DXR-60 Service Manual 6
DXR-60 Service Manual 6
DXR-60 Service Manual 6
Service Manual
209124 rev. 6
2 DIGORA Optime DXR-60
Copyright
Documentation, trademark and the software are copyrighted with all rights
reserved. Under the copyright laws the documentation may not be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or
machine readable form in whole or part, without the prior written permission
of PaloDEx Group Oy.
Manufacturer
www.kavokerrgroup.com
1. INTRODUCTION ............................................................................ 7
DISCLAIMER ......................................................................................7
SCOPE ..............................................................................................7
INSTRUCTION / COMMAND SYNTAX USED IN THE MANUAL: ...................... 7
ACRONYMS USED IN THIS MANUAL ....................................................... 7
2. OPERATING WARNINGS AND PRECAUTIONS ................................ 9
SERVICING PRECAUTIONS ................................................................. 10
3. SYSTEM DESCRIPTION ................................................................ 11
BLOCK DIAGRAM .............................................................................. 11
COVERS .......................................................................................... 12
SCANNER MECHANISM ASSEMBLY ...................................................... 13
ELECTRONICS.................................................................................. 15
FIRMWARE ...................................................................................... 17
4. USER INTERFACE STATUSES ....................................................... 18
UI PANEL ........................................................................................ 18
REAR PANEL LEDS ............................................................................ 19
5. SERVICE TERMINAL .................................................................... 20
ACCESSING SERVICE TERMINAL ......................................................... 20
NORMAL AND SERVICE MODE OF THE UNIT .......................................... 22
SERVICE COMMAND PRINCIPLES ........................................................ 22
SERVICE LOG .................................................................................. 23
6. CONFIGURATION PARAMETERS .................................................. 25
SHOWING / SETTING CONFIGURATION PARAMETERS ............................ 25
AUTOSTART <AUTOSCAN> ................................................................ 25
BEEPER ........................................................................................... 25
ENERGY SAVE MODE <IDLE> ............................................................. 25
AUTOMATIC SHUTDOWN <OFF>......................................................... 25
RESTORING FACTORY SETTINGS ........................................................ 25
7. TROUBLESHOOTING .................................................................... 26
ERROR MESSAGES ........................................................................... 26
ERROR 1, K100 ................................................................................ 26
ERROR 2, PMT .................................................................................. 27
ERROR 3, LASER .............................................................................. 28
ERROR 4, RESONATOR ...................................................................... 28
ERROR 12, K200 NOT CONNECTED ...................................................... 29
ERROR 13, K300 NOT CONNECTED ...................................................... 29
ERROR 23, K200 .............................................................................. 30
ERROR 24, LINEAR MOVEMENT ........................................................... 31
ERROR 34, PLATE SENSOR ................................................................. 33
ERROR 123, DOOR MOVEMENT ERROR................................................. 34
ERROR 124, SAFETY COVER ............................................................... 34
ERROR 234, UI ................................................................................. 36
ERROR 1234, SYNC TIMEOUT ............................................................. 36
4 DIGORA Optime DXR-60
8. MAINTENANCE ............................................................................ 37
ROUTINE CHECKS WITHOUT OPENING THE UNIT: ................................. 37
ROUTINE CHECKS WHEN UNIT OPENED: .............................................. 37
TENSION CHECK AND ADJUSTMENT ................................................. 38
9. INSTRUCTIONS FOR CHANGING SPARE PARTS ........................... 40
REMOVING AND INSTALLING THE COVERS ........................................... 40
BACK COVER ................................................................................ 40
FRONT COVER............................................................................... 41
CONTROL PANEL ........................................................................... 41
LIGHT COVER ............................................................................... 42
REPLACING MOVEMENT KIT PARTS (SP00912) ...................................... 44
REPLACING OUTER COVERS (SP00913) ............................................... 49
REPLACING INNER LIGHT PROTECT COVER (SP00914) ........................... 49
REPLACING CONTROL PANEL ASSEMBLY (SP00915)............................... 50
REPLACING K200 HOME-LASER-DETEC IPS-2013 (SP00916) ................... 51
REPLACING K300 PLATE DETEC REFL.SEN IPS-2013 (SP00917) ............... 54
REPLACING LASER AND MIRROR ASSEMLY IPS-2013 SP00918 ................ 55
INSTRUCTIONS 1: DEVICES BEFORE SN L1916033 ................................ 56
INSTRUCTIONS 2: DEVICES STARTING SN L1916033............................. 57
REPLACING PMT (SP00919) ............................................................... 59
REPLACING UNIT DOOR (SP00920) ..................................................... 60
10. FIRMWARE AND CORE UPGRADE .............................................. 61
FIRMWARE (=FW) UPGRADE .............................................................. 61
CORE UPGRADE ............................................................................... 62
11. TROUBLESHOOTING ................................................................. 63
ERRORS WITHOUT ERROR CODE ........................................................ 63
IMAGE ERRORS ................................................................................ 65
IMPROPER USE OF THE HYGIENE ACCESSORIES ................................ 65
IMPROPER X-RAY SETTINGS USED ................................................... 65
GHOST IMAGES, SHADOWS ............................................................ 66
IMPROPER AIMING OF THE X-RAY .................................................... 66
UNSHARP / BLURRED IMAGES ......................................................... 66
GEOMETRY DISTORTION ................................................................ 66
DECREASED CONTRAST, SHADOWS/SHADING, GHOST IMAGES… ........ 67
WHITE OR GREY DOTS/SPOTS/STAINS IN IMAGES ............................ 67
WEARING OF THE IMAGING PLATES ................................................. 67
IMAGE SHOWS DIFFERENT SIZE THAN THE IP USED .......................... 69
CIRCLE ON IMAGE ......................................................................... 69
VERTICAL STRIPE(S) ..................................................................... 70
WHITE BAR BEFORE IMAGE ............................................................ 71
PMT PROBLEMS ............................................................................. 71
INCORRECT DIMENSIONS ............................................................... 72
COMPRESSED (TOO NARROW) IMAGE .............................................. 72
RECTANGULAR / PIXELATED ARTEFACTS, ESPECIALLY
ON LIGHT AREAS........................................................................... 72
INCOMPATIBLE DRIVER OR SOFTWARE ............................................ 73
12. SPARE PARTS ........................................................................... 75
SCOPE
This manual provides the information necessary to perform field servicing and
maintaining of the DIGORA™ Optime Digital imaging plate system (hereafter
referred as a “scanner unit”).
CAUTION:
Do not move or knock the scanner when it is scanning.
The use of control or adjustment or performance of procedures other
than those specified herein may result in hazardous laser radiation
exposure.
This scanner must only be used to read image plates and must not be
used for any other purpose. Only use the imaging plates and protective
covers that are supplied by scanner unit manufacturer. NEVER use
imaging plates or protective covers from other manufacturers with the
scanner
This scanner, or its accessories, must not be modified, altered or
remanufactured in any way.
Only the manufacturer’s authorized service personnel are authorized to
carry out annual maintenance and repair. There are no user serviceable
parts inside.
This device can interfere with other devices due to its EMC
characteristics.
Other devices can interfere with this device due to their EMC
characteristics.
This device complies with IEC 60601-1 standard. Accessory equipment
connected to this device must be in compliance with the related
nationally harmonized IEC standards.
Equipment not suitable for use in the presence of flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
Only use the power supply unit that is supplied with the scanner. Do not
use any other power supply units with the scanner.
Use RJ45 UTP network cable, CAT 6
Other cables and accessories may negatively affect EMC
ESD
When servicing & unit covers being open, please note that inside the unit all
electrical components are ESD sensitive. Appropriate protective methods
(ESD wristbands, ESD safe working areas & ESD clothes etc…) must be used
against ESD problems.
Laser Radiation
There is a class 3B laser inside the scanner. The laser can be activated if the
light beam in the optic safety switch is blocked and the laser is switched on
from the service terminal. Avoid direct exposure to the laser beam or its’
reflections (for instance from shiny tools etc).
High Voltage
There is a high voltage circuit between the photo multiplier tube and amplifier
card. High voltage can be set on when the light beam of opto safety switch is
blocked. See service commands.
RECYCLING
INFORMATION
ITEM CODE DESCRIPTION MATERIAL
1 SP00913 OUTER COVERS OPTIME DXR-60 PC+ABS
FRAME: PC+ABS
2 SP00915 CONTROLPANEL ASSEMBLY OPTIME DXR-60 Other: E-Scrap
3 SP00973 IMAGING PLATE COLLECTOR DXR-60 PC+ABS
NOTICE!
Products without part numbers are not for sale.
RECYCLING
INFORMATION
ITEM CODE DESCRIPTION MATERIAL
1 SP00912 MOVEMENT KIT IPS-2013
2 SP00914 INNER LIGHT PROTECT COVER IPS-2013 PC+ABS
3 SP00916 K200 HOME-ERASE-LASERDETEC IPS-2013 E-SCRAP
4 SP00917 K300 PLATE DETECT REFL.SEN IPS-2013 E-SCRAP
5 SP00918 LASER AND MIRROR ASSEMBLY IPS-2013 E-SCRAP
6 SP00919 PMT, PHOTO MULTIPLIER TUBE IPS-2013 E-SCRAP
7 SP00920 UNIT DOOR IPS-2013 PBT/Steel
NOTICE!
Products without part numbers are not for sale.
RECYCLING
INFORMATION
PCS ITEM MATERIAL
2 SHAFT CCM08 CERAMIC COATED Aluminum, ceramic
coated
1 BELT PULLEY COVER POM
1 IMAGING PLATE CARRIAGE ASSEMBLY POM/NdFeB/ Aluminum/
Glass-fiber reinforced
polyarylamide
1 MOTOR HOLDER ASSEMBLY POM/Steel/E-scrap
1 READOUT HEAD BELT MXL025-G Rubber
1 BELT 264X8 CH XL GA152 MANDREL84.1 Rubber
1 BELT PULLEY 73.8 - 5.4 Aluminum
2 BELT PULLEY POM/Steel
2 BEARING 625 ZZ Steel
1 DOWEL PIN 4H6X24 DIN6325 Steel
2 LOCKRING 6 DIN6799 Steel
1 NEEDLE ROLLER NRB4X25,8-G2 Steel
LED D11 is activity led. D12 is speed led, it is on when Ethernet PHY
has established link at 100 Mbps. These are connected to the back
using light pipe.
Architecture
Workstation (PC) Ethernet The scanner unit
Imaging Application (optional)
Service
Terminal
(optional) User Cmd Msg
Imaging Application
End user’s interface to acquired images.
Handles image archiving and displaying.
Driver
Handles the connection/communication to the unit
Transfers and manipulates images.
Unit cannot operate without Driver
There are two alternatives to be used as a driver:
1. DSD driver. Run from imaging application.
2. s2terminal program. Run from the command prompt of Windows.
Service Terminal
Optional part of the driver.
Execute user commands
View the service output of the unit.
S2terminal program. Run from the command prompt of Windows.
NOTICE! If the opto safety switch light is blocked during start and covers
removed –the PMT may be damaged!
UI PANEL
1 3 2
1. ON/OFF key
2. START key
3. Error code indicators
Further details on the interface symbols, leds and buttons are described in the
user’s manual.
Be very careful not to use service commands without the necessary know-
how. Gain knowledge by going through all useful commands and train their
usage in practice with scanner unit.
Service terminal commands and functions are not explained in this manual.
They are instructed in the scanner firmware help (which is accessible through
service terminal). This arrangement is to keep the documentation up to date,
since the service terminal instructions for commands are always up to date in
the scanner firmware.
Therefore, this manual explains only the principles of the main commands.
Note the difference of actual command prompt and the s2terminal program
running in it.
The input and output of the s2terminal and of the command prompt look
similar. However, the functions and commands available in the command
prompt are not available in the s2terminal and vice versa.
NOTICE! If you get confused it is good idea to check the title bar of the
command prompt window - it should read “s2terminal” when the s2terminal
program is running.
1. Copy all mentioned files to one same and dedicated service folder that
you may need to create (in following example “c:\s2terminal” folder is
used to store s2terminal files)
20 DIGORA Optime DXR-60
2. Select from Windows: Start / Programs / Accessories / Command
Prompt
Or
Select from Windows: Start / run
4. Choose OK
Command prompt examples:
<cd\> (Changes to the root directory =C:\)
<cd s2terminal> (changes to directory “s2terminal”)
NOTICE! It is normal that you may have few times “login fail (no
connection)” –error also in normal functioning connection
Almost all of the service procedures described in this manual are done in the
service mode.
Remember to quit the service mode before trying to use the unit
normally.
<s> Activates the service mode – and “service>” prompt is shown in the
service terminal.
<quit> Quits the service mode (=resumes scanner into normal operation
mode).
<reset> Resets scanner unit. (ONLY SW RESET) Use if you get strange
behavior when trying to use the unit after exiting the service mode.
After reset you exit also the service mode
NOTICE! Do not execute the command if you are not sure about its
function. Improper use of some commands may result scanner unit to
become inoperative!
Help Command
The scanner firmware has built-in help.
The help functionality is similar in both (Normal / Service) modes of the
unit.
Example:
<h calpr> Means that typing: h calpr (and pressing enter) shows
the help for using calpr –options
Configuration info
<conf> Shows the configuration parameters of the scanner unit
SERVICE LOG
Service log is a very useful tool for troubleshooting, so please learn
how to use it!
You can get list of the error messages, which helps on troubleshooting.
(Every time an error is caught it is recorded into the service log).
You can see the performed service procedures
TIP! Copy the contents of the service log, paste it to Notepad, Word etc. and
send to manufacturer Technical Service by Email. This helps the manufacturer
staff to give troubleshooting support.
AUTOSTART <AUTOSCAN>
<autoscan 0> Configures scanner unit to start readout after
START is pressed
<autoscan 400> Configures scanner unit to start readout
AUTOMATICALLY 0.4s after plate cover is removed
BEEPER
<beeper 0> Disables extra beeps (when going to idle mode
etc…), but leaves all error and warning beeps
enabled
You can also restore individually each parameter to their factory settings by
commanding <parameter XXX>, where XXX is factory set value, which is
shown after each parameter in parenthesis (XXX).
1 - K100 error
2 - PMT error
3 - Laser error
4 - Resonator error
12 - K200 not connected
13 - K300 not connected
23 - K200 error
24 - Linear movement error
34 - Plate sensor error
123 - Door movement error
124 - Safety cover error
234 - UI error
1234 - Sync Timeout error
ERROR 1, K100
NOTICE! The K100 board cannot be changed in the field. These errors below can
in most cases be corrected only if the K100 board is changed in the factory.
1-1 ERROR_K100_LIN_AMP_NOT_OK
1-2 ERROR_K100_CFG_DATA_LOST
1-3 ERROR_K100_HV_NOT_OK
1-4 ERROR_K100_1V2_VCCD_PLL
1-5 ERROR_K100_1V2_VCCINT
1-6 ERROR_K100_1V8
1-7 ERROR_K100_2V5_VCCA
1-8 ERROR_K100_3V3
1-9 ERROR_K100_3V3A
1-10 ERROR_K100_5V
1-11 ERROR_K100_24V
Check power adapter output voltage. If below 22.8V change the
adapter.
1-12 ERROR_K100_DOOR_OPTOS_ACTIVE
This error comes if both door open and door closed opto lights are blocked
at the same time.
26 DIGORA Optime DXR-60
1-13 ERROR_K100_DOOR_OPEN_OPTO
Plate carrier is inside and light is blocked in the door open opto. (Door
should be Closed).
1-14 ERROR_K100_DOOR_CLOSED_OPTO
Plate carrier is at home position and light in the door closed opto is
blocked (Door should be open)
1-15 ERROR_K100_ETH_PHY
1-16 ERROR_K100_ADC
In cases 1-11, 12, 13 it is recommended first to check that there are no dust or
other particles that blocks the light in the door optos.
You can also use command <diagnose> and check if the door opens and
closes totally when carriage goes in and totally open when carriage goes out.
If door operates correctly and door open or door closed opto fails K100 board
is faulty.
K100 is not a spare part. You have to send the whole scanner back to factory
for repair.
If the door operation does not work correctly change the door (SP00920)
ERROR 2, PMT
PMT is tested during boot. Blue led at K200 board is used as the light source.
Error occurs if signal is too low.
Reason: PMT fail/K100 fail or led light source is not working on K200 board.
3-0 ERROR_LASER
3-1 ERROR_LASER_ALWAYS_ON
Fix: Detach power and Ethernet connectors. Open the covers and check the
laser connector. If it’s OK, you have to change the LASER AND MIRROR
ASSEMLY IPS-2013 (SP00918)
Laser connector
ERROR 4, RESONATOR
4-0 ERROR_MIRROR_NOT_MOVING
It is first recommended to check that mirror cable is connected to K100
board. If it’s connected but the mirror is still not working properly, you have
to change the LASER AND MIRROR ASSEMLY IPS-2013 (SP00918)
4-1 ERROR_MIRROR_UNSTABLE
This error comes if mirror is not stable. Too high magnetic field
(>10A/m) might cause this error. A mechanical vibration on table or floor
where the unit is placed can cause this error. First try the unit in different
place. If the unit is still not working change the LASER AND MIRROR ASSEMLY
IPS-2013 (SP00918)
28 DIGORA Optime DXR-60
4-2 ERROR_MIRROR_FREQ
This error comes if mirror frequency too high or too low. Change the LASER
AND MIRROR ASSEMLY IPS-2013 (SP00918)
23-0 ERROR_K200_ERASE_LED
This error comes if LED forward voltage level differs >10% from the
calibrated value. Check that all erase leds are light during erase. If leds are
ok and unit still gives this error, perform erase calibration with command
<calerase> in service mode.
If the leds are not lit correctly, replace K200 board and perform erase
calibration with command <calerase> in service mode.
23-1 ERROR_K200_LINEAR_OPTO_HOME
This error comes if home opto (near the door) light is blocked. First check
that there are no dust or other particles that blocks the light. If the opto is
clean and the unit is still not working, replace K200 board.
23-2 ERROR_K200_LINEAR_OPTO_START
This error comes if start opto (near the pmt tube) light is always blocked.
First check that there are no dust or other particles that blocks the light. If
the opto is clean and the unit is still not working, replace K200 board.
23-4 ERROR_K200_SYNC_OPTO_TOP
This error comes if firmware does not get any pulses from the upper photo
diode.
The pulse does not come if the photo diode is broken or the laser does not hit
the upper photo diode in K200 board. It is recommended first to check that
K200 board is properly connected to K100 board connector and board is
locked to the chassis.
23-5 ERROR_K200_SYNC_OPTO_BTM
This error comes if firmware does not get any pulses from the lower photo
diode.
The pulse does not come if the photo diode is broken or the laser does not hit
the lower photo diode in K200 board. It is recommended first to check that
K200 board is properly connected to K100 board connector and board is
locked to the chassis.
23-6 ERROR_K200_SYNC_OPTOS
This error comes if both sync optos (top & bottom) don’t produce pulses to the
firmware. Check the condition of the K200 board, usually this error occurs when
the K200 is broken and needs to be replaced.
24-0 ERROR_MOTOR
This error comes if motor connector short circuit wire is not connected. Check
the motor connector. Check that cable wires and contacts are not broken.
24-1 ERROR_MOTOR_POSITION
This error comes if plate carrier does not block the light in correct time from
the home or start optos in K200 board. That can happen if the motor is
running in wrong speed (check that the short circuit wire in the motor cable is
connected in correct pins). This can also happen if the motor belt is adjusted
too loose and then the shaft and belt will slip.
24-2 ERROR_MOTORHOME_TOO_EARLY
24-3 ERROR_MOTORHOME_TIMEOUT
These errors come if plate carrier does not reach the home position at the
specific time.
24-4 ERROR_MOTORHOME_STUCK.
Occurs if linear movement is stuck at home opto.
24-5 ERROR_MOTOR_NOT_CONNECTED.
Occurs if linear movement motor doesn’t get power. Check that the motor
cable is connected and that it’s not broken.
24-6 ERROR_MOTOR_BELT_SLIP.
Occurs if belts slip during the linear movement. Check the condition of the
belts and their tensions. Adjust the tension if needed.
If spare K300 is not available immediately, you can still use the unit by
setting autoscan to -2 in service mode. With this setting the scanning begins
when you press the start button but without checking that the image plate is
installed correctly.
234-0 ERROR_UI_NOT_CONNECTED.
Open the covers and check that the cable is properly connected to K100
board and user interface board. The cable contact pads must be against the
circuit board. The cable must be straight when it leaves the connector. Make
sure that the cable is pushed bottom of the connector and the connector
locking clip is properly closed.
NOTICE! In some cases, this error message can be shown in the panel and
sometimes not. It depends how many signals are still connected between
K100 and user interface. If the cable reconnecting does not help, then replace
control panel assembly.
If any of these errors occurs, it means that the laser synchronization is not
working properly. Check the condition of the laser, K200 board and mirror.
Replace the defected parts.
Check that the unit reads and shows all 4 IP sizes properly
Clean drive belt, motor shaft and drive wheel + conveyor belt and it’s
drive wheels
o Use lint-free cloth dampened with rubbing alcohol (ethanol)
o Clean the inside surface of the plate carrier belt. Clean also the
drive shaft of the plate carrier belt. Rotate from drive wheel back
and worth to ease the cleaning.
o Remove drive belt from motor shaft and drive wheel. Press the
motor towards the wheel and push the belt off. Clean the motor
shaft, drive wheel and belt. Reinstall the belt.
Tighten the plate carrier (conveyor) belt if gets too loose (=the plate
carrier not moving properly).
o See instructions “Plate carrier Belt (conveyor belt): Tension check
and adjustment” in this manual
Check that the plate carrier belt drive pulley (close to front plate) is
properly fastened
Push the tool against the shaft Belt tightness is now in maximum
and belt. limit (3 rings are shown).
Torque wrench
If you have torque wrench use 50Ncm moment.
Always scan test images after changing spare parts. See appendix 1.
You must also evaluate IEC 62353 compliance after repair. See appendix 2.
NOTICE! All clips and details in chassis base assembly are rigid and meant to
be bent only with great care. Do not use excessive force.
Before opening covers turn the power off and detach power and Ethernet
connectors
BACK COVER
2
1
Press the two tabs on the right When you install the front cover back
and left side to release the cover check that it is directed correctly to the
and remove it. slots. Ensure that the tabs lock the
panel.
CONTROL PANEL
LIGHT COVER
2 3
The tab in the light cover goes to the hole in the chassis. This is used in the
safety switch opto on K100 board to block the light.
If after reassembly safety switch error occurs the light cover might be raised
from the chassis.
Keep old and new parts separated so that you don’t mix them.
Remove the readout belt. Release the locking by pressing the tab (1) at the
back side of the plate carrier and simultaneously turning adjustment screw
slightly clockwise (2) and then releasing the black belt mechanism (3).
(Allen hex 3mm tool) at the front face of the plate carrier. After that,
remove the belt. See images below.
1
2
9. (In newer devices: Detach the bottom cover ESD tape from the belt
pulley cover.) Slide belt pulley cover away and remove the belt pulley
from front side.
4. Check also that the carriage can move smoothly from the back to the
door. It should not hit to K200 board.
5. Install the motor holder assembly. Remark that it goes only in the
correct position.
6. Connect motor cable to K100 and set the motor holder spring correctly.
See image.
7. Add grey belt around the belt pulley. Push the wheel to its place. Check
that the bearings goes perfectly to its correct place.
8. Add the drive belt first around the wheel then push the motor holder
towards the wheel and slide the belt on the motor shaft.
10. Add the dowel pin to the small wheel and install it to its place.
Pull both right and left side locking clips and lift
the control panel assembly up
Removing old control panel assembly:
a. Release the tabs from both sides and lift the panel.
b. Detach the cable from K100 board.
When you install the new panel, check the correct installation.
1. Remember to connect the cable back to K100 board
2. Install the covers back on
3. Attach the power adapter
4. Attach connection to PC
5. Turn the unit on and start S2terminal
6. Wait that the unit starts. This should happen without errors.
7. Take a test image
5. Remove PMT
6. Release light collector clip on the right side first and rotate light collector a
bit and lift light collector away.
A
B
7. Release the clip that locks the board to the chassis (marked B in image)
and lift the K200 board from the connector
8. Insert the new K200 board. Check that it goes to connector and that the
slot on the PCB goes to the holder (marked A). Check that B locks the PCB
correctly.
Correct installation of the light collector. Check that the clips lock the
collector. There is a small oval shape hole in the light cover. Check that the
PMT tube is pushed inside the light collector so that this hole is plugged.
18. Go to service mode and type <calerase>. The unit now calibrates erasing
LEDs.
20. After calibration, quit from service mode and turn the unit off and again
on. The device should now start without errors.
See Service manual for more instructions. There are two different ways to
replace the laser and mirror assembly, depending on the manufacturing date
of the device. Starting in 2019, new laser cable will have two ferrites to give a
better protection against EMC disturbances. Figures 1 and 2 shows the
difference. Choose upon the following two instructions based on the device
version you have. Figures 1 and 2 show the difference of the old and new
version. The spare part SP00918 will include 2 ferrites. However, these need
to be added only to the new version of the device. For the old version, the
ferrites can be scrapped.
Figure 2. New version of the device. See white ferrites in the laser cable.
Install the assembly. Check that the pin and cavity at the bottom go to
respective parts in the chassis. The assembly must be tightly on the chassis.
10. Open the two screws marked by arrows in Figure 3. Use 3mm hex key.
11. Lift assembly away.
Install the new assembly. Check that the pin and cavity at the bottom go to
respective parts in the chassis. The assembly must be tightly on the chassis.
8. Check door operation by rotating the big wheel. When carriage moves
forward it should open the door and when backwards the door should be
closed.
9. Insert the light collector, PMT and PMT keeper
10. Insert the light cover
11. Insert the back cover
Copy the new firmware file (.srec) to the dedicated folder (for example
c:\s2terminal) with the s2terminal files = from which you will start the
s2terminal.
Switch scanner unit off and on to see that there are no errors during start up.
1. Copy the core file to the same, dedicated folder with s2terminal files
(for example, C:\s2terminal).
2. CHECK THAT IMAGING APPLICATION IS CLOSED!
3. Execute the s2terminal program using the unit’s IP –address
4. Command <xc XXX> ( XXX = version number in the core filename)
>> Starts the core upgrade
5. Wait until s2terminal tells that upgrade is finished
6. Switch scanner unit off and on to see that there are no errors during start
up.
If the led at adapter is green when adapter not connected to K100 but turns
off when DC plug is connected K100 board is faulty. This is also the case
when power adapter led turns off when power button is pressed.
If power led is green but pressing the power button does not turn the unit on
check if control panel assembly is faulty.
Remove the main and front covers. Press start button at K100 board.
If the unit now turns on check that control panel flex cable is connected
correctly. If it is ok you have to change the control panel assembly.
If still not working change autoscan setting to -2 in service mode. After that
plates can be scanned in manual mode: image is scanned when start button
is pressed.
Motion artefact
Patient or X-ray cone has moved during the exposure
(maybe because too long exposure time has been
used)
Geometry distortion
- Scratches
Clean the plate(s), replace if cleaning does not help
Pay attention on IP handling, storing and maintenance
Circle on image
PMT problems
Issue: Noise reduction selected + very low signal from the plate.
New, noise reduction algorithm is improved for normal clinical images.
If the signal is very low due to some defect (very short exposure time, wrong
driver, hardware defect ….), image will appear pixelated.
Earlier versions of drivers will connect to the new scanner unit but with old
generation image processing parameters resulting unsatisfactory image
quality. DfW 2.8.112.480 and SCANORA 5.2.2 are the first compatible SW
versions.
How to detect:
Lack of grey levels and dynamics, “flat images”, poor contrast. Symptoms
may differ depending on how old the driver is.
- Increasing exposure time does not help.
- Can show image as wrong size, image smaller than the imaging plate
(=cropping error).
Check the Software version, DfW 2.8.112.480 and SCANORA 5.2.2 are the
first compatible SW version
Immediate correction
Install software from the media that came with the unit OR
run SOREDEX™ Driver Update 13.3 or newer.
A. Shafts 2 pcs
B. Imaging Plate Carriage
Assembly
C. Readout head belt
D. Drive Belt 264x8
E. Belt Pulley 73.8-5.4
F. Belt Pulley 2pcs
G. Bearing, 2pcs
H. Dowel Pin 4H6X24
I. Belt Pulley Cover
J. Motor Holder Assembly
K. Seeger lock ring 2pcs
L. Needle roller
NRB4X25,8-G2
NOTICE! Because back cover includes type label, the unit serial
number must be told to the customer service that they can make
correct label to the spare part.
Store performed service tasks into Unit service log by typing command
<logsign xxxxxx>, where xxxxxx.is free text inserted by service
technician to be stored into log.
o As an example <logsign John Smith: Replaced K200 board,
calibrated and endurance test performed. OK.>
All in-service and post repair electrical safety testing should be carried out in
accordance with the IEC 62353 standard and/or in accordance with any local,
state and national standards related to in-service and post repair electrical
safety testing of dental/medical devices.
According to the standard in-service and post repair electrical safety testing
should always be carried out after the device has been installed, repaired,
maintained or serviced.
According IEC 62353 limit for Class II equipment leakage current is 100uA
(direct method).
EQUIPMENT LEAKAGE CURRENT is defined as current flowing from MAINS
PARTS to earth via the protective earth conductor and ACCESSIBLE
CONDUCTIVE PARTS of the enclosure and APPLIED PARTS. The scanner unit
does not have protective earth conductor or accessible conductive parts or
applied parts so leakage current measurements are not needed.
The power adapter has accessible conductive part (DC plug) but this is newer
opened or serviced and it is always measured when delivered from factory.
That’s why repeated measurements are not needed.
More Information
[1] IEC 62353:2007, Medical electrical equipment – Recurrent test and test
after repair of medical electrical equipment
Read and familiarize yourself with the warnings and precautions listed in the
unit user's manual, Installation manual and Service manual.
Only service personnel trained and approved by the manufacturer of the unit
are allowed to service the unit.
VISUAL INSPECTION.
Inspection:
-Check that type label and Laser Warning label are present and complete on
the main cover
-Check that all covers are in good condition, no damage is allowed. The main
cover must be attached on the backside by the screw. See image below.
Type Label
Attach screw
-if you open the main cover check that the label on the light cover is ok.
If the unit does not comply with the requirements it must be corrected or
taken out of service until corrections are made.