Table of Contents
Title Page
Overview 3
Unpacking Inspection 4
Safety Information 5
Rules For Safe Operation 6
International Electrical Symbols 8
The Meter Structure 9
Rotary Switch 10
Functional Buttons 11
The Effectiveness of Functional Buttons 12
Display Symbols 13
Measurement Operation 15
A. DC Voltage Measurement 15
B. AC Voltage Measurement 16
C. Measuring Resistance 17
D. Testing Diodes 19
E. Testing for Continuity 20
F. Temperature Measurement 22
G. AC Current Measurement 23
Sleep Mode 25
Specifications 26
A. General Specifications 26
B. Environmental Restriction 26
Accuracy Specifications 27
A. AC Voltage 27
B. DC Voltage 27
C. Resistance 28
D. Continuity Test 28
E. Diode Test 28
F. Temperature 29
G. AC Current 29
Maintenance 30
A. General Service 30
B. Replacing the Battery 30
Overview
Warning
To avoid electric shock or personal injury, read the
“Safety Information” and “Rules for Safe Operation”
carefully before using the Meter.
Unpacking Inspection
Open the package case and take out the Meter. Check
the following items carefully to see any missing or damaged
part:
Safety Information
Warning
To avoid possible electric shock or personal injury,
and to avoid possible damage to the Meter or to the
equipment under test, adhere to the following rules:
AC (Alternating Current).
DC (Direct Current).
AC or DC.
Grounding.
Double Insulated.
Deficiency of Built-In Battery
Continuity Test.
Diode.
Capacitance Test
Fuse.
Warning. Refer to the Operating Manual.
Conforms to Standards of European Union.
( figure 1)
1. Input Terminals
2. LCD Display
3. Functional Buttons
4. Rotary Switch
5. Trigger: press the lever to open the transformer jaws.
When the pressure on the lever is released, the jaws
will close.
6. Hand Guards: to protect user’s hand from touching
the dangerous area.
7. Transformer Jaws: designed to pick up the AC current
flowing through the conductor. It could transfer current
to voltage.
Rotary Switch
Below table indicated for information about the rotary switch
positions.
Rotary Function
Switch
Position
OFF Power is turned off.
AC/DC voltage measurement.
: Diode test.
: Continuity test.
Ω Ω : Resistance measurement.
o o Temperature measurement
CF (UT202 only)
10
Functional Buttons(1)
Below table indicated for information about the functional
button operations.
AC voltage measurement.
l Press HOLD to enter and exit the
HOLD
Hold mode in any mode, the Meter beeps.
l Press and hold HOLD button
while turning on the Meter to display
full icons.
MAX Press MAX to start recording and updating
of maximum values.
SELECT Press SELECT button to switch between
Ω o o
and C F .
11
Model: UT201
Rotary Switch
Positions Functional Buttons
SELECT MAX HOLD
V N/A
V N/A
Ω N/A N/A
N/A
A 2/20A N/A
A 200/400A N/A
Model: UT202
Rotary Switch
Positions Functional Buttons
SELECT MAX HOLD
V N/A
V N/A
Ω N/A
o o
F C(K-Type) N/A
A 2/20A N/A
A 200/400A N/A
12
13
14
Measurement Operation(1)
A. DC Voltage Measurement (see figure 3)
red black
( figure 3)
Warning
To avoid harms to you or damages to the Meter from
eletric shock, do not attempt to measure voltages
higher than 600V AC/DC, although readings may be
obtained.
The DC Voltage ranges are:
200.0mV, 2.000V, 20.00V, 200.0V and 600V.
15
Measurement Operation(2)
l When DC voltage measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test and remove testing leads
from the input terminals.
red black
( figure 4)
Warning
To avoid harms to you or damages to the Meter from
eletric shock, do not attempt to measure voltages
higher than 600V AC/DC, although readings may be
obtained.
The AC Voltage ranges are:
2.000V, 20.00V, 200.0V and 600V.
16
Measurement Operation(3)
Note
l In each range, the Meter has an input impedance of
10MΩ. This loading effect can cause measurement
errors in high impedance circuits. If the circuit impedance
is less than or equal to 10kΩ, the error is negligible
(0.1 or less).
l When AC voltage measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test and remove testing leads
from the input terminals.
red black
( figure 5)
Warning
To avoid harms to you, do not attempt to input voltages
higher than 60V DC or 30V rms AC.
17
Measurement Operation(4)
The resistance ranges are:
200.0 , 2.000k , 20.00k ,200k ,2.000M and 20.00M .
18
Measurement Operation(5)
D.Testing Diodes (see figure 6)
red black
( figure 6)
Warning
To avoid damages to the Meter or to the devices under
test, disconnect circuit power and discharge all the
high-voltage capacitors before testing diodes.
19
Measurement Operation(6)
Note
l In a circuit, a good diode should still produce a forward
voltage drop reading of 0.5V to 0.8; however, the
reverse voltage drop reading can vary depending
on the resistance of other pathways between the
probe tips.
l Connect the test leads to the proper terminals as
said above to avoid error display.
l The LCD will display OL indicating either open circuit
or wrong polarity connection.
l The unit of diode is volt (V), displaying the forward
voltage drop readings.
l To remove the objects being tested from the circuit
when measuring can obtain a more accurate result.
l When diode testing has been completed, disconnect
the connection between the testing leads and the
circuit under test and remove testing leads from the
input terminals.
red black
( figure 7)
20
Measurement Operation(7)
Warning
To avoid damages to the Meter or to the devices under
test, disconnect circuit power and discharge all the
high-voltage capacitors before measuring continuity.
To test for continuity, connect the Meter as follows:
1. Insert the red test lead into the VΩ terminal and
the black test lead into the COM terminal.
2. Set the rotary switch to Ω and press SELECT
button to select measurement mode.
3. The buzzer sounds if the resistance of a circuit under
test is less than 50 Ω .
4. The buzzer may or may not sounds if the resistance
of a circuit under test is between 50 Ω to 120 Ω .
5. The buzzer does not sound if the resistance of a circuit
under test is higher than 120Ω.
Note
l The buzzer beeps once when pressing any buttons
at any rotary switch positions except at 2/20A positions
if the button is valid. If the button is not valid, it does
not beep. At 2/20A rotary switch position, the buzzer
is set not to beep.
l The buzzer beeps 5 times continuously on around
1 minute before entering the sleep mode. When it
is just before entering the sleep mode, it will have
one long beep to warn you.
l The LCD displays OL indicating the circuit being
tested is open.
l When continuity testing has been completed, disconnect
the connection between the testing leads and the
circuit under test and remove testing leads from the
input terminals.
21
Measurement Operation(8)
F. Temperature Measurement (UT202 Only)
(see figure 8)
red black
( figure 8)
o o
The temperature measurement ranges are -40 C~1000 C
o o
and -40 F~1832 F.
22
Measurement Operation(9)
G. AC Current Measurement (see figure 9)
( figure 9)
Warning
To avoid electric shock, never measure current while
the test leads are inserted into the input terminals
and disconnect test leads and tested circuit connection.
23
Measurement Operation(10)
Note:
l To obtain accurate reading, measure only one
conductor at each time.
l When current measurement has been completed,
disconnect the connection between the conductor
under test and the jaw, and remove the conductor
away from the transformer jaw of the Meter.
24
Sleep Mode
To preserve battery life, the Meter automatically turns
off if you do not turn the rotary switch or press any button
for around 15 minutes.
25
Specifications
A.General Specifications:
l Maximum voltage including transient overvoltage
between any terminals and grounding: 500V rms.
l Display: 3 1/2 digits LCD display, Maximum display 1999
l Auto Polarity Display
l Overloading : Display OL or –OL
l Battery Deficiency : Display
l Measurement Speed : Updates 3 times/second.
l Measuremnet Deviation : When the conductor being
meaured is not placed in a
correct position during AC
current measurement, it will
cause 1% reading deviation.
l Drop Test : 1 meter drop test passed.
l Max. Jaw Size : 28mm diameter.
l Projected Max.
Current conductor size : 26mm diameter.
l Power : 2pcs of 1.5V battery (AAA)
l Battery Life : typically 150hours (alkaline
battery)
l Sleep Mode (can be disabled)
l Dimensions (H x W x L) : 30mm x 76mm x 208mm.
Weight : Approximate 260g (battery
included)
B.Environmental Restrictions:
l The Meter is suitable for indoor use.
l Altitude : Operating: 2000m
Storage: 10000m
l Safety/ Compliances : IEC 61010 CAT.II 600V,
CAT.III 300V over voltage
and double insulation standard.
l Temperature and humidity:
o o
Operating: 0 C~30 C ( 75%
o o
R.H);30 C~40 C ( 70%R.H);
o o
40 C~50 C ( 45%R.H);
o o
Storage: -20 C~+60 C
( 75%R.H)
26
Accurate Specifications(1)
Accuracy: (a% reading + b digits), guarantee for 1 year.
o o
Operating temperature: 23 C 5 C
Relative humidity: 75%R.H
o
Temperature coefficient: 0.1x(specified accuracy)/1 C
Remarks:
l Input impedance: 10MΩ //<100pF”
l Displays effective value of sine wave (mean value
response).
l Frequency response: 40Hz~1kHz.
l To adjust reading in accordance with effective value
27
Accurate Specifications(2)
C. Resistance: Auto ranging
D. Continuity Test
Range Resolution Accuracy Overload
Protection
Around 600Vp50Ω the
100mΩ buzzer beeps
Remark:
l Open circuit voltage approximate 0.45V.
l The buzzer may or may not beeps when the resistance
of a circuit under test is between 50Ω and 120Ω
The buzzer may not beep when the resistance of a
circuit under test is greater than 120Ω.
E. Diode Test
Range Resolution Accuracy Overload
Protection
Display approximate
1mV forward voltage drop 600Vp
: 0.5V~0.8V
28
Accurate Specifications(3)
F. Temperature (UT202 Only):
-40~1000 C
o
1C
o
0~400 C:
(1%+3)
o
400~1000 C:
(2%+10) 600Vp
o
-40~32 F:
o
(3%+8)
o o o o
-40 F~1832 F 1F 32 F~752 F:
(1%+6)
o o
752 F~1832 F:
(2%+18)
Remarks:
l Displays effective value of sine wave (mean value
response).
To adjust reading in accordance with effective value
29
MAINTENANCE(1)
This section provides basic maintenance information
including battery replacement instruction.
Warning
Do not attempt to repair or service your Meter unless
you are qualified to do so and have the relevant
calibration, performance test, and service information.
A. General Service
Screw
( figure 10)
30
MAINTENANCE(2)
Warning
To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery
as soon as the battery indicator “ ” appears.
Make sure the transformer jaw and the tets leads are
disconected from the circuit being tested before
opening the case bottom.
** END **
This operating manual is subject to change without notice.
31
Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Dong Fang Da Dao
Bei Shan Dong Fang Industrial Development District
Hu Men Town, Dongguan City
Guang Dong Province
China
Postal Code: 523 925
Headquarters:
Uni-Trend Group Limited
Rm901, 9/F, Nanyang Plaza
57 Hung To Road
Kwun Tong
Kowloon, Hong Kong
Tel: (852) 2950 9168
Fax: (852) 2950 9303
Email: info@uni-trend.com
http://www.uni-trend.com
32