Ⅰ. Specification:
Rechargeable battery inside, you can use the scale during recharging.
Auto zero-setting when switch on the scale.
Auto zero-tracing.
Switch among weighing, counting, percentum. Function of choosing Weight range.
Power supply: AC 220V (±10%) / 50 Hz, 110V/60HZ, DC 4V / 4Ah (Rechargeable battery inside).
Working temperature range: 0 ℃ ~ +40 ℃/
Relative humidity: ≤ 85 % RH
Ⅱ. Keyboard instruction:
1. 【Cali.】 To Calibrate the scale when the scale is moved to a different place.
2. 【Mode】 To choose different function model: Weighing, Counting, percentum.
3. 【Unit】 To choose different unit: kg, lb.
4. 【Set】 To choose high and low limits range in weighing model.
To sample the weight in counting and percentum model.
5. 【Zero】 To make the display zero when there’s nothing on the platform but the display is not zero. And press this
key continuously for 2-4 second, the backlight will be switched on or off with the LCD model.
6. 【Tare】 To delete the tare.
7. 【 ↑ 】 To input number.
8. 【Enter】 To make sure.
Ⅲ. Operation:
The scale must to be placed on the stable and flat surface, adjust its four screws carrier pole to horizontal, if demand,
put the special plate, then switch on the power. The scale cannot be used for a long time under tare state or the function of
zero tracing will disappear and the zero will move.
The scale will go into the weighing model after the beginning display. The signals of “Zero” and “kg” set. You may
press【Mode】to choose the three model: weighing, counting, percentum circularly.
Backlight on/off: This just effective in the LCD model. Press【Zero】key continuously for 2-4 second, the backlight
will be switched on or off.
【Weighing mode】
Choosing Unit Function: Press【Unit】to choose the unit between “kg” and “lb”. The corresponding signal sets on display.
Tare Function:
Put the container on the platform. Press【Tare】to make the weight window display “Zero” after the weight is stable. When
the “Tare” signal sets, the weight displays is Net Weight. Remove the weight, the weight window displays negative.
Press【Tare】again, the weight window displays “Zero”. The “Tare” signal disappears.(The range of tare setting: ≤1/3F.S.)
Zero Function: You may press【Zero】to make display zero when the zero moved when in use.(Zero range: <=4%FS, it is
invalid under tare state.)
【Counting mode】
Press【Set】to sample weight in the counting model. It displays “SAP X” (X is the sample number).
1.Press【↑】to choose samples number from 10, 20, 50,100 and 200. Press【Enter】to make sure.
2.The weight window displays “LOAD-C” after the sample number is chosen. Put enough number of samples then
press【Enter】to finish sample.
3.There are two instance of unit weight lacking:
E-59
a. When the weight window displays “-LAC-”, it means the unit weight of sample is less than 80% of division. You
still can count but the counting may be inaccurate. The signal will disappear after about 3 seconds.
b. When the weight window displays “-CSL-”, it means the unit weight of sample is not enough at all. Please
press【Set】to re-sample. Press【Enter】to exit back to the counting model.
Press【Set】to exit from the sample state in the process of sample.
【Percentum mode】
1.Press【Set】in the percentum state to display “LOAD-P”. There are two ways to sample:
a. Put enough samples on the platform then press【Enter】to finish sample.
b. Press【Unit】 then the scale displays “000000”. Press【↑】and【Enter】to input the weight value.
2.When the sample weighed less than 0.1% FS., the weight window displays “-CSL-”. It means the sample is not enough.
You should re-sample. Press【Set】to re-sample. Press【Enter】to exit back to the weighing model.
Press【Set】to exit from the sample state in the process of sample.
【Calibration 】
Keep pressing【Cali.】for about 4 seconds. Calibrate after the weight window displays “CAL.”. There are two ways to
calibrate:
1. Full range load: Put enough weight on the platform according to the full scale. Press【Enter】to make sure. Calibration
finished.
2. Random load:
a. Press【Set】, the weight window displays the weight calibrated last time.
b. Press【↑】and【Enter】to input the weight you want to load. Then load the corresponding weight. Press【Enter】to
make sure. Calibration finished.
Press【Mode】to exit from the calibration state in the process of calibrating.
Note:
The high limit and low limit set in the weighing model are saved in the scale. Different unit corresponding to different
alarm range and method. The value inputted by pressing【↑】cannot be more than the full scale.
↑ ↑ ↑ ↑
9600 4800 2400 1200
2. RS232 interface
(1) Frame format
The scale have a standard RS232 UART signal, Each data frames has 10bits.The detail information of frames is as
follows:
bit1 bit2 bit3 bit 4 bit 5 bit 6 bit 7 bit 8 bit 9 bit 10
bit1:start bit
bit2~bit9:data bit
bit10:stop bit
(2)Data format:Every message include 6 bytes, they are following:
No. 1: D0~D7 —— 0FFH(Message Flag)
No. 2: D0~D2 —— Decimal point(0-5)
D3~D4 —— Current mode: 00 - weighting; 01- counting; 10 - percentum;
D5 —— 1 means weight is negative, otherwise is positive
D6 —— 1 means weight is stable, otherwise is unstable
D7 —— 1 means weight is overflow, otherwise is normal
No. 3: D0~D7 —— BCD1( LSB)
No. 4: D0~D7 —— BCD2(MSB)
No. 5: D0~D7 —— BCD3(HSB)
No. 6: D0~D7 —— Unit for weight: 1 - lb; 0 - Kg;