Technical specification:
D E A D W E I G H T
T E S T E R S
Model
Range
Model
Range
M 1900/1
M 1900/3
1 to 300 bar
10 to 4000 psi
W 1900/1
W 1900/3
M 2000/1
M 2000/3
1 to 350 bar
10 to 5000 psi
W 2000/1
W 2000/3
M 2100/1
M 2100/3
1 to 600 bar
10 to 8000 psi
W 2100/1
W 2100/3
M 2200/1
M 2200/3
1 to 700 bar
10 to 10000 psi
W 2200/1
W 2200/3
M 2800/1
M 2800/3
1 to 1100 bar
10 to 16000 psi
W 1800/1
W 1800/3
0.5 to 35 bar
5 to 500 psi
M 1800/1
M 1800/3
1 to 35 bar
10 to 500 psi
W 2000/1H
W 2000/3H
10 to 350 bar
200 to 5000 psi
M 4000/1
M 4000/3
1 to 120 bar
15 to 1800 psi
W 2100/1H
W 2100/3H
20 to 600 bar
400 to 8000 psi
M 2000/1H
M 2000/3H
10 to 350 bar
200 to 5000 psi
W 2200/1H
W 2200/3H
20 to 700 bar
400 to 10000 psi
M 2200/1H
M 2200/3H
20 to 700 bar
400 to 10000 psi
M 2800/1H
M 2800/3H
20 to 1100 bar
400 to 16000 psi
M 2820/1
M 2820/3
20 to 1400 bar
400 to 20000 psi
M 3800/1
M 3800/3
30 to 2500 bar
500 to 40000 psi
M 3820/1
M 3820/3
40 to 4000 bar
500 to 60000 psi
Optional extras
Built in hand pump
Large volume systems are easily primed.
Greatly speeds up priming and pressure generation.
Enables an effortless, small bore high pressure ram screw
to be fitted.
There is no additional valve operation involved.
Standard on Models M2800, M2820, M3800, and M3820
Motor drive
If a continuous pressure is required, a motor drive can be
provided to rotate the weights. Rotating the weights by
hand is satisfactory for standard calibrations.
The motor is isolated from the Deadweight Tester by a
thermal barrier, preventing warming of the measuring
piston. The motor is suitable for 210/250, 105/125 volts at
50/60 HZ, please specify requirement at time of order.
Not available on models M2800, M2820.
Specifications
GENERAL
WEIGHT INCREMENT
Accuracy
0.015% of reading
0.010% when specified
M3800, M3820
0.02% of reading
0.015% when specified
Total instrument weight 12 kg
Up to 1400 bar/20000 psi: 12 kg
M3800, M3820: 29 kg
Total boxed weight set
Up to 700 bar/10000 psi: 30 kg
M2800: 48 kg
M2820: 60 kg
M3800: 120 kg
M3820: 180 kg
Instrument size
47 x 32 x 20 cm L x D x H (including lid)
M3800, M3820: 63 x 41 x 33 x cm L x D x H
STARTING PRESSURE
Weight density
7.9 g/cm3
Piston material
Oil Operated: Tungsten Carbide
Water Operated: Stellite (up to 35 bar/500 psi)
Tungsten Carbide (above 35 bar/500 psi)
Cylinder material
Hardened martensitic steel, Tungsten Carbide (water
above 35 bar/500 psi)
D E A D W E I G H T
T E S T E R S
T5000 SERIES
Range
15 to 250 mbar
5 to 100 inH2O
15 to 600 mbar
5 to 250 inH2O
15 to 1000 mbar
5 to 400 inH2O
30 to 2000 mbar
12 to 800 inH2O
T9000 SERIES
PRESSURE MODELS
Model
T 1150/1
T 1150/3
T 1100/1
T 1100/3
T 2400/1
T 2400/3
T 2500/1
T 2500/3
Range
15 to 1000 mbar
5 to 400 inH2O
30 to 2000 mbar
12 to 800 inH2O
0.2 to 7 bar
3.0 to 100 psi
0.2 to 10 bar
3.0 to 150 psi
Model
T 1400/1
T 1400/3
T 2300/1
T 2300/3
Range
0.2 to 30 bar
3 to 400 psi
0.2 to 35 bar
3 to 500 psi
Range
1 to 70 bar
10 to 1000 psi
1 to 140 bar
10 to 2000 psi
Model
T 2700/1L
T 2700/3L
T 2900/1L
T 2900/3L
Range
1 to 70 bar
10 to 1000 psi
1 to 140 bar
10 to 2000 psi
VACUUM MODELS
Model
T 2600/1
T 2600/3
Range
0.03 to 1 bar
1 to 30 in Hg
Model
T 3550/1
T 3550/3
T 3580/1
T 3580/3
T 3400/1
T 3400/3
T 3500/1
MODEL T3550
T 3500/3
Range
0.03 to 1 bar V
15 to 1000 mbar P
1 to 30 inHg V
5 to 400 inH2O P
0.03 to 1 bar V
30 to 2000 mbar P
1 to 30 inHg V
12 to 800 inH2O P
0.03 to 1 bar V
0.2 to 30 bar P
1 to 30 inHg V
3 to 400 psi P
0.03 to 1 bar V
0.20 to 35 bar P
1 to 30 inHg V
3 to 500 psi P
Optional extras
V = Vacuum
Motor drive
Available for models T2400, T2500, T1400, T2300,
T2700, T2900 only.
Start point increases: Models T2400, T2500, T1400, T2300
to 0.25 bar. Models T2700, T2900 to 2 bar/40 psi.
P = Pressure
Specifications
GENERAL
Accuracy
0.015% of reading
0.010% when specified
MATERIALS OF CONSTRUCTION
Weight material
Series 3 non magnetic austenitic stainless steel
Weight density
7.9 g/cm3
Piston material
Tungsten Carbide
T5000, T1100, T1150, Vacuum: Ceramic
Cylinder material
Hardened martensitic steel
T2700, T2900: Tungsten Carbide
D E A D W E I G H T
T E S T E R S
LIQUID
This is the most portable and flexible version of our
comparison test pumps. The unit utilises our standard
Deadweight Tester system. The piston and weights are
replaced by a second test station to which either a test
gauge, transducer or transfer standard is fitted for
comparison with the instrument being tested. A hand pump
is fitted giving the following benefits:
MODEL T1300
0 to 20000 psi
MODEL T1200
LIQUID
Tested with water and suitable for use with:
water, mineral/vegetable oil, alcohol.
Fitted with a reservoir and valve, pressure is quickly
generated with the ram screw.
Model: T1200
0 to 10000 psi
MODEL T4100
AIR
Pressure is generated with the sensitive hand pump and is
vented with a fine control needle valve.
Model: T4100
Range: 0 to 20 bar
0 to 300 psi
VACUUM
Vacuum is generated quickly & easily with the hand pump
and vented with a fine control needle valve.
Model: T4200
MODEL T4200
Separators
GAS
LIQUID TO AIR
For high pressure pneumatic calibration, this unit interfaces with hydraulic deadweight testers, giving a rateless liquid to air
separation. A series of fluid traps ensures the instrument under test remains dry. Needle valves control the pressure, and the
deadweight tester ram screw controls the fluid level. Please specify if interface is for oil or water application.
Model: API5000/1
Model: API5000/2
Model: API5000/3
LIQUID TO LIQUID
This unit has a flexible "Viton" diaphragm, separating the
deadweight tester fluid and a non contaminating liquid.
Any liquid can be used that is compatible with "Viton" and
aluminium bronze. The unit also protects the deadweight
tester from contamination.
Model: T3600
Model: T3601
DIRT/MOISTURE TRAP
Protects pneumatic Deadweight Testers, when the
cleanliness of the instrument under test cannot be
guaranteed. The unit is mounted directly onto the test
station. Particles and moisture are trapped within the
body, which are visible on Model T4400.
MODEL T3600
MODEL T4400
Model: T4400
Model: T4401
Ancilliary equipment
ANGLE ADAPTOR
To calibrate back connection gauges, the angle adaptor fits
directly onto the test station, converting it through 90,
allowing the same adaptors to be used.
Model: T3700
POINTER REMOVER/PUNCH
To remove and refit the pointer of a pressure gauge. This
two in one tool has a spring loaded plunger to quickly &
consistently refit the pointer.
Model: T4600
MODEL T3700
MODEL T4600
D E A D W E I G H T
T E S T E R S
0 to 200 inH2O.
Two deadweight tester systems are built
into a single instrument with a crossconnecting valve. The two systems are
pressurised in parallel with the crossconnecting valve open, until both pistons and
weights are floating at the specific line
pressure. With the cross connecting valve
closed the differential pressure is generated by
each additional weight loaded to the measuring
piston. The balance piston remains floating,
ensuring that the line pressure is maintained.
MODEL M4000DA
Range
Differential
Uncertainty
M 1800/1D
M 1800/3D
2 to 35 bar
20 to 500 psi
0.0002 bar
0.003 psi
M 4000/1D
M 4000/3D
2 to 120 bar
30 to 1800 psi
0.0002 bar
0.003 psi
M 2000/1D
M 2000/3D
30 to 350 bar
400 to 5000 psi
0.00035 bar
0.005 psi
M 2200/1D
M 2200/3D
60 to 700 bar
800 to 10000 psi
0.007 bar
0.01 psi
Range
Differential
Uncertainty
T 1100/1D
T 1100/3D
0.0025 mbar
0.01 inH20
T 2500/1D
T 2500/3D
0.25 to 10 bar
7.0 to 150 psi
0.000025 bar
0.0004 psi
T 1400/1D
T 1400/3D
0.25 to 30 bar
7.0 to 400 psi
0.000035 bar
0.0005 psi
* M 4000/1DA
* M 4000/3DA
2 to 100 bar
30 to 1500 psi
0.0002 bar
0.003 psi
MODEL P7000
Shown with
optional fitted
handpump
C SERIES
10
Pressurements Limited
Unit 22, Apex Business Centre,
Boscombe Road, Dunstable,
Beds. LU5 4SB, U.K.
Telephone:
Fax:
Web site:
e-mail:
CALIBRATION
No. 0586
CAT. 110/5
A Druck Company
Due to Pressurements continuous improvement, specifications are subject to change without notice