In
Section 5.1, we discuss (and show with specific photos) the different elements of the experimental set-up: the phantom used to “wear” the wetsuit sample, the heater for steady state operation, the external water tank, the temperature sensors, and the data acquisition system. In all cases, the cylindrical phantom for measurement, designed and developed following the procedures in [
16], was closed by the wetsuit cylindrical sleeve and completely filled with water prior to its immersion inside the water tank. Two calibrated NTC sensors measured the temperatures
T2 and
T1 of the water inside and outside the phantom, respectively, and hence on the two sides of the wetsuit surface. In all of the measurements, the neoprene wetsuit sample used had a nominal specific thermal resistance per unit area
RT,A,NOM = 0.1 m
2·K/W, as measured by an independent metrological laboratory for other wetsuits from the same producer (same neoprene material and sample thickness of 5 mm at atmospheric pressure). To operate without using a powerful water chiller, which just recently was made available in our laboratory, the external tank’s water temperature was naturally regulated by the ambient temperature and stabilized by the large water volume
VT > 300 L, resulting in
T1 ≈ 20–22 °C, as it slightly changed over time.
5.1. Experimental Apparatus
In order to measure the wetsuit’s specific thermal resistivity, a custom phantom was realized in agreement with [
16] and following the schematic drawing in
Figure 1. The material used for construction of the upper and lower corks and for the corresponding four vertical rod spacers was plexiglass (thermal conductivity ~0.2 W/(K·m), specific heat capacity ~1500 J/(K·kg), and mass density ~1.2 kg/dm
3). Two photos of the phantom, “naked” and wearing the wetsuit’s cylindrical sleeve sample, are shown in
Figure 2. Each of the two corks was a Plexiglas cylinder with a diameter of 150 mm and height of 64 mm, while each of the four cylindrical rod spacers had a diameter of 25 mm and height of 210 mm. Note that with such a geometry, the thermal resistance of each cork along the phantom’s axis could be roughly estimated to be (5 m·K/W) × (0.064 m)/(π·(0.15/2)
2 m
2) ≅ 18 K/W. With this value for each cork, we obtained a total cork thermal resistance (parallel combination) significantly higher than the phantom’s thermal resistances measured in the experiments (
RT,C ~ 18/2 = 9 K/W >>
RT,P ~ 1 K/W), and thus, the assumption of
RT,C >>
RT,W could be considered valid. The electric heater was a cylindrical stainless steel rod with a diameter of 12 mm and a height of 100 mm. The heater was commercial with a high-power electric resistance of
RH ≅ 3.2 Ω, which could be driven by low current and voltage values (
U ≤ 30 V and
I ≤ 10 A, respectively) provided by a commercial power supply (RUZIZAO DC power supply, 30 V-10 A). This resulted in measured thermal powers
P =
U·
I up to a maximum value of 280 W, which were applicable to the phantom.
After the phantom was wearing the wetsuit sample, as one can see from
Figure 2b, by tightening the wetsuit to the corks by means of cylindrical mechanical clamps, the inner volume of the phantom was filled with water from an upper central hole (with a diameter of 20 mm), which was then hermetically sealed by a stainless steel stopper screw. We note that the inner wetsuit sleeve diameter (
DW = ~137 mm) for the available wetsuit samples was significantly smaller than the phantom’s cork diameter. This can also be seen in
Figure 2b, where the wetsuit was stretched to the larger 150 mm diameter in correspondence to the corks, but it rapidly bent back to the original wetsuit diameter and hence to some ~137 mm along the phantom’s inner lateral wall (see
Figure 2b and note the wetsuit shrinking from each cork to the center of the lateral wetsuit area), resulting in a wetsuit area crossed by the heat flux which was significantly smaller than the nominal area (
A = 0.1 m
2). The precise volume of water by which the phantom was filled without leaving air bubbles, as observable from the transparent top cork, was measured at the end of the wetsuit test by emptying the phantom’s inner water into a metrically graduated container.
Figure 3a shows a photo of the phantom immersed in a smaller and transparent water tank for a preliminary test, and
Figure 3b shows a photo of the large water tank (
VT ≅ 320 L~0.3 m
3).
5.2. Temperature Sensors and Data Acquisition System
The temperature sensors used to monitor the inner water temperature
T2 (inside the phantom) and the outer water temperature
T1 (inside the water tank) were commercial NTC “10k” thermistors (nominal values
R0 = 10 kΩ at 25 °C and
β25/100 = 3950 K). The thermistor resistance value was read with a precision digital multimeter (DMM; Hewlett-Packard HP3441A, 6½ digits) before exporting the data to a PC via the GPIB digital output using the LabVIEW
TM program. Numerical temperature values were read, recorded, and plotted using the PC with a sampling rate of 1 Sa/s. The
R vs.
T characteristic of the thermistors used in the measurements was previously calibrated by a precision Pt-100 digital thermometer (DOSTMAN Electronics, P655-LOG), with a rated accuracy of ±0.03 °C over the temperature range from −100 °C to +150 °C. The thermistors were calibrated by obtaining the specific values of
R0 and
β for each measurement chain, including also the effect of the DMM, including the wires connected to the DDM, and the DMM numerical reading using the data acquisition system (DAS). In the temperature sensor calibration, we measured the resistance values, as read by the DMM and DAS, in correspondence to different temperature values also measured by the reference platinum thermometer. A set of 20 calibration points was used in the temperature range of 10–50 °C. With a nonlinear fitting of the data to the thermistor equation
with
T0 = 25 °C over the 20 calibration points in the useful temperature range of 10–50 °C, we evaluated the specific values *
R0 and *
β for the calibrated thermistor reading (including the DMM and DAS). These two calibrated parameters were then used for evaluating the measured temperature
T, obtained by inverting the thermistor equation (Equation (12)), once the electrical resistance
R was measured in the experiments by the DMM and DAS. In this way, we estimated for our digital and thermistor-based temperature measurement system an uncertainty
u(
T) < 0.05 °C in the whole working range of 10–50 °C. Two photos showing the temperature acquisition system are given in
Figure 4.
With the described experimental apparatus, we could measure the inner and outer temperatures of the phantom (T2 and T1, respectively) while working in steady state or transient conditions. In all of the following experiments, the phantom was closed by the same 5 mm-thick neoprene wetsuit sample, and the tests were carried out at ambient pressure (pATM ≅ 1 bar ≅ 105 Pa) since a stainless steel water tank for performing measurements at a higher water pressure (up to six absolute bars) was not yet available.
5.3. Results When Operating in Steady State Condition
For the measurements in the steady state condition, the water inside the phantom was heated by the electric heater. By manually regulating the voltage (
U) and current (
I) of the power supply driving the electrical resistance, we empirically found that an input power value of
P = ~30 W for the wetsuit with
RT,A = ~0.1 m
2·K/W provided a temperature difference Δ
T =
T2 −
T1 ≈ 27 °C (thus in the range of 25–30 °C, as prescribed in [
16], which can be a typical temperature difference between the human body and the external cold water in recreational scuba diving immersions in cold waters). This experimental condition corresponded to a balance between the ingoing electric power
P, which is input in a controlled way into the phantom, and the outgoing thermal power
, leaving the phantom. Driven by the temperature difference Δ
T, this thermal power is leaving the phantom toward the external water tank, through the wetsuit’s thermal resistance. In the experiment, when the electrical power level was adjusted so that the temperature difference remained stable within ±0.1 °C for more than 15 min, both temperatures
T2 and
T1 were recorded.
Figure 5 shows the results for such steady state measurements obtained with an electric power
P = 26.08 W, as indicated on the power supply display. We can see from
Figure 5a that a rather long initial settling time was needed (Δ
t = ~10,000 s ≈ 3 h), where
T2 adapted to the manually set electric power value, if one wanted to reach highly stable temperature conditions. After this settling time, both temperatures remained stable—and hence their difference Δ
T did as well—over an extremely long observation time of ~70,000 s ≈ 20 h. The working regime was practically stationary, and all of the system’s parameters remained substantially constant:
P,
T2,
T1, and Δ
T. In particular,
Figure 5b shows an expanded view of the temperature difference Δ
T time diagram, where we can observe temperature stability as good as ±0.05 °C over a time interval of 15 min (900 s), as required by the guidelines in [
16].
From the measurement data in
Figure 5, and in particular from the values of the stable temperature difference Δ
T = 27.386 °C (average value over the time interval of 15 min, containing 900 measurement points), upon using the specific value of the electric input power
P = 26.08 W, one could find the whole-body thermal resistance of the phantom
RT = Δ
T/
P = 1.050 K/W. Then, with the nominal wetsuit surface
A = 0.1 m
2, as ideally obtained by the cork circumference (π·150 mm) multiplied by the inner phantom’s height (210 mm), one could find through Equation (6) the measured specific resistance per unit area
RT,A,ss =
A·(Δ
T/
P) = 0.104 m
2·K/W, measured in the steady state condition.
5.4. Results When Operating in Transient Condition
For the measurements in the transient condition, a water volume larger than the tank’s inner volume was preheated to some higher temperature (TH in the range of 40–50 °C) before pouring it into the phantom and sealing the phantom’s upper cork. In this case, there was no need for an electric heater inside the phantom nor precise regulation of its voltage, current, or power values. A few minutes after sealing the phantom, we waited for its thermalization, and the filled warm phantom was immersed inside the water tank which was kept at room temperature (T1 ≈ 20 °C), which was approximately constant over the whole measurement time. At this point, a temperature transient naturally occurred, with the inner tank’s water temperature T2(t) cooling down from near TH (the initial value is not important) down to the regime value T1. Both temperatures T2 and T1 were measured and recorded over time during the cooling process, which is a decreasing exponential given by Equation (10). Since the temperature T1 remained naturally constant over the measurement, then its value could be measured merely once or twice (e.g., in the beginning and at the end of the transient measurement to check its stability), and thus, only the data points T2(t) had to be recorded by the DAS with the usual sampling period of 1 Sa/s. In fact, for the subsequent fitting procedure used to extract the time constant of the exponential transient, just enough data points of T2(t) were needed, while both values of the initial and final temperatures of the transient were not important, and they could even be rather different from one experiment to the other. The lack of a need for precise values for the temperatures or adjusting to a specific value for the electric power (keeping the temperature difference well monitored) made the transient method far easier and faster than the stationary one.
Figure 6 shows the results obtained for the transient regime measurement, with the phantom starting from an initial inner temperature
T2,I and spontaneously cooling down to
T1 over several hours. In particular,
Figure 6a shows the recorded temporal evolution of both temperatures
T2 and
T1 over an extremely long observation time of ~80,000 s ≈ 20 h. This long observation time was indeed recorded only to provide a complete experimental analysis of the temperature’s transient behavior, but it is not needed for ordinary measurements, which can be performed in a much shorter time.
Figure 6b,c shows the graphs of the temperature difference evolution Δ
T(
t) =
T2 −
T1 corresponding to the experimental points in
Figure 6a together with the exponential fitting curve used to extract the time constant
τ and the variable experimental parameters Δ
TI and Δ
TF. With reference to Equation (10), Δ
TI = [
T2,I −
T2,F] is the initial temperature difference (the transient variable “step”), and Δ
TF =
T2,F −
T1 is the final temperature difference (theoretically equal to zero but potentially different from zero due to temperature measurement offsets) at the end of the temperature transient. From the black trace of the exponential fitting curves in
Figure 6b,c, we obtained a time constant value
τ = 12,039.5 s ≅ 12,040 s, evaluated as the average value of
τb = 12,028 s from
Figure 6b and
τc = 12,051 s from
Figure 6c.
After measuring the time constant of the phantom natural cooling transient, we could evaluate the phantom’s thermal capacitance CT and hence the phantom’s thermal resistance RT using Equation (5) to finally obtain the wetsuit’s specific thermal resistance per unit area RT,A following Equation (6) (i.e., simply multiplying by the phantom’s area A). The volume of the water contained in the phantom when immersed in the water tank was measured both during the phantom’s filling procedure and during the phantom’s emptying procedure, which resulted in VW = 3.01 L. The volume of the Plexiglas rods within the phantom was calculated using their geometrical dimensions, resulting in VP = 0.41 L, while the volume occupied by the heater was VH = 0.011 L.
Considering the specific heat capacity of water cT,W = 4180 J/kg·K, we could evaluate the heat capacity of the water in the phantom to be CT,W = mWVWcT,W = ρWVWcT,W = 12,582 J/K, where mW is the water mass and ρW ≅ 1 kg/L is the water mass density. Similarly, the heater’s thermal capacity (since, for simplicity of operation, the heater was not removed from the phantom during the transient regime measurements) was CT,H = mHVHcT,H = ρHVHcT,H = 40 J/K, where mH is the heater mass, ρH ≅ 7.7 kg/dm3 is the heater’s mass density, and cT,W = 460 J/kg·K is stainless steel’s specific heat capacity. Neglecting the small heat capacitance value of the heater, and also neglecting the heat exchanged by the Plexiglas rods within the phantom (since their thermal conductivity is much poorer than water’s thermal conductivity), we could estimate a total phantom thermal capacity approximately equal to the one of the contained water (CT ≅ CT,W). By using this thermal capacitance value together with the measured time constant (for example, τ = τ20,000 = 12,084 s), we found the thermal resistance of the phantom (whole body resistance) to be RT = τ/CT = 0.958 K/W. From this value, by once again multiplying it by the nominal wetsuit area A = 0.1 m2, one could finally obtain the measured specific resistance per unit area RT,A,tr = A·RT = Aτ/CT = 0.095 K·m2/W, as obtained by the transient condition measurement.
We note that in
Figure 6b,c, the initial fitting time is
tFIT,I = 1000. Instead, the final fitting times were
tFIT,F,b = 37,000 s for
Figure 6b and
tFIT,F,c = 3000 s for
Figure 6c. Thus,
Figure 6b refers to a data fitting procedure extended over as long as 10 h and 36,000 experimental points, while
Figure 6c refers to a much shorter data fitting procedure extended over only 2000 points (i.e., 2000 s). Through the exponential fitting procedure performed by the Excel
TM SOLVER function on the experimental data points of
Figure 6, we extracted the
τ parameters
τ36,000 = 12,028 s =
τb from the black trace exponential fitting curve in
Figure 6b and
τ2000 = 12,051 s =
τc from the black trace exponential fitting curve in
Figure 6c. The difference in the two time-constant values was negligible (0.2%), considering that the same percentage error on
τ would reflect an equal percentage error contribution to the indirectly measured thermal resistivity value, as one can see from Equation (11). For comparison, we note that extending the fitting procedure over other much different (and certainly long enough) time intervals also provided quite similar
τ values (e.g.,
τ10,000 = 12,048 s,
τ20,000 = 12,084 s, and
τ50,000 = 12,112 s). Instead, when working with shorter fitting time intervals (and less fitting points), the obtained
τ values were slightly underestimated. For example,
τ1000 = 11,953 s,
τ500 = 11,862 s, and finally
τ100 = 11,709 s with only 100 fitting points. We can conclude that fitting the exponential model of Equation (10) to a number of experimental points of the temperature transient to the order of 900–3600 points provided negligible fluctuations in the time-constant evaluation. These numbers of points correspond to time intervals from 15 min to 1 h for the temporal analysis of only the temperature
T2, which were quite practical times for the experiments and also for commercial tests on the wetsuits. As a comparison, we can highlight that with the steady state measurement in
Figure 5b (to be compared with the black fitting curve in
Figure 6c), the time needed for the measurement was ~10 h, compared with ~1 h for the transitory measurement.
We also finally note that by using this transient condition method, only a simple timing accuracy and reasonable linearity in the
T2 temperature measurement were needed. Any reasonable offsets, even one that varied day by day, in the only temperature (
T2) measurement system did not affect the result of
τ. After the NTC sensor calibration, both the offset and gain errors were made rather small, and their residual values resulted in negligible inaccuracy in the determination of the time constant
τ, since the temperature range of the
T2 measurement was limited (up to a maximum excursion of roughly 30 °C). Furthermore, the timing accuracy of the DAS was negligible with respect to the data point sampling time of 1 s and with measured values of
τ far larger than minutes if not hours. An uncertainty budget analysis, together with an intercomparison of the two methods, is performed in
Section 5.5.
5.5. Results Comparison and Uncertainty Budgets
As one can see from the results achieved in
Section 5.3 and
Section 5.4, with the stationary condition method and experiment, we found the specific thermal resistance of the wetsuit under testing (
RT,A,ss = 0.104 m
2·K/W), while with the transient condition method and experiment, we obtained the specific thermal resistance of the same wetsuit (
RT,A,tr = 0.095 m
2·K/W). The difference between the two measurements divided by their average value provided a relative error of 2.3%, and this was also the relative error between the two measured thermal resistance (
RT) values since in both methods, the same area
A was used for finding the resistance from the resistivity, and vice versa. Such a value for the relative error between the two measurement methods and the experimental setups is quite acceptable for these kinds of thermal resistance and resistivity measurements, since the thermal performance classes of wetsuits (see
Table 1) do change from one to the other by steps in the order of 17%, 18%, and 20% of their average value.
As previously highlighted, the steady state condition method is not simple to implement, requiring the availability and use of an electric heater and tedious manual adjustment of the electrical power used to achieve a useful and long-lasting ΔT stationary value. The stationary temperature difference value, as previously said, has to be in the limited range of 25–30 °C, and it strongly depends on the wetsuit’s thermal insulation, thus changing when the wetsuit is changed. The measurement can last from a minimum of about one hour if one already approximately knows the correct electrical power value to up to more than 10 h (depending on the time needed to manually adjust the value of the electric power). Most of this long adjustment procedure has to be repeated when the wetsuit sample is changed. Unless a custom electronic control loop is developed to adjust and regulate automatically the heater power P, the steady state measurement requires the constant presence and work of an operator for a rather long time. Furthermore, the simultaneous measurement of temperatures both inside (T2) and outside (T1) the phantom is needed, and it must be checked that the temperature difference ΔT = T2 − T1 remains constant to within ±0.1 °C over at least 15 min of observation time.
On the other side, the transitory state method is far simpler to implement, requiring just the automatic acquisition of one temperature transient T2(t), namely one than uses relatively easy numerical post-processing of the data acquired during the temperature transient, where T2 is naturally cooling down, without operator supervision from an initial (inessential) higher value toward the final value T1 (constant and, again, inessential) of the water in the outer water tank.
In terms of the measurement accuracy and uncertainty budget [
20], both measurement methods are indirect, and they rely on the measurements in Equations (9) and (11), respectively. When considering all the input and output uncertainties expressed in terms of relative uncertainties and assuming uncorrelated input variables, we can write the relative uncertainty of the measured thermal resistivity obtained by the two methods as follows:
Clearly, the uncertainty of the area of the wetsuit surface is responsible for the heat exchange process present in both expressions and with the same weight. At present, the wetsuit samples prepared for the experiments in the form of sewed cylindrical sleeves had a significantly smaller diameter (internal diameter DW = ~137 mm before putting the wetsuit onto the phantom) than the phantom’s cork diameter (DC = 150 mm). Thus, when the wetsuit was put on the phantom, the wetsuit surface was not the one of a simple cylinder with a circular base and constant diameter from the bottom cork up to the top cork. Instead, this cylindrical surface had a variable diameter, and in particular, its inner diameter started from a larger value DMAX = DC = 150 mm in proximity of the two corks and reduced itself to a significantly smaller value of DMIN ≅ DW ≅ 137 mm along most of the height of the phantom. The transition from the larger to the smaller wetsuit diameter was not constant all over the phantom’s height, but it happened mostly in the proximity of the corks such that at a few centimeters away from the corks, the wetsuit’s inner diameter was already at a value of ~137 mm. Due to this manufacturing defect of the available wetsuit, we worked with an uncorrected error to the order of −8% with respect to the nominal wetsuit sample area A = 0.1 m2, prescribed as the inner wetsuit heat exchange area. While waiting for better cut and sewed wetsuit sleeve samples, we can safely assume that the large area uncertainty ur(A) ≈ 8% for the wetsuit area.
Regarding the other input uncertainties of Equation (13), with our calibrated DAS measuring the two stationary temperatures
T2 and
T1, we could safely assume that each temperature was measured with an absolute uncertainty to the order of
u(
T) = ~0.2 °C. Now, considering the measurement of
T2 to be uncorrelated with the one for
T1 and a temperature difference to the order of Δ
T ≅ 27.5 °C, we obtained [
20]
u(Δ
T) =
u(
T) ≅ 0.28 °C, and hence we had a relative uncertainty of the measurement of the temperature difference of
ur(Δ
T) =
u(Δ
T)/Δ
T ≅ 1%. Regarding the measurements of the voltage
U and current
I, they both were simply measured by the power meter display, with each having an estimated uncertainty of
ur(
U) ≅
ur(
I) = ~2%. Of course, each of these two electrical quantity measurements could be significantly improved in terms of accuracy by using a good quality and already available DMM. However, in this case, when reducing the electrical variables’ uncertainties below a few percent, then the stability and repeatability of the moderate-quality electrical power supply should also be evaluated and taken into account. In the end, by combining the different uncertainties of the inputs, treated as uncorrelated quantities, one could find the uncertainty budget for the steady state operation measurement to be
ur(
RT,A,ss) ≅ 8.5%, with a dominant contribution from the area uncertainty.
Regarding the input uncertainties of Equation (14), we could estimate through the analysis of repeated τ measurements obtained upon changing the length of the fitting time interval a relative uncertainty of ur(τ) ≈ 1%, while the thermal capacitance uncertainty could roughly be estimated to be ur(CT) ≈ 4%. The resulting uncertainty budget, in the case of the transient operation, yielded ur(R,T,tr) ≅ 9.0%, again significantly determined by an area uncertainty of ur(A) ≅ 8%.
In the discussed preliminary tests on a commercial scuba diving wetsuit sample (thickness: 5 mm), the two measurement methods provided RT,A,ss = 0.104(9) m2·K/W and RT,A,tr = 0.0948(9) m2·K/W, with rather few degrees of freedom for both estimates of the standard uncertainties, henceforth expressed with one digit, as indicated in round brackets after the numerical values of RT,A. With such relatively large uncertainty intervals (≈9% of the measured value), the two measurements were highly compatible with each other. For now, the uncertainty budget of both measurement methods is strongly dominated by the wetsuit’s area uncertainty, which affects both measurements systematically in the same direction. As a next step, it is foreseen that the same measurements will be repeated and the uncertainty budgets re-evaluated when new wetsuit samples become available, with an inner diameter of 150 mm and different thicknesses. Anyway, the methodology and experimental set-up proposed to measure RT,A,ss and RT,A,tr, as well as the respective uncertainty budgets remain valid.
The results achieved were satisfactory, especially in regard to the main scope of this work (i.e., proving that a different measurement method, namely the transient state measurement, is also available and can be used as a valid alternative to the stationary state measurement method proposed in [
16]). With both methods, we could measure a 5 mm neoprene wetsuit’s thermal resistivity in a compatible way with what was previously described in the literature, as foam neoprene of this thickness has been shown to have a thermal resistivity of approximately 0.090–0.100 m
2·K/W [
18,
21].