Uv Visible Spectros
Uv Visible Spectros
Uv Visible Spectros
Spectroscopy
It is the branch of science that deals with the study of
interaction of matter with light.
OR
It is the branch of science that deals with the study of
interaction of electromagnetic radiation with matter.
Electromagnetic
Radiation
Electromagnetic Radiation
Electromagnetic radiation consist of discrete
packets of energy which are called as
photons.
Wavelength (λ):
It is the distance between two nearest parts of
the wave in the same phase i.e. distance
between two nearest crest or troughs.
Electromagnetic Radiation
3 • n → σ* transition
4 • n → π* transition
5 • σ → π* transition
6 • π → σ* transition
1 • σ → σ* transition
OH
Phenol λmax = 270 nm
-
OH
Alkaline
medium -
OH O
p-nitrophenol
λmax = 255 nm λmax = 265 nm
2 • Hypsochromic Shift (Blue Shift)
• When absorption maxima (λmax) of a
compound shifts to shorter wavelength, it is
known as hypsochromic shift or blue shift.
+ -
NH2 + NH 3 Cl
H
Acidic
medium
Aniline
λmax = 280 nm λmax = 265 nm
3 • Hyperchromic Effect
• When absorption intensity (ε) of a
compound is increased, it is known as
hyperchromic shift.
• If auxochrome introduces to the compound,
the intensity of absorption increases.
N N CH3
Pyridine 2methylpyridine
λmax = 257 nm λmax = 260 nm
ε = 2750 ε = 3560
4 • Hypochromic Effect
CH3
Naphthalene 2-methyl
naphthalene
ε = 19000 ε = 10250
Shifts and Effects
Hyperchromic shift
Blue Red
Absorbance ( A )
shift shift
Hypochromic shift
λmax
Wavelength ( λ )
Beer-Lambert’s Law
Statement: for a given material sample path length and concentration of the
sample are directly proportional to the absorbance of the light.
Beer-Lambert Law derivation helps us to define the relationship of the
intensity of visible UV radiation with the exact quantity of substance present.
Used in modern-day labs for testing of medicines, organic chemistry and to
test with quantification.
The Beer-Lambert law is expressed as:
A = εbc
where,
• A is the amount of light absorbed for a particular wavelength by the sample
• ε is the molar extinction coefficient
• b is the distance covered by the light through the solution
• c is the concentration of the absorbing species
Following is an equation to solve for molar extinction coefficient:
ϵ=A/bc
DERIVATION OF BEERS-LAMBERTS LAW
Experimental measurements are usually made in terms of transmittance (T), which is defined as:
T = I / Io
where I is the light intensity after it passes through the sample and Io is the initial light intensity.
The relation between A and T is:
A = -log T = - log (I / Io).
Absorption of light by a sample
NOTE: Modern absorption instruments can usually display the data as either
transmittance, %-transmittance, or absorbance. An unknown concentration of an
analyte can be determined by measuring the amount of light that a sample absorbs
and applying Beer's law. If the absorptivity coefficient is not known, the unknown
concentration can be determined using a working curve of absorbance versus
concentration derived from standards.
The Beer-Lambert law can be derived from an approximation for the absorption coefficient
for a molecule by approximating the molecule by an opaque disk whose cross-sectional
area, σ, represents the effective area seen by a photon of frequency w. If the frequency of
the light is far from resonance, the area is approximately 0, and if w is close to resonance
the area is a maximum. Taking an infinitesimal slab, dz, of sample:
Io is the intensity entering the sample at z=0, Iz is the intensity entering the infinitesimal slab
at z, dI is the intensity absorbed in the slab, and I is the intensity of light leaving the sample.
Then, the total opaque area on the slab due to the absorbers is σ * N * A * dz. Then, the
fraction of photons absorbed will be σ * N * A * dz / A so,
dI/ Iz = - σ * N * dz
Derivation continued…..
Integrating this equation from z=0 to z=b gives,
ln(I)-ln(Io) = - σ * N * b
OR -ln (I-Io) = σ * N * b
Since N(molecules/cm3) * (1 mole / 6.023x1023 molecules) * 1000 cm3 / liter
= c (moles/liter) and 2.303 * log(x) = ln(x) then,
-log (I-Io) = σ *(6.023x1020 / 2.303)* c * b
-log (I-Io) = A= ε * b * c
Where ε = σ *(6.023x1020 / 2.303) = σ* 2.61x1020
OR For each wavelength of light passing through the spectrometer, the intensity of the light passing through
the reference cell is measured. This is usually referred to as Io that’s I for Intensity.
A)Filters –
Two types of filters:
a) Interference Filters
b) Absorption Filters
Cont..
B. Monochromators
Wavelength selector that can continuously scan a
broad range of wavelengths.
Used in most scanning spectrometers including UV,
visible, and IR instruments.
Refractive type
PRISM TYPE
Reflective type
Diffraction type
GRATING TYPE
Transmission Type
SAMPLE COMPARTMENT
Spectroscopy requires all materials in the beam path
other than the analyte should be as transparent to the
radiation as possible.
The geometries of all components in the system should
be such as to maximize the signal and minimize the
scattered light.
The material from which a sample cuvette is
fabricated controls the optical window that can be
used.
Some typical materials are:
Optical Glass - 335 - 2500 nm
Special Optical Glass – 320 - 2500 nm
Quartz (Infrared) – 220 - 3800 nm
Quartz (Far-UV) – 170 - 2700 nm
Detectors
After the light has passed through the sample, we
want to be able to detect and measure the
resulting light.
These types of detectors come in the form of
transducers that are able to take energy from light
and convert it into an electrical signal that can be
recorded, and if necessary, amplified.
Three common types of detectors are used
Barrier layer cells
Photo emissive cell detector
Photomultiplier
SUMMARY
Types of source, sample holder and detector for
various EM region
A B
A)The Spectrum of compound in A(acid) and B(Base)
B) The difference spectrum of B relative to A
Conclusion:
Qualitative & Quantitative Analysis:
It is used for characterizing aromatic compounds and
conjugated olefins.
It can be used to find out molar concentration of the
solute under study.
Detection of impurities:
It is one of the important method to detect impurities
in organic solvents.
Detection of isomers are possible.
Determination of molecular weight using Beer’s law.
Reference Books
Introduction to Spectroscopy
Donald A. Pavia
Elementary Organic Spectroscopy
Y. R. Sharma
Practical Pharmaceutical Chemistry
A.H. Beckett, J.B. Stenlake