Nothing Special   »   [go: up one dir, main page]

HI 2210 HI 2211: Microprocessor-Based pH/mV/°C Bench Meters

Download as pdf or txt
Download as pdf or txt
You are on page 1of 15

Instruction Manual

HI 2210 HI 2211
Microprocessor-based
pH/mV/C
Bench Meters

w w w. h a n n a i n s t . c o m

Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the instruments.
This manual will provide you with the necessary information for correct use of
the instruments, as well as a precise idea of their versatility.
If you need additional technical information, do not hesitate to e-mail us at
tech@hannainst.com or view our worldwide contact list at www.hannainst.com.
These instruments are in compliance with
directives.

WARRANTY
HI 2210 and HI 2211 are guaranteed for two years against defects
in workmanship and materials when used for their intended purpose
and maintained according to instructions. Electrodes and probes are
guaranteed for six months. This warranty is limited to repair or
replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the problem. If the repair
is not covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization number from the Technical
Service department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packed for
complete protection.

TABLE OF CONTENTS
WARRANTY .................................................................................. 3
PRELIMINARY EXAMINATION ........................................................ 4
GENERAL DESCRIPTION ................................................................ 4
FUNCTIONAL DESCRIPTION .......................................................... 5
SPECIFICATIONS ........................................................................... 6
OPERATIONAL GUIDE ................................................................... 7
pH CALIBRATION ....................................................................... 10
TEMPERATURE CALIBRATION (for technical personnel only) ........... 13
pH BUFFER TEMPERATURE DEPENDENCE ...................................... 15
ELECTRODE CONDITIONING & MAINTENANCE ............................... 16
TROUBLESHOOTING GUIDE ........................................................ 19
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS ................. 20
ACCESSORIES ............................................................................. 21

PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during shipping.
If there is any damage, notify your Dealer or the nearest Hanna
Customer Service Center.
Each instrument is supplied with:
HI 1131B Glass-body Combination pH Electrode
HI 7662 Temperature Probe
HI 76404 Electrode Holder
pH 4.01 & 7.01 Buffer Solutions, 20 mL each
HI 7071S Electrolyte Solution
12VDC Power Adapter
Instruction Manual
Note: Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing with the supplied accessories.

FUNCTIONAL DESCRIPTION
Front Panel

Rear Panel

GENERAL DESCRIPTION
The Hanna HI 2210 and HI 2211 are microprocessor-based pH and
temperature bench meters.
HI 2211 can also be used for ion concentration (ISE) and Oxidation
Reduction Potential (ORP) in the mV range.
pH measurements are compensated for temperature effect manually or
automatically with the HI 7662 temperature probe.
The instrument is equipped with a large easy-to-read LCD which shows
the pH (or mV) and temperature simultaneously, together with graphic
symbols.
The calibration process is guided step by step through clear indications on
the LCD. A stability indicator makes the calibration procedure error-free.

Hanna Instruments reserves the right to modify the design,


construction and appearance of its products without advance notice.

1) Liquid Crystal Display (LCD).


2) CAL key, to enter or exit/escape calibration mode.
3) CFM key, to confirm different values.
4)
C key and C keys, to manually increase/decrease temperature or select pH buffer.
5) RANGE key, to select measurement range (HI 2211 only).
6) MEM key, to store a value into memory.
7) MR key, to recall the stored value.
8) Secondary LCD.
9) Primary LCD.
10) ON/OFF switch.
11) Power adapter socket.
12) BNC electrode connector.
13) Temperature probe socket.
14) Electrode reference socket.

OPERATIONAL GUIDE

SPECIFICATIONS

POWER CONNECTION
2.00 to 16.00 pH
399.9 mV (HI 2211 only)
2000 mV (HI 2211 only)

RANGE

9.9 to 120.0 C
0.01 pH
0.1 mV (HI 2211 only)
1 mV (HI 2211 only)

RE SOLUTION

0.1 C
0.01 pH
0.2 mV (HI 2211 only)
1 mV (HI 2211 only)

ACCURACY
@ 20C / 68F

0.5 C (0.0 100.0 C)


1 C (outside)
(excluding probe error)

pH Calibration

1 or 2 point calibration, 5 buffers


a va ila b le
(4.01, 6.86, 7.01, 9.18, 10.01)

Temperature compensation

Manual or Automatic from:


9.9 to 120.0 C (14.2 to 248.0 F)

pH E lectrode

HI 1131B (included)

Temperature probe

HI 7662 (included)

Input impedance

1012 ohms

Power supply

12 VD C adapter (included)

D imensions

240x182x74 mm (9.4x7.1x2.9)

Weight

1.1 Kg (2.5 lb);


kit with holder 2.5 Kg (5.5 lb)

E nvironment

0 to 50 C (32 to 122 F)
max. 95% RH non-condensing

Warranty

2 years

Plug the 12 VDC adapter into the power supply socket.


Notes: These instruments use non volatile memory to retain the
pH, mV, temperature calibrations and all other settings,
even when unplugged.
Make sure a fuse protects the main line.
ELECTRODE AND PROBE CONNECTIONS
For pH or ORP combination electrode connect to the BNC connector on
the back of the instrument.
For electrodes with a separate reference connect the electrodes BNC to the
BNC connector and the reference electrode plug to the reference socket.
For temperature measurements and automatic temperature compensation connect the temperature probe to the appropriate socket.
INSTRUMENT START-UP
Turn the instrument on by pressing the ON/OFF switch located
on the rear panel.
All LCD segments are displayed while the instrument performs a
self test.

pH

MEASUREMENTS

Make sure the electrode and the instrument have been


calibrated together before taking pH measurements.
Submerge the electrode and the temperature
probe approximately 4 cm (1) into the sample
to be tested and stir gently. Allow time for the
electrode to stabilize.
The pH is displayed on the primary LCD and the temperature on
the secondary LCD.
pH
C

If the pH reading is out of range, ---- will be displayed on the


LCD.
6

If measurements are taken successively in different samples, it is


recommended to rinse the electrode thoroughly with deionized water
or tap water and then with some of the next sample in order to
prevent cross-contamination.
The pH reading is affected by temperature. In order to measure the
pH accurately, the temperature effect must be compensated for.
To use the Automatic Temperature Compensation feature, connect
and submerge the HI 7662 temperature probe into the sample as
close as possible to the electrode and wait for a few seconds.
If the temperature of the sample is known, manual temperature
compensation can be performed by disconnecting the temperature probe.
The display will then show the default temperature
of 25 C or the last recorded temperature reading
with the C tag blinking. The temperature can now
be adjusted with the ARROW keys (from -9.9 C to
120.0 C).

TEMPERATURE MEASUREMENTS
Connect the HI 7662 temperature probe to the TEMP
socket and turn the instrument on.
Immerse the temperature probe into the sample and
allow the reading on the secondary LCD to stabilize.
MEMORY FUNCTION
Press and hold down MEM. The last reading will appear on the LCD
along with MEM tag until MEM is released.

Press MR and the previously memorized reading will be displayed


with MEM tag on the LCD.

ORP MEASUREMENTS (H
211 only
(HII 222
only))
An optional ORP electrode must be used to perform ORP measurements
(see Accessories).
Oxidation-reduction potential (REDOX) measurements provide the
quantification of the oxidizing or reducing power of the tested sample.
To correctly perform a redox measurement, the surface
of the ORP electrode must be clean and smooth.
Press RANGE to enter mV range.
Submerge the tip of the ORP electrode 4 cm
(1") into the sample to be tested and allow a
few seconds for the reading to stabilize.
The instrument displays the mV reading on the primary LCD and
the temperature on the secondary LCD.
C

mV

If the reading is out of range, ---- will be displayed on the


LCD.
8

p H CALIBRATION
Calibrate the instrument often, especially if high accuracy is required.
The instrument should be recalibrated:
Whenever the pH electrode is replaced.
At least once a week.
After testing aggressive chemicals.
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If
possible use plastic or glass beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two
beakers for each buffer solution: one for rinsing the electrode and one for
calibration.
If you are measuring in the acidic range, use pH 7.01 as first buffer
and pH 4.01 as second buffer. If you are measuring in the alkaline
range, use pH 7.01 as first buffer and pH 10.01 or pH 9.18 as
second buffer.
PROCEDURE
Calibration has a choice of five memorized buffers: pH 4.01, 6.86,
7.01, 9.18 and 10.01.
It is recommended to perform a two-point calibration. However, onepoint calibration is also permitted by the instruments.
TWO-POINT CALIBRATION
Submerge the pH electrode and the temperature probe approximately 4 cm (1") into a
buffer solution and stir gently. The temperature
probe should be close to the pH electrode.
Press CAL. The CAL and messages will appear and
7.01 buffer will be displayed on the secondary LCD.
pH

CAL

BUFFER pH
1

CAL

BUF

If necessary, press the ARROW keys to select a different buffer value.

The symbol will blink on the LCD until the reading is stable.
10

When the reading is stable and close to the selected buffer,


READY message will appear and CFM message will blink.
Press CFM to confirm calibration.
The calibrated value is then displayed on the
primary LCD and the secondary LCD will display
the second expected buffer value.
pH

CAL

BUFFER pH
BUF 2

Note: The instruments will automatically skip the buffer used for the
first point. It also skips 6.86 if 7.01 buffer was used and vice
versa. Likewise, it will skip 9.18 if 10.01 buffer was used and
vice versa.
After the first calibration point is confirmed,
immerse the pH electrode and the temperature
probe approximately 4 cm (1) into the
second buffer solution and stir gently. The
temperature probe should be close to the pH
electrode.
If necessary, press the ARROW keys to select a different buffer value.

The symbol will blink on the LCD until the reading is stable.
When the reading is stable, READY message will appear and
CFM message will blink.
Press CFM to confirm calibration. The instrument
will return to measurement mode.
Notes: If the value measured by the meter is not close to the
selected buffer, WRONG and WRONG messages
will blink alternately. In this case check if the correct buffer
has been used, or regenerate the electrode by following the
cleaning procedure (see page 18). If necessary, change the
buffer or the electrode.
The WRONG message and temperature value are displayed blinking if the temperature reading is out of the
defined temperature range of the buffer. Calibration cannot
be confirmed in this situation.

11

Press RANGE to display the temperature reading on the LCD during


calibration (HI 2211 only).
ONE-POINT CALIBRATION
Proceed as described in Two-point calibration section.
Press CAL after the first calibration point
was confirmed.
The instrument will return to measurement mode
and will memorize the one-point calibration data.
Note: Press and hold down CFM and then press CAL. The instrument
sets default calibration parameters, displays CLR message for a
few seconds and then returns to normal measurement mode.

TEMPERATURE CALIBRATION
(for technical personnel only)
All the instruments are factory calibrated for temperature.
Hannas temperature probes are interchangeable and no temperature calibration is needed when they are replaced.
If the temperature measurements are inaccurate, temperature
recalibration should be performed.
For an accurate recalibration, contact your dealer or the nearest
Hanna Customer Service Center, or follow the instructions bellow.
Prepare a vessel containing ice and water and another one
containing hot water (at a temperature of around 50 C). Place
insulation material around the vessels to minimize temperature
changes.
Use a calibrated thermometer with a resolution of 0.1 C as a
reference thermometer.
With the instrument off, press and hold down the CAL & MEM
keys, then power on the instrument. The CAL message will
appear and the secondary LCD will show 0.0 C.
CAL
C

Immerse the temperature probe in the vessel with ice and water
as near as possible to the reference thermometer. Allow a few
seconds for the probe to stabilize.
Use the ARROW keys to set the reading on the secondary LCD to
that of ice and water, measured by the reference thermometer.

When the reading is stable and close to the selected calibration


point, READY message will appear and CFM message will blink.
To confirm press CFM. The secondary LCD will show 50.0 C.
CAL
C

CFM

12

13

Immerse the temperature probe in the second vessel as near as


possible to the reference thermometer. Allow a few seconds for the
probe to stabilize.
Use the ARROW keys to set the reading on the secondary LCD to
that of the hot water.

p H BUFFER TEMPERATURE
DEPENDENCE
The temperature has an effect on pH. The calibration buffer solutions
are affected by temperature changes to a lesser degree than normal
solutions. During calibration the instrument will automatically calibrate
to the pH value corresponding to the measured or set temperature.
TEMP

When the reading is stable and close to the selected calibration


point, READY tag will appear and CFM tag will blink.
Press CFM to confirm. The instrument returns
to measurement mode.
Note: If the reading is not close to the selected
calibration point, WRONG tag will blink.
Change the temperature probe and restart
calibration.

pH BUFFERS

4. 01

6. 86

7. 01

9. 18

10. 01

32

4.01

6.98

7.13

9.46

10.32

41

4.00

6.95

7.10

9.39

10.24

10

50

4.00

6.92

7.07

9.33

10.18

15

59

4.00

6.90

7.05

9.27

10.12

20

68

4.00

6.88

7.03

9.22

10.06

25

77

4.01

6.86

7.01

9.18

10.01

30

86

4.02

6.85

7.00

9.14

9.96

35

95

4.03

6.84

6.99

9.11

9.92

40

104

4.04

6.84

6.98

9.07

9.88

45

113

4.05

6.83

6.98

9.04

9.85

50

122

4.06

6.83

6.98

9.01

9.82

55

131

4.08

6.84

6.98

8.99

9.79

60

140

4.09

6.84

6.98

8.97

9.77

65

149

4.11

6.84

6.99

8.95

9.76

70

158

4.12

6.85

6.99

8.93

9.75

75

167

4.14

6.86

7.00

8.91

9.74

80

176

4.16

6.87

7.01

8.89

9.74

85

185

4.17

6.87

7.02

8.87

9.74

90

194

4.19

6.88

7.03

8.85

9.75

95

203

4.20

6.89

7.04

8.83

9.76

During calibration the instrument will display the pH buffer value at


25 C.
14

15

ELECTRODE CONDITIONING &


MAINTENANCE

For refillable electrodes:


If the filling solution (electrolyte) is more than 2 cm (1") below the
fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for
double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte
Solution for single junction electrodes.
Unscrew the fill hole screw during measurements.
For AmpHel electrodes:
If the electrode does not respond to pH changes, the battery is dead
and the electrode should be replaced.
MEASUREMENT
Rinse the electrode tip with distilled water. Immerse the tip (bottom
4 cm /1) into the sample and stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested,
before taking measurements.
STORAGE PROCEDURE
To minimize clogging and assure a quick response time, the glass bulb
and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of HI 70300
or HI 80300 Storage Solution or, in its absence, Fill Solution (HI
7071 or HI 8071 for single junction and HI 7082 or HI 8082 for
double junction electrodes). Follow the Preparation Procedure on page
16 before taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED
WATER.
PERIODIC MAINTENANCE

PREPARATION PROCEDURE
Remove the protective cap of the pH electrode.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal
with electrodes. They will disappear when rinsed with water.
During transport, tiny bubbles of air may form inside the glass bulb
affecting proper functioning of the electrode. These bubbles can be
removed by shaking down the electrode as you would do with a
glass thermometer.
If the bulb and/or junction is dry, soak the electrode in HI 70300 or
HI 80300 Storage Solution for at least one hour.

16

Inspect the electrode and the cable. The cable must be intact and
well connected. No cracks should be seen on the electrode stem or
bulb. Connectors must be perfectly clean and dry. If any scratches or
cracks are present, replace the electrode. Rinse off any salt deposits
with water.
For refillable electrodes:
Refill the reference chamber with fresh electrolyte (HI 7071 or HI 8071
for single junction or HI 7082 or HI 8082 for double junction
electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.

17

CLEANING PROCEDURE

TROUBLESHOOTING GUIDE

General

Soak in Hanna HI 7061 or HI 8061 General


Cleaning Solution for approximately hour.
Protein
Soak in Hanna HI 7073 or HI 8073 Protein
Cleaning Solution for 15 minutes.
Soak in Hanna HI 7074 Inorganic Cleaning
Inorganic
Solution for 15 minutes.
Oil/grease Rinse with Hanna HI 7077 or HI 8077 Oil and
Fat Cleaning Solution.
IMPORTANT: After performing any of the cleaning procedures, rinse
the electrode thoroughly with distilled water, refill the reference
chamber with fresh electrolyte (not necessary for gel-filled electrodes)
and soak the electrode in HI 70300 or HI 80300 Storage Solution for
at least 1 hour before taking measurements.

18

SYMPTOMS
Slow reponse/excessive
drift.

PROBLE M
D irty pH electrode.

SOLUTION
Clean the electrode and
then soak the tip in HI
7061 or HI 8061 solution
for 30 minutes.

Readings fluctuate up Clogged/dirty junction.


and down (noise).
Low electrolyte level
(refillable electrodes
only).

Clean the electrode.


Refill with fresh solution
(for refillable electrodes
only). Check cable and
connector.

Out of range in the


mV scale.

D ry
membrane/junction.

Soak in HI 70300 or
HI 80300 storage
solution. Check cable and
connector.

The meter does not


accept the buffer
solution for
calibration.

pH electrode damaged.
Wrong buffer used.

Follow the cleaning


procedure. If still no
results, replace the
electrode. Replace Buffer.

The display shows


"pH" and "----".

Out of range in the pH a) Verify that the electrode


scale.
is connected.
b) Verify that the
shipping cap has been
removed.
c) Recalibrate the meter.
d) Make sure the pH
sample is in the specified
range.
e) Check the electrolyte
level and the general
state of the electrode.

The display shows


"mV" and "----".

Out of range in the


mV scale.

Verify that the electrode


is connected.

The meter does not


work with the
temperature probe.

Broken temperature
probe.
Wrong temperature
probe used.

Replace the temperature


probe.

The meter fails to


calibrate or gives
faulty readings.

Broken pH electrode.

Replace the electrode.

At startup the meter


displays all LCD tags
permanently.

One of the keys is


stuck.

Check the keyboard or


contact the vendor.

"E rr xx" error message


displayed.

Internal error.

Power off the meter and


then power it on. If the
error persists, contact the
vendor.

19

TEMPERATURE CORRELATION
FOR p H SENSITIVE GLASS
The resistance of glass electrodes partially depends on the temperature.
The lower the temperature, the higher the resistance. It takes more time
for the reading to stabilize if the resistance is higher. In addition, the
response time will suffer to a greater degree at temperatures below 25 C.

Since the resistance of the pH electrode is in the range of 50-200 Mohms,


the current across the membrane is in the pico Ampere range. Large
currents can disturb the calibration of the electrode for many hours.
For these reasons high humidity environments, short circuits and
static discharges are detrimental to a stable pH reading.
The pH electrodes life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.
Typical Electrode Life
Ambient Temperature
1- 3 years
90 C
Less than 4 months
120 C
Less than 1 month
Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline
solutions. The pH at which the interference starts to be significant
depends upon the composition of the glass. This interference is called
alkaline error and causes the pH to be underestimated. Hannas glass
formulations have the indicated characteristics.
Sodium Ion Correction for Glass at 20-25 C
Concentration
pH
13.00
0.1 Mol L-1 Na+
13.50
14.00
12.50
13.00
1.0 Mol L-1 Na+
13.50
14.00
20

Error
0.10
0.14
0.20
0.10
0.18
0.29
0.40

ACCESSORIES
pH BUFFER SOLUTIONS
HI 70004P
HI 70007P
HI 70010P
HI 7004L
HI 7006L
HI 7007L
HI 7009L
HI 7010L
HI 8004L
HI 8006L
HI 8007L
HI 8009L
HI 8010L

pH 4.01 Buffer Sachets, 20 mL, 25 pcs


pH 7.01 Buffer Sachets, 20 mL, 25 pcs
pH 10.01 Buffer Sachets, 20 mL, 25 pcs
pH 4.01 Buffer Solution, 500 mL bottle
pH 6.86 Buffer Solution, 500 mL bottle
pH 7.01 Buffer Solution, 500 mL bottle
pH 9.18 Buffer Solution, 500 mL bottle
pH 10.01 Buffer Solution, 500 mL bottle
pH 4.01 Buffer Sol. in FDA approved bottle, 500 mL
pH 6.86 Buffer Sol. in FDA approved bottle, 500 mL
pH 7.01 Buffer Sol. in FDA approved bottle, 500 mL
pH 9.18 Buffer Sol. in FDA approved bottle, 500 mL
pH 10.01 Buffer Sol. in FDA approved bottle, 500 mL

ELECTRODE STORAGE SOLUTIONS


HI 70300L Storage Solution, 460 mL bottle
HI 80300L Storage Solution in FDA approved bottle, 460 mL
ELECTRODE CLEANING SOLUTIONS
HI 70000P
HI 7061L
HI 7073L
HI 7074L
HI 7077L
HI 8061L
HI 8073L
HI 8077L

Electrode Rinse Sachets, 20 mL, 25 pcs


General Cleaning Solution, 460 mL bottle
Protein Cleaning Solution, 460 mL bottle
Inorganic Cleaning Solution, 460 mL bottle
Oil & Fat Cleaning Solution, 460 mL bottle
General Cleaning Sol. in FDA approved bottle, 460 mL
Protein Cleaning Solution in FDA approved bottle, 460 mL
Oil & Fat Cleaning Sol. in FDA approved bottle, 460 mL

ELECTRODE REFILL ELECTROLYTE SOLUTIONS


HI 7071
HI 7072
HI 7082
HI 8071
HI 8072
HI 8082

3.5M KCl+AgCl Electrolyte, 4x50 mL, for single junction


electrodes
1M KNO3 Electrolyte, 4x50 mL
3.5M KCl Electrolyte, 4x50 mL, for double junction
electrodes
3.5M KCl+AgCl Electrolyte in FDA approved bottle,
4x50 mL, for single junction electrodes
1M KNO3 Electrolyte in FDA approved bottle, 4x50 mL
3.5M KCl Electrolyte in FDA approved bottle, 4x50 mL
21

ORP PRETREATMENT SOLUTIONS


HI 7091L
Reducing Pretreatment Solution, 460 mL bottle
HI 7092L
Oxidizing Pretreatment Solution, 460 mL bottle

pH ELECTRODES
All electrodes part numbers ending with B are supplied with BNC
connector and 1 m (3.3') cable, as shown below:

HI 1330B
Glass-body, semimicro, single junction, refillable, combination pH
electrode. Use: laboratory, vials.
5mm DIA
0.2"

5mm
0.2"

HI 1330
120 mm
4.7"
"S" VERSION

HI 1043B
Glass-body, double junction, refillable, combination pH electrode.
Use: strong acid/alkali.
9.5mm DIA
0.37"

12 mm
0.5"

8 mm
0.3"

7.5mm DIA
0.29"

HI 1331
210 mm
8.25"

"S" VERSION

HI 1043
120 mm
4.7"

"S" VERSION

HI 1053B
Glass-body, triple ceramic, conic shape, refillable, combination pH
electrode. Use: emulsions.
12 mm
0.5"

HI 1053

HI 1230B
Plastic-body (PEI), double junction, gel-filled, combination pH electrode. Use: general, field.
12 mm
0.5"

HI 1230

"S" VERSION

120 mm
4.7"

"S" VERSION

HI 1083B
Glass-body, micro, Viscolene, non-refillable, combination pH electrode.
Use: biotechnology, micro titration.
12 mm
0.5"

HI 1331B
Glass-body, semimicro, single junction, refillable, combination pH
electrode. Use: flasks.

5 mm
0.2"

3 mm
0.12"

3.0 mm DIA
0.12"

120 mm
4.7"

HI 2031B
Glass-body, semimicro, conic, refillable, combination pH electrode. Use:
semisolid products.
6 mm
0.25"

HI 2031

HI 1083
120 mm
4.7"

75 mm
2.95"

"S" VERSION

HI 1131B
Glass-body, single junction, refillable, combination pH electrode.
Use: general purpose.
9.5mm DIA
0.37"

HI 1332B
Plastic-body (PEI), double junction, refillable, combination pH electrode. Use: general purpose.

12 mm
0.5"

12 mm
0.5"

HI 1131
HI 1332

120 mm
4.7"

"S" VERSION

22

"S" VERSION

120 mm
4.7"

23

FC 100B
Plastic-body (PVDF), double junction, refillable, combination pH
electrode. Use: general purpose for food industry.

HI 1413B
Glass-body, single junction, flat tip, Viscolene, non-refillable, combination
pH electrode. Use: surface measurement.
12 mm
0.5"

12 mm
0.5"
FC 100

HI 1413
120 mm
4.7"

110 mm
4.3"

FC 200B
Plastic-body (PVDF), open junction, conic, Viscolene, non-refillable,
combination pH electrode. Use: meat & cheese.
6 mm
0.25"

ORP ELECTRODES
HI 3131B
Glass-body, refillable, combination platinum ORP electrode.
Use: titration.

FC 200

12 mm
0.5"

75 mm
2.95"

HI 3131

FC 210B
Glass-body, double junction, conic, Viscolene, non-refillable, combination pH electrode. Use: milk, yogurt.

"S" VERSION

150 mm
5.9"

HI 3230B
Plastic-body (PEI), gel-filled, combination platinum ORP electrode.
Use: general purpose.

12 mm
0.5"

12 mm
0.5"

FC 210

HI 3230
120 mm
4.7"

FC 220B
Glass-body, triple-ceramic, single junction, refillable, combination pH
electrode. Use: food processing.
9.5mm DIA
0.37"

12 mm
0.5"

120 mm
4.7"

"S" VERSION

HI 4430B
Plastic-body (PEI), gel-filled, combination gold ORP electrode. Use:
general purpose.
12 mm
0.5"

FC 220

HI 4430
120 mm
4.7"

FC 911B
Plastic-body (PVDF), double junction, refillable with built-in amplifier,
combination pH electrode. Use: very high humidity.

120 mm
4.7"

"S" VERSION

Consult the Hanna General Catalog for more electrodes with screwtype or BNC connectors.

12 mm
0.5"

FC 911
110 mm
4.3"

24

25

EXTENSION CABLE FOR SCREW-TYPE ELECTRODES


(SCREW TO BNC ADAPTER)
HI 7855/1 Extension cable 1 m (3.3') long
HI 7855/3 Extension cable 3 m (9.9') long
HI 7855 SERIES CABLE CONNECTORS
CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC

CONNECT TO THE
BNC SOCKET
OF THE METER

CONNECT TO
SCREW TYPE
ELECTRODES

OTHER ACCESSORIES
HI 710005
HI 710006
HI 710012
HI 710013
HI 710014
HI 76405

Voltage adapter from 115 VAC to 12 VDC (USA plug)


Voltage adapter from 230 VAC to 12 VDC (European plug)
Voltage adapter from 240 VAC to 12 VDC (UK plug)
Voltage adapter from 230 VAC to 12 VDC (South Africa plug)
Voltage adapter from 230 VAC to 12 VDC (Australia plug)
Electrode holder

RECOMMENDATIONS FOR USERS


Before using these products, make sure they are entirely suitable for
the environment in which they are used.
Operation of these instruments in residential areas could cause
unacceptable interferences to radio and TV equipment, requiring the
operator to follow all necessary steps to correct interferences.
The glass bulb at the end of the pH electrode is sensitive to
electrostatic discharges. Avoid touching this glass bulb at all times.
During operation, ESD wrist straps should be worn to avoid possible
damage to the electrode by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may
degrade the instruments EMC performance.
To avoid electrical shock, do not use these instruments when voltages
at the measurement surface exceed 24 VAC or 60 VDC.
To avoid damage or burns, do not perform any measurement in
microwave ovens.

pH and ORP electrode simulator with 1 m (3.3') coaxial


cable ending in female BNC connectors
HI 931001 pH and ORP electrode simulator with LCD and 1 m (3.3')
coaxial cable ending in female BNC connectors
HI 7662 Temperature probe with 1 m (3.3') cable
HI 8427

Windows is registered Trademark of "Microsoft Co."

26

27

SALES AND TECHNICAL SERVICE CONTACTS


Australia:
Tel. (03) 9769.0666 Fax (03) 9769.0699

China:
Tel. (10) 88570068 Fax (10) 88570060

Egypt:
Tel. & Fax (02) 2758.683

Germany:
Tel. (07851) 9129-0 Fax (07851) 9129-99

Greece:
Tel. (210) 823.5192 Fax (210) 884.0210

Indonesia:
Tel. (21) 4584.2941 Fax (21) 4584.2942

Japan:
Tel. (03) 3258.9565 Fax (03) 3258.9567

Korea:
Tel. (02) 2278.5147 Fax (02) 2264.1729

Malaysia:
Tel. (603) 5638.9940 Fax (603) 5638.9829

Singapore:
Tel. 6296.7118 Fax 6291.6906

South Africa:
Tel. (011) 615.6076 Fax (011) 615.8582

Taiwan:
Tel. 886.2.2739.3014 Fax 886.2.2739.2983
MAN2211R1
02/08

Thailand:

Tel. (662) 619.0708.11 Fax (662) 619.0061

United Kingdom:
Tel. (01525) 850.855 Fax (01525) 853.668

USA:
Tel. (401) 765.7500 Fax (401) 765.7575
For e-mail contacts and a complete list of Sales and
Technical offices, please see www.hannainst.com.

28

You might also like