Unit I Bio Potential Generation and Electrodes Types: Prepared by Dr.K.Helenprabha, Mr.P.Arul
Unit I Bio Potential Generation and Electrodes Types: Prepared by Dr.K.Helenprabha, Mr.P.Arul
Unit I Bio Potential Generation and Electrodes Types: Prepared by Dr.K.Helenprabha, Mr.P.Arul
electrodes types
Prepared By
Dr.K.HelenPrabha, HOD/ECE
Mr.P.Arul, Assistant Professor
Introduction
• In order to measure and record potentials (currents) in the
body, it is necessary to provide some interface between the
body and the electronic measuring apparatus.
• Current flows in the measuring circuit for at least a fraction
of the period of time over which the measurement is
made.
• Bio potential electrodes is a transducer that convert the
body ionic current in the body into the traditional
electronic current flowing in the electrode.
• Current is carried in the body by ions, whereas it is carried
in the electrode and its lead wire by electrons.
Electrode change an ionic current into an electronic current
2
electrical behaviour
equivalent circuit
? ?
3
electrical behaviour
equivalent circuit
?
4
electrical behaviour
equivalent circuit
5
equivalent circuit
electrode-electrolyte
6
more precise approximation of
double layer – Randles circuit
electrode-electrolyte
7
Electrode-Electrolyte Interface
Oxidation reaction causes atom to lose electron
Reduction reaction causes atom to gain
electron
anion cation
a1 and a2 are the activity of the ions on each side of the membrane.
Ionic activity is the availability of an ionic species in solution to enter
into a reaction.
Note: ionic activity most of the time equal the concentration of the ion
For the general oxidation-reduction reaction
A + B C + D + ne−
The Nernst equation for half cell potential is
RT a a
E=E +
0
ln
C D
nF aA aB
1
1
Polarizable and Nonpolarizable Electrodes
1
2
chemical reactions
silver / silver chloride
13
The Silver/Silver Chloride Electrode
Advantage of Ag/AgCl is that it is stable in liquid that has large
quantity of Cl- such as the biological fluid.
For biological fluid where Cl- ion is relatively high
a − 1
Cl
E = EAg
0
+
RT
nF
ln aAg +
1
6
The Electrode-Skin Interface
For 1 cm2, skin impedance
reduces from approximately
200K at 1Hz to 200 at 1MHz.
Signal due to motion has low frequency so it can be filtered out when
measuring a biological signal of high frequency component such as
EMG or axon action potential. However, for ECG, EEG and EOG whose
frequencies are low it is recommended to use nonpolarizable
electrode to avoid signals due to motion artifact.
Must be considered:
–good adhesive connection to skin
–skin cleaning
–floating electrode
1
8
19
21
22
23
24
26
Disposable Foam-Pad Electrodes
2
7
28
30
32
NEEDLE ELECTRODES
Needle electrodes are generally used in clinical electro myography,
neuro graphy and other electrophysiological investigations under the
skin and in the deeper tissues.
3
7
38
Types of electrodes
39
Internal Electrodes
No electrolyte-skin interface
No electrolyte gel is required
➢The coating recedes after being used for several times and the
electrode must be discarded when this occurs. They are also color
coded.
BIOPOLAR (DOUBLE COAXIAL) NEEDLE ELECTRODE)
Contains two insulated wires within a metal cannula (is a tube that can be
inserted into the body, often for the delivery or removal of fluid).
The two wires are bared at the tip and provide the contacts to the patient.
The cannula acts as the ground
These electrodes are electrically symmetrical and have no sense of polarity.
CONCENTRIC(COAXIAL) CORE NEEDLE ELECTRODE
Contains both the active and reference electrode within the same structure.
Consists of an insulated wire contained within an hypodermic needle.
The inner wire is exposed at the tip which forms one electrode.
These needles have very stable electrical characteristics and are
convenient to use.
These electrodes are made by moulding fine platinum wire into hypodermic
needle having outside diameter less than 0.6mm.
One end is bevelled to expose he end of wire and provide easy penetration.
The surface area of the exposed tip is less than 00005mm sq.
44
45
46
Microelectrodes
•Electrode-electrolyte noise
•Noise at the electrolyte
•skin interface
•Motion artifact
•Electric and magnetic field interference
•Thermal noise
•Amplifier noise
•Noise from additional bioelectric events
•Other noise sources
55
Silver –Silver Chloride electrodes
Half cell potential is 2.5 mv only
Reduces the noise voltage and Increases the
stability electrochemically
Stabilizes the half cell potential- no movement
artifacts(variable electrochemical voltage)
Reduce the low frequency electrode- electrolyte
impedance
56
57
Contd..
Using large area electrode and bio
electric recorder of high input
impedance the distortion in the wave
form is reduced
58