Simplified and Low-Cost Characterization of Medium-Voltage Low-Power Voltage Transformers in the Power Quality Frequency Range
<p>Proposed measurement setup.</p> "> Figure 2
<p>Main features of the fluke calibrator.</p> "> Figure 3
<p>Results of the resistive divider characterization: (<b>a</b>) ratio and (<b>b</b>) phase displacement.</p> "> Figure 4
<p>Measurement setup adopted for the reference test.</p> "> Figure 5
<p>Ratio error 95% confidence intervals for different frequencies and for device A.</p> "> Figure 6
<p>Phase displacement 95% confidence intervals for different frequencies and for device A.</p> "> Figure 6 Cont.
<p>Phase displacement 95% confidence intervals for different frequencies and for device A.</p> "> Figure 7
<p>Ratio error 95% confidence intervals for different frequencies and for device P.</p> "> Figure 7 Cont.
<p>Ratio error 95% confidence intervals for different frequencies and for device P.</p> "> Figure 8
<p>Phase displacement 95% confidence intervals for different frequencies and for device P.</p> "> Figure 8 Cont.
<p>Phase displacement 95% confidence intervals for different frequencies and for device P.</p> ">
Abstract
:1. Introduction
2. IT4PQ Framework
2.1. The Consortium
2.2. The Project Goals
- Definition of accuracy limits for ITs and LPITs in PQ measurements. Such limits will be fixed after their experimental validation.
- Establishment of reference systems and testing conditions capable of characterizing ITs and LPITs in order for them to perform reliable PQ measurements.
- Evaluation of ITs, during PQ measurements, in the presence of multiple influence quantities and factors (temperature, humidity, proximity effects, vibrations, etc.).
3. Simplified and Low-Cost Characterization
3.1. Measurement Setup
- Function generator (FG). An arbitrary FG RIGOL DG1022 is used to generate sinusoidal low-voltage signals at different frequencies. Its output voltage is referred to as VFG. The FG can provide sinusoidal signals from 1 µHz to 25 MHz, with an accuracy of ±1 ppm of the setting value and a resolution of 1 µHz.
- Power amplifier. The analog device ADHV4702-1 precision operational amplifier (OA), used with its evaluation board, is used to increase the FG voltage to a more suitable level for the testing (referred to as VAMP). The main features of the OA are listed in Table 1.
- LPVT. In this work, two off-the-shelf MV LPVTs are used for testing. One is active and one is passive. Therefore, from hereon after, they are referred to as A and P, respectively. Their main characteristics are summarized in Table 2. As it can be noted from the table, the two devices, despite one being active and one being passive, are quite similar in terms of electrical parameters. They share the same accuracy class (AC), allowing the results of the characterization to be significantly compared and assessed.
- Reference divider. This is a custom-made active resistive divider. Its transformation ratio is 240, and it is used as a reference divider. Its characterization is described in Section 3.2. The choice of an active divider was made to avoid loading effect issues.
- Data acquisition system (DAQ). An NI 9239 DAQ board is used to acquire the voltage signals coming from the reference divider and the LPVTs. The two voltages are referred to as VREF and VVT, respectively (in accordance with Figure 1). The DAQ features a 24 bit analog-to-digital converter, an operating voltage range of ±10 V, simultaneous sampling at a maximum of 50 kS/s/ch, and an input impedance of 1 MΩ. The DAQ gain error is ±0.03%, while the offset error is 0.008%. As for the noise, it features 70 µV (rms).
3.2. Characterization of the Reference Divider
3.2.1. Measurement Setup
3.2.2. Experimental Characterization of the Divider
3.3. Test Procedure
3.3.1. Need for Testing
3.3.2. The Procedure
3.3.3. The Validation Procedure
3.4. Experimental Results
4. Uncertainty Evaluation
4.1. Sources of Uncertainty
4.2. MCM Application
4.3. MCM Results and Discussion
- Each graph illustrates one frequency.
- Blue dots are dedicated to the reference tests (with the calibrator), while green dots are dedicated to the proposed test (function generator plus amplifier).
- Next to each dot, for the sake of clarity, the voltage level of the test is given.
- The confidence intervals, obtained with the MCM, include both type A and type B methods to evaluate uncertainty.
- To keep the length of the results reasonable, not all frequencies are plotted for each device.
5. Conclusions
Author Contributions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Zavoda, F.; Lemire, R.; Abbey, C. Implementing predictive distribution maintenance using a universal controller. In Proceedings of the 2014 IEEE PES T&D Conference and Exposition, Chicago, IL, USA, 14–17 April 2014. [Google Scholar]
- Wang, X.; Wang, C.; Xu, T.; Meng, H.; Li, P.; Yu, L. Distributed voltage control for active distribution networks based on distribution phasor measurement units. Appl. Energy 2018, 229, 804–813. [Google Scholar] [CrossRef]
- Andreoni, R.; Macii, D.; Brunelli, M.; Petri, D. Tri-objective optimal PMU placement including accurate state estimation: The case of distribution systems. IEEE Access 2021, 9, 62102–62117. [Google Scholar] [CrossRef]
- Joseph, T.; Tyagi, B.; Kumar, V. Unbiased minimum variance filter-based generator state estimation using PMU measurements for unknown generator input. IEEE Syst. J. 2019, 13, 3176–3184. [Google Scholar] [CrossRef]
- Kolosok, I.; Korkina, E. Refinement of Transmission Line Equivalent Circuit Parameters at Power System State Estimation by PMU Data. Available online: https://link.springer.com/chapter/10.1007/978-3-030-39225-3_62 (accessed on 15 December 2021).
- Apostolopoulos, C.A.; Arsoniadis, C.G.; Georgilakis, P.S.; Nikolaidis, V.C. Evaluating a Fault Location Algorithm for Active Distribution Systems Utilizing Two-Point Synchronized or Unsynchronized Measurements. In Proceedings of the 2021 International Conference on Smart Energy Systems and Technologies (SEST), Vaasa, Finland, 6–8 September 2021. [Google Scholar]
- Fan, W.; Liao, Y.; Kang, N. Optimal fault location for power distribution systems with distributed generations using synchronized measurements. Int. J. Emerg. Electr. Power Syst. 2020, 21, 20200093. [Google Scholar] [CrossRef]
- Mingotti, A.; Ghaderi, A.; Mazzanti, G.; Peretto, L.; Tinarelli, R.; Valtorta, G.; Danesi, S. Low-cost monitoring unit for MV cable joints diagnostics. In Proceedings of the 2018 IEEE 9th International Workshop on Applied Measurements for Power Systems (AMPS), Bologna, Italy, 26–28 September 2018; 2018. [Google Scholar]
- Lu, P.; Byerly, K.; Buric, M.; Zandhuis, P.; Sun, C.; Leary, A.; Ohodnicki, P.R. Distributed fiber-optic sensor for real-time monitoring of energized transformer cores. In Proceedings of the SPIE Security + Defence, Anaheim, CA, USA, 18 May 2017. [Google Scholar]
- Zhao, Z.; See, K.Y. A multiprobe inductive coupling method for online impedance measurement of electrical devices distributed in multibranch cables. IEEE Trans. Instrum. Meas. 2020, 69, 5975–5977. [Google Scholar] [CrossRef]
- Douglin, R.H.; Ituzaro, F.A.; Butler-Purry, K.L. Using smart meter data to improve the performance of the overcurrent protective devices in distribution systems with DG. In Proceedings of the 2012 IEEE Power and Energy Society General Meeting, San Diego, CA, USA, 22–26 July 2012. [Google Scholar]
- Abdel-Majeed, A.; Braun, M. Low voltage system state estimation using smart meters. In Proceedings of the 47th International Universities Power Engineering Conference (UPEC), Uxbridge, UK, 31 December 2012. [Google Scholar]
- Bartolomei, L.; Mingotti, A.; Pasini, G.; Peretto, L.; Rinaldi, P.; Tinarelli, R.; Puddu, L. Performance evaluation of an energy meter for low-voltage system monitoring. J. Phys. Conf. Ser. 2018, 1065, 052032. [Google Scholar] [CrossRef] [Green Version]
- Brandolini, A.; Faifer, M.; Ottoboni, R. A simple method for the calibration of traditional and electronic measurement current and voltage transformers. IEEE Trans. Instrum. Meas. 2019, 58, 1345–1353. [Google Scholar] [CrossRef]
- Kaczmarek, M.A. Practical approach to evaluation of accuracy of inductive current transformer for transformation of distorted current higher harmonics. Electr. Power Syst. Res. 2015, 119, 258–265. [Google Scholar] [CrossRef]
- Faifer, M.; Laurano, C.; Ottoboni, R.; Toscani, S.; Zanoni, M.; Crotti, G.; Giordano, D.; Barbieri, L.; Gondola, M.; Mazza, P. Overcoming Frequency Response Measurements of Voltage Transformers: An Approach Based on Quasi-Sinusoidal Volterra Models. IEEE Trans. Instrum. Meas. 2019, 68, 2800–2807. [Google Scholar] [CrossRef]
- Budovsky, I. Calibration of precision current transformers and AC resistors by comparison using two sampling digital voltmeters. In Proceedings of the Conference on Precision Electromagnetic Measurements (CPEM), Denver, CO, USA, 24–28 August 2020. [Google Scholar]
- Mingotti, A.; Peretto, L.; Tinarelli, R.; Zhang, J. Use of COMTRADE Fault Current Data to Test Inductive Current Transformers. In Proceedings of the 2019 II Workshop on Metrology for Industry 4.0 and IoT (MetroInd4.0&IoT), Naples, Italy, 4–6 June 2019. [Google Scholar]
- Crotti, G.; Gallo, D.; Giordano, D.; Landi, C.; Luiso, M. Non-conventional instrument current transformer test set for industrial applications. In Proceedings of the 2014 IEEE International Workshop on Applied Measurements for Power Systems Proceedings (AMPS), Aachen, Germany, 24–26 September 2014. [Google Scholar]
- Mingotti, A.; Peretto, L.; Tinarelli, R. Smart characterization of rogowski coils by using a synthetized signal. Sensors 2020, 20, 3359. [Google Scholar] [CrossRef] [PubMed]
- Crotti, G.; Gallo, D.; Giordano, D.; Landi, C.; Luiso, M.; Cherbaucich, C.; Mazza, P. Low cost measurement equipment for the accurate calibration of voltage and current transducers. In Proceedings of the 2014 IEEE International Instrumentation and Measurement Technology Conference (I2MTC) Proceedings, Montevideo, Uruguay, 12–15 May 2014. [Google Scholar]
- Letizia, P.S.; Crotti, G.; Giordano, D.; Delle Femine, A.; Gallo, D.; Landi, C.; Luiso, M. Low Cost Procedure for Frequency Characterization of Voltage Instrument Transformers. In Proceedings of the 2019 IEEE International Instrumentation and Measurement Technology Conference (I2MTC), Auckland, New Zealand, 20–23 May 2019. [Google Scholar]
- Bolacell, G.S.; Venturini, L.F.; Da Rosa, M.A. Distribution System Reliability Evaluation Considering Power Quality Effects. In Proceedings of the 2018 IEEE International Conference on Probabilistic Methods Applied to Power Systems (PMAPS), Boise, ID, USA, 24–28 June 2018. [Google Scholar]
- Ratanatanyong, S.; Chalermwai, A.; Kanchanachinto, T.; Chantabutr, S.; Wasanapradit, T.; Chigbo, I. A novel investigation of power quality effects on premature ESP failures. In Proceedings of the Abu Dhabi International Petroleum Exhibition & Conference, Abu Dhabi, United Arab Emirates, 9 November 2020. [Google Scholar]
- IEC TR 61869-103:2012; Instrument Transformers—The Use of Instrument Transformers for Power Quality Measurement. International Standardization Organization: Geneva, Switzerland, 2016.
- IEC 61869-1:2010; Instrument Transformers—Part 1: General Requirements. International Standardization Organization: Geneva, Switzerland, 2010.
- IEC 61869-6; Part 6: Additional General Requirements for Low-Power Instrument Transformers. In Instrument Transformers. International Standardization Organization: Geneva, Switzerland, 2016.
- BS EN 50160:2010+A1; Voltage Characteristics of Electricity Supplied by Public Electricity Networks. BSI Standards Publication: London, UK, 2015.
- ISO/IEC Guide 98-3: 2008; Uncertainty of Measurement—Part 3: Guide to the Expression of Uncertainty in Measurement (GUM: 1995). International Standardization Organization: Geneva, Switzerland, 2008.
- Supplement 1 BIPM JCGM 101:2008; Evaluation of Measurement Data—Supplement 1 to the Guide to the Expression of Uncertainty in Measurement—Propagation of Distributions Using a Monte Carlo Method. International Standardization Organization: Geneva, Switzerland, 2008.
- Asprou, M.; Kyriakides, E.; Albu, M.M. Uncertainty bounds of transmission line parameters estimated from synchronized measurements. IEEE Trans. Instrum. Meas. 2019, 68, 2808–2818. [Google Scholar] [CrossRef]
- Goklani, H.; Gajjar, G.; Soman, S.A. Instrument transformer calibration and robust estimation of transmission line parameters using PMU measurements. IEEE Trans. Power Syst. 2021, 36, 1761–1770. [Google Scholar] [CrossRef]
- Cataliotti, A.; Cosentino, V.; Telaretti, E.; Di Cara, D.; Tine, G. Uncertainty evaluation of a Backward/Forward load flow algorithm for a MV smart grid. In Proceedings of the 2015 IEEE International Instrumentation and Measurement Technology Conference (I2MTC) Proceedings, Pisa, Italy, 11–14 May 2015. [Google Scholar]
- Mingotti, A.; Baldi, A.; Peretto, L.; Tinarelli, R. A general easy-to-use expression for uncertainty evaluation in residual voltage measurement. IEEE Trans. Instrum. Meas. 2020, 69, 1576–1584. [Google Scholar] [CrossRef]
- Nauta, S.; Serra, R. Zero-sequence current measurement uncertainty in three-phase power systems during normal operation. IET Gener. Transm. Distrib. 2021, 15, 3450–3458. [Google Scholar] [CrossRef]
- Mingotti, A.; Peretto, L.; Tinarelli, R.; Ghaderi, A. Uncertainty sources analysis of a calibration system for the accuracy vs. temperature verification of voltage transformers. J. Phys. Conf. Ser. 2018, 1065, 052041. [Google Scholar] [CrossRef] [Green Version]
- Isaiev, V.; Velychko, O. Metrological characterisation of current transformers calibration unit for accurate measurement. Acta IMEKO 2021, 10, 6–13. [Google Scholar] [CrossRef]
Dual supply | ±12 V to ±110 V | CMRR | 160 dB |
Slew rate | 74 V/µs | Input bias current | Max 2 pA |
Input noise | 8 nV/√Hz | Bandwidth | 10 MHz |
Feature | A (Active) | P (Passive) |
---|---|---|
Primary voltage (kV) | 20/√3 | 20/√3 |
Rated transformation ratio (-) | 20/√3 | 8200 |
Rated burden (MΩ) | – | 1 |
Accuracy class | 0.5 | 0.5 |
Rated frequency (Hz) | 50 | 50 |
Auxiliary supply (V) | ±12 | – |
Range (V) | Frequency (Hz) | Harmonic Distortion | Nonharmonic Noise Floor (%) | |
---|---|---|---|---|
% Setting | % Range | |||
90 | 16 to 850 | 0.016 | 0.003 | 0.025 |
850 to 6000 | 0.25 | 0.015 | 0.025 | |
180 | 16 to 850 | 0.016 | 0.003 | 0.016 |
850 to 6000 | 0.25 | 0.015 | 0.016 | |
360 | 16 to 850 | 0.016 | 0.003 | 0.05 |
850 to 6000 | 0.25 | 0.015 | 0.05 | |
650 | 16 to 850 | 0.016 | 0.003 | 0.1 |
850 to 6000 | 0.25 | 0.015 | 0.1 | |
1008 | 16 to 850 | 0.016 | 0.003 | 0.1 |
850 to 6000 | 0.25 | 0.015 | 0.1 |
Frequency (Hz) | Harmonic | Limit from [28] | Applied Voltage | |
---|---|---|---|---|
% | (V) | |||
150 | 3rd | 5.0% | ~0.5 | 61.5 |
1.0 | 120 | |||
2.1 | 250 | |||
4.3 | 500 | |||
5.0 | 580 | |||
250 | 5th | 6.0% | ~0.5 | 61.5 |
1.0 | 120 | |||
2.1 | 250 | |||
4.3 | 500 | |||
5.0 | 580 | |||
6.0 | 695 | |||
350 | 7th | 5.0% | ~0.5 | 61.5 |
1.0 | 120 | |||
2.1 | 250 | |||
4.3 | 500 | |||
5.0 | 580 | |||
450 | 9th | 1.5% | ~0.5 | 61.5 |
1.0 | 120 | |||
1.5 | 175 | |||
750 | 15th | 0.5% | ~0.5 | 61.5 |
1000 | 20th | 0.5% | ~0.5 | 61.5 |
1250 | 25th | 1.5% | ~0.5 | 61.5 |
1.0 | 120 | |||
1.5 | 175 | |||
1600 | 32nd | 0.5% | ~0.5 | 61.5 |
2000 | 40th | 0.5% | ~0.5 | 61.5 |
Frequency (Hz) | Applied Voltage | Test | A | P | |||
---|---|---|---|---|---|---|---|
% | (V) | Δφ (Rad) | ε (%) | Δφ (Rad) | ε (%) | ||
150 | ~0.5 | 61.5 | Prop | 1.9606·10−2 | −0.8090 | −3.89·10−3 | 0.560 |
Ref | 1.9239·10−2 | −0.6483 | −4.29·10−3 | 0.566 | |||
1.0 | 120 | Ref | 1.9218·10−2 | −0.6928 | −4.34·10−3 | 0.565 | |
2.1 | 250 | Ref | 1.91993·10−2 | −0.71793 | −4.308·10−3 | 0.5625 | |
4.3 | 500 | Ref | 1.92005·10−2 | −0.72828 | −4.288·10−3 | 0.5606 | |
5.0 | 580 | Ref | 1.91899·10−2 | −0.72936 | −4.290·10−3 | 0.5603 | |
250 | ~0.5 | 61.5 | Prop | 1.142·10−2 | −0.869 | −6.82·10−3 | 0.43 |
Ref | 1.066·10−2 | −0.700 | −7.6·10−3 | 0.461 | |||
1.0 | 120 | Ref | 1.064·10−2 | −0.749 | −7.65·10−3 | 0.460 | |
2.1 | 250 | Ref | 1.0618·10−2 | −0.7755 | −7.61·10−3 | 0.455 | |
4.3 | 500 | Ref | 1.0635·10−2 | −0.7866 | −7.59·10−3 | 0.453 | |
5.0 | 580 | Ref | 1.0614·10−2 | −0.7876 | −7.61·10−3 | 0.453 | |
6.0 | 695 | Ref | 1.0591·10−2 | −0.7878 | −7.622·10−3 | 0.4542 | |
350 | ~0.5 | 61.5 | Main | 8.063·10−3 | −0.9120 | −9.39·10−3 | 0.382 |
Ref | 6.96·10−3 | −0.745 | −1.038·10−2 | 0.389 | |||
1.0 | 120 | Ref | 6.942·10−3 | −0.7908 | −1.039·10−2 | 0.387 | |
2.1 | 250 | Ref | 6.924·10−3 | −0.8162 | −1.0397·10−2 | 0.3855 | |
4.3 | 500 | Ref | 6.933·10−3 | −0.8279 | −1.0362·10−2 | 0.3843 | |
5.0 | 580 | Ref | 6.902·10−3 | −0.8289 | −1.0388·10−2 | 0.3848 | |
450 | ~0.5 | 61.5 | Main | 6.360·10−3 | −0.9458 | −1.170·10−2 | 0.323 |
Ref | 4.879·10−3 | −0.7767 | −1.299·10−2 | 0.334 | |||
1.0 | 120 | Ref | 4.854·10−3 | −0.8233 | −1.298·10−2 | 0.334 | |
1.5 | 175 | Ref | 4.824·10−3 | −0.8385 | −1.3029·10−2 | 0.3358 | |
750 | ~0.5 | 61.5 | Main | 4.472·10−3 | −1.0168 | −1.813·10−2 | 0.214 |
Ref | 1.774·10−3 | −0.8501 | −2.062·10−2 | 0.216 | |||
1000 | ~0.5 | 61.5 | Main | 4.21·10−3 | −1.061 | −2.34·10−2 | 0.17 |
Ref | 4.6·10−4 | −0.952 | −2.64·10−2 | 0.13 | |||
1250 | ~0.5 | 61.5 | Main | 4.44·10−3 | −1.091 | −2.85·10−2 | 0.11 |
Ref | −2.1·10−4 | −0.987 | −3.27·10−2 | 8·10−2 | |||
1.0 | 120 | Ref | −1.2·10−4 | −1.006 | −3.29·10−2 | 0.1 | |
1.5 | 175 | Ref | −2.5·10−4 | −1.013 | −3.287·10−2 | 8.5·10−2 | |
1600 | ~0.5 | 61.5 | Main | 5.03·10−3 | −1.135 | −3.618·10−2 | 3.5·10−2 |
Ref | −7.2·10−4 | −1.022 | −4.126·10−2 | 6·10−3 | |||
2000 | ~0.5 | 61.5 | Main | 6.07·10−3 | −1.163 | −4.47·10−2 | −3·10−2 |
Ref | −1.22·10−3 | −1.055 | −5.07·10−2 | −6·10−2 |
Frequency (Hz) | Applied Voltage | A | P | |||
---|---|---|---|---|---|---|
% | (V) | ΔΦ (Rad) | Δε (%) | ΔΦ (Rad) | Δε (%) | |
150 | ~0.5 | 61.5 | 3.7·10−4 | −0.16 | 4.0·10−4 | −6.8·10−3 |
5.0 | 580 | 4.2·10−4 | −8.0·10−2 | 4.0·10−4 | −6.4·10−4 | |
250 | ~0.5 | 61.5 | 7.6·10−4 | −0.17 | 7.4·10−4 | −2.9·10−2 |
6.0 | 695 | 8.3·10−4 | −8.1·10−2 | 8.0·10−4 | −2.2·10−2 | |
350 | ~0.5 | 61.5 | 1.1·10−3 | −0.17 | 9.9·10−4 | −7.0·10−3 |
5.0 | 580 | 1.2·10−3 | −8.3·10−2 | 1.0·10−3 | −3.1·10−3 | |
450 | ~0.5 | 61.5 | 1.5·10−3 | −0.17 | 1.3·10−3 | −1.1·10−2 |
1.5 | 175 | 1.5·10−3 | −0.11 | 1.3·10−3 | −1.3·10−2 | |
750 | ~0.5 | 61.5 | 2.7·10−3 | −0.17 | 2.5·10−3 | −1.8·10−3 |
1000 | ~0.5 | 61.5 | 3.7·10−3 | −0.11 | 3.0·10−3 | 3.5·10−2 |
1250 | ~0.5 | 61.5 | 4.6·10−3 | −0.10 | 4.2·10−3 | 3.4·10−2 |
1.5 | 175 | 4.7·10−3 | −7.8·10−2 | 4.4·10−3 | 2.4·10−2 | |
1600 | ~0.5 | 61.5 | 5.7·10−3 | −0.11 | 5.1·10−3 | 2.9·10−2 |
2000 | ~0.5 | 61.5 | 7.3·10−3 | −0.11 | 6.1·10−3 | 3.3·10−2 |
Accuracy Class | Ratio Error at Harmonics (±%) | Phase Error at Harmonics (±mrad) | ||||||||
---|---|---|---|---|---|---|---|---|---|---|
2nd to 4th | 5th and 6th | 7th to 9th | 10th to 13th | Above 13th | 2nd to 4th | 5th and 6th | 7th to 9th | 10th to 13th | Above 13th | |
0.5 | 5 | 10 | 20 | 20 | +20, −100 | 87.2 | 174.5 | 349.1 | 349.1 | – |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
Mingotti, A.; Betti, C.; Peretto, L.; Tinarelli, R. Simplified and Low-Cost Characterization of Medium-Voltage Low-Power Voltage Transformers in the Power Quality Frequency Range. Sensors 2022, 22, 2274. https://doi.org/10.3390/s22062274
Mingotti A, Betti C, Peretto L, Tinarelli R. Simplified and Low-Cost Characterization of Medium-Voltage Low-Power Voltage Transformers in the Power Quality Frequency Range. Sensors. 2022; 22(6):2274. https://doi.org/10.3390/s22062274
Chicago/Turabian StyleMingotti, Alessandro, Christian Betti, Lorenzo Peretto, and Roberto Tinarelli. 2022. "Simplified and Low-Cost Characterization of Medium-Voltage Low-Power Voltage Transformers in the Power Quality Frequency Range" Sensors 22, no. 6: 2274. https://doi.org/10.3390/s22062274
APA StyleMingotti, A., Betti, C., Peretto, L., & Tinarelli, R. (2022). Simplified and Low-Cost Characterization of Medium-Voltage Low-Power Voltage Transformers in the Power Quality Frequency Range. Sensors, 22(6), 2274. https://doi.org/10.3390/s22062274