DC/DC Medical Converter TIM 6 Series, 6 Watt

• Compact DIP-24 plastic case


• I/O isolation 5000 VAC rated for 250
VAC working voltage
• Certification according to IEC/EN/ES
60601-1 3rd edition for 2 x MOPP
• Risk management process according to
ISO 14971 ES 60601-1 IEC 60601-1
UL 62368-1 IEC 62368-1
• Acceptance criteria for electronic
assemblies acc. to IPC-A-610 Level 3 The TIM 6 series is a cost efficient range of medical 6 Watt DC/DC converters
in DIP-24 plastic package with wide 2:1 input voltage range. They provide a re-
• Low leakage current <2 μA inforced isolation system for 5000 VAC and a very low leakage current of less
than 2 μA. The units are approved to IEC/EN/ES 60601-1 3rd ed. for 2 x
• Operating temperature –40°C to 95°C
MOPP and come along with an ISO 14971 risk management file. Design and
• EMC compliance to IEC 60601-1-2 4th production conform to the quality management system ISO 13485. With a high
edition and EN55032 class A efficiency of up to 89% and highest grade components the converters can reli-
ably operate in an ambient temperature range of –40°C up to +95°C. They
• Operating up to 5000m altitude
constitute a reliable solution not only for medical equipment but also for de-
• 5-year product warranty manding ranges of application such as transportation, control & measurement
or IGBT drivers.

Models
Order Code Input Voltage Output 1 Output 2 Efficiency
Range Vnom Imax Vnom Imax typ.
TIM 6-1211 5 VDC 1'200 mA 84 %
TIM 6-1212 12 VDC 500 mA 87 %
TIM 6-1213 9 - 18 VDC 15 VDC 400 mA 86 %
TIM 6-1221 (12 VDC nom.) +5 VDC 600 mA –5 VDC 600 mA 83 %
TIM 6-1222 +12 VDC 250 mA –12 VDC 250 mA 87 %
TIM 6-1223 +15 VDC 200 mA –15 VDC 200 mA 86 %
TIM 6-2411 5 VDC 1'200 mA 84 %
TIM 6-2412 12 VDC 500 mA 87 %
TIM 6-2413 18 - 36 VDC 15 VDC 400 mA 87 %
TIM 6-2421 (24 VDC nom.) +5 VDC 600 mA –5 VDC 600 mA 84 %
TIM 6-2422 +12 VDC 250 mA –12 VDC 250 mA 86 %
TIM 6-2423 +15 VDC 200 mA –15 VDC 200 mA 86 %
TIM 6-4811 5 VDC 1'200 mA 84 %
TIM 6-4812 12 VDC 500 mA 87 %
TIM 6-4813 36 - 75 VDC 15 VDC 400 mA 86 %
TIM 6-4821 (48 VDC nom.) +5 VDC 600 mA –5 VDC 600 mA 83 %
TIM 6-4822 +12 VDC 250 mA –12 VDC 250 mA 87 %
TIM 6-4823 +15 VDC 200 mA –15 VDC 200 mA 85 %

Options
on demand - Optional models with alternative pinning
(backorder with MOQ
non stocking item) - Optional models with alternative pinning and adjustable output voltage

www.tracopower.com January 12, 2024 Page 1 / 5


TIM 6 Series, 6 Watt

Input Specifications
Input Current - At no load 48 Vin models: 4 mA typ.
12 Vin models: 10 mA typ. (5 Vout model)
10 mA typ. (12 Vout model)
10 mA typ. (15 Vout model)
15 mA typ. (5 / -5 Vout model)
10 mA typ. (12 / -12 Vout model)
14 mA typ. (15 / -15 Vout model)
24 Vin models: 6 mA typ. (5 Vout model)
6 mA typ. (12 Vout model)
6 mA typ. (15 Vout model)
8 mA typ. (5 / -5 Vout model)
6 mA typ. (12 / -12 Vout model)
8 mA typ. (15 / -15 Vout model)
Surge Voltage 12 Vin models: 25 VDC max. (3 s max.)
24 Vin models: 50 VDC max. (3 s max.)
48 Vin models: 100 VDC max. (3 s max.)
Under Voltage Lockout 12 Vin models: 7 VDC min. / 8 VDC typ. / 8.8 VDC max.
24 Vin models: 15 VDC min. / 16 VDC typ. / 17.5 VDC max.
48 Vin models: 31.5 VDC min. / 33 VDC typ. / 34.5 VDC max.
Recommended Input Fuse 12 Vin models: 1'250 mA (slow blow)
24 Vin models: 630 mA (slow blow)
48 Vin models: 315 mA (slow blow)
(The need of an external fuse has to be assessed
in the final application.)
Input Filter Internal Pi-Type

Output Specifications
Output Voltage Adjustment –10% to +20% (15 Vout single models)
±10% (other single and dual output models)
(Only for optional models with adjustable output)
(By external trim resistor)
See application note: www.tracopower.com/overview/tim6
Output power must not exceed rated power!
Voltage Set Accuracy ±1% max.
Regulation - Input Variation (Vmin - Vmax) single output models: 0.2% max.
dual output models: 0.5% max.
- Load Variation (0 - 100%) single output models: 0.2% max.
dual output models: 1% max. (Output 1)
1% max. (Output 2)
- Voltage Balance dual output models: 2% max.
(symmetrical load)
- Cross Regulation dual output models: 5% max.
(25% / 100% asym. load)
Ripple and Noise - single output 5 Vout models: 50 mVp-p typ. (w/ 10 μF)
(20 MHz Bandwidth) 12 Vout models: 75 mVp-p typ. (w/ 10 μF)
15 Vout models: 75 mVp-p typ. (w/ 10 μF)
- dual output 5 / -5 Vout models: 50 / 50 mVp-p typ. (w/ 10 μF)
12 / -12 Vout models: 75 / 75 mVp-p typ. (w/ 10 μF)
15 / -15 Vout models: 75 / 75 mVp-p typ. (w/ 10 μF)
Capacitive Load - single output 5 Vout models: 1'500 µF max.
12 Vout models: 260 µF max.
15 Vout models: 210 µF max.
- dual output 5 / -5 Vout models: 860 / 860 µF max.
12 / -12 Vout models: 150 / 150 µF max.
15 / -15 Vout models: 110 / 110 µF max.
Minimum Load Not required

All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.

www.tracopower.com January 12, 2024 Page 2 / 5


TIM 6 Series, 6 Watt

Temperature Coefficient ±0.02 %/K max.


Start-up Time 35 ms typ. / 50 ms max.
Short Circuit Protection Continuous, Automatic recovery
Output Current Limitation 150% typ. of Iout max.
Overvoltage Protection 112 - 147% of Vout nom.
(depending on model)
5.6 VDC - 7 VDC (5 Vout single model)
13.5 VDC - 16 VDC (12 Vout single model)
18.3 VDC - 22 VDC (15 Vout single model)
5.6 VDC - 7 VDC (5 Vout dual model)
13.5 VDC - 18.2 VDC (12 Vout dual model)
17 VDC - 22 VDC (15 Vout dual model)
Transient Response - Peak Variation 150 mV typ. / 300 mV max. (25% Load Step)
(5.0 Vout models)
250 mV typ. / 350 mV max. (25% Load Step)
(other models)
- Response Time 250 µs typ. / 350 µs max. (25% Load Step)

Safety Specifications
Standards - IT / Multimedia Equipment EN 62368-1
IEC 62368-1
UL 62368-1
- Medical Equipment EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
2 x MOPP (Means Of Patient Protection)
- Certification Documents www.tracopower.com/overview/tim6
Pollution Degree PD 2
Over Voltage Category OVC II

EMC Specifications
EMI Emissions EN 60601-1-2 edition 4 (Medical Devices)
- Conducted Emissions EN 55011 class A (internal filter)
EN 55011 class B (with external filter)
EN 55032 class A (internal filter)
EN 55032 class B (with external filter)
- Radiated Emissions EN 55011 class A (internal filter)
EN 55011 class B (with external filter)
EN 55032 class A (internal filter)
EN 55032 class B (with external filter)
External filter proposal: www.tracopower.com/overview/tim6
EMS Immunity EN 60601-1-2 edition 4 (Medical Devices)
EN 55035 (Multimedia)
- Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A
Contact: EN 61000-4-2, ±8 kV, perf. criteria A
- RF Electromagnetic Field EN 61000-4-3, 10 V/m, perf. criteria A
- EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A
EN 61000-4-5, ±2 kV, perf. criteria A
Ext. input component: 3300µF, 25V, KZN || TVS 28V, 600W (12 Vin)
1200µF, 50V, KZN || TVS 58V, 600W (24 Vin)
390µF, 100V, KZN || TVS 120V, 600W (48 Vin)
- Conducted RF Disturbances EN 61000-4-6, 10 Vrms, perf. criteria A
- PF Magnetic Field Continuous: EN 61000-4-8, 100 A/m, perf. criteria A
1 s: EN 61000-4-8, 1000 A/m, perf. criteria A

General Specifications
Relative Humidity 95% max. (non condensing)

All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.

www.tracopower.com January 12, 2024 Page 3 / 5


TIM 6 Series, 6 Watt

Temperature Ranges - Operating Temperature –40°C to +105°C


- Approved Ambient Temp. +55°C max.
- Case Temperature +105°C max.
- Storage Temperature –55°C to +125°C
Power Derating - High Temperature Depending on model
See application note: www.tracopower.com/overview/tim6
Cooling System Natural convection (20 LFM)
Altitude During Operation 5'000 m max.
Regulator Topology Flyback Converter
Switching Frequency 225 - 275 kHz (PWM)
250 kHz typ. (PWM)
Insulation System Reinforced Insulation
Working Voltage (rated) 250 VAC
Isolation Test Voltage - Input to Output, 60 s 5'000 VAC
- Input to Output, 1 s 7'000 VDC
Creepage - Input to Output 8 mm min.
Clearance - Input to Output 8 mm min.
Isolation Resistance - Input to Output, 500 VDC 10'000 MΩ min.
Isolation Capacitance - Input to Output, 100 kHz, 1 V 15 pF typ.
17 pF max.
Leakage Current - Touch Current 2 µA max. (240 VAC / 60 Hz)
Distance Through Isolation 0.4 mm
Reliability - Calculated MTBF 3'942'000 h (MIL-HDBK-217F, ground benign)
Washing Process According to Cleaning Guideline
www.tracopower.com/info/cleaning.pdf
Environment - Vibration MIL-STD-810F
- Mechanical Shock MIL-STD-810F
- Thermal Shock MIL-STD-810F
Housing Material Non-conductive Plastic (UL 94 V-0 rated)
Base Material Non-conductive Plastic (UL 94 V-0 rated)
Potting Material Silicone (UL 94 V-0 rated)
Pin Material Tinned Copper
Pin Foundation Plating Nickel (2 - 3 µm)
Pin Surface Plating Tin (3 - 5 µm), matte
Housing Type Plastic Case
Mounting Type PCB Mount
Connection Type THD (Through-Hole Device)
Footprint Type DIP24
Soldering Profile Lead-Free Wave Soldering
260°C / 6 s max.
Weight 13.5 g
Thermal Impedance - Case to Ambient 18 K/W typ.
Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf
REACH SVHC list compliant
REACH Annex XVII compliant
- RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf
Exemptions: 7a, 7c-I
(RoHS exemptions refer to the component
concentration only, not to the overall
concentration in the product (O5A rule).)
- SCIP Reference Number 0df049e7-17e8-4c52-9786-9f341de06a8e

Supporting Documents
Overview Link (for additional Documents) www.tracopower.com/overview/tim6

All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.

www.tracopower.com January 12, 2024 Page 4 / 5


TIM 6 Series, 6 Watt

Outline Dimensions

Standard model
22.86 (0.900)

Pinout
31.8 (1.25)

20.3 (0.80)
10.6 (0.42) Pin Single Dual
15.24 (0.600)
1 +Vin
11 No pin Common
24 1
12 out No pin
(0.37)

23
9.5

2.54 (0.100)
2.0 (0.08)
13 +Vout out
15 No pin +Vout
Restricted area.
8.0 8.0 23
No copper traces in
(0.32) or components allowed. (0.32)
24 in
Ø0.60 (Ø0.024)
15
(0.37)

11
9.5

13 12

3.8 (0.15)
BOTTOM VIEW

Dimensions in mm (inch)
Tolerances: x.x ±0.5 (x.xx ±0.02)
x.xx ±0.25 (x.xxx ±0.010)
Pin diameter tolerance: ±0.10 (±0.004)

Alternative pinning
17.78 (0.700)

Pinout
31.8 (1.25)

20.3 (0.80)
10.6 (0.42) (without Trim option)
15.24 (0.600)
Pin Single Dual
2 in
10 No pin
(0.37)

23 2
9.5

22 11 NC out
2.54 (0.100)
4.5 (0.18)

14 +Vout
Restricted area.
8.0 No copper traces
8.0 16 out Common
(0.32) or components allowed. (0.32)
22 +Vin
23 +Vin
16 10
(0.37)
9.5

11
NC: Not connected
14
Ø0.60 (Ø0.024)
(with Trim option)
3.8 (0.15)
BOTTOM VIEW Pin Single Dual
Dimensions in mm (inch) 2 in
Tolerances: x.x ±0.5 (x.xx ±0.02) 10 Trim
x.xx ±0.25 (x.xxx ±0.010)
Pin diameter tolerance: ±0.10 (±0.004) 11 No pin out
14 +Vout
16 out Common
22 +Vin
23 +Vin

© Copyright 2023 Traco Electronic AG Specifications can be changed without notice.


Rev. January 12, 2024
www.tracopower.com Page 5 / 5