Data sheet

testo 570

Digital manifold

testo 570 – The manifold for


measuring, analysing and
documenting

Calculate superheating and subcooling simultaneously


bar micron
Record pressure, temperature, current and vacuum
measurement in one instrument
psi kPa
999 hours recording measurement values

40 refrigerants stored
°C/°F MPa

Report printout on site with Testo fast printer

hPa inch Hg

The digital manifold testo 570 is a robust tool for all The internal data memory replaces manual work steps.
imaginable measurements on a refrigeration system or This means that, for example, long-term measurements
heat pump. The instrument is capable of considerably can be carried out, or data can be recorded in real time
more than just measurement. Pressures, temperatures, and presented graphically. Errors can be found and
vacuum or current – testo 570 analyses them all. It stores rectified more quickly in cases of malfunction. Manual
all measurement data, right from commissioning, to service documentation is also a thing of the past. Measurement
and maintenance, through to the error analysis of a system. protocols can be produced on site and results traced at any
time. This ensures transparency and inspires trust in the
customer.

www.testo.com
testo 570

Technical data

General technical data

testo 570-1 Set Memory up to 999 h recording

Display LCD graphics display


testo 570-1 set digital manifold for all
measurement tasks on refrigeration systems Operating temperature -20 to +50 °C
and heat pumps, 4-way-valve block,
internal data memory, integrated vacuum Storage temperature -20 to +60 °C
measurement, 3 temperature probe Battery life approx. 40 h (w/o illumination)
connections, clamp probe,
demo version PC software, Dimensions 280 x 135 x 75 mm
calibration protocol and batteries included
Weight 1200 g
Part no. 0563 5701 Protection class IP42

Print media CFC, HFC, N, H2O, CO2

Refrigerant in the R12, R22, R123, R134a, R227, R290, R401A,


instrument R401B, R402A, R402B, R404A, R406A,
R407A, R407C, R408A, R409A, R410A,
testo 570-2 Set R411A, R413A,R414B, R416A, R417A,
R420A, R421A, R421B, R422A, R422B,
testo 570-2 set digital manifold for all R422D, R424A, R427A, R434A, R437A,
measurement tasks on refrigeration systems R438A, R502, R503, R507, R600, R600a,
and heat pumps, 4-way-valve block, internal R718 (H2O), R744 (only in the permitted
data memory, integrated vacuum measurement, measurement range up to 50 bar), R1234yf
3 temperature probe connections, calibration Refrigerant update possible by customer (via
protocol, 2x clamp probes, software (free „EasyKool“ software)
download) with USB cable, transport
case, mains unit and batteries included

Part no. 0563 5702

Sensor types

Pressure Temperature Vacuum

Measuring range 50 bar -50 to +150 °C –

Measuring range (rel.) – – -1 bar ... 0 bar


Accuracy (at 22 °C) 0.5% fs (±1 Digit) ±0.5°C (±1 Digit) 1 % fs

Resolution 0.01 bar / 0,1 psi 0.1 °C/0.1 °F 1 hPa / 1 mbar / 500 micron

Probe connections 3 x 7/16“ – UNF + 1 x 5/8‘‘ – UNF 3 x plug-in (NTC) –

Overload rel. (LP/HP) 52 bar / 52 bar – –


Low pressure rel. (LP) 50 bar / 50 bar – –
High pressure rel. (hP)
testo 570

Accessories

Accessories for measuring instrument Part no.

Transport case for testo 550 and accessories (also suitable for testo 570) 0516 0012

Plug-in mains adapter, 5 VDC 500 mA with European adapter, 100-250 VAC, 50-60 Hz 0554 0447

USB connection cable, instrument-PC 0449 0047

Software EasyKool for measuring data management (USB cable included) 0554 5604

Testo fast printer IrDA with wireless infrared interface; 1 roll thermal paper; 4 AA batteries 0554 0549

Spare valve set; exchange of 2 valve mechanisms with 4 valve knob covers (red, blue, 2 x black), 0554 5570
compatible with testo 549, testo 550, testo 557 and testo 570.

Probes

Probe type Dimensions Measuring Accuracy Part no.


Probe shaft/probe shaft tip range

Air probes

Efficient, robust NTC air probe -50 to +125 °C ±0.2 °C (-25 to +80 °C) 0613 1712
115 mm 50 mm ±0.4 °C (remaining range)

Ø 5 mm Ø 4 mm

Surface probes
Clamp probe for measurement on -40 to +125 °C ±1 °C (-20 to +85 °C) 0613 5505
pipes for diameter 6 to 35 mm,
NTC

Pipe wrap probe with Velcro for -50 to +70 °C ±0.2 °C (-25 to +70 °C) 0613 4611
±0.4 °C (-50 to -25.1 °C)
pipe diameter to max. 75 mm,
30 mm
Tmax. +75 °C, NTC,
Fixed cable 1.5 m 300 mm

Pipe clamp probe (NTC) for pipe -50 ... +120 °C ±0.2 °C (-25 ... +80 °C) 0613 5605
diameters 5 to 65 mm, Fixed cable
2.8 m

Waterproof NTC surface probe for -50 to +150 °C ±0.5% of m.v. (+100 to +150 0613 1912
115 mm 50 mm Long-term °C)
flat surfaces, meas. range ±0.2 °C (-25 to +74.9 °C)
Fixed cable 1.2 m +125 °C, short- ±0.4 °C (remaining range)
Ø 5 mm Ø 6 mm term +150 °C (2
minutes)

More probes

Current probe for measuring 0 to 20/200 A 0 to 9.9 A 4% 0554 5607


10 to 49.9 A 3%
current consumption of 50 to 200 A 2%
compressors, with switchable
measuring range
Fixed cable 2.9 m
Oil pressure probe to monitor 0 to 25 bar rel 1,5 % of fsv 0638 1742
Overload: 50 bar
the oil level with 2.9 m cable,
refrigerant and oil resistant
Fixed cable 2.9 m
testo 570

www.testo.com
Subject to change without notice. 0981 9194/msp/I/01.2020