ABC - logo İstifadəçi təlimatı
Versiya 1.0
AED CLIMATE BOX MODELS:
ABC-2-E, ABC-2-EH, ABC-2-EHV,
ABC-2-EH-GSM.ABC 2 E Aed Climate Box -

Giriş

The AED Climate Box is an innovative device designed specifically to protect and maintain AED defibrillators in optimal conditions, ready for use 24 hours a day. Due to the crucial role that defibrillators play in saving lives, their efficiency and availability in any conditions are critical. Unfortunately, improper storage conditions, such as extreme temperatures, can seriously damage the devices or even render them completely unusable, often only discovered when their use is most needed.
The solution to these challenges is the AED Climate Box, which offers advanced protective and monitoring features, including:

  •  Wi-Fi and GSM: Enable remote monitoring and comunication with the device,
  • Ventilation and heating: Automatically regulate the temperature inside, preventing damage caused by extreme temperature conditions. Users can adjust the heating temperature to ensure optimal conditions for AED,
  • Gravity ventilation: Ensures a constant airflow, maintaining proper circulation,
  • Mechanical ventilation: Two fans with a capacity of 60 m³/h each, with temperature regulation options for the user,
  • UV glass: Protection against the negative effects of UV radiation,
  • Accelerometer: Detects any unexpected movements or shocks,
  • AED presence sensor: Intelligently monitors whether the AED is in place and ready for use,
  • Acoustic alarm: Informs about various scenarios, such as door opening or other events requiring attention,
  • Internal LED lighting: Provides visibility inside the device.

Paket məzmunu

  • AED Climate Box device
  • Enerji təchizatı
  • İstifadəçi təlimatı

Tikinti

ABC 2 E Aed Climate Box - ConstructionABC 2 E Aed Climate Box - Construction1

Cihaz bağlantısı

The AED Climate Box comes with two LiFePO4 batteries, providing reliable backup power and are ready for use upon installation.
The device startup process is as follows:

  1. Activation of LiFePO4 Batteries:
    – locate the battery hole on internal control panel,
    – using screwdriver, switch both battery switches to the „ON” position to activate the batteries.
  2.  Connection of 24V 150W Power Supply:
    – The power supply is included with the device and should be connected to the AED Climate Box by a qualified electrician,
    – The 2-meter power cord can be extended as needed, but any changes to the electrical installation should be performed by a certified electrician.
  3. Connection to the Electrical Network:
    – After activating the batteries and connecting the power supply by an electrician, plug the power supply into the electrical outlet.

İlkin quraşdırma

After connecting the power supply, the device will start up in initial setup mode. This mode is characterized by the absence of any checks (no temperature or humidity measurements, no defibrillator presence check, etc.) and active an open Wi-Fi network for further device configuration. The active WiFi network name is unique for each device and looks as follows: AED_ Climate_Box_XXXXXX, where XXXXXX is replaced by a unique fragment of the device identifier.
The first startup mode is indicated by a brief activation of the speaker, pulsating internal LED lights, and alternating pulsation of the status LED in red, green, and blue colors.
To configure the device, connect your phone/tablet/computer, to the Wi-Fi network broadcasted by the device (for example: AED_Climate_Box_E0E664), and then open any web browser (Chrome browser recommended) and enter the IP address in the address bar: 192.168.4.1, then press Enter to proceed. It may happen that the page does not load because the browser automatically tries to connect using encryption (See Figure 1).

ABC 2 E Aed Climate Box - setup

To correctly load the configuration page, you need to replace „https” with „http” in the address bar and press Enter again (See Figure 2).

ABC 2 E Aed Climate Box - setup1

After loading the configuration page, the Wi-Fi network configuration tab will be displayed (See Figure 3). Following the field labels, you need to create your own network name, which should be entered in the field labeled „SSID,” and the Wi-Fi network password, which should be entered in the field labeled „PASSWORD.” This will allow you to control and view the parameters reported by the device. Restrictions: network name – maximum of 32 characters; password – minimum of 8 characters, maximum of 32 characters. Next, the user selects the cabinet model. The model depends on the cabinet’s equipment:

  • ACB-1-E/ ACB-2-E – cabinet with only the electronic module for communication with AED CONNECT,
  • ACB-1-EH/ ACB-2-EH – cabinet with the electronic module for communication with AED CONNECT and a heater,
  • ACB-1-EHV/ ACB-2-EHV – cabinet with the electronic module for communication with AED CONNECT, a heater, and fans,
  • ACB-1-EHV-GSM/ ACB-2-EHV-GSM – cabinet with the electronic module for communication with AED CONNECT, a heater, fans, and GSM module.

After entering the provided parameters, press the Confirm button. At this point, the device will go through the sensor initialization stage and start operating. In case of a mistake, press and hold the reset button located on the board for 10 seconds. Then, reconnect to the Wi-Fi network broadcasted by the cabinet and start the configuration process again.

Sensorun işə salınması

The sensor initialization stage is fully automatic for the user and is performed each time the cabinet is restarted. The following actions take place during this stage:

  • Activation of the temperature and humidity sensor, accelerometer, anddistance sensor,
  • Initiation of communication with theWi-Fi microcontroller – retrieving settings and sending current status,
  • Checking the presence of the heater – the presence of the heater is checked by measuring the supplyvoltage (reference measurement), then measuring the supply voltage during a short-term heater activation (10 ms, operation measurement), and comparing  both voltages. Əgər cildtages differ by more than 50 mV, the heater is considered fully functional,
  • Initialization of the GSM modem and sending a report after startup. (Only for cabinets with the GSM module).

İş rejimi

The device enters operating mode after startup, successful initialization, and loading of settings indicating the need to enter operating mode. The presence of the device in operating mode can be recognized by the status LED state – it will be steady or pulsating according to the state table. In the case of a full and correct connection, the LED will shine steady green.
In operating mode, all algorithms necessary for the proper operation of the device are executed:

  • Monitoring of temperature and humidity sensor,
  • Control and maintenance of temperature according to user settings,
  • Control of operation, faults, and control of fans and heater (implemented based on a PD controller, controller settings are configurable),
  • Control of digital and analog inputs and outputs, external devices connected to the AED Climate BOX,
  • Activation of acoustic and visual alarms according to scenarios,
  • Control of the USB socket – the USB socket is hardware-controlled in case of an attempt to draw too much current,
  • Control of external lighting (if connected), internal lighting, and status LED,
  • Listening to state sounds emitted by the defibrillator and reporting to AED Connect,
  • Communication with the Wi-Fi module and GSM module to report changes in the state of the AED Climate BOX cabinet to AED Connect. By default, changes via GSM are reported every 15 minutes, while alarms such as door opening or defibrillator removal are reported immediately. Changes via Wi-Fi are reported every 1 minute, and alarms are reported immediately.
AED Climate BOX configuration

When the device is in operating mode, it provides access to the configuration page via Wi-Fi (See Figure 4). Similar to the first configuration mode, to access the settings of the AED Climate BOX, you need to connect to the device active Wi-Fi network (previously set by the user’s network name) and enter the address 192.168.4.1 in the browser’s address bar.

ABC 2 E Aed Climate Box - configuration

Upon accessing the configuration page, where two configuration tabs are available in the menu, namely: STATUS and CONFIG, the contents of the STATUS tab will be displayed by default.

8.1 Device Status Preview Wi-Fi vasitəsilə
STATUS Tab:

  • MODE Section (See Figure 4): Displays information about the current operating mode of the device:
    – NORMAL mode: The cabinet operates in normal mode, meaning all messages are sent to AED Connect,
    – SERVICE mode: The cabinet operates in service mode, where no messages or alarms are sent to AED Connect.
  • STATUS Section: Displays the current status of the cabinet according to table number 1.

TABLE No. 1 – Possible Statuses:

Cihaz statusu Statusun təsviri
No errors or alarms BOŞ
Zəif batareya BATARYA AZ
No main power supply – operating on battery NO MAIN VOLTAGE
İstilik İSTİLMƏ
Havalandırma ventilyasiya
GSM bağlantısı yoxdur GSM NETWORK UNAVAILABLE
Wi-Fi bağlantısı yoxdur Wi-Fi NETWORK UNAVAILABLE
Açıq qapı QAPI AÇILDI
Defibrillator removed AED MISSING
Qızdırıcının nasazlığı HEATER NOT OPERATIONAL
Fan nasazlığı FAN 1/2 NOT OPERATIONAL
AED presence sensor calibration AED CALIBRATION
Defibrillator presence sensor turned off (10s) AED INACTIVE
Zərbənin aşkarlanması SHOCK DETECTION

The status field is not exclusive, meaning it can have more than one status, such as IDLE, BATTERY LOW, HEATING.

  • TEMPERATURE Section: Displays the current temperature (˚ C),
  • HUMIDITY Section: Displays the current humidity (%),
  • PERIPHERALS STATUS Section: Displays the following information:
    – (OFF – door closed, ON – door open),
    – AED defibrillator presence status (OFF – no AED defibrillator detected, OK – AED defibrillator detected).
  • CONNECTIVITY STATUS Section: Displays the current connection status with AED Connect:
    – NO CLOUD CONNECTION – no connection to AED Connect,
    – CLOUD CONNECTION OK – active connection to AED Connect.
    Depending on the equipment of the AED Climate BOX cabinet, information about the connection technology (Wi-Fi or GSM) is displayed below the AED Connect connection status, along with network parameters and RSSI signal strength.
  • DEVICE INFO Section: Contains information about the device’s serial number, MAC address, and software version.

8.2 Configuration of parameters and calibration
The CONFIG tab allows for the configuration of AED Climate BOX settings.
Due to its content and better understanding, the view has been divided into several graphics, as shown below.

ABC 2 E Aed Climate Box - configuration1

8.2.1 Heating and ventilation settings

The Temperatures section (See Figure 5) provides the possibility to configure the thresholds for activating and deactivating the heater and fans:

  • Critical low – the threshold for heater activation; heating is activated below this temperature.
  • Heater off – the threshold for heater deactivation; heating is turned off once the desired temperature is reached.
  • Critical high – the threshold for fan activation; fans are activated above this temperature.
  • Vents off – the threshold for fan deactivation; fans are turned off below this temperature.

Correctly set thresholds must satisfy the relation: critical low < heater off < vents off < critical high. Any changes made must be confirmed by pressing the Save temperature button.
Qeyd: Temperature settings should correspond to the minimum and maximum operating temperatures for the specific defibrillator model. Temperature settings directly affect energy consumption.

ABC 2 E Aed Climate Box - configuration2

8.2.2 AED Connect communication frequency
The Cloud reporting section enables the user to set the frequency of sending status messages from the AED Climate BOX cabinet to AED Connect (See Figure 6).
8.2.3 AED presence detector calibration
The AED calibration section allows for the calibration of the AED defibrillator presence sensor (See Figure 6). Calibration is required when the AED Climate BOX is set to notify with an acoustic alarm and report to AED Connect when the defibrillator is removed from the cabinet.
The Calibration status field provides information about the current calibration status:

  • not calibrated – indicates lack of calibration for the height of the placed defibrillator,
  • calibrated – indicates calibrated height.

Before clicking the Calibrate button, familiarize yourself with the following information and the calibration procedure.
To calibrate the defibrillator presence sensor:

  • Ensure that there are no objects in the AED presence sensor’s working area (the area between the top and bottom of the cabinet’s interior),
  • Click the Calibrate button, which triggers the device to enter calibration mode, removing old calibration data and deactivating active alarms during calibration,
  • Open the cabinet doors and hang the defibrillator on the hook within 10 seconds.
    Placing the defibrillator on the bottom of the cabinet is not permissible as it may damage the defibrillator due to the cabinet’s heating element.
    During calibration, a periodic intermittent acoustic signal will be emitted, indicating the ongoing procedure and allowing time for placing the AED in the cabinet and closing it. The intermittent acoustic signal will be emitted for 10 seconds and will remain insensitive to the absence of the AED in the cabinet. The cessation of the intermittent acoustic signal indicates the end of the calibration procedure,
  • Completion of the calibration procedure will change the status from not calibrated to calibrated.

ABC 2 E Aed Climate Box - configuration3

8.2.4 Lighting scenario
The Lighting scenario section provides the ability to change the external lighting scenario (according to the lighting operation scenarios). The Active scenario field indicates the currently active lighting scenario, while the Change to scenario X button allows switching to another scenario. By default, there is only one scenario regardless of the settings.
8.2.5 Service mode
The Mode section (See Figure 6) allows switching between operating mode and service mode. In service mode, the device does not transmit any alarm information to AED Connect. Pressing the Switch to service mode button will switch the device to Service mode, while pressing the Switch to normal mode button will switch the device to Normal mode (operating mode).
8.2.6 Settings for Connecting to AED connect
Depending on the equipment of the AED Climate BOX, the device can establish a connection with AED Connect via Wi-Fi or GSM technology.
a/ GSM settings (see Figure 7).

ABC 2 E Aed Climate Box - configuration4

If the AED Climate BOX was not factory-equipped with a GSM SIM card and it has GSM models, the following configuration settings should be performed.
The Modem Settings section provides an interface for configuring the GSM modem.
The SIM PIN field allows entering the SIM card PIN code (if required). The currently entered PIN code is displayed in the Current PIN field (no PIN is indicated as „—”).
The APN field allows configuring the Access Point Name (APN) to which the device should connect. The currently configured APN is displayed in the Current APN field (no APN is indicated as „–”).
Share modem connection with devices connected to local hotspot – enables or disables sharing the AED Climate BOX’s internet connection with other devices as a local hotspot. The Current state field indicates the current state of this option – „enabled” indicates sharing is enabled, while „disabled” indicates sharing is disabled.
The Connectivity technology selection field allows choosing the technology the modem will use to connect to the network. Available options include GSM, CAT-M, and AUTO. It is recommended to set the option to AUTO.
Each change in settings must be confirmed by clicking the Save Modem Settings button.
b/ Wi-Fi Settings (See Figure 8)
The SSID field allows entering the nameof the access point (the Wi-Fi network name to which the AED Climate BOX modem should connect), while the Password field allows entering the password for the selected Wi-Fi network. When entering values, similar restrictions apply as when setting up the SSID and password during the initial device configuration. However, in this section, the Password field can remain empty.
The Current SSID and Current password fields provide information about the current access data for the Wi-Fi network („—” indicates an empty field).

ABC 2 E Aed Climate Box - configuration5

8.2.7 Accelerometer Settings
The Shock detection section (See Figure 8) allows setting parameters related to shock detection (vandalism protection).
The Threshold field allows entering the threshold value for shock detection.
This field accepts values ranging from 10 to 30, where a value of 10 indicates high sensitivity to shocks, while 30 indicates low sensitivity.
Each change in settings must be confirmed by clicking the Save settings button.
8.2.9 Alarm reset
The Miscellaneous section (See Figure 8) allows the user to reset an alarm upon request. Pressing the Reset alarm button will deactivate the audible and visual signals, while the alarm information will still be displayed in the status field.
8.2.10 Zavod parametrlərinə sıfırlama
The Factory reset section (See Figure 8) provides the option to restore the device to its factory settings (i.e., the state before the first startup). Pressing the FACTORY RESET button will erase all settings and restart the device in the first startup mode.

Operation on emergency battery power

The device is capable of operating solely on emergency battery power. In the event of a main power supply failure, a timer is activated (3 seconds), after which the presence of the main power supply is checked again. Further absence of power triggers the device to enter partial sleep mode – the WiFi and modem modules are completely turned off, and only alarm reports are sent to AED Connect, while cyclic reports are ignored. Exiting partial sleep mode is achieved in reverse: upon detection of the main power supply presence, a timer is activated (3 seconds), after which the presence of the main power supply is checked again. Continued presence of the main power supply triggers the complete awakening of the Wi-Fi and modem modules. The following circuits do not operate on battery power: heater, fans, internal and external lighting, battery chargers, USB.

Məlumatların saxlanması

In the event of a disconnection from AED Connect, the data retention mode is activated. The device records the current timestamp (local) and the temperature inside the cabinet every 15 minutes. The device’s memory is sufficient to store 16384 data records. Upon reconnecting to AED Connect, a frame containing historical temperature data is sent.

Table of states

In the operating mode, the status LED and the sound signal behave according to the following state table.

Cihaz vəziyyəti LED Operation Mode LED Rəngi Sound Signal Operation Mode
Full power supply Davamlı ABC 2 E Aed Climate Box - icon yoxluq
Aşağı batareya Intermittent fast ABC 2 E Aed Climate Box - icon interrupted every 15 seconds
Batareya gücü Intermittent fast ABC 2 E Aed Climate Box - icon interrupted every 15 seconds
İstilik Davamlı ABC 2 E Aed Climate Box - icon1 yoxluq
Havalandırma Davamlı ABC 2 E Aed Climate Box - icon2 yoxluq
GSM bağlantısı yoxdur Intermittent fast ABC 2 E Aed Climate Box - icon2 interrupted every 15 seconds
Wi-Fi bağlantısı yoxdur Intermittent fast ABC 2 E Aed Climate Box - icon2 interrupted every 15 seconds
Açıq qapı Davamlı ABC 2 E Aed Climate Box - icon interrupted fast
No AED Davamlı ABC 2 E Aed Climate Box - icon3 yoxluq
Qızdırıcının nasazlığı Intermittent fast ABC 2 E Aed Climate Box - icon3 yoxluq
Fan nasazlığı Intermittent fast ABC 2 E Aed Climate Box - icon4 yoxluq

At any given moment, the AED Climate BOX can indicate only one state using the LED and sound signal.
The states are presented according to a specific priority:

1. Heater failure 6. Doors open
2. Fan failure 7. No main power supply
3. No Wi-Fi network 8. Batareya azdır
4. No GSM network 9. İstilik
5. No AED 10. Havalandırma

Siqnal vəziyyətləri

The device has three triggering sources for alarms:

  • Door opening (2-second delay after opening),
  • Defibrillator removal,
  • Shock detection.

Each alarm source has its own characteristic alarm operation. Alarms are also partially dependent on each other (door opening and defibrillator removal). The sound signal operation during an alarm state depends on its operation mode; the priority signal is continuous, meaning that opening the door during a shock alarm will not change the alarm operation mode to fast interrupted. Factory alarm scenarios:

  • Door opening – interrupted alarm, internal lighting in strobe mode,
  • Door opening and defibrillator removal – fast interrupted acoustic alarm after door opening, internal lighting in strobe mode. Continuous alarm after defibrillator removal (duration of 60 seconds). After 60 seconds, the alarm resets; the door alarm rearms after closing,
  • Door opening, defibrillator removal, door closing after 30 seconds – fast interrupted acoustic alarm after door opening, continuous alarm after defibrillator removal (for 60 seconds), internal lighting in strobe mode after door opening, continuously on after door closing, alarm resets after 60 seconds,
  • Door opening, defibrillator removal, door closing after 100 seconds – fast interrupted acoustic alarm after door opening, continuous alarm after defibrillator removal (for 60 seconds), internal lighting in strobe mode after door opening, continuously on after alarm duration, alarm resets after 60 seconds,
  • Door opening, shock detection above threshold, door closing – interrupted alarm after door opening, continuous alarm after shock detection (for 60 seconds), internal lighting in strobe mode after door opening, continuously on after door closing, alarm resets after 60 seconds,
  • Shock detection, door opening – continuous alarm (for 60 seconds), interrupted alarm after 60 seconds, internal lighting in strobe mode after door opening.

Problemlərin aradan qaldırılması

Problem Mümkün Səbəb Fəaliyyət
The configuration page does not load The web browser tries to connect to the page
using encryption
Change “https” to “http” and try again
AED Climate BOX reports a network
without internet access and returns to a network with internet access when attempting to connectto the Wi-Fi network
of the AED cabinet
Devices prioritize networks with internet access When trying to connect to the Wi-Fi
broadcasted by the AED cabinet, select to
connect to a network without internet
access always or only once
Unable to connect to the Wi-Fi network
broadcasted by the AED Climate BOX
The Wi-Fi of the AED Climate BOX is
connected to anotherdevice
Disconnect other devices previously connected to the Wi-Fi of the AED Climate BOX. Only
one external device can be connected to the Wi-Fi of the AED Climate BOX at a time (both in the first setup mode and in normal operation).

Qoşmalar

Qoşma 1
– Technical datasheets of the devices.

ABC - logo1

Bizimlə əlaqə saxlayın
GRAS PPPH
Sławieńska St. 12
77-231 Korzybie
POLŞA
telefon +48 59 857 73 03
faks +48 59 858 63 04
e-poçt: info@aedcb.com
www.aedcb.comABC - logo