Insert ALP2 0003333752190c501 V8 en
Insert ALP2 0003333752190c501 V8 en
Insert ALP2 0003333752190c501 V8 en
ALP2
Alkaline Phosphatase acc. to IFCC Gen.2
Order information
Analyzer(s) on which cobas c pack(s) can be used
03333752 190 Alkaline Phosphatase acc. to IFCC Gen.2 ALP2S 200 tests System‑ID 07 6761 1 Roche/Hitachi cobas c 311, cobas c 501/502
03333701 190 Alkaline Phosphatase acc. to IFCC Gen.2 ALP2L 400 tests System‑ID 07 6760 3 Roche/Hitachi cobas c 311, cobas c 501/502
10759350 190 Calibrator f.a.s. (12 x 3 mL) Code 401
10759350 360 Calibrator f.a.s. (12 x 3 mL, for USA) Code 401
12149435 122 Precinorm U plus (10 x 3 mL) Code 300
12149435 160 Precinorm U plus (10 x 3 mL, for USA) Code 300
12149443 122 Precipath U plus (10 x 3 mL) Code 301
12149443 160 Precipath U plus (10 x 3 mL, for USA) Code 301
10171743 122 Precinorm U (20 x 5 mL) Code 300
10171735 122 Precinorm U (4 x 5 mL) Code 300
10171778 122 Precipath U (20 x 5 mL) Code 301
10171760 122 Precipath U (4 x 5 mL) Code 301
05117003 190 PreciControl ClinChem Multi 1 (20 x 5 mL) Code 391
05947626 190 PreciControl ClinChem Multi 1 (4 x 5 mL) Code 391
05947626 160 PreciControl ClinChem Multi 1 (4 x 5 mL, for USA) Code 391
05117216 190 PreciControl ClinChem Multi 2 (20 x 5 mL) Code 392
05947774 190 PreciControl ClinChem Multi 2 (4 x 5 mL) Code 392
05947774 160 PreciControl ClinChem Multi 2 (4 x 5 mL, for USA) Code 392
04489357 190 Diluent NaCl 9 % (50 mL) System‑ID 07 6869 3
ALP2
Alkaline Phosphatase acc. to IFCC Gen.2
On‑board in use and refrigerated on the analyzer: 8 weeks Reagent pipetting Diluent (H2O)
Diluent NaCl 9 % R1 75 µL 25 µL
Shelf life at 2‑8 °C: See expiration date R2 17 µL 21 µL
on cobas c pack
label. Sample volumes Sample Sample dilution
On‑board in use and refrigerated on the analyzer: 12 weeks Sample Diluent (NaCl)
Normal 2.8 µL – –
Specimen collection and preparation
For specimen collection and preparation only use suitable tubes or Decreased 2.8 µL 20 µL 80 µL
collection containers. Increased 2.8 µL – –
Only the specimens listed below were tested and found acceptable.
Serum. cobas c 502 test definition
Plasma: Li‑heparin plasma.
Assay type Rate A
The sample types listed were tested with a selection of sample collection
tubes that were commercially available at the time of testing, i.e. not all Reaction time / Assay points 10 / 19‑48
available tubes of all manufacturers were tested. Sample collection systems Wavelength (sub/main) 480/450 nm
from various manufacturers may contain differing materials which could
affect the test results in some cases. When processing samples in primary Reaction direction Increase
tubes (sample collection systems), follow the instructions of the tube Units U/L (µkat/L)
manufacturer.
Centrifuge samples containing precipitates before performing the assay. Reagent pipetting Diluent (H2O)
R1 75 µL 25 µL
Stability:7 7 days at 15‑25 °C
R2 17 µL 21 µL
7 days at 2‑8 °C
Sample volumes Sample Sample dilution
2 months at (-15)‑(-25) °C
Sample Diluent (NaCl)
Materials provided
Normal 2.8 µL – –
See “Reagents – working solutions” section for reagents.
Decreased 2.8 µL 20 µL 80 µL
Materials required (but not provided)
▪ See “Order information” section Increased 5.6 µL – –
General laboratory equipment Calibration
Assay Calibrators S1: H2O
For optimum performance of the assay follow the directions given in this S2: C.f.a.s.
document for the analyzer concerned. Refer to the appropriate operator’s
manual for analyzer‑specific assay instructions. Calibration mode Linear
The performance of applications not validated by Roche is not warranted Calibration frequency 2‑point calibration
and must be defined by the user. • after reagent lot change
Applications for serum and plasma • as required following quality control
procedures
cobas c 311 test definition
Traceability: This method has been standardized against the proposed
Assay type Rate A IFCC formulation6 using calibrated pipettes together with a manual
photometer providing absolute values and the substrate‑specific
Reaction time / Assay points 10 / 13‑31 absorptivity, ε.
Wavelength (sub/main) 480/450 nm Quality control
Reaction direction Increase For quality control, use control materials as listed in the "Order information"
Units U/L (µkat/L) section.
In addition, other suitable control material can be used.
Reagent pipetting Diluent (H2O)
The control intervals and limits should be adapted to each laboratory’s
R1 75 µL 25 µL individual requirements. Values obtained should fall within the defined
R2 17 µL 21 µL limits. Each laboratory should establish corrective measures to be taken if
values fall outside the defined limits.
Sample volumes Sample Sample dilution Follow the applicable government regulations and local guidelines for
Sample Diluent (NaCl) quality control.
Normal 2.8 µL – – Calculation
Roche/Hitachi cobas c systems automatically calculate the analyte activity
Decreased 2.8 µL 20 µL 80 µL of each sample.
Increased 2.8 µL – – Conversion factor: U/L x 0.0167 = µkat/L
cobas c 501 test definition Limitations - interference
Criterion: Recovery within ± 10 % of initial value at an alkaline phophatase
Assay type Rate A activity of 100 U/L (1.67 µkat/L).
Reaction time / Assay points 10 / 19‑48 Icterus:8 No significant interference up to an I index of 60 for conjugated
Wavelength (sub/main) 480/450 nm and unconjugated bilirubin (approximate conjugated and unconjugated
bilirubin concentration: 1026 µmol/L or 60 mg/dL).
Reaction direction Increase Hemolysis:8 No significant interference up to an H index of 200
Units U/L (µkat/L) (approximate hemoglobin concentration: 124 µmol/L or 200 mg/dL).
ALP2
Alkaline Phosphatase acc. to IFCC Gen.2
ALP2
Alkaline Phosphatase acc. to IFCC Gen.2