De Ost et al., 1995 - Google Patents
KWARTA (Quality Assurance in the Radiotherapy centres of the Antwerp province): Quality control of the contract therapy machine and treatment planning systemDe Ost et al., 1995
View PDF- Document ID
- 10182380983113142529
- Author
- De Ost B
- Schaeken B
- Vanregemorter J
- Bellekens L
- Cardoen R
- Pieters D
- Goossens H
- Haest K
- Mertens N
- Publication year
External Links
Snippet
During the first year of the provincial QA project, joint procedures were set up for the routine quality control of linear accelerators, Cobalt treatment machines and simulators. A set of standard forms was produced for use in all centres, respecting the differences in each …
- 238000002560 therapeutic procedure 0 title abstract description 38
Classifications
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1048—Monitoring, verifying, controlling systems and methods
- A61N5/1075—Monitoring, verifying, controlling systems and methods for testing, calibrating, or quality assurance of the radiation treatment apparatus
- A61N2005/1076—Monitoring, verifying, controlling systems and methods for testing, calibrating, or quality assurance of the radiation treatment apparatus using a dummy object placed in the radiation field, e.g. phantom
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1048—Monitoring, verifying, controlling systems and methods
- A61N5/1049—Monitoring, verifying, controlling systems and methods for verifying the position of the patient with respect to the radiation beam
- A61N2005/1061—Monitoring, verifying, controlling systems and methods for verifying the position of the patient with respect to the radiation beam using an x-ray imaging system having a separate imaging source
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1048—Monitoring, verifying, controlling systems and methods
- A61N5/1064—Monitoring, verifying, controlling systems and methods for adjusting radiation treatment in response to monitoring
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/103—Treatment planning systems
- A61N5/1031—Treatment planning systems using a specific method of dose optimization
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N2005/1092—Details
- A61N2005/1096—Elements inserted into the radiation path placed on the patient, e.g. bags, bolus, compensators
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N2005/1085—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy characterised by the type of particles applied to the patient
- A61N2005/1087—Ions; Protons
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1077—Beam delivery systems
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1001—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy using radiation sources introduced into or applied onto the body; brachytherapy
- A61N5/1014—Intracavitary radiation therapy
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61N—ELECTROTHERAPY; MAGNETOTHERAPY; RADIATION THERAPY; ULTRASOUND THERAPY
- A61N5/00—Radiation therapy
- A61N5/10—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy
- A61N5/1042—X-ray therapy; Gamma-ray therapy; Particle-irradiation therapy with spatial modulation of the radiation beam within the treatment head
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B6/00—Apparatus for radiation diagnosis, e.g. combined with radiation therapy equipment
- A61B6/48—Diagnostic techniques
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B6/00—Apparatus for radiation diagnosis, e.g. combined with radiation therapy equipment
- A61B6/58—Testing, adjusting or calibrating devices for radiation diagnosis
- A61B6/582—Calibration
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61B—DIAGNOSIS; SURGERY; IDENTIFICATION
- A61B6/00—Apparatus for radiation diagnosis, e.g. combined with radiation therapy equipment
- A61B6/02—Devices for diagnosis sequentially in different planes; Stereoscopic radiation diagnosis
Similar Documents
Publication | Publication Date | Title |
---|---|---|
Glatstein et al. | The imaging revolution and radiation oncology: use of CT, ultrasound, and NMR for localization, treatment planning and treatment delivery | |
Nakayama et al. | Film measurement and analytical approach for assessing treatment accuracy and latency in a magnetic resonance‐guided radiotherapy system | |
Van Duyse et al. | Modification of a 3D-planning system for use with a multileaf collimator | |
De Vries | Spectrometric methods used in the calibration of radiodiagnostic measuring instruments | |
Colle et al. | Optimization of radiotherapy to target volumes with concave outlines: target-dose homogenization and selective sparing of critical structures by constrained matrix inversion | |
Schelfhout et al. | Planning and delivering high doses to targets surrounding the spinal cord at the lower neck and upper mediastinal levels: static beam-segmentation technique executed by a multileaf collimator | |
Wong et al. | Implementation of intensity modulation with dynamic multileaf collimation | |
Derycke et al. | Limits of dose escalation in lung cancer: a dose-volume histogram analysis comparing coplanar and non-coplanar techniques | |
De Ost et al. | KWARTA (Quality Assurance in the Radiotherapy centres of the Antwerp province): Quality control of the contract therapy machine and treatment planning system | |
Boellaard et al. | Transmission dosimetry with a liquid-filled electronic portal imaging device | |
Julia et al. | Evaluation of the vidars VXR-12 digitizer performances for film dosimetry of beams delimited by multileaf collimator | |
Danciu et al. | Film dosimetry in conformal radiotherapy | |
Dessy et al. | Tomodensitometry images: integration in radiotherapy | |
Rasch et al. | CT and MRI matching for radiotherapy planning in head and neck cancer | |
Scielzo et al. | Treatment planning in radiosurgery: parallel Monte Carlo simulation software | |
Thienpont et al. | Pulsed dose rate brachytherapy (PDR): an analysis of the technique at 2 years | |
Bosmans et al. | Validation of new 3D post processing algorithm for improved maximum intensity projections of MR angiography acquisitions in the brain | |
Vynckier et al. | In-vivo dosimetry in radiotherapy: a comparison of the response of semiconductor and thermoluminescence (TLD700) dosemeters | |
Van Herk et al. | Applications of three-dimensional image correlation in conformal radiotherapy | |
Van den Heuvel et al. | Immobilization for the radiation therapy treatment of the pelvic region | |
Scarlat et al. | Quality assurance and accuracy in Co-60 radiotherapy | |
Dirkx et al. | Application of an EPID for fast daily dosimetric quality control of a fully computer-controlled treatment unit | |
Trouette et al. | Three dimensional conformal radiation therapy may improve the therapeutic ratio of radiation therapy after pneumonectomy for lung cancer | |
Lafay et al. | Evaluation of the penumbras of a Philips multileaf collimator | |
Briot et al. | Penumbra characteristics of square photon beams delimited by a GEMS multi-leaf collimator |