DK1758 ch06
DK1758 ch06
DK1758 ch06
I. INTRODUCTION
Figure 1 illustrates the overall structure and many of the important structural parameters
for the 0.18-mm NMOSFET (1). The eight key electrical parameters listed and de®ned in
Table 1 were chosen to characterize the optimal device's electrical performance. In the
table, DIBL means ``drain induced barrier lowering,'' which is a short-channel effect. The
primary goal was to design a device that showed maximum drive current (at least 450 mA/
mm) while satisfying the targets in the table for peak off-state leakage, DIBL, peak sub-
strate current (to ensure hot-carrier reliability), etc. The optimal device was meant to be
broadly representative of industry trends, although this is a relatively low-power transistor
due to the 50-pA/mm limit on the leakage current. Due to the short gate length of the
device, it was necessary to include a boron ``halo'' implant as part of the device structure,
in order to obtain an optimal combination of turnoff and drive current performance for
the device. The effectiveness of the halo implant in suppressing short-channel effects as
well as maintaining hot-carrier reliability has been previously reported (2,3). This implant,
along with a boron VT adjust channel implant, was found to improve both the VT rolloff
with decreasing channel length and the device reliability while maintaining acceptable Idsat
vs. Ileak characteristics of the device. Because of the 1.8-V power supply
(Vdd 1:8 V nominal) assumed for this technology, consideration was given to ensuring
hot-carrier reliability. This was done through the use of a device in which shallow source-
drain (S/D) extensions were doped with a peak concentration of 4 1019 cm 3 and were
self-aligned to the edge of the oxide grown on the polysilicon gate (15 nm from the
polysilicon edge). The deep S/D regions were self-aligned to the spacer oxide edge and
had a junction depth of 150 nm, which was held constant throughout the analysis.
The process simulators, TSUPREM-3 (4) (one-dimensional) and TSUPREM-4 (5)
(two-dimensional), were used to generate the doping pro®les for the various regions of the
device. Due to the uncertain accuracy of two-dimensional diffusion models for arsenic-
implanted junctions with short thermal cycles, the one-dimensional vertical pro®le of both
the shallow and deep S/D junctions was simulated using TSUPREM-3. For each junction,
the two-dimensional pro®le was then generated by extending the vertical pro®le laterally
using a complementary error function with a characteristic length corresponding to 65%
of the vertical junction depth. Conversely, the two-dimensional halo implant pro®le was
directly simulated using TSUPREM-4. The VT adjust implant vertical pro®le was simu-
Figure 1 Schematic cross section of the 0:18-mm NMOSFET structure. The nominal values of the
structure parameters and the maximum variations that were used in the sensitivity analysis are listed
in Table 2. The polysilicon reoxidation thickness, tre-ox , was ®xed at 15 nm for all simulations.
Threshold voltage (from extrapolated linear I-V, @ Vd 0:05 V), VT (V) 0:5
Drive current (@ Vg Vd Vdd ), Idsat (mA=mm of device width) 450
Peak off-state leakage current (@ Vd 2 V, Vg 0, T 300K), Ileak 50
(pA/mm of device width)
DIBL (Vt @ Vd 0:05 V Vt @ Vd Vdd ), VT (mV) 100
Peak substrate current (@ Vd Vdd ), Isub (nA/mm of device width) < 200
Subthreshold swing (@ Vd 0:05 V), S (mV/decade of Id ) 90
Peak transconductance (@ Vd 2:0 V); gsm (mS/mm of device width) 300
Peak transconductance (@ Vd 0:05 V); glm (mS/mm of device width) 30
lated using TSUPREM-3, and was then extended laterally without change over the entire
device structure. A composite pro®le containing all the foregoing individual pro®les was
generated and imported to the device simulator, UT-MiniMOS (6) (where UT stands for
University of Texas at Austin and UT-MiniMOS is a version of the MiniMOS device
simulator with modi®cations from UT). UT-MiniMOS was chosen to simulate the device's
electrical characteristics because it has both the UT hydrodynamic (HD) transport model
based on nonparabolic energy bands and the UT models for substrate current (7), quan-
tum mechanical effects (8±10), and mobility in the inversion layer (11±13). Also, UT-
MiniMOS has adaptive gridding capability, and this capability was used to adapt the
grid to the potential gradient and the carrier concentration gradients during the
simulations.
The optimal device structure was determined by examining a large number of simu-
lated devices with different halo peak depths and doses. For each value of the halo peak
depth and dose, the boron VT adjust implant (also called the channel implant) dose was
adjusted to satisfy the requirement that the maximum off-state leakage current is 50 pA/
mm at room temperature (see Table 1). A number of simulations were performed to
examine the ranges of variation. The result of these simulations was the selection of a
boron halo implant dose of 1:5 1013 cm 2 with a peak doping pro®le depth of 80 nm and
a boron channel implant dose of 5:65 1012 cm 2 in order to obtain maximum drive
current while meeting all the other targets in Table 1.
In Table 2 the nine key structural and doping parameters for the optimal device are
de®ned, and the optimal value for each is listed. For Lg , Tox , Tsp , Xj sh , Nsh , and Rs , the
optimal values were selected from technology and scaling considerations, and the values
chosen are broadly representative of industry trends. For Nch , Nhalo , and d, the optimal
values were determined from simulations aimed at de®ning an optimal device structure,
as explained earlier.
A primary aim of this analysis was to obtain a set of complete, second-order empirical
model equations relating variations in the structural and doping (input) parameters of the
0:18-mm NMOSFET to the resulting variations in the key device electrical characteristics
listed in Table 1. (This technique is also known as response surface methodology (14).) In
this analysis, the nominal or design center device was identical to the optimal NMOSFET
from the previous section, and the variations were with respect to this nominal device.
Hence, the ``optimal'' values of the input parameters in Table 2 are also the ``nominal''
values for this analysis. Also listed in Table 2 are the maximum variation limits for each
input parameter. Since the model equations are accurate only for variations less than or
equal to these maximum limits, these limits were intentionally chosen to be large to give a
wide range of validity to the model equations. However, typical IC manufacturing lines
have manufacturing statistical variations considerably less than the maximum variations
listed in the table. In the next section, Monte Carlo simulations employing the model
equations were used to explore the impact of smaller, more realistic variations.
A three-level Box±Behnken design (15) was performed in order to obtain the
responses of the output parameters to the input parameters. Besides the centerpoint,
where all factors were maintained at their nominal values, the other data points were
obtained by taking three factors at a time and developing a 23 factorial design for
them, with all other factors maintained at their nominal values. The advantage of this
design was that fewer simulations were required to obtain a quadratic equation as com-
pared to other designs. A total of 97 simulations (96 variations plus the one nominal
device simulation) was required for this analysis for the case of nine input factors. One
drawback of this design, however, is that all of the runs must be performed prior to
obtaining any equation, and it is not amenable to two-stage analyses. Hence, there is
no indication of the level of factor in¯uence until the entire experiment has been
conducted.
In contrast to the nine input parameters that were varied (see Table 2), several device
parameters, such as the deep S/D junction pro®le and its peak doping, were held constant
throughout all of the simulations. In addition, a background substrate doping of
5 1015 cm 3 and an interface charge of 3 1010 cm 2 were uniformly applied. The
eight key device electrical characteristics listed in Table 1 were the response variables.
After the completion of the 97 simulations, two sets of model equations were generated
for each response, one in terms of the actual values of the input parameters, and the other
in terms of their normalized values. The normalized values were calculated using the
following equation:
Because the input variables are dimensionless, the coef®cients are independent of
units.
The relative importance of any term is determined solely by the relative magnitude of
the coef®cient of that term. For example, in the model equation for the satura-
tion drive current, Idsat is most sensitive to the normalized gate length (Lg ),
followed by the oxide thickness (Tox ), the shallow junction depth (Xsh ), the
spacer oxide width (Tsp ), and the channel dose, Nch . Also, this attribute of
the normalized equations simpli®es the generation of reduced equations by
dropping less signi®cant terms.
For all the normalized parameters, the mean value is zero and, as will be explained
later, the maximum value of the standard deviation is 1/3.
Monte Carlo
Response (key device Target value for simulated value for
electrical characteristics) Critical parameter critical parameter critical parameter
Note: The bold font indicates critical parameter values which do not meet their respective target.
run to generate the pdf's of the characteristics. The pdf's were then analyzed to obtain the
mean and standard deviation of each of the device electrical characteristics.
In Table 3, the de®nition of critical parameter and the target value for this parameter
are listed for each of the responses. The target values are meant to be broadly representa-
tive of industry trends. The critical parameter is either the 3s statistical variation or the
maximum or minimum value of the response, where the maximum value is calculated as
the mean value 3s statistical variation], while the minimum value is calculated as the
mean value 3s statistical variation]. In this set of Monte Carlo simulations, all the input
parameter variations, the siN 's were set to the maximum value of 1/3, corresponding to the
``Maximum variations'' in Table 2. Figure 2 shows a typical pdf, for the substrate current,
Isub . The mean value and the standard deviation, s, are listed at the top. The crosses
indicate the Monte Carlo±simulated pdf, while the solid curve is a ®tted Normal prob-
ability distribution with the same mean and s. The pdf is clearly not a Normal distribu-
tion, although the input parameter statistical distributions are Normal. The non-Normal,
skewed pdf for Isub (and the other responses) is due to the nonlinear nature of the model
equations (17), and the amount of skew and the departure from the Normal distribution
vary considerably from response to response. The Monte Carlo simulation results are
listed in Table 3, where the mean value and s from the Monte Carlo simulations were
used to calculate the ``simulated value'' in the last column. The targets for the ®rst four
responses in the table (VT , VT due to DIBL, Idsat , and Ileak ) were not met, but the targets
for the last four parameters in the table (Isub , S, gsm , and glm ) were met. To bracket the
problem, and to determine whether the targets for the ®rst four responses are realistic, all
the input parameter variations were reduced in two stages, ®rst to a set of more ``realistic''
values and second to a set of ``aggressive'' values. These sets are listed in Table 4; they
re¯ect the judgment of several SEMATECH experts (18). In the table, both the statistical
variations of the normalized input parameters, 3siN , and the corresponding statistical
variations in percentage terms of the input parameters are listed. A Monte Carlo simula-
tion was performed for each of these sets of variations, and the simulation results are listed
in Table 5. The targets for the third and fourth responses, Idsat and Ileak , were satis®ed with
the ``realistic'' input variations, but the targets for VT and VT were satis®ed only with
the ``aggressive'' input variations. The conclusion is that the targets for all the responses
can probably be met but that it will be especially dif®cult to meet them for VT and VT
(DIBL).
where yi is one of the eight key electrical device characteristics and Ai , Bij , Cijk , and Dij are
coef®cients. For the optimal value of yi from the design optimization (denoted by yi;opt ),
all the xj 's are zero, since yi;opt corresponds to all input parameters at their nominal
values and hence all xj 's set to zero. Then
yi;opt Ai 3
However, for the mean value of yi , denoted by hyi i:
X X X
hyi i Ai Bij hxj i Cijk xj xk Dij xj 2 4
j j;k;j6k j
However, for each xj the probability distribution is the Normal distribution centered
about zero. Because this distribution is symmetric about its center, hxj i
h xj xk i 0, since xj and xj xk are odd functions. On the other hand, xj 2
is an even function; hence h xj 2 i 6 0 19. In fact, s2jN h xj 2 i hxj i2 , and since
2 2
hxj i 0, h xj i sjN . Hence,
X
hyi i yi;opt Dij s2jN 5
j
Clearly, the nonlinear, second-order relationship between the responses and the input
parameters causes a shift in the mean value of the responses from their optimal values.
Using Eqs. (3) and (5):
P 2
hyi i yi;opt j Dij sjN
6
yi;opt Ai
The right-hand side of Eq. (6) can be used as a metric to evaluate the expected
relative difference between the mean value and the optimal value for any of the responses.
This metric, call it the expected shift of the mean, can be directly evaluated from the
normalized model equation before a Monte Carlo simulation is run to determine hyi i.
After such a simulation is run and hyi i is determined, the expected shift of the mean can be
compared to the ``actual shift of the mean,'' hyi i yi;opt =yi;opt . These calculations were
done for the case where all the normalized input parameter statistical variations are
maximum, 3siN 1. The results are listed in Table 6 for all the responses. For all the
responses except the leakage current, the absolute value of the actual shift of the mean is
small, at less than 5% in all cases and less than 1% in most, and the expected and actual
shifts are quite close to each other. Even for leakage current, the actual shift of the mean is
a tolerable 11%, but the expected and actual shifts are relatively far apart.
Next, Monte Carlo simulation was used to meet the targets for the output para-
meters with an optimal set of reductions in the input parameter statistical variations. Each
input parameter statistical variation was reduced in steps, as listed in Table 7. (Note that,
for each input parameter, the maximum variation is the same as that used in the previous
section [see Table 2] and earlier in this section [see Table 4], and the minimum is half or less
than half of the maximum.) The straightforward approach is to run a series of Monte
Carlo simulations covering the entire range of possible combinations for the input para-
meter variations. However, the number of simulations is 46,656 for each response (see
Table 7), an unreasonably high number. In order to reduce the number of simulations to a
more manageable total, the following procedure was used. For each response, the normal-
ized model equation was examined to select those input parameters that are either missing
from the equation or included only in terms with small coef®cients. Since, as noted pre-
viously, these inputs are unimportant in in¯uencing the response, the variation was held
®xed at its maximum value for each of these selected parameters. As shown in Table 8,
following this procedure, two parameters were selected for each response, and hence the
number of Monte Carlo simulations was reduced to a more manageable 3888 or 5184.
Since each Monte Carlo simulation took about 3 seconds to run on a Hewlett-Packard
workstation, the total simulation time was about 3±4 hours for each response. (Table 8
does not include listings for Isub , S, gml , or gsm , since those responses are within speci®ca-
tion for the maximum values for the variation of all the input parameters, as shown in
Table 3.) The outputs from the Monte Carlo simulations were imported to a spreadsheet
program for analysis and display. By utilizing the spreadsheet capabilities, the input
Lg , Xsh , Nsh , d, 1 (10%) 1/2 (5%) 1/4 (2.5%) 3 36 729
Nhalo , Nch
Lg , Tsp , Rs 1 (15%) 0.3 (4.5%) 0.233 (3.5%) 4 43 64
ÐÐÐÐ
Total number of 64 729 46,656
combinations
a
For each 3siN , the corresponding statistical variation as a percentage of the mean value of the non-normalized
input parameter is in parentheses.
No. of
Response Lg Tox Tsp Xsh Nsh d Nhalo Nch Rs combinations
VT 4 3 4 3 3 1 3 3 1 3888
VT 4 3 4 3 3 3 3 1 1 3888
Idsat 4 3 4 3 1 1 3 3 4 5184
Ileak 4 3 4 3 3 1 3 3 1 3888
Note: The bold font indicates that the number of steps has been reduced from the number in Table 7.
variations were then iteratively reduced from their maximum values to meet the targets for
the responses.
As already discussed, it was most dif®cult to meet the targets for VT and for VT
due to DIBL. Hence, these two were dealt with ®rst. From the size of the coef®cients in the
normalized model equation for VT , the terms containing Lg , Tox , Xsh , and Nch are
the most signi®cant. Thus, the statistical variations of only these parameters were reduced
to meet the VT target, while the variations of the other input parameters were held at their
maximum values. Contour plots of constant 3s variation in VT were determined using the
spreadsheet program. The results are shown in Figure 3, where the statistical variations of
Tox and Xsh were ®xed at their realistic values of 5% each (corresponding to 3siN 1=2),
and the statistical variations of Lg and Nch were varied. Along Contour 1, the 3s variation
in VT is 50 mV, and the variations of both Lg and Nch are less than 7.5%. Since these
variations are quite aggressive (see Table 4), the 50-mV target will be dif®cult to meet.
Along Contour 2, the 3s variation in VT is 60 mV. This target is realistic because the
variations of both Lg and Nch on the contour are achievable, particularly in the vicinity of
the point where the variations are about 9.5% for Lg and 7.5% for Nch (see Table 4).
Figure 4 also shows contours of constant 3s variation in VT ; the only difference from
Figure 3 is that the statistical variation of Xsh is 7.5%, not 5% as in Figure 3. The 60-mV
contour here, labeled Contour 3, is shifted signi®cantly to the left from the 60-mV contour
in Figure 3 and hence is much more dif®cult to achieve. For the case where the variation of
Tox is 7.5% while that of Xsh is 5%, the 60-mV contour is shifted even further to the left
than Contour 3. The contour plots can be utilized to understand quantitatively the impact
of the statistical variations of the key input parameters and how they can be traded off to
reach a speci®c target for VT variation. Looking particularly at Contour 2 in Figure 3, and
utilizing ``realistic'' values of the variations as much as possible (see Table 4), an optimal
choice for the variations is 5% for Tox and Xsh , 7.5% for Nch , and 9.5% for Lg .
Next, the requirements to meet the target for VT due to DIBL were explored.
From the size of the coef®cients in the normalized model equation for VT , the terms
containing Lg , Xsh , Tsp , and Tox are the most signi®cant. Thus, the variations of only
these parameters were reduced to meet the VT target, while the variations of the other
input parameters were held at their maximum values. Figure 5 shows contours of constant
(20), the gate CD (critical dimension) control is 10%. This approach can also be used to
determine tradeoffs. If, for example, the process variation of Tsp can only be controlled to
12%, it is evident from Contour 4 of Figure 5 that the control of Lg would have to be
tightened so that its process variation is 8% or less.
For process control purposes, let UL be the upper speci®cation limit for the process,
let LL be the lower speci®cation limit, and let MEAN be the mean value for the structural
and doping parameters. A very important quantity is the ``process capability,'' Cp . For the
ith input parameter, Cp UL LL=6si , where si is the standard deviation of the ith
(non-normalized) input parameter. The goal is to control the process variations and the
resulting si so that Cp 1. For Cp much less than 1, a non-negligible percentage of the
product is rejected (i.e., noticeable yield loss) because of input parameter values outside
the process limits, as illustrated schematically in Figure 8. For Cp 1, the statistical
distribution of the input parameter values is largely contained just within the process
limits, so the cost and dif®culty of process control are minimized, but very little product
is rejected because of input parameter values outside the process limits. Finally, for Cp
much larger than 1, the actual statistical distribution of the input parameter is much
narrower than the process limits, and hence very little product is rejected, but the cost
and dif®culty of process control are greater than for the optimal case, where Cp 1. In
practice, because of nonidealities in IC manufacturing lines and the dif®culty of setting
very precise process limits, the target Cp is typically somewhat larger than 1, with 1.3 being
a reasonable rule of thumb (21,22). In practical utilization of the previous Monte Carlo
results, especially the optimal set of variations, it makes sense to set
UL MEAN optimal 3si variation and LL MEAN optimal 3si variation) for
all of the key structural and doping parameters. Using these formulas, the values of LL
and UL are listed in Table 9 for all the input parameters.
Meeting the process control requirements for the statistical variation of the ®ve key
input parameters is dependent on the control at the process module level. For example, to
meet the channel implant dose (Nch ) requirement, the channel implant dose and energy as
well as the thickness of any screen oxide must be well controlled. As another example, to
meet the Tox requirement, the gas ¯ows, temperature, and time at temperature for a
furnace process must be well controlled. Through empirical data or simulations, the
level of control of the process modules necessary to meet the requirements on the input
parameter statistical variations can be determined. Of course the tighter the requirements
on the input parameter variations, the tighter the required level of control of the process
modules.
V. METROLOGY REQUIREMENTS
The metrology requirements are driven by the process control requirements, i.e., the UL
and LL in Table 9 for the input parameters. In-line metrology is used routinely in the IC
fabrication line to monitor these parameters, to ensure that they stay between the LL and
UL, or to raise an alarm if they drift out of speci®cation. For in-line metrology on a well-
established, well-characterized line, the most important characteristic is the measurement
precision, P, where P measures the repeatability of the measurement. For a measurement
of a given parameter with a particular piece of measurement equipment (for example, a
measurement of Tox using a particular ellipsometer), P is determined by making repeated
measurements of Tox on the same wafer at the same point. P is de®ned to be 6sMETROL ,
where sMETROL is the standard deviation of the set of repeated measurements (23). A key
parameter is the ratio of measurement precision, P, to the process tolerance, T, where
T UL LL. Then P=T 6sMETROL = UL LL). It is important that P=T 1, or
measurement errors will reduce the apparent Cp (or, equivalently, the apparent process
standard deviation will increase). This point is illustrated in Figure 9, where a simpli®ed
process control chart for one of the key doping and structural parameters, such as the gate
oxide thickness or gate length, is shown. The position of each X indicates the value of an
in-line measurement (or the mean value of multiple measurements on one or several wafers
from a lot) if there were no metrology errors. The cross-hatched area around each X
indicates the uncertainty introduced because of the random errors from the metrology,
i.e., the in¯uence of the nonzero sMETROL . In the following, let sPROC be the statistical
standard deviation due to random process variations, as discussed in the previous sections.
In Case 1 and Case 2 in the ®gure, Cp 1, so 6sPROC UL LL). In Case 1, P=T 1,
Nsh cm 3
Channel dose, Nch cm 2 5:65 1012 cm 2 7:5% 4:24 1011 cm 2 6:07 1012 cm 2
5:23 1012 cm 2
Halo dose, Nhalo cm 2 1:5 1013 cm 2 10% 1:5 1012 cm 2 1:65 1013 cm 2
1:35 1013 cm 2
! 1=2
CP;TOTAL P 2
1 CP;PROC 7
CP;PROC T
Since (UL LL) is ®xed by speci®cation, then CP;TOTAL =CP;PROC sPROC =sTOTAL . A
plot of CP;TOTAL =CP;PROC versus CP;PROC , with P=T as a parameter varying from 0.1 to
1.0, is shown in Figure 10. This plot illustrates the impact of the measurement variation as
characterized by the parameter P=T. For a given value of CP;PROC (and hence of sPROC ),
CP;TOTAL decreases (and hence sTOTAL increases) rapidly with P=T; and since the goal is to
maximize Cp and minimize s, the increase of P=T imposes a signi®cant penalty.
An alternate way to evaluate the impact of P=T variation is shown in Figure 11,
where CP;TOTAL and P=T are the independent variables and CP;TOTAL =CP;PROC is plotted
as in Figure 10, but versus the independent variable, CP;TOTAL . The parameter P=T varies
from 0.1 to 0.8. As in Figure 10, since (UL LL) is ®xed by speci®cation, then
CP;PROC =CP;TOTAL sPROC =sTOTAL . Using the de®nitions of P=T, CP;TOTAL , CP;PROC ,
sPROC , and sTOTAL , the equation is
!1=2
CP;PROC sPROC P 2
1 CP;TOTAL 8
CP;TOTAL sTOTAL T
perhaps 0.3 for this case, the reduction is relatively small and tolerable, but for large P=T,
the required reduction is intolerably large. Note that the curves for P=T 0:4 go to zero at
some value of CP;TOTAL Cp0 , where Cp0 < 2:5, the maximum value plotted. This occurs
for P=TCp0 1 [see Eq. (8)], which corresponds to sMETROL sTOTAL , and hence
sPROC 0, which is impossible to achieve. In this case, the entire budget for random
variation is absorbed by the metrology variation, leaving none available for the random
process variations.
Tying the foregoing considerations together with the process speci®cation limits (UL
and LL) in Table 9, the required metrology precision as re¯ected by the sMETROL values
for P=T 0:1 and for P=T 0:3 are listed in Table 10. Note that in most cases, the
required precision is quite high. For example, for Tox , with P=T 0:1, the sMETROL
value of 0.0075 nm is less than 0.1 AÊ; even for P=T 0:3, the 0.023-nm value is just
over 0.2 AÊ. Similarly, for P=T 0:1, the sMETROL value for Lg is just over 5 AÊ and for
xj sh , the sMETROL value is less than 1 AÊ.
The in-line metrology is limited by practical issues. Ideally, the in-line measure-
ments are rapid, nondestructive, direct, and done on product wafers. However, in a
number of cases, test wafers are used, and sometimes the measurement is destructive.
Of the ®ve key input parameters requiring ``tight control'' (i.e., 3sPROC < 10%, as
discussed in the previous section and listed in Table 9), four parametersÐthe gate length
(Lg ), the gate oxide thickness (Tox ), the spacer oxide width (Tsp ), and the channel dose
(Nch )Ðcan be routinely and directly measured on product wafers. Lg and Tsp are mea-
sured optically or via SEM after etch, Tox is measured via ellipsometry, and Nch is
measured via the thermal wave re¯ectance (25) technique. Alternatively, Nch is some-
times monitored on test wafers using secondary ion mass spectrosopy (SIMS) or four-
point sheet resistance measurements, and Tox is sometimes measured on test wafers. The
sMETROL sMETROL
Input parameter Mean UL LL (P=T 0:1) (P=T 0:3)
Nch cm 2
Halo dose, 1:5 1013 cm 2
3 1012 cm 2
5 1010 cm 2
1:5 1011 cm 2
Nhalo cm 2
Halo peak depth, 80 nm 16 nm 0.27 nm 0.8 nm
d (nm)
Series resistance 400 -mm 120 -mm 2 -mm 6 -mm
(external), Rs -mm
other parameter requiring tight control, the S=D extension (``shallow'') junction depth
(Xsh ), cannot currently be routinely measured on product wafers. Consequently, Xsh is
typically monitored on test wafers via SIMS. However, in the future, the Boxer-Cross
technique (26) shows promise of becoming practical for in-line measurements on product
wafers. For the four input parameters requiring ``looser control'' (i.e., 3sPROC 10%, as
discussed in the previous section and listed in Table 9), none are routinely measured on
product wafers. For Nhalo and the halo peak depth (d), given the loose control needed
and the typically tight control on the halo implant dose and energy, monitoring test
wafers via SIMS is generally adequate. The peak shallow-junction doping (Nsh ) can be
monitored using the same SIMS that is used to measure Xsh . Finally, at the end of the
wafer fabrication process, routine electrical monitoring of test transistors on product
wafers is used to measure Rs . Typically, for each lot in an IC manufacturing line, the in-
line measurements are made on a reasonable sample size, and good statistics for the
standard deviation (sTOTAL ) and direct veri®cation of meeting process control limits are
obtained.
As mentioned earlier, for in-line metrology used for process control in well-estab-
lished and well-characterized IC fabrication processes, the key requirement is precision
(i.e., repeatability), since there is a well-established, optimal baseline, and the main
requirement is to ensure that the process does not drift unacceptably far from the
optimal. However, for establishment of new or strongly modi®ed process modules or
process ¯ows, it is necessary to measure the values of key parameters, such as Tox and
Lg , to a relatively high degree of absolute accuracy in order to understand and estab-
lish the optimal baseline. For example, in establishing the 0.18-mm NMOSFET process,
it is important that Tox be accurately measured so that the real Tox of the process can
be set close to 4.5 nm. (Note that this metrology with high absolute accuracy is often
VI. CONCLUSIONS
A 0.18-mm NMOSFET device was designed and optimized to satisfy a speci®ed set of
electrical characteristics. This optimized device was the nominal design center for a simu-
lated sensitivity analysis in which normalized second-order polynomial model equations
were embedded within a special Monte Carlo code. Monte Carlo simulations with the code
were used to correlate the random statistical variations in key electrical device character-
istics to the random variations in the key structural and doping parameters. Using these
simulations, process control tradeoffs among the different structural and doping para-
meters were explored, and the level of process control required to meet speci®ed statistical
targets for the device electrical characteristics was analyzed. It turns out that meeting these
targets requires tight control of ®ve key structural and doping parameters: the gate length,
the gate oxide thickness, the shallow source/drain extension junction depth, the channel
dose, and the spacer width. Making process control tradeoffs based on estimates of
industry capability, an optimal set of 3s statistical variations was chosen for the ®ve
parameters: 9%, 5%, 5%, 7.5%, and 8%, respectively. If the estimates of industry cap-
ability were different, the tradeoffs would be changed and hence the optimal set of varia-
tions would be changed.
The optimal set of parameter statistical variations drives the in-line metrology. The
key requirement is that the metrology precision for any measured parameter be no more
than 10±30% of the optimal statistical variation for that parameter. Also, the impact of
not meeting the precision requirements was quantitatively analyzed, and it was found that
the more the metrology precision departs from the requirements, the tighter the process
control must be.
ACKNOWLEDGMENTS
In the following equations, each of the eight output responses is given as a function of the
normalized set of input factors. Each of these input factors, for example, Lg , will have a
range of values between 1 and 1. Referring to Table 2, a 1 value for Lg would
correspond to a gate length, Lg , of 0:18 mm 15% 0:153 mm, while a 1 value for Lg
would indicate an Lg 0:18 mm 15% 0:207 mm. For the nominal case, each normal-
ized input variable would have a value of 0.
VT mV 454:9 56:1 Lg 52:4 Tox 41:4 Nch 33:6 Xsh 9:9 Nsh
9:7 Tsp 7:1 Nhalo 1:6 d 19:9 L2g 2
5:9 Tox 2
3:8 Tsp
3:8 d 2 21:9 Lg Xsh 6:9 Lg Tsp 6:2 Lg Nsh
VT mV 72:3 39:4 Lg 17:8 Xsh 12:7 Tsp 8:4 Tox 5:9 Nhalo
4:7 Nch 3:0 Nsh 1:5 d 12:4 L2g 7:8 Tsp
2 2
2:3 Nsh
11:9 Lg Xsh 5:9 Lg Tsp 5:3 Lg Tox 4:4 Tsp
d 2:4 Lg Nhalo 2:3 Tsp Nhalo
Note: In the next two equations, a logarithmic transformation was used to achieve better
normality and ®t.
Idsat logmA=mm 2:721 0:060 Lg 0:052 Tox 0:028 Xsh 0:019 Tsp
0:016 Nch 0:008 Nhalo 0:007 Rs 0:005 Nsh
0:015 L2g 0:009 Tox
2 2
0:009 Tsp 0:013 Xsh
Nch 0:011 Tsp Nch 0:008 Lg Tox
0:006 Lg Xsh
Ileak logpA=mm 0:385 1:189 Lg 0:571 Xsh 0:508 Nch 0:417 Tox
0:241 Tsp 0:144 Nsh 0:127 Nhalo 0:011 d
0:424 L2g 0:104 Tsp
2
0:080 d 2 2
0:063 Nsh
2
0:055 Tox 0:449 Lg Xsh 0:156 Lg Tsp
0:112 Lg Nsh 0:088 Lg Tox
Isub nA=mm 35:4 17:6 Lg 1:9 Tsp 1:7 Nch 1:7 Rs 1:5 Nsh
1:5 ; Tox 1:2 Xsh 6:2 L2g 2:0 Tox
2 2
1:7 Tsp
2 2
1:7 Xsh 1:5 Nch 3:2 Lg Nch
Note: In the following equation, an inverse square root transformation was used to achieve
better normality and ®t.
gsm mS=mm 470:6 50:4 Tox 41:7 Lg 14:4 Xsh 6:0 Nhalo 2:9 Nch
2 2
14:3 Xsh 12:5 Nhalo 20:2 Tox Nch
glm mS=mm 58:63 9:44 Lg 3:93 Tox 2:74 Xsh 1:69 Tsp 1:53 Rs
0:85 Nch 0:47 Nsh 0:43 d 0:34 Nhalo 1:68 L2g
2 2 2 2
1:36 Tox 0:94 Nhalo 0:69 Nch 0:64 Tsp 1:33 Lg
Xsh 0:78 Lg Tsp
REFERENCES
1. P.M. Zeitzoff et al. Modeling of manufacturing sensitivity and of statistically based process
control requirements for a 0.18-mm NMOS device. Proc. of the 1998 International Conference
on Characterization and Metrology for ULSI Technology, Gaithersburg, MD, pp. 73±81,
1998.
2. A. Hori, A. Hiroki, H. Nakaoka, M. Segawa, T. Hori. Quarter-micrometer SPI (self-aligned
pocket implantation) MOSFET's and its application for low supply voltage operation. IEEE
Trans. Electron. Devices, Vol. 42, No. 1, Jan. 1995
3. A. Chatterjee, J. Liu, S. Aur, P.K. Mozumder, M. Rodder, I.-C. Chen. Pass transistor designs
using pocket implant to improve manufacturability for 256Mbit DRAM and beyond. IEDM,
pp. 87±90, 1994.
4. Avanti Corp. User's Manual for TSUPREM-3.
5. Avanti Corp. User's Manual for TSUPREM-4.
6. UT-MiniMOS 5.2±3.0 Information Package. Microelectronics Research Center, The
University of Texas at Austin, 1994.
7. V. Martin Agostinelli, T. James Bordelon, Xiaolin Wang, Khaled Hasnat, Choh-Fei Yeap,
D.B. Lemersal, Al F. Tasch, Christine M. Maziar. Two-dimensional energy-dependent models
for the simulation of substrate current in submicron MOSFETs. IEEE Trans. Electron.
Devices, vol. 41, no. 10, Oct. 1994.
8. M.J. van Dort, P.H. Woerlee, A.J. Walker. A simple model for quantization effects in heavily
doped silicon MOSFETs at Inversion Conditions. Solid-State Electronics, vol. 37, no. 3, pp.
411±414, 1994.
9. S.A. Hareland, S. Krishnamurthy, S. Jallepalli, C.-F. Yeap, K. Hasnat, A.F. Tasch, Jr., C.M.
Maziar. A computationally ef®cient model for inversion layer quantization effects in deep
submicron N-channel MOSFETs. IEDM, Washington, DC, pp. 933±936, December 1995.
10. S.A. Hareland, S. Krishnamurthy, S. Jallepalli, C.-F. Yeap, K. Hasnat, A.F. Tasch, Jr., C.M.
Maziar. A computationally ef®cient model for inversion layer quantization effects in deep
submicron N-channel MOSFETs. IEEE Trans. Electron. Devices, vol. 43, no. 1, pp. 90±96,
Jan. 1996.