MEK-9100 Service Training: Disassembly and Unit Description

Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1of 70

MEK-9100 Service Training

Disassembly
And Unit
Description
1
MEK-9100   Service Training
      
Required Steps Before Unit Replacement

(1) Run Clean Protein


The purpose of protein removal cleaning is to
remove all protein contamination and prevent
infection.
Some parts of the analyzer may be contaminated
with blood.
Run [Clean Protein] to protect yourself and ensure
your safety.
(2) Drain All

(3) Power OFF (Reset + Power OFF)

(4) Main Power OFF


2
Exterior [Front]

3
Exterior [Right Side]

4
Exterior [Left Side]

5
Exterior [Rear]

6
Composition Chart

7
MEK-9100   Service Training
Location of Units ①


⑰⑱⑲ Inside

①   PC-910W DATA PROCESSING UNIT ⑭
②   PV-910W FRONT PANEL UNIT ④
③   AUTO LOADER UNIT
④   MS-910W   SAMPLER UNIT
⑤   MS-911W   OPEN AIR UNIT
⑥   MS-912W OPEN LOADER UNIT
⑦   MP-911W ISO PUMP UNIT ⑧
⑧   MP-912W SAMPLE/RBC PUMP UNIT
⑨   MP-913W IWBC/OWBC PUMP UNIT
⑩   ZY-911W CUP HEATER UNIT ⑮
⑪   ZY-910W TANK HEATER UNIT
⑫   MH-910W   HGB MEASURING UNIT ⑫ ⑨
⑬   MP-910W   PNEUMATIC UNIT
⑭   MO-910W   LASER OPTICAL UNIT ⑥
⑮   MC-910W   CBC MEASURING UNIT ⑦
⑯   UT-7287 PRESSURE SENSOR BD
⑰   JQ-910W   ISO CHAMBER UNIT
⑱   JQ-911W   WASTE CHAMBER 1 UNIT ⑩
⑲   JQ-912W   WASTE CHAMBER 2 UNIT ⑪

⑤ 8
MEK-9100   Service Training
       Disassembly   
MIXING cover protects the mixing unit and barcode part
Removing the MIXING cover
(1) Loosen the screw using a coin or a large flat-
blade screwdriver.
   (Screw will remain on the cover)

Screw

9
MEK-9100   Service Training
       Disassembly   
Removing the MIXING cover
(2) Pull out the the cover while pressing the
plastic part on the left side.
There are hooks on the bottom side of the cover; pull the cover out and then lift it up.

Plastic part on
the side

10
MEK-9100   Service Training
       Disassembly
Removing the FRONT cover
  (1) Remove the screws at 2 locations.
    

Screw: BH (with washer) 3×8

Screw: BH (no washer) 3x8

11
MEK-9100   Service Training
       Disassembly
Removing the FRONT cover
Shift
   (2) Shift the cover towards the right

of the analyzer until it is close to

the plate.
   (3) Pull the cover slightly forward.
    

Pull the cover forward


12
MEK-9100   Service Training
       Disassembly
Removing the RIGHT cover
   (1) Remove the rubber cap and then remove the screw
inside it.

    

Screw: BH (no washer) 3x8

13
MEK-9100   Service Training
       Disassembly
Removing the RIGHT cover
   (2) Remove 9 screws.
    Reagent case-side:
Screw: BH (with washer) 3×8 7 pcs
   Open the reagent case:
Screw: BH (with washer) 3×8 2 pcs

14
MEK-9100   Service Training
       Disassembly
Removing the RIGHT cover
   (3) Shift the cover in the direction of the arrow
      shown in the illustration below to release the hooks.

15
MEK-9100   Service Training
       Disassembly
Removing the LEFT cover
   Remove 8 screws to remove the left cover.
    Screw: BH (with washer) 3×8 8 pcs

When removing the screws, take


care not to let the cover fall.
16
MEK-9100   Service Training
       Disassembly
Removing the TOP cover
   Remove 4 screws.
      Screw: BH (with washer) 3×8 4 pcs

Remove the top cover by shifting it in


the directions of the arrows (up then sideways)
as shown in the illustration below.

17
MEK-9100   Service Training
       Disassembly
Removing the SD access cover
   Remove 3 screws to remove the cover.
      Screw: BH (with washer) 3×8 3 pcs

18
MEK-9100   Service Training
       Disassembly
Removing the REAR panel
   Remove 14 screws and loosen 3 screws.
   Screw: BH (with washer) 3×8 17 pcs

Loosen 19
MEK-9100   Service Training
       Disassembly
Removing the AUTOLOADER cover
   Remove 3 screws to remove the cover.
      Screw: BH (no washer) 3×8 3 pcs

20
MEK-9100   Service Training
       Disassembly
PCBs on the BD HOLD PLATE
・ UT-7284   DRIVER BD
・ UT-7283   ANALOG BD
・ UT-7282   MAIN   BD

21
MEK-9100   Service Training
       Disassembly
Remove the BD HOLD PLATE.
  (1) Disconnect all the connectors from the MAIN BD and ANALOG BD.

  (2) Remove 6 screws that are securing the plate.


Screw: PS 3×8 6 pcs
  (3) Release the black latches at 2 locations.

22
MEK-9100   Service Training
       Disassembly
Remove the BD HOLD PLATE.
  (4) Shift the BD HOLD PLATE forward.
  (5) Disconnect all the connectors from the DRIVER BD.

There are 19 cables that connect to the DRIVER BD.


The connectors for cables are all on the interior side of the PCB except for
the power cable. 23
MEK-9100   Service Training
       Disassembly
Remove the BD HOLD PLATE.
  (6) Remove the stopper cables from the housing side.
      Screw: PS 3×8 2 pcs

Be careful not to drop the screws.

24
MEK-9100   Service Training
       Disassembly
DRIVER BD (UT-7284)
  Remove 9 screws from
the DRIVER BD.
    Screw: PS 3×8 6 pcs
    Screw: PS 2.5×6 3 pcs

Use a precision screwdriver when


removing PS 2.5×6 screws.

Pull the DRIVER BD straight up as it is connected


directly to the MAIN BD via the
connector.

25
UT-7284 - DRIVER BD
[Functions]
 Controls 29 valves, 8 stepping motors, and
1 compressor.
 Relays +5V and +24V power.
 Monitor and transmit signals to MAIN BD -
24 position sensor statuses, 4 float sensor
statuses, sensor statuses from
CONNECTION BD and sensor statuses
from SAMPLER SENSOR BD.
 Actuator control signals and motor driver
IC reset signals are supplied from MAIN
BD. Control signals are sent to the
respective driver circuits via a buffer.

26
MEK-9100   Service Training
       Disassembly
ANALOG BD (UT-7283)
  Remove 7 screws from the ANALOG BD.
      Screw: PS 3×8 7 pcs

Pull the ANALOG BD straight up as it is connected directly to the


MAIN BD via the connector.

* The ANALOG BD can be removed


without taking off the BD HOLD PLATE.
27
UT-7283 - ANALOG BD
[Functions]
 PCB that is responsible for loading
analog data.
 Data from CBC Measuring Unit, Laser
Optical Unit, HGB Measuring Unit and
Cup Heater Unit are loaded,
converted to digital data and
transmitted to MAIN BD.
 Amplifies the clog removal signals
from MAIN BD and applies the clog
removal voltage to CBC Measuring
Unit.

28
MEK-9100   Service Training
       Disassembly
MAIN BD (UT-7282)
  Remove the screws from the MAIN BD.
     Screw: PS 3×8 5 pcs
      Nut + Washer: M3 3 pcs  
     Spacer bolt: 2 pcs

MAIN BD cannot be removed before removing the ANALOG BD and DRIVER BD.29
UT-7282 - MAIN BD
[Functions]
 PCB that is responsible for controlling actuators, display and
indicators. Process measurement and external input/output signals.
 Motor control, valve control and sensor status monitoring, which
are responsible for unit operation via DRIVER BD.
 Heater control and temperature monitoring via PRESSURE
SENSOR BD.
 Autoloader control
 Performs autoloader control via serial communication. Inputs
status signals and controls output signals for power control,
etc.
 Interface with GUI control unit
 LCD and touch panel controls are performed by DATA
PROCESSING UNIT.
 The communication between this PCB and DATA
PROCESSING UNIT are achieved via Ethernet (10/100base-
TX).
 LED and key input monitoring is performed via the Indicator BD
and KEY BD.
 Analog control
 Controls the A/D converter implemented in ANALOG BD and
obtains the A/D conversion value for measurement data.
30
MEK-9100   Service Training
       Disassembly
CUSTOM POWER UNIT 2H213W
  (1) Remove 6 screws from
the cover.
     Screw: PS 3×8 6pcs
  (2) Remove the CUSTOM
POWER SUPPLY.
     Screw: PS 3×8 3pcs
     Spacer bolt: 6 pcs
     Connector: 7 pcs
     

31
MEK-9100   Service Training
       Disassembly
UT-7288    BACK PANEL BD
  (1) Remove the cables.
      USB:     2 pcs
      D-SUB:    1 pc
      LAN:     1 pc

Disconnect the USB cables from


the BACK PANEL BD side.

Disconnect the D-SUB and LAN ②


cables from the PC-910W side.
 
* The D-SUB cable is screwed on.
Use a precision screwdriver to remove the screws.
Be careful not to lose the screws as it is unique to the D-SUB cable.

32
MEK-9100   Service Training
       Disassembly
BACK PANEL BD
  (2) Loosen/Remove the screws.
    Screw: To be loosened PSW 3×8 2 pcs
    Screw: To be removed PS 3×8 2 pcs
(3) Slide the whole plate off with the
BACK PANEL BD attached.
(4) Disconnect 2 remaining cables.
      LAN:     1 pc
      D-SUB:   1 pc

33
UT-7288 - BACK PANEL BD
[Functions]
 PCB that is responsible for relaying signal input to and output from DATA
PROCESSING UNIT.
 USB (2ch)
 Ethernet (1ch)
 RS-232 (1ch)
 Connects to external connecting devices, including barcode reader,
printer, hub and other network devices.

34
MEK-9100   Service Training
       Disassembly
CONNECTION BD
  Remove the CONNECTION BD.
    Connector: 1 pc
    Screw: PS 3x8 2 pcs

35
UT-7292 - CONNECTION BD
[Functions]
 PCB that is responsible for transmitting
the ON/OFF signals from the waste
sensor (optional product) to DRIVER BD.

36
MEK-9100   Service Training
       Disassembly
PC-910W   DATA PROCESSING UNIT
  (1) Loosen 4 screws, and shift the ACCESS COVER in
the direction of the arrow to remove it.
     Screw: PSW 3×8 4 pcs

37
MEK-9100   Service Training
       Disassembly
PC-910W   DATA PROCESSING UNIT Press
(2) Disconnect 3 connectors
  * The 2 white connectors have hooks.
    Disconnect the connectors while pressing on the hooks.

Press

38
MEK-9100   Service Training
       Disassembly
PC-910W   DATA PROCESSING UNIT
  (3) Disconnect 2 relay connectors.

  (4) Disconnect 3 remaining cables.


LAN: 1 pc
USB: 2 pcs

39
MEK-9100   Service Training
       Disassembly
PC-910W   DATA PROCESSING UNIT
  (5) Remove 4 spacer bolts.
   

* Be careful not to catch or pinch


any cable when lifting up the unit.
40
PC-910W - DATA PROCESSING UNIT
[Functions]
 LCD screen control
 Touch panel control
 Measurement data analysis
 Input/Output with external interfaces
 Measurement data storage

[Specifications and External Interfaces]


 CPU : N2600 (1.6GHz) Intel
 Memory : 2GB PC3-8500 (DDR3-1066)
 LVDS : 800600 18bit
 SATA : Serial ATA 1.0a/II
 LAN : 10/100Base-T interface 2ch
 USB : USB2.0 2ch
 Serial Port : RS232C 3ch
 Battery : CR2032 FDK
 CPU cooler : DLR-4C11B-N2-GP CoolerMaster
 SSD capacity : 8 GB pre-installed with Windows7RT

41
MEK-9100   Service Training
       Disassembly
PV-910W   FRONT PANEL UNIT
  (1) Disconnect the relay
connector.
  (2) Remove 4 screws.
Screw: PSW (black) 3×16 3 pcs
Screw: BH (no washer) 3×8 1 pc

Remove the relay


connector

Hook

BH no washer 3×8
42
MEK-9100   Service Training
       Disassembly
PV-910W   FRONT PANEL UNIT
(3) Remove 4 screws.
  Screw: PSW (black) 3×8 4 pcs

(4) Remove the unit.


    

43
PV-910W - FRONT PANEL UNIT
[Functions]
 Displays data, emits buzzer sounds, allows switch
input,
LED indicators lights, etc.

44
MEK-9100   Service Training
       Disassembly
AUTOLOADER  
  (1) Remove the plate.
    Remove 2 screws from the front panel.
       Screw: PSW 3×8 2 pc  
Loosen 2 screws on the side of the analyzer, slide the plate forward and out.
Screw: PSW 3×16 2 pcs

Remove these Loosen these


2 screws 2 screws

Plate

45
MEK-9100   Service Training
       Disassembly
AUTOLOADER  
  (2) Disconnect 2 cables.
Open 2 clamps.
Disconnect 2 connectors.

46
MEK-9100   Service Training
       Disassembly
AUTOLOADER  
  (3) Remove the screws.
    Remove 8 screws that are securing the AUTOLOADER.
      Screw: PS 4×8 7 pcs
      Screw: PSW 4×8 1 pc
  (4) Remove the AUTOLOADER.
    Pull the unit forward.

47
AUTOLOADER
[Functions]
 Moves the rack to its designated position.
 Detects whether there are sample tubes in the
rack.
 Using the built in barcode reader, read the
barcodes on the sample tubes and rack.
 Grips the sample tubes to mix blood samples
(turns the tube upside down).

48
MEK-9100   Service Training
       Disassembly
MS-910W   SAMPLER UNIT  
  Remove MS-910W and MS-911W at the same time.

MS-911W cannot be removed


without removing MS-910W.
Separate the two
after removing them.

[NOTE]
Both MS-910W and
MS-911W have sharp
needle tip.
Take due caution to
prevent needle-stick
accidents.
49
MEK-9100   Service Training
       Disassembly
MS-910W   SAMPLER UNIT  
  (1) Remove the cup tray.
     Release the 3 black latches for the tray.
     Pull out the tray in the direction of the arrow.

Release the latches

50
MEK-9100   Service Training
      
MS-910W   SAMPLER UNIT  
Disassembly
  (2) Remove 3 connected tubes.
     1 tube connected to the sample needle
       Tube: 1-1
     2 tubes connected to sample needle rinsing unit
       Tube: A22 and A23
[Note] Do not mistake the tube connection.
Place a mark on the tube before removing.
A22

A23

1-1

51
MEK-9100   Service Training
       Disassembly
MS-910W   SAMPLER UNIT  
  (3) Move 3 tubes to the front side.
     Pass the 3 tubes through the grommet to the front.
  (4) Disconnect 3 sensor connectors.
     Disconnect 3 sensor connectors.
     Pass the cables outside of the sampler unit.
[Note] Take care of the needle when the sampler is moved.

52
MEK-9100   Service Training
       Disassembly
MS-910W   SAMPLER UNIT  
  (5) Disconnect the motor connector.
     Pass the cables outside of the sampler unit.
  (6) Disconnect the sampler sensor BD connector.
Pass the cables through the grommet and push them inside the analyzer.

53
MEK-9100   Service Training
      
MS-911W   OPEN AIR UNIT  
Disassembly
  (7) Remove the rinsing tubes.
Loosen the screw using a precision screwdriver in order to remove the tube.
Use a coin or a large flat-blade screwdriver to remove the screw.
   Disconnect 3 tubes.
Tubes: ORN-U, ORN-D, No Label
Pull the tubes out from the holes that
they pass through.
       
No
Label

ORN-D ORN-U 54
MEK-9100   Service Training
       Disassembly
MS-911W   OPEN AIR UNIT  
  (8) Disconnect 3 connectors.
Disconnect 2 sensor connectors.
Disconnect the motor connector.     

Sensor

Motor
55
MEK-9100   Service Training
       Disassembly
MS-910W   SAMPLER UNIT  
  (9) Remove 5 screws.
Screw: PS (black) 3×8 5 pcs    

56
MEK-9100   Service Training
       Disassembly
MS-910W   SAMPLER UNIT  
  (10) Release the hook.
Lift the unit up slightly to release it from the hook on the Hook
housing. Hold the unit diagonally as shown in the picture.
  (11) Disconnect the connector.
Disconnect the connector located on the back side of the
sampler unit.

57
MEK-9100   Service Training
       Disassembly
Remove MS-911W from MS-910W.
    Screw: PS (black) 3×8 3 pcs

58
MS-910W - SAMPLER UNIT MS-910W

[Functions]
 Dispenses diluted samples into cups in
coordination with the diluter unit.
 Moves the sample needle to its designated
positions.
 The sample needle has pencil shaped tip. It
prevents the suction of rubber dust and clogging
will be minimized.
 Rinse mechanism without O-ring. (Previous
Celltac models with O-ring). Sample needle
 One nozzle aspirates and dispense all blood         
sample. (MEK-8222 has two sampling nozzles.)

Symbol in flow diagram 59


MS-911W - OPEN AIR UNIT

[Functions]
 Adjusts the pressure inside the sample tube set in the
autoloader to 1 atmospheric pressure by puncturing it with a
vent needle.
  (The pressure inside of the sample tube must be adjusted to
the same pressure as outside because the amount of blood
sample in the sample tubes varies. If the pressure is not
adjusted to 1 atmospheric pressure, it may affect the amount
of blood being aspirated.)

Symbol in flowdiagram
60
MEK-9100   Service Training
       Disassembly
MS-912W   OPEN LOADER UNIT  
  (1) Remove 6 screws.
Screw: PS (black) 3×8 5 pcs
Screw: PSW (black) 3×12 1 pc
  (2) Pull the unit forward.
  (3) Disconnect 3 connectors.
For sensor: 2 connectors
For motor: 1 connector

61
MS-912W - OPEN LOADER UNIT

[Functions]
 Allow measurement using small-quantity sample
containers in cases where a vacuum sample tube
cannot be used for measurement.
 Allow measurement of STAT (urgent) samples.
 For Protein Cleaning.

62
MEK-9100   Service Training
       Disassembly
MP-911W   ISO PUMP UNIT   ISO-U
  (1) Remove 2 tubes.
    Tubes: ISO-U and ISO-D

  (2) Remove 3 screws.


    Screw: PS (black) 3×12 3 pcs
ISO-D

[Note]
Some fluid will come out
from the cylinder when
removing the tubes.

* Be careful not to bend the cylinder block connectors when removing tubes. 63
MEK-9100   Service Training
       Disassembly
MP-911W   ISO PUMP UNIT  
  (3) Pull the unit forward.
  (4) Disconnect 4 connectors.
    For sensor: 3 connectors
    For motor: 1 connector

[Note]
After the unit is removed, keep it
laid to prevent the unit from falling
which will result in the cylinder
block connectors being broken. 64
MEK-9100   Service Training
       Disassembly
MP-911W   ISO PUMP UNIT

・ CYLINDER BLOCK ASSY(ISO)


  (1) Remove 4 screws.
    Screw: BH 3×8 4 pcs
  (2) Pull the unit forward.

CYLINDER BLOCK ASSY(ISO)


Code: RPA-6112911268

65
MP-911W - ISO PUMP UNIT
[Functions]
 Moves the cylinder up and down using the motor to
dilute samples and aspirate / dispense diluent for
mixing.
 IWBC (about 220 times dilution)
 RBC (about 500 times dilution)
 OWBC (about 100 times dilution)
 Aspirates and dispenses diluent for cleaning cups, and
aspirates / dispenses detergent for cleaning.
 When the sample needle is raised, ISO pump unit
dispenses for the rinse.
 After the sample needle is raised, the positive
pressure of ISO chamber dispenses diluent for
rinse.
 During protein cleaning, this pump dispenses
CLEANAC810 to fluid path.
 When this pump unit is replaced,
it is necessary to calibrate the analyzer.
Symbol in flow diagram
66
MEK-9100   Service Training
       Disassembly
MP-912W SAMPLE/RBC PUMP UNIT
  (1) Remove the tubes.
    Tubes (SAMPLE PUMP): S-P-U and S-P-D
    Tubes (RBC PUMP): R-P-U and R-P-D
  (2) Remove the screws.
    Screw: PS (black) 3×12 3 pcs each
S-P-U R-P-U

R-P-D
S-P-D

* Be careful not to bend the cylinder block connectors when removing tubes. 67
MEK-9100   Service Training
       Disassembly
MP-912W SAMPLE/RBC PUMP UNIT
  (3) Pull the unit forward.
  (4) Disconnect the connectors.
    For sensor: 2 connectors each
    For motor: 1 connector each

68
MEK-9100   Service Training
       Disassembly
MP-912W SAMPLE/RBC PUMP UNIT

・ CYLINDER BLOCK ASSY(SAMPLE/RBC)


  (1) Remove the screws.
    Screw: BH 3×8 2 pcs each
  (2) Pull the assy out.
Code:
RPA-6112911271

69
MEK-9100   Service Training
       Disassembly
MC-910W   CBC MEASURING UNIT
  (3) Remove 2 tubes connected to the front valves.
     Tubes: 8-1, 9-1

8-1 9-1

70