LNE G-MED CE Guide
LNE G-MED CE Guide
LNE G-MED CE Guide
CE MARKING
OF MEDICAL DEVICES
Edition 2012
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
Table of Contents
1. Introduction to the guide ................................................ 3
....................................................................................................... 7
7. Conclusion ........................................................................ 23
Annexes ................................................................................ 24
4.3.1 EC Verification....................................................................... 17
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
93/42/EEC
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
93/42/EEC
98/79/EC
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
any medical device which is a reagent, regent product, calibrator, control material, kit,
instrument, apparatus, equipment, or system, whether used alone or in combination, intended
by the manufacturer to be used in vitro for the examination of specimens, including blood and
tissue donations, derived from the human body, solely or principally for the purpose of providing
information:
concerning a physiological or pathological state, or
concerning a congenital abnormality, or
to determine the safety and compatibility with potential recipients, or
to monitor therapeutic measures.
Specimen receptacles are considered to be in vitro diagnostic medical devices. Specimen
receptacles are those devices, whether vacuum-type or not, specifically intended by their
manufacturers for the primary containment and preservation of specimens derived from the
human body for the purpose of in vitro diagnostic examination.
Products for general laboratory use are not in vitro diagnostic medical devices unless such
products, in view of their characteristics, are specifically intended by their manufacturer to be used
for in vitro diagnostic examination.6
Further details on the interpretations of these definitions are detailed in the following MEDDEV guides
(cf 2.3):
MEDDEV 2.1/1
MEDDEV 2.1/2
MEDDEV 2.14/1
MEDDEV 2.14/2
Other devices that are not considered medical devices are:
Medicinal products covered by Directive 2001/83/CE;
Cosmetic products covered by Directive 76/768/EEC;
Human blood and human blood products, human plasma or blood cells of human origin or to
devices which incorporate at the time of placing on the market such blood products, plasma, or
cells;
Human transplants, tissues, cells, or products incorporating or derived from human tissues or cells;
Animal origin transplants, tissues, or cells unless a device is manufactured using animal tissues
rendered non-viable or non-viable products derived from animal tissues.
For the record, based on 14 years of experience, the regulatory recast that began in 2012 confirmed the
extension of the medical device regulations of medical devices to certain products without direct medical
purpose. These are products for esthetic purposes whose use may represent health hazards: non-corrective
contact lenses, implants for augmentation, fixation or sculpting of body parts, skin fillers, liposuction devices,
and lasers.
Regarding the field of diagnostic medical devices in vitro, software has been confirmed as part of the scope
of the draft, together with, for example, products intended to screen for predisposition to a medical condition
or disease.7
Editors note: this data may change during the year 2012.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
2.3 What can a manufacturer do when the regulatory texts dont answer his or her questions Where to find interpretations?
Despite the drafters best intentions, the reality of regulations and legislations is that sometimes they need to
be clarified to be understood and applied in a consistent manner. To help readers, the European Commission
develops guides and interpretive texts for the Directives to clarify some points. These guides are drafted by
experts from different Member States, although it should be noted that they arent legal texts as such. But
these guides provide significant help in interpreting, applying, and avoiding mistakes with the Directives.
These texts, the MEDDEV guides, are interpretive documents or consensus statements that do not have a
regulatory status. Therefore, the basis of a decision or action remains the Directives themselves and their
implementation into national law of the Member States. However, their use represents an important aid for
interpreting and applying the Directives and avoiding mistakes with very damaging consequences
These guides can be found on the European Commissions website,8 and others on NBOGs website.9
Finally, it is worth mentioning the important guide to the European Commission intended to explain the
concepts of the New Approach. This guide, the Blue Guide, covers the placing all CE marked products on
the European market and is not limited to medical devices.10
The experts from LNE/G-MED are able to help to analyze the contents of these different guides.
8
http://ec.europa.eu/health/medical-devices/regulatory-framework/index_en.htm.
http://ec.europa.eu/health/medical-devices/documents/consensus-statements/index_en.htm
http://ec.europa.eu/health/medical-devices/documents/interpretative-documents/index_en.htm
9
http://nbog.eu/index.html
10
http://ec.europa.eu/enterprise/policies/single-market-goods/files/blue-guide/guidepublic_en.pdf
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
2.4.1 Manufacturer
The manufacturer takes the responsibility for placing the medical device on the market. The Directives
define a manufacturer as:
The natural or legal person with responsibility for the design, manufacture, packaging and labelling
of a device before it is placed on the market under his own name, regardless of whether these
operations are carried out by that person himself or on his behalf by a third party.
The obligations of this directive to be met by manufacturers also apply to the natural or legal person
who assembles, packages, processes, fully refurbishes and/or labels one or more ready-made
products and/or assigns to them their intended purpose as devices with a view to their being placed
on the market under his own name. This subparagraph does not apply to the person who, while not
a manufacturer within the meaning of the first subparagraph, assembles or adapts devices already
on the market to their intended purpose for an individual patient.
If a manufacturer has no registered offices in Europe, it must appoint an authorized representative in one
of the Member States. The manufacturer could then instruct the authorized representative to take over
some of the conformity assessment procedures. So, the authorized representative would seek the help of a
notified body at the manufacturers request.
In the context of fulfilling its statutory responsibilities, the manufacturer must pay particular attention to its
relations with the subcontractors involved in the design and manufacturer of the product, as well as potential
distributors. A proxy must be appointed if the headquarters of the manufacturer is not the territory of the
European Economic Space or Switzerland.11
2.4.2 Other Parties
Sometimes, in specific situations, other parties are also responsible for putting a device on the market
along with the manufacturer. These parties are:
Authorized Representative of the manufacturer is a legal person established in the EEC who acts as
the face of the manufacturer to the Competent Authorities. More information about the Authorized
Representative, its role, and its responsibilities can be found in MEDDEV 2.5/10.
Subcontractors or suppliers of the manufacturer who may be involved when establishing the conformity of
the device. Information about the control of the suppliers and subcontractors may be found in the NBOG
Best Practice Guide 2010-1.
Distributors can also be involved when:
It is the European-based importer for a non-European manufacturer; and
The distributor can be intermediary for transmitting post-market surveillance information
While it may seem confusing, LNE/G-MED is here to help discern which parties must be involved in a devices
certification. To establish which parties must be involved in your devices certification, contact LNE/G-MED
to speak with a project manager or one of our regulatory experts.
11
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
12
http://www.team-nb.org/.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
Identifying the Manufacturer and, where applicable, its agent, its subcontractors and distributors;
Characterizing the products;
Determining applicable Directives;
Determining product classes for Directives 93/42/EEC and 98/79/EC;
Selecting the most appropriate procedures to determine compliance - Methods of proof
Collecting necessary data for the chosen procedures or, in particular for meeting the demands of
certain horizontal provisions (risk analysis, clinical evaluation, performance evaluation (IVD) etc.);
and
7. Consulting LNE / G-MED once the appropriate procedure is decided.
These milestones are the subject of the sub-chapters below.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
10
Duration of use;
Invasive or not, and type of invasiveness;
If the device is re-usable;
Therapeutic or diagnostic;
If the device depends on a source of energy; and
The part of the body in contact with the device.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
11
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
12
Choosing the procedure best suited to the manufacturers needs can be guided by the following criteria:
The operation of the business for the organization to ensure the required quality (quality system); and
How to design and manufacture of medical device taking into account the share of outsourced
processes.
For each of the three directives, the evaluation procedure is a combination of the following modules, taking
into account the devices class:
Directive
Modules
90/385/EEC (AIMD)
93/42/EEC (MDD)
98/79/EEC (IVDD)
CE Declaration of
Conformity Full Quality
Assurance System
Design examination
included or not
included
Production Quality
Assurance included
or not included
Annex II
Annex II
Annex IV
CE Type Examination
Annex III
Annex III
Annex V
CE Verification
Annex IV
Annex IV
Annex VI
CE Declaration of
Conformity Production
Quality Assurance
Annex V
Annex V
Annex VII
CE Declaration of
Conformity Product
Quality Assurance
NA
Annex VI
NA
CE Declaration of
Conformity
NA
Annex VII
Annex III
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
13
The Directives state which procedures are possible with each class. Flow charts showing the procedures are
included in Appendices A and B. Essentially, the conformity assessments have two aspects:
Evaluating the design of the medical device to determine if it meets the requirements of the
Directives; and
Assessing the ability of the manufacturer to mass produce conforming devices and to fulfill its
regulatory obligations.
The manufacturer has the obligation to respect the regulatory process it chose prior to placing the device
on the market and it must be able to demonstrate its approach. Also, depending on the class (or similar for
IVD), it may also be required to verify compliance of this process by a third party: the notified body (see 2.5).
Depending on the class (or similar for IVD) of the device, the choice of different possible procedures
(methods of proof), the association consists of various modules. And these procedures are shown in figures
in Appendices A and B.
The only procedure that doesnt require a Notified body is the CE Declaration of Conformity. This procedure
can only be used for Class I devices that are not sterile or that do not have a measuring function. Even in
these cases, the Notified Body only assesses the points related to securing and maintaining sterile conditions
or metrology.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
14
4.1 Design: Examination of the Medical Devices Design Conformity with the Essential
Requirements
The design examination can be, according to the manufacturers choice given the devices class, either:
The responsibility of the manufacturer as part of the Declaration of Conformity (for example, Annex
VII to the MDD or Annex III of 98/79/CE); or
By the Notified Body according to the Product Design Examination or the Type Examination.
4.1.1 Design Examination
This is part of the approval of the complete system of quality assurance module. LNE / G-MED review the
design from a file provided by the manufacturer. The design dossier must contain, in particular, data on the
following:
This information provides evidence of product conformity, such as reports of pre-clinical and clinical validation,
quality system-approved procedures, and records related to the design of the device.
This module will result in a Certificate of design examination with a 5-year validity period according to the
Directive. LNE/G-MED will also provide a design report review. The manufacturer must inform LNE / G-MED
of any plan for substantial changes of the products covered.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
15
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
16
4.4 Verification of Manufactured Products (The Only Assessment Procedure Under the
IVDD)
For IVDs that are in Annex II, List A of the IVDD, this procedure is mandatory. The manufacturer will give
LNE/G-MED the relevant batch reports immediately after the examinations and tests are performed. The
manufacturer will also provide to LNE/G-MED the samples of the batches it produced. If LNE/G-MED detects
any abnormalities in the samples that reveal a non-compliance with the Essential Requirements, LNE/GMED must notify the manufacturer within an agreed-upon timeframe. The manufacturer cannot market the
affected lots.
13
http://ec.europa.eu/health/medical-devices/files/meddev/2_1_3_rev_3-12_2009_en.pdf
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
17
MD & AIMD
IVDD
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
18
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
19
6. Maintaining CE Marking
This chapter describes the measures taken by the manufacturer, as well as the obligations laid on the
manufacturer, for the purpose of maintaining the CE marking for given devices over time.
Every device placed on the European market must conform to the essential requirements. Indeed, the
concepts of placing on the market and putting into service refer to each individual product and not to a
type of device.
This means that in this regard, two objectives must be met:
Objective 1:
Objective 2:
For this reason, the directives contain provisions that are intended to ensure the permanent
maintenance of this conformity.
14
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
20
It may be said that if one or other of the following circumstances were to arise, but not limited to, a declaration
should be made:
6.4. EC type-examination
Changes to the approved product must receive further approval from the Notified Body that issued the EC
type-examination certificate wherever the changes could affect conformity with the essential requirements
of the directive or with the conditions prescribed for use of the product.
As in the case of EC type-examination, the applicant, who here may be the manufacturers representative,
is responsible for declaration.
The extent of change in a product which may require such a declaration is not specified in the directives, but
in this regard the conclusions of paragraph 6.1.1 of this guide may be applied.
In addition, for diagnostic in vitro medical devices listed in annex II, list A or list B, the manufacturer shall
inform the Notified Body on changes of pathogen agent and infection markers, as stipulated in the paragraph
6.1.1 of this guide.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
21
6.5 EC verification
Note: This procedure must not be confused with the verification of manufactured products specific to
certain in vitro diagnostic devices.
In this procedure, the Notified Body intervenes either for each copy placed on the market (unit verification),
or for each batch released (statistical verification). The question of the maintenance, during the time, of CE
marking conditions does not occur in this case.
However, for devices placed on the market in a sterile condition, the conditions for maintaining CE marking
as described in 6.2 of this guide apply for aspects regarding the securing and maintenance of sterility.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
22
7. Conclusion
Marketing medical devices in Europe requires CE Marking and following the applicable regulations and
standards for the devices. Forming a deep understanding the regulations is key to forming the regulatory
strategy for the devices entry into the European market. This Guide, along with the expertise at LNE/G-MED,
gives manufacturers this deep understanding of the regulations for creating an effective strategy. LNE/GMED, as a Notified Body, enjoys a front-row seat to the European medical devices regulatory scheme and can
help to open the European market for medical device manufacturers. Manufacturers can use this expertise
and experience for their benefit by choosing LNE/G-MED as their Notified Body.
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
23
Annexes
Annex A: Conformity Assessment Routes for Medical Devices under Directives 93/42/
EEC (MDD) and 90/385/EEC (AIMDD)
Class I Medical Devices
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
24
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
25
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
26
Class III Medical Devices and Active Implantable Medical Devices (AIMD)
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
27
Annex B: Conformity Assessment Routes for In Vitro Diagnostic Medical Devices under
Directive 98/79/EEC (IVDD)
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
28
Copyright 2012. LNE/G-MED North America, Inc. All rights reserved - www.LNE-america.com
29