xylem SenTix 51 PH Electrodes
Specifications
Reference Electrolyte | KCl 3 mol/L, Ag+ free |
---|---|
Junction | Ceramic, Platinum |
Special Features | Plastic shaft, Longer shaft, Micro combination electrode |
pH Measuring Range | 0 to 14 |
Typical Applications | Field and Laboratory |
Shaft Material | Polyamide, Glass |
Cable Length | 1m |
Product Usage Instructions
Commissioning
The electrode comes pre-filled with reference electrolyte solution. To prepare for measuring:
- Open the refilling opening for the reference electrolyte solution. Ensure it remains open during calibration and measurement.
- Depending on the model, the stopper of the refilling opening may be an elastomer stopper or a slider.
Measuring
Follow these steps to measure pH using the electrode:
- Ensure the electrode is properly connected to the meter.
- Immerse the electrode in the sample solution.
- Allow time for the reading to stabilize.
- Record the pH value displayed on the meter.
Calibration
To calibrate the electrode:
- Prepare calibration solutions of known pH values (typically pH 4.01, 7.00, and 10.01).
- Immerse the electrode in each calibration solution and adjust the meter accordingly.
- Rinse the electrode between each calibration point.
Technical data
General data
Model | Reference electrolyte | Junction | NTC | Special features |
SenTix® 51 | KCl 3 mol/L, Ag+ free | Ceramic | NTC 30 kOhm | Plastic shaft |
SenTix® 52 | KCl 3 mol/L, Ag+ free | Ceramic | NTC 30 kOhm | Plastic shaft |
SenTix® 57 | KCl 3 mol/L, Ag+ free | Ceramic | NTC 10 kOhm | Plastic shaft |
SenTix® 60 | KCl 3 mol/L, Ag+ free | Platinum | No | |
SenTix® 61 | KCl 3 mol/L, Ag+ free | Platinum | No | |
SenTix® 62 | KCl 3 mol/L, Ag+ free | Platinum | No | |
SenTix® 81 | KCl 3 mol/L, Ag+ free | Platinum | NTC 30 kOhm | |
SenTix® 82 | KCl 3 mol/L, Ag+ free | Platinum | NTC 30 kOhm | |
SenTix® 83 | KCl 3 mol/L, Ag+ free | Platinum | Pt 1000 | |
SenTix® 84 | KCl 3 mol/L, Ag+ free | Platinum | Pt 1000 | |
SenTix® 85 | KCl 3 mol/L, Ag+ free | Platinum | Pt 1000 | |
SenTix® 86 | KCl 3 mol/L, Ag+ free | Platinum | NTC 30 kOhm | |
SenTix® 87 | KCl 3 mol/L, Ag+ free | Platinum | NTC 30 kOhm | |
SenTix® 91 | KCl 3 mol/L, Ag+ free | Platinum | NTC 30 kOhm | Longer shaft |
SenTix® Mic | KCl 3 mol/L, Ag+ free | Ceramic | No | Half micro combination electrode |
SenTix® MIC-B | KCl 3 mol/L, Ag+ free | Platinum | No | Micro combination electrode |
SenTix® MIC-D | KCl 3 mol/L, Ag+ free | Platinum | No | Micro combination electrode |
Measurement and application characteristics
Model | pH measur- ing range | Allowed tempera- ture range | Membrane resis- tance at 25 °C | Typical application |
SenTix® 51 | 0 … 14 | 0 … 80 °C | < 1 GOhm | Field |
SenTix® 52 | 0 … 14 | 0 … 80 °C | < 1 GOhm | Field |
SenTix® 57 | 0 … 14 | 0 … 80 °C | < 1 GOhm | Field |
SenTix® 60 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 61 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 62 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 81 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 82 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 83 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 84 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 85 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 86 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 87 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® 91 | 0 … 14 | 0 … 100 °C | < 600 MOhm | Laboratory |
SenTix® Mic | 0 … 14 | 0 … 100 °C | < 700 MOhm | Laboratory |
SenTix® MIC-B | 0 … 14 | -5 … 100 °C | < 1 GOhm | Laboratory |
SenTix® MIC-D | 0 … 14 | -5 … 100 °C | < 1 GOhm | Laboratory |
Shaft dimensions, shaft material, electrical connection
Model | Shaft | Electrical connection | ||||
Length [mm] | Ø [mm] | Material | Electrode connection | Meter con- nection | Cable length | |
SenTix® 51 | 120 | 12 | Polyamide | Fixed cable | DIN*+banana | 1 m |
SenTix® 52 | 120 | 12 | Polyamide | Fixed cable | BNC+banana | 1 m |
SenTix® 57 | 120 | 12 | Polyamide | Fixed cable | BNC+jack | 1 m |
SenTix® 60 | 120 | 12 | Glass | S7 plug-in connector | depending on S7 cable**** | |
SenTix® 61 | 120 | 12 | Glass | Fixed cable | DIN* | 1 m |
SenTix® 62 | 120 | 12 | Glass | Fixed cable | BNC | 1 m |
SenTix® 81 | 120 | 12 | Glass | Fixed cable | DIN*+banana | 1 m |
SenTix® 82 | 120 | 12 | Glass | Fixed cable | BNC+banana | 1 m |
SenTix® 83 | 120 | 12 | Glass | Fixed cable | DIN*+banana | 1 m |
SenTix® 84 | 120 | 12 | Glass | Fixed cable | BNC+banana | 1 m |
SenTix® 85 | 120 | 12 | Glass | Fixed cable | BNC+2x banana |
1m |
SenTix® 86 | 120 | 12 | Glass | Fixed cable | BNC+Cinch | 1 m |
SenTix® 87 | 120 | 12 | Glass | Fixed cable | BNC+jack | 1 m |
SenTix® 91 | 170 | 12 | Glass | Fixed cable | DIN*+banana | 1 m |
SenTix® Mic | 40/80** | 12/5** | Glass | S7 plug-in connector | depending on S7 cable**** | |
SenTix® MIC-B | 96*** | 3 | Glass | Fixed cable | BNC | 1 m |
SenTix® MIC-D | 96*** | 3 | Glass | Fixed cable | DIN* | 1 m |
- Coaxial plug according to DIN 19262
- Stage geometry
- Measured from upper level of built-in grinding (standard grinding NS 7,5)
- Connection cable not included in the scope of delivery of the combination electrode (see WEAR PARTS AND ACCESSORIES)
Commissioning, measuring, calibration
Commissioning
The electrode is filled with reference electrolyte solution in the factory. Prepare the electrode for measuring as follows:
- Open the refilling opening for the reference electrolyte solution. Depending on the model, the stopper of the refilling opening is an elastomer stopper or a slider.
- The refilling opening must always be open during calibration and measurement!
- Remove the watering cap from the electrode tip. Possible salt deposits in the area of the watering cap do not affect the measuring characteristics and can easily be removed with deionized water.
- Please keep the watering cap. It is required for the electrode to be stored. Always keep the watering cap clean.
- SenTix® 51, SenTix® 52, and SenTix® Mic: Remove any gas bubbles behind the pH membrane by shaking. With all other combination electrodes, gas bubbles behind the pH membrane are not a problem.
- Connect the combination electrode to the meter.
- Calibrate the electrode according to the operating manual of the meter and observe the following rules while doing so:
Calibration and measurement: General rules
- Make sure the refilling opening for the reference electrolyte solution is open.
- Avoid the displacement of any solution (sample or buffer solution) from one measurement to the next by taking the following measures:
- Shortly rinse the calibration and sample beakers with the solution the beakers are to be filled with next.
- Between measurements, rinse the electrode with the solution that follows.
- Alternatively, you can also rinse the electrode with deionized water and then carefully dab it dry. Immerse the electrode in the solution in a vertical or slightly tilted position.
- Make sure the immersion depth is correct. The junction must be completely submersed in the solution. The junction is in the area of the bottom end of the shaft (see figure). At the same time, the level of the reference electrolyte must be at least 2 cm above the level of the solution.
Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane.
Subsequent calibrations
The frequency of subsequent calibrations depends on the application. Many meters provide an option where you can enter a calibration interval. After the calibration interval has expired, the meter will automatically remind you of the due calibration.
Storage
During short measuring breaks
With the refilling opening open, immerse the electrode in reference electrolyte (KCl 3 mol/L, Ag+ free). Prior to the next measurement, shortly rinse the electrode with the test sample or deionized water. Prevent contact of the pH membrane to the beaker bottom to avoid scratches on the pH membrane.
Overnight or longer
Put the clean electrode in the watering cap that is filled with reference electrolyte (KCl 3 mol/L, Ag+ free) and close the refilling opening.
NOTE
pH electrodes must not be stored dry or in deionized water. The electrode could be permanently damaged by this. If the liquid in the watering cap has dried up, condition the electrode in reference electrolyte (KCl 3 mol/L, Ag+ + free) for at least 24 hours.
During longer storing periods, salt sediments may develop on the watering cap and on the refilling opening. They do not affect the measuring characteristics and can easily be removed with deionized water when the electrode is put into operation again.
Aging
pH electrodes are consumables. Every pH electrode undergoes a natural aging process. With aging, the responding behavior becomes slower and the electrode slope and asymmetry change. Moreover, extreme operating conditions can considerably shorten the lifetime of the electrode. These are:
- Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bromides, sulfides, iodides, proteins
- High temperatures
- High changes in pH and temperature.
The warranty does not cover failure caused by measuring conditions and mechanical damage.
Maintenance and cleaning
Refilling the reference electrolyte
During operation, a small amount of reference electrolyte leaks through the junction from the electrode into the test sample. If the level of reference electrolyte becomes too low with time, refill it through the refilling opening. Refilling is very easy with the enclosed dropping bottle. Proceed as follows:
- Cut off the tip of the dropping bottle at a right angle until the opening in the tip can be seen
- Open the refilling opening of the electrode
- Press the tip of the dropping bottle into the refilling opening while turning it slightly
- Pump the reference electrolyte in the shaft using the dropping bottle
- Pull the dropping bottle out of the refilling opening while turning it slightly if necessary.
Cleaning
Remove water-soluble contamination by rinsing with deionized water. Remove other contamination as follows:
Contamination | Cleaning procedure |
Fat and oil | Rinse with water containing household washing-up liquid |
Lime and hydroxide deposits | Rinse with citric acid (10 % by weight) |
Proteins | Immerse in pepsin cleaning solution PEP/pH for approx. 1 hour.
Note: Make sure the level of the reference electrolyte is above that of the cleaning solu- tion. |
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy the glass membrane.
After cleaning
Rinse the electrode with deionized water and condition it in reference electrolyte solution for at least 1 hour. Then recalibrate the electrode.
Wear parts and accessories
Description | Model | Order no. |
Reference electrolyte solution 250 mL (KCl 3 mol/L, Ag+ free) | KCl-250 | 109 705 |
Plastic arming for SenTix® pH electrodes | A pHLab/K | 903 841 |
Pepsin cleaning solution 3 x 250 ml | PEP/pH | 109 648 |
Connection cable S7 plug-in connector/DIN, 1 m | AS/DIN | 108 110 |
Connection cable S7 plug-in connector/DIN, 3 m | AS/DIN-3 | 108 112 |
Connection cable S7 plug-in connector/BNC, 1 m | AS/BNC | 108 114 |
Note: Detailed information on our wide range of buffer solutions and more accessories is given in the price list of the WTW catalog “Laboratory and field instrumentation”.
Disposal
Handle and dispose of all waste in compliance with local laws and regulations.
EU only: Correct disposal of this product — WEEE Directive on waste electrical and electronic equipment
- This marking on the product, accessories or literature indicates that the product should not be disposed of with other waste at the end of its working life.
- To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate these items from other types of waste and recycle them responsibly to promote the sustainable reuse of material resources.
- Waste from electrical and electronic equipment can be returned to the producer or distributor.
- The tissue in plants that brings water upward from the roots;
- a leading global water technology company.
We’re a global team unified in a common purpose: creating advanced technology solutions to the world’s water challenges. Developing new technologies that will improve the way water is used, conserved, and reused in the future is central to our work. Our products and services move, treat, analyze, monitor and return water to the environment, in public utility, industrial, residential and commercial building services settings. Xylem also provides a leading portfolio of smart metering, network technologies and advanced analytics solutions for water, electric and gas utilities. In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise with a strong focus on developing comprehensive, sustainable solutions.
For more information on how Xylem can help you, go to www.xylem.com.
Service and Returns: Xylem Analytics Germany Sales GmbH & Co. KG WTW Am Achalaich 11 82362Weilheim Germany
- Tel.: +49 881 183-325
- Fax: +49 881 183-414
- E–Mail wtw.rma@xylem.com
- Internet: www.xylemanalytics.com
- Xylem Analytics Germany GmbH Am Achalaich 11 82362Weilheim Germany
FAQ
Q: How often should I calibrate the pH electrode?
A: It is recommended to calibrate the pH electrode before each use for accurate measurements.
Q: Can I store the electrode dry?
A: No, always store the electrode in a storage solution to keep it hydrated and maintain its performance.
Documents / Resources
xylem SenTix 51 PH Electrodes [pdf] Owner's Manual BA-VAR-339336-EN, ba75737e05, SenTix 51 PH Electrodes, SenTix 51, PH Electrodes, Electrodes |