Teraohm 5 KV User Manual: Version 1.4, Code No. 20 750 722
Teraohm 5 KV User Manual: Version 1.4, Code No. 20 750 722
Teraohm 5 KV User Manual: Version 1.4, Code No. 20 750 722
MI 2077
User Manual
Version 1.4, Code No. 20 750 722
Distributor:
Producer:
METREL d.d.
Horjul 188
1354 Horjul
Slovenia
© 2002 Metrel
No part of this publication may be reproduced or utilized in any form or by any means
without permission in writing from METREL.
Table of contents
1. General introduction--------------------------------------------------------------------------- 4
1.1. Features ----------------------------------------------------------------------------------------4
1.2. Applied Standards----------------------------------------------------------------------------4
3. Warnings------------------------------------------------------------------------------------------- 8
5. Measurements ----------------------------------------------------------------------------------- 12
5.1. Generally about DC High voltage testing ----------------------------------------------12
5.2. Guard terminal --------------------------------------------------------------------------------15
5.3. Filter option-------------------------------------------------------------------------------------16
5.4. Voltage measurement-----------------------------------------------------------------------18
5.5. Insulation Resistance measurement ----------------------------------------------------18
5.6. Diagnostic test --------------------------------------------------------------------------------20
5.7. Step Voltage Insulation Resistance testing--------------------------------------------23
5.8. Withstanding voltage ------------------------------------------------------------------------25
7. Maintenance -------------------------------------------------------------------------------------- 31
7.1. Inspection---------------------------------------------------------------------------------------31
7.2. Battery Replacement ------------------------------------------------------------------------31
7.3. Cleaning ----------------------------------------------------------------------------------------32
7.4. Calibration --------------------------------------------------------------------------------------32
7.5. Service ------------------------------------------------------------------------------------------32
8. Specifications------------------------------------------------------------------------------------ 33
8.1. Measurements --------------------------------------------------------------------------------33
8.2. General specifications-----------------------------------------------------------------------35
1. General introduction
1.1. Features
The TeraOhm 5kV Tester is a portable battery / mains powered test instrument intended
for testing of Insulation Resistance by using high test voltages of up to 5 kV. It operates
on a SIMPLE and CLEAR basis.
The instrument is designed and produced with the extensive knowledge and experience
acquired through many years of dealing with similar test equipment.
A dot matrix LCD offers easy-to-read results and all associated parameters. Operation is
simple and clear; the operator does not need any special training (except reading and
understanding this Users Manual) to operate the instrument.
2. Instrument Description
2.1. Instrument Casing
The instrument is housed in a plastic casing that maintains the protection class defined
in the general specifications. The casing is equipped with a carrying strap, which is
intended for the instrument to be used hung around the operator’s neck. Short technical
specification is available on base of the housing.
Legend:
1 ..........START key to start any measurement.
2 ..........ON/OFF key to switch the instrument ON or OFF.
3 ..........MEM key to store, recall and erase results.
4 ..........SELECT key to enter set-up mode for the selected function or to select the
active parameter to be set.
5 ..........
5 cursor key to select an option upward.
6 ..........
6 cursor key to select an option downward.
2.3. Connectors
The TeraOhm 5 kV Tester contains the following connections:
- Connection of test leads to three banana safety sockets (1, 2, 3),
- Connection of the communication cable to the 9-pin RS 232 connector (4) and
- Mains supply cable connection to the mains socket (5).
Fig. 2. Connectors
Legend:
1 ..........Positive Insulation Resistance test terminal.
2 ..........Negative Insulation Resistance test terminal.
3 ..........GUARD test terminal intended to lead away potential leakage current while
measuring the Insulation Resistance.
4 ..........Galvanic separated RS 232 connector to connect the instrument to PC.
5 ..........Mains connector to connect the instrument to the mains supply.
5
DISCONNECT ALL TEST LEADS AND SWITCH OFF INSTRUMENT,
BEFORE REMOVING THIS BATTERY/FUSE COVER - HAZARDOUS VOLTAGE!
6 DO NOT CHARGE IF BATTERY COMPARTMENT CONTAINS ALKALINE CELLS!
Legend:
1 ..........Nylon strap (allows the operator to carry the instrument hung around the neck).
2 ..........Plastic cover (fixes the nylon strap to the instrument).
3 ..........Screw (unscrew to remove the carrying strap or to open the instrument).
4 ..........Label with measurement ranges.
5 ..........Battery/fuse compartment cover.
6 ..........Screw (unscrew to replace the batteries).
7 ..........Rubber foot.
2.5. Accessories
The accessories consist of standard and optional accessories. Optional accessories can
be delivered upon request. See attached list for standard configuration and options or
contact your distributor or see the METREL home page: http://www.metrel.si .
3. Warnings
In order to reach the highest level of operator’s safety while carrying out various
measurements and tests using the TeraOhm 5kV Tester, as well as to ensure that the
test equipment remains undamaged, it is necessary to consider the following warnings:
MEANING OF SYMBOLS
♦ ! symbol on the instrument means “Read the User Manual with special care!”
♦ symbol on the instrument means “Hazardous voltage higher than 1000 V
may be present at the test terminals!”!
GENERAL PRECAUTIONS
♦ If the test equipment is used in a manner not specified in this Users Manual,
the protection provided by the equipment may be impaired!
♦ Do not use the instrument and accessories, if any damage is noticed!
♦ Consider all generally known precautions in order to avoid the risk of electric
shock while dealing with electric installations!
♦ Service intervention or recalibration procedure can be carried out only by a
competent and authorized person!
♦ Only adequately trained and competent persons may operate the instrument.
BATTERIES
♦ Disconnect all test leads and switch the power off before opening the Battery
cover!
♦ Avoid connection to main supply when alkaline batteries are inserted, danger
of explosion! Do not charge when alkaline batteries are fitted!
EXTERNAL VOLTAGES
♦ Do not connect the instrument to a mains voltage different from the one
defined on the label adjacent to the mains connector, otherwise the instrument
may be damaged.
• Do not connect test terminals to an external voltage higher than 600 V AC or
DC (CAT III environment) to prevent any damage on the test instrument!
• In case of a capacitive test object (long tested cable etc.), automatic discharge
of the object may not be done immediately after finishing the measurement –
“Please wait, discharging” message will be displayed.
HANDLING WITH CAPACITIVE LOADS
♦ Note that 40 nF charged to 1 kV or 1.6 nF charged to 5 kV are hazardous live!
♦ Never touch the measured object during the testing until it is totally
discharged.
♦ Maximum external voltage between any two leads is 600 V (CAT III
environment).
4. Performing measurements
4. 1. Switching on the instrument
Autocalibration
The instrument is switched ON by pressing the ON/OFF key. After turning on, the
instrument first executes the autocalibration (Fig. 5).
It is necessary that the measuring test leads are disconnected during autocalibration. If
not the instrument will require disconnection of the test leads and repeat switching OFF
and ON.
After finishing the autocalibration, the instrument is prepared for normal operation.
Autocalibration prevents lowering of the accuracy when measuring very low currents. It
compensates the effects caused by ageing, temperature and humidity changes etc.
A new autocalibration is recommended when the temperature changes by more than
5°C.
Note:
If the instrument detects incorrect state during the autocalibration, the following warning
message will be displayed:
ERROR!
-TEST LEADS CONNECTED:
DISCONNECT AND SWITCH ON THE INSTRUMENT AGAIN
Possible reasons for out of range conditions are excessive humidity, too high
temperature, etc. In this case it is possible to perform measurements but results could
be out of technical specification.
Backlight operation
After turning the power on the LCD backlight is automatically turned on. It can be turned
OFF and ON by simply clicking the LIGHT key.
Backlight turns off automatically approximately 10s after switching it ON, if the
instrument is supplied by the internal battery.
10
Off function
The instrument can be switched OFF only by pressing the ON/OFF key. The auto-off
function is not available due to the possible long-term measurements.
4.2. Configuration
The configuration function enables the selection and adjustment of the parameters that
are not directly involved in the measurement procedure (Figure 6).
In the lower section of the display the supply status, date and time is shown (same in all
functions).
The following procedure must be carried out when adjusting some of the configuration
parameters:
Selected function
Function dependent
area
11
5. Measurements
5.1. Generally about DC High voltage testing
(FURTHER INFORMATION CAN BE FOUND IN METREL’S HANDBOOK
“INSULATION MEASURING TECHNIQUES”)
The purpose of insulation tests
Insulating materials are important parts of almost every electrical product. The material’s
properties depend not only on its compound characteristics but also on temperature,
pollution, moisture, ageing, electrical and mechanical stress, etc. Safety and operational
reliability require the regular maintenance and testing of the insulation material to ensure
it is kept in good operational condition. For testing the insulation materials high voltage
test methods are applied.
Itest
surface
+
material
Riss1
Cpi
Riso Ciso Itest
IPI
Guard Rpi
Riss2
ICiso
IRiso IRiss
-
12
Riss1 and Riss2 - the surface resistivity (position of optional guard connection)
Riso – the actual insulation resistance of material
Ciso – capacitance of material
Cpi, Rpi - represents polarization effects.
The right figure shows typical currents for that circuit.
Itest = overall test current (Itest= IPI+ IRISO+ IRISS)
IPI = polarization absorption current
IRISO = actual insulation current
IRISS = surface leakage current
Some application examples for using Teraohm 5 kV
Basic Insulation resistance test
Virtually every standard concerning the safety of electrical equipment and installations
requires the performance of a basic insulation testing. When testing lower values (in the
range of MΩ), Riso usually dominates. The results are adequate and stabilize quickly.
- The voltage, time and limit are usually given in the appropriate standard or
regulation.
- The suggested test voltage is often defined as 1.7 times the value of line voltage of
the tested device (e.g. motor), i.e. the test voltage for a 600 V device would be 1020
VDC. Measuring time should be set to 60 s or the minimum time required for the Ciso
to be charged up.
- Sometimes it is required to take into account the ambient temperature to adjust the
result for a standard temperature of 40°C.
- If surface leakage currents interfere with the measurements (see Riss above) use the
guard connection (see 5.2.). This becomes critical when measuring values in the GΩ
range.
Some examples:
Standard* Test voltage / time End value
EN/IEC 60204 500 V 1 MΩ
EN/IEC 60335-1 2000 V / 15 min
EN/IEC 60349-1 500 V 1000 Ω/V
EN/IEC 60598-1 500 V / 1 s 2 MΩ
* Only sample values are listed!
13
Example for minimum acceptable values for motor insulation (IEEE 43):
Class A =1.5, Class B = 2.0, Class F =2.0, Class H =2.0.
DIELECTRIC DISCHARGE
An additional effect of polarization is the recovered charge (from Cpi) after the regular
discharging of a completed test. This can also be a supplementary measurement for
evaluation of the quality of insulating material. This effect is generally found in insulating
systems with large capacitance Ciso.
14
15
IL
IM IM
IL
IA IA IL
Ut A Ut A
DIAGNOSTIC STEP
4..8 9 TEST VOLTAGE
+ (Red)
ESC
INSULATION WITHSTANDING
RESISTANCE
10 VOLTAGE
SELECT LIGHT VOLTAGE CONFIG
1
Guard
3 2
START MEM ON/OF
F
TeraOhm 5kV
- (Black)
Fig. 7. Connection of GUARD terminal to measured object (an example for measuring
cable insulation is shown)
where:
Ut ........Test voltage
IL .........Leakage current (resulted by surface dirt and moisture)
IM .........Material current (resulted by material conditions)
IA .........A-meter current
Result without using GUARD terminal: RINS = Ut / IA = Ut / (IM + IL) …incorrect result.
Note:
• Guard terminal is internal impedance (300 KΩ) protected.
16
Note:
Any of the selected filter options increases the settling time with Fil1 to 60 s, Fil2 to 70 s,
and Fil3 to 120 s.
- It is necessary to pay close attention to the selection of time intervals when using
filters.
- The recommended minimum measuring times when using filters are the settling
times of the selected filter option.
Example 1:
A noise current of 1 mA / 50 Hz adds approximately ±15 % distribution to the measured
result when measuring 1 GΩ.
By selecting FIL1 option the distribution will reduce to less than ±2 %.
In general using FIL2 and FIL3 will further improve the noise reduction.
Example 2:
Testing an object with the following parameters:
Utest =250 V, Inoise = 3 mA, R= 2.4 GΩ, C=220 nF
Filter options give following results:
OFF: distribution of results 30 %,
Fil1: distribution of results 3.5 %,
Fil2: distribution of results 1.5 %,
Fil3: distribution of results 0.9%.
For this example the setup time is also increased and for the Fil3 option, the time to
capture the first good result (Time1) is approximately 2 min.
17
Measurement procedure:
- Connect the test leads to the instrument and to the measured source.
- Press the START key to start the measurement, continuous measurement starts to
run.
- Press the START key again to stop the measurement.
- The result (see the right figure above) can optionally be saved by pressing the MEM
key twice, see the chapter 6.1. Store, Recall and Clear Operation.
•
Warning!
• Refer to Warnings chapter for safety precautions!
18
Measurement procedure:
- Connect the test leads to the instrument and to the tested object.
- Select INSULATION RESISTANCE function using the rotary switch knob.
- Press the START key and release it, continuous measurement starts to run.
- Wait until the test result has stabilized then press the START key again to stop the
measurement or until set timer runs out (if enabled).
- Wait for the object under test to be discharged.
- The result (see the right figure above) can optionally be saved pressing the MEM
key twice, see the chapter 6.1. Store, Recall and Clear Operation.
Notes:
- If the timer is disabled then OFF is displayed instead of the timer value.
- During the measurement the timer information displays the time needed for the
completion of the measurement (tr) while after the completion the test duration (tm) is
displayed.
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a possible dangerous test voltage.
- Value of capacitance is measured during the final discharge of the measured object.
19
Timer and Time1 are independent timers. Maximum time for each of them is
30 min 60 s.
Warning!
• Refer to Warnings chapter for safety precautions!
This is a long duration test for evaluating the quality of the insulation material under test.
The results of this test enable a decision to be made on the preventive replacement of
the insulation material.
POLARIZATION INDEX (PI)
PI is the ratio of Insulation Resistance values measured after 1 minute and after 10
minutes. The DC test voltage is present during the whole period of the measurement
(also Insulation Resistance measurement is running). On completion of the test the PI
ratio is displayed:
PI = R (10 min )
iso
R (1 min )
iso
Note: When determining Riso (1min) pay close attention to the capacitance of the object
under test. It has to be charged-up in the first time section (1 min). Approximate
maximum capacitance using:
20
t [s].103
[ ]
C max µF = . [V ] ,
3U
where:
t ...........period of first time unit (e.g. 1min)
U .........test voltage.
Idis1 min[mA]
DD =
U [V ].C [F ]
,
where:
Idis 1min...... discharging current measured 1 min after regular discharge
U ................. test voltage
C ................. capacitance of tested object.
Measurement procedure:
- Connect the test leads to the instrument and to the measured object.
- Press the START key to start the measurement.
- Wait until set timer runs out, result is displayed.
- Wait until the object under test has discharged
- The result (see the right part of the figure 11.) can optionally be saved by pressing
the MEM key twice, see the chapter 6.1. Store, Recall and Clear Operation.
21
Notes:
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a possible dangerous test voltage.
- The value of capacitance is measured during the final discharge of the measured
object.
- If enabled, the instrument measures DD when the measured capacitance is in the
range of 5 nF to 50 µF.
Time1, Time2 and Time3 are timers with the same start point. The value of each
presents the duration from the start of the measurement. Maximum time is 30 min. The
following figure shows the timer relationship.
Start Stop
R01min R02min R03min
PI1 PI2
(if enabled DD)
Time1 ≤ Time2
Time2 ≤ Time3
22
Insulation is measured in five equal time periods at test voltages from one fifth up to full
scale, see the figure 15. This function shows the level of the measured Insulation
Resistance versus the test voltage.
Measurement procedure:
- Connect the test leads to the instrument and to the measured object.
- Press the START key to start the measurement.
- Wait until set timer runs out, result is displayed.
- Wait until the object under test has discharged
- The result (see the right part of the figure 14.) can optionally be saved by pressing
the MEM key twice, see the chapter 6.1. Store, Recall and Clear Operation.
U
U
0.8U
0.6U
0.4U
0.2U
0 T 2T 3T 4T 5T t
Fig. 15. Step-up test voltage
23
Notes:
- Timer information is displayed from the start of the measurement until the completion
of each step measurement.
- Timer information shows the complete measurement period after the completion of
the measurement.
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a possible dangerous test voltage.
- The value of capacitance is measured during the final discharge of measured object.
24
Note:
- Maximum value for Step Time is 30 min.
Warning!
• Refer to Warnings chapter for safety precautions!
Both functions require breakdown current detection. The test voltage increases step by
step from the Start up to the Stop value over a predefined time and it is kept at the Stop
value for a predefined test time, see the figure below.
Ut Ut
Ustop
Ustep
Ustart Ustart
Ub
0 t 0 t
Tstep Tend
Ut ........Test voltage
Ustop...End test voltage
Ustep...Voltage step approx. 20 V (fixed value - not presetable)
Ustart ..Initial test voltage
Tstep ...Test voltage duration per one step
Tend....Constant test voltage duration after reaching End value
t ...........Time
Ub .......Breakdown voltage
Selecting this function displays the following states (initial state and state with results
after completion of the measurement).
25
Measuring procedure:
- Connect the test leads to the instrument and to the measured object.
- Press the START key to start the measurement.
- Wait until the set timers run out or until breakdown occurs, the result is displayed.
- Wait until the object under test is discharged.
- The result (see the right part of the figure 18.) can optionally be saved by pressing
the MEM key twice, see the chapter 6.1. Store, Recall and Clear Operation.
Note:
- Breakdown is detected when the measured current reaches or exceeds the set
current level Itrig.
Notes:
- Timer information shows the time needed to accomplish each step during the
measurement and it shows the total measurement period after the completion of the
measurement.
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a possible dangerous test voltage.
26
Tstep and Tend are independent timers. The maximum time for each timer is 30 min
60 s. Tend begins after the completion of the ramp period. Ramp period can be
calculated from:
If Tstep is set to 00min 00s, then the ramp voltage increases by approximately 20 V
per 2s.
Warning!
• Refer to Warnings chapter for safety precautions!
27
Recall: 0006
Figure 0006 presents the ser. No. of the stored result. By using the ↑ and ↓ keys
also the other stored results can be recalled. Press the ESC or Start key or turn the
rotary switch to exit this menu.
- To clear the last stored result: select CLR and press the MEM key.
In addition to the main result also the subresults and parameters of the selected function
are recorded. The following is a list of all data stored for each function.
28
29
Note:
• *Date and time of storing the test result are transferred to PC while date and time of
recalling are displayed when recalling results.
30
7. Maintenance
7.1. Inspection
To maintain the operator’s safety and to ensure the reliability of the instrument it is
advisable to inspect the instrument on a regular basis. Check that the instrument and its
accessories are not damaged. If any defect is found please consult your service center,
distributor or manufacturer.
Note:
• Operator does not need to disconnect the instrument from mains supply after the full
recharging period. The instrument can be connected permanently.
Battery cells are stored in the bottom section of the instrument casing under the battery
cover. In case of defective battery please note the following:
♦ All six cells have to be replaced and they have to be of the same type.
♦ Turn the power off and disconnect any measurement accessory connected to
the instrument before opening the battery cover to avoid electric shock.
♦ Do not operate the instrument by mains supply without rechargeable batteries
in order to avoid permanent damage of internal circuitry.
Nominal power supply voltage is 7.2 V DC. Use six NiCd or NiMH cells with size
equivalent to IEC LR14 (dimensions: diameter = 26 mm, height = 50 mm). See the next
figure for correct polarity of batteries.
Fully charged rechargeable battery can supply the instrument for approx. 20 hours.
Note:
• Standard IEC LR14 type alkaline cells can also be used to replace rechargeable
battery. Avoid connection to mains supply when alkaline batteries are inserted,
danger of explosion. High quality alkaline batteries can supply the instrument for
approx. 60 hours.
31
Notes!
• Insert cells correctly, otherwise test instrument will not operate and battery may be
discharged!
• If the instrument is not to be used for a long period of time remove all batteries from
battery compartment.
• Take into account handling and maintenance as defined by the manufacturer of
battery cells in use.
7.3. Cleaning
Use a soft cloth slightly moistened with soapy water or spirit to clean the surface of the
instrument and leave the instrument to dry totally before using it.
Notes!
• Do not use liquids based on petrol or hydrocarbons!
• Do not spill cleaning liquid over the instrument!
7.4. Calibration
It is essential that all measurement instruments be regularly calibrated. For occasional
daily use, we recommend an annual calibration to be carried out. When the instrument is
used continuously every day, we recommend calibrating the instrument every six
months.
7.5. Service
For repairing under or out of warranty period contact your distributor for further
information.
32
8. Specifications
8.1. Measurements
Insulation resistance
Nominal test voltage: Any within 250 and 5000 V
Current capability of test generator: >1 m A
Short-circuit test current: 1.4 m A max.
Automatic discharge of tested object: yes
33
Output 6
voltage *
[kV] 5
3
*
2
0
0,1 1 10 100
Resistance under test (MΩ )
Withstanding voltage
Voltage DC
Display range Withstanding voltage (V) Resolution Accuracy
0 ÷ 5500 1V ±(3 % of reading + 40 V)
Leakage current
Display range Itrigg (mA) Resolution Accuracy
0 ÷ 1.4 1 µA ±(3 % of reading + 3 digits)
Voltage
Voltage AC or DC
Display range External Voltage (V) Resolution Accuracy
0 ÷ 600 1V ±(3 % of reading + 3 V)
Input resistance: 3 MΩ ± 10 %
34
Capacitance
Measuring range C: 50 µF*
Display range C Resolution Accuracy
0 ÷ 99.9 nF 0.1 nF
100 ÷ 999 nF 1 nF ±(5 % of reading + 2 digits)
1 ÷ 50 µF 10 nF
*Full-scale value of capacitance is defined according to the following equation:
Polarization index PI
Display range PI Resolution Accuracy*
0 ÷ 99.9 0.01 ±(5 % of reading + 2 digits)
AUTOCALIBRATION
Auto-calibration of measuring system ............every time after turning power on
CONNECTING SYSTEM
Three safety banana sockets .........................+OUT, -OUT and GUARD.
Guard resistance............................................300 kΩ
35
DISCHARGING
Every time after measurement completion.
Discharging resistance:..................................100 kΩ ± 10 %
SERIAL COMMUNICATION
RS232 serial communication .........................galvanic separated
Baud rates:.....................................................2400, 4800, 9600, 19200 baud, 1 stop bit,
no parity.
Connector: .....................................................standard RS232 9-pin D female.
CLOCK
Built-in Real time clock...................................Displayed permanently and stored as a
parameter in combination with the result.
36