution at each standard temperature indicated a
good replication of the performance. The error distribution of this infrared thermometer at three
ambient temperatures is presented in Figure 5. The error bands were within 0.2 °C of the whole
measuring range except at 37 °C. The reading values of IRT-3020 thermometer could meet the
requirements of the ASTM standard in the range below 37 °C. Above the standard temperature
of 37 °C, the errors values of this thermometer were higher than the ASTM standard.
Sensors 2010, 10 3082

Figure 4. The relationship between reading values of BRAUN IRT-3020 thermometer


versus standard values.

Ta = 25 C Ta = 30 C Ta = 35 C

42

41
Reading values of infrared thermometer, C

40

39

38

37

36

35

34
34 35 36 37 38 39 40 41
Standard temperature, C

Figure 5. The error distribution of BRAUN IRT-3020 thermometer at three ambient


temperatures.

T a = 25 C T a = 30 C T a = 35 C

0. 5
0. 45
0. 4
0. 35
0. 3
Errors, C

0. 25
0. 2
0. 15
0. 1
0. 05
0
34 35 36 37 38 39 40 41
Standard temperature, C

The relationship between standard values and reading values of IRT-3020 thermometer was
established by regression analysis. The adequate equation is:
Tsta = 0.2637 + 0.9871Tred (12)

R 2 = 0.999, s = 0.0696
Sensors 2010, 10 3083

The predicted errors of Equation (12) are shown in Figure 6. All the predicted errors were located
within the requirement range of the ASTM standards except for the standard temperature of 37.5 °C.
However, the mean value of several replicates at this temperature could reduce the predictive errors.

Figure 6. The predicted errors of calibration equation of BRAUN IRT-3020 thermometer.


0. 25

0. 2
Res i d u al s o f p red i ct ed eq u at i o n , C

0. 15

0. 1

0. 05

-0. 05

-0. 1

-0. 15

-0. 2

-0. 25
34 35 36 37 38 39 40 41
St an d ard t emp erat u re, C

4.3. The Replicate of the Infrared Tympanic Thermometer

The standard deviation of five measurements of MC-510 thermometer at different standard


temperatures and three levels of ambient temperatures are shown in Figure 7.

Figure 7. The standard deviation of five measurements of OMRON MC-510 thermometer


at different standard temperatures and three leaves of ambient temperatures.
25 C 30 C 35 C

0.1

0.09

0.08

0.07
St an d ard d ev i at i o n , C

0.06

0.05

0.04

0.03

0.02

0.01

0
34 35 36 37 38 39 40 41
Ref. t emp w eat u re, C

At the standard temperature of 36 °C, the standard deviations for three ambient temperatures were
higher than that of other standard temperatures. To evaluate the factors affecting the standard
Sensors 2010, 10 3084

deviations of MC-510 thermometer, two-way ANOVA table was established and the significant tests
were executed by F test with a 95% confidence level. The result is listed in Table 2.

Table 2. Analysis of variance ANOVA table for the effect of standard temperature and
ambient temperature on the measurement of OMRON MC-510 thermometer.
Variance SS df MS F P -value F-critical
value
Standard temp. 0.003717 4 0.000929 5.9318 0.01614 3.8379
Ambient temp. 0.000832 2 0.000416 2.6552 0.1305 4.4590
Errors 0.001253 8 0.000157
Total 0.005802 14
Note: SS is the sum of square, df is the degree of freedom, MS is the mean square and
F-critical value is the smallest level of significance that be to reject the hypothesis of F test.

In comparison with the variations of measurements, the standard temperature has significantly
affected the variations of standard deviations. However, the ambient temperature showed no
significant effects on the variation of measurements (P = 0.1305>0.05). The ambient temperature did
not influence the variation of standard deviation of each measurement. The uncertainty source of
ambient temperature was not considered for the evaluation of uncertainty of the MC-510 thermometer.
The distribution of standard deviation of the Braun IRT-3020 measurement is presented in Figure 8.
The ambient temperature did not have the significant effect on the standard deviations. At the standard
temperature of 39 °C, the standard deviations were closed at 39 °C.

Figure 8. The standard deviation of five measurements of BRAUN IRT-3020 thermometer


at different standard temperatures and three leaves of ambient temperatures.

25 C 30 C 35 C

0. 05
0. 045
0. 04
St an d ard d ev i at i o n , C

0. 035
0. 03
0. 025
0. 02
0. 015
0. 01
0. 005
0
34 35 36 37 38 39 40 41
Ref. t emp erat u re, C
Sensors 2010, 10 3085

The two-way ANOVA table for the evaluation of the factors affecting the standard deviations of the
IRT-3020 thermometer is listed in Table 3. The results indicated that the standard temperatures
(P = 0.9355 > 0.05) and the ambient temperatures (P = 0.2274 0.05) did not have the significant
effects on the standard deviation of IRT-3020 measurements. The effect of ambient temperature on the
uncertainty of IRT-3020 was not considered in further analysis.

Table 3. Analysis of variance ANOVA table for the effect of standard temperature and
ambient temperature on the measurement of BRAUN IRT-3020 thermometer.
Source SS df MS F P-value F-critical
value
Standard temp. 0.000133 4 3.321E-05 0.1926 0.9355 3.8379
Ambient temp. 0.000617 2 0.000309 1.7924 0.2274 4.4590
Errors 0.001377 8 0.000172
Total 0.002127 14

4.4. Calculation of the Uncertainty of Infrared Tympanic Thermometer

The adequate calibration equation for OMRON MC-510 thermometer is the polynomial equation
and the best calibration equation for BRAUN IRT-3020 is a linear form. The uncertainty due to the
replications of measurement is the standard deviation of each measurement. However, the uncertainty
derived from the calibration equation is calculated by Equation (4). The type A uncertainty of several
observations for two-types of infrared tympanic thermometers are listed in Table 4. For the OMRON
MC-510 thermometer, the uncertainty of calibration equation did not reduce as compared to the
uncertainty calculated from standard deviations of each observation. The same results could be found
for the BRAUN IRT-3020 thermometers.

Table 4. The type A uncertainty of several observations for two infrared tympanic thermometers.
yobs of different temperatures
Calibration equations
34.5 36.0 37.5 39.0 40.5
OMRON None 0.1799 0.2193 0.1767 0.1408 0.1506
MC-510 Polynomial equation 0.2026 0.1990 0.1986 0.1989 0.2100

BRAUN None 0.0884 0.0535 0.0567 0.0594 0.05606


IRT-300 Linear equation 0.0711 0.0704 0.0701 0.0703 0.0709

The type B uncertainty analysis for the two infrared tympanic thermometers was calculated by
Equation (5), (6) and (7). This result is listed in Tables 5–6. Comparing these sources of uncertainty
for two-type infrared tympanic thermometers, the main source of the uncertainty included the
specification of nonlinear and repeatability (type B ) and the standard deviations (type A).
Sensors 2010, 10 3086

Table 5. The type B uncertainty analysis for OMRON MC-510 thermometer.


Description Estimate value Standard Probability
uncertainty distribution
Reference (uref) 0.03 °C 0.0153 Normal
Resolution (ures) 0.1 °C 0.0289 Rectangular
Nonlinear and repeatability
Unon1 36-39 °C 0.2 °C 0.1155 Rectangular
Unon2 ≤ 36 °C, ≥39 °C 0.3 °C 0.1732

Table 6. The type B uncertainty analysis for BRAUN IRT-3020 thermometer.


Description Estimate value Standard Probability
uncertainty distribution
Reference (uref) 0.03 °C 0.0153 Normal
Resolution (ures) 0.1 °C 0.0289 Rectangular
Nonlinear and repeatability Rectangular
Unon1 37-39 °C 0.1 °C 0.0577
Unon2 ≤ 37 °C, ≥39 °C 0.2 °C 0.1155

4.5. The Combined Standard Uncertainty

The uc values for two infrared tympanic thermometers at five observations are listed in Table 7.

Table 7. The combined uncertainty for two IR thermometers.


IR yobs of different temperature
thermomete Calibration equation 34. 36. 37. 39. 40.5
r 5 0 5 0
OMRON None 0.2 0.2 0.2 0.1 0.23
MC-510 Polynomial equation 519 499 136 851 18
0.2 0.2 0.2 0.2 0.27
BRAUN None 685 324 321 246 41
IRT-300 Linear equation
0.1 0.0 0.0 0.0 0.13
491 873 872 890 24
0.1 0.0 0.0 0.0 0.13
395 965 965 967 94

According to Equation (9), the values of uc1 are calculated at 34.5, 36.0, 37.5, 39.0 and 40.5 °C of
the standard temperature for MC-510 thermometer. The numeric values are 0.2519, 0.2499, 0.2316,
0.1851 and 0.2318 °C, respectively. For the polynomial form of calibration equation, the combining
standard uncertainty calculated by Equation (10) to evaluate at these standard temperatures were
0.2685, 0.2324, 0.2321, 0.2246 and 0.2741, respectively. The result indicated that the polynomial
calibration did not decrease the uncertainty for the MC-510 thermometer.
The values of uc are calculated at 34.5, 36.0, 37.5, 39.0 and 40.5 °C of the standard temperature for
IRT-3020 thermometer. They are found to be 0.1491, 0.0873, 0.0872, 0.089 and 0.1324 by
Equation (9), respectively. For the linear calibration equation, the combined standard uncertainty
Sensors 2010, 10 3087

evaluated at these standard temperatures was 0.1395, 0.0965, 0.0965, 0.0967 and 0.1394 by
Equation (10), respectively. This result indicated that the calibration equation did not improve the
uncertainty of the IRT-3020 thermometer.
This study evaluated the accuracy and uncertainty for two types of infrared tympanic thermometers.
If the reading values of both thermometers were recognized as the true values of tympanic
temperature, the accuracy of these thermometers could meet the requirement of the ASTM standards in
the temperature range below 37 °C. At the standard temperature of 37.5 °C, there was wide variation
in measurement in both thermometers. The average value of several measurements could be used to
improve the accuracy. That is, more data needs to be taken at this temperature point to ensure the
accuracy. At the temperature range above 37.5 °C, the reading values of both types of infrared
tympanic thermometer need to be transformed by calibration equation to improve their accuracy.
The errors of measurement could be classified as systematic and random errors. The systematic
errors could be expressed as errors and improved by its calibration equation. These procedures have
been studied [18,26]. The random errors can only be improved by selecting of sensing elements or
increasing sample numbers. The quantitative expression of this random error has been established in
the international standards [14].
In the previous study, the uncertainty of measurement for thermometer, hygrometer and rough rice
moisture meter could be improved by establishing the adequate calibration Equations [18, 19, 26]. In
this study, the measuring errors could be reduced by the calibration equations. That is the systematic
errors could be decreased. However, the uncertainty of measurement for two infrared tympanic
thermometers could not be improved by the calibration equations. The reason could be explained by
the form of predictive uncertainty [Equation (4)]. The estimated values of standard deviations for
calibration, Equations (11) and (12), were 0.1965 and 0.0696, respectively. These numeric values are
all almost higher than the standard deviation of replicates for different standard temperature and
ambient temperature. That is the reason that the predictive uncertainty calculated by Equation (4) is
almost higher than that of standard deviations.
Many studies have been reported that compare the performance of the rectal thermometer and
infrared tympanic thermometer [4,5,7]. However, the performance of accuracy and uncertainty of the
traditional mercury glass thermometer and the infrared tympanic thermometer were not mentioned by
these researchers. That may be the reason for the inconsistency of their results.

5. Conclusions

The uncertainty evaluation in this study included the Type A uncertainty for standard deviations of
replicates or predicative uncertainty of calibration and the Type B uncertainty for reference
temperature, nonlinear and repeatability and resolution source. The ambient temperature variation
effect was excluded. The uncertainty sources of different operators or different parts of human body
were not discussed in this study. The uncertainty analysis has become the basic information for the
instrument. No literature or related reports were found that mentioned the evaluation of infrared
tympanic thermometers. The method of uncertainty calculation for two types of infrared
tympanic thermometers in this study has been developed. This method could be applied for other
infrared thermometers.
Sensors 2010, 10 3088

Acknowledgements

The authors would to thank the National Science Council of the Republic of China for financially
supporting this research under Contract No NSC-95-2313-B-005-019.

References

1. McCarthy, P.W.; Heusch, A.I. The vagaries of ear temperature assessment. J. Med. Eng. Tech.
2006, 30, 242-251.
2. Cascetta, F. An evaluation of the performance of an infrared tympanic thermometer.
Measurement 1995, 16, 239-246.
3. Pusnik, I.; Simpson, R.; Drnovsek, J. Bilateral comparison of blackbody cavities for calibration of
infrared ear thermometers between NPL and FE/LMK. Physiol. Meas. 2004, 25, 1239-1247.
4. Dodd, S.R.; Lancaster, G..A.; Craig, J.V. In a systematic review, infrared ear thermometry for
fever diagnosis in children finds poor sensitivity. J. Clin. Epidemiol. 2006, 59, 354-257.
5. Craig, J.V.; Lancaster, G.A.; Taylor, S. Infrared ear thermometry compared with rectal
thermometry in children: a systematic review. Lancet 2002, 360, 603-609.
6. Kistemaker, J.A.; Den Hartog, E.A.; Daanen, H.A.M. Reliability of an infrared forehead skin
thermometer for core temperature measurements. J. Med. Eng. Tech. 2006, 30, 252-261.
7. Kocoglu, H.; Goksu, S.; Isik, M. Infrared tympanic thermometer can accurately measure the body
temperature in children in an emergency room setting. Int. J. Pediatr. Otorhinolaryngol. 2002, 65,
39-43.
8. Rosenthal, H.M.; Leslie, A. Measuring temperature of NICU patients – A comparison of three
devices. J. Neonatal Nurs. 2006, 12, 125-129.
9. Stavem, K.; Saxholm, H.; Smith-Erichsen, N. Accuracy of infrared ear thermometry in adult
patients. Intensive Care Med. 1997, 23,100-105.
10. Heusch, A.I.; Suresh. V.; McCarthy, P.W. The effect of factors such as handedness, sex and age
on body temperature measured by an infrared ‘tympanic’ thermometer. J. Med. Eng. Tech. 2006,
30, 235-241.
11. Pusnik, I.; Drnovsek, J. Infrared ear thermometers-parameters influencing their reading and
accuracy. Physiol. Meas. 2005, 26, 1057-1084.
12. Pusnik, I.; van der Ham, E.; Drnovsek, J. IR ear thermometers: what do they measure and how do
they comply with the EU technical regulation. Physiol. Meas. 2004, 25, 699-708.
13. Simpson, R.; Machin, G..; McEvoy, H. Traceability and calibration in temperature measurement:
a clinical necessary. J. Med. Eng. Tech. 2006, 30, 212-217.
14. International Organization for Standardization. Guide to the Expression of Uncertainty in
Measurement; ISO: Geneva, Switzerland, 1993.
15. Eurachem. EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement,
Eurachem: Middlesex, UK, 2000.
16. Mohamed, M.I.; Aggag, G.A. Uncertainty evaluation of shore hardness testers. Measurement
2003, 33, 251-257.
17. Fisicano, P.; Adriaens, A.; Ferrara, E. Assessment of the uncertainty budget for the amperometric
measurement of dissolved oxygen. Anal. Chim. Acta 2007, 597, 75-81.
Sensors 2010, 10 3089

18. Chen, C. Evaluation of measurement uncertainty for thermometers with calibration equations.
Accredit. Qual. Assur. 2006, 11, 75-82.
19. Lu, H.; Chen, C. Uncertainty evaluation of humidity sensors calibrated by saturated salt solutions.
Measurement 2007, 40, 591-599.
20. Nicholas, J.V.; White, D.R. In Radiation Thermometry Traceable Temperature, 1st ed.;
Sudenham, P.H., Sussex, T.W., Eds.; John Willey and Sons: Chichester, West Sussex, UK, 1994,
pp. 283-326.
21. Myers, R.H. Classical and Modern Regression with Application, 2nd ed.; PWS and Kent
Publishing: Boston, MA, USA, 1996.
22. Bruggemann, L.; Wennrich, R. Evaluation of measurement uncertainty for analytical procedures
using a linear calibration function. Accredit. Qual. Assur. 2002, 7, 269-273.
23. Krutchkoff, R.G. Classical and inverse regression methods of calibration. Technometrics 1967, 9,
25-439.
24. Grientschnig, D. Relationship between prediction errors of inverse and classical calibration.
Fresenius J. Anal. Chem. 2000, 367, 491-498.
25. Tellinghuisen, J. Inverse vs. classical calibration for small data sets. Fresenius J. Anal. Chem.
2000, 368, 585-588.
26. Chen, C. Uncertainty evaluation of conductance moisture meters for rough rice. Biosys. Eng.
2008, 99, 508-514.

© 2010 by the authors; licensee Molecular Diversity Preservation International, Basel, Switzerland.
This article is an open-access article distributed under the terms and conditions of the Creative