ISO TECH IPS 3303D Instruction Manual
ISO TECH IPS 3303D Instruction Manual
ISO TECH IPS 3303D Instruction Manual
EN FR
Table of Contents
Table of Contents
OVERVIEW ................................................................................ 9
Introduction ............................................................. 9
Series Lineup / Main Features ................................ 11
Principle of Operation ............................................ 12
CV/CC Crossover Characteristics ........................... 14
SETUP .................................................................................... 20
Power Up .............................................................. 20
Load Cable Connection .......................................... 21
Output On/Off ........................................................ 22
Beep On/Off .......................................................... 22
Switch between channels ....................................... 23
Front Panel Lock ................................................... 23
OPERATION ............................................................................ 24
CH1/CH2 Independent Mode .................................. 24
CH3 Independent Mode ......................................... 26
CH4 Independent Mode ......................................... 28
CH1/CH2 Tracking Series Mode ............................. 29
CH1/CH2 Tracking Parallel Mode ........................... 34
APPENDIX ............................................................................... 48
Fuse Replacement ................................................. 48
SPECIFICATIONS ..................................................................... 49
FAQ ......................................................................................... 52
INDEX ..................................................................................... 53
4
SAFETY INSTRUCTIONS
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow
when operating the IPS X303 series and when keeping it in storage.
Read the following before any operation to insure your safety and to
keep the best condition for the IPS X303 series.
Safety Symbols
These safety symbols may appear in this manual or on the IPS X303 series.
5
IPS X303 Series Instruction Manual
Safety Guidelines
General Do not place any heavy object on the IPS X303 series.
Guidelines Avoid severe impacts or rough handling that leads to damaging the
IPS X303 series.
CAUTION
Do not discharge static electricity to the IPS X303 series.
Do not block or obstruct the cooling fan vent opening.
Do not perform measurement at circuits directly connected to Mains
(see note below).
Do not disassemble the IPS X303 series unless you are qualified as
service personnel.
(Measurement categories) EN 61010-1:2001 specifies the measurement
categories and their requirements as follows. The IPS X303 series falls
under category I.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on circuits
not directly connected to Mains.
Power Supply AC Input voltage: 100V/120V/220V/230V ±10%, 50/60Hz
Connect the protective grounding conductor of the AC power cord to
WARNING
an earth ground, to avoid electrical shock.
Fuse Fuse type: 100V/120V: T6.3A/250V, 220V/230V: T3.15A/250V
Make sure the correct type of fuse is installed before power up.
WARNING
To ensure fire protection, replace the fuse only with the specified type
and rating.
Disconnect the power cord before fuse replacement.
Make sure the cause of fuse blowout is fixed before fuse replacement.
Cleaning the IPS Disconnect the power cord before cleaning.
X303 series Use a soft cloth dampened in a solution of mild detergent and water.
Do not spray any liquid.
Do not use chemicals or cleaners containing harsh products such as
benzene, toluene, xylene, and acetone.
6
SAFETY INSTRUCTIONS
7
IPS X303 Series Instruction Manual
8
OVERVIEW
OVERVIEW
This chapter describes the IPS X303 series in a nutshell, including its
main features and front / rear panel introduction. After going through the
overview, follow the Setup chapter (page 20) to properly power up and
set operation environment.
Introduction
Overview The IPS X303 regulated DC power supply series, are light weight,
adjustable, multifunctional work stations. The IPS 2303S has two
adjustable voltage outputs. The IPS 3303S has three independent
outputs: two with adjustable voltage level and one with fixed level
selectable from 2.5V, 3.3V and 5V. The IPS 4303S has four independent
voltage outputs that are all fully adjustable. The IPS X303 series can be
used for logic circuits where various output voltage or current are
needed, and for tracking mode definition systems where plus and minus
voltages with insignificant error are required.
Independent / The three output modes of IPS X303 series, independent, tracking
Tracking Series / series, and tracking parallel, can be selected through pressing the
Tracking Parallel TRACKING key on the front panel. In the independent mode, the output
voltage and current of each channel are controlled separately. The
isolation degree, from output terminal to chassis or from output terminal
to output terminal, is 300V. In the tracking modes, both the CH1 and CH2
outputs are automatically connected in series or parallel; no need to
connect output leads. In the series mode, the output voltage is doubled;
in the parallel mode, the output current is doubled.
9
IPS X303 Series Instruction Manual
Constant Voltage/ Each output channel is completely transistorized and well-regulated, and
Constant Current works in constant voltage (CV) or constant current (CC) mode. Even at
the maximum output current, a fully rated, continuously adjustable output
voltage is provided. For a big load, the power supply can be used as a
CV source; while for a small load, a CC source. When in the CV mode
(independent or tracking mode), output current (overload or short circuit)
can be controlled via the front panel. When in the CC mode (independent
mode only), the maximum (ceiling) output voltage can be controlled via
the front panel. The power supply will automatically cross over from CV
to CC operation when the output current reaches the target value. The
power supply will automatically cross over from CC to CV when the
output voltage reaches the target value. For more details about CV/CC
mode operation, see page 14.
Automatic tracking The front panel display (CH1, CH2) shows the output voltage or current.
mode When operating in the tracking mode, the power supply will automatically
connect to the auto- tracking mode.
Dynamic load When used in audio production lines, the power supply will provide a
continuous or dynamic load using a jumper connector (JP101/JP401)..
When the connectors are connected to the position “ON”, a stable DC
current power will be provided for audio power amplifiers.
10
OVERVIEW
Series Lineup
Main Features
11
IPS X303 Series Instruction Manual
Principle of Operation
Main
Current
“OR” Gate Amplifier Rectifier &
Comparator
Filter
O/P
Instant Over
Voltage Voltage Series
Load
Amplifier Comparator Regulator
Protection
Rectifier The auxiliary rectifiers D1011 to D1014 provide bias voltage filtered by
the capacitors C102 and C103, for the pre-regulators U101 and U102.
They provide a regulated voltage for other modules.
Main Rectifier The main rectifier is a full wave bridge rectifier. It provides the power
after the rectifier is filtered by the capacitor C101, and then regulated via
a series-wound regulator, which is finally delivered to the output terminal.
12
OVERVIEW
Current Limiter U104 acts as a current limiter. When the current is over predetermined
rating, U104 is activated and decreases the current. U208 provides a
reference voltage. U206 is the inverter amplifier. U103 is a comparator
amplifier which compares reference voltage and feedback voltage, and
then delivers to Q102, which then calibrates the output voltage.
Overload When the unit is overloaded, Q107 activates to control the current
magnitude of Q102, to limit the output current. The relay control circuit
controls the power dissipation in the series-wound regulated circuit.
13
IPS X303 Series Instruction Manual
Background The IPS X303 series automatically switches between constant voltage
mode (CV) and constant current mode (CC), according to load condition.
CV mode When the current level is smaller than the output setting, the IPS X303
series operates in Constant Voltage mode. The indicator on the front
panel turns green (C.V.) The Voltage level is kept at the setting and the
Current level fluctuates according to the load condition until it reaches
the output current setting.
CC mode When the current level reaches the output setting, the IPS X303 series
starts operating in Constant Current mode. The indicator on the front
panel turns red (C.C.) The Current level is kept at the setting but the
Voltage level becomes lower than the setting, in order to suppress the
output power level from overload. When the current level becomes lower
than the setting, the IPS X303 series goes back to the Constant Voltage
mode.
Diagram Vout
Vmax
Constant
Voltage
Constant
Current
Iout
Imax
14
FRONT AND REAR PANELS
BEEP
3 PARA
Parallel
/ INDEP CURRENT
Push
COARSE/ FINE
Keys
4 SER
/ INDEP
FINE
Series Keys
A A
Output Key
0 5V , 1A 0 30V , 3A 0 30V , 3A 0 5V , 3A
SLAVE MASTER 5 10V , 1A
Lock Key
C.V. C.C. C.V. C.C . PAR. C.V. C.C. C.V. C.C.
POWER
CH4 CH2 GND CH 1 CH3
CH4 Output CH2 Output GND Terminal CH1 Output CH3 Output
Power CH4 CV/CC CH2 CV/CC/PAR CH1 CV/CC CH3 CV/CC
Switch Indicator Indicator Indicator Indicator
Display
Volt Meter Displays output current of each channel.
IPS 4303S: CH1/CH3 and CH2/CH4
IPS 2303S/3303S: CH1 and CH2
IPS X303S (5 digits) IPS 3303D (3 digits)
15
IPS X303 Series Instruction Manual
Control Panel
MEMORY
RECALL / SAVE
Memory Keys Saves or recalls panel settings. Four settings, 1 to
1
4, are available. For save/recall details, see page
2
36.
3
CH2/4
BEEP
Parallel/Series PARA
/INDEP Activates Tracking Parallel operation or Tracking
Keys Series operation, For details, see page 28.
SER
/INDEP
Lock Key LOCK Locks or unlocks the front panel keys (excluding
UNLOCK the OUTPUT key). Pressing the LOCK key will
also exit remote mode if the machine is in remote
mode. For details, see page 23.
16
FRONT AND REAR PANELS
VOLTAGE
Voltage Knobs Push Adjusts the output voltage level for the selected
COARSE/FINE
channel. Pressing the knob switches coarse and
FINE
fine level setting.
CURRENT
Current Knobs Push Adjusts the output current level for the selected
COARSE/FINE
channel. Pressing the knob switches coarse and
FINE
fine level setting.
POWER
Power Switch Turns On or Off the main power. For power up
sequence, see page 20.
Terminals
GND
GND Terminal Accepts a grounding wire.
0 30V , 3A
CH1 Output MASTER Indicates CH1 Constant Voltage or Constant
CH1 CV/CC C.V. C.C.
Current state.
Indicator CH1
COM +
+
17
IPS X303 Series Instruction Manual
0 30V , 3A
CH2 SLAVE Indicates CH2 Constant Voltage, Constant
CV/CC/PAR C.V. C.C. PAR.
Current, or Tracking Parallel operation mode.
Indicator CH2
SERIES
0 5V , 3A
CH3 Overload 5 10V , 1A Indicates CH3 Constant Voltage or Constant
Indicator C.V. C.C.
Current state for the IPS 4303S.
CH3
CH3 Output Outputs CH3 voltage and current.
CH3 Overload OVER LOAD Indicates when CH3 (IPS 3303S) output current is
Indicator overloaded.
2.5V 3.3V 5V , 3A
CH3 Voltage Selects CH3 output voltage for the IPS 3303S:
Selector 2.5V, 3.3V, or 5V.
0 5V , 1A
CH4 CV/CC Indicates CH4 constant voltage or constant
Indicator current for the IPS-4303S.
C.V. C.C.
1 CH1/3 FINE
V V 2 CH2/4
BEEP
PARA
3 /INDEP CURRENT
Push
COARSE/FINE
SER
4 /INDEP FINE
A A
LOCK OUTPUT
UNLOCK
: Long Push
0 5V , 1A 0 30V , 3A 0 30V , 3A 0 5V , 3A
SLAVE MASTER 5 10V , 1A
18
FRONT AND REAR PANELS
Power Cord /
USB Connector Cooling Fan Fuse Socket AC Selector
WARNING
TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE GROUNDING
CONDUCTOR MUST BE CONNECTED TO GROUND.
FOR CONTINUED FIRE PROTECTION. REPLACE FUSE ONLY WITH 250V FUSE
OF THE SPECIFIED TYPE AND RATING.
NO OPERATOR SERVICEABLE COMPONENTS INSIDE. DO NOT REMOVE
COVERS. REFER SERVICING TO QUALIFIED PERSONNEL. AC SELECTOR
AC
SER. NO. LB
REPLACE FUSE
AS SPECIFIED
FUSE RATING
100V T 6.3A
120V 250V
220V T 3.15A
230V 250V
380 WATTS
490 VA MAX
50/60 Hz
Power Cord / Fuse The power cord socket accepts the AC mains:
Socket 115V/230V, 50/60Hz. For power up details, see page
20.
19
IPS X303 Series Instruction Manual
SETUP
This chapter describes how to properly power up and configure the IPS
X303 series before operation.
Power Up
Connect AC power Connect the AC power cord to the rear panel socket.
cord
POWER
Power On Press the Power switch to turn on the power. The
display shows the initialization screen with the
model name (IPS 3303S shown), followed by the
last recalled settings.
POWER
Power Off Press the Power switch again to turn off the power.
20
SETUP
Wire type When using load cables other than the attached, make sure they have
enough current capacity for minimizing cable loss and load line
impedance. Voltage drop across a wire should not excess 0.5V. The
2
following list is the wire current rating at 450A/cm .
Wire size (AWG) Maximum current (A)
20 2.5
18 4
16 6
14 10
12 16
21
IPS X303 Series Instruction Manual
Output On/Off
Panel operation Pressing the Output key turns on all channel OUTPUT OUTPUT
outputs.
The key LED also turns on. Pressing the Output key again turns the
output and the key LED off.
Automatic output Any of the following actions during output on automatically turns it off.
off They might involve sudden and harmful change in the output level.
Change the operation mode between independent / tracking series /
tracking parallel
Recalling other setups from the memory
Storing the setup into the memory
Beep On/Off
Panel operation By default, the beep sound is enabled. To turn off the CH2 CH2/4
beep, press the CH2 or CH2/CH4 key for 2 seconds. BEEP BEEP
A beep will be heard and the beep setting will be turned off. To enable
the beep, press the CH2 or CH2/CH4 key again for 2 seconds.
List of beep The following operations beep when the beep setting is on.
Power on Output on/off
INDEP – SER – PARA mode Panel lock/unlock
switching CH1/CH2 output level knob
Setup save/recall switching
Voltage/current knob fine/coarse Voltage/current level reaching
switching minimum (zero) level
22
SETUP
Panel operation Switching between channels only applies to IPS CH4 CH2 CH1 CH3
4303S.
Panel operation Press the LOCK key to lock the front panel key LOCK LOCK
turns on. To unlock, press the LOCK key for 2 seconds. The key LED
also turns off.
operation.
23
IPS X303 Series Instruction Manual
OPERATION
V V
A A
CH2 CH1
LOAD LOAD
2. Connect the load to the front panel terminals, CH1 +/−, CH2 +/−.
24
OPERATION
3. Set the CH1 output voltage and current. Press (For CH1)
the CH1 key (LED turns on) and then use the
CH1/3 CH1/3
Voltage and Current knob.
By default, the Voltage and Current knob work VOLTAGE
Push
CURRENT
Push
COARSE/FINE COARSE/FINE
(Fine control)
VOLTAGE
Push
COARSE/FINE
FINE
5. To turn on the output, press the output key. The OUTPUT OUTPUT
CH2 CH1
(CH1)
C.V. C.C. PAR.
CH2
(CH2)
25
IPS X303 Series Instruction Manual
Background / For the IPS 3303S the CH3 rating is fixed at 2.5V/3.3V/5V, 3A. CH3 for
Connection the IPS 4303S is variable: 0~5V,0~3A / 5.001~10V,0~1A .
V V
A A
CH3
LOAD
No Tracking CH3 does not have tracking series/parallel mode. Also, CH3 output is not
Series/Parallel affected by CH1 and CH2 modes.
Panel operation 1. Connect the load to the front panel CH3 +/−
terminal. (the diagram shows non-European
terminals)
26
OPERATION
2.5V 3.3V 5V , 3A
2. IPS 3303S: Select the output voltage,
2.5V/3.3V/5V using the CH3 voltage selector
key.
IPS 4303S: Press the CH1/3 key to switch to
CH3 (The CH3 indicator will light). Use the CH1/3
3. To turn on the output, press the output key. The OUTPUT OUTPUT
CH3
3.2A, the overload indicator turns red and CH3
operation mode switches from Constant Voltage to
Constant Current. OVER LOAD
CH3
C.V. C.C.
4303S: When the output value exceeds the set
CH3
value, the C.V./C.C. indicator turns red. This
indicates that CH3 has switched from the constant
voltage to constant current. C.V. C.C.
CH3
27
IPS X303 Series Instruction Manual
V V
A A
CH4
LOAD
No Tracking CH4 does not have tracking series/parallel mode. The CH4 output is not
Series/Parallel affected by CH1 and CH2 modes.
Panel operation 1. Connect the load to the front panel CH4 +/−
4
terminal. (the diagram shows non-European
terminals)
28
OPERATION
CV → CC When the output value exceeds the set value, the C.V. C.C.
CH4
Background Tracking series operation doubles the Voltage capacity of the IPS X303
series by internally connecting CH1 (Master) and CH2 (Slave) in series
and combining the output to a single channel. CH1 (Master) controls the
combined Voltage output level.
The following describes two types of configurations depending on the
common ground usage.
Connection
V V
A A
LOAD
2. Connect the load to the front panel terminals, CH1+ & CH2− (Single
supply).
3. Press the CH2 key (LED turns on) and then CH2/4 CH2/4
use the Current knob to set the CH2 output BEEP BEEP
FINE
4. Press the CH1 key (LED turns on) and then CH1/3 CH1/3
FINE FINE
5. To turn on the output, press the output key. The OUTPUT OUTPUT
30
OPERATION
CH2 CH1
Voltage level Double the reading on the CH1 Voltage meter. In the
above case, the actual output is 20.0 x 2 = 40.0V.
Current level CH1 meter reading shows the output Current. In the
above case, 2.000A. (CH2 Current control must be
in the Maximum position=3.0A).
Connection
V V
A A
COM
LOAD
31
IPS X303 Series Instruction Manual
2. Connect the load to the front panel terminals, CH1+ & CH2−. Use the
CH1 (−) terminal as the common line connection.
3. Press the CH1 key (LED turns on) and use the CH2/4 CH2/4
Voltage knob to set the master & slave output BEEP BEEP
FINE
(Fine control)
CURRENT
Push
COARSE/FINE
FINE
FINE
output key.
6. For the master (CH1) output level and CV/CC status, refer to the CH1
meter and indicator.
32
OPERATION
CH2 CH1
Master (CH1) CH1 meter reading shows the output voltage. In the
voltage level above case, 20.0V.
Master (CH1) CH1 meter reading shows the output current. In the
current level above case, 2.000A.
7. Press the CH2 key (LED turns on) and use the CH2/4 CH2/4
CURRENT
Push
COARSE/FINE
FINE
8. For the slave (CH2) output level and CV/CC status, refer to the
CH1/CH2 meter and CH2 indicator.
C.V. C.C. PAR.
CH2
Slave (CH2) The CH1 meter reading shows the output voltage.
voltage level In the above case, 20.0V.
Slave (CH2) The CH2 meter reading shows the output current.
current level In the above case, 3.000A.
33
IPS X303 Series Instruction Manual
Background / Tracking parallel operation doubles the current capacity of the IPS X303
Connection series by internally connecting CH1 and CH2 in parallel and combining
the output to a single channel. CH1 controls the combined output.
V V
A A
LOAD
3. To turn on the output, press the output key. The OUTPUT OUTPUT
34
OPERATION
CH2
5. Press the CH1 key (LED turns on) and then CH1/3 CH1/3
FINE FINE
control is disabled.
By default, the Voltage and Current knob work
in the coarse mode. To activate the fine mode,
(Fine control)
press the knob to turn the FINE LED on. CURRENT
Push
COARSE/FINE
FINE
6. For the output level and CV/CC status, refer to the CH1 meter and
indicator.
C.V. C.C. PAR. C.V. C.C.
CH2 CH1
Voltage level The CH1 meter reading shows the output voltage. In
the above case, 20.0V.
Current level Double the amount of CH1 current meter reading. In
the above case, 2.0A x 2 = 4.0A.
35
IPS X303 Series Instruction Manual
SAVE/RECALL SETUP
Save Setup
Background The front panel settings can be stored into one of the four internal
memories.
Contents The following list shows the setup contents.
Independent / tracking series / tracking parallel mode
CH1/CH2 knob selection
Fine/coarse editing mode
Output voltage/current level
The following settings are always saved as “off”.
Output on/off
Front panel lock/unlock
Buzzer on/off
Panel operation Press one of the 1 to 4 Memory keys for 2 MEMORY MEMORY
RECALL / SAVE RECALL / SAVE
4 4
36
SAVE/RECALL SETUP
Recall Setup
Background The front panel settings can be recalled from one of the four internal
memories.
Contents The following list shows the setup contents.
Independent / tracking series / tracking parallel mode
CH1/CH2 knob selection
Fine/coarse editing mode
Output voltage/current level
The following settings are always recalled as “off”.
Output on/off
Front panel lock/unlock
Buzzer on/off
Panel operation Press one of the 1 to 4 Memory keys, for MEMORY MEMORY
RECALL / SAVE RECALL / SAVE
turns off.
3 3
4 4
37
IPS X303 Series Instruction Manual
REMOTE CONTROL
Background The IPS 3303D and IPS X303S are capable of being remotely controlled
via the USB connection.
COM setting Set up the COM port inside the PC according to the following list.
Baud rate: 57600 or 115200
Parity bit: None
Data bit: 8
Stop bit: 1
Data flow control: None
Functionality Run this query command via the terminal application such as MTTTY
check (Multi-threaded TTY).
*IDN?
This should return the identification information: Manufacturer, model
name, serial number, firmware version.
Iso-tech, IPS X303x, SN: xxxxxxxx, Vx.xx
38
REMOTE CONTROL
3. The power supply also automatically enters the lock state (the Lock
key will become activated).
LOCK LOCK
UNLOCK UNLOCK
Leaving the 1. To exit remote mode either, 1) use the LOCAL command from the
remote control terminal connection, or 2) Press the LOCK key on the front panel to
mode return to local mode, or 3) disconnect the USB cable from the rear
panel.
2. The display shows “USB…NO” message.
LOCK LOCK
UNLOCK UNLOCK
39
IPS X303 Series Instruction Manual
Command Syntax
40
REMOTE CONTROL
Error Messages
The following error messages might appear when the IPS 3303D or X303S cannot accept
the command.
41
IPS X303 Series Instruction Manual
Command List
The “HELP” command shows all the below commands and their meanings, except for the
HELP command itself.
42
REMOTE CONTROL
Command Details
ISET<X>:<NR2>
ISET<X>?
VSET<X>:<NR2>
VSET<X>?
43
IPS X303 Series Instruction Manual
IOUT<X>?
VOUT<X>?
TRACK<NR1>
BEEP<Boolean>
OUT<Boolean>
44
REMOTE CONTROL
STATUS?
*IDN?
RCL<NR1>
SAV<NR1>
45
IPS X303 Series Instruction Manual
BAUD<NR1>
LOCAL
Description Exits remote mode and sets the instrument to local mode.
Response time Minimum 10ms
ERR?
Description Checks the error status of the instrument and returns the last error
message.
Response time Minimum 10ms
Contents See page 41 for the list of error messages.
46
REMOTE CONTROL
HELP?
47
IPS X303 Series Instruction Manual
APPENDIX
Fuse Replacement
Steps 1. Take off the power cord and remove the fuse socket using a minus
driver.
Rating 100V/120V:T6.3A/250V
220V/230V:T3.15A/250V
48
SPECIFICATIONS
SPECIFICATIONS
The specifications apply when the IPS X303 series are powered on for at least 30 minutes
under +20 to +30°C.
49
IPS X303 Series Instruction Manual
50
SPECIFICATIONS
Options
51
IPS X303 Series Instruction Manual
FAQ
Q1. I pressed the panel lock key but the output still turns on/off.
A1. The output key is not affected by the panel lock key operation, for ensuring safety.
A2. No, it simply means that the CH3 output current reached the maximum 3.0A and the
operation mode turned from CV (constant voltage) to CC (constant current). You can
continue using the power supply, although reducing the output load is recommended.
A3. Make sure that the power supply is powered on for at least 30 minutes, within +20 to
+30°C.
Q4. The internal memory is not recording the panel setting correctly – the output should be
on.
52
INDEX
INDEX
54
INDEX
Limited Warranty
This meter is warranted to the original purchaser against defects in material and
workmanship for 3 years from the date of purchase. During this warranty period, RS
Components will, at its option, replace or repair the defective unit, subject to
verification of the defect or malfunction. This warranty does not cover fuses,
disposable batteries, or damage from abuse, neglect, accident, unauthorized repair,
alteration, contamination, or abnormal conditions of operation or handling. Any
implied warranties arising out of the sale of this product, including but not limited to
implied warranties of merchantability and fitness for a particular purpose, are limited
to the above. RS Components shall not be liable for loss of use of the instrument or
other incidental or consequential damages, expenses, or economic loss, or for any
claim or claims for such damage, expense or economic loss. Some states or
countries laws vary, so the above limitations or exclusions may not apply to you.
For full terms and conditions, refer to the RS website.
55
IPS X303 Series Instruction Manual
Africa
Iso-Tech
1 & 2 Indianapolis Street
Kyalami Business Park
Kyalami, Midrand, South Africa
Asia
Iso-Tech
460 Alexandra Road, #15-01A
PSA Building
Singapore 119963
Europe
Iso-Tech
PO Box 99
Corby
Northamptonshire
NN17 9RS
United Kingdom
Japan
West Tower (12th Floor)
Yokohama Business Park
134 Godocho, Hodogaya
Yokohama, Kanagawa 240-0005 Japan
USA
7410 Pebble Drive
Fort Worth
Texas 76118-6961
Canada
1701 Woodward Drive
Ste 108 Ottawa
Ontario K2C 0R4, Canada
South America
Av. Pdte. Eduardo Frei M. 6001-71
Centro Empresas El Cortijo
Conchali, Santiago, Chile
56