BC-2800 Hematology Analyzer

BC-2800 Specification
7.5“ color LCD , 19 parameters and 3 histograms
3-part differential of WBC
automatically cleaning sample probe
Automatic diluting, lyzing, mixing, rinsing and clog-clearing
Count speed: 30tests/hr
3 counting mode : all parameters 、 WBC/HGB 、 RBC/PLT
Automatic flushing, zapping
remain reagents alarm
expired date alarm
build-in recorder
bar code scanner and keyboard option
Up to 10000 sample results (including histograms) can be stored
BC-2800 Installation---Environment

Environment
The environment should be as free as possible from dust, mechanical vibrations,
loud
noises and electrical interference.
Avoid proximity to brush-type motors, flickering fluorescent lights and electrical
contacts that regularly turn on and off.
Avoid placing the instrument in direct sun or in front of a source of heat.
BC-2800 Installation---Space

Space
Comfortable working height
At least 40cm on each side is the preferred access to perform service procedures.
At least 10cm behind for cabling and ventilation
There is enough space for reagents.
 
BC-2800 Installation---Power

Power
A properly grounded socket supplying 220 or 110 volts
The fluctuation should be in the range of ±10%.
The power frequency should be 50/60Hz.
Power consumption is less than 180VA.
 
 
Extension function in tabular review menu

Select some samples in the sample tabular review menu, and then p
ress [F5] to access “extension function” menu.
[PRINT] print the current displayed content
[HELP] help information
[MENU] exit menu
Extension function: “repetition” and “trend graph”
“trend graph” function only open to administrator
Trend graph

Password: 2826
Procedure: press F5 then press F1
Trend graph analysis mean: it is to monitor one patient’s illness trend.
BC-2800 sample and dispense

Blood needed:
whole blood: 13μL
pre-diluted: 20μL
Volume of volumetric testing: 300μL ×2times
Value of lyse: 0.5mL(whole blood) , 0.36mL(pre-dilute
d)
Ratio of whole blood Capillary
WBC/HGB 1:308 1:400
RBC/PLT 1:44862 1:44402
Aperture size : 80μm ( diameter ), 70mm ( depth )
Count time : less than 2 min / test
Range of linear and display

Parameter Linear Display


WBC 0 - 99.9 ×109/L 0.0-499.9 ×109/L
RBC 0 -9.99×1012/L 0.0-9.99 ×1012/L
HGB 0 - 300g/L 0- 300g/L
MCV ------ 0-250 fL
PLT 0 - 999×109/L 0-2999 ×109/L
Repetition

Parameter linear Repetition(CV%)


4.0-7.5×109/L ≤ 3.0
WBC
7.6-15.0×109/L ≤ 2.0
3.0-4.5×1012/L ≤ 2.0
RBC
4.5-6.5×1012/ L ≤ 1.5
HGB 110 – 180g/L ≤ 1.5
MCV 70.0 – 100.0fL ≤ 1.0
100-200 ×109/L ≤ 5.0
PLT
201 - 500×109/L ≤ 4.0

Diff
CV   100%
Mean
Power Supply

AC Input :
AC 220V±30V 50/60Hz±1Hz
Consumption :
150 VA
Fuse :
250 V 2A
Screen Saver

Without any operation in 10 minutes, the system will enter the


saver program.

LCD display turns off

Power indicator is still light.

Any key exists the saver program.


Front Panel

7.5” color LCD

Keypad

Record

Power indicator
Sample probe

Start key
Rear panel

Diluent coonector

Power switch and fuse Rinse connector


Lyse connector
Equipotential terminal Waste connector
Left side
Scanner interface

Printer or upgrade interface

Transition

Power or floppy

Keyboard
Display feature
Title Count Mode System System time
Status

Window

Error Messages Reagent Status

MENU On-line Help


Remain reagents

1 ) >99tests : remain reagents and remain waste


container are enough for 99tests

2 ) 0 test : any one of the remain reagents or remain


waste container is not enough for 1 test. The warning message
will be show in error area.
Actual tests : show the actual tests number
Keypad

注意: PS/2 标准键盘操作的区别

用 PS/2 标准键盘执行模式切换功能键是: [Ctrl + A]


Count

Menu tree Revi ew

Qual i t y cont r ol Cont r ol s

X- B anal ys i s

Set up Pas s wor d

Set t i ngs

Ser vi ce Mai nt enance

St at us

Sel f - t es t

Log

Conf i g

Pr i nt

Cal i br at i on Manual

Aut o

Aut o - f r es h bl ood

Shut down
Reagents consumption

E-Z
Diluent Rinse Lyse
cleanser
46.5ml
16.5ml 0.5ml ---
( clean & once background test)
Normal start-up
75ml
( clean & twice background test 26ml 1ml ---

All parameters
28.5ml 9.5ml 0.5ml ---
(whole blood)
All parameters 30.5ml
9.5ml 0.36ml ---
(pre-diluted)
Normal shut down 13.5ml 5.5 ml --- 1ml
Power on and shut down
1. Switch on BC-2800 , and the power indicator will be light.
2. The Initializing procedure will continue about 80s.
3. Self test procedure.
The last shut down status
Procedure

Normal shutdown clean tubing and background testing


Abnormal shutdown Reset motors, check software, clean
( Directly shutdown ) tubing and background test
Prime tubing, clean tubing several
Package or empty tubing
times, background test.

4 The system will enter “count” menu. If there are errors during startup, the
error message will be displayed on left bottom of screen.
Background test

Parameter Acceptable limit


WBC  0.3  109 / L
RBC  0.03  1012/ L
HGB  1g/L
HCT  0.5 %
PLT  10  109 / L
Sample information input

1. Only ID mode
2. All information
3. Batch information mode
Preparation of pre-diluted blood

1. Press [MODE] to select the prediluted mode (any mode preceded by PB).
2. Press [DILUENT] to enter the Add Diluent screen
3. Prepare a clean sample cup under the sample probe and press [START] t
o dispense 1.6ml diluent to the cup.

4. Take out the sample cup after the sample


probe rose up and press [ENTER] to exit the
Add Diluent screen.
5. Collect 20μL capillary blood specimen and
immediately add it to the sample cup and
shake to mix them.
[SETUP]-[Settings]

[F1] is to select the setting items.


[SETUP]-[Settings]--Reagents Remain

Reagent Valid volume


Diluent ( 0-30.0 ) L
Rinse ( 0-30.0 ) L
Lyse ( 0-999 ) mL
Waste ( 0-99.9 ) L
[SETUP]-[Settings]—Print & Communication
[SETUP]-[Settings]—Print & Communication

Printing
Device
Format1 Format2 Format3 Format4

Recorder Parameters + Parameters Parameters+ Parameters


histograms histograms

Printer Whole page with Whole page Whole page with Whole page
histograms without histograms without
histograms histograms
[SETUP]-[Settings]—Date & time
[Setup]-[Settings]---[Gain]
[Setup]-[Settings]---[Gain]--WBC

• too small WBC gain that should be increased

• too large WBC gain that should be decreased.


[Setup]-[Settings]---[Gain]--RBC

• When the difference between the actual MCV analysis result and the
expected result exceeds 6%, need to change the RBC gain.
• For example, assuming the expected MCV result is 90.0fL, while the
actual analysis result is 82.0fL, the gain factor as below:

ExpectedMCV 90.0
=  1.098
ActualMCV 82.0
[Setup]-[Settings]---[Gain]--HGB

• adjust the HGB gain to change the HGB blank voltage, which
usually should be set between 4.3 ~ 4.5V.
[Setup]-[Settings]--count
[Setup]-[Settings]--Reference Range
Patient Group Sex Age
Not specified Not specified
General
Not specified > 13 years
Male adults Man > 13 years
Female adults Woman > 13 years
Pediatric
Not specified >= 1 month and <= 13 years
patients
Neonates Not specified <1 month
[Setup]-[Settings]--other
[Service]-[Maintenance]--Prime
[Service]-[Maintenance]—Zap & Flush
[Service]-[Maintenance]—Probe cleanser clean

• If your analyzer is to run 24 hours a day, you should perform


this procedure every 3 days.

• If you follow the shutdown procedure to turn off your analyzer


everyday, you should perform this procedure every week.

• To make sure this analyzer functions normally, every time the


accumulated analyzed samples reach 200, a dialog box will pop
up to remind the user to perform the Probe cleanser cleaning
procedure.
[Service]-[Maintenance]—Lyse test

“Lyse test” is to check the injected volume for lyse is correct or not. W
hen shall we use this function?
When the WBC result or its histogram is abnormal, we should run this f
unction.

Screw
off
these
two
screws
[Service]-[Maintenance]—Lyse test

2ml lyse will be dispensed to the count bath.

Remove the The inject


Count bath ed lyse m
shield cover ust over t
he secon
d mark
[Service]-[Maintenance]—Drain

• Drain bath
Empty the count bath, and prime it again

• Drain Tubing
Empty all tubing, then turn off the
equipment
[Service]-[Maintenance]—Clean
• Clean bath
Use diluent to clean count bath

• Clean wipe block


After being used for a long time, the bottom of the probe wipe
block may be contaminated by blood.

When the accumulated analyzed samples reach 4,000, a dialog box


will
pop up to remind the user to clean the wipe block
[Service]-[Maintenance]—Prepare to ship

• Empty the tubing


• Wash the tubing
• Empty the tubing
• Turn off the equipment
[Service]-[System test]
[Service]-[self-test]
The available self-test items are divided into four groups, the
Tubing, Machine, Valve and Circuit.
[Service]-[self-test]—[Test tubing]
Count Time WBC ( S ) & Count Time RBC ( S )
It measures the duration of a WBC or RBC count
Aperture(v)
It measures the voltage (v) over the aperture.
Vacuum
It checks whether the vacuum system functions normally.
Pressure
It checks whether the system flushes the aperture at a normal press
ure.
Filter
It checks whether the filter functions normally.
[Service]-[self-test]—[Machine]
[Service]-[self-test]—[Valve]
[Service]-[self-test]—[Circuit]
It is to monitor that WBC, RBC and PLT signals can be properly converted
into digital signals or not.
[Service]-[Log]

The recorded events are divided into four groups, All, Para set, Error and Other
(including setting discriminators, system self-test and updating system software)

This analyzer can


save maximum 100
log files and once
the maximum
number has been
reached, the newest
log will
automatically cover
the oldest one.
[Service]-[Configuration]
Every screen displays 13 items
It only shows BC-2800 current configuration, no change.
[Service]-[Print management]

The printing tasks are queued in this screen


[Calibration]---[Manual]

Password: 2826

Valid range:
75%---125%
[Calibration]---[Auto]

Password: 2826

Valid range:
75%---125%
[Calibration]---[Fresh blood]

Password: 2826

Valid range:
75%---125%