DS-8400_CE_V07_1OM_4L011487I
DS-8400_CE_V07_1OM_4L011487I
DS-8400_CE_V07_1OM_4L011487I
Patient Monitor
Ver. 07
Emergo Europe
Westervoortsedijk 60, 6827 AT Arnhem, The Netherlands
UK Responsible Person
Fukuda Denshi UK Ltd.
Fukuda Denshi Co., Ltd.
Unit 6 Chambers Way Thorncliffe Park
3-39-4 Hongo, Bunkyo-ku, Tokyo 113-8483 Japan
Estate, Chapeltown, Sheffield, S35 2PH
CAUTION
● Only physician or persons instructed by physicians are allowed to use the device.
● The information contained in this document is subject to change without notice due
to improvement in the device.
Contents
Preface
Introduction..............................................................................................................................i
Important Notice ......................................................................................................................i
For Safe Operation of the Device.........................................................................................................i
Intended Use of this Device .................................................................................................................i
Copyright ............................................................................................................................................. ii
Maintenance, Repair, Replacement .................................................................................................... ii
Contact ................................................................................................................................................ ii
About This Manual ................................................................................................................ iii
Expression Used in This Manual........................................................................................................ iii
Composition of This Manual............................................................................................................... iv
Safety
About the Safety Precautions ..................................................................................................i
The Meaning of Each Safety Precaution..............................................................................................i
Warning Labels Attached to the Unit....................................................................................................i
Graphic Symbols .................................................................................................................................ii
Precautions for Safe Operation of Medical Device................................................................iv
Precautions about the Maintenance .......................................................................................v
Precautions about the Network System ................................................................................vi
Medical Telemetry .............................................................................................................................. vi
Precautions when Using with Other Device ......................................................................... vii
Pacemaker ........................................................................................................................................ vii
Non-Explosion Proof ......................................................................................................................... vii
Defibrillator ........................................................................................................................................ vii
Electrosurgical Instrument................................................................................................................ viii
MRI (Magnetic Resonance Imaging)................................................................................................ viii
Precautions about Connections to Peripheral Devices ...................................................................... ix
Precautions for Using the Device ..........................................................................................ix
This System........................................................................................................................................ ix
Wired Network (DS-LANII/ DS-LANIII) .............................................................................................. xx
Wireless Network System ................................................................................................................ xxi
RTC and Data Backup .................................................................................................................... xxii
Precautions about the Ventilator Monitoring....................................................................... xxii
Precautions about the SpO2 Sensor .................................................................................. xxii
Precautions about the Masimo Model ................................................................................ xxii
Precautions about the NIBP Cuff....................................................................................... xxiii
Precautions about Disposing of the Device, Accessories, or Components....................... xxiii
Precautions about Transportation ..................................................................................... xxiii
Monitoring after Power Failure .......................................................................................... xxiii
To Prepare for Emergency Use......................................................................................... xxiii
Electromagnetic Compatibility ........................................................................................... xxiv
Precautions for Safe Operation under Electromagnetic Influence ................................................. xxiv
EMC Guidance ................................................................................................................................ xxv
1
Contents
2
Contents
Chapter 4 Preparation
Daily Check ........................................................................................................................ 4-1
To Start Monitoring ............................................................................................................. 4-1
Check Discharge When Start Monitoring a New Patient................................................................. 4-2
Data Transfer Function Using the Super Unit ................................................................................. 4-3
Data Transfer Function Using the Transport Monitor (DS-8007) .................................................... 4-5
To Stop Monitoring ........................................................................................................... 4-11
Clock Setup ...................................................................................................................... 4-12
Installing the Recording Paper ......................................................................................... 4-13
Chapter 5 Admit/Discharge
To Display the "Admit/Discharge" Screen .......................................................................... 5-1
Admit .................................................................................................................................. 5-1
Entering the Patient Information...................................................................................................... 5-1
Entering Patient Information from the Magnetic Card ..................................................................... 5-4
Entering Patient Information from the Patient Data Server (When DS-LANIII is used)................... 5-4
Discharge ........................................................................................................................... 5-7
Discharging Procedure.................................................................................................................... 5-7
User Mode .......................................................................................................................... 5-8
To Select the User Mode................................................................................................................. 5-8
Suspend Monitoring ........................................................................................................... 5-9
To Suspend Monitoring ................................................................................................................... 5-9
To Resume Monitoring ..................................................................................................... 5-11
3
Contents
Chapter 7 Monitoring
To Display the Parameter Setup Screen ............................................................................ 7-1
ECG .................................................................................................................................... 7-2
Before Attaching the Electrodes ...................................................................................................... 7-2
Electrode Placement........................................................................................................................ 7-3
Type of Electrodes and Lead Cable ................................................................................................ 7-5
Connection to the Patient Monitor ................................................................................................... 7-5
ECG Parameter Setup..................................................................................................................... 7-6
Respiration ....................................................................................................................... 7-16
Respiration Monitoring (Impedance Method)................................................................................. 7-16
RESP Parameter Setup................................................................................................................. 7-17
BP ..................................................................................................................................... 7-20
BP Monitoring ................................................................................................................................ 7-21
Zero Balance of All Pressure Lines (User Key) ............................................................................. 7-26
Zero Balance of All Pressure Lines ([BP Zero] Key)...................................................................... 7-27
Zero Balance for Each Pressure Line............................................................................................7-27
BP Parameter Setup...................................................................................................................... 7-28
Non-Invasive Blood Pressure ........................................................................................... 7-35
Lineup of Cuffs............................................................................................................................... 7-35
NIBP Monitoring............................................................................................................................. 7-35
Inflation Mode Setup...................................................................................................................... 7-39
NIBP Auto Mode Setup.................................................................................................................. 7-39
Oscillation Graph Display .............................................................................................................. 7-42
Dyna Alert Function Status ............................................................................................................ 7-43
NIBP Parameter Setup .................................................................................................................. 7-45
Pulse Oximetry ................................................................................................................. 7-50
SpO2 Monitoring ............................................................................................................................ 7-50
SpCO, SpMet, SpHb, SpOC Measurement (Masimo)................................................................... 7-58
Precautions about the Masimo Sensors and Cables..................................................................... 7-58
SpO2 Parameter Setup (Medtronic)............................................................................................... 7-59
RR_SpO2 Parameter Setup (Medtronic) ...................................................................................... 7-62
SpO2 Parameter Setup (Masimo).................................................................................................. 7-63
Temperature ..................................................................................................................... 7-67
4
Contents
5
Contents
Chapter 9 Printing
Printing Setup ..................................................................................................................... 9-1
6
Contents
Chapter 11 Troubleshooting
Message List .................................................................................................................... 11-1
Vital Alarm Message ..................................................................................................................... 11-1
Vital Alarm Message (DS-LAN Standard Setup)........................................................................... 11-4
Device Status Alarm Message ...................................................................................................... 11-7
Numeric Data Box Message........................................................................................................ 11-13
Ventilator Alarm Message ........................................................................................................... 11-19
Ventilator Alarm Factor................................................................................................................ 11-19
Cardiac Output Message............................................................................................................. 11-20
Troubleshooting.............................................................................................................. 11-21
ECG............................................................................................................................................. 11-21
Respiration .................................................................................................................................. 11-26
Invasive Blood Pressure.............................................................................................................. 11-27
SpO2 Measurement (HS-8312N, DS-8007N, HG-820) ............................................................... 11-29
SpO2 Measurement (HS-8312M, DS-8007M, HG-810) .............................................................. 11-32
Non-Invasive Blood Pressure...................................................................................................... 11-35
Temperature................................................................................................................................ 11-40
Cardiac Output (CO) ................................................................................................................... 11-41
CO2 Measurement (HCP-810/HCP-820) ................................................................................... 11-44
CO2 Measurement (HPD-810/HPD-820) ................................................................................... 11-45
Recorder Unit (HR-800) .............................................................................................................. 11-46
Network Printer............................................................................................................................ 11-47
Wired Network (DS-LANII/ DS-LANIII) ........................................................................................ 11-48
7
Contents
8
Contents
Chapter 14 Specification
Specification ..................................................................................................................... 14-1
Main Unit: DSC-8400 Series ......................................................................................................... 14-1
Transport Monitor: DS-8007 and DSA-82 ..................................................................................... 14-2
Display Unit: LC-8018TC/LC-8016TC ........................................................................................... 14-3
Super Unit: HS-8312N/8312M and HSA-80/HSA-81 .................................................................... 14-4
Expansion Unit: MGU-800/810 Series and HR-800 ...................................................................... 14-5
Expansion Module: HM-800/801, HP-800, HG-810/820 ............................................................... 14-6
Gas Unit I/F: HPD-810 and CO2 Gas Unit: HCP-810.................................................................... 14-7
Gas Unit I/F: HPD-820 and CO2 Gas Unit: HCP-820.................................................................... 14-8
Input Box: IB-8004......................................................................................................................... 14-9
BISx I/F Unit: HBX-800................................................................................................................ 14-10
Performance ................................................................................................................... 14-11
Measurement Unit for Each Parameter .......................................................................... 14-21
Alarm Limit Range for Each Parameter.......................................................................... 14-25
About the SpO2 Clinical Test.......................................................................................... 14-27
9
Contents
10
Preface Introduction
Preface
Introduction
Thank you for purchasing this product. Read the "Safety Precautions" thoroughly before use to ensure correct and
safe use of the product.
Important Notice
(2) Fukuda Denshi cannot predict all the dangers which may be caused by misusage of this product or
environmental condition.
(3) For using this device, there are many items that "should be performed", "should not be performed", and "cannot
be performed". It is not possible to cover all these items in this manual or warning labels. Therefore, it is
necessary to also follow the general safety precaution other than the items described in this manual.
(4) To prevent accidents, usage other than intended, or usage, cleaning, and maintenance not described in this
manual should not be performed.
(5) When using this device, follow the respective regulation to minimize the probability of accidents.
REFERENCE
For specification of this device, refer to "Chapter 14 Specification" of this manual.
WARNING
This device is intended to be used by healthcare professionals. Users should have a
thorough knowledge of the function and operation before using this device. The
maintenance of this device should be performed by skilled personnel who received a training
of possible hazards and measures to avoid those hazards. Also, your local regulation must
i
Preface Important Notice
be followed. If this device is used for the purpose other than intended, or if the user does not
follow the safety instructions, the following hazard may result.
*Hazard to the Life and Health of the Patient or the User
*Damage to the device
*A Problem Related to Medical Practice
Copyright
(1) The copyright of this manual is owned by Fukuda Denshi. No part of this document may be copied or transmitted
in any form without the prior written permission of Fukuda Denshi Co., Ltd.
(2) This manual includes the description for the optional devices that can be connected.
(3) The illustration in this manual may differ with the actual device.
(4) If you lose or damage this manual, contact your nearest sales representative. Using the device without this
manual may cause accidents.
(5) When handing over this device, make sure to also pass this manual to the next owner.
Fukuda Denshi is liable for the safety, reliability, and performance of its device only if;
Maintenance, modifications, and repairs are carried out by authorized personnel or organization.
Components are used in accordance with Fukuda Denshi operating instructions.
A full technical description of the DS-8400 System is available from your local Fukuda Denshi sales representative.
Contact
If a serious incident has occurred in relation to this device, please report it to the manufacturer and to the
competent authority of the country where the user and/or the patient is established.
In case you need the contact information for your national competent authority, please ask the manufacturer or
the distributor from whom you purchased the device.
ii
Preface About This Manual
DANGER Failure to follow this message may cause immediate threat of death or serious injury.
WARNING Failure to follow this message may result in death or serious injury.
CAUTION Failure to follow this message may cause injury or failure to the device.
* Used in a table which indicates that there is detailed explanation outside the table.
The expressions displayed on the monitor screen are indicated by " " .
(Ex.: "Volume", "Admit/Discharge", etc.)
iii
Preface About This Manual
2.Names of Parts and Their Name and function of each part, external appearance
Functions
3.Operation Procedure and Screen Operation procedure, home display, window, procedure to return to the previous display,
Examples user key setup
4.Preparation Installing the recording paper, power ON/OFF, time/date, daily checks
5.Admit/Discharge Entering patient information (name, age, etc.) at admittance, discharging the patient, user
mode selection, suspend monitoring
7.Monitoring Measurement condition setup of the monitoring parameters, size/scale setup, etc.
Setup of the stop watch, connector
8.Review Function Arrhythmia analysis, 12-lead analysis, trend, recall, NIBP list, ST measurement,
hemodynamics, lung function, alarm history, other bed display, full disclosure waveform
10. System Configuration Setup of the display configuration, tone/volume, color, etc.
11. Troubleshooting Message list, maintenance and troubleshooting of this device
1.Installation of the Unit Precautions about the operating environment, system construction, mouse connection
4.Connection to the External External device connection/setup, magnetic card reader usage
Devices
5.Initial Settings Initial setup, administrator setup, alarm/measurement setup, user I/F, user mode
registration
7.Replacement Parts Precautions about the periodic replacement parts, consumable parts
9. Maintenance Check Daily and periodic checks, self-diagnosis function, software version
software install
iv
Safety About the Safety Precautions
Safety
About the Safety Precautions
Read this manual thoroughly before use to ensure correct and safe use of the product.
Be sure to follow the precautions indicated below, as these are important messages related to safety.
Type of Precaution Description
Indicates an imminently hazardous situation which, if not avoided, will result in death
DANGER or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in death
WARNING or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in minor
CAUTION or moderate injury.
Make sure to read the warning label attached to the device and comply with the requirements while operating the
device.
CAUTION
Do not damage or erase the warning label attached to the device.
This warning label contains important descriptions for handling and operating the device
properly and safely. A damaged label may compromise safe operation.
i
Safety About the Safety Precautions
Graphic Symbols
Refer to the following for the meaning of the symbol indicated on the device.
Symbol Description
Follow operating instructions (Warning); indicated in blue. Failure to follow operating instructions could
place the patient or operator at risk.
General precaution
Protective Earth
Indicates the protective earth inside the device.
Alternating Current
(Main Power Input Indicator)
ii
Safety About the Safety Precautions
Symbol Description
Signal Output
RS-232C Connector
Connects the related device.
Eject
Indicates the switch to pull out the paper tray.
Indicates prohibited actions. Refer to the instruction.
Battery
Alarm Silence
NIBP Start/Stop
Lock
Unlock
iii
Safety Precautions for Safe Operation of Medical Device
CAUTION
Users should have a thorough knowledge of the operation before using this device.
Do not use the device in an environment where protective earth and wiring is questionable.
iv
Safety Precautions about the Maintenance
WARNING
Never open the housing while the device is in operation or connected to hospital grade outlet as it may result
in electric shock.
CAUTION Precautions about Safety Check
For safe operation of the device, regular inspection and maintenance are required. Once a year, check all cables,
devices, and accessories for damage, earth impedance, earth and leakage currents, and all alarm functions.
Also, ensure that all safety labels are legible. Maintain a record of these safety inspections.
Immediate maintenance has to be carried out for the following case.
When the device was subjected to extreme mechanical stress, e.g. after a heavy fall.
When the device was subjected to liquid spill.
When the monitoring function is interrupted or disturbed.
When parts of the device enclosure are cracked, removed, or lost.
When any connector or cable shows signs of deterioration.
v
Safety Precautions about the Network System
Medical Telemetry
vi
Safety Precautions when Using with Other Device
Pacemaker
WARNING
Minute ventilation rate-adaptive implantable pacemakers can occasionally interact with certain cardiac
monitoring and diagnostic equipment, causing the pacemakers to pace at their maximum programmed rate. The
cardiac monitoring and diagnostic equipment may possibly send wrong information. If such event occurs,
please disconnect the cardiac monitoring and diagnostic equipment, or follow the procedures described in the
operation manual of the pacemaker. For more details, contact FUKUDA DENSHI personnel, your institution's
professionals, or your pacemaker distributors.
Rate meters may continue to count the pacemaker rate during occurrences of cardiac arrest or some
arrhythmias. Do not rely entirely upon rate meter alarms. Keep pacemaker patients under close surveillance.
Reference
"Minute Ventilation Rate-Adaptive Pacemakers"
FDA alerts health professionals that minute ventilation rate-adaptive implantable pacemakers can
occasionally interact with certain cardiac monitoring and diagnostic equipment, causing pacemakers to pace
at their maximum programmed rate.
[Based on a safety bulletin issued by FDA Center for Devices and Radiological Health on October 14, 1998]
Non-Explosion Proof
DANGER
Never operate the device in the presence of flammable anesthetics, high concentration of oxygen, or inside
hyperbaric chamber. Also, do not operate the device in an environment in which there is a risk of explosion.
Explosion or fire may result.
Defibrillator
WARNING
When defibrillating, keep away from the electrodes or medicament applied to the patient chest. If this is not
possible, remove the electrodes or medicament before defibrillating.
If the defibrillator paddles are directly in contact with the electrodes or medicament, an electrical shock may
result by the discharged energy.
When defibrillating, make sure that the electrodes, sensor cables, or relay cables are firmly connected to the
device.
Contacting the metal part of the disconnected cable may result in electrical shock from the discharged energy.
When defibrillating, do not touch the patient and the metal part of the device or cables.
Electric shock may result from the discharged energy.
This device will return to standard operating mode within 10 seconds after defibrillating. However, when in
diagnosis mode, it may require 10 seconds or more after defibrillation to display the normal ECG waveform as
the time constant setting is large.
The stored data will not be affected. The measurement accuracy will temporarily decrease during defibrillation,
but it will not compromise the safety of patient and the device.
The QRS synchronized signal is not intended to be used as synchronized signal for defibrillator.
vii
Safety Precautions when Using with Other Device
Electrosurgical Instrument
WARNING
The monitoring system contains protection against interference generated by electrosurgical instruments.
However, depending on the operating conditions, surgery site with respect to the location of ECG electrodes,
ground plate attachment condition, or the type of instrument used, it may cause burn injury at the electrode site
or noise on the ECG. The noise is generated at the tip of the electrosurgical knife and is difficult to completely
eliminate because of the frequency components of the ECG. To reduce electrosurgical interference, take the
following precautions:
Location:
Locate the electrosurgical unit as far as possible from this device and the patient cable. This will help reduce
interference on the ECG through the monitor or cables.
Power Supply:
Connect the electrosurgical unit to a power supply that is different from that of this device. This will help
prevent interference through the power cable.
Electrode Placement
The amount of noise interference is considerably different depending on the electrode position and surgery
site. Place the ECG electrodes as far away as possible from the surgery site and the ground plate. Do not
place electrodes in the path between the surgery site and the ground plate. If the electrodes are placed in
this path, the amount of interference will be quite large. Position (+) and (–) electrodes as close as possible
to each other.
Ground Plate
When using electrosurgical instruments, make sure the contact between the patient and the ground plate is
secure. If the connection is incomplete, the patient may suffer from burn at the electrode site.
The stored data will not be affected. The measurement accuracy will temporarily decrease during
electrosurgery, but it will not compromise the safety of patient and the device.
When using the electrosurgery-proof type ECG relay cable, the impedance respiration cannot be measured, and
its numeric data and waveform will not be displayed. When measuring in an environment where electrosurgery
is not performed, make sure to use the standard ECG relay cable.
As this device utilizes capacitive touch panel, the energy from the electrosurgical knife may pass through the
cable to the touch panel causing unintentional touch panel control. Locate the cables as far away as possible
from the touch panel.
WARNING
viii
Safety Precautions for Using the Device
To use the device safely and to ensure maximum performance of the device, connection of other manufacturer's
device to this device is not authorized, unless the connection is explicitly approved by Fukuda Denshi. It is the user's
responsibility to contact Fukuda Denshi to determine the compatibility and warranty status of any connection made
to another manufacturer's device.
WARNING
When multiple devices are connected to the patient, it may be necessary to take measures for connection (use
of separation device), power supply (use of isolation power), grounding (additional protective earth). If these
measures are not properly taken, a leakage current may flow between the devices, or the total amount of leakage
current may exceed the limit specified on EN 60601-1-1 or EN 60601-1.
Only the peripheral devices specified by Fukuda Denshi should be connected with the given procedure. Use of
an unspecified device may cause electric shock to the patient and/or operator due to excessive leakage current.
CAUTION
Although the peripheral device connectors on the DS-8400 System are, with some exceptions, isolated from
the power supply, the connecting peripheral devices should comply with EN 60601-1. It is the user's
responsibility to verify that the overall system complies with EN 60601-1-1 or EN 60601-1.
To prevent danger of electric shock, always position the peripheral devices away from the patient.
Network device including printer and hub should be located outside the "Patient Environment". If located
inside the "Patient Environment", it may result in electric shock to the patient or the operator.
Combinations of medical device with non-medical device must comply with EN 60601-1-1 or EN 60601-1 or
EN 60601-1. Never use a multiple portable socket-outlet or extension cable when connecting the devices unless
it is supplied specifically for use with that device.
This System
DANGER
When connecting to other devices, contact your nearest representative.
Danger such as electric shock may result to the patient and operator.
WARNING Warnings about the System
Do not connect any device or cable not authorized by Fukuda Denshi to any I/O connector. Also, do not connect
any damaged device or cable. The device cannot deliver its maximum performance, the device may be
damaged and safety cannot be ensured.
If the device is used under an environment not fulfilling the specified condition, not only that the device cannot
deliver its maximum performance, the device may be damaged and safety cannot be ensured. If using the
device under condition other than specified, contact your nearest representative.
Use only the supplied 3-way AC power cable. Use of other cables may result in electric shock to the patient
and the operator.
The power cable must be connected to a hospital grade outlet.
When using multiple ME device simultaneously, perform equipotential grounding to prevent potential
difference between the devices. Even a small potential difference may result in electric shock to the patient and
the operator.
ix
Safety Precautions for Using the Device
x
Safety Precautions for Using the Device
However, excessive waveform morphology change, motion artifact, or the inability to determine the waveform
pattern may cause an error, or fail to make adequate detection. Therefore, physicians should make final
decisions using printing and recall waveform for evaluation.
The RR/APNEA alarm will not be generated unless the numeric data box corresponded to the selected RR/
APNEA alarm source is displayed. Make sure to display the numeric data box for the RR/APNEA alarm
source.
The SpO2 respiration measurement is not intended for use as an APNEA monitor.
When selecting [0] for "Volume" or [Timer] for "Display" for the Night Mode, pay attention not to miss any
important alarm by simultaneously monitoring the patient on central monitor or other monitors.
When the alarm sound is suspended, the alarm sound will not generate for the fixed amount of time. Pay
attention not to miss any important alarm by simultaneously monitoring the patient on central monitor or other
monitors.
If the safety of the patient cannot be ensured, do not suspend the alarm or decrease the alarm volume.
The oxygenator mode is intended to prevent alarms during cardiopulmonary bypass surgery. Pay special
attention when using this mode as the alarm generation will not be the same as to the standard monitoring mode.
If the "Alarm Setting" under the Oxygenator Mode Setup is set to [All OFF], all vital alarm will not generate
regardless of the alarm setting of each parameter. Also, if [Sel. Parameter] is set, vital alarm for unselected
parameter will not generate. Pay attention to not miss any significant change of the patient's vital signs.
Once the cardiopulmonary bypass is finished, make sure to cancel the Oxygenator Mode and return to the
standard monitoring mode.
WARNING Warnings about the CO2 Monitoring (HCP-810/HCP-820, HPD-810/HPD-820)
Only one of either HCP-810/HCP-820/HPD-810/HPD-820 can be connected.
When using a sampling line for intubated patients with a closed suction system, do not place the airway adapter
between the suction catheter and endotracheal tube. This is to ensure that the airway adapter does not interfere
with the functioning of the suction catheter.
To prevent cross-infection, do not allow the sampling gas to return to the breathing system.
To protect the hospital staffs from unnecessary anesthetic agent when using the HCP-810/HCP-820, it is
strongly recommended to connect the exhaust hole to the gas exhaust system in the hospital.
Loose or damaged connections of the sampling line may compromise ventilation or cause an inaccurate
measurement of respiratory gases. Securely connect all components and check connections for leaks according
to standard clinical procedures.
Do not cut or remove any part of the sampling line. It could lead to erroneous readings.
If too much moisture enters the sampling line (i.e., from ambient humidity or breathing of unusually humid air)
when using the HCP-810/HCP-820, <Check Sample Line> will be displayed in the message area. Replace the
sampling line once this message is displayed.
Carefully route the sampling line to reduce the possibility of patient entanglement or strangulation.
Do not lift the HCP-810/HCP-820 by the sampling line, as the sampling line could disconnect from the device,
causing the device to fall on the patient.
CO2 readings and respiratory rate can be affected by sensor application errors, certain ambient environmental
conditions, and certain patient conditions.
WARNING Warnings about the Gas Monitoring (MGU-800/810)
Make sure to use only the specified Mindray Medical Sweden AB product.
( "AG Sensors for MGU-800/MGU-810 Manufacturer: Mindray" P13-11)
Be careful not to damage the water trap during operation as bacteria and/or mucus may contaminate the MGU-
800/810 series.
The airway adapter, sampling line, flow sensor are disposable products that are intended for single patient use
only. Do not reuse them on other patients as it may cause cross-infection.
xi
Safety Precautions for Using the Device
To prevent cross-infection, do not allow the sampling gas to return to the breathing system.
When using a sampling line for intubated patients with a closed suction system, do not place the airway adapter
between the suction catheter and endotracheal tube. This is to ensure that the airway adapter does not interfere
with the functioning of the suction catheter.
Do not use the MGU-800/810 series with the flammable anesthetic agents.
To protect the hospital staffs from unnecessary anesthetic agent, it is strongly recommended to connect the
exhaust hole to the gas exhaust system in the hospital.
The sampling line may get clogged by internal condensation.
The contents of the water trap should be handled as a potential infection hazard.
Do not use adult/pediatric type water traps and/or sampling lines with neonates to avoid high sampling flow.
Connect only DRYLINE gas sampling lines to the water trap. Note that there may be other compatible tubing
present, e.g. IV-lines.
Do not use DRYLINE neonatal sampling lines (blue luer lock nuts) with DRYLINE adult water traps as this
could result in incorrect measurement data.
Do not use DRYLINE adult sampling lines (colorless luer lock nuts) with DRYLINE neonatal water traps as
this could result in incorrect measurement data.
Only combine the SPIRIT Flow Sensors and DRYLINE Water Traps as specified. Other combinations might
lead to incorrect measurements.
( "Connecting to the Respiration Circuit" P7-88)
Use the adult flow sensor for a patient whose tidal volume is above 150 mL.
Use the pediatric flow sensor for a patient whose tidal volume is below 300 mL.
Make sure to use the correct flow sensor depending on the patient conditions, adult or pediatric and the tidal
volume.
Do not confuse the gas sampling line with other compatible tubing, e.g. IV-lines.
WARNING Warnings about the 12-Lead ECG Analysis Function
The 12-Lead ECG analysis function is designed to acquire and interpret ECG data from a resting, supine
patient. If ECG signals from moving or shaking patients are acquired, erroneous 12-lead interpretation result
may occur. Always ensure that the patient is kept motionless during 12-lead ECG signal acquisition and
analysis.
The 12-lead ECG analysis function is intended for use with adult and pediatric patients.
All computerized ECG analysis results should be reviewed by a physician before making decision for the
patient treatment.
WARNING Warnings about the BIS Monitoring
Clinical judgment should always be used when interpreting BIS in conjunction with other available clinical
signs. Reliance on BIS alone for intra-operative anesthetic management is not recommended.
BIS values should also be interpreted cautiously with certain anesthetic combinations, such as those relying
primarily on either ketamine or nitrous oxide/narcotics to produce unconsciousness.
As with any monitored parameter, artifacts and poor signal quality may lead to inappropriate BIS values.
Potential artifacts may be caused by poor skin contact (high impedance), muscle activity or rigidity, head and
body motion, sustained eye movements, improper sensor placement and unusual or excessive electrical
interference.
CAUTION Precautions about the System
Do not assess the patient's condition only with the information from this device. A clinical judgment based on
the information from the device should be made by a doctor who fully understands functions of the device, in
a comprehensive manner combined with clinical findings and other test results.
Do not assess the patient's condition only by the alarm generated on this device. When the alarm is set to OFF
or if the alarm priority is low, a sudden change of the patient may not be noticed.
xii
Safety Precautions for Using the Device
If an alarm generates, check the patient's condition first and ensure the safety. Depending on the alarm, take
appropriate measures to remove the problem. If the problem lies with the alarm setting, set the alarm properly.
When measuring for a long period of time, make sure not to compress the patient with the lead cables and the
electrodes. Compressing the same site for a long duration may inhibit the blood flow and generate compression
necrosis and burn injury.
Use only the products specified for this device. If unspecified products are used, proper function cannot be
executed.
Do not attach film to the touch panel. This may result in malfunction or failure.
For quality improvement purposes, specifications may be subjected to change without prior notice.
This device utilizes LED for the backlight. Since this LED deteriorates by the life cycle, the display may
become dark, scintillate, or may not light by the long term use. In such case, contact your nearest service
representative.
This device is intended to be used for only one patient.
The installation of this device should be performed by our service representative or a person who is well
acquainted with this device.
If not using the device for a long period, disconnect the power cable and lithium-ion battery.
Although the LCD utilizes highly accurate picture elements, occasionally, there may be a few pixels which do
not light or constantly light. Please note that this is not an device failure, and will not affect monitoring
operation.
Exposing LCD panel to intense light may deteriorate display property. Do not expose the panel to direct
sunlight or strong ultra violet (UV) light.
The lithium-ion battery can only be charged in the specified operating temperatures of the device. Refer to the
operation manual of the lithium-ion battery (BTO-005) for details.
CAUTION Precautions about the ECG Monitoring
If any electrodes get detached from the patient after being connected to the lead cable and patient monitor, pay
attention that the metal part of the electrode does not get in touch with any metal parts of the bed or any
conductive parts. Also, the operator should not touch any conductive parts with bare hands. Otherwise, it may
cause electric shock to the patient and/or operator due to excessive leakage current.
The indication for continuous use of the electrode is about one day.
Replace the electrode if the skin contact gets loosen due to perspiration, etc.
When an electrode is attached to the same location for a long period, some patients may develop skin irritation.
Check the patient's skin condition periodically and change the electrode site as required.
For stable arrhythmia detection and ECG monitoring, verify proper electrode placement, lead, waveform size,
and filter mode selection. If not properly selected, it may cause erroneous detection.
The threshold level for arrhythmia detection changes with ECG waveform size. Set a proper waveform size for
monitoring.
When the ECG waveform size is x1/4, x1/2, or x1, the arrhythmia detection level is 250 μV.
When the ECG waveform size is x2 or x4, the arrhythmia detection level is 150 μV.
The leads for arrhythmia detection, central monitor display, printing are fixed as ECG1 and ECG2. Set the most
appropriate leads with high QRS for ECG1 and ECG2, especially for arrhythmia detection. If the QRS
amplitude for the set lead is low, it may cause erroneous arrhythmia detection.
In ESIS Mode, artifacts such as electrosurgical noise or EMG can be largely reduced, but QRS amplitude
attenuation, waveform distortion, or ST segment change may occur compared with other filter modes.
The ESIS mode cannot completely reduce the electrical noise, and may erroneously detect the pacemaker
spike. This mode should be selected only when a high frequency noise largely affects the HR measurement.
There are some cases when the pacemaker pulse cannot be detected depending on the pacemaker type, pulse
voltage, pulse width, electrode lead type (unipolar, bipolar), or electrode placement which causes the
xiii
Safety Precautions for Using the Device
pacemaker pulse amplitude to decrease, and disables the pacemaker pulse detection.
If signals similar to a pacemaker pulse are present, such as electric blanket noise or excessive AC frequency
noise, these may be erroneously detected and displayed as a pacemaker pulse.
When a spontaneous QRS and pacemaker pulse overlap (ex. fusion beat, etc.), QRS detection cannot be
performed properly. In this case, the heart rate is degraded.
If a pacemaker pulse is continuously detected due to AC frequency interference, QRS detection will be
suspended and the heart rate will be reduced. Arrhythmia will not be detected either.
CAUTION Precautions about the ST Measurement
The ST algorithm has been tested for accuracy of the ST segment data. The significance of the ST segment
changes need to be determined by a clinician.
For the lead which the electrode is detached, the reference waveform setup cannot be performed. Check if the
electrode is appropriately attached, and perform the setup again.
CAUTION Precautions about the 12-Lead Analysis
Interpretation and Minnesota codes given by this device do not instruct the physician as to the kind and degree
of cardiac disease. Accordingly, four value judgments are given for ECG waveform and although "abnormal"
indicates a large possibility of organic cardiac disease, there are cases where no cardiac disease exists despite
an abnormal ECG (that is, an abnormal ECG may be caused by something other than heart).
On the other hand, care should be taken in the event any preclinical coronary arteriosclerosis could be present
despite a normal ECG interpretation.
Therefore, for a proper diagnosis, the ECG should be integrated with other interpretations.
ECG Recording by the Mason-Likar System
The 12-lead ECG recorded with the torso placement of the limb leads (Mason-Likar 12-lead system) may differ
from the standard 12-lead ECG. Moreover, waveforms may differ somewhat also in a supine position and a
standing position (sitting position).
Fukuda Denshi recommends to carry out the recording of the ECG by taking into consideration the waveform
differences according to electrode positions or postures.
For the model installed with ECG analysis program
The ECG analysis program is intended to analyze the standard 12-lead ECG waveforms. Therefore, the
analyzed result for waveforms recorded with the torso placement of the limb leads (Mason-Likar 12-lead
system) may differ from that of the standard 12-lead ECG waveforms.
Select "Used" for the pacemaker setting on the patient admit/discharge menu if a patient has a pacemaker.
The threshold values for classification of 12-lead ECG interpretation and Minnesota code are set by age and
sex as follows:
1. Male and Female of ages 19 years old and above
2. Male of age 12 through 18 years old
3. Female of age 12 through 18 years old
4. Male and Female of ages 3 through 11 years old
5. Male and Female of ages below 2 years old
If no patient information (i.e. Default : "Class." [Adult], "Sex": undetermined, and "Age" [0]) has been entered,
the system algorithm will handle the patient as a "35 years old male".
Before the analysis, make sure the patient classification ([Adult] / [Child]) is properly selected. If [Neonate] is
selected, the 12-lead ECG analysis will not function.
Enter the age of patient if known. If no age information (i.e. Default: [0]) has been entered, the system
algorithm will handle the patient as "35 years old".
Enter the sex of patient if known. If no sex information (i.e. Default: undetermined) has been entered, the
system algorithm will handle the patient as "Male".
If the patient classification is set as [Child] and no age (i.e. Default: [0] ) has been entered, the system algorithm
will handle the patient as "less than 2 years old""
xiv
Safety Precautions for Using the Device
xv
Safety Precautions for Using the Device
xvi
Safety Precautions for Using the Device
If the CO2 gas calibration is not performed at a specified interval, CO2 measurement accuracy may be affected
and also subsequent gas calibration may not be possible.
When the accumulated measurement time exceeds 1,200 hours from the first use.
However, if the first calibration was performed before the accumulated measurement time reaches 720 hours,
another calibration is required when the accumulated measurement time exceeds 1,200 hours from the first
calibration.
When 12 months has elapsed or the accumulated measurement time has exceeded 4,000 hours from the
previous calibration.
When EtCO2 measurement is not stable or accuracy is degraded compared with other measuring device.
When the patient monitor was not used for a while, or when EtCO2 was not measured for a while.
Perform the calibration 5 minutes after turning ON the power on the HCP-810/HCP-820.
Do not disconnect the sampling tube during calibration. If the sampling tube is disconnected, calibration will
cease.
Dispose of calibration gas according to the regulation of each medical institution.
The Microstream™ EtCO2 sampling lines are designed for single patient use, and are not to be reused. Do not
attempt to clean, disinfect, sterilize or flush any part of the sampling line as this can cause damage to the
monitor. It may cause crass-infection.
Dispose of sampling lines according to standard operating procedures or local regulations for the disposal of
contaminated medical waste.
Before use, carefully read the Directions for Use for the Microstream™ EtCO2 sampling line.
Use only the Microstream™ EtCO2 sampling line to ensure proper function of the monitor.
CAUTION Precautions about the CO2Monitoring (HPD-810/HPD-820 Gas Unit I/F)
The disposable airway adapter should be opened just before use.
Do not reuse the disposable airway adapter. If sterilized, it will become unusable.
The airway adapter should be attached with the thicker side facing to the patient. If attached oppositely, it may
damage the CO2 sensor or airway adapter.
Dispose of calibration gas according to the regulation of each medical institution.
CAUTION Precautions about the BIS Monitoring (HBX-800 BISx I/F Unit)
The conductive parts of sensors and connectors should not contact other conductive parts, including earth.
To reduce the hazard of burns in the high-frequency surgical neutral electrode connection, the BIS sensor
should not be located between the surgical site and the electrosurgical unit return electrode.
The BIS sensor must not be located between defibrillator pads when a defibrillator is used on a patient
connected to the patient monitor.
Continuous impedance checking may need to be disabled if the 1 nA, 128 Hz impedance check signal interferes
with other device (e.g., evoked potential monitors).
Considerations when using Electro-Convulsive Therapy (ECT) device during BIS monitoring: Place ECT
electrodes as far as possible from the BIS sensor to minimize the effect of interference. Certain ECT device
may interfere with the proper function of the BIS monitoring system.
CAUTION Precautions about the Alarm
Alarm messages will be displayed according to the priority. (Level S > Level H > Level M > Level L> Level N)
For the same alarm level, the alarm message for the newer alarm will be displayed.
The arrhythmia alarm message other than Tachy, Brady, Ext Tachy, Ext Brady will continue to be displayed
for 30 seconds after the alarm is resolved.
When "LEAD OFF", "Check Electrodes" is displayed, HR alarm or arrhythmia alarm will not function. If this
condition is left unresolved, a sudden change of the patient may not be noticed. Take prompt action when the
lead-off condition is detected.
xvii
Safety Precautions for Using the Device
When CO2 is measured on the HPD-810/HPD-820 Gas Unit I/F and HCP-810/HCP-820 CO2Gas Unit , the
upper EtCO2alarm will not generate if the upper limit is set to 100 mmHg/13.4 kPa and above as the
measurement range is 0 to 99 mmHg / 0 to 13.3 kPa.
Whether to use the SpO2 second alarm function and its threshold selection should be based on the patient's
clinical indication/portent and medical evaluation.
If the SpO2 alarm andsecond alarm setup is set to [OFF], the second alarm integral value will be set to 0.
Pay attention not to set the alarm volume too low to avoid missing any important alarms.
On a wired network, the alarm generated on the bedside monitor will be output to the network with a maximum
delay of 1 second, and to the central monitor with a maximum delay of 2.5 seconds.
If the NIBP alarm is turned OFF under the Oxygenator Mode, NIBP auto mode measurement and NIBP
measurement at alarm occurrence will not be performed.
If the same or similar devices with different alarm settings are used in the same facility or same department,
pay attention not to misjudge the alarms.
To ensure that the alarm setup is appropriate for the patient being monitored, check the setup each time this
device is used.
CAUTION Precautions about the System Setup
When the waveform and numeric data display for each parameter is set to OFF, the alarm and trend input will
be also suspended.
If the HR/PR source is set to [BP], and if BP waveform/numeric data is set to [Disp. OFF], the PR value will
not be displayed.
If the HR/PR source is set to [SpO2], and if SpO2 waveform/numeric data is set to [Disp. OFF], the PR value
will not be displayed.
If the RR source is set to [CO2/GAS], and if CO2 waveform/numeric data is set to [Disp. OFF], the RR value
will not be displayed.
If the RR source is set to [CO2/GAS], and if GAS waveform/numeric data is set to [Disp. OFF], the RR value
will not be displayed.
Do not set the same remote control ID to multiple monitors in the same floor. Otherwise, the remote control
operation may control multiple monitors at the same time.
After the remote control setup, check that the remote control unit is properly operating.
If the time/date is not correctly set, or changed during monitoring, malfunction may occur with NIBP
measurement, periodic printing, trend, NIBP list data, and age calculation from the birth date.
If the time/date is changed, the time/date for all the saved patient data (trend, list, recall, etc.) will also change.
The printed time/date before changing and the displayed time/date after changing will differ.Also, the data
transmitted to the central monitor before the time/date is changed will be displayed on the central monitor with
the previous time/date.
CAUTION Precautions about the Multigas Unit Data Monitoring
(MGU-800/MGU-810 Series)
When the multigas unit (MGU-800/MGU-810 series) and HPD-810/HPD-820, HCP-810/HCP-820 are
simultaneously used, the CO2 concentration measurement will be performed by the device selected for the
"CO2 Source Priority" under ( [Menu] > "Parameter" [CO2]) .
The MGU-800/MGU-810 series require warm up of about 10 minutes to correctly measure the data.
If the power supply is interrupted due to power failure, etc., MGU-800/810 series multigas unit will initialize
and enter into warm-up mode even if the power interruption is within 30 seconds.
About the Gas Calibration
The gas calibration will automatically start when the MGU-800/810 series multigas unit is connected.
After the warm-up completes, gas calibration will be performed every 4 hours during stable operation.
During warm-up, gas calibration interval will become shorter than during normal operation.
During gas calibration, measurement data will not be updated.
xviii
Safety Precautions for Using the Device
xix
Safety Precautions for Using the Device
Pay attention not to allow chemical solution to enter the device or connectors.
Do not use organic solvents, thinner, toluene or benzene to avoid damaging the resin case.
Do not polish the device with abrasive or chemical cleaner.
When disinfecting the entire room using a spray solution, pay close attention not to get any solution into the
device or connectors.
Use only neutral detergent to clean the device. The surface resin coating may damage, resulting in
discoloration, scratches, and malfunction.
Example:
chemical cloth, scrub brush, abrasive, polishing powder, hot water, volatile solvent and chemicals (cleanser,
thinner, benzine, benzol, and synthetic detergent for house and furniture), or sharp-edged tools
Do not open the housing.
Do not allow alcohol or other liquids to enter the device.
Replace the periodic replacement parts periodically as specified.
WARNING
Do not connect unspecified device to the wired network.
Do not mix devices with DS-LANII and DS-LANIII setting in the same wired network. The network may cease
and proper monitoring may not be possible.
CAUTION
If performing wired network transmission, configure the display so that the numeric data corresponded to the
waveform is displayed. If not, the displayed waveform or numeric data may not be transmitted.
The default setting of bed ID is "000". If connected to a wired network with the bed ID unchanged, monitoring
on the central monitor will not be possible.
When connecting to a wired network, make sure that there are no other bedside monitors with the same ID. If
there is more than one bedside monitor with the same bed ID, the duplicated bedside monitors cannot be
monitored on the central monitor.
When connected to the DS-LAN II network, set the bed ID in the range from "001" to "048" .
When connected to the DS-LAN III network, set the bed ID in the range from "001" to "100" .
The alarms that can be notified to the central monitor depend on the model type and software version of the
central monitor. For details, refer to the operation manual of the central monitor.
There are following restrictions when connecting the DS-8400 System to the wired network.
The BP measurement unit setting should be the same for all central monitors and bedside monitors. If the
setting is different among the monitors, data such as BP waveform, BP numeric data, NIBP numeric data,
NIBP list will not be transmitted. It will be treated as not measured data, and will not be displayed on the
central monitor. The alarm limit setup from the central monitor cannot be performed either.
On the DS-LAN II network, the following arrhythmia alarms will not be transmitted.
TACHY, BRADY, COUPLET, PAUSE, TRIGEMINY, TRIPLET, EXT TACHY, EXT BRADY, R on T,
MULTIFORM, VENT RHYTHM, SVT, IRREGULAR RR, PROLONGED RR, S FREQUENT, S
COUPLET, VPC, SVPC, PACER NOT CAPTURE, PACE NOT PACING
On the DS-LAN II network, arrhythmia alarm of "SLOW VT" will be transmitted as "VT" .
On the DS-LAN II network, waveform, numeric data, and alarm of BP7, BP8, TEMP3 to 8 will not be
transmitted. (These can be transmitted on the DS-LAN III network). Also, the displayable waveform,
numeric data, and alarm will differ depending on the central monitor model type. Refer also to the operation
manual for the respective central monitor.
The PR_IBP alarm will not be transmitted to the central monitor.
xx
Safety Precautions for Using the Device
If the "RR/APNEA alarm source" is other than [Impedance] (or, if [Auto] selects a setting other than
[Impedance]), the RESP waveform will not be transmitted on a wired network.
If the "RR/APNEA alarm source" is other than [CO2/GAS] (Or, if [Auto] selects a setting other than [CO2/
GAS]), the CO2 waveform will not be transmitted on a wired network.
For the numeric data displayed as "xxx", maximum or minimum value of measurable range will be
transmitted.
The numeric data displayed as "--- " will be treated as not measured data.
If the measurement unit of CO2 concentration is mmHg and [99mmHg] is selected for "Upper Limit of CO2
(mmHg) Transmission" ([Initial Settings]>[System]>[DS-LAN]), the CO2 value of 100 mmHg or above
will be transmitted as 99 mmHg.
As the DS-8400 System donot have the arrhythmia template display and 12-lead ST display function,
waveforms and other data will not be displayed for these displays on the central monitor connected to the DS-
LAN network.
When connected to the wired network, the time/date will synchronize with the central monitor. Even if the
time/date is changed on the DS-8400 System, it will be corrected to the time/date of the central monitor.
The ST display will be distorted on the central monitor if the ECG lead (ECG1 or ECG 2) is changed on the
DS-8400 System. Redrawing the ST display will return the display to normal.
On the central monitor, the respiration waveform and RR value based on the RR/APNEA alarm source selected
on the DS-8400 System will be displayed. The monitored RR and APNEA will be the same for the central
monitor and the DS-8400 System.
DANGER
When monitoring a patient using wireless telemetry, make sure the patient data is properly received at the
central monitor. Pay special attention when the channel ID at the bedside monitor is changed.
WARNING
A password can be set to access the channel ID setup menu to allow only the telemetry channel administrator
to change the channel ID.
Some type of wireless combinations may generate interference with other telemetry.
Before selecting a channel, verify it will not interfere with other channels.
Inform the supervisor of the use of telemetry channels to avoid interference with other telemetry.
If transmitters are used in a neighboring medical facility, your facility and the neighboring facility must make
agreements on the setting of the telemetry channels to prevent telemetry interference.
xxi
Safety Precautions about the Ventilator Monitoring
If the measurement unit of CO2 concentration is mmHg and [99mmHg] is selected for "CO2 (mmHg) Upper
Limit of Transmission" ([Initial Settings]>[System]>[Telemeter]), the CO2 value of 100mmHg or above will
be transmitted as 99mmHg.
CAUTION
This device is equipped with a built-in clock. When the power of this device is turned OFF, this clock is backed
up by a lithium primary battery. If incorrect time is displayed when turning ON the power, a low battery may
be the cause. In such case, contact Fukuda Denshi service representative for replacing the battery.
To protect the data during voltage dip, short interruptions and voltage variations on power supply input lines
or during short duration of power turned OFF, this device performs 5-minute (approx.) data backup using the
secondary battery. The data may not be protected if the power is turned OFF within 30 minutes from power ON.
WARNING
CAUTION
CAUTION
xxii
Safety Precautions about the NIBP Cuff
The purchase or possession of this product does not explicitly or implicitly permit the use of this product with
replacement parts. Whether the device is used alone or in combination with other parts, it is protected by the
patent rights of the related devices.
CAUTION
Some of the NIBP cuffs used for this device contain natural rubber latex which may cause allergic reactions.
(FDA: Medical Alert on Latex Products, "Allergic Reactions to Latex-Containing Medical Devices", Food &
Drug Administration, 10903 New Hampshire Avenue, Silver Spring, MD 20993, 1991.)
CAUTION
When disposing of this device, accessories, or components, use an industrial waste distributor. Do not dispose
of as ordinary waste.
When disposing of the battery, separate it from other wastes and contact your nearest service representative.
CAUTION
When transporting this device, pack it with specified packing materials.
Also, transport it under appropriate environment condition.
( Operation Manual "Specification" P14-1)
When the power failure is less than 30 seconds, monitoring will resume with the display mode and patient
information unchanged. When the power failure is 30 seconds or more, monitoring will resume with the default
display mode set by the user, or the display mode which was last set.
However, this is only if the device was operated for 30 minutes or more before the power failure.
MGU-800/810, HCP-810/820, HPD-810/820 will start up from the warm-up mode. The warm-up time differs for
each unit.
Accessories/Optional Accessories
The ECG electrodes are consumable products. Always prepare extra supplies of electrodes.
Verify that there is no wire break on the patient cable once a week.
Battery Pack
Even if the battery pack is not in use, the remaining capacity decreases due to self-discharge.Make sure to
verify once a week that the battery pack is fully charged.
To fully charge the empty battery pack, it takes 5.0 hours during operation, and 2.5 hours when the power is
OFF and AC cable is connected.
The performance of the battery deteriorates with repeated use. To ensure performance of the battery, it is
recommended to replace it once a year.
xxiii
Safety Electromagnetic Compatibility
Electromagnetic Compatibility
This equipment complies with EN 60601-1-2: 2015, safety standard regarding the electromagnetic disturbances of
medical electrical equipment. To ensure maximum performance against the electromagnetic disturbances, make sure
to follow the precautions for installation and usage described in this manual.
This equipment is intended for use in the medical facility (except in the vicinity of MRI device), and satisfies
the immunity level for professional healthcare facility environment stipulated in EN 60601-1-2: 2015.
When using this equipment, interference with other medical electrical equipments or non-medical electrical
equipments may occur. Make sure that no interference is present before usage.
This equipment is a ME equipment which intentionally receives RF energy of specific reception frequency. RF
electromagnetic radiation from other equipment for the intended specific reception frequency band may cause
radio interference. Make sure that the reception is properly made in the used environment.
To ensure basic safety and essential performance related to electromagnetic disturbances during the expected
service life of this equipment, "Daily Check" and "Periodic Check" must be performed.
(Refer to "Chapter 10 Maintenance Check" of the Maintenance Manual.)
If any sorts of electromagnetic wave, magnetic field, or static electricity exist around the device, noise interference
or malfunction of the device may occur. If any unintended malfunction or noise occurs during monitoring, check the
magnetic influence and take appropriate countermeasures.
The following are examples of the common cause and countermeasures.
xxiv
Safety Electromagnetic Compatibility
external antenna) with the specified cable should be used in a location at least 30 cm apart from any part of this
equipment. Otherwise, it may result in performance degradation of this equipment.
EMC Guidance
This equipment complies with EN 60601-1-2: 2015. However, if portable transmitter or wireless LAN equipment is
used extremely nearby, the electromagnetic influence may largely exceed the compliance level and may cause
unexpected phenomenon such as noise interference on the waveform, etc.
Also, if this equipment is installed close to, or stacked with other equipment, malfunction may occur. Make sure to
verify that the equipment operates properly in a used location.
This equipment should be used in a location specified by each medical institution.
If any unexpected noise interference on the waveform or failure to the peripheral device occurs, stop using the
equipment and follow the instruction of the technical engineer.
This equipment complies with EN 60601-1-2: 2015 for the following system configuration.
Main Unit: DSC-8410
Display Unit: LC-8018TC
Patient Monitor: DS-8007
Multi Module: HM-801
Multi Module: HM-800 x 2
Recorder Unit: HR-800
Telemetry Transmitter Module: HLX-801
Gas Unit I/F: HPD-820
Input Box: IB-8004
Multiport Module HP-800
Multigas Unit: MGU-810
BISx I/F Unit: HBX-800
BISx Module: BISx
Lithium-Ion Battery: BTO-005
xxv
Safety Electromagnetic Compatibility
Harmonic Emissions
Class A
IEC 61000-3-2
CAUTION
The emission performance of this equipment is suitable for use in industrial environment and
hospital environment (EN 55011 Group 1 Class A). Do not use in home environment
(generally, EN 55011 Group 1 Class B is required).
xxvi
Safety Electromagnetic Compatibility
CAUTION
The assumed service TETRA 400 of the test frequency of 385 MHz is a public wireless
device, and this product, which is intended for use in a professional healthcare facility
environment, has not been tested as it will not be radiated in close proximity.
The assumed service GMRS 460, FRS 460 of the test frequency of 450 MHz is a US
standard wireless device, and this product, which is intended for use in Europe, has not been
tested as it will not be radiated in close proximity.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
xxvii
Safety Electromagnetic Compatibility
xxviii
Chapter 1 General Description of the Device Composition of the System
The DS-8400 system is composed of Display Unit (LC-8016TC/LC-8018TC), Main Unit (DSC-8410), Super Unit
(HS-8000 series/DS-8007 series), Multigas Unit (MGU-800 series/MGU-810 series), Recorder Unit (HR-800),
expansion modules and Input Box (IB-8004).
DSC-8410 Yes No No
DSC-8410 (with CC-84) Yes 2ch 1ch
1-1
Chapter 1 General Description of the Device Composition of the System
*: RR_SpO2, SpMet, SpCO, SpHb, SpOC, and PVI measurements are optional functions.
Lineup of Adapter
Model Type Remarks
HSA-80 Adapter for HS-8000 series (Connects to the patient monitor with module connection cable.)
Adapter for HS-8000 series (Connects to the patient monitor by sliding on the rail at the rear side
HSA-81
of the patient monitor.)
DSA-82 Adapter for DS-8007 series (Connects to the patient monitor with module connection cable.)
Yes
MGU-811P Yes Yes Yes
Paramagnetic
Lineup of Recorder
Model Type Remarks
HR-800 50 mm Roll Paper, 3 waveforms printing
IB-8004 4
SpO2 Module M
HG-810 SpO2, SpCO*, SpMet*, SpHb*, PVI *, PR
(Masimo)
SpO2 Module N
HG-820 SpO2, PR
(Nellcor)
1-2
Chapter 1 General Description of the Device Features
HBX-800 BISx I/F Unit BIS, SQI, EMG, SR, EEG, SEF, TOTPOW
Features
Various displays such as enlarged numeric data, trend, or ventilator can be selected according to monitoring
conditions.
The operation can be performed with the touch panel. Also, frequently used keys can be programmed as user
key.
It is possible to connect two types of display units in addition to the main display unit to extend the display.
An optional mouse can be connected allowing mouse operation. (However, some operations are not possible.)
The alarm indicator notifies the alarm with different flashing patterns corresponding to the alarm level so that
the users can easily identify the alarm level of the generating alarm.
Remote control is possible using the optional remote control unit.
Using the CF-820 IR Remote Control Unit allows to remotely control the patient monitor.
By using the multiparameter amplifier, the HS-8000 series Super Unit is capable of monitoring parameters in
combination of BP (maximum 6 channels), temperature (maximum 6 channels), and CO (maximum 1channel).
In addition to ECG, respiration, SpO2 (pulse wave), BP, NIBP, temperature and CO, it is also possible to
measure CO2 and BIS as optional function.
By using the multiparameter amplifier, the DS-8007 is capable of monitoring basic parameters of BP (max. 4
ch.), temperature (max. 6ch.), CO (max. 1ch.), and optional parameters of CO2 and BIS.
27 types of arrhythmia can be analyzed.
By using the optional SpO2 Module (HG-810/HG-820), arterial oxygen saturation can be also measured. By
using the system with the Super Unit, arterial oxygen saturation measured at 2 different sites can be monitored
as additional parameter.
This system uses pulse oximetry to measure and display functional oxygen saturation in the blood. There are
two model types with different built-in SpO2 modules, which are Covidien/Nellcor and Masimo.
SpCO, SpMet, SpHb, PVI measurement are optional function available when using the HS-8312M/DS-8007M
and HG-810 with built-in Masimo SpO2 module.
SpOC measurement is optional function available when using the DS-8007M.
RR_SpO2 is optional parameter which can be measured on the DS-8007N with the built-in Covidien SpO2
module.
By using the optional HP-800 Multiport Module, or connecting the ventilator to Status II port on the main unit,
airway flow, airway pressure waveform, minute ventilation, airway resistance, etc. can be monitored. Also,
ventilator alarm can be notified to the central monitor via wireless and wired network. The following
ventilators can be connected.
SV-300/300A
1-3
Chapter 1 General Description of the Device Features
1-4
Chapter 1 General Description of the Device Menu Configurations
Menu Configurations
Menu Screen
The menu screen is a group of shortcut keys to jump to each menu.
The menu is composed of the following 10 groups and can be accessed from the menu screen.
Function Groups Displayed Menu
Admit/Discharge Admit/Discharge
Other Bed Other bedside monitors connected to the DS-LAN will be displayed.
REFERENCE
Other than the "Initial Settings", the items to be displayed on the menu screen can be
customized by groups.
(
Maintenance Manual "Menu Setup" P5-21)
Admit/Discharge
Admit/Discharge Mode Selection
ID, Name, Classification, Sex, Team, Birth Date, Age, Height, Weight, BSA,
Blood Type (ABO, Rh), Pacemaker, Impedance Meas., Admit Date/Time
Monitor Suspend
Discharge
Basic Setup
Display Configuration Layout, Background, Palette, Detail Setup, Meas., Wave (Sweep Speed), Short Trend,
User Key
Manual Printing Basic (Printer, Waveform, Print Duration, Delay Time), 12-Lead (12-Lead Waveform/
Analysis Format, Position, Wave Format, Print Calibration, Printer Auto Scale, Lead
Boundary), Other Setup (Graphic Printing, Recall Printing), Common (QRS Classific.,
Speed, Calibration: Print Calibration, Print NIBP Data)
Auto Printing Alarm Printing (Print, Printer, Waveform, Print Duration), Periodic Printing (Print, Printer,
Periodic Interval, Waveform, Print Duration), Common (QRS Classific, Speed, Print
Calibration, Print NIBP Data)
Tone/Volume Vital Alarm Sound, Ventilator Alarm Sound, Status Alarm Sound, Tone Source, Key
Sound, Other Bed Alarm Sound, Boot/Shutdown, Other
Brightness Brightness
Night Mode Night Mode, Detail Setup (Volume, Display, Alarm Indicator)
1-5
Chapter 1 General Description of the Device Menu Configurations
Alarm
Basic The parameters to be displayed are selectable.
Circulatory Alarms for HR, Ext Tachy, Ext Brady, PR, SpO2, Ext SpO2, NIBP, BP,TEMP can be set.
List List of alarm ON/OFF setting and lower/upper limits, Meas. List/All List, Print Setup, Recall
Setup
Detail Setup Alarm Suspend Time, Alarm Silence Time, Alarm Silence, Alarm Sound Suspend, Status
Alarm Control, Alarm Limit Display
Parameter
ECG Arrhy., Arrhy. Alarm Setup, ST Setup, HR
Detail Setup
Filter, Synchronized Mark/Tone, Pacemaker, Pacemaker Pulse, Pace Pulse Mask Time,
HR Average, HR Delay, ECG Drift Filter, AC Filter, Auto Lead, 3Lead Override, ST/VPC/
Arrhy. Alarm Display, ECG Analog Output, ECG waveform display during Lead-OFF,
Noise Detection
RESP Size, Common Setup (RR Synchronized Mark, RR/Apnea Alarm Source), RR, Apnea,
Alarm Assist, Display ON/OFF
Impedance Setup
CVA Detection, Impedance Measurement, Impedance Detection Lead, Impedance
Detection Level
Detail Setup
Patient Classification, Dyna Alert, Oscillograph, PR Display, NIBP Erase Time, Measure
at Alarm, Quick Measurement, Sight Inflation, MAP, End Tone, User Interval, Auto Mode
with Start/Stop key, Time Display, Periodic Measurement Starting Time, Oscill. Print,
Target Inflation Value
BP BP Zero (BP1 to BP8), Scale Selection, Label, Alarm Assist, Display ON/OFF
Detail Setup
Synchronized Mark/Tone, Display Type, Wave Filter, Mean Wave, Respiration Filter, IBP
Analog Output, Alarm during NIBP, ART Catheter Check Message
1-6
Chapter 1 General Description of the Device Menu Configurations
Detail Setup
EtCO2 Peak Duration, N2O Compensation, Atmospheric Pressure, O2 Compensation,
Anesthetic Compensation, CO2 Source Priority
BIS Scale, Alarm Assist, TREND-E
Common Setup
Short Trend 2nd Parameter, Smoothing Rate, Continuous Impedance Check, EEG Filter
Data Review
Graphic Trend Latest Data, Alarm Review, Trend Group, Alarm Disp. Sel., Print
Tabular Trend Latest Data, Alarm Review, List Group, Setup, Print, Print (All)
Waveform Review
Zoom Wave Latest Data, Alarm Review, Meas., Print, Delete
ST ST Waveform, Reference Waveform, Setup, Slide Show, Size, Latest Data, Print
Full Disclosure Waveform Latest Data, Alarm Review, Slide Show, Time Search, Size/Scale, Setup, Alarm Display,
Print
Calculation
Hemodynamics Input Data, Edit, calculation results list, New Regist., Index Display, Print
Lung Function Input Data, Edit, calculation results list, New Regist., Index Display, Print
Other Bed
Other Bed Display Area Selection (Area 1 to 5), Other Bed Alarm Sound, Alarm Display, Area Setup (Area 1
to 5), Bed List
Other Bed Display Area Selection (Area 1 to 5), Other Bed Alarm Silence, ON/OFF of menu title display,
Waveform Selection
Initial Settings
Alarm - Alarm System, Basic Alarm Parameter, Asystole, VF, VT Alarm,
Oxygenator Mode Setup, Buzzer Tone at Speaker Failure, Suspend Arrhy.
Analysis during Noise Interference, Lower Limit for Alarm Volume, Alarm
Indicator, Alarm Level, HR/PR Lower Limit during Alarm Auto Setting,
Alarm Threshold Limit, Auto Alarm Setup
1-7
Chapter 1 General Description of the Device Menu Configurations
User I/F Display/Print Date Format, BP Alarm Increment, Trend Clip, BP Printing Scale, Night
Mode Cancel, ST Display Lead Setup, VENT Display Parameters, Hemo/
etc Display Parameters, Auto Display Configuration, Dim All Data Other
than Numeric, All Window Opaque, Printer Message Display, Message
Icon, Operation Guide Display, Notification when Changing Device
Configuration, Sync wave size/scale of extended display with main unit,
12-lead Analysis Filter Display, Waveform Size Display, Shift Time (Day
Shift, Twilight Shift, Night Shift), Key Group Setup, Event Label Setup, RR
Alarm Increment, Patient Name on the Information Display Area, External
Device Numeric Data Box Operation, Drug Calculation, Monitor Suspend
Label, Monitor Suspend Timer
Power ON/ Discharge Check Discharge at Power ON, Discharge Mode, NIBP Resume Auto
Mode with Manual Measurement., Backup Setting at Power ON/Discharge
Key Mask Items not to be displayed on the menu screen can be selected.
Remote Control Remote Control Key Function, Room ID, Bed ID
External Device Main Unit Port COM, Status II, U-LINK, Numbering of HP-800
HP-800 Ventilator (SV-300, SERVO-i/s, SERVO-U/n/air, PB, Evita) , SvO2/CCO
(Vigilance, PiCCO, PulsioFlex), GAS/SPIRO (MGU-800, MGU-810), Other
(PC Comm., HLX, Barcode, Magnetic Card, BIS, aepEX, INVOS, FLOW-
i, TCM4/TCM5, PC Comm. (DS-5000), Anes.GE, Anes.Draeger, Apollo
Draeger), Trend Data Setup
Magnetic Card Reader Data digits for each patient information, Auto Reference to Central Monitor
when Reading Patient ID
Other AC filter, Search Patient ID, Data Transfer, Data Selection for Transfer,
Numeric Data External Output, HLX, HR-800, Extended Display Unit
Administrator Setup Key Lock Key lock for each function can be set.
Maintenance
Maintenance Program Version, External Media, GAS Accuracy Check, Parts Usage Time, Install,
Module Install, Test Menu
1-8
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
WARNING
Do not connect any device or cable not authorized by Fukuda Denshi to any I/O connector.
If unspecified device or cable is connected, not only that this device cannot deliver its
maximum performance, the device may be damaged and safety cannot be ensured.
Front Side
1 Display Unit Attaching Position
Attaches the display unit (LC-8018TC/LC-8016TC).
1
2 Display Unit Connector
Connects the display unit.
Rear Side
1 HLX Fixing Position
Fixes the Telemetry Transmitter Module (HLX-801). 1
2-1
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Left Side
1 Super Unit Connector
Connects the specified device.
2
2 Module Connector
Connects the specified device.
1
3 Super Unit Release Lever
Releases the Super Unit from the main unit.
3
Right Side
The illustration shown on right is when the optional CC-84 and HR-800 are installed.
No. 11 to No. 14 described below can be used only when the
CC-84 is installed.
1 Card Slot (CF, CFast)
A slot to insert the specified external media 15
1
2 Card Access Indicator (CF, CFast) 2
9 U-LINK Connector
Connects the Recorder Unit (HR-800) and Multigas Unit (MGU-800/MGU-810 series).
10 module-LAN Connector
Connects the HS Adapter (HSA-80), DS-8007 Adapter (DSA-82), or the Input Box (IB-8004).
2-2
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
This section explains about the LC-8018TC (18.5 inch), LC-8016TC (15.6 inch).
Front Side
1 Remote Control Sensor 5 5
4 Standby Switch
Sets ON/OFF the standby condition.
5 Alarm Indicator
Lights/blinks when the alarm generates.
Red: Level H (Urgent Alarm, Alarm Priority/High)
Yellow: Level M (Cautionary Alarm, Alarm Priority/Medium)
Blue: Level L (Status Alarm, Alarm Priority/Low)
6 Fixed Keys
(
"Fixed Keys" P3-1)
NOTE
If the battery charging LED flashes, battery charging error is occurring. Remove the battery
and install it again. If the error persists, contact your nearest service representative.
2-3
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Rear Side
1 Display Unit Attaching Position
Attaches to the main unit.
1
2 Mouse/Keyboard Connector
Connects the optional mouse.
CAUTION
The display unit utilizes LED for the backlight. Since this LED deteriorates by the life cycle,
the display may become dark, scintillate or may not light by the long term use. In such case,
contact your nearest service representative.
Front Side
1 NIBP Start/Stop Key
Starts/stops the NIBP measurement.The indicator 1
lights during the NIBP measurement. 2
7
3
2 BP Zero Balance Indicator 4
Performs BP zero balance.The indicator lights during
the BP zero balancing. 5 8
5 ECG Connector
Connects the ECG cable.
6 AUX Connector
Connects the Gas Unit I/F (HPD-810), CO2 Gas Unit (HCP-810), or BISx I/F Unit (HBX-800).
7 Multiparameter Connector
Connects the input cables for BP, TEMP or CO.
8 NIBP Connector
Connects the NIBP air hose.
9 SpO2 Connector
Connects the SpO2 sensor, or relay cable (patient cable).
2-4
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Rear Side
1 HS Adapter Connector
Connects the HSA-80 HS Adapter.
HS Adapter: HSA-80
Front Side
1 Super Unit Connector
1
Connects the HS-8000 series.
2 Release Lever
Releases the HS-8000 series from the HS
Adapter.
Rear Side
1 module-LAN Connector
Connects the Main Unit or IB-8004.
2-5
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side
1 Super Unit Connector 1
Connects the HS-8000 series.
2 Release Lever
Releases the HS-8000 series from the
HSA-81.
2
Rear Side
1 DS I/F Connector
Connects the HSA-81 to the DSC-8410.
NOTE
When the DS-8007 is connected to this device via DSA-82, the DS-8007 fixed keys (MENU,
PRINT START/STOP, NIBP START/STOP) and alarm silence key are enabled on the host
monitor.
When this device is connected, the standby switch, HOME key, alarm indicator and alarm
sound suspend key on the DS-8007 will not function.
For the operation procedure when DS-8007 is used without connection to other system,
refer to the DS-8007 system Operation Manual.
Front Side
1 Standby Switch
Sets ON/OFF the standby condition.
4
2 Power Supply LED 1 5
Indicates the power supply status. Light will be off 2
when the AC power is not supplied to the monitor. 3
Orange: Standby Mode
Green: In normal operation
Light Off: During battery operation (AC power cable is
not connected.)
2-6
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
NOTE
If the battery charging LED flashes, battery charging error is occurring. Remove the
battery and install it again. If the error persists, contact your nearest service
representative.
4 Fixed Key
HOME Key
MENU Key
PRINT START/STOP Key
NIBP START/STOP Key
Rear Side
1 CO2 I/F Connector
Attaches the Gas Unit I/F (HPD-820) or CO2 Gas Unit
(HCP-820).
1
Right Side
1 USB Memory Slot
Insert a USB memory.
2 Battery Cover
Stores the specified lithium-ion battery.
2
3 DS I/F Connector
Connects to the DSA-81 AC Unit, DSA-82 DS-8007 Adapter, 1
or DSC-8410.
2-7
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Left Side
1 Analog Output Connector
Outputs the ECG, BP, synchronized signal.
2 ECG Connector
Connects the specified ECG relay cable. 1
5
3 NIBP Connector 6
Connects the NIBP air hose. 2
4 AUX Connector
7
Connects the Gas Unit I/F (HPD-810), CO2 Gas Unit (HCP-
810), BISx I/F Unit (HBX-800).
3
5 Temperature Connector
Connects the TEMP sensor cable. 4
6 Multiparameter Connector
Connects the input cables for BP, TEMP or CO.
7 SpO2 Connector
Connects the SpO2 sensor, or relay cable (patient cable).
NOTE
When operating as an input module by connecting to the rear side of the host monitor (DS-
8400 system), standby switch, fixed key, alarm indicator will not function..
Bottom Side
The illustration shown on right is when the cover is removed.
1 SD Card Indicator
Lights when the SD card is accessed.
2 SD Card Slot
Card slot for the specified SD card 1
2
1 DS I/F Connector
Connects the DSA-82 to the DS-8007. 1
2 DS-8007/HS-8000 Switch
This is a switch to change between DS-8007 and HS-
8000.The selected unit and the host monitor will
communicate.The LED for the selected unit will light in 2
green.Switching to the left will communicate with the DS-
8007.Switching to the right will communicate with the
HS-8000. 3
3 Lock Lever
Lever to release the DS-8007 series from the DSA-82
2-8
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
5 HS-8000 Connector
Connects the HS-8000 via HSA-80.
4 5
NOTE
When the DS-8007 is connected via DSA-82, the NIBP auto mode setting of the DS-8007
will be displayed on the DS-8007, but the actual measurement will be performed according
to the NIBP auto mode setting of the DS-8400.
Front Side
1 Power Supply LED
Indicates the power ON/OFF status. 1
2
2 BP Zero Balance LED 3
Lights during BP zero balancing.
4 Multiparameter Connector
Connects the input cables for BP, TEMP or CO.
Rear Side
1 Power Input Connector
Supplies power while connecting to the input box .
1
2 Infrared Communication Port
Communicates with the input box via IrDA.
2-9
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side
1 Power Supply LED
1
Indicates the power ON/OFF status. 2
3
2 BP Zero Balance Indicator
Lights during BP zero balancing.
4 Multiparameter Connector
5
Connects the input cables for BP, TEMP or CO.
5 AUX Connector 6
Connects the Gas Unit I/F (HPD-810), CO2 Gas Unit (HCP-
810), BISx I/F Unit (HBX-800). 7
7 Release Lever
Press here to remove the expansion modules from the input box.
Rear Side
1 Power Input Connector
Supplies power while connecting to the input box.
1
2 Infrared Communication Port
Communicates with the input box via IrDA.
2-10
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
HG-810 (Masimo model) and HG-820 (Nellcor ) can be used for the SpO2 modules. The following shows the
example of HG-810.
Front Side
1
1 Power Supply LED
Indicates the power ON/OFF status.
2 SpO2 Connector
Connects the SpO2 sensor, or relay cable (patient cable).
2
3 Release Lock Button
Press to lock the release lever.
4 Release Lever
Press here to remove the expansion modules from the input
box.
3
4
Rear Side
1 Power Input Connector
Supplies power while connecting to the input box . 1
2-11
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side
1 Power Supply LED 1
Rear Side
1 Power Input Connector
Supplies power while connecting to the input box .
1
2 Infrared Communication Port
Communicates with the input box via IrDA.
2-12
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side
1 2 3 4 5 6
1 Power Supply Indicator
Indicates the power status.
2 Printing Indicator
Lights during printing.
3 Print Key
Starts/stops the printing.
6 Open/Close Lever
Press to open the paper holder.
Rear Side
1 U-LINK Connector
Connects to the MGU-800/810 series Multigas Unit or Main
Unit.
2-13
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side
1 LAN-ID Setting Indicator 1
Indicates the assigned LAN-ID. 2
3
2 Power Output Connector
Supplies power to the expansion module.
Rear Side
1 LAN ID Setting Dial
Sets the LAN ID to 1.
2 module-LAN Connector x3
Connects the Main Unit, IB-8004, or HSA-80 1
HS Adapter.
2
REFERENCE
For the connection procedure, refer to the operation manual of the IB-8004.
2-14
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side 3
1 Power Supply LED
Indicates the power ON/OFF status. It will light in green 1
while the power is ON.
Rear Side
1 AUX Connector
Connects to the AUX connector of HS-8000/DS-8007/HM-
801 with AUX connection cable.
2 Exhaust Hole 2
NOTE
The usable AUX connection cable differs depending on the connecting device. For the
combination of the AUX connection cable and the connecting device, refer to the section on
"Optional Accessories".
CAUTION
Do not block the exhaust hole as it may cause damage to the device.
2-15
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side
1 Sampling Tube Connector
Connects the sampling tube manufactured by Covidien.
Rear Side
1 Gas Unit Connector
Connects to the CO2 I/F connector of the DS-8007. 1
2 Exhaust Hole
Connects the gas exhaust system and exhausts
sampling gas.
2
CAUTION
Do not block the exhaust hole as it may cause damage to the equipment.
2-16
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side 3
1 Power Supply LED
Indicates the power ON/OFF status. It will light in green 1
while the power is ON.
2 CO2 Connector
Connects to the Capnostat 5 (Philips).
3 Clip 2
Attaches to the bedside rail or headboard for bedside use.
Rear Side
1 AUX Connector
Connects to the AUX connector of HS-8000/DS-8007/HM-
801 with AUX connection cable.
NOTE
The usable AUX connection cable differs depending on the connecting device. For the
combination of the AUX connection cable and the connecting device, refer to the section on
"Optional Accessories".
2-17
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side
1 CO2 Connector
Connects to the Capnostat 5 (Philips).
Rear Side
1 Gas Unit Connector
Connects to the CO2 I/F connector on the DS-8007. 1
2-18
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
Front Side
1 BISx Module Connector
Connects the BISx (Covidien).
1
2 Power Supply LED
Indicates the power ON/OFF status.
Lights in orange when the power is ON, and BISx is not
connected.Lights in green when the BISx is connected.
2
Rear Side
1 AUX Connector
Connects to the AUX connector of HS-8000/DS-8007
with AUX connection cable.
NOTE
The usable AUX connection cable differs depending on the connecting device. For the
combination of the AUX connection cable and the connecting device, refer to the section on
"Optional Accessories".
2-19
Chapter 2 Names of Parts and Their Functions Names of Parts and Their Functions
2 Inhale Port 2
5
Connects the sampling tube to inhale sampling 3
gas. 4
3 Water Trap with Reservoir
Removes the water from the sampling tube
connected to the patient. When the reservoir is
more than half full with water, empty the water.
(
Maintenance Manual "Water Trap
(Multigas Unit)" P8-5)
4 Exhaust Hole
Connects gas exhaust system and exhausts sampling gas.
WARNING
To protect the hospital staffs from unnecessary anesthetic agent, it is strongly
recommended to connect the exhaust hole to the gas exhaust system in the hospital.
CAUTION
Do not block the exhaust hole as it may cause damage to the device.
Rear Side
1 External Device Connector 1
Connects the main unit.
2-20
Chapter 3 Operation Procedure and Screen Examples Operation Procedure
All operation of this device is performed using the fixed keys, touch keys, and mouse (optional). The device can also
be operated remotely using the optional remote control.
Fixed Keys
4 Standby Switch
Enters into standby mode or cancels the standby mode.
Touch Key
3
1
1 Pressing the [Menu] key will switch the display with a pip sound.
2 The touch key will respond by pressing any part of the key.
3 Pressing the [Home] key at any time will return the display to the home display.
3-1
Chapter 3 Operation Procedure and Screen Examples Operation Procedure
REFERENCE
The above is an example of the screen. The user keys can be customized and can be placed
to any position.
(
"To Configure the Display" P10-4)
2 Pressing the [Home] key at any time will return the display to the home display.
REFERENCE
Frequently used touch keys can be programmed as user key. The user keys can be
positioned to the user key display area and also to the numeric data area.
(
"For Easier Use" P3-27)
Mouse
An optional mouse can be connected allowing touch key control using the mouse.
By moving the pointer on the displayed keys, and left-clicking the mouse, the operation can be performed just the
same as by directly touching the displayed keys.
The pointer will be hidden if the mouse is not used for 5 minutes.(default operation)
The hidden mouse pointer will be displayed again by moving the mouse.
NOTE
It is necessary to set the mouse function (ON/OFF, pointer shape, moving speed) in
advance. (
Maintenance Manual "Operation Related Setup" P5-26)
Some mouse operation such as dragging operation is not possible on this equipment.
3-2
Chapter 3 Operation Procedure and Screen Examples Home Display
Home Display
Bottom 2 rows to 6 rows The waveform display area or numeric data can be
enlarged.
By connecting maximum of 2 extended display units, maximum of 2 displays independent from the main display can
be used. (extended display function)
Also, the user key location can be selected from left, right, or bottom. By selecting the layout of 2 columns on left/
right, more user keys can be displayed.
REFERENCE
The display layout can be configured/registered according to the monitoring waveform and
numeric data as necessary.
(
"To Configure the Display" P10-4)
Display Example
On this system, 9 main modes and 6 sub modes can be preprogrammed according to the monitoring purpose.
By registering the configuration to each mode, the display configuration setups at admittance of patient can be
simplified by just selecting one of the modes.
3-3
Chapter 3 Operation Procedure and Screen Examples Home Display
( "To Select the User Mode" P5-8)
It is recommended to program the display mode in rough classification such as patient’s condition, monitoring
purpose (ICU or surgery), and if necessary, perform unique setup for each patient.
Oxygenator Mode
WARNING
The oxygenator mode is intended to prevent alarms during cardiopulmonary bypass
surgery. Pay special attention when using this mode as the alarm generation will not be the
same as to the standard monitoring mode.
If the "Alarm Setting" under the Oxygenator Mode Setup is set to [All OFF], all vital alarm will
not generate regardless of the alarm setting of each parameter. Also, if [Sel. Parameter] is
set, vital alarm for unselected parameter will not generate. Pay attention to not miss any
significant change of the patient's vital sign as the alarms will not be generated during the
Oxygenator Mode.
Once the cardiopulmonary bypass is finished, make sure to cancel the Oxygenator Mode
and return to the standard monitoring mode.
CAUTION
If the NIBP alarm is turned OFF under the Oxygenator Mode, NIBP auto mode measurement
and NIBP measurement at alarm occurrence will not be performed.
Oxygenator mode can be used to prevent frequent alarm generation when oxygenator is used for extracorporeal
circulation during cardiac surgery.
During oxygenator mode, "Oxygenator Mode" will be displayed on the screen, alarm generation will be stopped, and
low priority parameter will be displayed with decreased brightness.
The main difference of standard monitoring mode and oxygenator mode is as follows.
Standard Monitoring Mode Oxygenator Mode
Vital Alarm will be generated. will not be generated,
or only the alarm for specified parameter will be
generated.*
Device Status Alarm will be generated. will be generated for specified parameter.
NIBP Periodic will be performed. If [NIBP] is not selected on Oxygenator Mode Setup,
Measurement periodic measurement will not be performed.
It will not be performed even if NIBP measurement is
requested from the central monitor.
Night Mode Night mode can be used. Night mode cannot be used.
*It is also possible to set the same alarm function with the standard monitoring mode.
REFERENCE
The oxygenator mode setup can be performed under [Menu>Initial
Settings>Alarm>Oxygenator Mode Setup].
(
Maintenance Manual "Alarm Related Setup" P5-4)
3-4
Chapter 3 Operation Procedure and Screen Examples Home Display
2 Press the [OK] key to change the monitoring mode to oxygenator mode.
3 Press the [Standard Mode] inside the message window to return to the standard monitoring mode.
REFERENCE
The message window can be dragged to any position within the waveform area.
The message window will not be displayed by selecting [OFF] for "Oxygenator Mode
Message" (Menu>Initial Settings>Oxygenator Mode Setup).
Displayed Items
Other than waveforms and numeric data, patient name, alarm message, status message, etc. will be displayed on the
screen.
3-5
Chapter 3 Operation Procedure and Screen Examples Home Display
2 3 4 5 6 8 9 12 10 11 13 7
1
2 Room/Bed ID
Displays the 3-digit Room ID and 3-digit (000–999) Bed ID.
4 Patient Name
The patient name set on the "Admit/Discharge" menu will be displayed.
The patient name can be hidden from the display area by selecting [OFF] for "Patient Name on the Information
Display Area" (Initial Settings>User I/F>Display/Print).
5 Pacemaker Usage
When [Used] is set for "Pacemaker" on the "Admit/Discharge" menu, <Pacemaker> will be displayed.
6 Patient Classification
The patient classification (Adult, Child, Neonate) set on the "Admit/Discharge" menu will be displayed.
7 Date/Time
Displays the current date (month, day) and time (hour, minute).
8 Set Mode
The currently selected user mode will be displayed. Sub mode
will be also displayed if selected.
When using the data transfer function with the Super Unit,
alarm settings and parameter settings can be also transferred.
When the settings are changed by the data transfer function,
the mode name will be highlighted to notify that the setting has
been changed. Pressing the highlighted mode name will
display the confirmation message window (shown on right).
Pressing the [OK] key will clear the highlight.
When the alarm settings are changed, the alarm settings list will be displayed.
10 Drift Filter
When drift filter is set to ON, "Drift-F ON" will be displayed.
11 Message Area
When an alarm generates, a message will be displayed.
By pressing the message display area, the alarm message history can be verified.
3-6
Chapter 3 Operation Procedure and Screen Examples Home Display
13 Power Supply/Battery
The power supply status and battery charging status (when battery is installed) will be displayed.
Waveform Area
1 ECG
2 ECG Lead
3 ECG Size 1
The waveform size display of ECG, RESP, SpO2 can be 2
3
selected from [Numeric]/[Bar]/[Bar (10 mm)].
[Initial Settings > User I/F > Display/Print > Waveform Size
4
Display] (
13)
Maintenance Manual "Display/Print Setup" P5-
5
6
4 SpO2 Waveform 7
8
5 SpO2 Size
6 BP Label
9
7 BP Scale 10
8 BP Waveform
9 Respiration Waveform
11
10 RESP Size
12 CO2 Scale
12
13 CO2 Waveform 13
14 EEG Waveform 14
15
15 EEG Scale
Enlarged Waveform
By selecting [ON] for "Zoom" under
[Menu>Basic Setup>Display Config.], the
waveform display can be enlarged. The
waveform display can be enlarged for ECG,
pulse wave, and respiration waveforms.
<200%> will be displayed for the enlarged
waveform. Also, the sweep speed will be
doubled.
3-7
Chapter 3 Operation Procedure and Screen Examples Home Display
HR, HR/PR 1
1 HR/PR Synchronization Mark
When HR or PR according to the setting of "Synchronized Mark/
Tone" is detected, HR/PR synchronized mark will be displayed 3 2
inside the corresponding numeric data box.
2 HR/PR Value
The HR/PR value will be displayed. When the value exceeds the
measurable range, "xxx" will be displayed.
3-8
Chapter 3 Operation Procedure and Screen Examples Home Display
PR, HR/PR 3
1
1 Pulse Rate (BP)
SpO2 3
2
1 SpO2 Value 4
The arterial oxygen saturation will be displayed.
2 SpO2 Label 1 5 6
The label set for SpO2 will be displayed.
4 Pulse Rate
The pulse rate is displayed. When the value exceeds the measurable range, "xxx" will be displayed.
VPC
1 VPC (1 min)
1
The VPC rate for the last 1 minute will be displayed. <---> will be 2
displayed during arrhythmia learning.
ST
ST Level
The ST value for 4 leads can be displayed in the ST data box.
3 groups (A, B, C) of lead combination can be programmed.
For the following case, "---" will be displayed.
Learn
During lead-off condition
3-9
Chapter 3 Operation Procedure and Screen Examples Home Display
REFERENCE
The leads displayed inside the ST level box can be changed.
(
Maintenance Manual "Display/Print Setup" P5-13)
2 RR Synchronized Mark
When the respiration of the set RR source is detected, a synchronized mark will be displayed inside the
corresponding numeric data box.
3 Respiration Rate
The impedance RR, CO2 RR, and ventilator RR will be displayed. When the value exceeds the measurable
range, "xxx" will be displayed.
When the impedance measurement is set to OFF, impedance RR will not be displayed.
NOTE
To display the respiration rate for each respiration source (impedance/CO2/ventilator/
SpO2), set the corresponding numeric data box on the display configuration setup menu.
NIBP
3 4 5
1 NIBP Value/Cuff Pressure
The NIBP measurement value (SYS / DIA / MAP) will be displayed.
The mean NIBP display can be set to ON or OFF on the NIBP setup 1
menu. The value will be displayed as "---" when the preprogrammed
NIBP erase time has elapsed.
2
During measurement, a cuff pressure will be displayed.
5 Oscillation Graph
The horizontal axis in the graph shows the cuff pressure, and vertical axis shows the pulse amplitude with
reference to maximum pulse amplitude.
6 NIBP List
The NIBP list of the latest 3/6/9/12/18 data and measured date/ 6
time will be displayed. The number of displaying data depends on
the size of numeric data box.
3-10
Chapter 3 Operation Procedure and Screen Examples Home Display
Blood Pressure
1 BP Label
The label set for the blood pressure will be displayed.
2 <MEAN_WAVE>
<MEAN_WAVE> is displayed when [ON] is set for "Mean Wave" under [Menu>Parameter>BP>Detail Setup].
3 Blood Pressure
The BP measurement value (SYS/DIA/MEAN) will be displayed. On the BP setup menu, the display type (S/
D/M, S/D, M) can be selected. When the value exceeds the measurable range, "xxx" will be displayed. If BP
zero balance is not performed, "---" will be displayed, and if transducer is not connected, nothing will be
displayed.
2 PDP Value
When the BP label is IAP, PDP (Peak Diastolic Pressure) of IABP can
be measured. Note that Systolic Pressure (SYS) = Peak Systolic
Pressure (PSP).
3 CPP Value
When the BP label is ICP, labeling the artery pressure as ART will allow measuring the CPP (Cerebral
Perfusion Pressure). CPP = Mean Arterial Pressure – Mean Intracranial Pressure If the CPP value is negative
value, or zero balance has not been performed for ICP or ART, "---" will be displayed, and if ICP or ART has
not been measured, nothing will be displayed. Also, alarm cannot be set for CPP.
Temperature
1 TEMP Label
The label set for the temperature will be displayed.
2 TEMP Value
The temperature will be displayed. 400 series temperature sensor can be used. When the value exceeds the
measurable range, "xxx" will be displayed. When 700 is used, "---" will be displayed.
Blood Temperature
By using the thermodilution catheter for the CO measurement, blood
temperature can be displayed. When the value exceeds the measurable
range, "xxx" will be displayed.
EtCO2/ InspCO2
InspCO2 Value/EtCO2 Value
The end-tidal CO2 concentration and inspiratory CO2 concentration
measurement value will be displayed.
The measurement unit can be selected from mmHg / kPa / % under the
"Initial Settings" menu.
Ventilator
Ventilator Data
When a ventilator is connected, ventilator measurement data will be
displayed.
3-11
Chapter 3 Operation Procedure and Screen Examples Home Display
P-V, F-V
P-V, F-V Loop
By connecting the ventilator, multigas unit (MGU-810 with SPIRO unit), or
anesthesia delivery system (FLOW-i, GE, Draeger), P-V loop (airway
pressure / ventilation) and F-V loop (airway flow/ventilation) can be
monitored on the ventilator display.
SvO2/CCO Data
When SvO2/CCO Measurement Device (Vigilance/Vigilance CEDV/VigilanceII/Vigileo/HemoSphere/ PiCCO2/
EV-1000/PulsioFlex) is connected, the measurement data (SvO2, CO, etc.) acquired from these devices will
be displayed. The displayed data will differ depending on the used SvO2/CCO Measurement Device and
measurement mode.
NOTE
When measuring StO2 by connecting HemoSphere, the measurement data will be displayed
as rSO2. rSO2 will be displayed on the INVOS numeric data box.
Hemodynamic Data
Hemodynamic Data (Vigilance)
Based on the CCO data measured by the Vigilance (or Vigilance CEDV/
Vigilance/Vigileo), the following hemodynamic data are calculated and
displayed every second based on the following condition. However the
following condition should be met.
It is measured on Vigilance with CCO mode. (It will not be displayed
during ICO mode.)
SvO2 parameter key (oximeter numeric data box) is displayed.
BP label is set as ART, PAP, CVP.
(If the unit is "kPa", the data is converted to "mmHg" for calculation.)
Data Description Formula
3-12
Chapter 3 Operation Procedure and Screen Examples Home Display
NOTE
The hemodynamic data based on Vigilance data will not be stored in the list. For the
Vigilance list, the data directly acquired from the Vigilance will be stored.
TIMER
Stopwatch Key
Functions as stopwatch.
BIS
BIS Value
By connecting the BISx module through the HBX-800, or by connecting the
BIS monitor, BIS data (BIS, SQI, EMG, SR) will be displayed.
If SQI value is below 50%, the BIS value will be displayed in gray.
If SQI value is below 15%, the BIS value and SR value will disappear.
EMG and SQI will be displayed in bar graph.
2 21 to 40 39 to 47
3 41 to 60 48 to 55
4 61 to 80 55 and above
5 81 to 100 -
The alarm bar will be displayed only when measurement is performed on BISx using the HBX-800.
AEP
AEP Monitor Data (AEP Index)
When connected to AEP Monitor, aepEX PLUS, the AEP index will be
displayed.
A circle graph is displayed until the measurement data is obtained.
INVOS
INVOS 5100C Measurement Data
When connected to INVOS 5100C, regional cerebral oxygen saturation value
will be displayed.
Lt- indicates left brain, and Rt- indicates right brain.
3-13
Chapter 3 Operation Procedure and Screen Examples Home Display
NOTE
Regional cerebral oxygen saturation value will also be displayed when HemoSphere is
connected.
1
2
1
3
2
SYS
DIA
MEAN
The alarm limit can be displayed beside each numeric data.The display type can be selected from [Graph]/
[Numeric]/[OFF] for "Alarm Limit Display" under [Menu>Alarm>Detail Setup].
If ON is selected for the individual alarm, the alarm limit will be displayed.
The upper and lower limit will be displayed at upper and lower row respectively.
For BP and NIBP, each alarm limit of systolic BP (SYS), diastolic BP (DIA), mean BP (MEAN/MAP) will be
displayed from the top.
ON/OFF of alarm limit display can be selected. (
"List of Alarm Settings" P6-5)
1 Upper Alarm Limit
3-14
Chapter 3 Operation Procedure and Screen Examples Home Display
NOTE
If the alarm limit display for BP is [Graph], systolic value will be displayed.
Depending on the numeric data box type, alarm limit may not be displayed.
If the SpO2 lower alarm limit is set to 85%SpO2 or below, the alarm limit value will be
displayed regardless of the "Alarm Limit Display" setting.
2 Trend Scale
The short trend scale will be displayed between the short trend and numeric data. The displayed scale will be
in accordance with the scale set on the "Trend" screen.
For the following parameters, the short trend scale can be synchronized with the corresponding waveform
scale by selecting [Waveform] for "Short Trend Scale" under [Menu>Display Config.>Detail Setup].
BP, PEAK, VT, CO2, O2, Agent
For operation procedure on the short trend display, refer to
"Short Trend" P8-10.
Displayed number of waveform and numeric data
Layout Maximum Displayed Display Duration (25 mm/s) Maximum Displayed
Waveforms Boxes
3-15
Chapter 3 Operation Procedure and Screen Examples Home Display
Alarm OFF
Indicates the alarm is OFF.
Alarm Silence
Indicates the alarm is silenced when the alarm system is IEC mode.
HR Synchronized Mark
This mark flashes synchronizing to the heartbeat.
RR Synchronized Mark
This mark flashes synchronizing to the inspiration.
Message Icon
Indicates that an alarm message is present for that parameter. Whether or not to display this icon can
be selected under "Initial Settings".
Key Lock Mark
Indicates that the item requires a password to change the setting.
This icon (2/3 green) indicates that the battery is less than full, but still usable.
This icon (1/3 yellow) indicates that the battery level is low and needs to be charged.
This icon (1/3 red) indicates that the battery level is very low and flashes to alert the low battery status.
Immediate battery charge is required.
Technical alarm will generate.
This icon (red frame) indicates that the battery is almost depleted and it flashes to alert that charging is
necessary. Make sure to charge the battery immediately.
This icon (black frame with a slash) indicates that the battery is not installed. Pay attention as power
will not be supplied if AC power cable is disconnected during this state.
3-16
Chapter 3 Operation Procedure and Screen Examples Home Display
This section explains about the message displayed on the home display.
There are vital alarm message and device status alarm message which will be displayed at the top of the home
display.
The alarms are classified to Level S (top priority), Level H (high priority, urgent), Level M (medium priority,
caution), Level L (low priority, status), and Notification, and the message will be displayed according to the priority
of Level S > Level H > Level M > Level L > Notification.
The displayed messages will flash in red and white for Level S, red for Level H, yellow for Level M, blue for Level
L, and white for Notification.
Alarm Priority, Level Description Tone/Volume Displayed Color
CAUTION
When more than one alarms of the same priority are generated, the newer alarm message
will be displayed.
There are 2 types of vital alarm messages; numeric data alarm and arrhythmia alarm. If both alarms occur at the same
time, the numeric alarm message and arrhythmia alarm message will be displayed alternately in 2-seconds intervals.
The alarm messages will be displayed according to the priority. If the priority level is the same, the newer alarm
message will be prioritized.
CAUTION
The arrhythmia alarm message other than Tachy, Brady, Ext Tachy, Ext Brady will continue
to be displayed for 30 seconds after the alarm is resolved.
3-17
Chapter 3 Operation Procedure and Screen Examples Home Display
Lead-Off Message
If the ECG electrodes used for HR measurement orarrhythmia analysis are detached, the status will be notified.
WARNING
When <Lead Off> is displayed, HR alarm or arrhythmia alarm will not generate. If this
condition is left unresolved, a sudden change of the patient may not be noticed. Take prompt
action when the lead-off condition is detected.
WARNING
The ventilator alarm sound is set to OFF (factory default).
The alarm sound can be turned ON under [Menu>Tone/Volume].
(
"Tone/Volume" P10-19)
3-18
Chapter 3 Operation Procedure and Screen Examples Home Display
CAUTION
For the SV-900, VELIA, ASTRAL, VS ULTRA, ventilator alarm factor will not be notified to
the central monitor.
Depending on the central monitor type and software version, ventilator alarm factor may not
be displayed. For details of the central monitor type and software version, refer to your
nearest service representative.
The ventilator alarm factors are displayed only on the central monitor. These will not be
displayed on the bedside monitor.
CAUTION
Check occasionally the communication status of this device and the ventilator.
Verify that a ventilator alarm is not generated, and that the <Vent. Comm.> message is
displayed.
This confirmation window will be displayed until the displayed key is pressed or proper
communication with the ventilator is resumed. When the communication is resumed, the
window will automatically close.
When disconnecting the ventilator and this device, make sure to select [OFF] on the
confirmation window which will be displayed when the power of the ventilator is turned OFF,
or when the cable is disconnected.
3-19
Chapter 3 Operation Procedure and Screen Examples Window Display
Window Display
The screens that are displayed when operating this system are referred to as windows. (The windows that appear by
pressing the numeric data area are called floating windows, as they can be moved to any desired position.)
The target window can be displayed by using various method, such as selecting the menu items, pressing a parameter
key or using a short cut key such as user key.
Display
4 Minimize Key
Pressing this key will minimize the currently displayed window and will be stored to the user key.
To restore the minimized window, press the Restore key in the user key and select the window to restore.
NOTE
The color of each key lock icon indicates its administrative level, and a higher level
password must be entered to unlock it.
3-20
Chapter 3 Operation Procedure and Screen Examples Window Display
6 Setup Item
Most of the setups can be performed by selecting from the dropdown list.
The dropdown list will close when a selection has been made.
Pressing the item again or selecting a different item will also close the dropdown
list.
Some menu may display a subwindow to perform the setup.
To close the subwindow, press either the key, [Home] or [Prev. Disp.] key.
Pressing the key with the " " icon will display another window.To return to
the original display, press the key.
Example of an item which displays another screen
7 Dropdown List
Select one from the displayed selection list.
3-21
Chapter 3 Operation Procedure and Screen Examples Window Display
The descriptions of the floating window which is displayed by pressing the numeric data area are as follows.
The displayed items on the floating window depends on the parameter, but there are some common items as follows.
1 Window Title 1
The windows can be moved to any desired position by dragging the 2 3
window title.
4
2 Alarm Assist Key
The alarm assist screen will be displayed. On the alarm assist
screen, maximum of 24 hours of trend data for the corresponding
parameter will be displayed, and alarm threshold can be adjusted by
checking the trend data.
( "Alarm Assist Screen" P6-13)
3 Close Key
Press the key to close the window.
4 Detail Key
On the floating window, minimum items are displayed. Press the key to display more detailed items.
3-22
Chapter 3 Operation Procedure and Screen Examples Window Display
Minimize Window
To temporarily display the home display during the setup, press the (Minimize) key. The current window will
be minimized. By pressing the [Restore Window] key, the window will be redisplayed.
NOTE
To use the restore window function, the [Restore Window] key needs to be preprogrammed
as user key.
1 Press .
NOTE
Maximum of 8 windows can be minimized. If exceeded, the oldest window will be deleted.
To delete all minimized window, press the [Delete All] key which will be displayed when
is pressed for more than 1 second.
The window which has been automatically erased after fixed amount of time can be
remained minimized by selecting [ON] for "Auto Minimize" under [Initial Settings>User I/F
>Operation].
3-23
Chapter 3 Operation Procedure and Screen Examples Window Display
Transfer Window
The floating window displayed by pressing the numeric data area can be moved by dragging the window title. This
operation is possible on the touch panel.
2 Place the finger on the window title and drag to the desired position.
NOTE
The floating window cannot be overlapped to the numeric data area or information display
area.
The window which is displayed from "Menu" cannot be moved.
The displayed position of the floating window will be stored until the power is turned OFF.
3-24
Chapter 3 Operation Procedure and Screen Examples Procedure to Return the Display
Operation Restriction
3-25
Chapter 3 Operation Procedure and Screen Examples To Enter Characters
To Enter Characters
Alphanumeric characters and symbols can be entered using the displayed keyboard.
The procedure to enter characters is explained below using the example of patient admit menu.
Entering Numerics
For age, telemetry channel ID, etc., only numbers can be
entered.
In such case, only numeric keys will be displayed.
Enter the numbers.
3-26
Chapter 3 Operation Procedure and Screen Examples For Easier Use
The user keys and menu can be customized according to the monitoring purpose.
REFERENCE
From the preprogrammed user mode, the display configuration and alarm settings can be
selected according to the monitoring purpose.
(
Maintenance Manual "User Mode Registration" P5-30)
User Key
By assigning the [ ] to the user key area, 2 pages of user keys can be registered, and pressing the [ ] allows to
switch the pages.The user key can be enlarged by using 2 display areas.
The user key can be also assigned to the numeric data area. It is useful if the key related to numeric data is assigned
near the numeric data.
Menu Screen
The key position can be changed and unnecessary keys can be deleted on the "Menu" screen.
( Maintenance Manual "Display/Print Setup" P5-13)
3-27
Chapter 3 Operation Procedure and Screen Examples Display on the External Monitor and Extended Display Unit
For the DS-8400 system, in addition to the main display, another display unit can be used for extended display.
REFERENCE
Use only the specified 19-inch display unit. For details, refer to our service representative..
Displayable Screen
Model
External Monitor Display Extended Display 1 Extended Display 2
DSC-8410 Yes No No
3-28
Chapter 3 Operation Procedure and Screen Examples Display on the External Monitor and Extended Display Unit
The same monitoring display can be displayed on other display unit. However, only the messages related to
arrhythmia alarm will be displayed.
Other messages, menu, and setup screens are not displayed. Also, operation is not possible on the external monitor.
Extended Display 1
On the extended display 1, the independent display from the main display can be displayed.
On the extended display 1, the following operations are possible on the touch panel.
*Selection of Preprogrammed Display Layout (3 types) *NIBP Start/Stop
3-29
Chapter 3 Operation Procedure and Screen Examples Display on the External Monitor and Extended Display Unit
*Alarm Suspend
CAUTION
As a speaker is not equipped for the extended display unit, key sound and alarm sound will
not be generated. Alarm sound will be generated from the main display.
The same setup window cannot be opened simultaneously on the main display and the
extended display. If the same setup window is opened, the previously opened window will
close.
REFERENCE
The waveform size for the main display and the extended display is independent.
On the "Initial Settings" menu, whether or not to synchronize the waveform size/scale of
extended display with the main unit can be selected.
(
Maintenance Manual "Display/Print Setup" P5-13)
The waveform size for the extended display 1 and 2 is common.
By setting the [Scale (Extended Display)] key as user key, the waveform scale on the
extended display can be changed on the main unit.
(
"User Key Selection" P10-15)
The display configuration for the extended display 1 can be also changed on the main
display.
(Menu > Initial Settings > User Mode Regist. > Select mode for "Extended Display 1" >
Change the setting)
The key mask function, key lock function, auto hide function cannot be used.
Extended Display 2
On the extended display 2, the independent display from the main display can be displayed.
On the extended display 2, the following operations are possible on the touch panel.
Selection of Preprogrammed Display Layout (3 types)
ON/OFF of Short Trend Display
3-30
Chapter 3 Operation Procedure and Screen Examples Display on the External Monitor and Extended Display Unit
CAUTION
As a speaker is not equipped for the extended display unit, key sound and alarm sound will
not be generated. Alarm sound will be generated from the main display.
REFERENCE
On the extended display unit 2, operation such as alarm setup is not possible.
The waveform size for the main display and the extended display is independent.
On the "Initial Settings" menu, whether or not to synchronize the waveform size/scale of
extended display with the main unit can be selected.
(
Maintenance Manual "Display/Print Setup" P5-13)
The waveform size for the extended display 1 and 2 is common.
By setting the [Scale (Extended Display)] key as user key, the waveform scale on the
extended display can be changed on the main unit.
(
"User Key Selection" P10-15)
The display configuration for the extended display 2 can be changed on the main display.
(Menu > Initial Settings > User Mode Regist. > Select mode for "Extended Display 2" >
Change the setting)
3-31
Chapter 3 Operation Procedure and Screen Examples Display on the External Monitor and Extended Display Unit
3-32
Chapter 4 Preparation Daily Check
Chapter 4 Preparation
Daily Check
To Start Monitoring
This section explains about the procedure to turn the power ON and start monitoring.
CAUTION
If not using the device for a long period, disconnect the power cable and lithium-ion battery.
During transportation, firmly grasp the handle and make sure that the device does not fall.
Otherwise, it may cause injury to the operator or damage to the device.
WARNING
Do not connect a battery other than the lithium-ion battery (BTO-005).
4-1
Chapter 4 Preparation To Start Monitoring
NOTE
The operation after the power is turned ON will be according to the setting made on [Initial
Settings] > [User I/F] > [Power ON/Discharge]. However, if the power was turned OFF for
less than 30 seconds, the setting before the power was turned OFF will remain.
If the battery charging LED flashes, battery charging error is occurring. Remove the battery
and install it again. If the error persists, contact your nearest service representative.
REFERENCE
The power ON/OFF operation of the main unit, Super Unit, expansion unit, expansion
module, and input box synchronizes with the standby switch operation (ON/OFF) on the
display unit.
Check Discharge
1 Select from [Discharge] / [Continue].
NOTE
If the standby switch was turned OFF for less than 30 seconds, the discharge confirmation
screen will not be displayed. To perform the discharge procedure, press the [Discharge] key
on the "Admit/Discharge" screen.
(
"Discharge" P5-7)
To start monitoring a new patient, select [Discharge] and enter the new patient information
on the "Admit/Discharge" screen.
REFERENCE
Whether or not to display the discharge confirmation window can be selected.
(
Maintenance Manual "Power ON/Discharge" P5-18)
4-2
Chapter 4 Preparation To Start Monitoring
REFERENCE
The parts which the replacement period will be notified are the NIBP unit in the Super Unit
and the CO2 unit in the HCP-810/HCP-820.
(
Maintenance Manual "Periodic Replacement" P7-1)
Even if it is set not to display the discharge confirmation screen, the confirmation message
for parts replacement will be displayed when the replacement period approaches.
The patient data and settings are stored in the HS-8000 Series Super Unit.
When transferring the patient to another bed, the same monitoring condition such as patient data and settings can be
used on new bed by transferring the HS-8000 along with the patient.
CAUTION
This function can be used when the HS-8000 series Super Unit is connected. When the DS-
8007 is connected, refer to
"Data Transfer Function Using the Transport Monitor (DS-
8007)" P4-5, as the operation will differ.
2 Connect the HS-8000 to the DS-8400 system of the new bed, and turn ON the power of the DS-8400 system.
4-3
Chapter 4 Preparation To Start Monitoring
CAUTION
After the data transfer process, make sure that the setting and patient data are correct.
NOTE
During the data update process, the patient name on the home display will flash.
When [Continue monitoring] is selected, the stored data on the main unit will be overwritten
with that of the HS-8000.
If a central monitor is connected, the data on the central monitor will be also deleted.
The alarm settings and parameter settings can be also transferred. When the settings are
changed by the data transfer function, the mode name will be highlighted to notify that the
setting has been changed. Pressing the highlighted mode name will display the confirmation
message window, and pressing the [OK] key will clear the highlight. When the alarm settings
are changed, the alarm settings list will be displayed.
When [Discharge] is selected, both data on the main unit and the HS-8000 will be deleted/
initialized.
When [Cancel] is selected, the stored data on the HS-8000 will be overwritten with that of
the main unit.
The data on the HS-8000 will be updated if any of the [Continue monitoring]/[Discharge]/
[Cancel] is selected. Do not disconnect the HS-8000 during the update process. If
disconnected, the data consistency may be lost.
The BP zero balance value on the HS-8000 will not be cleared. After transferring the data,
make sure to verify the BP zero balance value.
The recall event generated during the data update process will not be stored.
If the time setting is different between the data transferring monitors, the time of the recall
data and trend data may not be correctly displayed on the monitor which the data was
transferred.
Do not disconnect the HS-8000 while setting up the extended display.
REFERENCE
ON/OFF of data transfer function and the data selection to be transferred can be performed
on the "Initial Settings" menu.
(
Maintenance Manual "System Setup" P5-28)
4-4
Chapter 4 Preparation To Start Monitoring
The DS-8007 system can be used as a transport monitor by using the stored patient data and settings.
For the monitoring system using the DS-8007, DS-8400, DS-8500, DS-8900, the data during transport (when not
connected to the central monitor) can be transferred between the monitors using the transport monitor (DS-8007).
CAUTION
This function can be used only when the monitoring system is constructed with the DS-8007,
DS-8400, DS-8500, DS-8900. To use this function, refer also to the DS-8900 Operation
Manual.
1 2 3 4
4-5
Chapter 4 Preparation To Start Monitoring
1 When the patient information matches between the DS-8400 and DS-8007
When the patient information matches between the DS-8400 and DS-8007, the patient selection window will
not be displayed, and monitoring will continue.
[Monitor New Patient]: Starts monitoring a new patient by discharging the patients for both DS-8400 and DS-
8007.
Both data on the DS-8400 and DS-8007 will be deleted/initialized.
4-6
Chapter 4 Preparation To Start Monitoring
CAUTION
After the data transfer process, make sure that the patient information is correct.
REFERENCE
To transfer the alarm settings and parameter settings when [Monitor Patient of the Transport
Monitor] or [Monitor Patient of This Equipment] is selected, it is necessary to set [Transport]
for "Data Transfer" under [Menu > Initial Settings > System > Other], and set [ON] for "Alarm
Setup" and "Parameters" under [Menu > Initial Settings > System > Other > Data Selection
for Transfer].
To transfer the patient review data, refer to
"Transferring the Patient Review Data" P4-9.
When EMR Link Function is Used
When the EMR link function is used on the central monitor, the patient admit/discharge process will be performed
through the EMR.
4-7
Chapter 4 Preparation To Start Monitoring
CAUTION
Do not disconnect the DS-8007 while <Uploading> is displayed. Otherwise, upload process
cannot be completed.
CAUTION
If [OFF] is set for alarm setting and parameter setting on the "Data Selection for
Transfer" screen, a confirmation window to transfer the settings will not be displayed,
and the settings will not be transferred.
When the alarm settings and parameter settings are transferred, make sure the
settings are properly transferred to the DS-8400.
NOTE
When the alarm settings of the transport monitor is transferred:
Even if the alarm threshold of the transport monitor exceeds the alarm threshold limit
of the DS-8400, the exceeded alarm threshold will be applied to the DS-8400. Make
sure to check the alarm setting on the DS-8400 as the alarm threshold limit status will
be changed to "Limit Deactivating Mode".
4-8
Chapter 4 Preparation To Start Monitoring
1 For the following case, the review data of the DS-8007 will be saved to the DS-8400.
When the patient information matches between the DS-8400 and DS-8007
Trend, recall, full disclosure waveform, numeric data of the latest transferring data
When [Monitor Patient of the Transport Monitor] is selected on the confirmation window
Trend, recall, full disclosure waveform, numeric data for the selected patient from admittance
When the bed is transferred using the EMR link function
Trend, recall, full disclosure waveform, numeric data while the patient has been transferred
When the EMR link function is used, and the DS-8007 is temporarily disconnected
Trend, recall, full disclosure waveform, numeric data of the latest transferring data
2 The saved waveform data will be displayed on the full disclosure waveform screen as shown below. On the
explanation area on the review screen, the starting and ending time of transfer will be displayed.
(Orange): Indicates the starting time ( ) and ending time ( ) of transfer. For starting point and
ending point, one (1) second blank display is shown.
(Purple): Indicates the starting time ( ) and ending time ( ) of data while the patient was
monitored on another bed.
NOTE
To transfer the patient review data, it is necessary to use the CFast card to save the full
disclosure waveform data. If the CFast card is not used, the patient review data will not be
transferred.
4-9
Chapter 4 Preparation To Start Monitoring
Cancellation of Uploading
The uploading will be canceled under the following condition.Once canceled, the uploading will not resume.
To upload, disconnect and reconnect the DS-8007 to the DS-8400, and manually upload the data by selecting the
corresponding data on the central monitor. For procedure, refer to the operation manual of the central monitor.
<Cancellation of Uploading to the Central Monitor>
The DS-8007 was disconnected from DS-8400 during uploading.
The patient of the DS-8400 was discharged during uploading.
On the central monitor, discharge process or bed transfer was performed, or bed registration was canceled for
the uploading bed.
The power of the DS-8400 or central monitor was turned OFF.
DS-LAN cable was disconnected.
<Cancellation of Uploading to the Host Monitor (DS-8400)>
The DS-8007 was disconnected from DS-8400 during uploading.
The patient of the DS-8400 was discharged during uploading.
The power of the DS-8400 was turned OFF.
Uploading to the central monitor was canceled.
REFERENCE
For details of the MPDR function of the DS-8007, refer to the DS-8007 System Operation
Manual.
4-10
Chapter 4 Preparation To Stop Monitoring
To Stop Monitoring
3 When 10 seconds has elapsed without pressing the [Cancel] key, the
display will turn OFF and monitoring will stop.
The operation of the Super Unit and the Input Box will also stop.
Using the standby switch to stop monitoring will allow to easily resume monitoring by turning ON the standby
switch again.
CAUTION
If the remaining battery capacity becomes extremely low during battery operation,
monitoring will automatically stop.
If not using the device for a long period, disconnect the power cable and lithium-ion
battery.
NOTE
When the power is turned OFF, graphic/tabular trend data (Vigilance, ventilator), recall,
ST measurement, OCRG data will be erased after 5 minutes.
4-11
Chapter 4 Preparation Clock Setup
Clock Setup
CAUTION
If the time/date is not correctly set, or changed during monitoring, malfunction may occur
with the NIBP measurement, periodic printing, graphic/tabular trend data, and age
calculation from the birth date.
The time/date can not be set while connected to a wired network system. The time/date will
synchronize with the central monitor.
If the time/date is changed, the time/date for all the saved patient data (trend, list, recall, etc.)
will also change.
The printed time/date before changing and the displayed time/date after changing will differ.
Also, the data transmitted to the central monitor before the time/date is changed will be
displayed on the central monitor with the previous time/date.
The entered date/time will be set.(The number of seconds will be set to "00".)
Press [Cancel] to cancel the time/date setup.
4-12
Chapter 4 Preparation Installing the Recording Paper
CAUTION
About the Recording Paper
Use only "OP050-02TDR" for the recording paper.
If the surface treatment and thickness of the recording paper are different, it may result
in poor print quality.
Storing the Recording Paper
Since the recording paper is thermal type, inappropriate storage may change the quality of
the printed content, and make it illegible.
When storing the recording paper, follow the precautions below.
Store in a place where light is shut off and avoid direct sunlight.
Do not leave the paper in a high temperature (50 °C/122 °F and above).
Do not store the paper in a polyvinyl chloride bag.
Do not superpose the papers until the diazo copy is completely dried.
Do not expose the paper to alcohol, hydrochloric acid, or ester ketone.
Avoid using adhesive agents other than water based glue.
Installing the Recording Paper
When installing the recording paper, pay attention not to touch the thermal head or
sensor. The temperature of those parts rises immediately after printing and may cause
burn injury. Also, it may cause failure to the thermal head and sensor.
Do not operate the device with wet hand. Doing so may short the thermal head.
4-13
Chapter 4 Preparation Installing the Recording Paper
NOTE
Place the paper so that the "FUKUDA DENSHI" logo is outside and facing up.
NOTE
Push until it locks into place with a click sound.
4-14
Chapter 5 Admit/Discharge To Display the "Admit/Discharge" Screen
Chapter 5 Admit/Discharge
This menu allows setup of admitting, discharging, suspend monitoring of a patient, and selection of the user mode
(display configuration) according to the monitoring purpose.
CAUTION
If monitoring of new patient is started without performing a discharge procedure of the
previous patient, new data will be added to the previous data which will result in inaccuracy.
Admit
2 3
1 4
6 5
The selected patient classification and icon will be displayed on the home display.
For adult:
For child:
For neonate:
The patient classification affects the accuracy of NIBP measurement, HR measurement, and RR
measurement. It also affects the delay time to generate the measurement data alarm.
The alarm delay time is the function to prevent frequent generation of the measurement data alarm by holding
the alarm generation for the duration of each delay time.
The alarm delay functions for HR/PR, BP, RR, SpO2, TEMP, EtCO2/InspCO2, Tachy, Brady, Ext Tachy,
Ext Brady.
Adult Child Neonate
30 mmHg to 280 30 mmHg to 180 30 mmHg to 130
SYS
mmHg mmHg mmHg
15 mmHg to 235 15 mmHg to 160 15 mmHg to 100
NIBP Measurement Range MAP
mmHg mmHg mmHg
10 mmHg to 200 10 mmHg to 150 10 mmHg to 90
DIA
mmHg mmHg mmHg
0 bpm, 30 bpm to
HR 0 bpm, 12 bpm to 300 bpm
300 bpm
Monitor 0.5 Hz to 40 Hz 1.6 Hz to 40 Hz
ESIS 1.6 Hz to 15 Hz 1.6 Hz to 15 Hz
Filter Mode
3 electrodes: 0.05 Hz to 100 Hz
Diagnosis
4, 5, 10 electrodes: 0.05 Hz to 150 Hz
Impedance Respiration 1.5 Hz 2.5 Hz
Alarm delay time 5 sec. 0 sec.
WARNING
The patient classification selection influences the precision of the QRS detection and
NIBP measurement. Make sure the correct selection is made.
The NIBP air hose corresponded to the set patient classification must be used to perform
NIBP measurement. (However, if the patient classification is child, NIBP air hose for adult
can be used.)
5-2
Chapter 5 Admit/Discharge Admit
CAUTION
When "Link with Patient Class." is set to [ON], and patient classification is changed, the
main mode will change to the selected mode on the "Link Settings". (
Manual "To Program the User Mode" P5-31)
Maintenance
When [Used] is selected, <Pacemaker> will be displayed at the upper part of the home display.
5-3
Chapter 5 Admit/Discharge Admit
By using the magnetic card reader, patient information can be entered from the magnetic card. The admittance
process will speed up compared to manually entering each information.
NOTE
To automatically enter the patient information from the magnetic card or barcode, it is
necessary to perform the setup in advance. (
Card Reader" P4-25)
Maintenance Manual "Using the Magnetic
[Change only patient info.] : Replaces the current patient information with the newly acquired information.
[Cancel] : Cancels the acquired data.
NOTE
Make sure the patient is discharged before replacing the patient information.
The item which the information was not acquired from the magnetic card or barcode will
be left blank. For the blank item, manually enter the information.
Entering Patient Information from the Patient Data Server (When DS-LANIII is used)
When the central monitor is connected to the patient data server, and the DS-LANIII network is used, patient
information can be automatically entered by searching on the patient data server.
NOTE
When a DS-LAN II network is used, patient information cannot be entered from the patient
data server.
CAUTION
When "Link with Patient Class." is set to [ON], and patient classification is changed by
acquiring patient information from the patient data server, the Main Mode will change to the
selected mode on the "Link Settings". (
Mode" P5-31)
Maintenance Manual "To Program the User
5-4
Chapter 5 Admit/Discharge Admit
When Using the Patient Data Server and Magnetic Card Reader (or Barcode Reader)
NOTE
Select [ON] for "Auto Reference to Central Monitor when Reading Patient ID" under [Initial
Settings>Magnetic Card Reader] in advance.
(
Maintenance Manual "Magnetic Card Reader Setup" P4-25)
2 The acquired patient information from the patient data server will be displayed in the "New Information" area.
If there is no applicable patient information, current patient information will be displayed in the "New
Information" area.
Select from [Change only patient info.] / [Discharge and admit as new patient.] / [Cancel].
[Change only patient info.] will replace the current patient information to the newly acquired information.
[Discharge and admit as new patient.] will initialize the current patient data/monitoring condition and admit the
searched patient as new patient.
[ Cancel ] will invalidate the acquired data.
NOTE
The item not acquired from the patient data server will be left blank.
For the blank item, manually enter the information.
5-5
Chapter 5 Admit/Discharge Admit
3 Press the [Search ID] key and start searching on the patient
data server.
2 Based on the entered patient ID, patient information will be searched on the patient data server through the
DS-LANIII network.
The searched patient information will be displayed under "New Information" .
Select from [Change only patient info.] / [Discharge and admit as new patient.] / [Cancel]. [Change only
patient info.] will replace the current patient information to the newly acquired information. [Discharge and
admit as new patient.] will initialize the current patient data/monitoring condition and admit the searched patient
as new patient. [ Cancel ] will invalidate the acquired data.
The item not acquired from the patient data server will be left blank. For the blank item, manually enter the
information.
NOTE
The item not acquired from the patient data server will be left blank.
For the blank item, manually enter the information.
5-6
Chapter 5 Admit/Discharge Discharge
Discharge
Discharging Procedure
CAUTION
If monitoring of new patient is started without discharging the previous patient, the
measurement data of the previous and new patient will become mixed up on the recall and
trend data.
When the discharge process is performed, patient data such as recall and trend will be
initialized. The parameter and alarm settings will be reset according to the settings made
under [Menu>Initial Settings>User I/F>Power ON/Discharge).
When the discharge process is performed on the central monitor, alarm will be reset
according to the setting on "Admit Setup" of the central monitor.
(
Maintenance Manual "Power ON/Discharge" P5-18)
If the power is turned OFF or if the system enters into standby mode soon after the discharge
procedure, the patient may not be discharged on the central monitor.
If it is necessary to turn OFF the power or enter into standby mode after the discharge
procedure, select [Standby] for "Discharge Mode" under [Initial Settings>User I/F>Power
ON/Discharge].
NOTE
Depending on the setting made for "At Discharge" under ([Initial Settings>User I/F>Power
ON/Discharge], some items may not be initialized.
(
Maintenance Manual "Power ON/Discharge" P5-18)
If the discharge procedure is performed during stopwatch operation, the counting will stop
and will be reset to "00:00:00" .
The monitoring condition after discharge can be set on "Discharge Mode" under [Initial
Settings>User I/F>Power ON/Discharge].
Data Description
Patient Data Trend, Tabular Trend, Recall, ST, OCRG, CO, Hemodynamics, Lung Function,
P-V/F-V control data will be erased.
The settings for recall, tabular trend, graphic trend, vigilance list will remain.
5-7
Chapter 5 Admit/Discharge User Mode
Data Description
Patient Information Erases the data of patient name, ID, sex, age.
The patient classification will not be initialized.
REFERENCE
For details, refer to [Initial Settings>User I/F>Power ON/Discharge].
(
Maintenance Manual "Power ON/Discharge" P5-18)
User Mode
CAUTION
The selected user mode will be stored even after the power is turned OFF or discharge
process is performed.
Before monitoring, make sure the current user mode is suitable for the patient's condition.
(
Maintenance Manual "User Mode Registration" P5-30)
REFERENCE
For the user mode, up to 9 main modes of display configuration and alarm settings can be
registered according to the patient's age and monitoring purpose.
Also, for temporarily changing the display configuration (ex. when checking the 12-lead
ECG), 6 sub modes of display configuration can be registered.
For the extended display, 3 modes for each extended display (1, 2) can be registered.
(
Maintenance Manual "User Mode Registration" P5-30)
3 When the extended display is used, select the mode for the
extended display.
5-8
Chapter 5 Admit/Discharge Suspend Monitoring
WARNING
After changing the mode, make sure that the monitoring setting is appropriate.
When the mode is changed, patient classification, alarm settings, etc. will be changed.
REFERENCE
The selected user mode will be stored even after the power is turned OFF. If a new patient
is admitted without changing the user mode, the monitoring will start with the previous user
mode.
The mode setting after the discharge operation can be set under [Initial Settings>User I/
F>Power ON/Discharge].
To change from the sub mode to the main mode, press [Return to Main Mode].
Refer to "Setup Item/Default Value" for the default setting of each mode.
(
Maintenance Manual "User Mode Registration" P5-30)
Suspend Monitoring
To Suspend Monitoring
On the home display, numeric data and waveform display will be
suspended.
REFERENCE
When the monitoring is suspended, telemetry transmission will cease. Note that the
5-9
Chapter 5 Admit/Discharge Suspend Monitoring
square wave will be displayed on the central monitor indicating the too far condition of the
telemetry.
The stopwatch counting will continue even when the monitoring is suspended.
The setting can be changed even when the monitoring is suspended.
When Both "Monitor Suspend Label" and "Monitor Suspend Timer" are set
On the home display, the time will start counting for the set duration.
When the set duration completes, alarm sound will generate (5 sec.
interval), and alarm indicator will light.
REFERENCE
To extend the monitoring suspended duration, press [Extend] to display the timer
selection.
When "Monitor Suspend Label" is set, but "Monitor Suspend Timer" is not set:
1 Press the [Menu], "Admit/Discharge" icon, [Monitor Suspend] keys.
5-10
Chapter 5 Admit/Discharge To Resume Monitoring
To Resume Monitoring
CAUTION
Resuming monitoring will also resume the suspended alarm.
The "Monitoring is suspended" message will disappear and monitoring will resume.
5-11
Chapter 5 Admit/Discharge To Resume Monitoring
5-12
Chapter 6 Alarm Function Alarm
The arrhythmia alarm can be turned ON or OFF, and arrhythmia detection level can be set.
WARNING
Set the appropriate upper and lower alarm limit for each parameter according to the
monitoring condition.
When the system alarm is suspended, all the alarms will be suspended even if the
parameter alarm is set to ON. Also, the alarms will not be stored as recall events.
If the upper/lower alarm limit of the parameter is set to OFF, or if arrhythmia alarm is set to
OFF, alarm will not function even if the system alarm is enabled. Pay attention when setting
them OFF.
NOTE
<Arrhythmia alarm OFF> will be displayed when the Asystole, VF, VT, Slow_VT, Tachy,
Brady, Ext Tachy, Ext Brady, or HR alarm is OFF.
If [Always ON] is selected for "Asystole, VF, VT Alarm" under "Initial Settings", the
Asystole, VF, VT alarm cannot be set to OFF.
(
Maintenance Manual "Alarm Related Setup" P5-4)
If [Check when OFF] is selected for "Asystole, VF, VT Alarm" under "Initial Settings", a
confirmation window will be displayed when the Asystole, VF, VT alarm is set to OFF.
6-1
Chapter 6 Alarm Function Alarm
REFERENCE
The arrhythmia detection level for tachycardia (Tachy), bradycardia (Brady), extreme
tachycardia (Ext Tachy), extreme bradycardia (Brady) alarms link with the upper and
lower alarm limit for HR/PR.
The tachycardia (Tachy) alarm generates when the value exceeds the HR/PR upper
alarm limit. When the upper alarm limit is OFF, alarm will not generate.
For the Ext Tachy alarm, the alarm threshold level cannot be set below that of Tachy
alarm.
The bradycardia (Brady) alarm generates when the value exceeds the HR/PR lower
alarm limit. When the lower alarm limit is OFF, alarm will not generate.
For the Ext Brady alarm, the alarm threshold level cannot be set above that of Brady
alarm.
Ext Brady 20 bpm to 295 bpm Pacer Not Pacing 20 bpm to 200 bpm
4 Press the [Detail Setup] key, and set HR Lower Limit for VT, HR Lower Limit
for RUN and HR Lower Limit for SVT.
6-2
Chapter 6 Alarm Function Alarm
The SpO2 second alarm function is available when the Super Unit HS-8312N, HS-8412N, DS-8007N, HG-820 is
connected.
When the SpO2 value is unstable around the lower alarm limit, the frequently generated alarm may be bothersome.
The second alarm function controls these frequent alarms.
This function generates the alarm only when the integral value (the accumulation of difference between the alarm
limit and SpO2 value at every second) reaches the preprogrammed second alarm threshold value.
NOTE
The SpO2 SEC alarm function utilizes SatSeconds TM technology of Medtronic.
SatSecondsTM is a trademark of Medtronic.
1 Integral Value
2 Alarm Generation
3 SpO2 Value
4 Alarm Limit
CAUTION
Whether to use the second alarm function and its threshold selection should be based on
the patient's clinical indication portent and medical evaluation.
If the SpO2 alarm and SEC alarm setup is set to [OFF], the second alarm integral value will
be set to 0.
6-3
Chapter 6 Alarm Function Alarm
1 Press the [Menu], [SpO2] ("Parameter") keys to display the "SpO2" setup screen.
[10]/ [25]/ [50]/ [100]: A circular second alarm indicator will be displayed
inside the numeric data box.
As the integral value increases, the indicator will begin to fill, and when it
is completely filled, an alarm will be generated.
ST Alarm Setup
Set the ST upper limit and lower limit for the reference waveform.
The alarm limit should be set for each measurement unit (mm/mV). The upper and lower limit can be set in 1 mm/
0.1 mV increments.
2
3
5
4
[OFF]: Alarms will not generate even if the alarm for each lead is set to ON.
6-4
Chapter 6 Alarm Function Alarm
5 Slide the / and set the upper, lower limit (±20 mm / ±2.0 mV) .
Alarm will be set to OFF if the value -20 mm / +2.0 mV or lower is selected.
Alarm will be set to OFF if the value +20 mm / +2.0 mV or above is selected.
The alarm settings can be verified in list format. The alarm settings for each parameter can be changed on this list.
3
2 Select from [All List] / [Meas. List].
[All List]: The settings for all the parameters will be displayed.
[Meas. List]: The settings for only the measured parameters will be displayed.
1 Select a parameter.
The alarm setup window will be displayed.
6-5
Chapter 6 Alarm Function Alarm
Detail Setup
The alarm-related setup such as alarm suspend time and alarm silence time can be performed.
2
3
4
5
6
7
2 Select [1 min.] / [2 min.] for "Suspend Time".
5 Select the "Alarm Sound Suspend Time" from 1 min., 2min., 5 min., 10min., 30 min., 60 min. ,90 min., 120 min.,
240 min., 360 min.
REFERENCE
The alarm silence time for the level L device status alarm ("Check electrodes", "NIBP
Check patient type, air hose", etc.) can be set.
(
"Device Status Alarm Message" P11-7)
[Link to Alarm Silence Time]: When the [Alarm Silence] key is pressed at occurrence of device status alarm,
alarm will be silenced for fixed amount of time set for "Silence Time".
If the alarm factor still remains at completion of silence time, the alarm sound will generate again.
If the same alarm occurs during the alarm silence time, the alarm sound will not generate.
If a new alarm occurs during the alarm silence time, the alarm sound for the new alarm will generate.
[Link to each new occurrence]: When the [Alarm Silence] key is pressed at occurrence of device status
alarm, the alarm will be silenced as long as the alarm factor remains regardless of the "Silence Time" setting.
While the same device status alarm is generated, the alarm will remain silenced.
If the alarm factor is resolved during the alarm silence time, the alarm will be canceled.
If the same alarm generates again during the alarm silence time, the alarm sound will generate.
6-6
Chapter 6 Alarm Function Alarm Limit Setup
The upper and lower alarm limit will be displayed on the home display.
NOTE
The alarm limit for the parameter with the alarm turned OFF will not be displayed
regardless of this setup.
If the alarm limit display for BP is [Graph], systolic value will be displayed.
Depending on the numeric data box type, alarm limit may not be displayed.
If the SpO2 lower alarm limit is set to 85%SpO2 or below, the alarm limit value will be
displayed regardless of the "Alarm Limit Display" setting.
This section explains the procedure to enable/suspend the system alarm, and to set the upper/lower alarm limit for
each parameter.
On this system,9 modes can be preprogrammed according to the monitoring purpose. By preprogramming the alarm
setting to each mode, the alarm setups at admittance of patient can be simplified by just selecting a mode. It is
recommended to program the mode in rough classification such as patient’s age, monitoring purpose (ICU or
surgery), and if necessary, perform unique setup for each patient.
WARNING
When the system alarm is suspended, all the alarms will be suspended even if the
parameter alarm is set to ON. Also, the alarms will not be stored as recall events.
If the upper/lower alarm limit of the parameter is set to OFF, or if arrhythmia alarm is set to
OFF, alarm will not function even if the system alarm is enabled. Pay attention when setting
them OFF.
6-7
Chapter 6 Alarm Function Alarm Limit Setup
<xxx s> indicates the remaining time. The system alarm will be enabled when the suspended time
completes.
The alarm sound can be suspended for fixed amount of time. There are two ways to suspend the alarm sound, which
are "Alarm Silence" and "Alarm Sound Suspend".
The "Alarm Silence" function suspends the alarm sound for fixed amount of time (1 min. / 2 min.).
The "Alarm Sound Suspend" function suspends the alarm generation in advance such as during surgery when the
alarm generation is expected. Alarm monitoring will continue even while the alarm sound is suspended. The alarm
sound suspend duration can be selected from 1 min., 2min., 5 min., 10min., 30 min., 60 min. ,90 min., 120 min., 240
min., 360 min.
1 To silence the alarm, press the [Alarm Silence] key (fixed key).
6-8
Chapter 6 Alarm Function Alarm Limit Setup
2 To suspend the alarm sound, press the Alarm Silence key (fixed key) for more than 3 seconds.
NOTE
If the [Alarm Silence] key is pressed while the alarm sound is generated, it will bring the
system to "Alarm Silence" condition and not the "Alarm Sound Suspend" condition.
During the "Alarm Sound Suspend" duration, other bed alarm sound will not generate.
6-9
Chapter 6 Alarm Function Alarm Limit Setup
The alarm for each parameter can be turned ON or OFF, and upper and lower alarm limit can be set.
WARNING
Set the appropriate upper and lower alarm limit for each parameter according to the
monitoring condition.
When the system alarm is suspended, all the alarms will be suspended even if the
parameter alarm is set to ON. Also, the alarms will not be stored as recall events.
If the upper/lower alarm limit of the parameter is set to OFF, or if arrhythmia alarm is set to
OFF, alarm will not function even if the system alarm is enabled. Pay attention when setting
them OFF.
When the numeric data acquired from an anesthesia delivery system is displayed, the
following alarms cannot be set. Also, alarm will not generate.
InspCO2/EtCO2, InspO2/ExpO2, InspN2O/ExpN2O, InspAgent/ExpAgent, MAC, ExpMV,
PEAK, PEEP
2
3
REFERENCE
The standard parameters will be displayed on the Menu screen. The parameters to be
displayed here are selectable.
(
Maintenance Manual "Alarm Related Setup" P5-4)
6-10
Chapter 6 Alarm Function Alarm Limit Setup
By releasing the finger from the key, fine-tune keys will appear for a fixed period of time.
REFERENCE
indicates the current measurement value.
REFERENCE
[Auto] key will be displayed only when [Enable] is set for "Auto Alarm Setup" under "Initial
Settings".
To maintain the alarm setting even after the power is turned OFF or after the discharge
procedure, store the setting to one of the alarm modes, or select "Backup" for "Alarm"
on the "Backup at Discharge" menu (Monitor Setup).
(
Maintenance Manual "Display/Print Setup" P5-13.)
By setting the alarm threshold limit ("Initial Settings") in advance, the alarm threshold can be limited within the
preprogrammed range. When the alarm threshold limit function is enabled, threshold limit will be displayed beside
the alarm bar.
( Maintenance Manual "Alarm Related Setup" P5-4)
Above is an example of alarm threshold limit setting where HR is set to [Enable], and upper and lower limits are set
to 180 bpm and 40 bpm respectively.
NOTE
The alarm threshold limit can be set for each parameter. When enabling this function, make
sure the upper and lower limits are set appropriately.
When the alarm threshold limit function is enabled, pressing the [Auto] key for alarm settings
will set the alarm threshold within the limit range.
6-11
Chapter 6 Alarm Function Alarm Limit Setup
Above is an example of HR upper threshold limit being deactivated. The upper limit keys are turned to blue
indicating that the upper limit 180 bpm can be exceeded.
NOTE
If the alarm threshold set on the central monitor exceeds the threshold limit set on the DS-
8400, the alarm threshold set on the central monitor will be applied. Make sure to check the
alarm setting on the DS-8400 as the alarm threshold limit status will be changed to "Limit
Deactivating Mode".
If the alarm threshold of the transport monitor exceeds the alarm threshold limit of the DS-
8400, the exceeded alarm threshold will be applied to the DS-8400. Make sure to check the
alarm setting on the DS-8400 as the alarm threshold limit status will be changed to "Limit
Deactivating Mode".
If the alarm threshold of "Setup at Discharge" exceeds the alarm threshold limit, the
exceeded alarm threshold will be applied. Make sure to check the alarm setting at
admittance as the alarm threshold limit status will be changed to "Limit Deactivating Mode.
f the monitor mode is changed, and the alarm threshold of the current monitor mode exceeds
the threshold limit, this alarm setting will be applied. Make sure to check the alarm setting
on the DS-8400 as the alarm threshold limit status will be changed to "Limit Deactivating
Mode".
6-12
Chapter 6 Alarm Function Alarm Assist Screen
On the alarm assist screen, maximum of 24 hours of trend data for the corresponding parameter will be displayed.
Alarm limit can be set by using the past trend data as reference.
1 To display the alarm assist screen, press [Menu], select a parameter, and press on the corresponding
parameter setup screen.
Or, press the numeric data box on the home display, and press on the corresponding parameter setup
screen.
2 4
3
Dragging the slider to the right will display newer data, and dragging it to the left will display older data.
Pressing [24h] will switch the display by 24 hours.
6-13
Chapter 6 Alarm Function Alarm Assist Screen
6-14
Chapter 7 Monitoring To Display the Parameter Setup Screen
Chapter 7 Monitoring
To Display the Parameter Setup Screen
This section explains how to display the "Parameter Setup" of monitoring parameters.
1 Press the [Menu], and then select the parameter to perform the setup.
Or, press the numeric data box on the home display, and press on the corresponding parameter setup
screen.
When the numeric data box on the home display is pressed, a floating window
for the basic setup such as size/scale will be displayed.
To display the "Parameter Setup" screen for detailed setup, press .
7-1
Chapter 7 Monitoring ECG
ECG
This section explains the procedure for ECG measurement preparation and monitoring condition setup.
CAUTION
Do not use different types (materials) of electrodes at the same time.
The difference between the polarization potential from each electrode may interfere
monitoring.
ECG measurement part is Type CF applied part, but it is not intended to directly apply on
patient's heart.
2 Clean the electrode sites with alcohol wipes or other skin preparation.
REFERENCE
Use the recommended skin cleaner.
3 After opening the package, peel off the backing of electrode, and attach to the patient.
NOTE
After opening the package, pay attention not to touch the electrode gel.
7-2
Chapter 7 Monitoring ECG
Electrode Placement
Depending on the lead cable type, 3-electrode/4-electrode/5-electrode/10-electrode placements are available. Using
the 4-electrode, 5-electrode or 10-electrode application allows simultaneous monitoring of 2 ECG waveforms, and
high accuracy of arrhythmia analysis can be attained. (1 to 12 waveforms can be displayed depending on the number
of electrodes.)
Also, the displayed lead type can be changed.
7-3
Chapter 7 Monitoring ECG
NOTE
Electrode Placement for 12-Lead ECG Analysis
When acquiring 12-lead ECG signals, Fukuda Denshi recommends placing the limb
electrodes anywhere along the arms and legs as shown below.
However if it is difficult, use the Mason-Likar 12-lead system.
To reduce the waveform differences from the standard 12-lead, Fukuda Denshi
recommends that the torso placement of the R and L electrodes be near as possible to each
arm, in the infraclavicular fossae, within the area unaffected by myoelectricity. )
7-4
Chapter 7 Monitoring ECG
There are various types of disposable electrodes for ECG measurement depending on the connection method with
the lead cable and materials which the electrodes are made of. Make sure to use the appropriate electrodes which will
make full use of the characteristics.
Do not reuse/resterilize the disposable electrodes.
For details of usable lead cables, refer to
"ECG Accessory" P13-4
CAUTION
The indication for continuous use of the electrode is about one day.
Replace the electrode if the skin contact gets loosen due to perspiration, etc.
When an electrode is attached to the same location for a long period, some patients may
develop skin irritation. Check the patient's skin condition periodically and change the
electrode site as required.
When using the electrosurgery-proof type ECG relay cable, the impedance respiration
cannot be measured, and its numeric data and waveform will not be displayed. When
measuring in an environment where electrosurgery is not performed, make sure to use the
standard ECG relay cable.
NOTE
Use only the specified relay cables, lead cables, and electrodes.
The conductive parts of electrodes and associated connectors for applied parts, including
the neutral electrode, should not contact other conductive parts including earth.
7-5
Chapter 7 Monitoring ECG
4 Plug in the relay cable to the ECG input connector (green) of the HS-8000.
5 Adjust the waveform size and position, and change the monitoring lead as necessary.
(
"ECG Parameter Setup" P7-6)
ECG Parameter Setup
Press the [Menu], [ECG] keys to display the "ECG" setup screen.
7-6
Chapter 7 Monitoring ECG
CAUTION
The threshold level for arrhythmia detection changes with ECG waveform size. Set a proper
waveform size for monitoring. When the ECG waveform size is x1/4, x1/2, or x1, the
arrhythmia detection level is 250 μV. When the ECG waveform size is x2 or x4, the
arrhythmia detection level is 150 μV.
Automatic size/position of the ECG is effective only at the time the [Auto] key is pressed.
This does not continuously adjust the size and position.
The waveform size and position cannot be set if the waveform is not displayed. Refer to "To
Configure the Display" P10-4, and change the display configuration as necessary.
The threshold level for HR detection changes with ECG waveform size. Set a proper
waveform size for monitoring.
REFERENCE
By setting the [ECG Size (All Leads)] key as user key, ECG size for all leads can be changed
at once.
(
"User Key Setup" P10-10)
1 Press the key for "ECG1" to "ECG12", and display the "Size" selection
window.
When the display layout is "12-Lead", the waveform size can be set
differently for limb leads and chest leads.
3 If the waveform is difficult to see due to ECG amplitude, press / and set the baseline position to 0 mV.
The baseline position for the waveform display and printing will be adjusted.
7-7
Chapter 7 Monitoring ECG
Lead Selection
Set the monitoring lead.
CAUTION
The leads for arrhythmia detection, central monitor display, printing are fixed as ECG1 and
ECG2. Set the most appropriate leads with high QRS for ECG1 and ECG2, especially for
arrhythmia detection.
The alarms for HR, Tachy, Brady will not be generated when the electrode for ECG1 or
ECG2 lead is detached, and for 30 seconds after the electrode is reattached.
1 Press the key for "ECG1" to "ECG12", and display the "Lead" selection
window.
When the display layout is "12-Lead", select the lead for ECG1 and ECG2
on the lead selection window.
REFERENCE
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under "Initial
Settings". (
Maintenance Manual "Alarm Related Setup" P5-4)
When [Auto] is set, the upper and lower limit will be automatically set to +40 bpm and -40
bpm to the current value respectively. The lower limit will be clipped to the setting made for
"HR/PR Lower Limit during Alarm Auto Setting" (Menu>Initial Settings>Alarm).
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto setting is
used, the upper and lower alarm limit will be clipped to the set limit.
However if "HR/PR Lower Limit during Alarm Auto Setting" is also set, HR lower alarm limit
will be clipped to the larger value.
Ext Tachy will be set to HR upper limit+10 bpm, Ext Brady will be set to HR lower limit-10
bpm. When the set value exceeds 300 bpm for the upper limit and 20 bpm for the lower limit,
the setting will be clipped to 300 bpm and 20 bpm respectively
When [Auto] is set for Ext Tachy, Ext Brady, the same setting, HR upper limit+10 bpm, HR
lower limit-10 bpm, will be set respectively.
7-8
Chapter 7 Monitoring ECG
Detail Setup
1 3 5 7
2 4 6 8
9 11 13 15
10 12 14
1 Set the filter mode.
CAUTION
The ESIS mode cannot completely reduce the electrical noise, and may erroneously
detect the pacemaker spike.
The ESIS mode should be selected only when a high frequency noise largely affects the
HR measurement.
In ESIS Mode, artifacts such as electrosurgical noise or EMG can be largely reduced, but
QRS amplitude attenuation, waveform distortion, or ST segment change may occur
compared with other filter modes.
7-9
Chapter 7 Monitoring ECG
NOTE
When the filter mode is changed, a notch will appear on the ECG waveform due to the
change in frequency characteristic as shown below.
[Used]: Pacemaker pulse will be detected and pace pulse mask function will be performed for set duration.
[Not Used]: Pacemaker pulse will not be detected.
[ECG]: HR synchronized mark will be displayed. The synchronized tone will turn ON.
[SpO2-1]/[SpO2-2]: SpO2 synchronized mark will be displayed.
The synchronized tone will turn ON.
[BP]: BP synchronized mark will be displayed. The synchronized tone will turn ON.
CAUTION
Precautions about Pacemaker Pulse Detection
There are some cases when the pacemaker pulse cannot be detected depending on
the pacemaker type, pulse voltage, pulse width, electrode lead type (unipolar, bipolar),
or electrode placement which causes the pacemaker pulse amplitude to decrease,
and disables the pacemaker pulse detection.
If signals similar to a pacemaker pulse are present, such as electric blanket noise or
excessive AC frequency noise, these may be erroneously detected and displayed as
a pacemaker pulse.
When a spontaneous QRS and pacemaker pulse overlap (ex. fusion beat, etc.), QRS
detection cannot be performed properly. In this case, the heart rate is degraded.
If a pacemaker pulse is continuously detected due to AC frequency interference, QRS
detection will be suspended and the heart rate will be reduced. Arrhythmia will not be
detected either.
7-10
Chapter 7 Monitoring ECG
REFERENCE
Pacemaker Pulse Detection Algorithm 3
WARNING
If the QRS pace mask function is set to [OFF]/[10ms]/[20ms]/[40ms], the pace pulse may
be erroneously be detected as a QRS complex and HR alarm or asystole alarm may not
generate due to incorrect HR (counting pace pulse as QRS complex). Select [OFF]/
[10ms]/[20ms]/[40ms] only if you are sure that pacing failure will not occur, or when the
patient can be constantly monitored.
REFERENCE
For the patients using pacemakers, there are cases when the pacing waveform may not
occur in spite of the pacing stimulus. This condition is called "pacing failure".To avoid
detecting pacemaker pulses as a QRS complex, this monitor has a function to suspend
QRS detection for a fixed amount of time starting from the detection of the pacing
stimulus. This function is called "pace pulse mask".
But if the pacemaker does not detect the patient's spontaneous heartbeat (sensing
failure), and the pacing stimulus is applied at the same timing as QRS, this pace mask
function may erroneously mask the QRS and cause the heart rate measurement to
decrease.
To avoid this, QRS pace pulse mask function can be set to [OFF]/[10ms]/[20ms] for
correct measurement of the heart rate. (Default: Auto)
7-11
Chapter 7 Monitoring ECG
1 2 3 4
1 Pacemaker Pulse
2 Pacing waveform caused by pacemaker pulse
3 No waveform in spite of pacing stimulus
4 Pacemaker pulse and spontaneous heartbeat occurring at the same time
[ON]: Only the amplitude with frequency component under 1 Hz will be attenuated to prevent the ECG
baseline drift.
The patient signal display will delay about 0.5 seconds.
On the home display, "Drift-F ON" will be displayed in the information area, and "DF" will be displayed in the
waveform area.
7-12
Chapter 7 Monitoring ECG
ECG2.
If artifact is present on one of the waveforms, HR will be calculated using only the stable ECG waveform.
If artifact is present on both of the waveforms, HR value will be displayed as "---".
When ECG electrodes are detached, arrhythmia analysis cannot be performed, and <Lead OFF> message
will be displayed. Alarm sound will also generate.
NOTE
When the patient classification is set to [Neonate], "HR Delay" will be set to [OFF].
When the patient classification is set to [Adult] or [Child], and "HR Delay" is set to [ON],
"Inst." or "Av." will not be displayed inside the HR or HR/PR numeric data box.
[ON]: AC filter which attenuates the AC noise of 50 Hz to 60 Hz will be set. "AC" will be displayed in the
waveform area.
10 Set the "Auto Lead". The automatic lead switching will be performed for ECG 1 and ECG 2.
During Lead OFF
R III III
4-electrode
L II II
5-electrode L/L+C II II
C II aVR
10-electrode L/L+C II II
C,C2 to C6 II aVR
[ON]: When lead-off condition occurs, the lead will automatically switch.
Also, <Check Electrodes> will be displayed.
[OFF]: The lead will not automatically switch even when lead-off condition occurs.
[ON]: If 2 or more boxes are used for ECG numeric data display, ST level, VPC, arrhythmia alarm factor will
be displayed inside the ECG numeric data box.
[OFF]: ST level, VPC, arrhythmia alarm factor will not be displayed inside the ECG numeric data box.
NOTE
When a relay cable for 5-lead or 10-lead is used with a 3-lead cable, it will be judged as
lead-off condition and <LEAD OFF> message will be displayed.
If a 3-lead cable is intentionally used, select [ON] for "3lead Override" to avoid displaying
the <LEAD OFF> message.
If [ON] is selected for "3-lead Override" even though 4-lead, 5-lead, or 10-lead relay
cable is used with all the lead cables and electrodes connected, it will be acknowledged
as only 3 electrodes are used and only one waveform will be displayed.
Also, artifact may interfere to the waveform or lead-off information may become incorrect.
7-13
Chapter 7 Monitoring ECG
When using the "3lead Override" function, use only 3 electrodes of L, R and F.
[ON]: The input waveform will be displayed even during lead-off condition.
[OFF]: Baseline will be displayed during lead-off condition.
NOTE
If chest lead is set for ECG1/ECG2, chest lead OFF condition will be notified by an alarm
generation even if [Disable] is set for "Chest Lead-OFF".
[ON]: HR data before the noise detection will be retained, and synchronizing source will switch to SpO2, BP.
[OFF]: HR data before the noise detection will not be retained, and synchronizing source will not switch to
SpO2, BP.
NOTE
Even if the synchronizing source is switched to SpO2, the ECG tone will remain and not
change.
7-14
Chapter 7 Monitoring ECG
CAUTION
When the waveform and numeric data display is set to OFF, the alarm generation and
tabular trend input will also cease.
When ECG electrodes are attached to the patient with the ECG display set to OFF, the ECG waveform and
numeric data will be automatically displayed after 10 seconds.
7-15
Chapter 7 Monitoring Respiration
Respiration
This section explains about the respiration measurement by the impedance, CO2, ventilator, or SpO2 method and the
measurement condition settings.
WARNING
The SpO2 respiration measurement is not intended for use as an APNEA monitor.
CAUTION
When a defibrillator is used during respiration monitoring, a large offset voltage will be
placed on the ECG electrodes, which may cause interruption of monitoring for a few
seconds.
When using the electrosurgery-proof type ECG relay cable, the impedance respiration
cannot be measured, and its numeric data and waveform will not be displayed. When
measuring in an environment where electrosurgery is not performed, make sure to use the
standard ECG relay cable.
2 Verify that the respiration waveform and respiration rate is displayed on the home display.
NOTE
Adjust the detection lead, waveform size, baseline position, and sweep speed for
optimum waveform display.
(
"To Configure the Display" P10-4)
To change the lead, press the lead name on the waveform area, and display the lead
selection window.
7-16
Chapter 7 Monitoring Respiration
Press the [Menu], [RESP] keys to display the "RESP" setup screen.
The example when the HS-8312 is used is shown below.
1 2
4 3
5
6
7
9
8
1 Press the key for "Size" to adjust the waveform size and baseline position.
REFERENCE
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under "Initial
Settings". (
Maintenance Manual "Alarm Related Setup" P5-4)
When [Auto] is set, the upper and lower limit will be automatically set to +20 Bpm and -
20 Bpm to the current value respectively.
The adjustable increment for upper and lower limit depends on the patient classification
7-17
Chapter 7 Monitoring Respiration
and "RR Alarm Increment" setting under "Initial Settings" > "User I/F".
Normal Small
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto setting
is used, the upper and lower alarm limit will be clipped to the set limit.
REFERENCE
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under "Initial
Settings". (
Maintenance Manual "Alarm Related Setup" P5-4)
When [Auto] is set, the apnea alarm setting registered for the currently selected mode will
be applied.
The upper limit can be set in 1 second increment. There is no lower limit.
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto setting
is used, the upper and lower alarm limit will be clipped to the set limit.
WARNING
The RR/APNEA alarm will not be generated unless the numeric data box corresponded
to the selected RR/APNEA alarm source is displayed. Make sure to display the numeric
data box for the parameter set as the RR/APNEA alarm source.
The SpO2 respiration measurement is not intended for use as an APNEA monitor.
7-18
Chapter 7 Monitoring Respiration
CAUTION
If the "RR/APNEA Alarm Source" setting is other than [Impedance] (or, if [Auto] selects a
setting other than [Impedance]), the RESP waveform will not be transmitted on a wired
network.
RR_SpO2 is a parameter which can be measured on the DS-8007N.
[Impedance]: RR alarm will be generated based on the impedance respiration curve. The RR synchronized
mark based on impedance respiration will be displayed.
[CO 2/ GAS]: When multigas unit/anesthesia delivery system is used, RR alarm will be generated based on
the RR measured by the multigas unit/ anesthesia delivery system.
If multigas unit is not used, RR alarm will be generated based on the RR measured by the HPD-810/HPD-
820 (Capnostat 5) or HCP-810/HCP-820. The RR synchronized mark based on CO2 waveform will be
displayed.
[Ventilator]: RR alarm will be generated based on the RR measured by the ventilator. The RR synchronized
mark based on ventilator measurement will be displayed.
[SpO2]: RR alarm will be generated based on the RR measured by the SpO2 module. The RR synchronized
mark will not be displayed.
REFERENCE
When the amplitude of the respiration waveform decreases due to causes such as
respiratory pause, the ECG waveform may be superimposed on to the respiration
waveform, making the RR equal to the HR. This condition is called CVA (Cardio-Vascular
Artifact), and is detected using the CVA detection function.
This function will be effective only when [Impedance] is set as the "RR/APNEA Alarm
Source" or, when [Auto] selects impedance respiration.
If the ECG waveform is superimposed on to the respiration waveform with HR (RR) of 30
Bpm or above for 20 seconds (10 seconds for neonates) or more and if the "CVA Detect"
is set to [ON], the <CVA detected> message will be displayed, and an alarm sound will
be generated.
[ON]: When CVA is detected, alarm will generate and message will be displayed.
[OFF]: CVA detection will not be performed.
WARNING
If a patient is using an adaptive (minute ventilation) pacemaker, "Impedance
Measurement" should be set to OFF.
The respiration measurement using the impedance method conducts high-frequency and
weak current between the ECG electrodes attached to the patient, and measures the
potential difference between the electrodes caused by thoracic movement using the
synchronous rectification system. For the patient using the adaptive (minute ventilation)
pacemaker, the pacemaker measurement signal and the high-frequency current of this
device interferes with each other which causes incorrect respiration measurement.
7-19
Chapter 7 Monitoring BP
RR data will not be displayed. A high-frequency current which is a measurement signal will not be
conducted.
This section explains about the procedure of BP1 to BP 8 measurement preparation and measurement condition
setup.
CAUTION
Do not reuse / re-sterilize the disposable type transducers.
If using a reusable blood pressure transducer, disinfect it according to the manufacturer's
guidelines.
The long-term use of the blood pressure transducer, tube and catheter may increase the risk
of infection. Perform periodic replacement with new one. The guidelines of the CDC
(Disease Control and Prevention) recommend replacing within 96 hours.
If the ambient temperature of the blood pressure transducer has changed greatly, the zero
balance may cause the drift. Perform the zero balance again.
An operator must not get away from a patient during the BP measurement. However, when
getting away from the patient is necessary, do not activate the Alarm Suspend and Silence
functions in order not to miss any sudden changes in the patient's condition.
(
"To Set the System Alarm (ON or Suspend)" P6-7)
(
"To Silence or Suspend the System Alarm Sound" P6-8)
Be sure to perform Daily Check. Use of faulty device might harm the patient or operator.
(
"Daily Check" P4-1)
If the Device Status Alarm occurs or if you feel the unusual operation of the device, perform
the inspections to confirm the safety or contact our service representative.
(
"Device Status Alarm Message" P11-7)
The BP value will not be displayed until zero balance is performed after the power is turned
ON. Make sure to perform the zero balance.
Once the zero balance is performed, the zero balance information will be maintained, and
the BP value will be displayed.
7-20
Chapter 7 Monitoring BP
BP Monitoring
The HS -8000 utilizes multiparameter amplifier input method which allows monitoring of 2 channels of BP through
the 2ch BP conversion cable, CJO-P01B-DJ0.5. The BP relay cable can be directly connected to the multiparameter
connector.
The measurement is also possible using the HM-800/HM-801 Multi Module
( "Multiparameter Connector Setup for BP, TEMP, CO Measurement" P7-114)
1 Multiparameter Connector
2 2ch BP Conversion Cable CJO-P01B-
DJ0.5 1
3
2
7-21
Chapter 7 Monitoring BP
REFERENCE
The following procedure explains the case when a BP transducer (LS575 series) is used.
If using other transducers, refer to the operation manual for the corresponding
transducer.
4 Align the bracket to patient’s heart position (about 1/2 of the chest depth).
5 Inject 1000 units of heparin into the saline bag, mix thoroughly and puncture the infusion line through the
same hole.
6 Set the saline bag to pressure bag, and hang from the infusion device. Fill saline to about 1/3 of the drip.
7-22
Chapter 7 Monitoring BP
7 After loosening the zero-port plug, push the flash button to perform priming to remove air bubbles.
8 Verify that all air bubbles are removed, and tighten the zero-port plug. Turn on the zero-port plug side of
the open-air three-way valve.
7-23
Chapter 7 Monitoring BP
2 Press the BP numeric data box (parameter key) on the home display.
3 Press the [Zero] key on the BP floating window or BP parameter setup screen.
7-24
Chapter 7 Monitoring BP
5 Verify that the BP waveform and numeric data is displayed on the home display.
CAUTION
The zero balance procedure is required for the following case.
When starting the measurement.
When the position of the heart has changed due to body movement.
When the position of the transducer has changed.
When measuring for a long period of time and there is a possibility of measurement
error due to change in ambient temperature, etc.
When a connector is connected/disconnected, or a transducer is replaced.
7-25
Chapter 7 Monitoring BP
The zero balance for all the displayed BP can be performed using the user key.
If any of the BP is in progress of measurement, perform the zero balance on each BP parameter setup screen.
3 Verify the BP waveform is positioned at zero, and "0" is displayed for the BP value.
NOTE
If <FAILED> is displayed, the three-way valve may not be opened to air, artifact is
present, or the transducer may be defective. Check the cause and try the zero balance
procedure again.
If <DRIFT> is displayed, verify that all the connections are secure.
CAUTION
When the transducer or tubing is replaced, make sure to perform the zero balance.
Otherwise, accurate measurement will not be performed.
<READY> will not be displayed unless the three-way valves of all pressure transducers
are opened to air. If the status is not displayed, or if <MEASURE> is displayed, check if
the three-way valve of pressure transducers are opened to air.
7-26
Chapter 7 Monitoring BP
By using the [BP Zero] key on the HS-8000 or Multi Module, zero balance can be performed for all the BP even if
not displayed.
When the BP zero balance properly completes, a beep sound will generate for 1 second and LED will light
in blue.
When the BP zero balance fails, a beep sound will generate for 3 seconds and LED will flash in blue.
NOTE
Using the [BP Zero] key will allow to perform zero balance for all the BP even if not displayed
on the home display.
For the BP channel with the transducer in progress of measurement, zero balance will not
be performed.
2 Verify that "Zero ready" is displayed on the BP parameter setup screen for BP1 to BP8, and press the [Zero]
key.
3 Verify the BP waveform is positioned at zero, and "0" is displayed for the BP value.
NOTE
If a message, "Zero failed" is displayed, the three-way valve may not be opened to air,
artifact is present, or the transducer may be defective. Check the cause and try the
zero balance procedure again.
If a message, "Zero drift" is displayed, verify that all the connections are secure.
7-27
Chapter 7 Monitoring BP
BP Parameter Setup
%3 7(03 %3
%3 %3
When the DS-8007 is used, the temperature measured on temperature jack is fixed as T1, T2.
NOTE
The channel settings can be changed on the "Multi-amplifier" screen ([Initial Settings >
System > Unit Module]).
(
Maintenance Manual "Unit Module Setup" P4-19)
Label Setup
REFERENCE
Description of Each Label:
ART (Arterial Pressure)
PAP (Pulmonary Artery Pressure)
CVP (Central Venous Pressure)
ICP (Intra-cranial Pressure)
IAP (Intra-aortic Balloon Pumping Pressure)
LVP (Left Ventricular Pressure)
US1 to US5: User labels (3 characters) which can be set on the "Initial Settings".
(
Maintenance Manual "User Label Setup" P5-9)
NOTE
US3 to US5 cannot be selected for the device connected to DS-LANII.
7-28
Chapter 7 Monitoring BP
CAUTION
The default setting of “ART Catheter Check Message” is [OFF].
When “ART Catheter Check Message” is set to [ON], alarm will generate when the
transducers are opened to air.
CAUTION
Note that Systolic Pressure (SYS)=Peak Systolic Pressure (PSP) when reviewing graphic
trend, data base, or when setting the alarm.
When ECG is not measured, PDP cannot be calculated.
7-29
Chapter 7 Monitoring BP
PCWP Measurement
When PAP is set as BP label, the mean value can be displayed as PCWP (Pulmonary Capillary Wedge Pressure).
The PCWP value will be displayed inside the PAP (BP label) numeric data box with the measurement time.
It will be also displayed on the trend data.
7-30
Chapter 7 Monitoring BP
Scale Setup
CAUTION
When wireless network is used, BP waveform with a scale above the set scale will not be
properly transmitted. The displayed BP scale should be within the set scale.
NOTE
Select the full scale for displaying and printing.
The scale selection will differ depending on the label as shown below.
Change the scale before the freeze operation, as the waveform will be deleted if the scale
is changed after the freeze operation.
Scale
5 10 15 20 30 40 50 75 100 150 200 250 300 mmHg
BP Label 1 2 3 4 5 6 8 12 16 20 24 32 40 kPa
20 40
cmH2O
BP1 to BP8
Yes Yes Yes Yes Yes Yes Yes Yes
User Label
ART,
IAP, Yes Yes Yes Yes Yes Yes Yes
LVP
PAP Yes Yes Yes Yes Yes Yes Yes Yes Yes
CVP Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes
ICP Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes
1 Press the key for "Scale Selection", and display the scale selection window.
Alarm Setup
1 Set the BP alarm.
(
"Alarm Limit Setup for Each Parameter" P6-10)
NOTE
Set the upper limit in the range of 2 mmHg to 300 mmHg / 0.2 kPa to 40.0 kPa. If a value
above 300 mmHg / 40.0 kPa is set, the upper alarm will turn OFF.
Set the lower limit in the range of 0 mmHg to 295 mmHg / 0 kPa to 39.5 kPa. If a value
below 0 mmHg / 0 kPa is set, the lower alarm will turn OFF.
Alarm will not generate until 30 seconds has passed after the zero balance or after the
transducer has been opened to air.
REFERENCE
Select ON/OFF of BP alarm and set the upper and lower alarm limit for systolic (S),
diastolic (D), and mean (M) BP.
The alarm limit should be set for each unit (mmHg/kPa).
The adjustable increment will be according to the "BP Alarm Increment" setting. (Normal/
Small).
7-31
Chapter 7 Monitoring BP
(
Maintenance Manual "Display/Print Setup" P5-13)
The adjustable increment for upper and lower limit changes from 50 mmHg / 7 kPa.
When [Auto] is set for the BP label of BP1/ART, the upper and lower limit will be
automatically set to +40 mmHg / +5 kPa and -20 mmHg / -3 kPa respectively to the
current value.
When [Auto] is set for the BP label other than BP1/ART, the upper and lower limit will be
automatically set to +20%, –20% respectively to the current value.
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under "Initial
Settings". (
Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto setting
is used, the upper and lower alarm limit will be clipped to the set limit.
1 4
2 5
3 6
7
8
9
7-32
Chapter 7 Monitoring BP
[Auto]: The synchronized mark will be displayed in the priority of "ECG > SpO2-1 > SpO2-2 > BP".
[ECG]: HR synchronized mark will be displayed.
[SpO2-1]/[SpO2-2]: SpO2 synchronized mark will be displayed.
[BP]: BP synchronized mark will be displayed.
[OFF]: Synchronized mark will not be displayed.
NOTE
If the corresponding BP (BP1/ART) is not measured, PR (BP) will be displayed as "---".
CAUTION
The undisplayed BP data will not generate a BP alarm or be displayed in the tabular
trend. Select the appropriate display type according to the monitoring purpose.
NOTE
The display type of numeric data can be selected from [S/D/M]/[S/D]/[M]. The BP alarm
will not be generated unless the data is displayed.
If the BP label is CVP, IAP, PAP, ICP, the display type is fixed.
NOTE
Select the appropriate filter from 6 Hz, 8 Hz, 12 Hz, 40 Hz. An artifact may interfere on
the BP waveform depending on the combination of BP measurement circuit.
7-33
Chapter 7 Monitoring BP
[ON]: When the BP label is "ART" and the catheter is disconnected, check message will be displayed.
[OFF]: ART catheter check message will not be displayed.
CAUTION
The setting is common for all BP channels. When setting is changed for BP1, the same
setting will be applied for BP2 to 8.
The default setting of "ART Catheter Check Message" is [OFF].
When "ART Catheter Check Message" is set to [ON], alarm will generate when the
transducers are opened to air.
Selecting [Auto] will measure the pulse rate from ART or BP1.
7-34
Chapter 7 Monitoring Non-Invasive Blood Pressure
The procedure of NIBP measurement and measurement condition setup are explained.
CAUTION
For the following situation, measurements will be terminated.
When the measurement time has exceeded 160 seconds for adult and child, 80 seconds
for neonate.
When the inflation value has exceeded 300 mmHg for adult, 210 mmHg for child, and 150
mmHg for neonate.
If used with the incorrect patient classification, it will not only cause erroneous
measurement, but the inflating level for the adult may be applied to child or neonate causing
dangerous situation to the patient.
The following factors may affect the NIBP value.
Body motion, arrhythmia, convulsion, low pulse pressure, slow pulse
Continuous noise such as cardiac massage
Noise from the electrosurgical instrument
Lineup of Cuffs
REFERENCE
According to the AHA (American Heart Association) guideline, the appropriate cuff width is
40% of the arm circumference.
Select the appropriate cuff from the following selections.
For other usable cuffs, refer to
"NIBP Cuffs" P13-7
NIBP Monitoring
WARNING
Before the NIBP measurement, make sure the patient classification ([Adult]/[Child]/
[Neonate]) is properly selected on the "Admit/Discharge" menu. Otherwise, correct
measurement cannot be performed, and congestion or other injury may result.
CAUTION
Correct NIBP measurement cannot be performed if oxygenator is used or if the pulse is
difficult to detect.
Pay attention when measuring the NIBP of patient with bleeding disorders or
hypercoagulation. The cuff inflation constricting the arm may cause petechia or circulatory
failure with blood clot.
Do not apply the cuff to the arm or thigh where vein is secured. The blood may backflow
causing the chemical injection to cease.
Properly arrange the cuff and air hose.
Check the condition of cuff-applied part on the patient during measurement so that the blood
circulation will not be blocked over long period of time by the squashed or bent cuff hose.
7-35
Chapter 7 Monitoring Non-Invasive Blood Pressure
Check the patient's condition constantly while measuring over a long period of time with
interval of 2.5 minutes or less. Also, periodically check the blood circulation while performing
periodic measurement over a long period of time. Congestion or rash may occur at the
measuring site.
Make sure to check the patient's condition constantly when repeatedly using continuous
measurement as it may cause dysfunction of patient's circulation.
When the cuff is not applied to the patient, pay attention not to leave the cuff unattended. If
periodic or continuous measurement is set, the cuff will automatically inflate and may cause
the rubber bag inside the cuff to burst. When not performing the NIBP measurement, set the
NIBP measurement interval OFF and disconnect the air hose from the NIBP connector.
The following factors may affect the NIBP value.
Body motion, arrhythmia, convulsion, low perfusion
Continuous noise such as cardiac massage
Periodic electromagnetic noise
If the cuff inflation may adversely affect the patient's blood flow or wound, attach the cuff to
an appropriate position under physician's instruction.
Do not apply the NIBP cuff to the arm of the mastectomized side. It may cause swelling or
other circulatory failure.
Do not perform NIBP measurement to patient who is pregnant including preeclampsia.
Accurate measurement may not be possible.
NOTE
When the [NIBP Start/Stop] key is pressed or when the NIBP measurement interval is
changed, the standby mode will be canceled and the NIBP periodic measurement will start.
7-36
Chapter 7 Monitoring Non-Invasive Blood Pressure
CAUTION
Make sure that the cuff and air hose connection is secure.
If there is any air leakage, correct NIBP measurement cannot be performed.
NOTE
The neonate cuff should be connected to air hose for neonate. Other cuffs should be
connected to air hose for general use.
The HS-8000automatically determines the patient classification (neonate or adult/child)
according to the connected air hose. If the air hose is not connected to the cuff connector,
the measurement will not start.
NOTE
Position the ARTERY mark over the artery on the patient’s arm and wrap the cuff
around.
One or two fingers should just fit in between the cuff and arm.
REFERENCE
Align the cuff height and heart position to eliminate an error caused by the blood weight.
It is most appropriate to measure with the patient lying down or comfortably seated, and
arms naturally extended.
7-37
Chapter 7 Monitoring Non-Invasive Blood Pressure
REFERENCE
About the Oscillometric Method
The oscillometric method measures the blood pressure by detecting the pulse oscillation
change by the cuff pressure. The cuff connects to the NIBP connector via the air hose. The
air pressure inside the cuff is converted to voltage by the pressure sensor, converted to
digital signal (A/D conversion), and transmitted to the CPU.
The measurement is performed with the following process.
The cuff inflates to the set value and inhibits the arterial blood flow at the measured site.
The cuff gradually deflates.
The arterial blood flow of the patient will return when the cuff pressure is decreased
sufficiently.
The oscillation (pulse signal) caused by the restricted blood circulation is transmitted to
the pressure sensor via the air hose, and converted to an electric signal.
From the pulse signal and cuff pressure detected at the pressure measurement circuit,
the systolic, diastolic, average blood pressure and pulse rate will be measured at the
CPU.
The systolic, diastolic, mean blood pressure will be displayed on the monitor. The
measurement will start with the following factor.
When the [NIBP Start/Stop] key (fixed key or user key) is pressed.
At the selected measurement interval.
For fixed amount of time after the NIBP Cont. key (user key) is pressed. (Max. 15 min.)
If "NIBP Measurement at Alarm Occurrence" is set ON, and the set parameter generates
an alarm.
When the change in patient's circulation condition is detected from the time difference of
ECG and SpO2 waveform.
7-38
Chapter 7 Monitoring Non-Invasive Blood Pressure
The maximum inflation value and measurement duration needs to be changed according to the patient classification.
The inflation mode will automatically change according to the patient classification setting. Set the appropriate
patient classification on "Admit/Discharge" menu or "Detail Setup" menu under NIBP parameter setup.
The NIBP measurement on this device is provided with forced exhaust system for safety purpose. When the
maximum inflation value is reached or when the fixed measurement duration is exceeded, the system will
automatically start to exhaust. The maximum inflation value, maximum measurement duration, initial inflation
value, measurement range, and alarm limit range for this exhaust system is set according to the patient classification
setting.
Maximum Measurement
Patient Classification Target Inflation Value Maximum Inflation Value
Duration
7-39
Chapter 7 Monitoring Non-Invasive Blood Pressure
CAUTION
When [1min] is selected, the 1-minute interval measurement will start from the time the
selection is made.
The 1-minute interval measurement will automatically stop after 12 minutes (maximum of
15 minutes when re-measured), and 2.5-minutes interval measurement will start.
The continuous mode will continuously measure for 12 minutes (maximum of 15 minutes
when re-measured). When the measurement completes, 2.5 minute interval
measurement will start.
When "Auto Mode with Start/Stop Key" is set to [ON], the auto mode measurement needs
to be started manually.
When using the continuous mode or Lumbar mode for measurement, make sure that the
setting is according to the intended purpose.
(
"About the Lumbar Mode" P7-41)
The Lumbar mode is recommended for use during spinal anesthesia. It should be used
with sufficient safety measures.
NOTE
1-minute interval measurement cannot be stopped by pressing the [NIBP Start/Stop] key
(fixed key or user key). To stop the 1-minute interval measurement, select [OFF] or other
interval on "NIBP Auto Mode" window.
When the NIBP auto mode interval is [Cont.]/[1min]/[2min]/[2.5min]/[5min]/[Lumbar
Mode], NIBP measurement cannot be started from the central monitor.
When the [NIBP Start/Stop] key is pressed or when the NIBP measurement interval is
changed, the standby mode will be canceled and the NIBP periodic measurement will
start.
REFERENCE
Select [OFF] if not performing the auto mode measurement.
The measurement starting point can be selected from [Time] (start from 0 min.) or [Meas.]
(start from actual measured time).
(
"NIBP Parameter Setup" P7-45)
measurement interval when "NIBP Start 5 min. early" is set to [ON], the measurement will
start 5 minutes before the set time. If outputting the data to PC or other external device
using the PC communication function of this system, an error may be generated to the
NIBP measurement time depending on the input interval of the external device. This
system outputs the data at completion of NIBP measurement, and if the external device
inputs the data at 60 minutes interval, 60 minutes time lag will occur. By starting the
measurement 5 minutes early, this time lag between the external device can be
minimized.
[Menu > Initial Settings > Meas. > Other]
7-40
Chapter 7 Monitoring Non-Invasive Blood Pressure
On the "Initial Settings", whether or not to backup the NIBP measurement interval at
discharge/power ON can be selected. (OFF/Backup/OFF→2.5min./OFF→5min.)
1 min. 2 min. 2 min. 2 min. 2 min. 2 min. 2 min. 2 min. 5 min. 5 min. 5 min.
If [Lumbar] is selected when the measurement is not performed, the first measurement will start.
If [Lumbar] is selected during the measurement, the current measurement will be counted as the first measurement.
The second measurement will start after 1 minute, and after 7 times of 2-minute interval measurement, the Lumbar
mode will end. The Lumbar mode can be manually stopped by selecting other interval or selecting [Lumbar] again.
When the Lumbar mode ends, 5-minute interval measurement will automatically start.
CAUTION
Pressing the [NIBP Start/Stop] key during measurement will only stop the measurement and
not the Lumbar mode. To stop the Lumbar mode, select other interval or select [Lumbar]
again.
The manual measurement can be performed in between the Lumbar mode measurement.
The Lumbar mode measurement will not start if the manual measurement is still in progress
when the next Lumbar mode measurement time arrives.
7-41
Chapter 7 Monitoring Non-Invasive Blood Pressure
When the NIBP numeric data box size is 2-box size or larger, and "Oscillograph" is set to ON on the "NIBP" setup
screen, the oscillation graph will be displayed inside the NIBP numeric data box. ( "NIBP Parameter Setup" P7-
45)
The horizontal axis shows the cuff pressure, and vertical axis shows the pulse amplitude with reference to maximum
pulse amplitude.
The bar graph shown at left indicates the size of maximum pulse amplitude compared with the reference value.
For example, if the maximum pulse amplitude is 1/2 of the reference value, the bar graph will be half Fill in.
1 Bar Graph
2 DIA Value
3 SYS Value
When the NIBP measurement result seems not match to the patient, the reliability of the NIBP measurement
can be checked by the oscillation graph.
Oscillation Graph (When NIBP is correctly measured.)
When noise interferes, the amplitude will be measured higher than actual value and the measurement value may not
be displayed correctly.
7-42
Chapter 7 Monitoring Non-Invasive Blood Pressure
When pulse is not detected, amplitude may not detected and the measurement value may not be displayed correctly.
The Dyna Alert function is a technology to prevent accidents which may occur by sudden BP change during the non-
measured duration by estimating the variation of circulatory dynamics.
This function is available for the HS-8312N, or DS-8007N with the Medtronic SpO2 module.
When [ON] is selected for "Dyna Alert", NIBP measurement will automatically start when the Dyna Alert estimated
value exceeds the alarm limit. The function will activate with the following condition.
( "Dyna Alert" P7-46)
Patient Classification: Adult (20 kg or above)
Cuff Applied Site: Upper Arm
SpO2 Sensor Attachment Site: Fingertip
NIBP Measurement Interval: 5 minutes to 60 minutes
CAUTION
When the SpO2 sensor is applied to the toe or forehead, the circulatory dynamics variation
monitoring by the Dyna Alert may not properly function.
The circulatory dynamics variation monitoring by the Dyna Alert is effective only on the HS-
8312N, or DS-8007N with the Medtronic SpO2 module.
In the NIBP numeric data box, the following mark and message indicating the status of the Dyna Alert function will
be displayed.
7-43
Chapter 7 Monitoring Non-Invasive Blood Pressure
Dyna Alert
Message Status Function
Color of Mark Status*1
Interv.: <5min. NIBP interval is set to Cont., 1min, 2min, or 2.5min. Suspended
Yellow Measure NIBP Initialization of Dyna Alert is complete, and the NIBP Suspended
measurement has not been performed since the power is
turned ON.
Poor ECG Signal ECG signal failure due to lead-off, noise, etc. Disable
Poor PTG Signal PTG (Photoplethysmograph) signal failure due to sensor off, Disable
noise, severe low perfusion, etc.
DA-NIBP Suspended Within 2.5 minutes from previous Dyna Alert NIBP Suspended
measurement.
Measuring NIBP NIBP measurement other than Dyna Alert is in progress. Disable
Initializing Waiting for stable signal after starting Dyna Alert. Disable
Green PTG Low Perfusion PTG amplitude is 200 unit or above, and below 800 unit. Enable
Suspended: Circulatory dynamics variation is monitored. But the display suspends the measurement when NIBP
measurement is requested. When the suspending factor is resolved, the measurement will resume as
quickly as possible.
Enable: Circulatory dynamics variation is monitored. The display control software responds to NIBP
measurement request as quickly as possible.
*2: "Measuring BP" indicates the status when IBP (BP1 or ART) measurement is possible and can be
displayed on the monitor.
CAUTION
When using the Dyna Alert function, be aware of these risks and do not increase the NIBP
interval time by relying only on the Dyna Alert function.
After the Dyna Alert NIBP measurement, the next Dyna Alert NIBP measurement cannot be
performed for 2.5 minutes.
The Dyna Alert will not properly function for the following cases.
If peripheral circulatory insufficiency or very low BP is developed.
If highly-frequent arrhythmia is generated.
If an oxygenator is used.
If a large noise from body movement or electric surgery device is interfering.
If autonomic nerve or circulatory dynamics is largely affected by medication.
7-44
Chapter 7 Monitoring Non-Invasive Blood Pressure
Press the [Menu], [NIBP] keys to display the "NIBP" setup screen.
2
1
3 8
4 9
5 10
6 11
7 12
13
14
15 18
16
17
1 NIBP Alarm
(
"Alarm Limit Setup for Each Parameter" P6-10)
NOTE
Set the upper limit in the range of 15 mmHg to 300 mmHg / 2.0 kPa to 40.0 kPa. If a value
above 300 mmHg / 40.0 kPa is set, the upper alarm will turn OFF.
Set the lower limit in the range of 10 mmHg to 295 mmHg / 1.5 kPa to 39.5 kPa. If a value
below 10 mmHg / 1.5 kPa is set, the lower alarm will turn OFF.
REFERENCE
Set ON/OFF of NIBP alarm, upper and lower alarm limits of systolic (S), diastolic (D),
mean (M) NIBP.
When [Auto] is set, the upper and lower limit will be automatically set to +40 mmHg / +5
kPa and -20 mmHg / -3 kPa respectively to the current value.
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under "Initial
Settings". (
Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto setting
is used, the upper and lower alarm limit will be clipped to the set limit.
The alarm limit should be set for each unit (mmHg/kPa).
The upper/lower limit can be set in 5 mmHg / 0.5 kPa increment.
7-45
Chapter 7 Monitoring Non-Invasive Blood Pressure
The patient classification setting is linked with that on the "Admit/Discharge" screen. The inflation value and
measurement duration will differ according to the patient classification setting.
(
"Inflation Mode Setup" P7-39)
WARNING
The patient classification selection influences the precision of the QRS detection and
NIBP measurement. Make sure the correct selection is made.
The NIBP air hose corresponded to the set patient classification must be used to
perform NIBP measurement. However, if the patient classification is child, NIBP air
hose for adult can be used.
4 Dyna Alert
[ON]: Dyna Alert function will turn ON when HS-8312N or DS-8007N, HG-820 is used.
CAUTION
When the PTG (SpO2) sensor is applied to the toe or forehead, the circulatory
dynamics variation monitoring by the Dyna Alert may not properly function.
The circulatory dynamics variation monitoring by the Dyna Alert is effective only on the
HS-8312N or DS-8007N, HG-820 with the NellcorTM SpO2 module.
REFERENCE
About the Dyna Alert: Using a cuff allows to measure the blood pressure
noninvasively, but on the other hand, there is a demerit of not being able to perform
the measurement continuously.Therefore, there is always a risk of sudden blood
pressure change in between the periodic measurements.
5 Oscillograph
[ON]: Oscillation graph will be displayed inside the numeric data box.
[Oscill. Print] key will be also displayed.
[Oscill. Print]: Oscillation graph will be output on the HR-800 Recorder Unit.
[OFF]: Oscillation graph will not be displayed.
7-46
Chapter 7 Monitoring Non-Invasive Blood Pressure
NOTE
The oscillation graph can be displayed when the NIBP numeric data box size is 2-box size
or larger, and "Oscillograph" is set to [ON] on the "NIBP" setup screen.
6 PR Display
NOTE
PR will be displayed only. It will not generate alarm, or be displayed for the tabular
trend.
8 Measure at Alarm
CAUTION
If the NIBP measurement has not been performed since the power was turned ON,
NIBP measurement at alarm occurrence will not be performed.
9 Quick Measurement
[ON]: NIBP measurement will be performed in duration of about 20 seconds to 25 seconds in case of adult
patient.
NOTE
The quick measurement can be performed only if the patient classification is adult or
child. For neonate, normal measurement will be performed regardless of this setting.
10 Sight Inflation
NOTE
The sight inflation function can be used only during the NIBP auto mode
measurement.
The sight inflation function cannot be used when the patient classification is "Neonate".
The sight inflation function cannot be used when performing the 1-minute interval
measurement or continuous measurement.
7-47
Chapter 7 Monitoring Non-Invasive Blood Pressure
CAUTION
If the mean BP (MAP) value is not displayed, the mean BP (MAP) alarm will not be
generated.
12 End Tone
[ON]: A buzzer tone will be generated when the NIBP measurement completes.
13 User Interval
[ON]: When the power is turned ON, NIBP auto mode will resume by starting a manual measurement
for the newly admitted patient. Until the NIBP auto mode is resumed or the interval is changed,
"Standby" will be displayed inside the NIBP numeric data box.
NOTE
If the power OFF duration was within 30 seconds, the NIBP auto mode will resume at
power ON even when the above setting is [ON].
When the [NIBP Start/Stop] key is pressed or when the NIBP measurement interval is
changed, the standby mode will be canceled and the NIBP periodic measurement will
start.
15 Time Display
7-48
Chapter 7 Monitoring Non-Invasive Blood Pressure
[Meas.]: The periodic measurement will start from the actual starting time.
Measurement time when [Time] is Measurement time when [Meas.]
selected: is selected:
For the following case, the target inflation value will be automatically set to the default value of each patient
classification.
At Discharge
When the patient classification is changed
CAUTION
When the "Sight Inflation" is [OFF], the target inflation value after the first periodic
measurement will be automatically set based on the previous measurement value.
When the "Sight Inflation" is [ON], the target inflation value after the first periodic
measurement will be set based on the sight inflation function. (
"Sight Inflation" P7-47)
18 Cancel Error
NOTE
Make sure that the NIBP measurement can be properly performed after solving the
cause of the NIBP system error message. If the message still remains, device failure
can be considered.
(
"Non-Invasive Blood Pressure" P11-35)
7-49
Chapter 7 Monitoring Pulse Oximetry
Pulse Oximetry
SpO2 Monitoring
WARNING
When measuring the SpO2 of patient with high fever or peripheral circulatory insufficiency,
check the sensor attachment periodically and change the attachment site.The temperature
of the attachment site will rise due to the sensor heat which may result in burn injury.
For the following case, accurate measurement may not be possible.
Patient with excessive abnormal hemoglobin (COHb, MetHb)
Patient with the pigment injected to the blood
Patient receiving CPR treatment
When a sensor is applied to a limb with NIBP cuff, arterial catheter, or intracatheter
When measuring at site with venous pulse
Patient with body motion
Patient with small pulse
When a patient is receiving a photodynamic therapy, measuring SpO2 on a same site for a
long duration may cause blisters from the irradiation light of the SpO2 sensor. Make sure to
periodically change the sensor attachment site.
Do not connect unspecified sensor or cable to any I/O connector. If done so by mistake, not
only that the device cannot deliver its maximum performance, the device may be damaged
and safety cannot be ensured.
If any measurement seems questionable, first check the patient's vital signs by alternate
means and then check the pulse co-oximeter for proper functioning.
For the following case, accurate measurement of SpO2 may not be possible.
Improper sensor application
Elevated levels of COHb or MetHb: High levels of COHb or MetHb may occur with a
seemingly normal SpO2. When elevated levels of COHb or MetHb are suspected,
laboratory analysis (CO-Oximetry) of a blood sample should be performed.
Elevated levels of bilirubin
Elevated levels of dyshemoglobin
Vasospastic disease, such as Raynaud's, and peripheral vascular disease
Hemoglobinopathies and synthesis disorders such as thalassemias, Hb s, Hb c, sickle
cell, etc.
Hypocapnic or hypercapnic conditions
Severe anemia
Very low arterial perfusion
Extreme motion artifact
Abnormal venous pulsation or venous constriction
Severe vasoconstriction or hypothermia
7-50
Chapter 7 Monitoring Pulse Oximetry
CAUTION
If irritation such as skin reddening appears with the sensor use, change the attachment site
or stop using the sensor.
When attaching the sensor with tape, do not wrap the tape too tight. At the same time, check
the blood flow constantly so that congestion is not generated at the peripheral site.
Even attachment for a short duration may inhibit the blood flow and generate compression
necrosis or burn injury. Also, blood flow inhibition may prevent correct measurements.
If SpO2 values indicate hypoxemia, a laboratory blood sample should be taken to confirm
the patient's condition.
If the <SpO2 Low Perfusion> message is frequently displayed, find a better perfused
monitoring site. In the interim, assess the patient and, if indicated, verify oxygenation status
through other means.
Change the application site or replace the sensor and/or patient cable when a <Replace
Sensor>, <Replace Cable>, <Low Signal IQ> is displayed on the monitor. These messages
may indicate that patient monitoring time is exhausted on the patient cable or sensor.
Replace the cable or sensor when a <Replace Sensor> or <Low Signal IQ> message is
consistently displayed while monitoring consecutive patients after completing
troubleshooting steps listed in this manual.
If using pulse oximetry during full body irradiation, keep the sensor out of the radiation field.
If the sensor is exposed to the radiation, the reading might be inaccurate or the device might
read zero for the duration of the active irradiation period.
The device must be configured to match your local power line frequency to allow for the
cancelation of noise introduced by fluorescent lights and other sources.
Variation in hemoglobin measurements may be profound and may be affected by sampling
technique as well as the patient's physiological conditions. Any results exhibiting
inconsistency with the patient's clinical status should be repeated and/or supplemented with
additional test data. Blood samples should be analyzed by laboratory devices prior to clinical
decision making to completely understand the patient's condition.
Check the sensor attachment site constantly in every 4 hours when probes or reusable
sensor are used, and at least every 8 hours when single patient use sensors are used.Be
especially careful of a patient with bad perfusion. If the sensor attachment position is not
7-51
Chapter 7 Monitoring Pulse Oximetry
changed constantly, skin irritation or skin necrosis due to compression may be developed.
For the patient with bad perfusion, check the sensor attachment position at least every 2
hours.
As skin for neonate, premature infant is immature, change the sensor attachment site more
frequently depending on the condition.
Direct sunlight to the sensor area can cause a measurement error. Place a black or dark
cloth over the sensor if using in direct sunlight.
When not measuring, unplug the relay cable and sensor from the SpO2
connector.Otherwise, the outside light may affect to falsely display measurements.
If "– – –" is displayed for the numeric data, make sure that the sensor is properly attached.
Before bathing the patient, make sure to remove the sensor and device from the patient.
7-52
Chapter 7 Monitoring Pulse Oximetry
7-53
Chapter 7 Monitoring Pulse Oximetry
personnel or device damage could occur. Return this device for servicing if necessary..
CAUTION
Do not place this device where the controls can be changed by the patient.
Electrical shock and flammability hazard: Before cleaning, always turn off the device and
disconnect from any power source.
When patients are undergoing photodynamic therapy, they may be sensitive to light
sources. Pulse oximetry may be used only under careful clinical supervision for short time
periods to minimize interference with photodynamic therapy.
Do not place this device on electrical device that may affect the operation, preventing it from
working properly.
If SpO2 values indicate hypoxemia, a laboratory blood sample should be taken to confirm
the patient’s condition.
If the <SpO2 Low Perfusion> message is frequently displayed, find a better perfused
monitoring site. In the interim, assess the patient and, if indicated, verify oxygenation status
through other means.
If using this device during full body irradiation, keep the sensor out of the radiation field. If
the sensor is exposed to the radiation, the reading might be inaccurate or the device might
read zero for the duration of the active irradiation period.
The device must be configured to match your local power line frequency to allow for the
cancelation of noise introduced by fluorescent lights and other sources.
To ensure that alarm limits are appropriate for the patient being monitored, check the limits
each time this device is used.
Variation in hemoglobin measurements may be profound and may be affected by sampling
technique as well as the patient's physiological conditions. Any results exhibiting
inconsistency with the patient’s clinical status should be repeated and/or supplemented with
additional test data. Blood samples should be analyzed by laboratory instruments prior to
clinical decision making to completely understand the patient’s condition.
Do not submerge this device in any cleaning solution or attempt to sterilize by autoclave,
irradiation, steam, gas, ethylene oxide or any other method. This will seriously damage this
device.
Electrical Shock Hazard: Carry out periodic tests to verify that leakage currents of patient-
applied circuits and the system are within acceptable limits as specified by the applicable
safety standards. The summation of leakage currents must be checked and in compliance
with IEC 60601-1 and UL 60601-1. The system leakage current must be checked when
connecting external device to the system. When an event such as a component drop or a
spillage of blood or other liquids occurs, retest before further use. Injury to personnel could
occur.
Disposal of product - Comply with local laws in the disposal of the device and/or its
accessories.
To minimize radio interference, other electrical device that emits radio frequency
transmissions should not be in close proximity to this device.
NOTE
A functional tester cannot be used to assess the SpO2 accuracy.
High-intensity extreme lights (such as pulsating strobe lights) directed on the sensor, may
not allow this device to obtain SpO2 readings.
When the pulse detection sensitivity is set to [High], performance of the "Sensor Off"
detection may be compromised. If the device is in this setting and the sensor becomes
dislodged from the patient, the potential for false readings may occur due to environmental
"noise" such as light, vibration, and excessive air movement.
7-54
Chapter 7 Monitoring Pulse Oximetry
NOTE
SpCO, SpMet, SpHb, PI, and PVI are parameters which can be measured by the Masimo
unit.
SpOC can be measured on the DS-8007M/ HS-8412M.
High-intensity extreme lights (such as pulsating strobe lights) directed on the sensor, may
not allow this device to obtain SpO2 readings.
Do not loop the patient cabling into a tight coil or wrap around the device, as this can damage
the patient cabling.
Additional information specific to the Masimo sensors compatible with this device, including
information about parameter/measurement performance during motion and low perfusion,
may be found in the sensor's directions for use.
2 Insert the sensor into the SpO2 relay cable connector, and lock it with the transparent cover.
7-55
Chapter 7 Monitoring Pulse Oximetry
CAUTION
The SpO2 patient cables (LNOP®, LNCS®, M-LNCSTM, Rainbow®) are for Masimo SET
sensor, Rainbow SET sensor only. Connect them only to the HS-8312M or DS-8007M,
HG-810. Otherwise, the device will not properly function.
NOTE
Pull the connector slowly to ensure it is securely connected.
If necessary, secure the cable to the patient.
CAUTION
If the nail is rough, dirty, or manicured, accurate measurement will not be possible.
Change the finger or clean the nail before attaching the probe and sensor.
Probe Type
1 As shown below, the probe cable should be on the nail side.
1
2
3 Press the probe lightly so that the finger and the rubber cover are appressed.
This is to stabilize the probe, and to avoid ambient light.
7-56
Chapter 7 Monitoring Pulse Oximetry
Single-Patient-Use Type
1 Clean the attachment site with alcohol, etc., and allow to dry.
2 Align the light emitting element and light receiving element of the sensor with the measuring site in between
when attaching the sensor to patient.
1 Light Emitting 1
Element
2 Light Receiving
Element
3 Secure the cable with surgical tape so that the sensor does not come off when the cable is pulled.
4 Verify that the SpO2 measurement and SpO2 waveform are displayed on the home display.
7-57
Chapter 7 Monitoring Pulse Oximetry
This section explains the SpCO, SpMet, SpHb, SpOC measurement procedure when using the HS-8312M or DS-
8007M, HG-810.
CAUTION
The SpCO, SpMet, SpHb, SpOC measurements are provided only with specific rainbow
sensors supporting specific parameter combinations. SpHb/SpOC/SpMet and SpCO/SpMet
are each valid sensor combinations which also support PVI.
SpHb/SpCO is not a valid sensor combination.
For details, contact your nearest service representative.
REFERENCE
SpCO, SpMet, SpHb, SpOC measurements are optional function.
A technology called X-Cal for patient safety and reinforcement of efficiency in a clinical site is implemented for
Masimo sensors and cables.
X-Cal is designed to address the following three common factors that can impact measurement accuracy and patient
safety due to reliability risks.
1 Imitation Masimo sensors and cables
7-58
Chapter 7 Monitoring Pulse Oximetry
This section explains the measurement procedure when using the HS-8312N or DS-8007N, HG-820.
Press the [Menu], [SpO2] keys to display the "SpO2" setup screen.
3
4
1
5
2
6
7
8
When Using the HS-8312N
2 Select the label from [None]/ [Auto]/ [RH]/ [LH]/ [RF]/ [LF]/ [OT].
7-59
Chapter 7 Monitoring Pulse Oximetry
REFERENCE
Also, when the SpO2 value is unstable around the lower alarm limit, the frequently
generated alarm can be corrected by setting the second alarm function.
(
"SpO2 Second Alarm Setup" P6-3)
When [Auto] is set, the upper limit will be turned OFF and the lower limit will be set to
90%SpO2.
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under "Initial
Settings". (
Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto setting
is used, the upper and lower alarm limit will be clipped to the set limit.
The upper/lower limit can be set in 1%SpO2 increment.
indicates the current measurement value.
The following delay occurs for the SpO2 alarm depending on the patient classification and
second alarm setting. (For Nellcor)
REFERENCE
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under "Initial
Settings". (
Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto setting
is used, the upper and lower alarm limit will be clipped to the set limit.
When [Auto] is set, the lower limit will be set to "SpO2 lower limit - 10%SpO2".
The lower limit can be set in 1%SpO2 increment.
indicates the current measurement value.
7-60
Chapter 7 Monitoring Pulse Oximetry
The following delay occurs for Ext SpO 2 alarm depending on the patient classification
and second alarm setting. (For Nellcor)
Patient Classification
Adult/Child Neonate
SpO2 Alarm Condition Delay About 7 sec. to 9 sec. About 7 sec. to 9 sec.
REFERENCE
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under "Initial
Settings". (
Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto setting
is used, the upper and lower alarm limit will be clipped to the set limit.
When [Auto] is set, the upper and lower limit will be automatically set to +40 bpm and -40
bpm to the current value respectively.
The upper and lower limit can be set in 5 bpm increments.
It can be set in 1 bpm increment for 25 bpm and below.
The following delay occurs for the PR alarm depending on the patient classification. (For
Nellcor)
PR Alarm Condition Delay: <Adult/Child/Neonate> About 5 sec. to 6 sec.
PR Alarm Signal Delay: <Adult/Child> About 5 sec., <Neonate> 0 sec.
NOTE
During the NIBP measurement, the cuff inflation restricts the blood flow which disables
the correct detection of the SpO2 and PR, and may generate an improper alarm.
Selecting [OFF] for "Alarm during NIBP" will not generate the SpO2, Ext SpO2, PR, SpCO
(Masimo only), SpMet (Masimo only), SpHb (Masimo only) alarm until the NIBP
measurement is complete.
REFERENCE
This setup can be used when the SpO2 sensor and the NIBP cuff is placed on the same
limb for measurement.
7-61
Chapter 7 Monitoring Pulse Oximetry
REFERENCE
When SpO2 sensor is attached to the patient with the SpO2 display set to OFF, and SpO2
is measured for 10 seconds, the pulse wave and numeric data will be automatically
displayed.
This section explains the RR_SpO2 measurement procedure when using the DS-8007N.
CAUTION
The RR_SpO2 can be measured only when using the Medtronic Respiratory Sensor.
For details, contact your nearest service representative.
7-62
Chapter 7 Monitoring Pulse Oximetry
This section explains the measurement procedure when using the HS-8312M or DS-8007M, HG-810. Press the
[Menu], [SpO2] keys to display the "SpO2" setup screen.
REFERENCE
This setting is available when using the HS-8312M or DS-8007M, HG-810. PVI, SpCO,
SpMet, SpHb, SpOC measurements are an optional function.
1
2 5
6 4 10
7 3 11
8 12 13
9
Patient Classification
SpO2 Averaging
Adult/Child Neonate
SpO2 Alarm Condition Delay For all settings About 7 sec. to 9 sec. About 7 sec. to 9 sec.
SpO2 Alarm Signal Delay For all settings About 5 sec. 0 sec.
7-63
Chapter 7 Monitoring Pulse Oximetry
WARNING
Be careful when setting the "SpO2 Averaging" duration as the SpO2 alarm is based on
the displayed SpO2 value which is averaged from the duration set in "SpO2 Averaging".
The alarm occurrence time will be affected or may not occur for the transient value of
SpO2 depending on the set duration.
CAUTION
If [High] is selected for pulse sensitivity, probe-off detection will become somewhat
inaccurate.
NOTE
To improve the low perfusion condition, or to perform fast tracking when the SpO2
value changes abruptly, select [High].
For standard use, select [Normal].
If there is a high possibility of sensor getting disconnected, select [APOD].
When the pulse detection sensitivity is set to [High], performance of the "Sensor Off"
detection may be compromised. If the device is in this setting and the sensor becomes
dislodged from the patient, the potential for false readings may occur due to
environmental "noise" such as light, vibration, and excessive air movement.
NOTE
To pick up the abrupt change of the value sooner, and to take advantage of the qualities
of FAST SAT mode, SpO2 averaging time will be fixed as [2- 4 sec.] when FAST SAT is
set ON.
NOTE
The perfusion index is calculated by pulsatile signal divided by apulsatile signal times
100, and indicates patient’s circulation condition at the monitoring site.
7-64
Chapter 7 Monitoring Pulse Oximetry
The perfusion index assists clinicians in determining optimal placement of the SpO2
sensor. This parameter is also useful as a troubleshooting tool by helping a clinician rule
out whether a questionable value may be due to low perfusion and/or a low signal to noise
condition. Also, it can be used as diagnosis index to predict the patient’s critical condition
when at low perfusion.
REFERENCE
Perfusion Index (PI) is a relative assessment of the pulse strength at the monitoring site.
It is a ratio of the pulsatile and the non-pulsatile blood flow at the monitoring site. It can
be used to find the most appropriate sensor application site by finding the site with the
highest PI. Perfusion Index (PI) is displayed in the range from 0.02% to 20%.
Pleth Variability Index (PVI) is an index of the change in PI that occurs during the
respiratory cycle. It is calculated by measuring the changes in PI over a time interval
where one or more complete respiratory cycles have occurred. Pleth Variability Index
(PVI) is displayed in the range from 0% to 100%.
NOTE
The signal IQ wave cannot be printed.
REFERENCE
The signal IQ wave indicates the signal confidence and pulse beat. The vertical length
indicates the signal confidence. A low vertical line indicates a lower signal confidence.
7-65
Chapter 7 Monitoring Pulse Oximetry
15 16
14
CAUTION
Set the upper limit in the numeric range of 1% to 40% SpCO. If a value above 40% SpCO
is set, the upper alarm will turn OFF.
The lower limit cannot be set.
The automatic alarm cannot be set.
CAUTION
Set the upper limit in the range of 1% to 15%SpMet. The upper limit alarm will turn OFF
if a value above 15%SpMet is set.
The lower limit cannot be set.
The automatic alarm cannot be set.
16 Set the SpHb measurement condition. Press the [SpHb] key to display the SpHb setup screen.
1
2
CAUTION
Set the upper limit in between 2.0 g/dL to 24.5 g/dL. The upper limit alarm will turn OFF
if the value above 24.5 g/dL is set.
7-66
Chapter 7 Monitoring Temperature
Set the lower limit in between 1.0 g/dL to 24.0 g/dL. The lower limit alarm will turn OFF if
the value below 1.0 g/dL is set.
The automatic alarm cannot be set.
Temperature
This section explains the measurement procedure and measurement condition setup of temperature (T1 to T8).
TEMP Monitoring
CAUTION
Before the measurement, make sure that the specified probe/relay cable is used.
If unspecified probe/relay cable is used, measurement error may occur.
Stop using the probe if it is damaged..
NOTE
700 series temperature probe cannot be used.
REFERENCE
The HS-8000 or module utilizes multiparameter amplifier input method which allows
monitoring of 2 channels of temperature through the 2ch temperature relay cable (CJO-
P01T-DA**) connected to the HS-8000 or module connector.
The measurement is also possible using the HM-800/HM-801 Multi Module inserted to
the input box.
1 Connect the 2ch temperature relay cable (CJO-P01T-DA**) to the multiparameter connector on the HS-
8000.
2 Connect the temperature probe to the 2ch temperature relay cable.
7-67
Chapter 7 Monitoring Temperature
NOTE
The probe location shown above is an example. Adjust the probe location according
to the patient's condition.
7-68
Chapter 7 Monitoring Temperature
%3 7(03 %3
%3 %3
When the DS-8007 is used, the temperature measured on temperature jack is fixed as T1, T2.
NOTE
The channel settings can be changed on the "Multi-amplifier" screen ([Initial Settings >
System > Unit Module]).
(
Maintenance Manual "Unit Module Setup" P4-19)
Parameter Setup
Press the [Menu], [TEMP] keys to display the "TEMP" setup screen.
1 4
2
1 TEMP Label
REFERENCE
Description of Each Label:
T1-T8 (Default)
Tsk (Skin Temperature)
Tre (Rectal Temperature)
Tes (Esophageal Temperature)
Tco (Core Temperature))
US1 to US7: User labels (3 characters) which can be set on the "Initial Settings".
(
Maintenance Manual "User Label Setup" P5-9)
7-69
Chapter 7 Monitoring Temperature
NOTE
US3 to US7 cannot be selected for the device connected to DS-LANII/III.
2 Temperature Alarm
(
"Alarm Limit Setup for Each Parameter" P6-10)
NOTE
Set the upper limit in the range of 31.0°C to 45.0°C. If a value above 45.0°C is set, the
upper alarm will turn OFF.
Set the lower limit in the range of 30.0°C to 44.0°C. If a value below 30.0°C is set, the
lower alarm will turn OFF.
REFERENCE
The upper and lower limit can be set in 0.5°C increments.
When [Auto] is set, the upper and lower limit will be automatically set to +2.0 °C and -
2 °C to the current value respectively.
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under
"Initial Settings". ( Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto
setting is used, the upper and lower alarm limit will be clipped to the set limit.
3 Display ON/OFF
(
"ECG Parameter Setup" P7-6)
CAUTION
When the parameter display is set to OFF, the alarm generation and tabular/graphic
trend input will also cease.
4 ΔT Display
NOTE
To display on the home display, the setup on the "Display Config." is necessary.
(
"To Configure the Display" P10-4)
The alarm cannot be set for ΔT.
7-70
Chapter 7 Monitoring Cardiac Output and Blood Temperature
When thermodilution catheter is used to measure the cardiac output, the blood temperature (Tb) can be monitored.
The CO measurement can be performed using the multiparameter connector on the HS-8000 or module.
The measurement is also possible using the HM-800/HM-801 Multi Module inserted to the input box.
( "Cardiac Output (CO)" P8-45)
NOTE
The usable catheter relay cable depends on the injectate temperature measurement
method.Select the appropriate cable according to the used measurement method.
2 Connect the catheter relay cable to the multiparameter connector on the HS-8000/DS-8007 or HM-800/HM-
801 Multi Module, and connect the catheter to the catheter relay cable.
7-71
Chapter 7 Monitoring Cardiac Output and Blood Temperature
Thermodilution Method
The thermodilution catheter is inserted from the vein through the right atrium, right ventricle, and pulmonary artery.
From the side hole near the catheter tip, injectate is injected quickly to the right atrium. At this time, the heart
contraction and heat diffusion mixes the injectate with blood, and causes blood temperature fall.
Variable initiated by these effects are measured as time function at the pulmonary artery, and the following
thermodilution curve can be drawn.
Cardiac output is calculated by applying this to the Stewart-Hamilton formula shown below.
60 : seconds
7-72
Chapter 7 Monitoring Cardiac Output and Blood Temperature
As shown above, cardiac output is directly proportional to the Injectate Volume (Vi) and the difference between
Blood Temperature and Injectate Temperature (Tb - Ti), and is inversely proportional to the area of the
thermodilution curve (S).
Hematocrit Value
Hematocrit value of 45%, (Si*Ci)/(Sb*Cb) = 1.08 is programmed for this device.
NOTE
If the hematocrit value is different, an error may be caused in cardiac output measurement.
2 Select ON/OFF of blood temperature alarm and set the upper and lower alarm limits.
(
"Alarm Limit Setup for Each Parameter" P6-10)
NOTE
Set the upper limit in the range of 31.0°C to 45.0°C. If a value above 45.0°C is set, the
upper alarm will turn OFF.
Set the lower limit in the range of 30.0°C to 44.0°C. If a value below 30.0°C is set, the
lower alarm will turn OFF.
REFERENCE
The upper and lower limit can be set in 0.5°C increments.
When [Auto] is set, the upper and lower limit will be automatically set to +2.0 °C and -2
°C to the current value respectively.
[Auto] key will be displayed only when [Enable] is set for "Auto Alarm Setup" under "Initial
Settings".
7-73
Chapter 7 Monitoring CO2 Concentration (Mainstream Method)
This section explains about the CO2 concentration measurement procedure and measurement condition setup when
using the Philips Capnostat 5 (Mainstream Method, Gas Unit I/F HPD-810/HPD-820) .
CAUTION
When the multigas unit (MGU-800/MGU-810) and HPD-810/HPD-820 are simultaneously
used, the CO2 concentration measurement will be performed by the equipment selected for
the "CO2 Source Priority" under ([Menu] > "Parameter" [CO2]).
By using the HPD-810/HPD-820 CO2 Gas Unit I/F, CO2 measurement by the Philips Capnostat 5 (Mainstream
Method) can be performed.
2 Connect the CO2 sensor (Capnostat5) to the CO2 connector on the HPD-810/HPD-820.
The CO2 sensor will automatically begin warming up. The CO2 sensor requires a warming up process to
achieve stable operating temperature. Warm up process will require minimum of 2 minutes.
During the warm up period, <CO2 Warm Up> message will be displayed on the monitor.
When the warm up completes, the message will disappear.
NOTE
When using more than one HM-801, only one AUX connector on the HM-801 can be used
for measurement of CO2, BIS (with HBX-800).
When the CO2 is measured on both the HS-8000/DS-8007 and HM-801, the
measurement of the HS-8000/DS-8007 will be prioritized.
7-74
Chapter 7 Monitoring CO2 Concentration (Mainstream Method)
CAUTION
The disposable airway adapter should be opened just before use.
Do not reuse the disposable airway adapter. Do not disassemble, clean, disinfect, or
sterilize it.
NOTE
There are 4 types of airway adapters. Select the appropriate adapter according to the
used endo-tracheal tube size and operating environment.
4 Verify that the warm up is complete, and attach the CO2 sensor to the airway adapter until a click sound is
heard.
2 Window 1
3 Airway Adapter
A: Thick Side 3
B: Thin Side
B
2
A
CAUTION
The airway adapter should be attached with the thicker side facing to the patient. If
attached oppositely, it may damage the CO2 sensor or airway adapter.
5 Perform the setting for the O2 compensation, N2O compensation, anesthetic gas compensation, atmospheric
pressure Set these items each time the condition changes. (
"CO2 Parameter Setup" P7-77)
6 Press the [Menu], [CO2] ("Parameter"), [Calibrate Airway Adapter] keys to calibrate the airway adapter.
7-75
Chapter 7 Monitoring CO2 Concentration (Mainstream Method)
NOTE
The airway adapter calibration must be performed before connecting to the respiration
circuit.
The airway adapter calibration should be also performed for the following case.
When the airway adapter is replaced.
When <Zero the CO2 Adapter> or <Check airway adapter.> is displayed.
A clean airway adapter must be used.
If reusing an airway adapter, clean and air-dry it. Then, wipe the window with a swab, and
sterilize (EOG, etc.) before use.
During the calibration, the measurement data will be displayed as "---". The measurement
data during calibration may be included in the trend data causing discontinuity.
Calibration cannot be performed if respiration is detected within 20 seconds before
calibration. In such case, wait for 20 seconds and perform the calibration again.
When <Cal. error> is displayed, perform the airway adapter calibration again.
The EtCO2 accuracy is checked according to the test method of ISO 80601- 2-55: 2011
(Medical electrical equipment-Part 2-55: Particular requirements for the basic safety and
essential performance of respiratory gas monitors) .
The waveform sampling rate is 100 Hz.
The CO2 measurement accuracy is not guaranteed for all humidity levels (non-
condensing).
The CO2 measurement accuracy is tested at 35°C.
The respiration rate accuracy was verified by using a solenoid test setup to deliver a
square wave of known CO2 concentration to the device. 5% and 10% CO2 concentrations
were used and respiration rate was varied over the range of the device. Pass/Fail criteria
was comparison of the respiratory rate output from the sensor to the frequency of the
square wave. EtCO2 measurements at those rates were compared to the CO2 readings
under static flow conditions.
7 Verify that the airway adapter calibration is properly completed, disconnect the CO2 sensor from the airway
adapter temporarily, and attach the airway adapter to the patient’s respiration circuit.
2 Y-Piece 1
A: Patient Side
B: Device Side
3
A
7-76
Chapter 7 Monitoring CO2 Concentration (Mainstream Method)
NOTE
Attach the airway adapter between the patient's circuit Y-piece and intubation tube.
The CO2 sensor should be facing upward.
9 Verify that the CO2 waveform, EtCO2 value, InspCO2 value are displayed.
NOTE
Set the scale, measurement unit, alarm, etc. as necessary.
Press the [Menu], [CO2] keys to display the "CO2" setup screen.
4
1
3
2
5 8
6 9
7 10 11
Select from [0- 50]/[0- 100] if the measurement unit is mmHg, and from [0- 4]/[0- 8]/[0- 10] if the unit is kPa or %.
(
"Alarm Limit Setup for Each Parameter" P6-10)
NOTE
The EtCO2 alarm will not generate unless 2 or more respiration is detected within 30
seconds after the power is turned ON, monitoring is resumed, CO2 unit is connected,
or a patient is discharged.
7-77
Chapter 7 Monitoring CO2 Concentration (Mainstream Method)
Set the upper limit in the range of 3 mmHg to 100 mmHg/0.3 kPa to 13.3 kPa/0.3% to
13.3%.
Setting a value above 100 mmHg/3.3 kPa/13.3% will turn OFF the alarm.
Set the lower limit in the range of 1 mmHg to 98 mmHg/0.1 kPa to 13.1 kPa/0.1% to
13.1%.
Setting a value below 1 mmHg/0.1 kPa/0.1% will turn OFF the alarm.
When Capnostat 5 is used, EtCO2 alarm will not generate unless 2 or more respiration
is detected after completion of the airway adapter calibration.
REFERENCE
The alarm limit should be set for each unit (mmHg/kPa/%).
The upper/lower limit can be set in 1 mmHg/0.1 kPa/0.1% increment.
When [Auto] is set, the upper and lower limit will be automatically set to +10 mmHg /
+1.3 kPa / +1.3%, and -10 mmHg / -1.3 kPa / -1.3% respectively to the current value.
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under
"Initial Settings". ( Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto
setting is used, the upper and lower alarm limit will be clipped to the set limit.
NOTE
The InspCO2 alarm will not generate unless 2 or more respiration is detected within 30
seconds after the power is turned ON, monitoring is resumed, CO2 unit is connected,
or a patient is discharged.
Set the upper limit in the range of 1 mmHg to 4 mmHg/0.1 kPa to 0.4 kPa/0.1% to
0.4%.
Setting a value equal to or above 4 mmHg/0.4 kPa/0.4% will turn OFF the alarm.
When Capnostat 5 is used, InspCO2 alarm will not generate unless 2 or more
respiration is detected after completion of the airway adapter calibration.
REFERENCE
The alarm limit should be set for each unit (mmHg/kPa/%).
The upper limit can be set in 1 mmHg/0.1 kPa/0.1% increments. There is no lower limit.
When [Auto] is set, the upper limit will be set to 3 mmHg / 0.4 kPa / 0.4%.
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under
"Initial Settings". ( Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto
setting is used, the upper and lower alarm limit will be clipped to the set limit.
7-78
Chapter 7 Monitoring CO2 Concentration (Mainstream Method)
[10sec]/[20sec]: Maximum EtCO2 value for the selected duration will be displayed.
[OFF]: EtCO2 value for each respiration will be displayed.
NOTE
As the EtCO2 value display is updated each second, EtCO2 value for each respiration
cannot be displayed if respiration rate is 60 Bpm and above.
For the InspCO2 value, minimum value of 20 seconds will be displayed regardless of
the setting.
6 O2 Compensation
By entering the used O2 concentration value, compensation can be made to display more accurate value.
Enter the O2 compensation value on the "O2" screen, and press the [Set] key.
NOTE
The value cannot be changed if the total value of O2 compensation and anesthetic
agent compensation exceeds 100%. In such case, change the O2 compensation value
after changing the anesthetic agent compensation value.
7 N2O Compensation
NOTE
If N2O is present in the respiration circuit, the CO2 value tends to be displayed higher
than the actual value. By setting the N2O compensation to [ON], this can be adjusted.
By entering the used anesthetic agent concentration value, compensation can be made to display more
accurate value.
Enter the anesthetic compensation value on the "Agent" screen, and press the [Set] key.
NOTE
The value cannot be changed if the total value of O2 compensation and anesthetic
agent compensation exceeds 100%. In such case, change the anesthetic agent
compensation value after changing the O2 compensation value.
7-79
Chapter 7 Monitoring CO2 Concentration (Mainstream Method)
9 Atmospheric Pressure
By entering the atmospheric pressure, the pressure difference will be compensated and allows more accurate
measurement.
Enter the atmospheric pressure value on the "Atmos. Pressure" screen, and press the [Set] key.
[MGU-800]: CO2 value measured by the MGU-800/810 Multigas Unit will be prioritized.
[HS-8000]: CO2 value measured by the HS-8000/DS-8007/HM-801 will be prioritized.
NOTE
When the HS-8000/DS-8007 and HM-801 are simultaneously used, the CO2
measurement of the HS-8000/DS-8007 will be prioritized.
11 Display ON/OFF
(
"ECG Parameter Setup" P7-6)
CAUTION
When the waveform and numeric data display is set to OFF, the alarm generation and
tabular trend input will also cease.
When the waveform and numeric data display is set to OFF, the respiration rate
measured by CO2 will not be displayed either.
REFERENCE
When the filter line is applied to the patient during the "Display OFF" condition, the
waveform and numeric data will be automatically displayed when 2 or more
respirations are detected in 30 seconds.
7-80
Chapter 7 Monitoring CO2 Concentration (Sidestream Method)
The HCP-810/HCP-820 is a CO2 Gas Unit which measures CO2 concentration. The HCP-810/HCP-820 CO2 Gas
Unit incorporates Microstream technology of Covidien for EtCO2 (End-tidal CO2 concentration) and InspCO2
(Inspiratory CO2 concentration) measurement. This section explains about the procedure and setup of the CO2
concentration measurement of the HCP-810/HCP-820.
WARNING
When using a sampling line for intubated patients with a closed suction system, do not place
the airway adapter between the suction catheter and endotracheal tube. This is to ensure
that the airway adapter does not interfere with the functioning of the suction catheter.
Loose or damaged connections may compromise ventilation or cause an inaccurate
measurement of respiratory gases. Securely connect all components and check
connections for leaks according to standard clinical procedures.
Do not cut or remove any part of the sampling line. It could lead to erroneous readings.
If too much moisture enters the sampling line (i.e., from ambient humidity or breathing of
unusually humid air), <Check Sample Line> will appear in the message area. Replace the
sampling line when this message appears.
Carefully route the filter line to reduce the possibility of patient entanglement or
strangulation.
Do not lift the HCP-810/HCP-820 by the filter line, as the filter line could disconnect from the
device, causing the device to fall on the patient.
CO2 readings and respiratory rate can be affected by sensor application errors, certain
ambient environmental conditions, and certain patient conditions.
CAUTION
When the multigas unit (MGU-800/MGU-810) and HPD-810/HPD-820 are simultaneously
used, the CO2 concentration measurement will be performed by the device selected for the
"CO2 Source Priority" under ([Menu] > "Parameter" [CO2]).
The Microstream EtCO2 sampling lines are designed for single patient use, and are not to
be reused. Do not attempt to clean, disinfect, sterilize or flush any part of the sampling line
as this can cause damage to the monitor.
Dispose of sampling lines according to standard operating procedures or local regulations
for the disposal of contaminated medical waste.
Before use, carefully read the Directions for Use for the Microstream EtCO2 sampling line.
Use only the Microstream EtCO2 sampling line to ensure proper function of the monitor.
During nebulization or suction for intubated patient, remove the sampling line from the HCP-
810/HCP-820 to avoid moisture buildup and sampling line occlusion.
Replace the sampling line according to hospital protocol or when a blockage is indicated on
the device. Excessive patient secretions or a buildup of liquids in the airway tube may
occlude the sampling line, requiring more frequent replacement.
NOTE
When connecting a sampling line to the HCP-810/HCP-820, screw the sampling line
clockwise into the connector firmly to avoid inaccurate measurement which may be caused
by gas leak from the connection point.
When <Check Sample Line> appears on the screen indicating that the filter line connected
to the HCP-810/HCP-820 is blocked, the CO2 pump will stop pumping the patient's breath
to the monitor. In such case, follow the instructions in the "Troubleshooting" section of this
7-81
Chapter 7 Monitoring CO2 Concentration (Sidestream Method)
manual. First, disconnect and reconnect the filter line. If the message still appears,
disconnect and replace the filter line. Once a working filter line is attached, the pump will
automatically resume operation.
After connecting the CO2 sampling line to the HCP-810/HCP-820 and patient, check that
CO2 values appear on the monitor display.
The EtCO2 accuracy is checked according to the test method of ISO 80601- 2-55: 2011
(Medical electrical equipment 2-Part 2-55: Particular requirements for the basic safety and
essential performance of respiratory gas monitors) .
The waveform sampling rate is 20 samples per second.
The minimum value and maximum value of the CO2 waveform are used for the InspCO2
value and EtCO2 value respectively.
When using the HCP module with a ventilator, under high over pressures close to 10 kPa
(100 cmH2O), the module may enter into a blockage mode in order to protect the module
from damage.
The respiration rate test simulates breaths for use in respiration rate measurement with a
system which uses a tank of N2 (representing no CO2 for inhalation) and a tank of CO2 (of
the %CO2 required for the particular test). A control board, which is triggered by a computer,
uses solenoids to switch the module input between the 2 tanks of gas, creating a gas CO2
square wave. This system can create simulated breaths over the full required range of
specified respiration rates.
The CO2 concentration can be measured by using the HCP-810/HCP-820 CO2 Gas Unit.
NOTE
Accurate CO2 concentration measurement can be acquired after 40 seconds from turning
the power ON.
NOTE
When using more than one HM-801, only one AUX connector on the HM-801 can be used
for measurement of CO2, BIS (with HBX-800).
When the CO2 is measured on both the HS-8000/DS-8007 and HM-801, the
measurement of the HS-8000/DS-8007 will be prioritized.
7-82
Chapter 7 Monitoring CO2 Concentration (Sidestream Method)
2 Attach the airway adapter, oral/nasal sampling line or nasal sampling line to the patient.
1 Attach the nasal or oral/nasal patient interface of the sampling line to the patient as described in the
sampling line directions for use.
2 Connect the sampling line to the connector on the HCP-810/HCP-820.
Verify that all the tubes are properly connected.
Verify that the CO2 waveform, EtCO2 value, InspCO2 value are displayed.
CAUTION
If the power supply is interrupted due to power failure, etc., HCP-810/HCP-820 will be
initialized even if the power interruption was within 30 seconds.
7-83
Chapter 7 Monitoring CO2 Concentration (Sidestream Method)
NOTE
Connecting a sampling line or nasal prong to the HCP-810/HCP-820 will automatically
start the sampling pump. To prevent the pump from deteriorating, disconnect the
sampling line and nasal prong from the HCP-810/HCP-820 when not measuring the CO2
concentration.
Set the scale, measurement unit, alarm, etc. as necessary.
When ambient temperature or atmospheric pressure changes significantly, auto zeroing
will function. During auto zeroing, "---" will be displayed inside the CO2 numeric data box
and CO2 measurement cannot be performed.
Press the [Menu], [CO2] keys to display the "CO2" setup screen.
1 3
4 2
5
6 7
8
1 Scale
Select from [0- 50]/[0- 100] if the measurement unit is mmHg, and from [0- 4]/[0- 8]/[0- 10] if the unit is kPa or %.
(
"Alarm Limit Setup for Each Parameter" P6-10)
NOTE
The EtCO2 alarm will not generate unless 2 or more respiration is detected within 30
seconds after the power is turned ON, monitoring is resumed, CO2 unit is connected,
or a patient is discharged.
Set the upper limit in the range of 3 mmHg to 100 mmHg/0.3 kPa to 13.3 kPa/0.3% to
13.3%.
Setting a value above 100 mmHg/3.3 kPa/13.3% will turn OFF the alarm.
Set the lower limit in the range of 1 mmHg to 98 mmHg/0.1 kPa to 13.1 kPa/0.1% to
13.1%.
Setting a value below 1 mmHg/0.1 kPa/0.1% will turn OFF the alarm.
REFERENCE
The alarm limit should be set for each unit (mmHg/kPa/%).
The upper/lower limit can be set in 1 mmHg/0.1 kPa/0.1% increment.
When [Auto] is set, the upper and lower limit will be automatically set to +10 mmHg /
+1.3 kPa / +1.3%, and -10 mmHg / -1.3 kPa / -1.3% respectively to the current value.
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under
"Initial Settings". ( Maintenance Manual "Alarm Related Setup" P5-4)
7-84
Chapter 7 Monitoring CO2 Concentration (Sidestream Method)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto
setting is used, the upper and lower alarm limit will be clipped to the set limit.
(
"Alarm Limit Setup for Each Parameter" P6-10)
NOTE
The InspCO2 alarm will not generate unless 2 or more respiration is detected within 30
seconds after the power is turned ON, monitoring is resumed, CO2 unit is connected,
or a patient is discharged.
Set the upper limit in the range of 1 mmHg to 4 mmHg/0.1 kPa to 0.4 kPa/0.1% to
0.4%.
Setting a value equal to or above 4 mmHg/0.4 kPa/0.4% will turn OFF the alarm.
REFERENCE
The alarm limit should be set for each unit (mmHg/kPa/%).
The upper limit can be set in 1 mmHg/0.1 kPa/0.1% increments. There is no lower limit.
When [Auto] is set, the upper limit will be set to 3 mmHg / 0.4 kPa / 0.4%.
To use the auto setting, [Enable] should be selected for "Alarm Auto Setup" under
"Initial Settings". ( Maintenance Manual "Alarm Related Setup" P5-4)
When [Enable] is set for "Alarm Threshold Limit" under "Initial Settings", and auto
setting is used, the upper and lower alarm limit will be clipped to the set limit.
[10sec]/[20sec]: Maximum EtCO2 value, minimum InspCO2 value for the selected duration will be displayed.
[OFF]: EtCO2 value, InspCO2 value for each respiration will be displayed.
NOTE
As the EtCO2 value display is updated each second, EtCO2 value for each respiration
cannot be displayed if respiration rate is 60 Bpm and above.
[MGU-800]: CO2 value measured by the MGU-800/810 Multigas Unit will be prioritized.
[HS-8000]: CO2 value measured by the HS-8000/DS-8007/HM-801 will be prioritized.
NOTE
When the HS-8000/DS-8007 and HM-801 are simultaneously used, the CO2
measurement of the HS-8000/DS-8007 will be prioritized.
6 CO2 Calibration
7-85
Chapter 7 Monitoring Multigas Unit/SPIRO
7 Suspend CO2
[Suspend]: The pump operation will stop, CO2 waveform and numeric data display will disappear, and
"Suspended" will be displayed inside the CO2 numeric data box.
[Resume]: Resumes CO2 monitoring. This key will be displayed when the measurement is suspended.
CAUTION
When the measurement is suspended, the alarm generation and trend input will be also
suspended.
8 Display ON/OFF
(
"ECG Parameter Setup" P7-6)
CAUTION
When the waveform and numeric data display is set to OFF, the alarm generation and
tabular trend input will also cease.
When the waveform and numeric data display is set to OFF, the respiration rate
measured by CO2 will not be displayed either.
REFERENCE
When the filter line is applied to the patient during the "Display OFF" condition, the
waveform and numeric data will be automatically displayed when 2 or more
respirations are detected in 30 seconds.
Multigas Unit/SPIRO
The MGU-800/810 series Multigas Unit can be connected to the DS-8400 system via U-LINK connector.
( Maintenance Manual "Connection of Multigas Unit" P1-11)
When the multigas unit is connected, monitoring conditions for CO2 concentration, anesthetic gas concentration, O2
concentration, and N2O concentration, respiration (SPIRO) can be set.
The MGU-800/810 series have an internal barometer and thermistor that allow compensation for changes over a
range of temperature and atmospheric pressures.
WARNING
Make sure to use only the specified Mindray Medical Sweden AB product.
(
"AG Sensors for MGU-800/MGU-810 Manufacturer: Mindray" P13-11)
Be careful not to damage the water trap during operation as bacteria and/or mucus may
contaminate the MGU-800/810 series.
The airway adapter, sampling line, flow sensor are disposable products that are intended for
single patient use only. Do not reuse them on other patients as it may cause cross-infection.
Do not use the MGU-800/810 series with the flammable anesthetic agents.
To protect the hospital staffs from unnecessary anesthetic agent, it is strongly
recommended to connect the exhaust hole to the gas exhaust system in the hospital.
The sampling line may get clogged by internal condensation.
CAUTION
When the multigas unit (MGU-800/MGU-810 series) and HPD-810/HPD-820, HCP-810/
HCP-820 are simultaneously used, the CO2 concentration measurement will be performed
by the device selected for the "CO2 Source Priority" under ( [Menu] > "Parameter" [CO2]) .
7-86
Chapter 7 Monitoring Multigas Unit/SPIRO
NOTE
The MGU-800/810 series uses a fixed correction of 11hPa (22ºC@40% RH) to compensate
for the influence of water vapor in the gas sample, when converting the gas readings to
ATPD. An increase in the ambient H20 partial pressure to 30 hPa (28ºC@80% RH or
33ºC@60% RH) will cause a general error for all gases of only -2% REL.
The EtCO2 accuracy is checked according to the test method of ISO 80601-2-55: 2011
(Medical electrical equipment-Part 2-55: Particular requirements for the basic safety and
essential performance of respiratory gas monitors) under the condition of 60 Bpm and below
with I:E ratio of 1:1.
When the RR exceeds 60 Bpm, the EtCO2 accuracy cannot be specified. (Depends on the
I:E ratio.)
The data sampling rate is 25 Hz.
The minimum value and maximum value of the CO2 waveform are used for the InspCO2
value and EtCO2 value respectively.
For the gas measurement data, "0" will be displayed if the value becomes below the
following threshold for 3 seconds or more. (Full Accuracy/during warm-up)
CO2: 0.1/0.3[vol%]
N2O: 3/3[vol%]
O2: 0/0[vol%]
Volatile Anesthetic: 0.15/0.3[vol%]
The ability to properly resolve end-tidal values can be measured by using the set-up
described in ISO 80601-2-55: 2011, figure 201. 101.
In short, the method consists of sampling gas from two different sources connected to an
electrically controlled pneumatic valve to permit rapid switching between the two sources.
During the test, the valve is set to switch gas source at a number of frequencies (simulating
the range of specified breath rates) and for each frequency, the end-tidal value presented
by the gas analyzer is noted. From a diagram of end-tidal value over frequency, the
frequency at which the gas analyzer is no longer able to resolve end-tidal values according
to specification is identified.
7-87
Chapter 7 Monitoring Multigas Unit/SPIRO
NOTE
If [Adult] or [Child] is selected as patient classification on the "Admit/Discharge" screen,
install the DRYLINE Adult/Child Water Trap (60-13100-0).
If [Neonate] is selected as patient classification on the "Admit/Discharge" screen, install the
DRYLINE Neonatal Water Trap (60- 13200-0).
If the used water trap and the set patient classification does not match, <GAS Check Water
Trap Class> will be displayed.
1 Install the DRYLINE Water Trap (Adult/Child: 60-13100-00, Neonate: 60-13200-00), DRYLINE II Water Trap
(Adult/Child: 100-000080-00, Neonate: 100-000081-00) aligning the lugs with the corresponding holes in the
receptacle and pushing gently into place. (See below.)
Make sure that both barbs on the lugs are fully engaged by pulling the water trap, which should be firmly
seated.
2 Connect the DRYLINE Airway Adapter (Straight: 60-14100-00, or Elbow: 60-14200-00) to the patient breathing
system.
3 Remove the protective cap from the airway adapter and connect it to the sampling line (for adult/child: 60-
15200-00, for neonate: 60- 15300-00).
4 Connect the other side of the sampling line to the inhale port of the water trap. When the water trap is half full,
empty the water trap's reservoir.
(
Maintenance Manual "Water Trap (Multigas Unit)" P8-5)
WARNING
The contents of the water trap should be handled as a potential infection hazard.
7-88
Chapter 7 Monitoring Multigas Unit/SPIRO
Patient
Classification
Patient Selection on
SPIRIT Flow Sensor DRYLINE Water Trap DRYLINE II Water Trap
Category "Admit/
Discharge"
Screen
Adult Adult Adult (60- 16100-00) Adult/Child (60- 13100-00) Adult/Child (100- 000080-00)
Child Child Child (60- 16200-00) Neonate (60- 13200-00) Neonate (100- 000081-00)
Neonate Neonate Child (60- 16200-00) Neonate (60- 13200-00) Neonate (100- 000081-00)
Use the adult flow sensor for a patient whose tidal volume is above 150 mL.
Use the pediatric flow sensor for a patient whose tidal volume is below 300 mL.
Make sure to use the correct flow sensor depending on the patient conditions, adult or
pediatric and the tidal volume.
Do not confuse the gas sampling line with other compatible tubing, e.g. IV-lines.
2 Connect the end (for adult: 22/15 mm, pediatric: 15 mm) of the flow sensor, marked to the patient tracheal
tube or similar.
3 Connect the end of the flow sensor to the patient breathing system. For best results, a heat and moisture
exchanger (HME) or similar should be put between the flow sensor and the breathing system.
4 Connect the pressure line of the flow sensor to the flow sensor connector on the MGU-810.
5 Connect the gas sampling line of the flow sensor (for adult: colorless, for pediatric: blue) to the gas inlet of the
water trap. When the water trap is half full, empty the water trap's reservoir.
(
Maintenance Manual "Water Trap (Multigas Unit)" P8-5)
WARNING
The contents of the water trap should be handled as a potential infection hazard.
6 To prevent accumulation of condensed fluid, the flow sensor shall be always be positioned a few degrees off
the horizontal level towards the ventilator side. For the same reason, the pressure tubes shall exit the flow
sensor upwards.
7 The pressure tubes should be routed in such a way that a water lock is formed by a section of tubing being
positioned lower than the flow sensor connector on the MGU-810.
8 A patient breathing system leakage test shall be performed according to the recommendations of the ventilator
manufacturer.
7-89
Chapter 7 Monitoring Multigas Unit/SPIRO
CAUTION
The adult flow sensor dead space is 6.9 mL and the flow resistance is 1.8 cmH2O at 60 L/
min.
The pediatric flow sensor dead space is 0.75 mL and the flow resistance is 0.9 cmH2O at 10
L/min.
Adjust ventilation accordingly.
To prevent condensation, the patient breathing circuit, flow sensor and pressure tubing
should not be directly exposed to cooling device such as fans or cooling blankets.
Leakage of gas from the patient breathing system may occur if the pressure or gas sampling
lines are not connected to the MGU-810.
The pressure tube and gas sampling lines of the flow sensor should always be routed from
the patient circuit to the MGU-810 such a way as to avoid kinking.
Flow sensors that have suffered damage to sensor head, tubing or tubing connector must
not be used.
If liquid has entered the pressure tubes, it can be removed by gently tapping or shaking the
flow sensor.
NOTE
Even though the CO2 measurement can be done in several units or modules, setups for the
alarm limit, measurement unit and scale are common for all the units and modules.
When a measurement unit is changed, make sure to set the alarm condition for that unit. Set
the alarm for each measurement unit.
The data of currently set measurement unit will be displayed on the graphic/tabular trend.
7-90
Chapter 7 Monitoring Multigas Unit/SPIRO
The gas data display when a respiration is not detected can be selected from [None] (bar display) or [Insp. Only]
(displays only the inspiratory data).
NOTE
When [Insp. Only] is selected for "GAS Display during Undetected Breath" and if only
inspiratory data is displayed, inspiratory and expiratory data display on the central
monitor will become invalid.
When [Insp. Only] is selected for "GAS Display during Undetected Breath" and if only
inspiratory data is displayed, the GAS alarm will not be generated.
7-91
Chapter 7 Monitoring Multigas Unit/SPIRO
2 4 3 4
5 5
6 6
7 7
8 11 11
GAS_N2O Screen (MGU-800 series ) GAS_AGT Screen (MGU-800 series )
5 4 4
9 6
6
7
11
GAS_MAC Screen (MGU-800 series) GAS_RESP Screen (MGU-800 series)
4 4
11
GAS_SPIRO Screen (MGU-810 series )
4
10
11
7-92
Chapter 7 Monitoring Multigas Unit/SPIRO
EtCO2 Alarm
NOTE
Set the upper limit in the range of 3 mmHg to 100 mmHg/0.3 kPa to 13.3 kPa/0.3% to
13.3%. Setting a value above 100 mmHg/3.3 kPa/13.3% will turn OFF the alarm.
Set the lower limit in the range of 1 mmHg to 98 mmHg/0.1 kPa to 13.1 kPa/0.1% to
13.1%. Setting a value below 1 mmHg/0.1 kPa/0.1% will turn OFF the alarm.
REFERENCE
The alarm limit should be set for each unit (mmHg/kPa/%).
The upper/lower limit can be set in 1 mmHg/0.1 kPa/0.1% increment.
InspCO2 Alarm
NOTE
Set the upper limit in the range of 1 mmHg to 4 mmHg/0.1 kPa to 0.4 kPa/0.1% to 0.4%.
Setting a value above 4 mmHg/0.4 kPa/0.4% will turn OFF the alarm.
REFERENCE
The alarm limit should be set for each unit (mmHg/kPa/%).
The upper/lower limit can be set in 1 mmHg/0.1 kPa/0.1% increment.
ExpO2 Alarm
NOTE
Set the upper limit in the range of 18% to 100%.
The alarm will turn OFF if a value above 100% is set.
Set the lower limit in the range of 18% to 100%.
The alarm will turn OFF if a value below 18% is set.
REFERENCE
The upper/lower limit can be set in 2% increment.
InspO2 Alarm
NOTE
Set the upper limit in the range of 18% to 100%.
7-93
Chapter 7 Monitoring Multigas Unit/SPIRO
REFERENCE
The upper/lower limit can be set in 2% increment.
NOTE
Set the upper/lower limit in the range of 0% to 100%.
The upper limit and lower limit will turn OFF if a value above 100% and below 0% is set
respectively.
REFERENCE
The upper/lower limit can be set in 2% increment.
NOTE
The adjustable range of the upper limit differs depending on the anesthetic gas label.
ISO, HAL, ENF: 0.5% to 6.0%
SEV: 0.5% to 8.0%
DES: 0.5% to 18.0%
The alarm will turn OFF if a value above the range is set.
The adjustable range of the lower limit differs depending on the anesthetic gas label.
ISO, HAL, ENF: 0.5% to 6.0%
SEV: 0.5% to 8.0%
DES: 0.5% to 18.0%
The alarm will turn OFF if a value below the range is set.
REFERENCE
The upper/lower limit can be set in 0.5% increment.
MAC Alarm
NOTE
Set the upper limit in the range of 0.1 to 9.9.
The upper limit alarm will turn OFF if a value below 9.9 is set.
REFERENCE
The upper limit can be set in 0.1 increments.
RR/Apnea Alarm
NOTE
Set the upper limit of the RR alarm in the range of 10 to 150Bpm. The upper limit alarm
will turn OFF if a value above 150Bpm is set.
Set the APNEA upper limit in the numeric range of 10 to 60 sec. The upper limit alarm will
turn OFF if a value above 60 seconds is set.
7-94
Chapter 7 Monitoring Multigas Unit/SPIRO
Set the lower limit of the RR alarm in the range of 5 to 145Bpm. The alarm will turn OFF
if a value below 5Bpm is set.
REFERENCE
The adjustable increment for RR alarm depends on the patient classification and "RR
Alarm Increment" setting. (Initial Settings>User I/F).
Normal Small
ExpMV/PEAK/PEEP Alarm
NOTE
Set the upper/lower limit of ExpMV alarm in the range of 2.0 L/minute to 20 L/minute for
Adult, 0.5 L/minute to 5.0 L/minute for Child/Neonate.
Set the upper/lower limit of PEAK alarm in the range of 8 cmH2O to 100 cmH2O.
Set the upper/lower limit of PEEP alarm in the range of 2 cmH2O to 50 cmH2O.
REFERENCE
The upper/lower limit can be set as followings.
ExpMV alarm can be set in 0.5 L/minute increment.
PEAK/PEEP alarm can be set in 1 cmH2O increment.
5 Perform a zero calibration. Press the [Zero Cal.] key to start the zero calibration.
NOTE
While performing the zero calibration, the baseline waveform is displayed.
REFERENCE
On the patient monitor, a zeroing (zero calibration) of the multigas unit is periodically
performed, but it can also be performed manually when necessary.
6 Set the "Flow Rate" (sampling flow rate for the multigas unit ).
The selectable "Flow Rate" value differs depending on the type of used water trap (adult/child or neonate) and
sampling line.
NOTE
If the used water trap and the set patient classification does not match, <GAS Check
Water Trap Class> will be displayed.
If <GAS Pump Regulating> is displayed, the gas sampling flow rate may be insufficient.
Check the sample line for any blockage or bent. If the message is still displayed, adjust
the flow rate.
Select the appropriate water trap, sampling line, or flow sensor from 2 types according to
7-95
Chapter 7 Monitoring Multigas Unit/SPIRO
[ON]: The exceeded part of the waveform will be displayed in straight line at the upper or lower scale limit.
[OFF]: The whole part of the waveform will be displayed even if it exceeds the scale. However, the exceeded
part may not be displayed depending on the sweep speed of the waveform displayed above or below the
gas waveform.
NOTE
When the HS-8000/DS-8007 and HM-801 are simultaneously used, the CO2
measurement of the HS-8000/DS-8007 will be prioritized.
[Auto]: The label will be automatically set according to the detected anesthetic gas.
10 When the MGU-810 series is used, set the respiratory waveform scale.
7-96
Chapter 7 Monitoring Multigas Unit/SPIRO
MAC Display
The MAC value can be displayed in the numeric data display area.
Minimum Alveolar Concentration (MAC) is defined in ISO-80601-2-55 as the concentration of anesthetic at which
50% of patients do not move in response to a surgical stimulus.
NOTE
The MAC value will be displayed only if [ON] is set for "MAC Value". Perform the setting if
necessary.
Age compensation can be performed from 1 to 95 years old. Effective MAC range is 0.6 to 1.6.
[ON]: MAC will be calculated by the compensation formula.
[OFF]: The measurement will not be compensated.
The MAC value for MAC correction is calculated from the following formula.
MAC(age)
K(age)N2O K(age)PAGT K(age)SAGT
7-97
Chapter 7 Monitoring BIS Data (BISx)
N2O 104
ISO 1.17
HAL 0.75
ENF 1.63
SEV 1.80
DES 6.60
* Reference for constant number: “Age-related iso-MAC charts for iso¯urane, sevo¯urane and des¯urane in man [R.
W. D. Nickalls and W. W. Mapleson] 2003”
This section explains about the BIS measurement and setup procedure when using the BISx with the BIS I/F Unit,
HBX-800.
WARNING
Clinical judgment should always be used when interpreting BIS in conjunction with other
available clinical signs. Reliance on BIS alone for intra-operative anesthetic management is
not recommended.
BIS values should also be interpreted cautiously with certain anesthetic combinations, such
as those relying primarily on either ketamine or nitrous oxide/narcotics to produce
unconsciousness.
As with any monitored parameter, artifacts and poor signal quality may lead to inappropriate
BIS values. Potential artifacts may be caused by poor skin contact (high impedance),
muscle activity or rigidity, head and body motion, sustained eye movements, improper
sensor placement and unusual or excessive electrical interference.
CAUTION
Generally, the BIS value decreases with the decrease of brain activity. When a patient is in
hypothermia state during cardiac bypass surgery, the suppression of brain wave will cause
the BIS value to decrease.
Pay attention when artifact interferes or signal quality decreases, as it may cause incorrect
BIS measurement.
Pay attention when AC disturbing signal interferes during Filter OFF condition, as it may
cause incorrect BIS measurement.
Pay attention when a pacemaker pulse is displayed in the brain wave, as it may cause
incorrect BIS measurement.
The BIS value tends to increase with the EMG interference. The patient's shivering during
recovery from anesthesia increases the EMG and may cause the BIS value to increase.
When attaching the BIS sensor, lightly apply pressure to the electrode part for about 5
seconds to decrease the electrode impedance.
7-98
Chapter 7 Monitoring BIS Data (BISx)
By connecting the BISx module using the HBX-800 BIS I/F Unit, BIS data can be monitored.
NOTE
When using more than one HM-801, only one AUX connector on the HM-801 can be used
for measurement of CO2, BIS (with HBX-800).
When the BIS data is measured on both the HS-8000/DS-8007 and HM-801, the
measurement of the HS-8000/DS-8007 will be prioritized.
2 Connect the HBX-800 to the AUX connector on one of the following equipments, and connect the BISx to the
serial communication connector on the HBX-800.
*HS-8000 Super Unit
*DS-8007
*HM-801
4 When the system detects the sensor, "Sensor Check" window will be displayed, and impedance for all the
electrodes will be automatically measured.
REFERENCE
Pressing the [Sensor Check] key will also start the sensor check process.
In this display, the impedance value for each electrode, in kilo ohms, appears on the screen along with its
status.
7-99
Chapter 7 Monitoring BIS Data (BISx)
<PASS>: An electrode passes if the impedance for that electrode is less than 7.5 kilo ohms, and the
ground electrode (electrode #2) is less than 30 kilo ohms.
<HIGH>: The impedance value is above 7.5 kilo ohms.
As long as the combined impedance of electrodes #1 and #3 and the combined impedance of
electrodes #1 and #4 are less than 15 kilo ohms, and the ground electrode is less than 30 kilo ohms, the
sensor check will be considered successful.
<LEAD OFF>: The electrode is detached from the patient.
<NOISE>: The signal from the electrode is outside the measurable range.
NOTE
During the sensor check process, EEG waveform will become unstable.
5 If the impedance for all the electrodes are within variable range, <Sensor Check Passed> will be displayed on
the "Sensor Check" window.
6 Press the key on the "Sensor Check" window to end the sensor check process.
BIS measurement will automatically start when the "Sensor Check" window is closed.
NOTE
If the "Sensor Check" window is closed before <Sensor Check Passed> is displayed,
<BIS Perform "Sensor Check"> will be displayed. Press the [Sensor Check] key and start
the sensor check again.
BIS Setup
Press the [Menu], [BIS] keys to display the "BIS" setup screen.
2
1
5
3
6
4
1 Scale
2 Alarm
7-100
Chapter 7 Monitoring BIS Data (BISx)
CAUTION
The conductive parts of sensors and connectors should not contact other conductive
parts, including earth.
To reduce the hazard of burns in the high-frequency surgical neutral electrode
connection, the BIS sensor should not be located between the surgical site and the
electrosurgical unit return electrode.
The BIS sensor must not be located between defibrillator pads when a defibrillator is used
on a patient connected to the patient monitor.
Continuous impedance checking may need to be disabled if the 1 nA, 128 Hz impedance
check signal interferes with other device (e.g., evoked potential monitors).
Considerations when using Electro-Convulsive Therapy (ECT) device during BIS
monitoring: Place ECT electrodes as far as possible from the BIS sensor to minimize the
effect of interference. Certain ECT device may interfere with the proper function of the
BIS monitoring system.
NOTE
During the continuous impedance check, the following impedance will be measured.
A) Combined Impedance of Signal Electrode and Reference Electrode
This check process will not affect the EEG waveform. If the impedance value is within
the allowable range, the check result will not be notified.
B) Impedance of Ground Electrode
This check process will be performed every 10 minutes. During this process, <Ground
Check in Progress> will be displayed, as artifact interferes to the EEG waveform.
5 Smoothing Rate
6 EEG Filter
7-101
Chapter 7 Monitoring BIS Data (A-2000/A-3000)
Press the [Menu], [BIS] ("Parameter") keys to display the BIS setup screen.
1
2
2 Trend E
Ventilator
By connecting a ventilator, numeric data and waveform measured by the ventilator can be displayed on the DS-8400
System.
( Maintenance Manual "Ventilator Connection" P4-3)
By assigning [P-V/F-V] to numeric data box, P-V (pressure-volume) loop/F-V (flow-volume) loop can be also
displayed.
This section explains about the AWP/AWF/AWV scale setup procedure and P-V/F-V screen operation.
7-102
Chapter 7 Monitoring Ventilator
Press the [Menu], [Ext. Device], ("Parameter"), [VENT] key to display the "VENT" screen.
The ventilator measurement will be displayed, and AWF / AWP / AWV / P-V, F-V scale can be set.
1 3
2 4
5
REFERENCE
The scale setup window can be also displayed by pressing the scale on the waveform
display area or [Scale] on the user key.
1 Press the key for "AWF Scale", and set the AWF scale.
2 Press the key for "AWP Scale", and set the AWP scale.
3 Press the key for "AWV Scale", and set the AWV (Airway Volume) waveform scale.
4 Press the key for "P-V, F-V Scale", and set the P-V (Pressure-
Volume) scale.
5 Press the key for "P-V, F-V Scale", and set the F-V (Flow-
Volume) scale.
7-103
Chapter 7 Monitoring Ventilator
CAUTION
For PURITAN-BENNETT ventilator, P-V loop and F-V loop cannot be displayed or printed.
4
2
3
P-V (pressure-volume) loop/F-V (flow-volume) loop is sampled each 60ms and displayed for each
respiration. The beginning of the loop is displayed in cyan, and the rest of the loop is displayed in white.
For the P-V loop, the horizontal axis shows AWP (unit: cmH2O), and vertical axis shows volume (unit: mL).
For the F-V loop, the horizontal axis shows volume (unit: mL), and vertical axis shows AWF (unit: L/min).
Press the [Regist] key to store the displayed P-V/F-V loop as a control loop.
Press the [Review] key to display the registered control loop.
The control loop 1 will be displayed in yellow, and control loop 2 will be displayed in green.
7-104
Chapter 7 Monitoring FLOW-i Data
FLOW-i Data
The FLOW-i can be connected to the serial port, status port of the DS-8400 System or to the HP-800.
( Maintenance Manual "Connection with the FLOW-i" P4-14)
When the FLOW-i is connected, monitoring conditions for CO2 concentration, anesthetic gas concentration, O2
concentration, N2O concentration, and respiration can be set.
WARNING
When the numeric data acquired from FLOW-i is displayed, the following alarms cannot be
set. Also, these alarms will not generate.
InspCO2/EtCO2, InspO2/ExpO2, InspN2O/ExpN2O, InspAgent/ExpAgent, MAC, ExpMV,
PEAK, PEEP
CAUTION
The FLOW-i and MGU-800/MGU-810 cannot be used simultaneously.
The FLOW-i and ventilator cannot be used simultaneously.
NOTE
The CO2 measurement unit is not linked between the FLOW-i and this device.
When the FLOW-i is connected, CO2 alarm cannot be set. Also, the alarm will not generate.
The data of currently set measurement unit will be displayed on the graphic/tabular trend.
7-105
Chapter 7 Monitoring FLOW-i Data
FLOW-i Setup
2 3
5 5
6
5 4
5
FLOW-i_VENT Setup
7-106
Chapter 7 Monitoring FLOW-i Data
Select from [0-50]/[0-100] if the measurement unit is mmHg, and from [0-4]/[0-8]/[0-10] if the unit is kPa or %.
[ON]: The exceeded part of the waveform will be displayed in straight line at the upper or lower scale limit.
[OFF]: The whole part of the waveform will be displayed even if it exceeds the scale. However, the exceeded
part may not be displayed depending on the sweep speed of the waveform displayed above or below the
gas waveform.
NOTE
When the HS-8000/DS-8007 and HM-801 are simultaneously used, the CO2
measurement of the HS-8000/DS-8007 will be prioritized.
NOTE
Set the upper limit of RR alarm in the range of 10 Bpm to 60 Bpm. If a value above 150
Bpm is set, the upper alarm will turn OFF.
Set the upper limit of apnea alarm in the range of 10 sec. to 60 sec. If a value above 60
sec. is set, the upper alarm will turn OFF.
Set the lower limit of RR alarm in the range of 5 Bpm to 145 Bpm. If a value below 5 Bpm
is set, the lower alarm will turn OFF.
REFERENCE
The adjustable increment for RR alarm depends on the patient classification and "RR
Alarm Increment" setting. (Initial Settings>User I/F).
Normal Small
7-107
Chapter 7 Monitoring FLOW-i Data
MAC Display
The MAC value can be displayed in the numeric data display area.
NOTE
The MAC value will be displayed only if [ON] is set for "MAC Value". Perform the setting if
necessary.
[ON]: The MAC value will be displayed in the numeric data display area.
[OFF]: The MAC value will not be displayed in the numeric data display area.
7-108
Chapter 7 Monitoring FLOW-i Data
The ventilator measurement will be displayed, and AWF / AWP / AWV / P-V, F-V scale can be set.
3
1 4
2 5
REFERENCE
The scale setup window can be also displayed by pressing the scale on the waveform
display area or [Scale] on the user key.
2 AWP Scale
3 AWV Scale
7-109
Chapter 7 Monitoring FLOW-i Data
4
2
3
P-V (pressure-volume) loop/F-V (flow-volume) loop is sampled each 60ms and displayed for each
respiration. The beginning of the loop is displayed in cyan, and the rest of the loop is displayed in white.
For the P-V loop, the horizontal axis shows AWP (unit: cmH2O), and vertical axis shows volume (unit: mL).
For the F-V loop, the horizontal axis shows volume (unit: mL), and vertical axis shows AWF (unit: L/min).
Press the [Regist] key to store the displayed P-V/F-V loop as a control loop.
Press the [Review] key to display the registered control loop.
The control loop 1 will be displayed in yellow, and control loop 2 will be displayed in green.
7-110
Chapter 7 Monitoring SvO2/CCO Data
SvO2/CCO Data
The DS-8400 System can display the monitoring data of oximeter/CCO measurement device, Vigilance, Vigilance
CEDV, Vigilance II, Vigileo, EV-100, HemoSphere (Edwards Lifescience) or the hemodynamic monitoring device,
PiCCO2, PulsioFlex (PULSION Medical Systems).
( Maintenance Manual "SvO2/CCO Monitor Connection" P4-7)
On the SvO2/CCO data screen, the displayed numeric data can be switched.
STAT Mode: When the Vigilance is in CCO mode, STAT mode display can be set ON or OFF.
Index Display: When the Vigilance is in CCO mode, Index Display can be set ON or OFF.
When the Vigilance is in ICO mode, the 6 latest data of ICO (Intermittent Cardiac Output) and ICI (Intermittent
Cardiac Index) will be displayed.
NOTE
STAT Mode: When Vigilance is in CCO mode, STAT mode display can be set ON or OFF.
Index Display: When Vigilance is in CCO mode, Index display can be set ON or OFF.
When the Vigilance is in ICO mode, the 6 latest data of ICO (Intermittent Cardiac Output)
and ICI (Intermittent Cardiac Index) will be displayed.
STAT Mode [OFF], Index Display [OFF]: SvO2 (or ScvO2) , CCO, EDV, BT will be displayed inside the
SvO2+CO numeric data box.
7-111
Chapter 7 Monitoring INVOS Data
STAT Mode [OFF], Index Display [ON]: CCI and EDVI will be displayed instead of CCO and EDV.
STAT Mode [ON], Index Display [OFF]: CCO_STAT and EDV_STAT will be displayed instead of CCO and
EDV.
STAT Mode [ON], Index Display [ON]: CCI_STAT and EDVI_STAT will be displayed instead of CCO and
EDV.
NOTE
ON/OFF of STAT mode can be changed only when Vigilance is connected.
INVOS Data
In the INVOS numeric data box, measurement data of Lt-rSO2/Rt-rSO2 will be displayed.
On the INVOS screen, the channel for Lt-rSO2/Rt-rSO2 data can be selected.
2 Press the [ch*] key for the INVOS label ("Lt-rSO2" / "Rt-rSO2" / "S1-rSO2" / "S2-rSO2") to set the channel.
7-112
Chapter 7 Monitoring Stopwatch
Stopwatch
The stopwatch function can be used by setting the [Stopwatch] key on the numeric data box or on the user key.
1 Press the [Stopwatch] key on the numeric data box or on the user key.
Label Setup
Start/Stop
NOTE
If the discharge procedure is performed during stopwatch operation, the counting will
stop and will be reset to "00:00:00" .
The stopwatch counting will continue even when the monitoring is suspended.
7-113
Chapter 7 Monitoring Multiparameter Connector Setup for BP, TEMP, CO Measurement
On the Super Unit and Multi Module, multiparameter connectors are provided. The quantity of multiparameter
connectors are as follows.
Multiparameter Connectors Super Unit Multiparameter Connectors Super Unit
3 ports 2 ports
2 ports 1 ports
By using the multiparameter connector, any combination of BP, TEMP and CO measurement can be performed
according to the monitoring purpose.
By using the 2ch TEMP relay cable, 2ch BP relay cable, or 2ch BP conversion cable, 2 channels of temperature and
BP can be monitored through one multiparameter connector.
By using the Multi Module with the Input Box, up to 8 channels of BP, 8 channels of TEMP and 1 channel of CO
can be measured.
The multiparameter connector setup can be performed on the "Initial Settings" menu.
( Maintenance Manual "Unit Module Setup" P4-19)
Blood Pressure
4ch
Blood Pressure 2ch -
(2ch)
Temperature
Blood Pressure
2ch
Temperature 4ch -
(1ch)
Temperature
Temperature
Temperature - 6ch -
Temperature
Blood Pressure
2ch
Temperature 2ch 1ch
(1ch)
CO
Blood Pressure
4ch
Blood Pressure - 1ch
(2ch)
CO
7-114
Chapter 7 Monitoring Multiparameter Connector Setup for BP, TEMP, CO Measurement
Temperature
CO
The numbers in parenthesis shows the channels when using the 1ch BP conversion cable.
Temperature
- 4ch -
Temperature
Temperature
- 2ch 1ch
CO
The numbers in parenthesis shows the channels when using the 1ch BP conversion cable.
On the DS-8007, there are 2 temperature connectors in addition to the multiparameter connectors.
7-115
Chapter 7 Monitoring Multiparameter Connector Setup
It is necessary to manually set the measuring parameter for each multiparameter connector.
Example:
To assign BP5 to multiparameter connector 1 for the HM-800 Multi Module inserted to the built-in slot:
1
1 Press the [Menu], [Initial Settings], [System], [Unit Module], [Multiamplifier] keys.
2 Select the multiparameter connector location. The selected location will be displayed in blue.
3 Assign the parameter to the selected location. In this case, select [5] for "BP".
The parameter will be assigned to the selected connector.
CAUTION
The same parameter cannot be set to more than one connectors.
By setting [OFF] for the unused connector, the parameter will become selectable on another
connector.
If the parameter assigned to the multiparameter connector and the connected relay cable
does not match, the connector location will be displayed in red and the connected relay
cable type will be displayed.
NOTE
The temperature measured on the temperature connectors on the DS-8007 are fixed as T1,
T2.
7-116
Chapter 8 Review Function Common Operation
The common operations for all the review screens are explained below.
1 Time Bar
Changing the time span, scrolling the time, displaying the latest data can be performed.
1 2 3 4 2 1 5
With the displayed date/time, the review data display can be switched.
Other review data (graphic/tabular trend) can be displayed without moving the current cursor time.
8-1
Chapter 8 Review Function Arrhythmia Analysis
Arrhythmia Analysis
Arrhythmia Definition
QRS
Learn Judgment Pattern Matching
Waveform Arrhythmia
VPC? VPC!
The arrhythmia detection is performed by learning the normal waveform of the patient, and determines the VPC by
comparing the waveform (QRS pattern) and R-R interval for each heartbeat.
The parameters such as QRS amplitude, QRS width, QRS polarity, RR interval are compared with the normal
waveform to extract the abnormal QRS.
Then, the QRS with suspected VPC is pattern matched. The noise and VPC are distinguished to determine the VPC,
and generates the arrhythmia alarm.
WARNING
Objective and constant arrhythmia detection is possible through the fixed algorithm
incorporated in this unit.
However, excessive waveform morphology change, motion artifact, or the inability to
determine the waveform pattern may cause an error, or fail to make adequate detection.
Therefore, physicians should make final decisions by closely checking the data obtained by
printing and recall waveform.
CAUTION
For stable arrhythmia detection and ECG monitoring, verify proper electrode placement,
lead, waveform size, and filter mode selection. If not properly selected, it may cause
erroneous detection.Waveform Size: ECG
QRS Classification
Each QRS will be classified to the following pattern.
N (Normal) Normal QRS beat
8-2
Chapter 8 Review Function Arrhythmia Analysis
Arrhythmia Type
With the QRS judgment, the following types of arrhythmia alarm will be generated.
Arrhythmia Description Detection Criteria
Run (Consecutive VPC) ON, OFF Continuous VPC exceeding the preprogrammed value (2
2 beats to 8 beats, 1 beat increments beats to 8 beats) is detected.*3
Couplet (Couplet VPC) ON, OFF 2 continuous VPC beats are detected.
Frequent (Frequent VPC) ON, OFF VPC exceeding the preprogrammed value is detected within
1 bpm to 50 bpm, 1 beat increments 1 minute.
Ext Tachy (Extreme ON, OFF The upper alarm limit of extreme tachycardia is exceeded.
Tachycardia) 22 bpm to 300 bpm,
22 bpm to 60 bpm, 1 beat increments
60 bpm to 300 bpm, 5 beat increments
Ext Brady (Extreme ON, OFF The lower alarm limit of extreme bradycardia is exceeded.
Bradycardia) 20 bpm to 295bpm
20 bpm to 60 bpm, 1 beat increments
60 bpm to 295 bpm, 5 beat increments
R on T (R on T VPC) ON, OFF VPC is detected within the preprogrammed RR interval (200
200 ms to 600 ms, 8 ms increments ms to 600 ms).
Multiform (Multiform ON, OFF 2 different forms of VPC beats are detected within 4
VPC) minutes.
Vent Rhythm (Ventricular ON, OFF Continuous VPC beats with HR below the set value for "HR
Rhythm) Lower Limit for Run" (0 bpm to 100 bpm), and same or
above value of the set beats for Run (2 beats to 8 beats) are
detected.
SVT (Supraventricular ON, OFF Continuous SVPC exceeding the preprogrammed value (2
Tachycardia) 2 beats to 10 beats, 1 beat increments beats to 10 beats) is detected.
Irregular RR (Irregular ON, OFF RR interval variability exceeding the preprogrammed value
RR Interval) 10% to 20%, 5% increments (10% to 20%) is detected.
Prolonged RR ON, OFF RR interval of 1.75 times longer than the normal RR interval
(Prolonged RR Interval) is detected.
Pacer Not Capture (Non- ON, OFF HR is not detected from the pacing pulse within the set
Capture) 80 ms to 480 ms, 8 ms increments duration.
Pacer Not Pacing ON, OFF Pacing pulse and HR are not detected during the set instant
(Oversensing) 20 bpm to 200 bpm, HR.
20 bpm to 150 bpm, 5 beat increments
150 bpm to 200 bpm, 10 beat
increments
Triplet (Triplet VPC) ON, OFF 3 continuous VPC beats are detected.
S Frequent (Frequent ON, OFF SVPC exceeding the preprogrammed value is detected
SVPC) 1 bpm to 50 bpm, 1 beat increments within 1 minute.
8-3
Chapter 8 Review Function Arrhythmia Analysis
*3: HR of same or above the set value of "HR Lower Limit for RUN" (0 bpm to 100 bpm)
*4: x indicates N, P, F, ?.
NOTE
If the patient classification is "Adult" or "Child", Asystole, VF, VT alarm cannot be turned OFF
unless [ON/OFF] is selected for "Asystole, VF, VT Alarm" under "Initial Settings".
If the patient classification is "Neonate", VF, VT can be turned OFF regardless of the setting
for "Asystole, VF, VT Alarm" under "Initial Settings".
8-4
Chapter 8 Review Function Arrhythmia Analysis
Set the Run analyzing condition for the arrhythmia analysis. Run alarm will generate if the HR is same or
above the set value.
Press the / keys for "HR Lower Limit for Run" to set the HR in the range from 0 bpm to 100 bpm.
Set the SVT analyzing condition for the arrhythmia analysis. SVT alarm will generate if the HR is same or
above the set value.
Press the / keys for "HR Lower Limit for SVT" to set the HR in the range from 100 bpm to 250 bpm.
Arrhythmia Learn
Learning the normal ECG largely affects the accuracy of arrhythmia analysis.
If any error occurs in arrhythmia detection and QRS judgment, performing arrhythmia learning will recover the
original analyzing accuracy.
Arrhythmia learning will be performed for about 20 beats for the normal ECG, but it may take longer if the heartbeat
is unstable.
During arrhythmia learning, arrhythmia alarm other than Asystole, VF, Tachy, Brady, Ext Tachy, Ext Brady will not
generate.
8-5
Chapter 8 Review Function Graphic Trend
NOTE
If [Used] is selected for "Pacemaker", the [Learn] key will not change to blue and
<LEARN> will not be displayed, but the learning process will be performed.
Pressing the key while arrhythmia learning is in process will not stop the process.
Graphic Trend
This section explains the graphic trend function and printing procedure.
If the numeric data is displayed on the home display, 24 of data will be automatically stored and displayed as trend
data.
2 2
5
3 6
4 7
8
2 graphs are displayed on each page, and graphic trend of 4 parameters can be displayed simultaneously
on each graph.
2 Changing the time span, scrolling the time, displaying the latest data
(
"Common Operation" P8-1)
3 Set the parameter, display type, scale.
1 Press the scale area for each parameter, and display the scale selection window.
8-6
Chapter 8 Review Function Graphic Trend
2 Press the key for "Parameter Selection", and select the parameter.
NOTE
The selected parameter will be also registered for the trend group.
The APNEA duration will be stored when it exceeds the upper alarm threshold level.
If lower than the alarm threshold level, it will be stored as "0 (zero)".
1 Pressing the center part of will display the trend data at the cursor position.
2 The cursor will move to left and right by dragging.
3 Press / to adjust the cursor position.
The data display at cursor position will be automatically erased after fixed duration.
4 Press to display the 10-minute trend data before and after the cursor position.
5 Press to return the display to the previous time range.
REFERENCE
The displayed data is compressed as follows depending on the display interval.
VPC: Maximum value within the display interval
APNEA: Maximum value within the display interval
Other than above: Latest value within the display interval
For example, if the 24-hour trend for the parameter with minimum resolution of 1 minute
is displayed, one mark will be displayed for the 12-minute (720-second) data.
If the display resolution is higher than the minimum resolution of the data, the same data
is repeated to match the display resolution.
Refer to the following table for resolution. The data resolution differs according to the
parameter.
Display Resolution
Minimum Resolution
8-7
Chapter 8 Review Function Graphic Trend
Display Resolution
Minimum Resolution
Data Resolution
Minimum
Parameter
Resolution
3
1 Time Bar 1 4
Select the time bar display interval from [4h]/[8h]/[12h]/
[16h]/[20h]/[24h]/[36h]/[48h]. 2
8 Press [Print].
To print the trend data, press the [Print] key, select the parameter, and press the [Enter] key.
8-8
Chapter 8 Review Function Graphic Trend
PR_SpO2-1,
SpO2 Pulse Rate 100, 200, 300 bpm
PR_SpO2-2
BP1~8 Blood Pressure (Systolic / Mean / Diastolic) 4, 8, 16, 20, 24, 40 kPa
20, 40 cmH2O
20, 50, 100, 150, 200, 300 mmHg
PDP Peak Diastolic Pressure of IABP
4, 8, 16, 20, 24, 40 kPa
BIS Bispectral Index (BIS Monitor Measurement) 25, 50, 75, 100 -
8-9
Chapter 8 Review Function Short Trend
PEEP*1 Peak End Expiratory Pressure 10, 20, 50, 100 cmH2O
*2
Lt-rSO2
Rt-rSO2*2
Regional Cerebral Oxygen Saturation 20 to 100 %
S1-rSO2*2
S2-rSO2*2
*1: When an anesthesia delivery system (FLOW-i, GE, Draeger) is used, the measurement by the anesthesia delivery system will
be used.
*2: The external device parameters to be displayed on the graphic trend/tabular trend needs to be selected in advance on the
"Trend Data Setup" window ([Data Review>Graphic Trend or Tabular Trend] or [Initial Settings>External Device>Main Unit/HP-
800]).
NOTE
The apnea duration will be stored when it exceeds the upper alarm threshold level. If lower
than the alarm threshold level, it will be stored as "0 (zero)".
Short Trend
8-10
Chapter 8 Review Function Short Trend
8-11
Chapter 8 Review Function Short Trend
The data within the reference lines (including the parameters without the reference line display) will be displayed
with lower brightness.
The data outside the reference lines will be displayed with higher brightness.
The reference lines can be displayed by selecting [Enable] for "Reference Line Function". (Menu>Display
Config.>Detail Setup)
However, it cannot be displayed for the overlapped short trend. And, when the reference line function is enabled, the
function to display the alarm occurrence point on the graph in red cannot be used.
When [Enable] is set for "Reference Line Function", ON/
OFF and upper/lower limit of reference line display can be
selected on the "Short Trend Setup" window for each
parameter.
The "Short Trend Setup" window can be displayed by
pressing the short trend scale area.
Pressing the center part of will display the review data (tabular trend/graphic trend/zoom wave)
at the cursor point.
(However, zoom wave can be displayed only when the full disclosure waveform function is enabled.)
The cursor cannot be displayed for the overlapped short trend. And, when the cursor function is enabled, the function
to highlight the alarm generated data cannot be used.
When the cursor function is enabled, the function to enlarge/reduce the short trend display area cannot be used.
During the cursor display, the short trend data will not be updated. When the cursor is not used for 10 seconds or
when other window is displayed, the cursor will be automatically cleared.
8-12
Chapter 8 Review Function Tabular Trend
Tabular Trend
This section explains the tabular trend function and printing procedure.
If the numeric data is displayed on the home display, 24 of data will be automatically stored and displayed in 10
seconds/30 seconds interval.
2
2 3
4
5
8
6
7
2 Changing the displayed time, scrolling the time, updating the data
(
"Common Operation" P8-1)
3 The list will be scrolled up or down to display other parameters.
NOTE
The minimum resolution of NIBP is 20 seconds. If multiple measurements have been
performed within 20 seconds, the latest value will be displayed.
The data resolution differs according to the parameter.
24 hours of data will be stored regardless of the time bar display range.
If the display resolution is higher than the minimum resolution of the data, the same data
is repeated to match the display resolution.
Data Resolution
Minimum Resolution Parameter
8-13
Chapter 8 Review Function Tabular Trend
(
"Parameter Setup for Tabular Trend" P8-15)
7 Press the [Print]/[Print (All)] key.
8 The displayed data time can be scrolled by dragging the data display area.
For the data when the measurement was not performed (before admittance) or when the monitoring was suspended,
the time will be displayed as " : ".
Also, if the measured data is not displayed on the home display, or BP zero balance is not performed, the data will
not be displayed.
The alarm generated data will be displayed with red background.
The date column of alarm generated data will be also displayed with red background.
NOTE
The red background will be displayed for the alarm generated parameter.
The alarm display for the expiratory and inspiratory parameter such as EtCO2 and InspCO2
will be the same.
For example, if the alarm is generated for BP-S, the background color of BP1-S, BP1-M,
BP1-D will be displayed in red.
On the left side of the parameter, the color assigned for the corresponding parameter will be displayed.
8-14
Chapter 8 Review Function Tabular Trend
4
3 2
5
4
6
2 Press the [Fix x param.] key to set the fixed parameters.
3 Select the display location for the parameter. The selected location will be displayed with blue frame and
will be displayed at the side.
To change the location, directly press the desired location or drag the key up or down.
To change the displayed page, press the / keys on the left.
SvO2 / CCO SvO2, ScvO2, SaO2, O2EI, B-Temp, CCO, CCO-STAT, CCI, CCI-STAT, DO2, RVEF, RVEF-
STAT, VO2, SV, SV-STAT, SVI, SVI-STAT, SVR, SVRI, SVV, EDV, EDV-STAT, EDVI, EDVI-
STAT, MAP, ESV, ESVI, CFI, iCO, iCI, iSV, iSVI, iSVR, iSVRI, GEDV, GEDI, GEF, EVLW, ELWI,
PVPI, ITBV, ITBI, VO2e, VO2I, VO2Ie, iB-Temp, SQI, MAP, CVP, HR, PR, SpO2, iMAP, iCVP,
iAvgPR, DO2I, HGB, dPmx, CO CAL
8-15
Chapter 8 Review Function Recall
Other BIS, SQI, EMG, SR, SEF, TOTPOW, IMP, AEP, Lt-rSO2, Rt-rSO2, S1-rSO2, S2-rSO2, tcpO2,
tcpCO2
NOTE
The APNEA duration will be stored when it exceeds the upper alarm threshold level.
If lower than the alarm threshold level, it will be stored as "0 (zero)".
The external device parameters to be displayed on the graphic trend/tabular trend
needs to be selected in advance on the "Trend Data Setup" window ([Data
Review>Graphic Trend or Tabular Trend] or [Initial Settings>External Device>Main
Unit/HP-800]).
The measurement unit of tcpO2, tcpCO2 can be set on the TCM4 or TCM5 FLEX.
When the measurement unit is changed, the tabular trend data of tcpO2 and tcpCO2
on the bedside monitor will be deleted.
REFERENCE
[H Module] is a generic term for HS-8000, HM-800/HM-801, HP-800, HG-810/HG-820.
When the FLOW-i, GE or Draeger anesthesia delivery system is connected, the
display will change to [H Module/Anes.].
Recall
This section explains about the recall function and the setup procedure.
2 Recall Factor
4 Mark
When the alarm for the specified recall factor occurs, waveforms (max. 2 waveforms/12 seconds) and numeric data
for each recall factor will be stored up to 200 data. On the display selection menu, the data to be displayed can be
selected from the stored recall data. 5 compressed recall waveforms will be displayed. Pressing the waveform area
will display the enlarged waveform.
If the recall data exceeds 200, the data will be erased from the oldest one.
8-16
Chapter 8 Review Function Recall
The recall waveform will be acquired from the point prior to alarm occurrence so that alarm-generated point will be
displayed at 7 to 8 seconds point on the 12-seconds recall waveform. mark indicates the alarm generated point.
2
3
4
5
2 Changing the displayed time, scrolling the time, updating the data
(
"Common Operation" P8-1)
3 Press the [Display Selection] key, and set the recall display.
NOTE
The "Display Selection" setting will be also applied to the recall list display on the numeric
data display area.
(
"Extended Function (Recall List)" P3-14)
8-17
Chapter 8 Review Function Recall
On the enlarged recall waveform display, the recall waveform will be displayed in 25mm/s and by using the cursor,
the data before and after the alarm occurrence can be checked.
5
2
2 Measurement
The measurement value will be displayed.
5 The waveform can be scrolled by dragging the waveform area to left and right.
8-18
Chapter 8 Review Function Recall
Recall Setup
The storing condition at alarm occurrence can be set for the recall function.
The recall waveform and recall factor (numeric data, arrhythmia) can be selected.
2 3
Neonate
Adult Child
Numeric Data Alarm Arrhythmia Alarm
For the parameters measured on the multigas unit, the delay time is 8 seconds.
8-19
Chapter 8 Review Function OCRG
OCRG
2
3
4
7 5
8 6
2 / will become effective by using the optional CFast card. The cursor will move to the alarm
generated time.
4 OCRG Setup
5 4
8-20
Chapter 8 Review Function OCRG
5 Alarm Display
The color of the alarm occurrence point can be changed for display according to the alarm factor selected on
the Setting > Alarm Display Selection.
6 Printing
The currently displayed trend and compressed waveform on the OCRG screen will be printed.
1 Pressing the center part of will display the trend data at the cursor position.
2 The cursor will move to left and right by dragging.
3 Press / to adjust the cursor position.
The numeric data on the cursor position will be displayed. The display will not be updated when the numeric
data is displayed by the cursor. Press [Real Time] to resume.
8-21
Chapter 8 Review Function Alarm History
Alarm History
This section explains the alarm history function and printing procedure.
The alarm generation of numeric data, arrhythmia, device status and change in alarm settings can be stored as alarm
history. Maximum of 1599 data can be stored.
NOTE
The alarm history cannot be deleted manually. When 1600 data is exceeded, the data will
be deleted from the oldest one.
2
2
3
5 4
2 Changing the displayed time, scrolling the time, updating the data
(
"Common Operation" P8-1)
3 Set the alarm history display.
5 The displayed data can be scrolled by dragging the display area up and down.
8-22
Chapter 8 Review Function Alarm History
Item Details
Time The alarm generated time or alarm setting changed time will be displayed.
Code The code related to alarm generation or alarm setting change will be displayed in
hexadecimal.
Factor The factor for alarm generation and alarm setting change will be displayed.
In case of numeric data/arrhythmia alarm, the numeric data and alarm setting at
alarm generation will be also displayed.
In case of alarm setting change, the changed value will be also displayed.
Duration (sec.) The duration of numeric data/arrhythmia/device status alarm generation, alarm
suspend, monitor suspend, night mode will be displayed in seconds. The
maximum displayable value is 99999 sec.
It will not be displayed for the alarm setting change.
8-23
Chapter 8 Review Function Zoom Wave
Zoom Wave
This section explains about the "Zoom Wave" window. (When using the optional CFast card)
Maximum of 6 waveforms (9.8 seconds each) can be displayed.
The "Zoom Wave" window can be also displayed by pressing the waveform area on the "Full Disc. Wave" window.
If the optional CFast card is not used, the latest enlarged recall waveform will be displayed.
3 Changing the displayed time, scrolling the time, updating the data
(
"Common Operation" P8-1)
4 The numeric data of the displayed time will be displayed.
9 The waveform can be scrolled by dragging the waveform area to left and right.
8-24
Chapter 8 Review Function ST Measurement
ST Measurement
On the ST display, ECG for the selected time duration (10 sec./1 min./5 min./10 min.) will be displayed overlapped
in 1 block.
If 3-lead cable is used, maximum of 8 hours of ST waveform will be displayed.
NOTE
If 3-lead cable is used, the measurement will be performed for only the displayed leads.
For the following case, ST level will not be displayed.
When learning arrhythmia.
When the lead is off.
When the reference waveform is not set.
When "N" or "S" is not detected for QRS within 30 seconds.
2
3
4
5
2 Changing the displayed time, scrolling the time, updating the data
(
"Common Operation" P8-1)
3 Changing the Displayed Waveform Size
Select from [x1/4]/[x1/2]/[x1]/[x2]/[x4].The same waveform size will be applied to all the leads. The selected
size will not be applied to the ECG waveform on the home display.
The "Setup" window will be displayed and "Slide Show" (1 sec./5 sec./10 sec./20 sec./30 sec.) can be selected.
REFERENCE
When 3-lead cable is used, 36 blocks will be displayed. When 4, 5, 10-lead cable is
used, 3 blocks for each lead will be displayed.
The duration of each block can be selected from [10 sec.]/[1 min.]/[5 min.]/[10 min.].
For the selections other than [10 sec.], the overlapped waveform for the selected
duration will be displayed.
8-25
Chapter 8 Review Function ST Measurement
5 Printing
The ST reference waveform will be automatically set after learning the arrhythmia.
The reference waveform can be updated manually.
2 Update the ST reference waveform. Press the [Update Ref. Wave] key.
CAUTION
For the lead which the electrode is detached, the reference waveform cannot be
set.Check if the electrode is correctly attached, and perform the setup again.
16 beats average of the ECG judged as normal QRS by arrhythmia analysis will be set as the reference
waveform.
While updating the reference waveform, the [Update Ref. Wave] key will be displayed in blue.
The updated time of the reference waveform will be displayed.
NOTE
While learning arrhythmia, or if VPC is present, it will take more than 16 beats to set the
reference waveform.
When the electrode quantity is changed, the reference waveform will be automatically
updated.
In case such as when the patient is discharged, the reference waveform will be
automatically set.
1 Slide the reference point to right and left using the key.
2 Slide the measurement point to right and left using the key.
NOTE
Set the reference point in the range of –240 ms to 0 ms in increments of 10 ms from
the peak of QRS to the P wave direction.
8-26
Chapter 8 Review Function ST Measurement
ST Alarm Setup
Set the ST upper limit and lower limit for the reference waveform.
5
3
4
[OFF]: Alarms will not generate even if the alarm for each lead is set to ON.
REFERENCE
The upper and lower limit can be set in 1 mm/0.1 mV increments.
6 The ST level display can be selected from [Absolute Value] / [Relative Value].
8-27
Chapter 8 Review Function 12-Lead Analysis (Optional Function)
This section explains about the 12-lead analysis function. By using the 10-electrode cable, 12-lead ECG can be
displayed, analyzed, stored, and printed. Maximum of 10 analyzed results can be saved.
WARNING
The 12-Lead ECG analysis function is designed to acquire and interpret ECG data from a
resting, supine patient. If ECG signals from moving or shaking patients are acquired,
erroneous 12-lead interpretation result may occur. Always ensure that the patient is kept
motionless during 12-lead ECG signal acquisition and analysis.
The 12-lead ECG analysis function is intended for use with adult and pediatric patients.
All computerized ECG analysis results should be reviewed by a physician before making
decision of the patient treatment.
CAUTION
Interpretation and Minnesota codes given by this device do not instruct the physician as to
the kind and degree of cardiac disease. Accordingly, four value judgments are given for ECG
waveform and although "abnormal" indicates a large possibility of organic cardiac disease,
there are cases where no cardiac disease exists despite an abnormal ECG (that is, an
abnormal ECG may be caused by something other than heart).
On the other hand, care should be taken in the event any preclinical coronary
arteriosclerosis could be present despite a normal ECG interpretation.
Therefore, for a proper diagnosis, the ECG should be integrated with other interpretations.
ECG Recording by the Mason-Likar System
The 12-lead ECG recorded with the torso placement of the limb leads (Mason-Likar 12-lead
system) may differ from the standard 12-lead ECG. Moreover, waveforms may differ
somewhat also in a supine position and a standing position (sitting position).
We recommend to carry out the recording of the ECG by taking into consideration the
waveform differences according to electrode positions or postures.
About the ECG Analysis Program
The ECG analysis program is intended to analyze the standard 12-lead ECG waveforms.
Therefore, the analyzed result for waveforms recorded with the torso placement of the limb
leads (Mason-Likar 12-lead system) may differ from that of the standard 12-lead ECG
waveforms.
When a pacemaker is used, select [Used] for "Pacemaker" under "Admit/Discharge" menu.
The threshold values for classification of 12-lead ECG interpretation and Minnesota code
are set by age and sex as follows.
1. Male and Female of ages 19 years old and above
2. Male of age 12 through 18 years old
3. Female of age 12 through 18 years old
4. Male and Female of ages 3 through 11 years old
5. Male and Female of ages below 2 years old
If no patient information (i.e. Default: "Class.": [Adult], "Sex": undetermined) has been
entered, the system algorithm will handle the patient as a "35 years old male".
Before the analysis, make sure the patient classification ([Adult] / [Child]) is properly
selected. If [Neonate] is selected, the 12-lead ECG analysis will not function.
Enter the age of patient if known. If no age information (i.e. Default: [0]) has been entered,
the system algorithm will handle the patient as "35 years old".
Enter the sex of patient if known. If no sex information (i.e. Default: undetermined) has been
entered, the system algorithm will handle the patient as "Male".
If the patient classification is set as [Child] and no age (i.e. Default: [0] ) has been entered,
the system algorithm will handle the patient as "less than 2 years old."
8-28
Chapter 8 Review Function 12-Lead Analysis (Optional Function)
NOTE
Electrode Placement for 12-Lead ECG Analysis
When acquiring 12-lead ECG signals, it is recommended to place the limb electrodes
anywhere along the arms and legs. (
"Electrode Placement" P7-3)
If it is difficult, use the Mason-Likar 12-lead system. To reduce the waveform differences
from the standard 12-lead, it is recommended that the torso placement of the R and L
electrodes be near as possible to each arm, in the infraclavicular fossae, within the area
unaffected by myoelectricity.
When the 12-lead analysis function is enabled on the DS-8007, the function will be also
enabled on the DS-8400.
2
3
4
5
6
2 Changing the displayed time, scrolling the time, updating the data
(
"Common Operation" P8-1)
3 The real-time waveforms are displayed.
The 12-lead analysis will be performed based on the displayed waveforms.
REFERENCE
A pacemaker pulse will not be displayed on the 12-lead analysis screen even if [ON] is
set for "Pacemaker Pulse".
4 The [Chest Lead]/[Limb Lead] keys will switch the display between chest lead and limb lead.
5 Setup
The 12-lead waveform size, filter, analysis method can be set.
(
"12-Lead Analysis Setup" P8-30)
6 Printing
8-29
Chapter 8 Review Function 12-Lead Analysis (Optional Function)
2 ECG Analysis 2
The timing to read the waveform for ECG analysis can be
set.
3
[Real Time]: The waveform of 10 seconds after the [Start
Analyze] key is pressed will be analyzed. 4
[Review]: The waveform of 10 seconds before the [Start
Analyze] key is pressed will be analyzed.
5
3 Waveform Size 6
The waveform size for the real-time waveform displayed on
the 12-lead screen can be set.
Limb Lead: The waveform size for the limb lead can be changed.
Chest Lead: The waveform size for the chest lead can be changed.
4 Filter
The setup for the AC Filter, EMG Filter, Drift Filter can be performed.
AC Filter: If AC noise is present, select [ON]/ [OFF] for "AC Filter".
If [ON] is selected, cut-off frequency will be 75 Hz.
EMG Filter: If EMG noise is present, select [Strong (25Hz)]/ [Weak (35Hz)]/ [OFF].
Drift Filter: If base line drift is present, select [Strong (0.50Hz)]/ [Weak (0.25Hz)]/ [OFF].
CAUTION
A baseline or notch will be generated on the ECG waveform (display, print, recall) during
the filter setting (up to about 2.4 seconds).
This device complies to the distortion test of EN 60601-2-25 when all the filters are set to
OFF. The frequency characteristic is 0.05 Hz to 150 Hz when all the filters are set to OFF.]
When a pacemaker is used, the baseline fluctuation becomes large and it may be difficult
to read the electrocardiogram. Set the drift filter by checking the electrocardiogram.
5 Background Color
The background color for the 12-lead display can be set.
6 Time Bar
Select the time bar display interval from [4h]/[8h]/[12h]/[16h]/[20h]/[24h]/[36h]/[48h].
8-30
Chapter 8 Review Function 12-Lead Analysis (Optional Function)
3 Dominant Waveform
The reference waveform used for the analysis will be displayed.The dominant waveform is the waveform at
the point of mark on the rhythm waveform.
On the analyzed result, the abnormal lead with the highest grade finding will be highlighted in red.
The dominant waveform display can be switched by pressing the [Chest Lead]/[Limb Lead] keys.
4 Filter Information
5 Analyzed Result
For the analyzed result, overall judgment, numeric data, finding will be displayed.
8-31
Chapter 8 Review Function 12-Lead Analysis (Optional Function)
2 Numeric Data: Main numeric data used for ECG analysis will be displayed.
The abnormal numeric data with the highest grade finding will be highlighted in red.
3 Finding: The findings by the ECG analysis will be displayed. These will be classified by colors according to
the grade specified for each finding.
Grade 6: Red
Grade 4: Blue
Grade 2, 0: Black
The highest grade finding will be highlighted in color specified for each abnormality level.
6 Panorama Display
By pressing the [Panorama] key, overall judgment, finding, abnormal site will be indicated by heart
illustration.
1
2
8-32
Chapter 8 Review Function 12-Lead Analysis (Optional Function)
Press the [Print] key on the analyzed result screen or analyzed waveform screen.
There are following 3 types of analyzed result printing.
Keys displayed when Printer Selection for Graphic Printing Key Note
[Print] key is pressed Display
Waveform Report 12-Lead Waveform Bedside Yes Standard 12-lead waveform printing
Prints the analyzed waveform.
Laser Yes
Analyzed Report 12-Lead Analysis Result Bedside Yes Standard analyzed result printing
Prints the waveform and analyzed
Laser Yes
result.
NOTE
If no patient information has been entered, "Adult", "35 years old", and "Male" will be printed.
If the patient classification is set as "Child", and no age and sex information have been
entered, "Child", "2 years old", and "Male" will be printed.
The output printer will be according to the setting made for "12-Lead Analysis Result"
([Bedside]/[Laser]) under [Manual Printing>Printer Sel. (Graphic Printing)]. (
Printing (Other Setup)" P9-6)
"Manual
Printed Data
The following basic data will be printed.
Heart Rate Heart rate obtained by basic arrhythmia measurement
QRS Interval QRS interval of basic waveform measurement. Average value of measurements of leads I to V6. The
equipotential part (I wave) at the beginning of QRS and the equipotential part (K wave) at the end of
QRS are not included in QRS interval.
QTc Interval QTc interval of basic arrhythmia measurement. This value is calculated from the following equation:
QRS Axis QRS axis of basic arrhythmia measurement. This value is calculated from the following equation:
where, I, II, and III are the sums of the maximum amplitude values (signed) of Q, R, S, R', and S'
waves from each lead.
ST Amplitude from the baseline. Measurement position: End of QRS wave + (QT/10) sec.
8-33
Chapter 8 Review Function 12-Lead Analysis (Optional Function)
The following is the output example when [Analyzed Report] key is pressed.
Pressing the [Waveform Report] key will print the analyzed waveform in a standard format.
8-34
Chapter 8 Review Function 12-Lead Analysis (Optional Function)
Pressing the [Analyzed Report] will print the analyzed result in a format set for "12-Lead Analysis Format"
under [Menu>Setup>Manual Printing]. (
"Manual Printing (12-Lead)" P9-4)
Pressing the [Waveform Report] will print the waveform in a format set for "12-Lead Waveform Format"
under [Menu>Setup>Manual Printing]. (
"Manual Printing (12-Lead)" P9-4)
The following is the output example when [Panorama Report] key is pressed.
NOTE
To print out the 12-lead analysis panorama report in color, use a laser printer with the
page description language in LIPS IV. If a printer with other page description language
is used, the printout will be in black and white.
8-35
Chapter 8 Review Function Full Disclosure Waveform (Optional Function)
By using the optional CFast card, 240 hours of wave data can be saved.
Up to six waveforms can be displayed. The alarm event and time will be also saved which allows to search the
waveform by each factor.
CAUTION
Use only the specified CFast card.
Turn OFF the power before removing the CFast card.
Check that the CFast card indicator is not lit in orange when turning OFF the power. When
using the CFast card, use the standby switch.
The CFast card can be used only on the device where it was formatted.
The CFast card formatted for the central monitor full disclosure waveform data cannot be
used on the DS-8400 System.
NOTE
When the full disclosure waveform data exceeds the capacity of the CFast card, the data will
be deleted from the old one.
To delete the full disclosure waveform data, perform the discharge procedure.
(
"Discharge" P5-7)
REFERENCE
To save the full disclosure waveform, the CFast card needs to be formatted for the full
disclosure waveform. (
Card" P3-4)
Maintenance Manual "Formatting the Full Disclosure Waveform
Waveform Setup
The displaying/printing waveform quantity and type of storing waveform, display duration (sec.) per line for the full
disclosure waveform can be preprogrammed.
2 4
3 5
6
8-36
Chapter 8 Review Function Full Disclosure Waveform (Optional Function)
NOTE
The maximum waveform quantity that can be printed differs depending on the output
printer.
4 Set the display duration per line from [10 sec.]/[30 sec.]/[1 min.].
5 Select the slide show interval from [1 sec.]/[5 sec.]/[10 sec.]/[20 sec.]/
[30 sec.].
NOTE
On the full disclosure waveform display, the arrhythmia occurrence point will be displayed
7 seconds before the actual arrhythmia occurrence.
5 Press [Print].
REFERENCE
The parameter selected for "Waveform" will be printed. The waveform quantity that can
be printed differs depending on the output printer. Maximum of 3 waveforms for the
bedside monitor printer, and maximum of 6 waveforms for the laser printer can be printed.
Press the desired waveform area. The Zoom Wave window will be displayed.
8-37
Chapter 8 Review Function Hemodynamics
To Search by Time
1 Press the [Time Search] key on the full disclosure waveform display.
2 Enter the searching date/time using the numeric keys and press the
[Search] key.
Hemodynamics
This section explains the procedure for hemodynamics calculation and printing.
NOTE
When the DS-5700 system central monitor is connected via DS-LAN II network, the system
will function as follows.
If the device is connected to DS-LAN, and [ON] is selected for "Synchronize
Hemodynamic Data with the Central Monitor", the latest 5 hemodynamic data will be
enabled and other hemodynamic data will be deleted. For the latest 5 data, the edited
hemodynamic data will synchronize between this monitor and the central monitor.
If the device is connected to DS-LAN, and [OFF] is selected for "Synchronize
Hemodynamic Data with the Central Monitor", the latest 5 data will be transmitted to the
central monitor, but the data will not synchronize between this monitor and the central
monitor. The data edited on the central monitor will be deleted. The data edited on this
monitor will be transmitted to the central monitor.
Calculation Data
This device can calculate and display the following parameters of hemodynamics.
Data Item Formula
h0.725xw0.425x71.84x10-4
BSA Body Surface Area (m2)
(Dubois Formula)
8-38
Chapter 8 Review Function Hemodynamics
NOTE
The blood pressure unit for hemodynamics is "mmHg". If the unit is "kPa" or "cmH2O", it will
be converted to "mmHg" when calculating.
3 [Print] key
2
The currently displayed hemodynamic data 3
will be printed.
8-39
Chapter 8 Review Function Hemodynamics
The hemodynamics calculation can be performed using the newly entered data.
The data can be entered manually using the numeric keys or automatically using the current data.
2
3
NOTE
If the height, weight, BSA is changed on the "Admit/Discharge" screen, average CI will
be recalculated. However, the hemodynamic will not be recalculated with the new CI
data.
Input Data
Data Item (Unit) Editing Range
ART M Mean Arterial Pressure (mmHg / kPa) 0 mmHg to 350 mmHg / 0 kPa to
46.6 kPa
8-40
Chapter 8 Review Function Hemodynamics
Input Data
Data Item (Unit) Editing Range
[Regist.]: The calculation will be performed using the newly entered data, and the entered data and
calculation result will be registered on the list.
REFERENCE
The calculation result will not be displayed if sufficient data is not entered.
Maximum of 10 data can be registered. If exceeded, the oldest data will be deleted.
The edited data will be also displayed in blue on the list.
The entered data which has been already calculated can be edited or deleted.
1 Press the [Menu], [Hemodynamics] ("Calculation"), and then the date/time display area for the data to edit.
8-41
Chapter 8 Review Function Lung Function
Lung Function
This section explains the procedure for lung function calculation and printing.
Calculation Data
AaDO2=PAO2-PaO2
Qs/Qt=(CćO2-CaO2)/(CćO2-CvO2)
Qs/Qt Shunt Rate (%)
CćO2=1.34xHb+0.003xPAO2
REFERENCE
The blood pressure unit for lung function calculation is "mmHg". If the unit is other than
"mmHg", it will be converted to "mmHg" when calculating.
3 [Print] key
2
The currently displayed lung function data
will be printed. 3
8-42
Chapter 8 Review Function Lung Function
The lung function calculation can be performed using the newly entered data.
The data can be entered manually using the numeric keys or automatically using the current data.
2
3
Press the [Latest Data] key to display the entered data of "HEIGHT", "WEIGHT", "CO".
Press the key for the editing data to display the numeric keys. Edit the data using the numeric keys, and
press the [Set] key.
NOTE
If the height, weight, BSA is changed on the "Admit/Discharge" screen, average CI will
be recalculated. However, the lung function calculation result will not be recalculated with
the new average CI.
Input Data
Data Item (Unit)
8-43
Chapter 8 Review Function Lung Function
[Regist.]: The calculation will be performed using the newly entered data, and the entered data and
calculation result will be registered on the list.
REFERENCE
The calculation result will not be displayed if sufficient data is not entered.
Maximum of 256 data can be registered. If exceeded, the oldest data will be deleted.
The edited data will be also displayed in blue on the list.
The entered data which has been already calculated can be edited or deleted.
1 Press the [Menu], [Lung Function] ("Calculation"), and then the date/time display area for the data to edit.
8-44
Chapter 8 Review Function Cardiac Output (CO)
This section explains about the cardiac output measurement using the thermodilution method, setup procedure for
catheter type, etc., and procedure for editing the measurement result.
2 Thermodilution Curve 4
8
3 Time Scale
5
4 Cardiac Index (CI)
5 Blood Temperature 6
6 Injectate Temperature 7
7 Status Message 1 2
8 Cardiac Output (CO)
8-45
Chapter 8 Review Function Cardiac Output (CO)
Before measuring the cardiac output, set the measurement condition such as ON/OFF of auto start, time scale for
thermodilution curve, injection condition, etc.
2
3
REFERENCE
Even when [ON] is selected, the measurement can be manually started by pressing
the [Start] key.
[Manual Input]: The computation constant for the used catheter can be manually input with the numeric
keys.
8-46
Chapter 8 Review Function Cardiac Output (CO)
4
2
1 3
REFERENCE
ARGON: Argon Medical Devices Japan, K.K. (formerly Becton, Dickinson and Company)
The manufacturer name can be changed on "Catheter Manufacturer for CC Input" setting
(Menu>Initial Settings>Meas.>Other).
2 Select the "Catheter Size (F) from [5]/ [6]/ [7]/ [7.5].
3 Select the "Injectate Volume (mL) " from [3]/ [5]/ [10].
When the above items are selected, the computation constant will be automatically set.
NOTE
If the CC value does not correspond to the used catheter, or to use the previous CC
value, press the [Cancel] key, and enter the value manually.
To automatically enter the computation constant, the catheter relay cable needs to be
connected.
8-47
Chapter 8 Review Function Cardiac Output (CO)
4
2
1
3
1 Select the catheter manufacturer from [BIOSENS]/[ARGON]/[EDWARDS].
If previously entered value is present, press the key for "History".
If the previously entered value is not present, enter the CC value using the numeric keys.
CO Measurement
NOTE
While "WAIT" is displayed, the measurement cannot be started. Wait until "READY" is
displayed.
8-48
Chapter 8 Review Function Cardiac Output (CO)
REFERENCE
If "Auto Start" is ON, the measurement will automatically start at injection by detecting the
blood temperature.
NOTE
When "WAIT" message is continuously displayed, verify that catheter relay cable is
properly connected to the cardiac output module, and thermodilution catheter is securely
connected.
Before injecting, check that the Ti (injectate temperature) setting is correct.
When repeatedly performing the measurement, inject at intervals of 30–60 seconds
The CI value will not be displayed unless height/weight or BSA value is input on the
"Admit/Discharge" screen.
(
"Entering the Patient Information" P5-1)
For the following cases, measurements may be inaccurate.
Shunt disease, tricuspid regurgitation or pulmonic regurgitation.
During exercise stress
As body temperature varies non-continuously and unevenly by exercise, constant CO
value cannot be measured.
Excessive Arrhythmia
As blood volume varies non-continuously due to arrhythmia, accurate CO value
cannot be measured.
8-49
Chapter 8 Review Function Cardiac Output (CO)
The average CO and average CI can be calculated by performing the CO measurement continuously and editing the
measurement result.
3
4
2 To Change the Selected Status
The selected data for the average value will be displayed in blue.
Press the graph area to change the selected status.
V Mark: VPC detected during CO measurement.
*: CO value exceeding the average CO value ±10%.
3 [Average CO Input]
NOTE
If the height, weight, BSA is changed on the "Admit/Discharge" screen, average CI will
be recalculated.
As the CI will not be recalculated after the hemodynamic calculation, save the average
CI by hemodynamic calculation before changing the height, weight, and BSA.
The [Delete Sel.] key will change to [Delete] key, and the data can be deleted.
x mark will be displayed for the data to be deleted, and pressing the [Delete OK] key will delete the data.
8-50
Chapter 8 Review Function Drug Calculation
Drug Calculation
3 4
If the weight is entered on the "Admit/Discharge" menu, the entered weight will be displayed.
NOTE
The weight can be changed on this menu, but the changed weight will be used only for
the drug calculation and will not be reflected on the "Admit/Discharge" menu.
The list of registered drugs will be displayed. Select the drug to administer to the patient.
When a drug is selected, the drug amount, diluent amount, dosing rate/unit preset for that drug under "Initial
Settings" will be automatically entered.
NOTE
The flow rate will be automatically calculated when the value for each item is updated.
8-51
Chapter 8 Review Function Drug Calculation
On the initial display of the drug calculation menu, the previous calculation data will be
displayed. The calculation data will be cleared when the patient is discharged.
To change the automatically entered value, press the key for each item and manually enter the value.
The dosing rate and flow rate can be adjusted by pressing the [+], [-] keys.
NOTE
If the selected unit for the dosing rate requires weight, the flow rate cannot be calculated
if the weight is not entered.
4 Press the [Update "End by"] key to update the estimated time of completion.
Pressing the [Update "End by"] key will recalculate the time from the pressed time and update the estimated
time of completion.
mg/min
mg/hr
mg/kg/min
mg/kg/hr
μg/min
μg/hr
μg/kg/min
μg/kg/hr
units/hr
IU/hr
mg/min
mg/hr
mg/kg/min
mg/kg/hr
μg/min
μg/hr
8-52
Chapter 8 Review Function Drug Calculation
μg/kg/min
μg/kg/hr
units/hr
IU/hr
Unit and Setting Range (Dosing Rate, Drug Amount, Diluent Amount, Flow Rate, Weight)
Dosing Rate Drug Amount
Drug Unit
Setting Range Unit Setting Range
(Selectable)
AMRINONE
AMINOPHYLLINE
BRETYLIUM
DOBUTAMINE
DOPAMINE
mg/min,
EPINEPHRINE mg/hr,
mg/kg/min,
ISOPROTERENOL mg/kg/hr,
mg
LIDOCAINE μg/min,
μg/hr,
NITROGLYCERIN μg/kg/min,
μg/kg/hr
NITROPRUSSIDE
NOREPINEPHRINE
PHENYLEPHRINE
0.01 to 1500000.00 0.01 to 1500000.00
PROCAINAMIDE
tPA
HEPARIN
units/hr units
INSULIN
STREPTOKINASE IU/hr IU
DRUG-A to G mg/min,
mg/hr,
mg/kg/min,
mg/kg/hr,
mg
μg/min,
μg/hr,
μg/kg/min,
μg/kg/hr
units/hr units
IU/hr IU
NOTE
The setting is not possible if it cannot be correctly calculated by the entered value.
8-53
Chapter 8 Review Function Other Bed Display
This section explains about the function to display the waveform and numeric data and to set alarms for other bedside
monitors.
The other bed alarm function generates the alarm sound for the other bed on this monitor. To use this function, wired
network (DS-LAN II or DS-LAN III) connection is required.
CAUTION
On the DS-LANII network system, maximum of 3 monitors (including the central monitor)
can display the data of this monitor using the other bed display function.
However, there is no restriction of numbers for the DS-7000 series central monitors and DS-
5700. These monitors will be counted as 1 monitor regardless of the numbers.
Ex. 1) In case of 1 central monitor and 5 bedside monitors (A to E):
The total number of monitors that can display the data of Bedside Monitor A is 3 monitors
which consist of 1 central monitor and 2 out of 4 bedside monitors (B to E).
Ex. 2) In case of 3 central monitors (DS-7000 series or DS-5700) and 5 bedside monitors
(A to E):
The total number of monitors that can display the data of Bedside Monitor A is 5 monitors
which consist of 3 central monitors and 2 out of 4 bedside monitors (B to E).
If the number of bedside monitors displaying the same bed exceeds the limit, the bedside
monitor with smaller ID will be prioritized.
If monitoring 12-lead waveform on the central monitor, the total numbers of monitors that can
display the same bed will be reduced by 1.
NOTE
This device cannot connect to a wired network of AU-5500N 8ch Recorder set as the
administrator.
Even if connected, other bed display, printing and other function cannot be used.
The other bed display can be accessed from the menu or from the preprogrammed user key.
Also, by setting the other bed alarm to [ON], [Other Alarm] will be displayed when other bedside monitor generates
an alarm.By pressing this [Other Alarm] key, the display for the other bed can be accessed.
2
4 6
5
3
On the other bed selection menu, select the bed to display from maximum of 100 beds
(in case of DS-LAN III) connected to the wired network. The Room / Bed ID for the alarm
8-54
Chapter 8 Review Function Other Bed Display
generating bed will be displayed in red. The other bed alarm generating bed will be
indicated by an icon inside the Room/Bed ID key.
1 3
5
2
7
6
1 Message Area
The message for the other bed will be displayed.
2 Waveform Display Area
Maximum of 6 waveforms for the DS-LAN III network, and maximum of 2 waveforms for the DS-LAN II
network can be displayed.
3 By pressing the [Other Bed Alarm Silence] key on the other bed display, the alarm sound for the displayed
bed can be silenced.
4 Pressing this key will switch ON/OFF of menu title display.
5 Numeric Data Area
The numeric data at the bottom of the screen can be switched by using the / keys.
6 Press the [Waveform Selection] key to select the waveforms.
Waveform 1 is fixed as ECG, but other waveforms can be selected.
Maximum of 6 waveforms for the DS-LAN III network, and maximum of 2 waveforms for the DS-LAN II
network can be displayed.
Select the waveform from the waveform selection window.
7 Press the [Numeric Selection] key to display [Numeric Data Selection] window. The parameters to display
on the right side of the screen can be selected.
8-55
Chapter 8 Review Function Other Bed Display
Press the [Alarm Display] key to change the screen to other alarm setup mode. When the mode is changed,
the [Alarm Display] key will be displayed in blue. To return to the original mode, press the [Alarm Display] key
again.
Select the bed to generate the other bed alarm.
Select the Room/Bed ID for the bed to generate the alarm. The selected bed will be indicated by blue
frame. To cancel the selection, press the key for the bed again.
[Select All], [Cancel All]: Selection/cancellation for all the beds can be performed at once.
[Enter]: The selection will be finalized.
All the beds connected to the network can be displayed, but it is also possible to divide the beds by areas,
which allows to display the beds by each area.
1
3
4
2
1 Press the key for "Area Setup" to change the screen to area setup mode. When the mode is changed, the
key for selected area will be displayed in blue. To return to the original mode, press the key again.
2 Select the Room/Bed ID for the bed to assign to the area. The selected bed will be indicated by blue frame.
To cancel the selection, press the key for the bed again.
[Select All], [Cancel All]: Selection/cancellation for all the beds can be performed at once.
[Enter]: The selection will be finalized.
3 Press the key for "Area Setup" to change the screen to area setup mode.
4 Press the [Area Name/Color] key.
8-56
Chapter 9 Printing Printing Setup
Chapter 9 Printing
Printing Setup
REFERENCE
The printed HR/PR data depends on the ECG/SpO2/BP selection for "Synchronized Mark/
Tone" under [Menu>Parameter>ECG (SpO2, BP)]. ( "Synchronized Mark/Tone Setup"
P7-10)
Under the following condition, the amplitude value will be printed for the ECG calibration
waveform.
*[Bar (10mm)] is set for "Waveform Size Display" under [Initial Settings>User I/F>Display/
Print].
*[ON] is set for "Print Calibration" under [Manual Printing>Common]
1 Press the [Menu], [Manual Printing] or [Auto Printing] ("Basic Setup") keys.
The manual printing can be set to start from the time the key is pressed, or 8 sec./16 sec. prior to the time the key is
pressed.
Also, the printing can be set to automatically stop after 24 seconds, or continue to print until the "Print Start/Stop"
key is pressed again.
The printer can be selected from bedside monitor printer or central monitor printer.
1 2
4 3
1 Printer
[Bedside]: Data will be printed on the HR-800 of the bedside monitor.
[Central]: Data will be printed on the central monitor printer.
9-1
Chapter 9 Printing Printing Setup
2 Waveform
On the "Select Wave" window, 3 waveforms can be selected for printing.
The key for the selected waveform will be displayed in blue.
3 Delay Time
[None]: Printing will start from the point the [Print Start/Stop] key is pressed.
[8 sec.] / [16 sec.]: Printing will start 8 sec. or 16 sec. prior from the point the [Print Start/Stop] key is pressed.
NOTE
If [None] is selected for the manual printing delay time, QRS classification symbol will
not be printed. To print the QRS symbol, set the delay time to [8 sec.] or [16 sec.].
The HR-800 can be inserted to the internal slot, or connected using the U-LINK cable.
Select which HR-800 to use from [Built-in] or [U-LINK] under [Initial Settings > System
> Other].
4 Print Duration
[24sec.]: Printing will automatically stop after 24 seconds.
[Cont.]: Printing will continue until the [Print Start/Stop] key is pressed again or until paper runs out.
REFERENCE
For print duration of recall enlarged waveform, refer to
"Zoom Wave" P8-24
To Start/Stop the Printing
Pressing this key during periodic printing, alarm printing, graphic printing, or recall printing will cease the
printing in process.
Inside the [Print Start/Stop] key, the output printer status for manual printing will be displayed.
Message Description
No Normal Operation
9-2
Chapter 9 Printing Printing Setup
6 7 8 9 10 11
1
2
3
4
5
12 13 14 15 16 17 18 19
10 Patient Classification
The 21-digit number printed at the bottom of the paper indicates the settings of the device. At the 14th digit from the
left, filter setting (AC filter, drift filter) is printed in hexadecimal number.
0 8 Drift Filter ON
1 9 Drift Filter ON
4 C Drift Filter ON
5 D
9-3
Chapter 9 Printing Printing Setup
The monitoring 12-lead waveform can be printed on the bedside monitor printer.The delay time is 6 seconds. The
12-lead waveform cannot be printed on the central monitor printer.
2
3
4
1
5
1 Waveform Format
[Regular]: Printing will start from the limb leads. (In the order of I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6)
[Reverse]: Printing will start from the chest leads. (In the order of V1, V2, V3, V4, V5, V6, I, II, III, aVR, aVL,
aVF)
2 Print Calibration
[ON]: Calibration waveform will be printed. If [Bar (10mm)] is set for "Waveform Size Display" under [Initial
Settings>User I/F>Display/Print], the amplitude value corresponding to the displayed waveform size will be
printed.
[OFF]: Calibration waveform will not be printed.
3 Position
[Center]: Equalizes the printing width of each lead so that the waveform baseline will be at the center. The
printing scale of the waveform will be also automatically adjusted.
[Proportional]: Equalizes the blank space between each lead to avoid overlapping of the waveforms. The
printing scale of the waveform will be also automatically adjusted.
[OFF]: Waveform position will not be adjusted when printing.
NOTE
The printer scale will be adjusted in the range of x1, x1/2, x1/4. It will not be adjusted to
x2 or x4 even if the amplitude is small.
9-4
Chapter 9 Printing Printing Setup
5 Lead Boundary
This setting will be displayed only when [Laser] is selected as the printer for "12-Lead Waveform", "12L
Analysis Result".
[ON]: Lead boundary between the leads will be printed.
[OFF]: Lead boundary will not be printed.
Length of
Output Example Waveform Layout Each
Waveform
3 waves x 4
First column:
I,II,III
1 2
Second column:
aVR, aVL, aVF
6 sec.
Third column:
V1, V2, V3
3 4 Fourth column:
V4, V5, V6
2 waves x 6
First column:
I,II
1 2
Second column:
III, aVR
Third column:
aVL, aVF
3 4 6 sec.
Fourth column:
V1, V2
Fifth column:
V3, V4
5 6 Sixth column:
V5, V6
9-5
Chapter 9 Printing Printing Setup
1 2
NOTE
The HR-800 can be inserted to the internal slot, or connected using the U-LINK cable.
Select which HR-800 to use from [Built-in] or [U-LINK] under [Initial Settings > System >
Other].
REFERENCE
Graphic printing is a printing performed from the data review screen such as graphic trend
and tabular trend.
To select laser printer, it is necessary to select [ON] or [DS-LAN] for "Network Printer"
under [Menu > Initial Settings > External Device > Network] in advance.
(
Maintenance Manual "Laser Printer Setup" P4-30)
2 Recall Printing
[Graphic Printing]: Recall data will be output on the printer selected for "Graphic Printing".
[Manual Printing]: Recall data will be output on the printer selected for "Printer" under "Basic".
9-6
Chapter 9 Printing Printing Setup
When numeric data alarm or arrhythmia alarm occurs, printing will automatically start.
NOTE
The alarm detection is performed each second, and if more than one alarm occurs at the
same time, one data will be stored according to the alarm priority.
Maximum of 3 alarm data can be stored. If more than 3 alarms generate, the higher priority
alarm will replace the previously stored lower priority alarm. The stored data will be deleted
once it is printed.
Priority of alarm printing factor ;
ASYSTOLE > VF > VT > Ext Tachy > Ext Brady > SLOW VT > TACHY > BRADY > RUN >
HR (HR / PR_SpO2 / PR_IBP) > APNEA > BP1 (or ART) > SpO2 >
NIBP > RR (RR_IMP / RR_CO2 / RR_GAS / RR_VENT) > EtCO2 >
GAS (CO2-E / CO2-I / AGT-E / AGT-I / O2-E / O2-I / N2O-I) > MAC > MV > PAUSE >
COUPLET > BIGEMINY > TRIGEMINY > FREQUENT > SVT > IRREGULAR RR >
PROLONGED RR > S FREQUENT > S COUPLET > VPC > SVPC > NOT CAPTURE
> NOT PACING > BP2 > BP3 > BP4 > BP5 > BP6 > BP7 > BP8 > ST > TEMP > Tb >
InspCO2 > SpCO > SpMet > SpHb > RR_SpO2 > PEAK > PEEP > BIS
1 3
2 4
5
1 Alarm Printing
3 Printer
4 Print Duration
(
"Manual Printing (Basic)" P9-1)
9-7
Chapter 9 Printing Printing Setup
NOTE
The delay time differs depending on the print duration.
The HR-800 can be inserted to the internal slot, or connected using the U-LINK cable.
Select which HR-800 to use from [Built-in] or [U-LINK] under [Initial Settings > System >
Other].
Delay Time
5 Waveform
(
"Manual Printing (Basic)" P9-1)
[Alarm]: Prints the waveform of the alarm factor.
1 3
2 4
5
1 Periodic Printing
[ON]: Printing will automatically start at fixed interval.
[OFF]: Turns OFF the periodic printing function.
2 Printer
[Bedside]: Data will be printed on the HR-800 of the bedside monitor.
[Central]: Data will be printed on the central monitor printer.
9-8
Chapter 9 Printing Printing Setup
REFERENCE
If [5 min.] is selected for [Interval], the time will be displayed in real time such as 10:00,
10:05, ...10:25. If [60 min.] is selected, it will be displayed as 10:00, 11:00, 12:00.
4 Print Duration
The printing will automatically stop after the selected duration.
5 Waveform
(
"Manual Printing (Basic)" P9-1)
The printing condition common for manual printing and automatic printing can be set.
1 2
4 3
1 QRS Classification
[ON]: QRS classification symbol will be printed with the ECG waveform.
Symbol Description
9-9
Chapter 9 Printing Printing Setup
Symbol Description
NOTE
The QRS symbol cannot be printed for manual printing if the "Delay Time" is set to
[None], and for periodic printing. To print the QRS symbol, set the "Delay Time" to [8
sec.] or [16 sec.] for manual printing.
The "S" (QRS symbol) will be printed as "N" on the central printer.
2 Printing Speed
[25mm/s]: The printing speed will be set to 25mm/s.
[50mm/s]: The printing speed will be set to 50mm/s.
4 Print Calibration
[Top]: Calibration waveform will be printed at the beginning of the waveform.
[Each Page]: Calibration waveform will be printed in 18.75 cm interval.
[OFF]: Calibration waveform will not be printed.
Freeze Printing
The waveform trace can be suspended and printed from 12 seconds prior to the point the waveform trace was
stopped.
The waveform selected for manual printing will be printed. The print duration is 12 seconds.
To freeze the waveform display, the [Freeze] key needs to be assigned as user key.
( "To Configure the Display" P10-4)
9-10
Chapter 9 Printing Printing Setup
When the display layout is "12-Lead", pressing the [12-Lead Print] key will start 12-lead waveform printing.
6Wavesx2*1 5 sec.
When printed on the laser printer: 10 sec.
3 wavesx4+Rhythm*1 12.5 sec.
12 Waves*2 10 sec.
*1 [CONTINUOUS]: The waveform output will be in the time sequence of waveform block order.
*2 [COHERENT]: The waveform output will be in the same time phase for all waveforms.
9-11
Chapter 9 Printing Printing Setup
9-12
Chapter 10 System Configuration Display Configuration
This section describes about the display configuration type and the procedure to configure the display.
The waveform/numeric data display can be configured according to the monitoring purpose. Basic display layouts
are as follows.
Standard
12-lead
Numeric Data/Bottom
When ECG cascade or block cascade is selected, a full disclosure waveform can be displayed. The user keys can be
also assigned to the numeric data area.
If extended board (optional) is equipped, up to 2 extended displays can be used. (extended display function)
Display Example
On this system, 9 main modes and 6 sub modes can be preprogrammed according to the monitoring purpose.
By registering the configuration to each mode, the display configuration setups at admittance of patient can be
simplified by just selecting one of the modes.
( "To Select the User Mode" P5-8)
It is recommended to program the display mode in rough classification such as patient’s condition, monitoring
purpose (ICU or surgery), and if necessary, perform unique setup for each patient.
10-1
Chapter 10 System Configuration Display Configuration
The numeric data to be displayed can be selected on the "Numeric Data Selection" window.
The parameters of the "Numeric Data Selection" window can be assigned to the numeric data box on the home
display.
( "Numeric Data Box Display (for each parameter)" P3-8)
Example: Page 1
HR/PR x ○ ○ ○ ○ ○ ○
HR x ○ ○ ○ ○ ○ ○
PR_SpO2 x ○ ○ ○ ○ ○ ○
PR_IBP ○ ○ ○ ○ ○ ○ ○
VPC, PACE x ○ ○ ○ ○ ○ ○
ST, VPC x ○ ○ ○ ○ ○ ○
BP1 to BP8 x ○ ○ ○ ○ ○ ○
NIBP x ○ ○ ○ ○ ○ ○
NIBP List x ○ ○ ○ ○ ○ ○
SpO2 x ○ ○ ○ ○ ○ ○
SpO2, PR x ○ ○ ○ ○ ○ ○
SpCO x ○ ○ ○ ○ ○ ○
SpMet x ○ ○ ○ ○ ○ ○
SpHb x ○ ○ ○ ○ ○ ○
Sp* x ○ ○ ○ ○ ○ ○
T1 to T8, Tb ○ ○ ○ ○ ○ ○ ○
VENT x x ○ ○ x ○ ○
10-2
Chapter 10 System Configuration Display Configuration
Height*2 H1 H1 H2 H3 H1 H2 H3
P-V, F-V x x ○ ○ x ○ ○
SvO2, CO x x ○ ○ x ○ ○
SvO2, CO, CI x x ○ ○ x ○ ○
BIS x ○ ○ ○ ○ ○ ○
AEP x ○ ○ ○ ○ ○ ○
INVOS x ○ ○ ○ ○ ○ ○
CO2 x ○ ○ ○ ○ ○ ○
O2 ○ ○ ○ ○ ○ ○ ○
N2O ○ ○ ○ ○ ○ ○ ○
Agent x ○ ○ ○ ○ ○ ○
Agent, N2O x ○ ○ ○ ○ ○ ○
GAS, SPIRO x x ○ ○ x ○ ○
SPIRO x x ○ ○ x ○ ○
HEMO x x ○ ○ x ○ ○
HEMO-I x x ○ ○ x ○ ○
STOPWATCH x ○ ○ ○ ○ ○ ○
VENT-A x ○ ○ ○ ○ ○ ○
VENT-B x ○ ○ ○ ○ ○ ○
Hemo/etc-A x ○ ○ ○ ○ ○ ○
Hemo/etc-B x ○ ○ ○ ○ ○ ○
*3
Extended Function-A x x ○ ○ x ○ ○*3
*1: For LC-8016TC, W1/2 is about 30 mm, W1 is about 60 mm, W2 is about 120 mm
*2: For LC-8016TC: H1 is about 16 mm, H2 is about 32 mm, H3 is about 48 mm (H1 is the same length as waveform areax2)
For LC-8018TC: H1 is about 17mm, H2 is about 36mm, H3 is about 55mm (H1 is the same length as waveform areax2)
10-3
Chapter 10 System Configuration Display Configuration
2
1 Layout
(
"Changing the Layout" P10-4) 1
2 Numeric Data 3
(
"Changing the Displayed Numeric Data"
P10-5) 4
5
3 Waveform 6
(
"Changing the Displayed Waveform" P10-7)
7
4 Sweep Speed 8
(
"Sweep Speed" P10-9)
5 Short Trend
(
"Short Trend Display" P10-8)
6 Zoom Wave
(
"Enlarged Waveform Setup" P10-10)
7 User Key
(
"User Key Setup" P10-10)
8 Detail Setup
(
"Detail Setup" P10-11)
3 Select the user key location from [Right], [Left], [Bottom], and select the number of columns.
When bottom 2 rows for "Right/Left & Bottom" is selected for display layout, "Bottom" cannot be selected.
If there are parameters which cannot be displayed due to display area,
"Delete Confirmation" window will be displayed.
(shown on right)
Pressing the [Set] key will set the layout with some parameters not
displayed.
Pressing the [Cancel] key will return to the "Layout" window.
10-4
Chapter 10 System Configuration Display Configuration
NOTE
For both [Type-1] and [Type-2], the waveform layout is equivalent to that when the [Same
as Numeric] key is pressed.
When [Auto] is selected for the display layout, the following changes are not possible.
Changing the Displayed Waveform
Changing the Displayed Numeric Data
Changing the short trend parameters
CAUTION
When performing the telemetry or wired network transmission, configure the display so
that the numeric data corresponding to the waveform is displayed. If not, the displayed
waveform or numeric data may not be transmitted.
NOTE
For HR/PR data, an alarm will be generated only for the current parameter displayed in
the HR/ PR numeric data box. The Parameter alarm will not be generated unless the data
is displayed.
The parameter for the HR/PR numeric data box can be selected by pressing the key for
"HR/PR" on the ECG, BP, SpO2 parameter setup window/floating window or by pressing
the [HR/PR] user key.
10-5
Chapter 10 System Configuration Display Configuration
3 2
4
By pressing the selected area again, the selection will be canceled.
To start again from the beginning, press the [Reselect Area] key.
Adjust the size of the selected area which is indicated by blue box.
On the continuous setup mode, the numeric data box area can be sequentially selected.
NOTE
The selected parameter may not be displayed depending on the combination of the
parameters and size.
In such case, "Size Error" will be displayed in numeric data area. Adjust the size.
(
"Numeric Data Selection" P10-2)
10-6
Chapter 10 System Configuration Display Configuration
CAUTION
When performing the telemetry or wired network transmission, configure the display so
that the numeric data corresponding to the waveform is displayed. If not, the displayed
waveform or numeric data may not be transmitted.
3
4 2
5
By pressing the selected area again, the selection will be canceled.
To start again from the beginning, press the [Reselect Area] key.
Adjust the size of the selected area which is indicated by blue box.
On the continuous setup mode, the waveform display area can be sequentially selected.
10-7
Chapter 10 System Configuration Display Configuration
1
2
3
NOTE
The short trend can be displayed when the numeric data layout is "Right"/"Right&Bottom"/
"Left"/"Left&Bottom"/"Bottom".
When 12-lead layout is displayed, ST value of each lead can be displayed in short trend.
4 [Same as Numeric]: The same parameters for the currently displayed numeric data will be set as the short
trend parameters.
5 [Same as Waveform]: The same parameters for the currently displayed waveform will be set as the short
trend parameters.
6 Press [Continuous Setup] to switch to continuous setup mode. On the continuous setup mode, the short
trend display area can be sequentially selected.
NOTE
The [Change] key will be displayed when [User Setup] is selected for "Short Trend"
(Display Config.>Detail Setup).
[Same as Numeric], [Same as Waveform] will be applied for the displayed parameters at
the point when the key is pressed. The short trend parameters will not automatically
change when the displayed parameters are changed.
10-8
Chapter 10 System Configuration Display Configuration
3 When [ON] or [Overlap] is selected, set the time span. The selectable duration differs depending on the short
trend data resolution and display width (7 levels).
0 1 2 3 4 5 6
2 The trend display time will change to the time of the pressed position.
NOTE
When an alarm is generated for the recall alarm factor, recall screen will be displayed.
When the cursor function is enabled, a cursor will be displayed. The display duration can
be changed under "Short Trend" (Menu > Display Config.)
Sweep Speed
The sweep speed can be set with the following procedure. The sweep speed can be set differently for the circulatory
system waveforms (ECG, BP) and respiratory system waveforms.
10-9
Chapter 10 System Configuration Display Configuration
NOTE
When the sweep speed is set to [50 mm/s], "Zoom" cannot be set to [ON].
Scale will not be enlarged.
3 2
4
5
By pressing the selected area again, the selection will be canceled.
To start again from the beginning, press the [Reselect Area] key.
Adjust the size of the selected area which is indicated by blue box.
3 Select the function to assign to the user key on the "User Key Selection" window.
NOTE
The displayed user key can be switched between 2 displays using the [User Key Up] and
[User Key Down] keys.
Press the
15)
/ keys to switch the user key selection. (
"User Key Selection" P10-
10-10
Chapter 10 System Configuration Display Configuration
Detail Setup
1 Press the key for "Detail Setup".
5 13
14
6
9 10
15 19
16
17
18
20 24
21 25
22 26
23 27
2 At Alarm Occurrence
The numeric data display format at alarm occurrence can be selected.
[Reversed]: The numeric data will be displayed in reversed color at alarm occurrence.
[3D]: The numeric data will be displayed in 3D at alarm occurrence.
3 Grid
The ECG waveform can be displayed on the grid.
[ON]: Grid will be displayed.
[Bold]: Grid will be displayed in bold format.
[OFF]: Grid will not be displayed.
REFERENCE
Short trend and grid cannot be displayed overlapped.
10-11
Chapter 10 System Configuration Display Configuration
4 Scale
The scale can be selected from [ON]/[Bold1]/[Bold2].
5 Thickness
The thickness of the displayed waveforms can be selected from [Thin] / [Regular] / [Thick].
6 Clip
Whether or not to clip the overlapped waveforms of the neighboring display area can be selected.
8 Fill O2 Waveform
Whether or not to fill in the O 2 waveform from the baseline can be selected.
10 BP Overlap
The overlapping BP waveforms can be set for each overlap group 1 to 3.
11 RR Overlap
The overlapping RR waveforms can be set.
12 12-Lead ST Wave
The ST waveform to be displayed for the 12-Lead layout can be set.
[Ref.]: The ST reference waveform will be displayed.
[Average]: The average waveform will be displayed.
15 Block Cascade
The waveform combination for block cascade display can be set.
17 Extended Function
[Graphic/Tabular Trend]: Graphic trend and tabular trend will be displayed in the waveform display area.
[OCRG]: OCRG will be displayed in the waveform display area.
NOTE
When the extended function is selected, the waveform set to the same display area with
the graphic/tabular trend will not be displayed.
NOTE
OCRG size for the extended function is fixed to [small].
10-12
Chapter 10 System Configuration Display Configuration
20 Short Trend
The short trend parameters can be linked to the displayed numeric data or waveform.
[Link with Numeric]: The short trend layout will be linked to the displayed numeric data on the home display.
[Link with Waveform]: The short trend layout will be linked to the displayed waveform on the home display.
[User Setup]: User settings will be applied for the short trend layout.
22 Display Parameter
Whether or not to display the parameter name of the displayed short trend can be set.
[ON]: Displays the parameter name with the corresponding color of the parameter.
[Gray]: Displays the parameter name in gray.
[OFF]: Parameter name will not be displayed.
NOTE
The reference line function cannot be used for the overlapped short trend display.
When [Enable] is selected, the function to highlight the alarm generated data cannot
be used.
24 Cursor Function
Whether or not to display a cursor can be selected. By displaying a cursor, the measured data and review
data (tabular trend/graphic trend/zoom wave) at the time of cursor position can be displayed.
[Enable]: The cursor function will be enabled. However, the function to enlarge/reduce the display duration
by pressing the short trend area will be disabled.
[Disable]: The cursor function will be disabled.
NOTE
The cursor function cannot be used for the overlapped short trend display.
When [Enable] is selected, the function to highlight the alarm generated data cannot
be used.
The cursor will be displayed when the short trend area is pressed, and will be
automatically cleared after a short while.
25 Cursor Linkage
When [Enable] is selected for "Cursor Function", the review data to be displayed can be selected from
[Tabular Trend] / [Graphic Trend] / [Zoom Wave].
The zoom wave can be displayed only when the full disclosure waveform function is enabled.
10-13
Chapter 10 System Configuration Display Configuration
NOTE
If the numeric data box is configured at the bottom of display, user keys cannot be
assigned to the numeric data box area.
After configuring the display, make sure to verify the configured display by pressing the
[Home] key.
To maintain the configured display even after the power is turned OFF or after the
discharge procedure, store the configuration to one of the user modes, or select [Backup]
for "Display Configuration" under Initial Settings>User I/F>At Power ON/At Discharge.
(
"To Select the User Mode" P5-8)
Waveform Selection
1 2 2 3 4
Page 1 Page 2 Page 3
1 ECG1 to ECG12
The ECG waveform of the specified channel will be displayed. Minimum of 2 blocks are required to display the
ECG waveform.
3 BP Overlap 1 to 3
The BP waveform (BP1 to BP8) set on "BP Overlap Setup" will be displayed.
If the waveform display area is too small to display the assigned BP waveforms, it will be displayed in the
priority from smaller channel numbers.
4 RR Overlap 1 to 3
The RR waveform (CO2, O2, Agent) set on "RR Overlap Setup" will be displayed.
If the waveform display area is too small to display the assigned waveforms, it will be displayed in the priority
10-14
Chapter 10 System Configuration Display Configuration
of CO2>O2>Agent.
5 Block Cascade
The waveforms (2 to 6) set on the "Block Cascade Setup" will be displayed in one block.
Other than the waveforms explained above, the selected waveform on the "Waveform Selection Window" will be
displayed.
The user keys can be set on the "User Key Selection" window.
This section explains the function for each user key.
Example: Page 1
Page 1
OFF Blank key will be displayed.
Home The display will return to the home display. The [Home] key is also available as fixed key.
Menu The menu screen will be displayed. The [Menu] key is also available as fixed key.
Minimize Window Pressing this key will minimize the currently displayed window and will be stored to the user
key.
User Key The first and second page of the user key area will switch.
This key will be located at the same position for both first and second page.
Alarm Silence Alarm sound will be suspended for fixed amount of time. The [Alarm Silence] key is also
available as fixed key. By pressing the key for more than 3 seconds while the alarm is not
generated, it will bring the system to "Alarm Sound Suspend" condition.
Alarm Suspend Alarm (sound and display) will be suspended for fixed amount of time.
Freeze Waveform trace will freeze for fixed amount of time. Pressing the [Print Start/Stop] key while
in freeze condition will print the frozen waveform. Holding down the key will start the
waveform trace again.
Key Lock Touch key operation will turn ON/OFF. It can be used when cleaning the display panel.
10-15
Chapter 10 System Configuration Display Configuration
HR/PR The HR/PR numeric data box will be switched between HR and PR.
Page 2
HR/PR Source The parameter for HR/PR Source will be automatically selected.
Scale (Extended Display) The extended display waveform size/scale setup menu will be displayed. This setting can
be performed only when [OFF] is set for "Sync wave size/scale of extended display with
main unit".
GAS Display ON/OFF Multigas unit data display will turn ON/OFF.
Auto Display Config. The display will be automatically configured with the currently measured parameters.
Enlarged Display For the standard display layout, the numeric data box width will change.
Graphic/Tabular Trend ON/OFF The graphic/tabular trend display will turn ON/OFF.
Trend (Group) List of trend groups will be displayed, and selecting a trend group will display the graphic
trend.
Tabular Trend The tabular trend will be displayed.
Tabular Trend (Group) List of tabular trend groups will be displayed, and selecting a trend group will display the
tabular trend.
Page 3
10-16
Chapter 10 System Configuration Display Configuration
Alarm Setup (All) Alarm settings for all parameters will be displayed.
Alarm Setup (Basic) Alarm settings for basic parameters will be displayed.
Page 4
Main Mode 1(Initial) Main mode 1 (Initial) will be set as the monitoring mode.
Main Mode 2 (Hemo.) Main Mode 2 (Hemo.) will be set as the monitoring mode.
Main Mode 3 (Cardiac) Main Mode 3 (Cardiac) will be set as the monitoring mode.
Main Mode 4 (Local) Main mode 4 (Local) will be set as the monitoring mode.
Main Mode 5 (Full) Main mode 5 (Fulll) will be set as the monitoring mode.
Main Mode 6 (Heart) Main mode 6 (Heart) will be set as the monitoring mode.
Main Mode 7 (Neo.) Main mode 7 (Neo.) will be set as the monitoring mode.
Main Mode 8 (Recovery) Main mode 8 (Recovery) will be set as the monitoring mode.
Main Mode 9 (Cardiac) Main Mode 9 (Cardiac) will be set as the monitoring mode.
Sub Mode 1 (Induct.) Sub Mode 1 (Induct.) will be set as the monitoring mode.
Page 5
Sub Mode 2 (Surgery) Sub Mode 2 (Surgery) will be set as the monitoring mode.
Sub Mode 3 (Waking) Sub Mode 3 (Waking) will be set as the monitoring mode.
Sub Mode 4 (12-Lead) Sub Mode 4 (12-Lead) will be set as the monitoring mode.
WARNING
After changing the mode, make sure that the monitoring setting is appropriate.
When the mode is changed, patient classification, alarm settings, etc. will be changed.
NOTE
If the [Minimize] key is not set to the user key area, pressing the (Minimize) key on
the window will function the same as the (Close) key.
10-17
Chapter 10 System Configuration Display Configuration
When using the LC-8016TC/LC-8018TC Display Unit, holding down the home display area will change the mode
to display layout change mode.
On the display layout change mode, the following layout change can be performed.
Changing the displayed position of the waveform/numeric data
Changing the size of the waveform/numeric data
Adding the waveform/numeric data
Deleting the waveform/numeric data
10-18
Chapter 10 System Configuration Tone/Volume
Tone/Volume
This section explains the tone/volume setup procedure for alarm sound, HR synchronized tone, key sound, and boot/
shutdown sound. The tone/volume setup screen also allows to turn OFF the ventilator alarm sound.
The volume of the sound which notifies the completion of BP zero balance and NIBP measurement can be adjusted
on "Other" setting.
NOTE
The tone setup for the synchronized tone is effective only for HR and BP synchronized tone.
The tone for SpO2 synchronized sound will change according to the SpO2 value. The tone
will increase as the SpO2 value increases, and vice versa.
When the DS-8400 is activated with the DS-8007 connected and soon enters into standby
mode, a shutdown sound may generate on the DS-8007.
The start-up sound will always generate for the DS-8400 even when the "Boot/Shutdown"
sound is set to minimum level.
WARNING
Changing the setting for "Alarm System" (Initial Settings > Alarm) will also change the
alarm volume and tone setting. Make sure to check the volume and tone when the setting
is changed.
CAUTION
Pay attention not to set the alarm volume too low to avoid missing any important alarms.
The alarm sound for ECG, SpO2, CO2 will be different from the test sound. The set
volume will be applied but the set tone will not be applied to these parameters.
When [Standard Tone] is set for the "Alarm System", the alarm volume and tone for the
ventilator alarm and device status alarm will be the same with that of the vital alarm.
REFERENCE
The volume above the set minimum volume can be set.
(
Maintenance Manual "Alarm Related Setup" P5-4)
10-19
Chapter 10 System Configuration Color
2 Press .
The volume will be adjusted.
REFERENCE
The order of alarm priority is Urgent (H) > Caution (M) > Status (L).
The volume is also set according to the alarm priority.
The volume for high priority alarm cannot be set lower than the lower priority alarm,
and vice versa.
NOTE
The tone selection is different for the synchronized tone, alarm sound, and key sound.
For the "Sync. Tone", [Selected Tone] will generate the HR synchronized tone with the
selected tone. [Sync. With SpO2 Value] will generate the HR synchronized tone with the
same tone with the SpO2synchronized tone. If the SpO2 value is invalid, [Tone 2] will be
applied.
Color
In this section, setup procedure for the color of background, numeric data, waveform is explained.
The colors of the background, numeric data, waveform, user key can be customized.
The colors can be customized according to the various monitoring scene such as recognizable colors from a far
distance or colors which will not strain your eyes by the long time monitoring.
2
3
3 4
10-20
Chapter 10 System Configuration Color
2 Press the key for "Background", and set the background color.
5 Select a color.
The selected color for the parameter will be applied to the waveform, numeric data, graphic trend, and
tabular trend.
10-21
Chapter 10 System Configuration Brightness
Brightness
CAUTION
This device utilizes LED for the backlight. Since this LED deteriorates by the life cycle, the
display may become dark, scintillate, or may not light by the long term use. In such case,
contact your nearest service representative.
Night Mode
10-22
Chapter 10 System Configuration Night Mode
During the night mode, "Night Mode Active" message will be displayed.
NOTE
When the timer is set, the night mode will automatically start at the set "Start Time".
Night Mode
The time to start and end the night mode, and the night mode display can be set.
3 6
4
5
10-23
Chapter 10 System Configuration Night Mode
2 Set the "Start Time" and "End Time" for the night mode.
[Manual]: The night mode can be turned ON or OFF manually using the user key.
[Timer]: The night mode will automatically turned ON or OFF at the preprogrammed time.
The night mode can be manually turned ON from the user key or from the remote control unit even when
the [Timer] is set.
WARNING
When selecting [Silence], pay attention not to miss any important alarm by
simultaneously monitoring the bed on other monitors such as central monitor.
WARNING
When selecting [Time Only], pay attention not to miss any important alarm by
simultaneously monitoring the patient on other monitors such as central monitor.
[ON]: The alarm indicator will light even during the night mode.
[OFF]: The alarm indicator will not light during the night mode.
10-24
Chapter 10 System Configuration Night Mode
10-25
Chapter 10 System Configuration Night Mode
10-26
Chapter 11 Troubleshooting Message List
Chapter 11 Troubleshooting
Message List
CAUTION
The arrhythmia alarm message other than Tachy, Brady, Ext Tachy, Ext Brady will continue
to be displayed for 30 seconds after the alarm is resolved.
The alarm level shown below is the standard level set by Fukuda Denshi.
The alarm level can be changed under "Initial Settings > Alarm Setup > Alarm Level" .
<VF>
<VT>
<Slow VT>
<Tachy>
<Brady>
<Ext Tachy>
<Ext Brady>
11-1
Chapter 11 Troubleshooting Message List
<Upper HR Alarm>
<Upper PR Alarm>
<Upper MV Alarm>
BIS (When HBX-800 is used) <Lower BIS Alarm>
Arrhythmia <Pause>
<Run>
11-2
Chapter 11 Troubleshooting Message List
<Lower Tb Alarm>
Arrhythmia <Couplet>
<Bigeminy>
<Trigeminy>
<Frequent>
<Triplet>
<R on T>
<Multiform>
<Vent. Rhythm>
<SVT>
<Irregular RR>
<Prolonged RR>
<S Frequent>
<S Couplet>
<VPC>
<SVPC>
*1: # indicates the channel number of BP, TEMP, SpCO, SpMet, SpHb.
For SpCO, SpMet and SpHb, N1/N2/M1/M2/HR/HL/FR/FL/OT will be displayed for #.
*2: When the numeric data acquired from FLOW-i, GE or Draeger is displayed, alarm will not generate. The alarm will not
generate on the central monitor either.
11-3
Chapter 11 Troubleshooting Message List
Notification Alarm
Measuring Parameters Message
NOTE
(xxx sec) of the <Alarm Suspend (xxx sec)> message indicates the remaining time of alarm
suspended duration.
(xxx min.) of the <Alarm Silence (xxx min.)> message indicates the remaining time of alarm
sound suspended duration.
The <ARRHY OFF> message will be displayed when the Asystole, VF, VT, Slow_VT,
TACHY, BRADY, EXT TACHY, EXT BRADY and HR alarm is OFF.
WARNING
The SpO2 respiration measurement is not intended for use as an APNEA monitor.
CAUTION
The arrhythmia alarm message other than Tachy, Brady, Ext Tachy, Ext Brady will continue
to be displayed for 30 seconds after the alarm is resolved.
The alarm level shown below is the standard level set by Fukuda Denshi.
<Upper HR Alarm>
Pulse Rate (SpO2) <Lower PR Alarm>
<Upper PR Alarm>
<Upper PR Alarm>
11-4
Chapter 11 Troubleshooting Message List
<Apnea>
*1
Gas <Lower CO2-E Alarm>
Arrhythmia <Asystole>
<VF>
<VT>
<Slow VT>
<Tachy>
<Brady>
<Run>
<Ext Tachy>
<Ext Brady>
11-5
Chapter 11 Troubleshooting Message List
Blood Pressure <Lower BP2 to 8 Alarm> or <Lower (label other than ART) Alarm>*1
<Lower Tb Alarm>
Arrhythmia <Pause>
<Couplet>
<Bigeminy>
<Trigeminy>
<Frequent>
<Triplet>
<R on T>
<Multiform>
<Vent. Rhythm>
<SVT>
<Irregular RR>
<Prolonged RR>
<S Frequent>
<S Couplet>
<VPC>
<SVPC>
*1: # indicates the channel number of BP, TEMP, SpCO, SpMet, SpHb.
For SpCO, SpMet and SpHb, N1/N2/M1/M2/HR/HL/FR/FL/OT will be displayed for #.
*2: When the numeric data acquired from FLOW-i, GE or Draeger is displayed, alarm will not generate. The alarm will not
generate on the central monitor either.
11-6
Chapter 11 Troubleshooting Message List
Notification Alarm
Measuring Parameters Message
NOTE
(xxx sec) of the <Alarm Suspend (xxx sec)> message indicates the remaining time of alarm
suspended duration.
(xxx min.) of the <Alarm Silence (xxx min.)> message indicates the remaining time of alarm
sound suspended duration.
The <ARRHY OFF> message will be displayed when the Asystole, VF, VT, Slow_VT,
TACHY, BRADY, EXT TACHY, EXT BRADY and HR alarm is OFF.
<VENT COMM> 1
GAS
<GAS Unit Failure> 1
(MGU-800/MGU-810)
<BISx Incompatible> 3
11-7
Chapter 11 Troubleshooting Message List
BIS (When HBX-800 is used) <Check BIS Sensor, Perform Sensor Check> 3
<IB-8000-# Failure>*2 3
*1: On "Initial Settings" menu, the alarm level can be selected from Level M, L, N (Notification). (Default: Level M)
If [Alarm Silence] key is pressed during Level M, L alarm generation, the alarm level will change to Level N
(notification).
# indicates an error code.
*2: # indicates the input box number.
*3: # indicates the input box number, and the slot number of input box.
11-8
Chapter 11 Troubleshooting Message List
<Cannot Analyze> 1
<ECG Artifact> 3
11-9
Chapter 11 Troubleshooting Message List
<CO Disconnected> 3
<CO2 Disconnected> 3
<Fan Failure> 3
11-10
Chapter 11 Troubleshooting Message List
*6: On "Initial Settings" menu, the alarm level can be selected from Level M/L/N. (Default: Level L)
If [Alarm Silence] key is pressed during Level M/L alarm generation, the alarm level will change to Level N
(notification).
*7: # indicates the input box number.
*8: # indicates the input box number, and the slot number of input box.
*9: On "Initial Settings" menu, the alarm level can be selected from Level M/L. (Default: Level L)
NOTE
<NIBP meas. failed>, <Check NIBP cuff, hose>, <Connector Off>, <ECG Only 5 electrodes
are used.>, <Check xx Conn.>, <Check xx Comm.>, <SPIRO Check Flow Sensor> alarms
will be canceled when [Alarm Silence] key is pressed. Pay attention not to cancel the
important alarm.
Notification Alarm
Item Message Delay Time (sec.)
*1
Operation <Waveform Frozen (xxsec.)> 1
*1
<Key Locked (xxsec.)> 1
<Oxygenator Mode> 1
<GAS Zeroing> 1
<SPIRO Zeroing> 1
11-11
Chapter 11 Troubleshooting Message List
<BIS Noise> 3
<Check Cassette>*6 3
*6
Central Printer <Check Paper (Central)> 3
*6
<Check Cassette> 3
<Some parameters are not displayed due to the display layout setting.> 3
<Upload Standby> 1
11-12
Chapter 11 Troubleshooting Message List
HR
Message
<Unit Failure>
<Upper HR Alarm>
<Lower HR Alarm>
<Lower ST Alarm>
<Upper ST Alarm>
<Cannot Analyze>
<Check Electrodes>
<Out of Range>
<Low Amplitude>
<Noise Interference>
<Artifact>
ST
Message
<Lower ST Alarm>
<Upper ST Alarm>
BP1 to 8
Level H for BP1 and ART, Level M for other label
Message
<Lower BP Alarm>
<Upper BP Alarm>
<Zero Required>
<Out of Range>
<Out of Range>
11-13
Chapter 11 Troubleshooting Message List
NIBP
If <NIBP Meas. Error> is displayed, the message can be canceled by pressing [Cancel Error] on the NIBP setup
screen, [NIBP Start/Stop] key (user key), or HS-8000/DS-8007 [NIBP START/STOP] key.
If the same message is repeatedly displayed, a failure of the device can be considered. Cease the measurement, and
contact your nearest service representative.
( "<NIBP Unit Error (E**-**)> is displayed." P11-39)
Message
<Measurement Failed.>
<Initializing>
<Out of Range>
<Unit Failure>
<Replace Sensor>
<Check Sensor Attach.>
<Motion Artifact>
<Pulse Search>
<Replace Sensor>
<Low Confidence>
<Pulse Search>
<Noise Interference>
<Check Sensor>
<Replace Cable>
11-14
Chapter 11 Troubleshooting Message List
Message
<Check Cable>
<Zeroing>
<Low Confidence>
<Unit Failure>
<RR Interference>
<Unable to calculate>
<Calculating>
<Outside Range>
<Out of Range>
PR-SpO2
Message
TEMP1 to 8
Message
<Unknown Sensor>
<Out of Range>
Tb
Message
<Lower Tb Alarm>
<Upper Tb Alarm>
<Out of Range>
RR (Impedance)
Message
<Apnea Alarm>
<Upper RR Alarm>
<Lower RR Alarm>
11-15
Chapter 11 Troubleshooting Message List
Message
<CVA detected>
<Out of Range>
<Suspended>
RR (Ventilator)
Message
<Apnea Alarm>
<Upper RR Alarm>
<Lower RR Alarm>
RR (Gas)
Message
<Apnea Alarm>
<Upper RR Alarm>
<Lower RR Alarm>
<Out of Range>
<Zeroing>
<Warming Up>
<Unknown Sensor>
<Out of Range>
CO2 (HCP-810/HCP-820)
Message
<Initializing>
<Zeroing>
<Perform calibration.>
<Out of Range>
<Upper CO2-E>
<Lower CO2-E>
<Upper CO2-I>
11-16
Chapter 11 Troubleshooting Message List
<Upper RR Alarm>
<Lower RR Alarm>
<Apnea Alarm>
<GAS Zeroing>
11-17
Chapter 11 Troubleshooting Message List
<SPIRO Zeroing>
<Upper RR>
<Lower RR>
<Apnea>
<Upper MV>
<Lower MV>
<Upper PEAK>
<Lower PEAK>
<Upper PEEP>
<Lower PEEP>
<Check Sensor>
<Expired Sensor>
<Invalid Sensor>
<High Impedance>
<Artifact>
<Lead OFF>
<Artifacts>
<BISx Failure>
<BISx Incompatible>
11-18
Chapter 11 Troubleshooting Message List
WARNING
When the VELIA, ASTRAL, VS ULTRA ventilator is connected, and the ventilator power is
turned OFF, alarm will not generate on the DS-8400. If the connection cable is disconnected
from the ventilator, <Vent. Alarm> will generate, but <VENT_COMM> alarm will not
generate on the DS-8400.
The ventilator alarm sound is set to OFF (factory default).
The alarm sound can be turned ON on the "Tone/Volume" menu. (
19)
"Tone/Volume" P10-
CAUTION
For the ventilators other than Servo ventilators, ventilator alarm factor will not be notified to
the central monitor.
Depending on the central monitor type and software version, ventilator alarm factor may not
be displayed. For details of the central monitor type and software version, refer to your
nearest service representative.
The ventilator alarm factors are displayed only on the central monitor. These will not be
displayed on the bedside monitor.
11-19
Chapter 11 Troubleshooting Message List
Status Message
Message Details
Result Status
The result status will be displayed for 30 seconds after completion of measurement.
Message Details
• After the injection, the blood temperature is out of the measurement range.
• The thermistor connector and relay cable are not securely connected.
• The thermistor connector and relay cable are not securely connected.
• The blood temperature has not returned to stable condition after the measurement.
• The thermistor connector and relay cable are not securely connected.
• The sensor or relay cable is defective.
• The thermistor connector and relay cable are not securely connected.
11-20
Chapter 11 Troubleshooting Troubleshooting
Troubleshooting
ECG
Cause 2
Even though the "3-lead Override" (ECG Parameter Setup) is set to [ON], electrodes other than L, R, F are
connected.
Solution
Set the "3-lead Override" to [OFF].
Or, detach the electrodes other than L, R, F.
NOTE
Using 4-electrode or 5-electrode instead of 3-electrode allows more accurate QRS
detection.
Cause 2
The electrode contact is poor.
Electrical blanket or other noise source is near the patient.
Solution
Attach the electrodes firmly. Or, replace the electrodes.
If the lead cable or relay cable is defective (wire break, etc.), replace it.
If any noise source is near the patient, move it away from the patient as far as possible.
Cause 3
Even though the "3-lead Override" (ECG Parameter Setup) is set to [ON], electrodes other than L, R, F are
connected.
Solution
Set the "3-lead Override" to [OFF].Or, detach the electrodes other than L, R, F.
11-21
Chapter 11 Troubleshooting Troubleshooting
Cause 2
EMG is interfering.
Solution
Change the electrode site to a location where the myoelectricity will be less likely to interfere.
Select ESIS for the filter mode.
CAUTION
Selecting ESIS for the filter mode will decrease the QRS amplitude and may result in not
counting the heart rate.
Cause 3
Even though the "3-lead Override" (ECG Parameter Setup) is set to [ON], electrodes other than L, R, F are
connected.
Solution
Set the "3-lead Override" to [OFF].Or, detach the electrodes other than L, R, F.
Cause 1
Electrode is detached.
Solution
Place the electrodes again.If the electrode contact is poor, replace the electrode.
(
"Before Attaching the Electrodes" P7-2)
(
"Electrode Placement" P7-3)
Cause 2
The lead cable is disconnected from the electrode terminal.
Solution
Securely connect the lead cable.
REFERENCE
If the error persists, wire break of the lead cable or relay cable can be considered. Contact
your nearest service representative.
11-22
Chapter 11 Troubleshooting Troubleshooting
Cause 2
Even though the "3-lead Override" (ECG Parameter Setup) is set to [ON], electrodes other than L, R, F are
connected.
Solution
Set the "3-lead Override" to [OFF].Or, detach the electrodes other than L, R, F.
CAUTION
Using 4-electrode or 5-electrode instead of 3-electrode allows more accurate QRS
detection.
Also, if large amount of noise is interfering, the noise may be erroneously detected as
QRS.Change the electrode site and increase the ECG amplitude.
Solution 2
Increase the displayed waveform size. By increasing the waveform size, small QRS wave will become detectable.
However, noise may be also detected.
11-23
Chapter 11 Troubleshooting Troubleshooting
Cause 2
Even though the "3-lead Override" (ECG Parameter Setup) is set to [ON], electrodes other than L, R, F are
connected.
Solution
Set the "3-lead Override" to [OFF].Or, detach the electrodes other than L, R, F.
Cause 2
"Pacemaker Pulse" is set to [OFF] (ECG Parameter Setup).
Solution
Select [ON] for "Pacemaker Pulse" .
Cause 3
The electrode attachment site is not appropriate.
Solution
Check the electrode attachment site.
(
"Before Attaching the Electrodes" P7-2)
(
"Electrode Placement" P7-3)
11-24
Chapter 11 Troubleshooting Troubleshooting
11-25
Chapter 11 Troubleshooting Troubleshooting
Respiration
Green
Red Yellow
Cause 2
The electrode contact impedance is high.
Solution 1
Reattach the electrode. If the electrode contact is poor, replace the electrode.
(
"Before Attaching the Electrodes" P7-2)
(
"Electrode Placement" P7-3)
Solution 2
Change the lead for respiration measurement.
11-26
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The impedance respiration measurement is ceased.
Solution
Select [ON] for "Impedance Measurement" on "Admit/Discharge" or "RESP" setup screen.
CAUTION
If the pacemaker with the minute ventilation measuring function is used, turn OFF the
impedance respiration measurement. Otherwise, both the pacemaker and this monitor will
not be able to perform accurate measurement.
11-27
Chapter 11 Troubleshooting Troubleshooting
Solution 2
Connect the transducer.
Solution 3
The BP relay cable or transducer may be defective. Replace the BP relay cable or transducer.
11-28
Chapter 11 Troubleshooting Troubleshooting
Cause
The sensor is detached from the patient.
Solution 1
Check if the sensor is properly attached to the patient.
Solution 2
Check that the light emitting and receiving parts of the sensor LED are aligned.
11-29
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The sensor has not been attached long enough to obtain stable measurement.
Solution
After the sensor attachment, wait and see for about one minute until the waveform stabilizes.
Cause 1
The amplitude of the pulse waveform is too low, or the sensor is not positioned correctly.
Solution
Check that the light emitting and receiving parts of the sensor LED are aligned.
Cause 2
The sensor is defective.
Solution
Replace the sensor.
Cause 3
SpO2 sensor is not firmly connected to the connector.
Solution
Make sure the SpO2 sensor is firmly connected.
Cause 4
Sensor is exposed to light.
Solution
Place a black or dark cloth over the sensor to avoid direct sunlight. When not using the sensor for measurement,
avoid placing the sensor in light or unplug the sensor from the connector.
11-30
Chapter 11 Troubleshooting Troubleshooting
Solution 2
Relocate the sensor, or change the sensor to which the body motion will have less influence.
Cause 2
The probe size is not appropriate.
Solution
Select a probe size which is appropriate for the patient.
Cause 3
Sensor is exposed to light.
Solution
Place a black or dark cloth over the sensor to avoid direct sunlight.
Cause 2
Communication error has occurred with the SpO2 unit.
Solution
A defective cable or SpO2 unit failure can be considered.
Contact your nearest service representative.
Cause 3
The system was started with the sensor and cable connected.
Solution
Disconnect the SpO2 cable and sensor from this device, and press the standby switch to enter into standby mode.
Then, press the standby switch again to cancel the standby mode, and when the monitoring screen is displayed,
connect the cable and sensor.
Cause 2
The sensor is defective.
Solution
Replace the sensor.
Cause 3
A wrong sensor is used.
Solution
Replace the sensor.
For details of the usable sensors, refer to your nearest service representative.
11-31
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The sensor is defective.
Solution
Replace the sensor.
Cause 3
A wrong sensor is used.
Solution
Replace the sensor.
(
"SpO2 Sensors for Masimo Manufacturer: Masimo Corporation" P13-8)
Cause 4
The sensor is used beyond its expected life.
Solution
Replace the sensor.
NOTE
The X-Cal function automatically monitors the Active Monitoring Time (actual time of SpO2
monitoring) for each sensor and cable.
Even if the sensor is used beyond its expected life, the measurement will not cease unless
the power is turned OFF, sensor is disconnected from the cable, cable is disconnected from
the monitor, or the sensor is reattached.
When a measurement with a sensor that has reached its end of life is suspended for certain
amount of time, and resumed with the same sensor, a message to replace the sensor will
be displayed.
Depending on the device, some sensors may not be recognized.
11-32
Chapter 11 Troubleshooting Troubleshooting
Cause 1
The sensor is detached from the patient.
Solution 1
Check if the sensor is properly attached to the patient.
Solution 2
Check that the light emitting and receiving parts of the sensor LED are aligned.
Cause 2
The sensor is exposed to too much ambient light. The detecting part of the sensor is not covered appropriately.
Solution 1
Turn down or turn off the light.
Solution 2
Avoid the sensor from exposure to ambient light.
Solution 3
Relocate the sensor position.
Cause 2
The sensor has not been attached long enough to obtain stable measurement.
Solution
After the sensor attachment, wait and see for about one minute until the waveform stabilizes.
11-33
Chapter 11 Troubleshooting Troubleshooting
<SpO2 Check Sensor>, <SpO2 Replace Cable>, or <SpO2 Check Cable> is displayed.
Cause 1
Unrecognizable sensor is connected.
A wrong patient cable is used.
When attached to the patient, the sensor was exposed to high-intensity light which lead to false recognition.
Solution
Reattach the SpO2 sensor and patient cable.
Replace with a Fukuda Denshi specified patient cable and sensor.
(
"SpO2 Sensors for Masimo Manufacturer: Masimo Corporation" P13-8)
Cause 2
The cable is used beyond its expected life.
Solution
Replace the patient cable.
NOTE
The X-Cal function automatically monitors the Active Monitoring Time (actual time of SpO2
monitoring) for each sensor and cable.
Even if the cable is used beyond its expected life, the measurement will not cease unless
the power is turned OFF or the cable is reconnected.
When a measurement with a cable that has reached its end of life is suspended for certain
amount of time, and resumed with the same cable, a message to replace the cable will be
displayed.
Depending on the device, some cable may not be recognized.
11-34
Chapter 11 Troubleshooting Troubleshooting
Solution 2
Remove the sensor or patient cable from the HS-8000 or module, and then reconnect it to the SpO2 connector.
Cause 2
The cuff size does not match the selected patient type.
Solution
Use the cuff with correct size for the selected patient type.
11-35
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The pulse is too small to acquire reliable measurement accuracy.
Solution
Check if the cuff application is proper, and if the cuff size corresponds with the selected patient type.
Cause 3
The air hose is disconnected.
Solution
Check if the air hose is tightly connected, and then measure again. If the same message is displayed again, air
leakage inside the HS-8000/DS-8007 can be considered.
Contact your nearest service representative.
Cause 2
The cuff is compressed.
Solution
Make sure that the cuff is not subjected to compression, and measure with the patient arms extended as much as
possible.
If the same message is repeatedly displayed, air system may be clogged. Cease the measurement, and contact
your nearest service representative.
Cause 3
The cuff size is not suitable for the patient.
Solution
Check that the cuff size is appropriate for the patient, and that the cuff is properly attached, and measure again.
Cause 4
The cuff size and the patient classification setting do not match.
Solution
Make sure that the appropriate cuff size is used according to the patient classification setting.
11-36
Chapter 11 Troubleshooting Troubleshooting
Error code condition (phenomenon, or situation) and its cause are indicated below.
C02-00 When "Quick Measurement" is [OFF], the data could not be measured.
Cause 1
The blood pressure may not be correctly measured due to the patient's condition.
Solution
Check the patient's condition, and measure again.
Cause 2
The cuff application has become loose.
Solution
Check that the cuff size is appropriate for the patient, and then measure again after attaching the cuff properly.
C02-01 When "Quick Measurement" is [ON], the data could not be measured.
Cause 1
The blood pressure may not be correctly measured due to the patient's condition.
Solution
Check the patient's condition, set "Quick Measurement" to OFF, and measure again.
Cause 2
The cuff application has become loose.
Solution
Check that the cuff size is appropriate for the patient, and then measure again after attaching the cuff properly.
C02-02 The air hose was disconnected from the NIBP connector during the measurement.
Cause
The air hose was disconnected from the NIBP connector during the measurement.
Solution
Connect the air hose to the NIBP connector, and then measure again.
C03-xx The exhaust ventilation has ceased, or the target deflation speed was not achieved.
Cause 1
During measurement, an artifact such as body motion may have interfered.
Solution
Keep the patient still as much as possible, and measure while the patient is not moving. When performing the
measurement during surgery, avoid artifact caused by the surgery.
Cause 2
During the measurement, air hose was bent or occluded by the compression.
Solution
Make sure that the air hose is not bent or compressed before the measurement.
11-37
Chapter 11 Troubleshooting Troubleshooting
If the error persists and C03-xx error is frequently displayed, contact your nearest service representative and notify
the error code.
C04-xx The cuff inflation was insufficient for the patient's blood pressure.
Cause
The blood pressure has significantly increased from the previous measurement.
Solution
Check the cuff application and size and perform the manual measurement.
C06-xx The pulse signal detected during the measurement was unstable.
Cause 1
During the measurement, the patient has trembled or moved.
Solution
Keep the patient still as much as possible, and measure while the patient is not trembling or moving.
Cause 2
Arrhythmia has frequently occurred during the measurement.
Solution
If arrhythmia occurs many times, correct measurement cannot be performed.Measure when arrhythmia is not
frequently occurring.
Cause
Measurement is automatically repeated due to body motion or insufficient inflation.
Solution
Check the cuff application and size, and measure while keeping the patient still as much as possible.
Cause
The patient has trembled or moved.
Solution
Keep the patient still as much as possible, and measure while the patient is not moving.
C09-00 The inflation value has exceeded the allowable maximum value.
Cause
The cuff was subjected to compression.
Solution
Make sure that the cuff is not subjected to compression, and measure with the patient arms extended as much as
possible.
11-38
Chapter 11 Troubleshooting Troubleshooting
Cause
The cuff size is not suitable for the patient.
Solution
Check that the cuff size is appropriate for the patient, and that the cuff is properly attached, and measure again.
The time of measurement disappears and the numeric data is displayed as " - - - ".
Cause
The preprogrammed time to clear the NIBP data has elapsed.
Solution
The "NIBP Erase Time" can be selected from [60 min.], [120 min.], and after the set duration, the NIBP data will
be displayed as "---".
Select the appropriate time which best fits the monitoring purpose.
Solution
When <NIBP Meas. Error (Exx-xx)> is displayed, the NIBP periodic measurement will be canceled. To resume the
measurement, press the [NIBP Start/Stop] key and check that the measurement is properly performed.
NOTE
When the NIBP Start/Stop key is pressed or when the NIBP measurement interval is
changed, the standby mode will be canceled and the NIBP periodic measurement will start.
11-39
Chapter 11 Troubleshooting Troubleshooting
Solution 1
These errors can be cleared by pressing the [Cancel Error] on the NIBP setup menu or [NIBP Start/Stop] key (fixed
key or user key).If the same message is repeatedly displayed, a failure of the device can be considered. Cease
the measurement, and contact your nearest service representative.
Solution 2
When <NIBP Unit Error (Exx-xx)> is displayed, make sure that the congestion is not generated, and remove the
cuff if necessary.
Temperature
Cause 1
700 series temperature probe is used.
Solution
Use the 400 series temperature probe for measurement.
Cause 2
There is a contact failure of the temperature probe.
Solution
Check if the temperature probe is properly inserted.
11-40
Chapter 11 Troubleshooting Troubleshooting
Cause 2
Injection temperature is not appropriate.
Solution
If iced injectate is used, pay attention not to warm the injector with hands.
Cause 3
The thermistor location is not appropriate.
Solution
Reposition the thermistor.
Cause 4
Arrhythmia event has occurred during the measurement.
Solution
Wait until the patient has stable heart rhythm.
Cause 5
There was patient's body movement during the measurement.
Solution
Have the patient stay still during the measurement.
Cause 6
The patient's hemodynamics changed during the measurement.
Solution
Wait until the patient has stable hemodynamics.
11-41
Chapter 11 Troubleshooting Troubleshooting
The baseline of the thermodilution curve is displaced to the minus side. <LOWER FAULT> is
displayed.
Cause
The blood temperature has not returned to stable condition after the measurement.
The thermodilution curve did not return to the cut off point soon enough. The temperature must return to a point
that is 30% of the peak value within 30 seconds (or 60 seconds depending on the setup).
Solution
If performing continuous measurement, wait for 30 to 60 seconds and check that "Ready" is displayed before
performing the next measurement.
After the measurement is started, the peak of the thermodilution curve was not determined within 22 seconds
(when the time scale is "30 sec" ) or 45 seconds (when the time scale is "60 sec" ).
Solution
The thermistor may be contacting the pulmonary artery wall. Reposition the thermistor and measure again.
11-42
Chapter 11 Troubleshooting Troubleshooting
1.8 F
After the measurement is started, the change in blood temperature is less than 0.1°C /1.8°F for more than 15
seconds (when the time scale is "30 sec" ) or 30 seconds (when the time scale is "60 sec" ).
Solution
Use the iced injectate, and measure again.
Cause 1
The thermistor connector and relay cable is not securely connected.
Solution
Correct measurement cannot be performed unless the thermistor connector and relay cable is securely connected.
Check the connection and perform the measurement again.
Cause 2
The sensor or relay cable is defective.
Solution
If the sensor or cable is defective, measurement can not be performed. Replace the sensor or cable and perform
the measurement again.
11-43
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The sampling line is bent or pinched.
Solution
Make sure that the sampling line is properly allocated.
<Initializing> displayed inside the numeric data box does not disappear.
Cause
An error has occurred during the initialization at power ON.
Solution
Reconnect the cable of HCP-810/HCP-820 and reboot.
If the message is still displayed, CO2 unit failure can be considered. Contact your nearest service representative.
Cause 2
The CO2 calibration value is not appropriate.
Solution
Perform the CO2 calibration again.
11-44
Chapter 11 Troubleshooting Troubleshooting
Cause 1
The CO2 sensor temperature has increased above 40°C/104°F.
Solution
Remove any heat generating source around the sensor.
Cause 2
The CO2 sensor is malfunctioning.
Solution 1
Replace the CO2 sensor.
Solution 2
If the error persists, the failure of HPD-810/HPD-820 can be considered. Stop using the unit and contact our
service representative.
Cause 2
The airway adapter is disconnected from the sensor.
Solution 1
Securely connect the airway adapter to the sensor.
Solution 2
If error persists, perform the airway adapter calibration again.
11-45
Chapter 11 Troubleshooting Troubleshooting
Cause
There is no paper in the printer.
Solution
Set the paper in the paper holder.
Cause
The paper holder is open.
Solution
Firmly close the paper holder.
The second and third waveforms are not printed for manual printing or alarm printing.
Cause
The second and third waveforms are not set on the printing setup screen.
Solution
Set the second and third waveform on the corresponding printing setup screen.
11-46
Chapter 11 Troubleshooting Troubleshooting
The power supply LED on the HR-800 is lit in orange, and [Print Start/Stop] key does not function.
Cause 1
The U-LINK setting is incorrect.
Solution
Press the [Initial Settings]> [External Device]> [U-LINK] keys.
If HR-800 is connected via MGU-800, select [MGU-800].
If HR-800 is not connected via MGU-800, select [OFF].
Cause 2
The HR-800 setting is incorrect.
Solution
Press the [Menu]>[Initial Settings]>[System]>[Other]>[HR-800] keys.
If the HR-800 is installed to the internal slot, select [Built-in].
If connected to U-LINK, select [U-LINK].
Cause 1
The paper is jammed.
Solution
Open the paper holder and properly set the paper.
Cause 2
The thermal head temperature has increased or other failure exists.
Solution
Damage to the thermal head or other failure can be considered. Contact your nearest service representative.
Network Printer
11-47
Chapter 11 Troubleshooting Troubleshooting
Cause 2
A central monitor which is not compatible is used.
Solution
The following central monitors can not be used on the DS-LAN III network.
DS-5700
DS-5800N/NX/NXMB
DS-7600/7600W with software version V05 and prior
When using these central monitors, all monitors in the same network should be set to DS-LAN II.
Cause 3
Inappropriate HUB is used.
Solution
For the DS-LAN II network, use the specified repeater HUB.
For the DS-LAN III network, use the specified switching HUB.
Cause 4
The bed ID is duplicated in the same network.
Solution
If bedside monitors with the same bed ID exist in the same network, communication is not possible. Make sure to
set a unique bed ID for each bedside monitor.
Cause 5
A device not specified by Fukuda Denshi is connected to the network.
Solution
Do not connect PC, printer, or other unspecified device to the DS-LAN network.
Cause 6
The DS-LAN cable is not properly connected.
Solution
The DS-LAN connection will be performed by our service representative. Contact your nearest service
representative.
11-48
Chapter 11 Troubleshooting Troubleshooting
The CO2 waveform is not displayed on the central monitor although the CO2 numeric data is
displayed.
Cause 1
[Impedance] is selected for "RR/APNEA Alarm Source" on the RESP setup menu.
Cause 2
[Ventilator] is selected for "RR/APNEA Alarm Source" on the RESP setup screen.
Solution
Select [CO2] for "RR/APNEA Alarm Source" on the RESP setup menu.
In this case, RR and apnea alarm will be generated based on CO2 measurement.
The impedance respiration waveform is not displayed on the central monitor although the RR
numeric data is displayed.
Cause 1
[CO2] is selected for "RR/APNEA Alarm Source" on the RESP setup menu.
Cause 2
[Ventilator] is selected for "RR/APNEA Alarm Source" on the RESP setup screen.
Solution
Select [Impedance] for "RR/APNEA Alarm Source" on the RESP setup menu.
NOTE
The impedance waveform will not be displayed if [CO2] is set for "RR/APNEA Alarm
Source". AWF, AWP waveform will be displayed.
The CO2 waveform will not be displayed if [Impedance] is set for "RR/APNEA Alarm
Source". AWF, AWP waveform will be displayed.
The CO2 waveform and impedance waveform will not be displayed if [Vent.] is set for
"RR/APNEA Alarm Source" .
11-49
Chapter 11 Troubleshooting Troubleshooting
Telemeter (HLX-801)
Cause 2
The HLX connection setting is incorrect.
Solution
Press the [Menu]>[Initial Settings]>[System]>[Other]>[HLX Connection] keys.
If connected to internal port, select [Internal Port].
If connected to COM port, select [COM Port].
After the setting, make sure that the set channel is displayed as "CHxxxx" at the upper left of the screen.
Cause 2
[Ventilator] is selected for "RR/APNEA Alarm Source" on the RESP setup screen.
Solution
Select [Impedance] for "RR/APNEA Alarm Source" on the RESP setup menu.
11-50
Chapter 11 Troubleshooting Troubleshooting
Remote Control
Cause 2
The section number is not correct.
Solution
Set the correct section number.
General
Even though the numeric data displayed on the extended display unit or central monitor is
exceeding the alarm limit, alarm does not generate.
Cause
The parameters not displayed on the display unit (LC-8016TC/8018TC) are displayed on the central monitor/
extended display unit as [All Data] is selected for "Numeric Data External Output" under [Initial Settings] > [System]
> [Other].
Solution 1
For the parameters which requires alarm monitoring on the extended display unit/central monitor, make sure to
display those on the display unit (LC-8016TC/8018TC).
Solution 2
For the extended display unit/central monitor, if monitoring is necessary for only the parameters displayed on the
display unit (LX-8016TC/8018TC), select [Displayed Data] for "Numeric Data External Output" under [Initial
Settings] > [System] > [Other].
Nothing is displayed on the screen, and the power supply LED is not lit.
Cause 1
The display unit is not properly attached to the main unit.
Solution
Connect the apparatus correctly.
(
Maintenance Manual "System Construction" P1-2)
Cause 2
The main unit or display unit is malfunctioning.
11-51
Chapter 11 Troubleshooting Troubleshooting
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
Cause 2
The main unit or display unit is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
Cause 2
The battery for the backup memory is depleted.
Solution
The battery needs to be replaced. Contact your nearest service representative.
Cause 2
The service life of the LCD backlight has expired.
Solution
The LCD unit needs to be replaced. Contact your nearest service representative.
CAUTION
The display unit utilizes LED for the backlight.
Since this LED deteriorates by the life cycle, the display may become dark, scintillate, or
may not light by the long term use. In such case, contact your nearest service
representative.
11-52
Chapter 11 Troubleshooting Troubleshooting
The system does not start although the power supply cable is connected.
Cause 1
Incorrect CF card is inserted.
Solution
Remove the CF card, turn OFF the power, and turn ON the power again.
Cause 2
The main unit or display unit is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
CAUTION
The calibration will be performed by our service representative.Users should not perform
this procedure as incorrect calibration may cause malfunction of the device.
Cause 2
The LCD unit is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
<DSC-8410 Out of Operating Temp. Range>, <Display Unit Out of Operating Temp. Range> is
displayed.
Cause
The main unit or display unit is used outside the specified environment condition.
Solution
Use the device in the specified environment condition (10°C to 40°C).
11-53
Chapter 11 Troubleshooting Troubleshooting
<The settings have been changed. Reboot the unit.> is displayed when the power is turned ON.
Cause
Rebooting of the system is required.
Solution
Power cycle the system. If the same message is repeatedly displayed, turn OFF the power and contact your
nearest service representative.
<Some parameters are not displayed due to the display layout setting.> is displayed.
Cause 1
The measured parameter is not set to be displayed.
Solution
On the "Display Config." setting, select the measured parameter to be displayed.
Cause 2
During auto display configuration, the quantity of measured parameters exceeded the displayable parameters.
Solution
If there are parameters which measurements are not actually performed, please disconnect their probes/cables.
11-54
Chapter 11 Troubleshooting Troubleshooting
Cause 2
On the "External Device" setting, the set external device is duplicated.
Solution
Check the "External Device" setting, and make sure that the selected external device is not duplicated. The
external devices other than Vigilance, INVOS, BIS, AEP cannot be duplicated.
The combinations of FLOW-i/ GE/ Draeger and MGU-800/ MGU-810, FLOW-i/ GE/ Draeger and ventilator are not
possible.
Super Unit
The system does not start although the power is turned ON.
The power supply LED on the Super Unit does not light in green.
<Super Unit Check Conn.> is displayed.
Cause 1
The power cable of the main unit is not connected.
Solution
Turn OFF the power and connect the power cable.
Cause 2
The module connection cable is not properly connected.
Solution
Turn OFF the power, and make sure that the module connection cable is securely connected. If the knob is loose,
tighten it securely.
Cause 3
The fuse inside the Super Unit has blown out.
Solution
Immediately turn off the power and stop using the device. Contact your nearest service representative.
Cause 4
The Super Unit is not properly connected to the adapter (HSA-80/HSA-81 or DSA-82).
Solution
Insert the Super Unit into the HSA-80/HSA-81 or DSA-82 until a click sound is heard.
11-55
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The module connection cable is not properly connected.
Solution
Turn OFF the power, and make sure that the module connection cable is securely connected. Reconnect the cable
if necessary. If the knob is loose, tighten it securely.
11-56
Chapter 11 Troubleshooting Troubleshooting
This is a normal operation. To not suspend the alarm sound, set the alarm sound suspend function OFF.
Cause 2
The SD card is not formatted.
Solution
Format the SD card.
Cause 2
On the central monitor, "Data Transfer" function is set to [OFF].
Solution
Check the setting on the central monitor.
NOTE
For the software version and model type of the central monitor compatible to data transfer
function, refer to your nearest service representative.
Alarm settings, parameter settings are not transferred from the transport monitor.
The confirmation window to apply the alarm settings, parameter settings of the transport monitor
is not displayed.
Cause
The "Data for Transfer" setting is set to [OFF].
Solution
Check if the "Data for Transfer" setting is set to [ON]. If not, set it to [ON].
11-57
Chapter 11 Troubleshooting Troubleshooting
The system does not start although the power of the DS-8400 is turned ON.
<IB-8000-* Check Conn.> is displayed.
Cause 1
The power cable of the main unit is not connected.
Solution
Turn OFF the power and connect the power cable.
Cause 2
The module connection cable is not properly connected.
Solution
Turn OFF the power, and make sure that the module connection cable is securely connected.
Cause 2
The slot used in the IB-8004 is malfunctioning.
Solution
Insert the expansion module into another slot. If it operates properly, the IB-8004 needs to be repaired.
Cease the measurement, and contact your nearest service representative.
Cause 3
IB-8004 or expansion module is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
Cause 2
The IB-8004 is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
11-58
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The slot used in the IB-8004 is malfunctioning.
Solution
Insert the expansion module into another slot. If it operates properly, the IB-8004 needs to be repaired.
Cease the measurement, and contact your nearest service representative.
Cause 3
IB-8004 or expansion module is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
11-59
Chapter 11 Troubleshooting Troubleshooting
Expansion Module
The system does not start although the power of the DS-8400 is turned ON.
Cause 1
The power cable of the main unit is not connected.
Solution
Turn OFF the power, and make sure that the module connection cable is securely connected.
Cause 2
The module connection cable is not properly connected.
Solution
Turn OFF the power, and make sure that the module connection cable is securely connected.
Cause 3
The standby switch of the display unit is set to OFF.
Solution
Turn ON the standby switch on the display unit.
Cause 2
The slot used in the IB-8004 is malfunctioning.
Solution
Insert the expansion module into another slot. If it operates properly, the IB-8004 needs to be repaired.
Cease the measurement, and contact your nearest service representative.
Cause 3
IB-8004 or expansion module is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
Cause 2
The slot used in the IB-8004 is malfunctioning.
Solution
Insert the expansion module into another slot. If it operates properly, the IB-8004 needs to be repaired.
Cease the measurement, and contact your nearest service representative.
11-60
Chapter 11 Troubleshooting Troubleshooting
Cause 3
IB-8004 or expansion module is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
The following error messages related to the expansion module are displayed on the monitor.
<IB* Slot* Module Failure>, <IB* Slot* Analog Unadjusted>,
<IB* Slot* Check Module>, <IB* Slot* Out of Operating Temp. Range>
Cause
The module connected to the IB-8004 slot is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
Cause 2
The slot used in the IB-8004 is malfunctioning.
Solution
Insert the expansion module into another slot. If it operates properly, the IB-8004 needs to be repaired.
Cease the measurement, and contact your nearest service representative.
Cause 3
IB-8004 or expansion module is malfunctioning.
Solution
Immediately turn OFF the power and cease the operation. Contact your nearest service representative.
Nothing is displayed on the extended display unit. The same display with the main unit is
displayed.
Cause
The video cable of the extended display unit is connected to the external monitor connector of the main unit.
Solution
Connect the cable to the extended display unit connector on the main unit.
The touch panel does not function on the extended display unit.
Cause
The serial communication cable is not connected.
Solution
Connect the serial communication cable of the extended display unit to the extended serial connector (COM A,
COM B) of the main unit.
11-61
Chapter 11 Troubleshooting Troubleshooting
Ventilator
Cause 1
The cable between the DS-8400 System and the ventilator is disconnected or not securely connected.
Solution
Make sure the cable is properly connected.
Cause 2
The power of the ventilator is turned OFF.
Solution
Turn ON the power of the ventilator.
Cause 3
The ventilator is in standby mode.
Solution
Start the ventilation on the ventilator.
Cause 4
The network setting of the monitor does not match with the ventilator.
Solution
Make sure that the network setting of the connecting devices are as follows.
SV-300/SERVO-i/SERVO-s
No network setting.
SERVO-U/n/air
No network setting.
PB-740/760/840
Baud Rate: 9600 bps
Parity Bit: None
Stop Bit: 1
Data Bit: 8
11-62
Chapter 11 Troubleshooting Troubleshooting
Evita4/2dura/XL
Communication Protocol: Medibus
Baud Rate: 19200 bps
Parity Bit: Even
Stop Bit: 1
Multigas Unit
11-63
Chapter 11 Troubleshooting Troubleshooting
11-64
Chapter 11 Troubleshooting Troubleshooting
SvO2/CCO Monitor
Connection Cable
SvO2/CCO Monitor
For Status II Connector For Serial Connector
PulsioFlex - CJ-725*1
*1: To connect the PulsioFlex PC4000, USB to RS-232C connector cable (UC-232A, ATEN) specified by PULSION
Medical Systems is required.
Cause 2
The "External Device" setting is not correct.
Solution
Select [Vigilance/Vigileo], [PiCCO] or [PulsioFlex] for the port function on the "External Device" setup screen.
Cause 3
The measurement data is not displayed on the corresponding external device.
Solution
The measurement data of SvO2, CO, etc. will not be displayed on the monitor unless the data is displayed on the
used external device.Check if the data is displayed on the used external device.
Cause 4
The CCO is not measured.
Solution
The CCO/CCI data will be displayed on this device only when CCO is measured on each external device.
Cause 5
The network setting of the monitor does not match with each external device.
Solution
The network setting of the monitor is fixed to the default setting of each external device and cannot be changed.
Make sure that the network setting of the connecting device is in default setting.
In Case of Vigilance/Vigileo:
Make sure that the network is set as follows.
For procedure to check the Vigilance/Vigileo network setting, refer to the operation manual for the Vigilance/
Vigileo.
11-65
Chapter 11 Troubleshooting Troubleshooting
Cause 6
The software version of Vigilance does not correspond.
Solution
If the Vigilance without the STAT function is connected, the STAT data will not be displayed. Check the software
version of the Vigilance.
Cause 7
The software version of PiCCO does not correspond.
Solution
The compatible version of PiCCO2 is from V3.0. Check the version of the PiCCO2.
Cause 8
The software version of PulsioFlex does not correspond.
Solution
Check the software version of PulsioFlex. The compatible version is PulsioFlex V1.0.
11-66
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The communication setting of the BIS monitor is incorrect.
Solution
ASCII should be set to communicate with this system.
Make sure that ASCII is set on the BIS monitor communication setting.
Refer to the BIS monitor operation manual for procedures.
Cause 2
The BISx cable is defective.
Solution
Check the cable including the connector part, and replace the cable if necessary.
Cause 3
The BISx is defective.
Solution
Replace the BISx.
11-67
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The sensor application is incorrect.
Solution
Read instructions on sensor package to ensure correct sensor placement.
Cause 3
The patient interface cable (PIC cable) is defective.
Solution
Replace the patient interface cable (PIC cable).
Cause 4
The BISx is defective.
Solution
Replace the BISx.
Cause 2
Poor or contaminated connection between the sensor and patient interface cable (PIC cable).
Solution
Clean the connection part, and connect them properly.
Cause 3
The patient interface cable (PIC cable) is disconnected.
Solution
Connect the patient interface cable (PIC cable) correctly.
Cause 4
The patient interface cable (PIC cable) is defective.
Solution
Replace the patient interface cable (PIC cable).
Cause 5
The BISx is defective.
Solution
Replace the BISx.
11-68
Chapter 11 Troubleshooting Troubleshooting
Cause 2
The sensor application is incorrect.
Solution
Read instructions on sensor package to ensure correct sensor placement.
Cause 3
The sensor is not properly connected.
Solution
Verify that the sensor is properly connected.
Cause 4
The patient interface cable (PIC cable) is defective.
Solution
Replace the patient interface cable (PIC cable).
Cause 5
The BISx is defective.
Solution
Replace the BISx.
<Artifacts> is displayed.
Situation The signal quality is less than half of the level desirable for optimal monitoring conditions.
NOTE
This message may occur as the results of artifact (non-EEG signal) such as those generated
from motion (patient movement or eye blinks) or the presence of electrocautery, warming
blankets, or other devices.
Cause 1
Artifact, such as those generated by motion or eyeblinks, is causing loss of EEG recognition.
Solution
If <Artifacts> appears on the display, attempt to identify and eliminate artifact source.
Cause 2
EMG bar indicates electrical activity that may be interfering with EEG recognition.
Solution
If EMG bar is illuminated, attempt to determine and eliminate cause.
Cause 3
The patient interface cable (PIC cable) is defective.
Solution
Make sure that sensor check passes. If not, replace the patient interface cable (PIC cable).
11-69
Chapter 11 Troubleshooting Troubleshooting
Cause 4
The BISx is defective.
Solution
Replace the BISx.
Cause 1
Artifact, such as those generated by motion or eyeblinks, is causing loss of EEG recognition.
Solution
If <BIS SQI < 15%> appears on the display, attempt to identify and eliminate artifact source.
Cause 2
EMG bar indicates electrical activity that may be interfering with EEG recognition.
Solution
If EMG bar is illuminated, attempt to determine and eliminate cause.
Cause 3
The patient interface cable (PIC cable) is defective.
Solution
Make sure that sensor check passes. If not, replace the patient interface cable (PIC cable).
Cause 4
The BISx is defective.
Solution
Replace the BISx.
11-70
Chapter 11 Troubleshooting Troubleshooting
Replace the patient interface cable (PIC cable), then perform "Sensor Check".
Solution 4
Replace the BISx, then perform "Sensor Check".
<Replace BIS Sensor, Too Many Uses>, <Replace BIS Sensor, Invalid Sensor> is displayed.
Cause 1
Sensor has been connected and disconnected too many times.
Solution
Replace the sensor.
Cause 2
The sensor is invalid.
Solution
Replace the sensor.
AEP Monitor
11-71
Chapter 11 Troubleshooting Troubleshooting
INVOS
FLOW-i
Cause 2
The GE anesthesia delivery system is in standby mode.
Solution
The numeric data will be displayed when the measurement is started on the GE anesthesia delivery system.
Cause 3
11-72
Chapter 11 Troubleshooting Troubleshooting
The software version of the GE anesthesia delivery system is not compatible with the DS-8400.
Solution
The compatible software version of DS-8400 is from 07-01.
The GE anesthesia delivery systems compatible with DATEX-OHMEDA COM1.2 SERIAL PROTOCOL protocol
can be connected.
Cause 4
The anesthesia delivery system is not a connectable model type to the DS-8400.
Solution
The connectable model types are as follows.
Avance
Aysis
Cause 2
The Draeger anesthesia delivery system is in standby mode.
Solution
The numeric data will be displayed when the measurement is started on the Draeger anesthesia delivery system.
Cause 3
The software version of the Draeger anesthesia delivery system is not compatible with the DS-8400.
Solution
The compatible software version of DS-8400 is from 07-01.
The Draeger anesthesia delivery systems compatible with Medibus.X protocol can be connected to the DS-8400
system. Set the transmission speed of the anesthesia delivery system to 9600bps for Apollo or 38400 bps for other
devices.
Cause 4
The anesthesia delivery system is not a connectable model type to the DS-8400.
Solution
The connectable model types are as follows.
Fabius
Perseus
Atlan
Apollo
11-73
Chapter 11 Troubleshooting Troubleshooting
PC Communication
Cause 2
Communication with the PC is not performed. The communication is ceased.
Solution
Resume the communication with the PC. The communication time out period is about 1 minute.
TCM4/TCM5 FLEX
*1: To connect the TCM4, the cable specified by Radiometer Medical ApS is required.The communication
will be enabled by connecting the CJ-726 and the cable specified by Radiometer Medical ApS.
*2: To connect the TCM5 FLEX and CJ-725, D-sub 9-pin male to male gender changer (inch screw) is
required.
Cause 2
The "External Device" setting is not correct.
Solution
Select [TCM4/TCM5] for the port function on the "External Device" setup screen.
Cause 3
The measurement data is not displayed on the corresponding external device.
Solution
The measurement data of tcpO2, tcpCO2 will not be displayed on the monitor unless the data is displayed on the
used external device. Check if the data is displayed on the used external device.
Cause 4
The network setting of the monitor does not match with each external device.
Solution
The network setting of the monitor cannot be changed. Make sure that the network setting of the connecting device
11-74
Chapter 11 Troubleshooting Troubleshooting
is as follows.For details of the network setting on the TCM4 and TCM5 FLEX, refer to Radiometer Medical ApS.
In case of TCM4:
RS-232C Protocol: Monlink
In case of TCM5 FLEX:
RS-232C Protocol: Monlink2.0
Cause 5
The software version of TCM4 or TCM5 FLEX does not correspond.
Solution
Check the software version of TCM4 or TCM5 FLEX.
For details of the network setting on the TCM4 and TCM5 FLEX, refer to Radiometer Medical ApS.
TCM4: Version 3.04
TCM5 FLEX: Version 1.18
Cause 2
The power of the external device has been turned OFF.
Solution
Turn ON the power of the external device.
Cause 3
The TCM series device other than TCM4, TCM5 FLEX is connected.
Solution
Only TCM4, TCM5 FLEX can be connected. Check the model type of the external device.
External Media
11-75
Chapter 11 Troubleshooting Troubleshooting
Solution
Set the CF/CFast card into the CF/CFast card slot.
<Data Read Error. Model type or software version is not compatible. Do you want to read only the
common data?> is displayed.
Cause 1
There is no data on the CF card.
Solution
Check if the CF card is readable. Or, check if the data is present on the CF card. Pressing "Yes" will not start
reading the compatible data.<Card access error.> will be displayed.
Cause 2
Error is detected during the read process.
Solution
The data may not be correctly written on the CF card. Format the card again on the used device and try the write/
read process again. Pressing "Yes" will not start reading the compatible data.
Cause 2
Error is detected during the write process.
Solution
Make sure that the CF card is properly inserted and try the write process again.
Format the card again on the used device and try the write/read process again.
Cause 3
Unspecified CF card is used.
Solution
Use the specified CF card.
<Wrong CF card for full disclosure.>, <Failed to read full disclosure from the CF card.>
Cause
Specified memory card is not used.
The card is not formatted.
The data stored in the card is damaged. The card has been already used on another device.
11-76
Chapter 11 Troubleshooting Troubleshooting
Solution 1
Use the recommended memory card.
Disconnect and connect the full disclosure waveform card again to make sure that it is properly inserted.
Format the card on the used device. (All previous data will be deleted.)
Solution 2
If the error persists, contact your nearest service representative.
11-77
Chapter 11 Troubleshooting Troubleshooting
11-78
Chapter 12 Setup Item/Default Value Patient Admit / Discharge
Patient Name Numeric, Alphabet, Symbol (16 characters) Blank Backup Initialize
Age 0 year to 150 years or 0 day to 999 days 0 year Backup Initialize
Blood Type A, B, O, AB
Blank Backup Initialize
Rh +/-
Depends on the setting under
Pacemaker Used, Not used Not Used [Initial Settings>User I/F>Power
ON/Discharge].
Alarm
12-1
Chapter 12 Setup Item/Default Value Alarm
Pacer not Pacing ON, OFF 20 beats to 200 beats/ OFF, 50 bpm
min.
5 beat increments
12-2
Chapter 12 Setup Item/Default Value Alarm
HR Lower Limit for VT 120, 140 bpm 120 Depends on the "Main Mode"
setting under [Initial
0 beats to 100 beats beats/min Settings>User I/F>Power ON/
HR Lower Limit for Run *4 40 bpm
10 beat /min. increments
Discharge].
If "Main Mode" setting is
100 beats to 250 beats/min [Backup]; Depends on the
HR Lower Limit for SVT 150 bpm
10 beat /min. increments "Alarm" setting under [Power
ON/Discharge].
ST All Alarm ON, OFF ST All Alarm OFF
*2 Individual Alarm ON, OFF Individual Alarm OFF
ST1 to ST12(mm)
±20 mm OFF to OFF
1 mm increments
ON
ON, OFF
*5 SYS: 80 to 180
BP1 (mmHg) 0 mmHg to 300 mmHg
DIA: OFF to OFF
8 mmHg increments
MEAN: OFF to OFF
ON Depends on the "Main Mode"
ON, OFF
SYS: 10.0 to 24.0 setting under [Initial
BP1 (kPa) *5 0 kPa to 40.0 kPa
DIA: OFF to OFF Settings>User I/F>Power ON/
0.5 kPa increments
MEAN: OFF to OFF Discharge].
OFF If "Main Mode" setting is
ON, OFF [Backup]; Depends on the
*5 SYS: OFF to OFF
BP2 to BP8 (mmHg) 0 mmHg to 300 mmHg "Alarm" setting under [Power
DIA: OFF to OFF
8 mmHg increments ON/Discharge].
MEAN: OFF to OFF
OFF
ON, OFF
SYS: OFF to OFF
BP2 to BP8 (kPa) *5 0 kPa to 40.0 kPa
DIA: OFF to OFF
0.5 kPa increments
MEAN: OFF to OFF
*1: Select [ON/OFF] for "Asystole, VF, VT Alarm" under [Menu<Initial Settings<Alarm] in advance.
*3: For HR, Ext Tachy, Ext Brady, 60 bpm or lower can be set in 1 bpm increments. For PR_SpO2, 25 bpm or lower can be set in
1 bpm increments.
*4: "HR Lower Limit for Run" can be set in 5 bpm increments for 50 bpm and above.
*5: For BP, 50 mmHg/7.0 kPa or lower can be set in 2 mmHg/0.2 kPa increments.
12-3
Chapter 12 Setup Item/Default Value Alarm
ON
ON, OFF
SYS: 80 to 180
NIBP (mmHg) 10 mmHg to 300 mmHg
DIA: OFF to OFF
5 mmHg increments
MEAN: OFF to OFF
ON
ON, OFF
SYS: 10.0 to 24.0
NIBP (kPa) 1.5 kPa to 40.0 kPa
DIA: OFF to OFF
0.5 kPa increments
MEAN: OFF to OFF
12-4
Chapter 12 Setup Item/Default Value Alarm
Alarm Settings Alarm Suspend 1 min., 2 min. 2 min. Depends on the "Main Mode"
(Setup) Time setting under [Initial
Settings>User I/F>Power ON/
Alarm Silence 1 min., 2 min. 2 min.
Discharge].
Time
Status Alarm Link to alarm silence time, Link to Link to each new
Control each new occurrence occurrence
12-5
Chapter 12 Setup Item/Default Value Parameter
*1: For CVP, 50 mmHg/7.0 kPa or lower can be set in 2 mmHg/0.2 kPa increments.
*2: For RR, 1 Bpm increments may be applied depending on the "RR Alarm Increment" settings. (
I/F" P5-13)
Maintenance Manual "User
*3: When the numeric data acquired from FLOW-i, GE or Draeger anesthesia delivery system is displayed, the following alarms
cannot be set. Also, these alarms will not generate.
NOTE
By selecting [Backup] for "Power ON" and "Discharge" under [Initial Settings>User I/F >Power
ON/Discharge], the settings will be retained at "Power ON" and "Discharge" respectively. If
[Initialize] is selected, the settings will be initialized at "Power ON" and "Discharge".
Parameter
ECG
Leads I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, ECG1: II
V6 ECG2: aVR
ECG3: I
ECG4: III
ECG5: aVL
ECG6: avF
*1
ECG7: V1
ECG8: V2
ECG9: V3
ECG10: V4
ECG11: V5
ECG12: V6
ECG1 to ECG12
Size Auto, x1/4, x1/2, x1, x2, x4 *1
x1
Filter Mode Monitor, Diagnosis, ESIS Monitor Backup Backup
Synchronized Mark/Tone ECG, SpO2-1, SpO2-2, BP, Auto, OFF Auto Backup Backup
Pace Pulse Mask Time Auto, 10 ms, 20 ms, 40 ms, OFF Auto Depends on
the "Main
Mode" setting
under [Initial
Initialize
Settings>Use
r I/F>Power
ON/
Discharge].
ST/VPC/Arrhy. Alarm
ON, OFF ON Backup Backup
Display
ECG Analog Output Disp. Lead, Selected Lead Disp. Lead Backup Backup
12-6
Chapter 12 Setup Item/Default Value Parameter
ECG
*1: Depends on the "Main Mode" setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup] ; Depends on the "ECG1, ECG2 Size" setting under [Setup>Initial Settings>User I/F>Power ON/
Discharge].
RESP
RR/APNEA Alarm Source Auto, Impedance, Vent., CO2/GAS Auto Backup Backup
CVA Detect ON, OFF OFF *1
*1: Depends on the "Main Mode" setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup]; Depends on the "CVA Detect" setting under [Power ON/Discharge].
SpO2 (General)
12-7
Chapter 12 Setup Item/Default Value Parameter
SpO2 (Medtronic)
SpO2 Averaging 2-4 sec, 4-6 sec, 8 sec, 10 sec, 12 sec, 14 8 sec.
*1
sec, 16 sec
*1: Depends on the "Main Mode" setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup] ; Depends on the "SpO2 Averaging" setting under [Power ON/Discharge].
NIBP
NIBP Auto Mode Cont., 1min, 2min, 2.5min, 5min, 10min, 15 OFF
min, 20min, 30min, 60min, 120min, Lumbar *1
Mode, OFF
12-8
Chapter 12 Setup Item/Default Value Parameter
NIBP
Time Display Elapsed, Meas. Elapsed Time Depends on the "Main Mode"
setting under [Initial
Periodic Measurement Time, Meas. Time Settings>User I/F>Power ON/
Starting Time
Discharge].
Target Inflation Value Adult: 100 mmHg to 290 mmHg Adult: 180 mmHg
Child: 100 mmHg to 200 mmHg Child: 140 mmHg
Backup Initialize
Neonate: 100 mmHg to 140 mmHg Neonate: 110
mmHg
*1: Depends on the "Main Mode" setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup] ; Depends on the "NIBP Auto Mode" setting under [Setup>Initial Settings>User I/F>Power ON/
Discharge].
BP1 to 8
12-9
Chapter 12 Setup Item/Default Value Parameter
TEMP1 to TEMP8
Label T*, Tsk, Tre, Tes, Tco, User 1 to User 7 T* (T1 to T8) Depends on the "Main Mode"
setting under [Initial
Settings>User I/F>Power ON/
Discharge].
ΔTEMP-A to TEMP-D
0- 4, 0- 8, 0- 10 kPa 0- 4 *1
*1: Depends on the Main Mode setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup]; Depends on the "CO2 Scale" setting under [Power ON/Discharge].
*2: Depends on the "Main Mode" setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup]; Depends on the "EtCO2 Peak Duration" setting under [Power ON/Discharge].
CO2 (Medtronic/HCP-810/HCP-820)
0- 4, 0- 8, 0- 10 kPa 0- 4 *1
*1: Depends on the Main Mode setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup]; Depends on the "CO2 Scale" setting under [Power ON/Discharge].
*2: Depends on the "Main Mode" setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup]; Depends on the "EtCO2 Peak Duration" setting under [Power ON/Discharge].
12-10
Chapter 12 Setup Item/Default Value Parameter
AWF Scale 5, 10, 20, 50, 180 L/min 50 L/min Depends on the "Main Mode"
AWV Scale 50, 250, 500, 1000, 3000 mL 500 mL setting under [Initial
Settings>User I/F>Power ON/
Pressure: 0, 20, 30, 50, 120 Pressure: 30 Discharge].
P-V, F-V Scale Volume: 250, 500, 750, 1000 Volume: 500
Flow: ±20, ±50, ±180 Flow: ±50
0- 4, 0- 8, 0- 10 kPa 0-4kPa *2
*1: This setting is enabled when FLOW-i, GE or Draeger anesthesia delivery system is connected.
*2: Depends on the Main Mode setting under [Initial Settings>User I/F>Power ON/Discharge]. If "Main Mode"
setting is [Backup]; Depends on the "CO2 Scale" setting under [Power ON/Discharge].
BIS (A-2000/A-3000)
Scale (EEG1, EEG2) ± 25, ± 50, ± 100, ± 250 ±50μV Depends on the "Main Mode"
setting under [Initial
Short Trend 2nd SR
SR, EMG, SQI, OFF Settings>User I/F>Power ON/
Parameter Discharge].
12-11
Chapter 12 Setup Item/Default Value Parameter
EEG (HA-800)
Time Constant 0.1 sec., 0.3 sec. 0.3 sec. Backup Backup
Stopwatch
SvO2/CCO
Item Description Default At Power ON At Discharge
INVOS
12-12
Chapter 12 Setup Item/Default Value Data Review
Data Review
Graphic Trend
Trend A HR, ST (I to V6), SpO2, PR_ SpO2, VPC, NIBP, Upper Row:
BP1 to 8, PR_IBP, PDP, CPP, HR, NIBP
TEMP1 to 8, Tb, ΔTEMP-A to D, RR_IMP, Lower Row:
APNEA, EtCO 2, InspCO2, RR_ GAS, ExpN2O, SpO2, TEMP1, RR_IMP
InspN2O, O2, ExpAGT, InspAGT, MAC, BIS,
Trend B Upper Row:
SR, EMG, SQI, AEP, SvO2, ScvO2, CCO, CCI,
BT, RR_VENT, RR_SpO2, ExpO2, Insp O2, PI, HR, BP1, TEMP1, NIBP
Lower Row:
PVI, SpCO, SpMet, SpHb, SpOC, PEAK,
SpO2, EtCO2, ST (II) , Depends on the "Main Mode"
PEEP, ExpMV
RR_GAS setting under [Initial
Settings>User I/F>Power
Trend C Upper Row:
ON/Discharge].
HR, TEMP1, BP1, NIBP
Lower Row:
SpO2, InspO 2, EtCO2,
InspAGT
Trend D N/A
Time 10min, 1h, 2h, 4h, 8h, 12h, 16h, 24h 4 hours Depends on the "Main Mode"
setting under [Initial
Settings>User I/F>Power
ON/Discharge].
Display , , , , , ,
Selection
, , , ,
, , ,
BP1 to BP8 20, 50, 100, 150, 200, 300 200 mmHg 24 kPa
mmHg
4, 8, 16, 20, 24, 40 kPa
Depends on the "Main Mode"
PDP, CPP 20, 50, 100, 150, 200, 300 200 mmHg 24 kPa
setting under [Initial
mmHg Settings>User I/F>Power
4, 8, 16, 20, 24, 40 kPa
ON/Discharge].
NIBP 100, 150, 200, 300 mmHg 200 mmHg 24 kPa
16, 20, 24, 40 kPa
12-13
Chapter 12 Setup Item/Default Value Data Review
Graphic Trend
12-14
Chapter 12 Setup Item/Default Value Data Review
Graphic Trend
Tabular Trend
12-15
Chapter 12 Setup Item/Default Value Data Review
Tabular Trend
Parameter [H Module]
Selection OFF, HR, VPC, ST (I to V6), SpO2,
PR_ SpO2, NIBP-S/D/M, BP1 to 8- S/D/M,
PR_IBP, PDP, PCWP, CPP, TEMP1 to 8, Tb,
CO, EtCO2, InspCO2, RR_GAS, RR_IMP,
RR_VENT, APNEA, MAC, O2-E, O2-I, N2O-E,
N2O-I, AGT-E, AGT-I, AGT2-E, AGT2-I,
PI, PVI, SpCO, SpMet, SpHb, SpOC,
E-VT, I-VT, E-MV, I-MV, P-PEAK, P-PAUSE,
PEEO, P-MEAN, RES, COMP,
VT1sec, I/E RATIO
[Ventilator]
E-VT, I-VT, MV, SMV, P-PEAK, P-PAUSE,
PEEP, P-MEAN, E-RES, I-RES, COMP, FiO2,
P-MIN, S-COMP, D-COMP, S-RR, I/E RATIO,
RES
[Other]
BIS, SQI, EMG, SR, SEF, TOTPOW, IMP,
AEP, Lt-rSO2, Rt-rSO2, S1-rSO2, S2-rSO2,
tcpO2, tcpCO2
Filtering
(Sampling 10sec., All All Initialize Initialize
Interval)
12-16
Chapter 12 Setup Item/Default Value Data Review
OCRG
Recall
ST Measurement
ST Waveform
10 sec., 1 min., 5 min., 10 min. 10 sec.
Interval
12-17
Chapter 12 Setup Item/Default Value Basic Setup
NOTE
The graphic trend, tabular trend, alarm history will be saved even after the power is turned
OFF.
The recall, ST, OCRG data will be saved even after about one hour of standby mode (AC
power or battery operation).
12-lead Display
Basic Setup
Tone/Volume
Item Description Default At Power ON At Discharge
Tone: 5 types 1
Tone: 3 types 1
12-18
Chapter 12 Setup Item/Default Value Basic Setup
Tone/Volume
Tone: 1 type 1
Tone: 1 type 1
* When [Fukuda Tone] is selected for "Alarm System", the tone can be selected from 8 levels.
Display
Configuration
Numeric Data OFF, HR/PR, HR, PR_IBP, VPC/PACE, ST/ HR, SpO2-1, NIBP, BP1,
VPC, ST-A to C, BP1 to 8, NIBP, NIBP LIST, RR_IMP, CO2
SpO2-1, SpO 2-1/PR_SpO2-1, PR_SpO2-1,
RR_IMP, RR_CO2, RR_VENT, RR_SpO2,
TEMP1 to 8, TEMP1/2, TEMP3/4, TEMP5/6,
TEMP7/8, SpO2-2, SpO2-2/PR_SpO2-2,
Depends on the setting under
PR_SpO2-2 ΔTEMP-A to D, VENT, P-V F-V,
[Initial Settings>User I/
SvO2/CO, BIS, AEP, CO2, O2, N2O, Agent, RR/
F>Power ON/Discharge].
CO2/Agent/O2/N2O, CO2/Agent/O2/N2O, RR/
Agent/O2/N2O, Agent/O2/N2O, Agent/N2O,
HEMO, HEMO-I, STOPWATCH, SpCO,
SpMet, SpHb, GAS/SPIRO, SPIRO, VENT-A,
VENT-B, Hemo/etc-A, Hemo/etc-B, Extended
Function-A
Waveform FF, ECG1 to ECG12, ECG1 Cascade to ECG1, SpO2-1, BP1, RESP,
ECG12 Cascade, BP1 to BP8, BP Overlap 1 to CO2
BP Overlap 3, SpO2-1, SpO2-2, RESP, AWF, Depends on the setting under
AWP, AWV, CO2, O2, Agent, Block Cascade, [Initial Settings>User I/
RR Overlap 1 to 3, EEG1 to 8 F>Power ON/Discharge].
Enlarged ON, OFF OFF
Waveform
Sweep Speed Circ.: 6.25, 12.5, 25, 50 Circ.: 25 Depends on the setting under
Vent: 6.25, 12.5, 25 Vent: 6.25 [Initial Settings>User I/
F>Power ON/Discharge].
Short Short Trend Selection OFF Depends on the setting under
Graphic Trend ON, OFF, Overlap 15 min. [Initial Settings>User I/
Display Length: 0, 5, 10, 15, 20, 25, 30 min. F>Power ON/Discharge].
12-19
Chapter 12 Setup Item/Default Value Basic Setup
Display
Configuration
BP Overlap 2, 3 N/A
BP Overlap 2, 3 N/A
12-20
Chapter 12 Setup Item/Default Value Basic Setup
Display
Configuration
Detail Setup Short Trend Link with Numeric, Link with Numeric
(Short Trend) Link with Waveform,
User Setup
NOTE
By selecting [Backup] for "Power ON" and "Discharge" under [Initial Settings>User I/F
>Power ON/Discharge], the display configuration settings will be retained at "Power ON"
and "Discharge" respectively. If [Initialize] is selected, the settings will be initialized at
"Power ON" and "Discharge".
Manual Printing
Waveform ECG1, ECG2, ECG3, BP1 to 8, SpO2, ECG1 Depends on the "Main Mode"
RESP, CO2, O2, Agent, AWF, AWP, setting under [Initial
EEG1 to 8 Settings>User I/F>Power
Print Duration 24 sec., Cont. 24 sec. ON/Discharge].
12-21
Chapter 12 Setup Item/Default Value Basic Setup
Manual Printing
Zoom Wave
Bedside, Central, Laser Bedside
(Recall, Full Disc.)
Other Setup: Recall Printing Graphic Printing, Manual Printing Graphic Printing Depends on the "Main Mode"
setting under [Initial
Settings>User I/F>Power
ON/Discharge].
Auto Printing
12-22
Chapter 12 Setup Item/Default Value Basic Setup
Other Setup
Volume No Change, 3, 1, 0 1
Color Background Color Black, Gray, Light Gray Numeric Data: Black Depends on the "Main Mode"
(Meas.) Waveform: Black setting under [Initial
Settings>User I/F>Power ON/
Background Color Discharge].
(Wave)
Color Palette Light, Clear, Deep, Vivid Vivid Depends on the "Main Mode"
setting under [Initial
Settings>User I/F>Power
ON/Discharge].
HR 12 colors + White 6
ST 6
VPC White
PACE White
NIBP 8
SpO2(Ch1, Ch2) 4
SpCO (Ch1, Ch2) 4
CO2 8
RESP White
BP1, ART 1
PAP 4
CVP 8
ICP 7
IAP 12
LVP 2
BP2 8
BP3 4
BP4 6
BP5 2
BP6 12
BP7 9
BP8 7
TEMP1 to 8, Tb 2
12-23
Chapter 12 Setup Item/Default Value Basic Setup
Other Setup
AWF 6
AWP 4
AWV 8
VENT White
BIS 2
AEP White
INVOS White
MAC White
SvO2+CO White
Stopwatch White
12-24
Chapter 13 System Components Medical Device
DSC-8410 Yes No No
Accessories
The following products are available as accessories for the DS-8400 System.
Purchase them as necessary.
CAUTION
Use only the spare parts specified for this device. If unspecified products are used, proper
function cannot be executed.
For quality improvement purposes, specifications may be subjected to change without prior
notice.
Accessory
HSA-80 HS Adapter
HSA-81 HS Adapter
IB-8004 Input Box
13-1
Chapter 13 System Components Accessories
OAO-100A Mounting Bracket for Gas Unit/External For attaching the gas unit/external output box to the HSA-
Output Box 81
13-2
Chapter 13 System Components Accessories
13-3
Chapter 13 System Components Accessories
OTO-16L Trolley(L)
ECG Accessory
CAUTION
When using the electrosurgery-proof type ECG relay cable, the impedance respiration
cannot be measured, and its numeric data and waveform will not be displayed. When
measuring in an environment where electrosurgery is not performed, make sure to use the
standard ECG relay cable.
13-4
Chapter 13 System Components Accessories
Temperature Accessory
NOTE
700 series temperature probe cannot be used.
CO Accessory
NIBP Accessory
13-5
Chapter 13 System Components The Other Medical Devices
DS-8007 System
13-6
Chapter 13 System Components The Other Medical Devices
ECG Sensors
CMF-700-3 ECG Clip Type Lead Cable 3-electrode 3-electrode (clip type)
CMF-700-4 ECG Clip Type Lead Cable 4-electrode 4-electrode (clip type)
CMF-700-5 ECG Clip Type Lead Cable 5-electrode 5-electrode (clip type)
CMF-700-5C ECG Clip Type Lead Cable 5-electrode 10-electrode (clip type, for chest-5lead 90cm)
CMF-702-5 Limb-5lead For 10 electrode ECG 10-electrode (clip type, for limb-5lead 90/150cm)
CMF-703-5 Limb-5lead For 10 lead ECG 10-electrode (clip type, for limb-5lead 150/220cm)
CMF-704-5C Chest-Lead For 10 lead ECG 10-electrode (clip type, for chest-5lead 150cm)
CMC-700-3 3 electrode ECG 3-electrode (clip type)
CMC-703-5 10 electrode ECG Limb-5lead 10-electrode (clip type, for limb-5lead 150/220cm)
CMC-704-5C 10 electrode ECG Chest-5lead 10-electrode (clip type, for chest-5lead 150cm)
CMC-700-5C 10 electrode ECG for Chest lead 10-electrode (clip type, for chest-5lead 150cm)
NIBP Cuffs
NOTE
Use only the specified cuff for the inflation measurement.
13-7
Chapter 13 System Components The Other Medical Devices
Nellcor Respiratory Sensor 10068119 For adult weighing 30 kg and above, usable only with DS-8007N
NOTE
There are various types of sensors available. For details, refer to your nearest service
representative.
Masimo Rainbow Sensor Rainbow® DCI-dc3 Reusable Direct Connect Sensor for Adult (0.9 m)
®
Masimo Rainbow Sensor Rainbow DCI-dc8 Reusable Direct Connect Sensor for Adult (2.4 m)
®
Masimo Rainbow Sensor Rainbow DCI-dc12 Reusable Direct Connect Sensor for Adult (3.6 m)
®
Masimo Rainbow Sensor Rainbow R25 Adhesive Sensor for Adult
®
Masimo Rainbow Sensor Rainbow R25-L Adhesive Sensor (L-Shape) for Adult/Neonate
®
Masimo Rainbow Sensor Rainbow R20 Adhesive Sensor for Pediatric
®
Masimo Rainbow Sensor Rainbow R20-L Adhesive Sensor (L-Shape) for Pediatric/Infant
Masimo Rainbow ReSposable Sensor Rainbow ReSposable R2-25 ReSposable Sensor Cable (For Adult) x1
ReSposable Sensor (Adhesive Tape for Adult)
x10
13-8
Chapter 13 System Components The Other Medical Devices
Masimo Rainbow ReSposable Sensor Rainbow ReSposable R2-25a To be used with ReSposable sensor (adhesive
tape for adult), ReSposable sensor cable (for
adult), 25 per box
Masimo Rainbow ReSposable Sensor Rainbow ReSposable R2-20a To be used with ReSposable sensor (adhesive
tape for child), ReSposable sensor cable (for
child), 25 per box
Masimo Rainbow ReSposable Sensor Rainbow ReSposable R2-25r To be used with ReSposable sensor tape (for
adult), 5 per box
Masimo Rainbow ReSposable Sensor Rainbow ReSposable R2-20r To be used with ReSposable sensor tape (for
child), 5 per box
Masimo RD SET Sensor RD SET NeoPt Adhesive Sensor for Premature Neonate
LNCS Red Patient Cable Red LNC-04 For LNCS sensor, 1.2 m
LNCS Red Patient Cable Red LNC-10 For LNCS sensor, 3.0 m
LNCS Red Patient Cable Red LNC-14 For LNCS sensor, 4.2 m
Masimo Rainbow Patient Cable Rainbow RC-1 For Rainbow Sensor, 0.3 m
Masimo Rainbow Patient Cable Rainbow RC-4 For Rainbow Sensor, 1.2 m
Masimo Rainbow Patient Cable Rainbow RC-12 For Rainbow Sensor, 3.6 m
RD Rainbow Patient Cable RD Rainbow SET MD20-1.5 For RD SET sensor, 0.5 m
RD Rainbow Patient Cable RD Rainbow SET MD20-05 For RD SET sensor, 1.5 m
RD Rainbow Patient Cable RD Rainbow SET MD20-12 For RD SET sensor, 3.7 m
NOTE
SpCO and SpHb cannot be measured at the same time for all the sensors.
NOTE
There are various types of sensors available.For details, contact your nearest service
representative.
13-9
Chapter 13 System Components The Other Medical Devices
Smart CapnoLine Plus (Adult / Intermediate) 009818 For oral nasal, short term use
Smart CapnoLine Plus (Adult / Intermediate 009822 For oral nasal, short term use
O 2)
Smart CapnoLine (Pediatric) 007266 For oral nasal, short term use
Smart CapnoLine H Plus (Adult / Intermediate 010433 For oral nasal, long term use
O 2)
Smart CapnoLine H (Pediatric) 010581 For oral nasal, long term use
Smart CapnoLine H (Pediatric O2) 010582 For oral nasal, long term use
Microstream VitaLine H set (Adult / Pediatric) 010787 For long term use
Microstream VitaLine H set (Infant / Neonate) 010807 For long term use
FilterLine H Set (Adult / Pediatric) XS04624 For long term use
Airway adapter (adult) 7007-00, 7007-01 Reusable, for ET tube sizes > 4.0 mm
(7007-00: 10 per box, 7007-01: 1 per box)
Airway adapter (neonate) 7053-00, 7053-01 Reusable, for ET tube sizes = < 4.0 mm
(7053-00: 10 per box, 7053-01: 1 per box)
NOTE
There are various types of sensors available. For details, refer to your nearest service
representative.
13-10
Chapter 13 System Components The Other Medical Devices
CAUTION
Avoid liquid ingress to the patient interface cable (PIC). Contact of fluids with the PIC sensor
connector can interfere with PIC performance.
To minimize the risk of patient strangulation, the patient interface cable (PIC) must be
carefully placed and secured.
When installing the BISx, it should not be closely attached to the patient. Secure it on the
bedside rail or pole using a clip.
BIS sensor is disposable. Do not reuse it.
Do not reuse the BIS sensor to other patients. It may cause cross-infection.
The duration for one usage should be within 24 hours.
SPIRIT Flow sensor, Adult 60-16100-00 For MGU-810 series, single-use only
SPIRIT Flow sensor, Pediatric 60-16200-00 For MGU-810 series, single-use only
13-11
Chapter 13 System Components The Other Products
The other non-medical products to be used in combination with this device as a system are shown below.
Product Name / Description Model Remarks
13-12
Chapter 14 Specification Specification
Chapter 14 Specification
Specification
Size
Weight
7.5 kg (not including the accessory)
Environmental Conditions
14-1
Chapter 14 Specification Specification
Safety
Power Supply
Voltage 100-240 V AC
Frequency 50 Hz / 60 Hz
Operation Time 1.5 hours or more (at 25°C, DSC-8410, LC-8018TC, power saving mode, DS-8007 standard
measurement, continuous NIBP measurement of 15 min.)
Usable Life
6 years According to self-certification
(
Maintenance Manual "Periodic Replacement" P7-1)
Size
DS-8007 200 (W) mm x 108 (D) mm x 185 (H) mm (not including the protrusion)
Weight
Environmental Conditions
14-2
Chapter 14 Specification Specification
Safety
Power Supply
Usable Life
Size
LC-8018TC 475 (W) mm x 307 (H) mm x 62.5 (D) mm (not including the protrusion)
LC-8016TC 410 (W) mm x 265 (H) mm x 62.5 (D) mm (not including the protrusion)
Weight
Environmental Conditions
Safety
14-3
Chapter 14 Specification Specification
Power Supply
Usable Life
Size
HS-8312N/8312M 85 (W) mm x 100 (H) mm x 200 (D) mm (not including the protrusion)
Weight
Environmental Conditions
Safety
Degree of protection against ECG /RESP (Impedance), SpO2, SpCO*, SpMet*, SpHb*, TEMP, BP, CO: Type CF Applied Part
electric shock NIBP: Type BF Applied Part
*: For HS-8312M only
14-4
Chapter 14 Specification Specification
Power Supply
Usable Life
Size
MGU-801P 125 (W) mm x 110 (H) mm x 200 (D) mm (not including the protrusion)
MGU-811P 125 (W) mm x 108.5 (H) mm x 200 (D) mm (not including the protrusion)
HR-800 87 (W) mm x 108.5 (H) mm x 100 (D) mm (not including the protrusion)
Weight
Environmental Conditions
Safety
General Standard EN 60601-1:2006+A1:2013
(Medical electrical equipment - Part 1: General requirements for basic safety and essential
performance)
EMC Standard EN 60601-1-2: 2015
(Medical electrical equipment - Part 1-2: General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic disturbances - Requirements and tests)
Type of protection against Class I Equipment (During AC power operation)
electric shock Internally Powered Equipment (During battery operation)
14-5
Chapter 14 Specification Specification
Usable Life
Size
Weight
HM-800/HM-801 0.5 kg (not including the accessory)
Environmental Conditions
Operating Temperature 10°C to 40°C
Safety
Power Supply
Voltage DC 12 V (Supplied from the DSC-8410 Main Unit via IB-8004 Input Box)
Usable Life
14-6
Chapter 14 Specification Specification
Size
Weight
Environmental Conditions
Operating Temperature 10°C to 40°C
Safety
General Standard EN 60601-1:2006+A1:2013
(Medical electrical equipment - Part 1: General requirements for basic safety and essential
performance)
EMC Standard EN 60601-1-2: 2015
(Medical electrical equipment - Part 1-2: General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic disturbances - Requirements and tests)
Type of protection against Class I Equipment (During AC power operation)
electric shock Internally Powered Equipment (During battery operation)
Power Supply
Usable Life
14-7
Chapter 14 Specification Specification
Size
Weight
Environmental Conditions
Operating Temperature 10°C to 40°C/50°F to 104°F
Safety
General Standard EN 60601-1:2006+A1:2013
(Medical electrical equipment - Part 1: General requirements for basic safety and essential
performance)
EMC Standard EN 60601-1-2: 2015
(Medical electrical equipment - Part 1-2: General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
Type of protection against Class I Equipment (DS-8007 System)/Internally Powered Equipment (DS-8007 System)
electric shock
Waterproof/Dustproof IPX0
Power Supply
Voltage HCP-820: DC 12 V
HPD-820: DC 12 V / 5 V
(Via DS-8007 Main Unit)
Usable Life
14-8
Chapter 14 Specification Specification
Size
IB-8004 180 (W) mm x 137.5 (H) mm x 160 (D) mm (not including the protrusion)
Weight
Environmental Conditions
Safety
Power Supply
Usable Life
14-9
Chapter 14 Specification Specification
Size
Weight
Environmental Conditions
Safety
Type of protection against Class I equipment (DS-8400 System )/Internally Powered Equipment (DS-8400 System )
electric shock
Degree of protection against BIS: Type BF Applied Part (When connected to BISx)
electric shock
Waterproof/Dustproof IPX0
Voltage DC 12 V / 5 V
Usable Life
14-10
Chapter 14 Specification Performance
Performance
This section states the performance of the DS-8400 system. The EMC essential performance is indicated with .
Display Panel
Operation
Touch Panel Capacitive Touch Panel
Fixed Keys 3 keys (NIBP Start/Stop, NIBP Auto Mode, Alarm Silence)
Sound Pressure
Clock Accuracy
Alarm
Alarm Function For each alarm level, the respective alarm sound generates, and the alarm indicator flashes.
ECG
Accuracy of Input Signal Overall system error and frequency response is set using method A, B, C, and D.
Reproduction
14-11
Chapter 14 Specification Performance
Tall T-wave Rejection 1.2 mV T-wave can be removed when tested according to IEC 60601-2-27.
Capability
Transient Characteristic 3.2 sec, 0.3 sec, 0.1 sec (time constant can be changed)
14-12
Chapter 14 Specification Performance
Respiration (HS-8000)
Respiration (DS-8007)
Temperature
Medtronic Unit
Measurement Accuracy Adult: ±3%SpO2 when 70%SpO2 to 100%SpO2 (When DS-100A is used)
Neonate: ±2%SpO2 when 70%SpO2 to 100%SpO2
PR Measurement Range 20 bpm to 250 bpm
14-13
Chapter 14 Specification Performance
NOTE
Respiration Rate (Pulse Wave Analysis)is an optional function of theDS-8007.
Masimo Unit
SpO2
Measurement Range 1%SpO2 to 100%SpO2
Measurement Accuracy Adult: ±2%SpO2 when 70%SpO2 to 100%SpO2 (When DS-100A is used)
Neonate: ±3%SpO2 when 70%SpO2 to 100%SpO2
SpCO
Resolution 1%SpCO
Measurement Accuracy ±3%SpCO (SpCO: 1%SpCO to 40%SpCO)
SpMet
SpHb
Measurement Range 0 g/dL to 25.0g/dL
Resolution 0.1g/dL
Measurement Range 0.02% to 20.0% (when disposable sensor is used) / 0.05% to 20.0% (when reusable sensor is
used)
SpOC
Pulse Rate
Measurement Accuracy ±3 bpm when 26 bpm to 239 bpm (without body motion)
Measurement Response 7 levels: 2 to 4 sec., 4 to 6 sec., 8 sec., 10 sec., 12 sec., 14 sec., 16 sec. (averaging duration)
Time
14-14
Chapter 14 Specification Performance
NOTE
The SpO2 measurement accuracy is determined based on the values of the root-mean-
square (rms) difference between SpO2 readings of the pulse oximeter device and values of
SaO2 determined with a CO oximeter, by healthy adult volunteers. The pulse oximeter
device measurements are statistically distributed; ±2% measurement accuracy means that
only about two-thirds of pulse oximeter device measurements can be expected to fall within
±2% of the value measured by a CO-oximeter.
PVI, SpCO, SpMet, SpHb, SpOC measurements are optional functions.
Blood Pressure
Resolution 1 mmHg
PR Accuracy ±5%
14-15
Chapter 14 Specification Performance
Covidien Unit
CO
BIS
BISx (Medtronic)
Resolution 1%
Resolution 0.1%
14-16
Chapter 14 Specification Performance
EMG
Resolution 1 dB
Frequency Characteristic
Sidestream Method
Measure CO2, N2O, Volatile Anesthetic: Infra-Red Solid-State Method
ment O2: Paramagnetic Method
Method
Warm-Up Time Multi Gas Module
ISO Accuracy: 45 sec.
Full Accuracy: 10 min.
Auto Zeroing Multi Gas Module
ISO Accuracy: 30 sec.
Full Accuracy: 4 hours
Measure CO2: 0vol% to 10.0vol% (0mmHg to 76mmHg, 0kPa to 10kPa)
ment
Range
N2O: 0vol% to 100vol%
O 2: 0vol% to 100vol%
N2O: 0 to 20[vol%]:±2[vol%]
20 to 100[vol%]:±3[vol%]
O 2: MGU-801P/MGU-811P0 to 25[vol%]:±1[vol%]
25 to 80[vol%]:±2[vol%]
80 to 100[vol%]:±3[vol%]
14-17
Chapter 14 Specification Performance
0 to 1[vol%]:±0.15[vol%]
1 to 5[vol%]:±0.2[vol%]
Sevoflurane
0 to 1[vol%]:±0.15[vol%]
1 to 5[vol%]:±0.2[vol%]
5 to 8[vol%]:±0.4[vol%]
Desflurane
0 to 1[vol%]:±0.15[vol%]
1 to 5[vol%]:±0.2[vol%]
5 to 10[vol%]:±0.4[vol%]
10 to 15[vol%]:±0.6[vol%]
15 to 18[vol%]:±1.0[vol%]
*1: This is the maximum influence within the gas level of specified measurement accuracy. The total influence will not
exceed 5% of the gas level.
*2: For CO2, N2O, O2, the influence from mixed agent is the same as that from single agent.
14-18
Chapter 14 Specification Performance
N2O 250ms
O2 At flow rate of 200 mL/min: 500 ms (15% to 21%) , 700 ms (21% to 60%)
At flow rate of 120 mL/min: 600 ms (15% to 21%) , 800 ms (21% to 60%)
Enflurane 350ms
AWP [cmH2O]
Measurement Range: 1.5 L/min to 100 L/min (Adult) , 0.25 L/m to 25 L/min (Child*)
Compliance [mL/cmH2O]
Measurement Range: 4 m/cmH2O to 100 m/cmH2O (Adult) , 1 m/cmH2O to 100 m/cmH2O (Child*)
I:E Ratio
14-19
Chapter 14 Specification Performance
Output Voltage ECG Output 1 V/mV (fixed), BP Output 1 V/100 mmHg (fixed)
BP Output: DC to 40 Hz
Pacemaker Pulse No
Delay Time 35 ms and below (when the "Filter" setting is [Monitor] or [Diag.])
NOTE
The delay time of analog waveform output and QRS synchronization output depends on the
filter setting and the input waveform type. For details, refer to your nearest service
representative.
The QRS synchronized signal is not intended to be used as synchronized signal for
defibrillator. When using the QRS synchronized signal, refer to your nearest service
representative.
14-20
Chapter 14 Specification Measurement Unit for Each Parameter
The measurement units of the displayed numeric data for this device are as follows.
Description Parameter Display Unit Default Unit
PACE beat/minute
CO CO CO L/minute
Blood Temperature Blood Temperature Tb Celcius
14-21
Chapter 14 Specification Measurement Unit for Each Parameter
14-22
Chapter 14 Specification Measurement Unit for Each Parameter
14-23
Chapter 14 Specification Measurement Unit for Each Parameter
Electromyograph EMG dB
Suppression Ratio SR %
Stroke Volume SV mL
Stroke Volume Index SVI mL/m2
Systemic Vascular
SVRI dyn x sec x cm-5 x m2
Resistance Index
14-24
Chapter 14 Specification Alarm Limit Range for Each Parameter
Pulmonary Vascular
PVPI ---
Permeability Index
Transcutaneous Oxygen
tcpO2 mmHg, kPa *
TCM4, TCM5 FLEX Partial Pressure
Data Transcutaneous Carbon
tcpCO2 mmHg, kPa *
Dioxide Partial Pressure
*: The measurement unit of tcpO2, tcpCO2 can be set on the TCM4 or TCM5 FLEX. When the measurement unit is
changed, the tabular trend data of tcpO2 and tcpCO2 on the bedside monitor will be deleted.
Adjustable Increments
1 mm increments
2 Bpm increments
14-25
Chapter 14 Specification Alarm Limit Range for Each Parameter
Adjustable Range
Adjustable Increments
RR_SpO2 (Adult) 5 Bpm to 30 Bpm 10 Bpm to 35 Bpm
5 Bpm increments
N/A
RR_SpO2 (Child) 6 Bpm to 32 Bpm 8 Bpm to 34 Bpm
2 Bpm increments
1 cmH2O increments
1 mmHg increments
Upper: current value +10 mmHg (+1.3 kPa
0.1 kPa to 13.1 kPa 0.3 kPa to 15.0 kPa / +1.3%)
0.1 kPa increments Lower: current value -10 mmHg (-1.3 kPa
/ -1.3%)
0.1% to 13.1% 0.3% to 15.0%
0.1% increments
1 mmHg increments
- 0.1% to 3.0%
0.1% increments
1.0°F increments
14-26
Chapter 14 Specification About the SpO2 Clinical Test
Adjustable Range
Adjustable Increments
SpHb 1.0 g/dL to 24.0 g/dL 2.0 g/dL to 24.5g/dL
N/A
0.1 g/dL increments
BIS 1 to 98 2 to 99
N/A
increments of 1
*: If the value exceeds the adjustable range, the limit within the range will be set.
The automatic setup will not be performed for the turned OFF limit.
Medtronic Unit
The SpO2 and pulse rate measurement accuracy have been validated for each range by testing on healthy adult male
and female volunteers against a laboratory CO-Oximeter.
The SpO2 accuracy has been validated for the range from 70% to 100% by testing on healthy adult male and female
volunteers (age: 19 to 48 years old) with light to dark skin pigmentation. The standard deviation is ±2% which
encompasses 68% of the population.
The pulse rate accuracy has been validated for the range from 70% to 100% by testing on healthy adult male and
female volunteers (age: 19 to 48 years old) with light to dark skin pigmentation The standard deviation is ±3 bpm
which encompasses 68% of the population.
These clinical test data are disclosed based on the data provided from Medtronic.
Masimo Unit
The SpO2, SpCO, SpMet, and pulse rate measurement accuracy have been validated for each range by testing on
healthy adult male and female volunteers against a laboratory CO-Oximeter.
SpO2 and SpMet accuracy have been validated by testing on 16 neonatal NICU patients ranging in age from 7 days
to 135 days old and weighing between 0.5 kg to 4.25 kg. Seventy-nine (79) data samples were collected over a range
of 70% to 100% SpO2 and 0.5% to 2.5% HbMet with a resultant accuracy of 2.9% SpO2 and 0.9% SpMet.
The SpO2 accuracy has been validated for the range from 70% to 100% by testing on healthy adult male and female
volunteers (age: 21 to 36 years old) with light to dark skin pigmentation. Without body motion, the standard
deviation is ±2% which encompasses 68% of the population. With body motion, the standard deviation is ±3% which
encompasses 68% of the population. For the validation, frictional or contact motion of 1 cm to 2 cm, and random
vibration of 1 Hz to 5 Hz were tested.
The pulse rate accuracy has been validated for the range from 70% to 100% by testing on healthy adult male and
female volunteers (age: 24 to 37 years old) with light to dark skin pigmentation The standard deviation is ±3 bpm
which encompasses 68% of the population.
The SpCO accuracy has been validated for the range from 0% to 40% by testing on healthy adult male and female
volunteers with light to dark skin pigmentation. The standard deviation is ±3% which encompasses 68% of the
population.
The SpMet accuracy has been validated for the range from 0% to 15% by testing on healthy adult male and female
volunteers with light to dark skin pigmentation. The standard deviation is ±1% which encompasses 68% of the
population.
The SpHb accuracy has been validated for the range from 8 g/dL to 17 g/dL by testing on healthy adult male and
female volunteers with light to dark skin pigmentation. The standard deviation is ±1 g/dL which encompasses 68%
of the population.
These clinical test data are disclosed based on the data provided from Masimo.
14-27
Chapter 14 Specification About the SpO2 Clinical Test
14-28
: SD Logo is a trademark of SD-3C, LLC.
Microstream, FilterLine and CapnoLine are trademarks or registered trademarks of Oridion Medical 1987 Ltd.
X-Cal is a trademark of Masimo Corporation.
Other company and product names used in this manual are trademarks or registered trademarks of respective companies.
3-39-4 Hongo, Bunkyo-ku, Tokyo 113-8483 Japan
Tel: +81-3-5684-1455 Fax: +81-3-3814-1222
https://fukuda.com/