Biorad Error Maintenance
Biorad Error Maintenance
Biorad Error Maintenance
Basic Power
Supply
Instruction
Manual
Catalog Number
164-5050
For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-424-6723
Table of Contents
Page
Safety.................................................................................................................i
Section 1
1.1
1.2
Introduction ......................................................................................................1
Overview ..........................................................................................................1
Unpacking ........................................................................................................2
Section 2
Section 3
Section 4
4.1
4.2
4.3
4.4
Appendix A
Specifications .................................................................................................10
List of Figures
1. Front View.
2. Rear View.
3. Front View with Legs in Lowered Position.
4. Front Panel.
5. Power Leads Connected Correctly.
6. Power Leads Connected Incorrectly.
7. Rear View Showing Fuse Drawer with Notches.
Safety
Caution/Warning
!
PowerPac power supplies use high output voltages that are electrically isolated from earth
ground to minimize the risk of electrical shock to the user. The following guidelines should
be observed and followed when using a PowerPac power supply.
PowerPac power supplies have passed test for operation at temperatures between 0 and
40 C, with relative humidity between 0 and 95% non-condensing. Operating the power
supply outside these conditions is not recommended by Bio-Rad and will void the
warranty.
1. To ensure adequate cooling of the power supply, be sure that there is at least 6 cm
clearance around the power supply. Do not block the fan vents at the rear of the unit.
2. Always connect the power supply to a 3-prong, grounded AC outlet, using the
3-prong AC power cord provided with the power supply.
3. Bio-Rad electrophoresis cells have molded two-prong plugs that are inserted into the
power supply's high voltage output jacks. These plugs have been EN 61010* certified
for safety compliance for use with PowerPac power supplies. Use of other plugs or
banana jacks is done at the user's own risk and is not recommended by Bio-Rad.
When inserting and removing the molded two-prong plug, always grasp the plug by the
molded support at the rear of the plug. Do not grasp the individual prong ends.
4. Do not operate the power supply in extreme humidity (>95%) or where condensation
can short the internal electrical circuits of the power supply.
5. When taking the power supply into a cold room, the unit can be operated
immediately. However, when removing the power supply from the cold room, let the
unit equilibrate to room temperature for a minimum of 2 hours before using it.
6. Never connect a high voltage output lead to earth ground. This defeats the floating
electrical isolation of the power supply and exposes the user to potentially lethal high
voltages.
Important
This instrument is intended for laboratory use only.
This product conforms to the class A standards for Electromagnetic Emissions, intended
for laboratory equipment applications. It is possible that emissions from this product may
interfere with some sensitive appliances when placed nearby or on the same circuit as
those appliances. The user should be aware of this potential and take appropriate
measures to avoid interference.
Bio-Rad's PowerPac power supplies are designed and certified to meet EN 61010* safety
standards. Certified products are safe to use when operated in accordance with the
instruction manual. This safety certification does not extend to electrophoresis cells or
accessories that are not EN 61010 certified, even when connected to this power supply.
This instrument should not be modified or altered in any way. Alteration of this instrument
will void the manufacturer's warranty, void the EN 61010 certification, and create a
potential safety hazard for the user.
Bio-Rad is not responsible for any injury or damage caused by the use of this instrument
for purposes other than those for which it is intended, or by modifications of the
instrument not performed by Bio-Rad or an authorized agent.
*EN 61010 is an internationally accepted electrical safety standard for laboratory instruments.
Section 1
Introduction
1.1 Overview
The PowerPac Basic provides constant voltage or constant current to instruments used in
electrophoresis. The power supply operates at the values specified for the constant
parameter. However, to prevent damage to the electrophoresis cell, the PowerPac Basic
provides automatic crossover to constant current or constant voltage, depending on
which set value is first reached. When the set limit of the non-constant parameter is
reached, and the power capability of the unit is not exceeded, the power supply will
switch, making the non-constant parameter the new constant parameter.
Output specifications:
Voltage:
Current:
Power:
75 watts (maximum).
FRONT PANEL
OUTPUT JACKS
AC INLET
POWER SWITCH
FUSE
Stackable case with adjustable viewing angle via flip down legs (see Figure 3)
LEGS
(LOWERED POSITION)
1.2 Unpacking
When you receive the power supply, carefully inspect the container for any damage which
may have occurred in shipping. Severe damage to the container may indicate damage to
the power supply itself. If you suspect damage to the unit, immediately file a claim with the
carrier in accordance with their instructions before contacting Bio-Rad Laboratories.
After unpacking the PowerPac Basic, remove the plastic film from the translucent green
top case. The plastic film may leave a residue. If so, clean with a soft, damp cloth.
Contents include:
Power cord
Instruction manual
Warranty card
Declaration of conformity
Section 2
Control Features
3 DIGIT LED
SCROLL KEYS
STOP KEY
CONSTANT
PARAMETER KEY
Key
Description
Constant Parameter Key:
Selects either constant voltage or current.
The LED indicates the selected parameter. During a run, maximum
power is indicated when both LEDs are lit.
Display Mode Key:
Selects the parameter to be displayed (volts, milliamperes or minutes).
The LED displays the value of the indicated parameter.
Scroll Keys:
Changes the value of the selected parameter. If the Scroll Key is
pressed constantly for more than 5 units in either direction, +/-, the
values will increase/decrease in increments of 10 to reach the desired
value faster.
Run & Pause Key:
Starts and pauses a run. Pausing allows editing of the constant
parameter and the parameter values.
Corresponding LED indicates the status of the power supply.
Stop key:
Stops run. Constant parameter and limit parameter values are
preserved. Timer is reset to zero.
Section 3
Setup and Operation
STEP
PROCEDURE
DESCRIPTION
1.
2.
3.
4.
5.
For volts
For milliamperes
STEP
PROCEDURE
DESCRIPTION
6.
7.
8.
STEP
PROCEDURE
DESCRIPTION
9.
10.
Note:
Editing the limit value is possible in the
pause mode. To change from an un-timed
to a timed run, stop the run and re-program.
11.
Pause mode
Note:
To change from an un-timed to a timed run,
stop the run and re-program.
12.
End of run.
13.
14.
Powering down
Section 4
Maintenance and Troubleshooting
4.1 Maintenance
The PowerPac Basic requires little maintenance to assure reliable operation. To clean the
case, first unplug the power supply. Use a damp cloth to wipe down the outer case.
4.2 Troubleshooting
Problem
No display/lights/fan
Cause
1. No AC power.
2. Blown fuse.
3. Power switch
exercised rapidly to
the on and off
positions.
Hardware failure
Output terminals for the
PowerPac Basic are
recessed 16 mm to
meet safety regulations.
Some leads are not
long enough to make
electrical connection.
No load detected
E1 error code
displayed
E2 error code
displayed
E3 error code
displayed
E5 error code
Detection played
E6 error code
displayed
E7 error code
displayed
Instrument not
connected to PS
The current load is
below 4mA
Over current (load
current greater then
400 mA)
Solution
1. Check if PowerPac Basic is
unplugged, or problem with AC
power source, or power switch is in
off position.
2. Replace fuse. See section 4.3 for
details.
3. The unit needs to be reset. Turn
power switch to the off position,
wait 510 seconds, then turn
power switch to the on position to
resume
normal operation.
Contact Bio-Rad Technical Resources.
Use the PowerPac Adaptor, which
accommodates most standard 4 mm
banana plugs, to make a secure
electrical connection.
Problem
E8 error code
reset.
displayed
Cause
Regulation error
E9 error code
displayed
Change in Load
Resistance.
Solution
Turn power supply off, then on to
If problem persists, contact Bio-Rad
Technical Resources
Check and correct any potential
resistance problem then,
press run/pause key to resume run
or,
press any key, other than the
run/pause
key, to clear the error code.
The PowerPac
Hardware failure
Hardware failure
Internal system error
Internal system error
FUSE DRAWER
NOTCHES
Appendix A
Specifications and Ordering Information
Input Power
nominal
actual
Fuses
Input Power Cord
Output (Programmable)
Voltage
Voltage Accuracy
Current
Current Accuracy
Power (maximum)
Terminals
Timer Control
Ripple
Line Regulation
Load Regulation
Drift
Noise
Readout Stability
Volts
Current
Safety Features
No load detection
Sudden load change detection
Overload/short circuit protection
Input line protection
Auto power up after power failure
Safety Compliance
EMI
Display Functions
Function Modes
Environmental
Operating Temperature
Humidity
Dimensions
Unit is stackable
Weight
100120/220240 VAC, 50 or 60 Hz
90132 or 198264 VAC, 50 or 60 Hz
2.5 A, 250 VAC, 5mm x 20mm, Type T
3-wire; grounded
10 V to 300 V, fully adjustable in 1 V steps
2% of reading or 3 volts, whichever is more
4 mA to 400 mA, fully adjustable in 1 mA steps
2% of reading or 3 mA, whichever is more
75 W
4-pair recessed banana jacks, floating in parallel
001 to 999 minutes, fully adjustable
< 1% @ 300 V and 70 W
< 1% @ 300 V and 70 W, 90132 or
198264VAC
<1% @ 300 V for a 50% change in output load
< 1% @ after 15 min warmup at 300 V and 70 W
55 dBA at 1 meter
1 V
1 mA
Indicated by error message on display
Indicated by error message on display
Automatic power limit
Fuse on both hot and neutral
User-selectable, setup values maintained
EN-61010
Conforms to CE standards for Emissions and
Immunity class A, tested only at 230V. See
Declaration of Conformity for details. TUV EMC
certification
3-digit LED displays voltage, current, or time
Constant voltage, constant current, timer, pause
User-selectable automatic power-up after power
failure
040C
095%, in the absence of condensation
25 cm (L) x 21 cm (W) x 8.5 cm (H)
1.1 kg
10
Appendix B
Warranty and Ordering Information
Warranty
The PowerPac Basic power supply is covered by a standard Bio-Rad Laboratories
warranty. Contact your local Bio-Rad representative for details of the warranty. If any
defects should occur during this warranty period, Bio-Rad Laboratories will replace the
defective parts without charge. However, the following defects are specifically excluded:
1. Defects caused by improper operation.
2. Repair or modification done by anyone other than Bio-Rad Laboratories or their
authorized agent.
3. Use with cables or connectors not specified by Bio-Rad Laboratories for this power
supply.
4. Deliberate or accidental misuse.
5. Damage caused by disaster.
For inquiry or request for repair service, contact your local Bio-Rad office.
Warranty Information
Model: _____________________________________________________________________
Serial Number: ______________________________________________________________
Date of Delivery: _____________________________________________________________
Warranty Period: _____________________________________________________________
Ordering Information
Catalog
Number
Description
164-5050
165-5061
165-5066
900-7283
11
Bio-Rad
Laboratories, Inc.
Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austria 01 877 89 01 Belgium 09 385 55 11 Brazil 55 11 5044 5699
Canada 905 364 3435 China 86 21 6169 8500 Czech Republic 420 241 430 532 Denmark 44 52 10 00 Finland 09 804 22 00
France 01 47 95 69 65 Germany 089 31 884 0 Greece 30 210 9532 220 Hong Kong 852 2789 3300 Hungary 36 1 459 6100 India 91 124 4029300
Israel 03 963 6050 Italy 39 02 216091 Japan 03 6361 7000 Korea 82 2 3473 4460 Mexico 52 555 488 7670 The Netherlands 0318 540666
New Zealand 64 9 415 2280 Norway 23 38 41 30 Poland 48 22 331 99 99 Portugal 351 21 472 7700 Russia 7 495 721 14 04
Singapore 65 6415 3188 South Africa 27 861 246 723 Spain 34 91 590 5200 Sweden 08 555 12700 Switzerland 026 674 55 05
Taiwan 886 2 2578 7189 Thailand 800 88 22 88 United Kingdom 020 8328 2000