BC-2800 Specification
7.5“ color LCD , 19 parameters and 3 histograms
3-part differential of WBC
automatically cleaning sample probe
Automatic diluting, lyzing, mixing, rinsing and clog-clearing
Count speed: 30tests/hr
3 counting mode : all parameters 、 WBC/HGB 、 RBC/PLT
Automatic flushing, zapping
remain reagents alarm
expired date alarm
build-in recorder
bar code scanner and keyboard option
Up to 10000 sample results (including histograms) can be stored
BC-2800 Installation---Environment
Environment
The environment should be as free as possible from dust, mechanical vibrations,
loud
noises and electrical interference.
Avoid proximity to brush-type motors, flickering fluorescent lights and electrical
contacts that regularly turn on and off.
Avoid placing the instrument in direct sun or in front of a source of heat.
BC-2800 Installation---Space
Space
Comfortable working height
At least 40cm on each side is the preferred access to perform service procedures.
At least 10cm behind for cabling and ventilation
There is enough space for reagents.
BC-2800 Installation---Power
Power
A properly grounded socket supplying 220 or 110 volts
The fluctuation should be in the range of ±10%.
The power frequency should be 50/60Hz.
Power consumption is less than 180VA.
Extension function in tabular review menu
Select some samples in the sample tabular review menu, and then p
ress [F5] to access “extension function” menu.
[PRINT] print the current displayed content
[HELP] help information
[MENU] exit menu
Extension function: “repetition” and “trend graph”
“trend graph” function only open to administrator
Trend graph
Password: 2826
Procedure: press F5 then press F1
Trend graph analysis mean: it is to monitor one patient’s illness trend.
BC-2800 sample and dispense
Blood needed:
whole blood: 13μL
pre-diluted: 20μL
Volume of volumetric testing: 300μL ×2times
Value of lyse: 0.5mL(whole blood) , 0.36mL(pre-dilute
d)
Ratio of whole blood Capillary
WBC/HGB 1:308 1:400
RBC/PLT 1:44862 1:44402
Aperture size : 80μm ( diameter ), 70mm ( depth )
Count time : less than 2 min / test
Range of linear and display
Diff
CV 100%
Mean
Power Supply
AC Input :
AC 220V±30V 50/60Hz±1Hz
Consumption :
150 VA
Fuse :
250 V 2A
Screen Saver
Keypad
Record
Power indicator
Sample probe
Start key
Rear panel
Diluent coonector
Transition
Power or floppy
Keyboard
Display feature
Title Count Mode System System time
Status
Window
X- B anal ys i s
Set t i ngs
St at us
Sel f - t es t
Log
Conf i g
Pr i nt
Cal i br at i on Manual
Aut o
Aut o - f r es h bl ood
Shut down
Reagents consumption
E-Z
Diluent Rinse Lyse
cleanser
46.5ml
16.5ml 0.5ml ---
( clean & once background test)
Normal start-up
75ml
( clean & twice background test 26ml 1ml ---
)
All parameters
28.5ml 9.5ml 0.5ml ---
(whole blood)
All parameters 30.5ml
9.5ml 0.36ml ---
(pre-diluted)
Normal shut down 13.5ml 5.5 ml --- 1ml
Power on and shut down
1. Switch on BC-2800 , and the power indicator will be light.
2. The Initializing procedure will continue about 80s.
3. Self test procedure.
The last shut down status
Procedure
4 The system will enter “count” menu. If there are errors during startup, the
error message will be displayed on left bottom of screen.
Background test
1. Only ID mode
2. All information
3. Batch information mode
Preparation of pre-diluted blood
1. Press [MODE] to select the prediluted mode (any mode preceded by PB).
2. Press [DILUENT] to enter the Add Diluent screen
3. Prepare a clean sample cup under the sample probe and press [START] t
o dispense 1.6ml diluent to the cup.
Printing
Device
Format1 Format2 Format3 Format4
Printer Whole page with Whole page Whole page with Whole page
histograms without histograms without
histograms histograms
[SETUP]-[Settings]—Date & time
[Setup]-[Settings]---[Gain]
[Setup]-[Settings]---[Gain]--WBC
• When the difference between the actual MCV analysis result and the
expected result exceeds 6%, need to change the RBC gain.
• For example, assuming the expected MCV result is 90.0fL, while the
actual analysis result is 82.0fL, the gain factor as below:
ExpectedMCV 90.0
= 1.098
ActualMCV 82.0
[Setup]-[Settings]---[Gain]--HGB
• adjust the HGB gain to change the HGB blank voltage, which
usually should be set between 4.3 ~ 4.5V.
[Setup]-[Settings]--count
[Setup]-[Settings]--Reference Range
Patient Group Sex Age
Not specified Not specified
General
Not specified > 13 years
Male adults Man > 13 years
Female adults Woman > 13 years
Pediatric
Not specified >= 1 month and <= 13 years
patients
Neonates Not specified <1 month
[Setup]-[Settings]--other
[Service]-[Maintenance]--Prime
[Service]-[Maintenance]—Zap & Flush
[Service]-[Maintenance]—Probe cleanser clean
“Lyse test” is to check the injected volume for lyse is correct or not. W
hen shall we use this function?
When the WBC result or its histogram is abnormal, we should run this f
unction.
Screw
off
these
two
screws
[Service]-[Maintenance]—Lyse test
• Drain bath
Empty the count bath, and prime it again
• Drain Tubing
Empty all tubing, then turn off the
equipment
[Service]-[Maintenance]—Clean
• Clean bath
Use diluent to clean count bath
The recorded events are divided into four groups, All, Para set, Error and Other
(including setting discriminators, system self-test and updating system software)
Password: 2826
Valid range:
75%---125%
[Calibration]---[Auto]
Password: 2826
Valid range:
75%---125%
[Calibration]---[Fresh blood]
Password: 2826
Valid range:
75%---125%