UV/Vis Spectrophotometer
Operation Manual
■ Safety
The safety statements in this manual comply with the requirements of the HEALTH AND SAFETY
AT WORK ACT, 1974.
Read following instructions before installing and using the instrument and its accessories. The
apparatus should be operated by appropriate laboratory technicians.
■ General
The apparatus described in this manual is designed to be used by properly trained personnel in a
suitable equipped laboratory. For the correct and safe use of this apparatus it is essential that
laboratory personnel follow generally accepted safe procedures in addition to the safety
precautions called for in this manual.
The covers on this instrument may be removed for servicing. However, the inside of the power
supply unit is a hazardous area and its cover should not be removed under any circumstances.
There are no serviceable components inside this power supply unit. Please avoid touching the
high voltage power supply at all times.
Some of the chemicals used in spectrophotometer are corrosive and/or inflammable and samples
may be radioactive, toxic, or potentially infective. Care should be taken to follow the normal
laboratory procedures for handling chemicals and samples.
■ Electrical
Before switching on the apparatus, make sure it is set to the voltage of the local power supply.
The power cord shall be inserted in a socket provided with a protective earth contact. The
protective action must not be negated by the use of an extension cord without a protective
conductor.
■ Warning
Any interruption of the protective conductor inside or outside the apparatus or disconnection of the
protective earth terminal is likely to make the apparatus dangerous. Intentional interruption is
prohibited.
Whenever it is likely that the protection has been impaired, the apparatus shall be made
inoperative and be secured against any unintended operation.
NEVER touch or handle the power supply on due to the high voltage!
The protection is likely to be impaired if, for example, the apparatus
Shows visible damage
Fails to perform the intended measurements
Has been subjected to prolonged storage under unfavorable conditions
Has been subjected to severe transport stresses
■ Radio Interference
For compliance with the EMC standards referred to in the EC Declaration of Conformity, it is
necessary that only shielded cables supplied by us are used when connecting the instrument to
computers and accessories.
Contents
Chapter 1 Introduction ...........................................................................................................1
1.1 Measurement Principle ...................................................................................................... 1
1.2 Performance and features ................................................................................................. 1
1.3 Application ....................................................................................................................... 2
1.4 Technical Specifications .................................................................................................... 2
1.5 Packing List ......................................................................................................................3
1.6 Symbols and Notices .........................................................................................................3
1.7 Product Design ................................................................................................................. 4
Chapter 2 Installation ............................................................................................................ 7
2.1 Unpacking ........................................................................................................................ 7
2.2 Requirements ................................................................................................................... 7
2.3 Installation ........................................................................................................................7
Chapter 3 Instrument Operation ............................................................................................. 8
3.1 Power On & System Initialization ........................................................................................8
3.2 Photometric Measurement ................................................................................................. 9
3.2.1 Photometric measurement ............................................................................................ 10
3.3 Quantitative Measurement ............................................................................................... 12
3.3.1 Standard curve measurement ........................................................................................14
3.3.2 Measurement based on experienced calculation formula ................................................. 16
3.3.3 Data processing ........................................................................................................... 17
3.4 Wavelength Scan ............................................................................................................ 17
3.4.1 Wavelength scan ..........................................................................................................19
3.4.2 Data processing ........................................................................................................... 20
3.5 Kinetic Analysis ...............................................................................................................20
3.5.1 Kinetic measurement .................................................................................................... 22
3.5.2 Data processing ........................................................................................................... 23
3.6 DNA/Protein Measurement .............................................................................................. 24
3.6.1 DNA/Protein measurement ............................................................................................25
3.6.2 Data processing ........................................................................................................... 26
3.7 Multi-wavelength Measurement ........................................................................................ 27
3.7.1 Multi-wavelength measurement ..................................................................................... 28
3.7.2 Data processing ........................................................................................................... 29
3.8 System Settings .............................................................................................................. 30
3.8.1 Wavelength calibration ..................................................................................................30
3.8.2 Printer setting ...............................................................................................................31
3.8.3 Lamps management ..................................................................................................... 32
3.8.4 Time setting ................................................................................................................. 33
3.8.5 Dark current calibration ................................................................................................. 34
3.8.6 Photometric accuracy verification ...................................................................................34
3.8.7 Wavelength verification ................................................................................................. 37
3.8.8 Connect to PC ..............................................................................................................39
3.8.9 Beeper on/off ............................................................................................................... 39
3.8.10 System baseline calibration ......................................................................................... 40
3.8.11 File system format ...................................................................................................... 40
3.8.12 Restore factory default ................................................................................................ 40
Chapter 4 Maintenance ....................................................................................................... 41
4.1 Maintenance ................................................................................................................... 41
4.2 Fuse Replacement .......................................................................................................... 42
4.3 Lamps Replacement ........................................................................................................43
4.3.1 Deuterium lamp replacement .........................................................................................43
Chapter 5 Troubleshooting .................................................................................................. 47
Chapter 1 Introduction
1
Data saving and reading is much flexible, and the data saved on the instrument
memory can be exported to PC as necessary.
1.3 Application
The UV/Vis spectrophotometer is a common analytical instrument in chemistry laboratory,
and it is widely used in pharmaceutical, medicine & health, chemical, energy, machinery,
metallurgy, environmental protection, geology, food, biology, materials, agriculture, forestry,
fisheries and other industries. It’s also applied in the fields of higher education, metrology,
teaching and scientific research, and has advantages in quality control, raw material and
product inspection in production.
UV-9000 series are double beam UV/Vis spectrophotometers. Thanks to their stable
performance, accurate measurement and powerful functions, they have strong advantages
in various fields of scientific research and quality control.
2
1.5 Packing List
No. Item Unit Qty Note
2 Power Cord pc 1
5 Dust Cover pc 1
7 User’s Manual pc 1
8 Software Manual pc 1
9 Quality Certificate pc 1
10 Packing List pc 1
: HIGH VOLTAGE.
Caution the danger of high voltage, and be careful of the risk of electric shock.
: HOT SURFACE
Caution the hot surface, and avoid the risk of burn.
: ULTRAVIOLET RADIATION
Caution the emission of UV radiation.
: NOTICE.
Pay attention to the notice.
: SPECIAL EXPLANATION.
Pay additional attention to the special explanation.
3
1.7 Product Design
1. Configuration
The profile of UV-9000 series UV/Vis Spectrophotometer is shown in Fig. 1-1:
Fig. 1-1
The back side of UV-9000 series UV/Vis Spectrophotometer is shown in Fig. 1-2:
Fig. 1-2
The compartment configuration of UV-9000 series UV/Vis Spectrophotometer is shown in
Fig. 1-3:
4
Fig. 1-3
The schematic diagram of the instrument’s internal structure is shown in Fig. 1-4:
Fig. 1-4
2. Control Panel and Keys
The control panel of UV-9000 series UV/Vis Spectrophotometer is shown in Fig. 1-5.
Fig. 1-5
The keys include number & character keys, function keys and shortcut keys. Following are
the description of the keys.
1) Number & character keys
, , , , , , , , , , :
User can input the value such as wavelength, concentration, and date. It’s also used for
file name edit.
2) Function keys
: Calibrate the blank, to adjust to 0.000 Abs or 100.0 %T, or to run baseline calibration.
5
: Start the measurement.
: Esc key, to exit the current interface, and return to the previous interface. It’s also for
Stop key, to stop the measurement during the wavelength scanning or kinetic
scanning.
: For cell position control. It’s only available for the instrument with the accessory of
automatic cells holder.
, , , :
Up, down, left and right keys are used for selection. Up and down keys are also used for
ordinate setting, and for wavelength scanning data retrieval with peak or valley. And also
for caps lock when inputting the letter. Left and right keys are also used for abscissa
setting, and for wavelength or kinetic scanning data retrieval point by point.
3) Shortcut keys
, , , :
The shortcut key operation is based on the information on the screen.
6
Chapter 2 Installation
Please carefully read the instruction in this chapter before unpacking and installation UV-
9000 series UV/Vis Spectrophotometer.
2.1 Unpacking
Please check the outer packing and make sure that it is intact before unpacking UV-9000
series UV/Vis Spectrophotometer. Then, check the instrument and its accessories
according to the packing list and make sure they are completely well. If you have any
questions, or anything lost or damaged, please contact us in time.
2.2 Requirements
A laboratory should be prepared, and following requirements should be met:
1) The instrument should be placed in a dry room, and the room temperature should be in
the range of 5 ℃~ 35 ℃. The relative humidity should be no more than 85%.
2) Power supply requirement: The rated voltage should be 220 V ± 22 V AC (110 V ± 11 V
AC is also optional), and the frequency should be 50 Hz (60 Hz is also optional). Well
grounding is also required. An electronic AC regulator or AC regulator with the power
more than 1000 W is suggested to enhance the anti-interference performance of the
instrument.
3) Other requirements: Be far away from strong or continuous vibration. Neither setting up
the instrument near electromagnetic field, nor exposing the instrument to direct sunlight
or the radiation of heaters. It should be free of dust, as well as corrosive vapors. The
instrument should be placed on a stable workbench. And for well cooling and ventilation,
a clearance of at least 15 mm to the wall is suggested.
2.3 Installation
Install the instrument as following steps:
Step 1: Place the instrument onto a stable bench after unpacking.
Step 2: Connect the power cord to the instrument. If a printer is equipped, connect the
power cord of the printer and connect the instrument to the printer with the
communication cable.
7
Chapter 3 Instrument Operation
Before switching on the power, make sure that all connections work well, the power supply
is with well grounding and met the requirement, neither sample in the sample compartment
nor any other block in the light path.
Fig. 3-1
Please don’t open the lid of the sample compartment during the system initialization.
2. Warm up
It will cost 15 minutes for warm up process. User can press to skip the process. If the
beeper is on, the instrument will beep three times when the warm up process is completed.
given. Press or to select “Yes” and press , the system will continue the
calibration process and then enter the main interface (Fig. 3-2).
8
User also can skip the system calibration by pressing or directly to enter the
main interface.
Fig. 3-2
For UV-9000S, its bandwidth is adjustable, and there will additionally show “F1 Slit Width: 2.0nm”
in the main interface. User can press to set the bandwidth, when it shows “Set Slit: 2.0nm”,
press or (or or ) to select certain bandwidth, and press to make sure the
setting.
The system calibration process includes “get dark current”, “search 656.1nm”, “check energy” and
“baseline calibration”. If the data in memory has been lost, the instrument will automatically
perform system calibration. And for measurement accuracy, the system calibration is usually
suggested.
It’s ready for operation. Functions such as photometric measurement (Basic mode),
quantitative measurement, wavelength scan, kinetic analysis, DNA/Protein measurement,
Multi-wavelength measurement, and system settings (Utility) are available with the
instrument operation. Just press the corresponding number key to enter the operation
interface.
When the system shows the main interface, press on the key panel to enter the
photometric measurement interface (Fig. 3-3).
9
Fig. 3-3
There are four function shortcuts at the bottom of the screen: Unit, Mode, F Factor, and
Standard.
Unit: Press to enter the setup, it shows “Please select unit: ”. press or (or
or ) to select the unit, such as mg/L, g/L, ppb, ppm, %, other, IU, mM/L, M/L, g/mL
and mg/mL. If “other” is selected, there further shows “Please input self defined
unit: ”, user can edit the unit.
Mode: Press to select the mode, it shows “please select mode:”. press or
to select the mode, and there are three modes for selection: Abs, T% and
Conc.
F Factor: Press , it shows “Please input Factor: ”. User can input the value between
0.000 and 9999.
Standard: Press , it shows “Please input Conc. of standard: ”. User can input the
value between 0.000 and 9999.
When the system shows the main interface (Fig. 3-2), press to enter the photometric
measurement interface (Fig. 3-3).
10
Step 2: Set the measurement wavelength.
Press , it shows “Please input wl:” (Fig. 3-4), input the value and press to confirm
the setting, the instrument will move wavelength to the designated spot, and adjust blank
automatically.
The valid wavelength range of UV-9000 series is between 190 nm and 1100 nm. If the input value
is out of the range, it is invalid, and user needs input again.
User can press to clear the input when an error is found, then input the target value again.
It also works in the process of digital setting in subsequent operations.
Step 3: Mode selection & Sample measurement.
Press to select the mode (Fig. 3-5), and begin to measure sample, use liquid sample
as an example.
a. Under the mode of T%, or Abs, put the blank solution or reference solution separately
into the reference light path and sample light path, press to calibrate blank. Then
replace the blank solution or reference solution with the sample solution only in the
sample light path, read out the value directly.
b. Under the mode of Conc., press to select the unit. Then put the blank solution or
reference solution separately into the reference light path and sample light path, press
to calibrate blank. For the sample that has known F-factor, replace the blank
solution or reference solution with the sample solution only in the sample light path, and
input F-factor. The system will measure the Absorbance, and user can read the
concentration directly. For the sample that F-factor hasn’t known, replace the blank
solution or reference solution with the standard solution only in the sample light path,
11
and input the standard concentration. The system will measure the absorbance, and
calculate F-factor as well. (F = Conc. / Abs). Next, replace the standard solution with the
sample solution, read the concentration directly (Fig. 3-6).
Fig. 3-6
Photometric measurement is usually for simple measurement, the result can be directly
read out, and the data saving is unavailable.
Different absorbance linearity range will cause different measurement error. The best absorbance
linearity range is between 0.2 and 0.8.
12
Although the standard curve is shown with the absorbance as ordinate and the concentration as
abscissa, the linear fit equation of the instrument is shown as C=K0+K1×A. However, it won’t
affect the sample measurement and the final result display.
UV-9000 series UV/Vis Spectrophotometer has two operation states with the quantitative
measurement function, Quantitative test and Standard calibration curve. User can do
sample measurement based on the standard calibration curve. User also can utilize
sample measurement referring with the experienced calculation formula.
When the system shows the main interface, press on the key panel to enter the
quantitative measurement interface (Fig. 3-7).
There are two function shortcuts at the bottom of the screen: Unit, Standard Curve.
Unit: Press to enter the setup, it shows “Please select unit: ”. press or (or
or ) to select the unit, such as mg/L, g/L, ppb, ppm, %, other, IU, mM/L, M/L, g/mL
and mg/mL. If “other” is selected, there further shows “Please input self defined
unit: ”, user can edit the unit.
Standard curve: Press to enter the standard curve operation interface (Fig. 3-8),
continue the next setup.
the value and press , then shows “Input r =”. When the method
Cubic fit is selected, step by step input the factors: K0, K1, K2, K3.
When the method Square fit is selected, step by step input the factors:
K0, K1, K2. When the method Linear fit is selected, there shows “Input
K0 =”, input the value and press , then shows “Input K1 =”, input
the input, then input the next value accordingly. Press to end the
setting. Then begin the testing.
Curve: After completing the testing press to view the standard curve.
When the system shows the main interface (Fig. 3-2), Press to enter the quantitative
measurement interface (Fig. 3-7).
Step 2: Set the measurement wavelength.
Press , it shows “Correction method:”. press or , and there are three methods
for selection: Single WL, 3 Points, Isoabsorbance. For example, Single WL is selected,
press and it shows “Please input wl 1:”, input the value, then press to confirm the
setting.
Step 3: Standard blank calibration.
Put the blank solution of standard separately into the reference light path and sample light
Press to enter the unit setup, press or to select the unit and press to
confirm the setting. Then press to enter the standard curve operation interface (Fig.
3-8). Press to enter the method setup, press or to select “Linear fit” (for
example) and press . Then press to enter the standard setting interface (3-9), it
shows “Edit the number… ”, the cursor points to the setting position, press , it shows
“Input standard Conc: ”, input the concentration value and press to make sure the
input, then input the next value accordingly. Press after completing all the standards
setting.
When press to end the standard setting, it shows “Place Standard #1”, replace the
blank solution with the first standard solution only in the sample light path, and press
or , accordingly to measure other standards and record the data (Fig. 3-10).
If some data need to be reacquired, user can press , and press or , let the cursor
After completing the standard curve measurement, press to return to the quantitative
measurement interface. Put the blank solution of sample separately into the reference light
path and sample light path, press to calibrate blank. Then, replace the blank solution
with sample solution only in the sample light path, press , the measurement data will
be given (Fig. 3-11). More than one hundred testing data can be shown on the table list.
Following are brief operation steps for quantitative measurement based on experienced
calculation formula.
Step1: Enter the quantitative measurement interface (Fig. 3-7).
Step 2: Set the measurement wavelength.
Step 3: Put the blank solution separately into the reference light path and sample light path,
Press to set the unit and press to enter the standard curve operation interface.
Press to set the fitting method (“Linear fit through zero” for example) and press to
enter the parameters setting interface (Fig. 3-12), set each parameter according to the
prompts.
16
Step 5: Sample measurement.
Press after completing all parameters setting, return to the quantitative measurement
interface. Replace the blank solution with sample solution only in the sample light path,
User can do data processing such as data saving, loading and retrieval after completing
quantitative measurement.
In the standard curve operation interface, press to save standard measurement data
and the standard curve, it shows “Please input File Name:”, input the file name and
press , the file will be saved, use “.fit” as suffix. User can press to view the curve.
the file will be saved, use “.qua” as suffix. User can press or (or or ) to
browse the sample data one by one. While pressing or , it shows “Search
sample:”, input the testing serial number, and press to view the designated sample
data.
In the standard curve operation interface or the quantitative measurement interface, press
, it will show “Open file:”, press or (or or ) to select certain file with the
right suffix, and the saved data can be load.
User should enter the right interface to load the file with the right suffix. Otherwise, the file can’t be
opened, and the system will give the prompts “File Type Error”.
When the system shows the main interface, press on the key panel to enter the
wavelength scan interface (Fig. 3-13).
17
Fig. 3-13
There are four function shortcuts at the bottom of the screen: Setup, Mode, Search, and
Smooth.
Setup: Press to enter the setup, it shows “Scan from:” (the maximum of the
wavelength), make sure the input, and shows “Scan to:” (the minimum of the
Mode: Press to select the mode, shows “Please select mode:”, press or
to select the mode, and there are two modes for selection, Abs, and T%.
Search: After completing the wavelength scan, press to enter the data retrieval
Smooth: User can press to perform spectrum smoothing after completing the
wavelength scanning.
18
3.4.1 Wavelength scan
When the system shows the main interface (Fig. 3-2), press to enter the wavelength
scan interface (Fig. 3-13).
Step 2: Select the scanning mode.
Press to enter the setting, input the wavelength scanning range respectively with Scan
from and Scan to, press to confirm the setting. Successively press or to
select the scanning step and scanning speed, press to confirm the setting and return
to the wavelength scan interface. Then, press or to set the ordinate, it shows
“Min Y: ”, next shows “Max Y: ”. Input the displaying range and press to confirm the
setting.
Scan step can be selected between 0.1nm, 0.2nm, 0.5nm, 1.0nm, 2.0nm and 5.0nm at most. Due
to the memory limitation of the system (each scanning curve with 1001 data at most), it provides a
reasonable scan step for selection according to the actual wavelength scanning range. And there
are three scanning speeds for selection, HI, MEDIUM, and LOW.
Step 4: Sample scanning.
Put the reference solution separately into the reference light path and sample light path,
press to calibrate baseline. Then replace the reference solution with the sample
solution only in the sample light path, press to begin sample scanning, and the
spectrum will be shown at the same time.
The scanning sequence is from the maximum wavelength to the minimum wavelength. If the
beeper is on, the instrument will beep three times after completing the baseline calibration and
sample scan, and it will return to the maximum wavelength at the same time.
19
3.4.2 Data processing
User can do data processing such as data saving, loading, retrieval and spectrum
smoothing after completing wavelength scan.
1. Data saving
In the wavelength scan interface, press to save the scanning data, it shows “Please
input File Name:”, input the file name and press , the file will be saved, use “.wav” as
suffix.
2. Data loading
In the wavelength scan interface, press , it will show “Open file:”, press or
(or or ) to select certain file with the right suffix, and the saved data can be load.
3. Data retrieval
In the wavelength scan interface, user can press to enter the data retrieval interface.
Press , it shows “Please input peak height:”, input the value and press , then press
or to search the peak. User also can press or to retrieve the scanning
The peak height retrieval is only available with absorbance mode and the spectrum should
contain obvious characteristic peaks.
4. Spectrum smoothing
In the wavelength scan interface, user can press to perform spectrum smoothing, and
it will directly show a smoothed spectrum. If pressing several times, it will perform
spectrum smoothing several times.
20
When the system shows the main interface, press on the key panel to enter the
kinetic measurement interface (Fig. 3-14).
Fig. 3-14
There are four function shortcuts at the bottom of the screen: Setup, Mode, Process, and
Search.
Setup: Press to enter the setup, it shows “Total Time:”, make sure the input, and
shows “Delay Time:”, make sure the input, then shows “Time interval:”, press
Mode: Press to select the mode, shows “Please select mode:”, press or
to select the mode, and there are two modes for selection, Abs, and T%.
Process: User can press to perform data processing after completing the kinetic
measurement, it shows “Begin Time:”, make sure the input, and shows “End
Time:”, make sure the input, then shows “Factor:”, make sure the input and the
Search: After completing the kinetic measurement, press to perform data retrieval.
User can press or to retrieve the kinetic measurement data point by point.
21
3.5.1 Kinetic measurement
When the system shows the main interface (Fig. 3-2), press to enter the kinetic
measurement interface (Fig. 3-14).
Step 2: Select the kinetic measurement mode.
Press , it shows “Please input wl:”, input the value and press to confirm the setting,
the instrument will move wavelength to the designated spot.
Step 4: Set the kinetic measurement parameters.
Press to enter the setting, input Total Time and Delay Time, then press or
to select the time interval, press to confirm the setting and return to the kinetic
analysis interface. Then, press or to set the ordinate, it shows “Min Y: ”, next
shows “Max Y: ”. Input the displaying range and press to confirm the setting.
The maximum of Total Time is 30000s. The Time interval can be selected between 0.5s, 1.0s,
2.0s, 5.0s, 10s, 30s and 1min. Due to the memory limitation of the system (each kinetic
measurement with 4000 data at most), it provides a reasonable time interval for selection
according to the total time.
Step 5: Kinetic measurement.
Put the reference solution separately into the reference light path and sample light path,
press to calibrate blank. Then replace the reference solution with the sample solution
only in the sample light path, press to begin kinetic measurement, and the kinetic
curve will be shown at the same time.
If the beeper is on, the instrument will beep three times after completing the kinetic measurement,
and it will return to the maximum wavelength at the same time.
22
3.5.2 Data processing
User can do data processing such as data saving, loading, kinetic analysis and data
retrieval after completing kinetic measurement.
1. Data saving
In the kinetic measurement interface, press to save the kinetic measurement data, it
shows “Please input File Name:”, input the file name and press , the file will be saved,
use “.kin” as suffix.
2. Data loading
In the kinetic measurement interface, press , it will show “Open file:”, press or
(or or ) to select certain file with the right suffix, and the saved data can be load.
3. Kinetic analysis
In the kinetic measurement interface, user can press to perform data processing
(kinetic analysis) after completing the kinetic measurement. Successively input Begin Time,
End Time, and Factor. The system will give the kinetic analysis result based on the setting
time range. The value I.U. will be calculated with the calculation formula I.U.=Factor×Δ
A/min, and be displayed on the right of the screen (Fig. 3-15). The average straight line
between the Begin Time and End Time will be calculated. The gradient of this line gives
the rate of change of ΔA/min.
Fig. 3-15
23
4. Data retrieval
In the kinetic measurement interface, user can press to perform data retrieval. press
or to retrieve the kinetic measurement data point by point. Press to exit the
data retrieval.
When the system shows the main interface, press on the key panel to enter the
DNA/Protein measurement interface (Fig. 3-16).
Fig. 3-16
There are four function shortcuts at the bottom of the screen: Coeff, Method, Unit, and
Default.
Coeff: Press to enter the coefficients setup, it shows “Input f1=62.90”. User can use
the default value, or input the setting value, and gradually input the factors f2, f3
and f4 according to the prompts.
24
select Yes or No. If selecting Yes, the measurement will be taken together with the
two testing wavelengths and the reference wavelength 320nm.
Unit: Press to enter the setup, it shows “Please select unit: ”. Press or (or
or ) to select the unit, such as mg/L, g/L, ppb, ppm, %, other, IU, mM/L, M/L,
g/mL and mg/mL. If “other” is selected, there further shows “Please input self
defined unit:”, user can edit the unit.
Default: Press to restore the default coefficients values. For the measurement method
"Absorbance difference 1", the default coefficients are: f1 =62.90, f2 =36.00, f3
=1552, f4 =757.3, and for the measurement method "Absorbance difference 2",
the default coefficients are: f1 =49.10, f2 =3.480, f3 =183.0, and f4 =75.80.
When the system shows the main interface (Fig. 3-2), press to enter the
DNA/Protein measurement interface (Fig. 3-16).
Step 2: Test method selection
Press to select the test method, it shows “Measurement: Absorbance difference 1”,
press or to select the test method, and there are two methods for selection,
Absorbance difference 1 (with the testing wavelengths 260nm and 280nm), and
Absorbance difference 2 (with the testing wavelengths 260nm and 230nm). Then shows
“With reference: Yes”, press or to select Yes or No. If selecting Yes, the
measurement will be taken together with the two testing wavelengths and the reference
wavelength 320nm.
25
Step 3: Coefficients & Unit setting.
Press to enter the coefficients setup, it shows “Input f1=62.90”. User can use the
default value, or input own experienced value, and gradually input the factors f2, f3 and f4
according to the prompts.
Then, press to enter the unit setup, it shows “Please select unit: ”. Press or
to select the unit, or input the self-defined unit, press to confirm the setting.
Step 4: DNA/Protein measurement.
Put the reference solution separately into the reference light path and sample light path,
press to calibrate blank. Then replace the reference solution with the sample solution
only in the sample light path, press to begin DNA/Protein measurement. The
absorbance values and DNA/Protein concentration results will be recorded (Fig. 3-17).
Fig. 3-17
User can do data processing such as data saving, loading, and retrieval after completing
DNA/Protein measurement.
1. Data saving
measurement data, it shows “Please input File Name:”, input the file name and press ,
the file will be saved, use “.dna” as suffix.
26
2. Data loading
In the DNA/Protein measurement interface, press , it will show “Open file:”, press
or (or or ) to select certain file with the right suffix, and the saved data can be
load.
3. Data retrieval
In the DNA/Protein measurement interface, the current display only shows one sample
record. For more sample records, user can press or to browse the sample data
one by one. While pressing or , there shows “Search sample:”, input the testing serial
When the system shows the main interface, press on the key panel to enter the
multi-wavelength measurement interface (Fig. 3-18).
Fig. 3-18
There are two function shortcuts at the bottom of the screen: WL Setup, Mode.
WL Setup: Press to enter the setup, it shows “Edit the number… ”, press , it
shows “Please input wl: ”, make sure the input, then input the next value
27
Mode: Press to select the mode, it shows “please select mode:”. Press or
to select the mode, and there are two modes for selection: Abs, T%.
When the system shows the main interface (Fig. 3-2), press to enter the multi-
wavelength measurement interface (Fig. 3-18).
Step 2: Multi-Wavelength parameters setting
Press to enter the setup, it shows “Edit the number… ”, press , it shows “Please
input wl: ”, make sure the input, then input the next value accordingly. Press to end
the setting.
Up to 10 wavelengths point can be set, and the wavelength parameters will be remembered. If
user want to reduce wavelength points for next measurement, after entering the wavelength
setting, press first to delete the last wavelength parameters one by one when the system
prompting " Edit the number…", then reset the wavelength parameters.
Step 3: Mode selection.
press to calibrate blank. Then replace the reference solution with the sample solution
only in the sample light path, press to begin multi-wavelength measurement. The
measurement data will be recorded (Fig. 3-19).
28
Fig. 3-19
For multi-wavelength measurement, the instrument performs wavelength movement following the
setting sequence. If the beeper is on, the instrument will beep three times after completing blank
calibration or sample measurement. And it will return to the first wavelength position.
User can do data processing such as data saving, loading, and retrieval after completing
multi-wavelength measurement.
1. Data saving
measurement data, it shows “Please input File Name:”, input the file name and press ,
the file will be saved, use “.mul” as suffix.
2. Data loading
In the multi-wavelength measurement interface, press , it will show “Open file:”, press
or (or or ) to select certain file with the right suffix, and the saved data can
be load.
3. Data retrieval
In the multi-wavelength measurement interface, the current display only shows one
sample record. For more sample records, user can press or to browse the
sample data one by one. While pressing or , it shows “Search sample:”, input the
testing serial number, and press to view the designated sample data.
29
3.8 System Settings
In the system settings interface, user can carry out wavelength calibration, dark current
calibration, photometric accuracy verification, wavelength verification and system baseline
calibration. Other operations such as printer setting, lamps management, time setting,
beeper setting, file system format, restore factory default, and PC connection are also
available.
When the system shows the main interface, press on the key panel to enter system
settings interface (Fig. 3-20).
Fig. 3-20
There are twelve operation items shown in the system settings interface: WL Reset,
Printer , Lamp, Clock, Dark current, Accu Validity, WL Validity, Connect to PC, Beeper
on/off, System Baseline, Delete entire saved files, and Restore default. User can press
the corresponding number key or shortcut key to enter the relevant operation interface or
directly perform certain operation. In the system settings interface, user can press to
return to the main interface.
The wavelength calibration is necessary when user doubts that there is a deviation of the
30
3.8.2 Printer setting
In the system settings interface, press to enter the printer setting interface (Fig. 3-
21).
Fig. 3-21
There are four setting items: Reset printer, Select print port, Select printer, and Print report.
1. Reset printer
Press , it shows “Select the print port:”. There are two print port for selection, LPT
(parallel port, for printer only), and Comm (serial port, for multifunction interface). Press
Press , it shows “Printer:”. There are eight printer types for selection, SPRT SP _T,
SPRT SP _POS76II, HP PCL (black mode), HP PCL (1 color cartridge), Epson ESC/P,
Epson ESC/P2 or above, Epson Stylus C4x, and Epson Stylus C6x. Press or
Press to switch the print mode between “print report” and “print screen”. A small icon
" " will appear on the upper right of the screen when print screen is selected. And it will
disappear when print report is selected.
31
3.8.3 Lamps management
User can switch on or off the lamp and set the lamp conversion wavelength by lamps
Fig. 3-22
There are five setting items: Switch D2: on/off, Reset D2 lamp usage time, Switch W:
on/off, Reset W lamp usage time and Switch point.
1. Switch D2: on/off
usage time to zero as necessary. Press , it shows such as “5 hrs used. Are you sure?
No”. Press or to select Yes or No, and press to confirm the setting. User
usage time to zero as necessary. Press there shows such as “5 hrs used. Are you
sure? No”. Press or to select Yes or No, and press to confirm the setting.
User also can press to exit the interface without any operation.
32
5. Switch point
It’s used for lamp conversion wavelength setting. The valid setting range is between
300nm and 360nm, and the default lamp conversion wavelength is 339nm. Press ,
there shows “Input the D2/W switch point:”, input the conversion wavelength value as
For measurement accuracy, please don’t measure just under the coeversion wavelength. Please
set the conversion wavelength properly before measurement.
User can set the system displaying time. In the system settings interface, press to
enter the time setting interface (Fig. 3-23).
Fig. 3-23
There are four setting items: Set time, Set date, Display time, and Display date.
1. Set time
Press , it shows “Please input the time:”. The input format “**.**.**” means hour,
minute and second respectively. Input the time values and press to confirm the
setting. The time will be displayed on the upper right corner of the screen with the display
format “** : ** : **”.
2. Set date
Press , it shows “Please input the date:”. The input format “**.**.**” means year,
month and day respectively. Input the date values and press to confirm the setting.
33
3. Display time
User can press to display time, and it will be displayed on the top right corner of the
screen.
4. Display date
User can press to display date, it will be displayed on the top right corner of the
screen with the format “** - **- **”.
Display time or display date, user can only choose one. If display time is performed, the date
won’t display.
The dark current may changes when the instrument runs for a long time, or the wavelength
is set again, or any other influences. For measurement accuracy, the dark current
calibration is necessary before measurement.
Photometric accuracy verification is performed with the certified sample (with known
absorbance or transmittance) to test the photometric validity of the instrument.
In the system settings interface, press to enter the photometric accuracy verification
interface (Fig. 3-24).
Fig. 3-24
34
There are three function shortcuts at the bottom of the screen: Standard, Mode, and
Tolerance.
Standard: Press to enter the standard setting, it shows “Edit the number… ”,
Mode: Press to select the test method, it shows “Please select mode:”, press or
to select the mode, and there are two modes for selection, Abs, and T%.
Tolerance: Press to enter the setup, it shows “Input tolerance:”, input the tolerance
value and confirm the setting.
In the system settings interface (Fig. 3-20), press to enter the photometric accuracy
verification interface (Fig. 3-24).
Step 2: Set the test wavelength.
Press , it shows “Edit the number… ”, press , it shows “Please input wl:”, input the
value and press to confirm the setting. Several wavelength points can be set,
accordingly input the next value. Otherwise, press to end the setting.
Step 3: Select the test mode.
Press to enter the mode setup. Press or to select Abs or T% as the test
Press to enter the setting, it shows “Edit the number… ”, press , it shows “Input
standard:”, input the certified value of the standard and press to confirm the setting. If
several wavelength points are set, it shows “Edit the number… ”, press , it shows
“Input standard:”, accordingly input the next value. After completing all the settings, it
35
shows “Edit the number… ”, press to return to the photometric accuracy verification
interface.
Press to enter the setup, it shows “Input tolerance:”, input the tolerance value and
confirm the setting.
press to calibrate blank. Then replace the reference solution with the certified sample
solution only in the sample light path, press to begin the test. The system will
determine the test result (pass or fail) according to the tolerance. At last, all the test date
will be recorded (Fig. 3-25).
Fig. 3-25
User can do data saving and loading after completing the photometric accuracy verification.
In the photometric accuracy verification interface, press to save test data, and the file
will be saved using “.phv” as suffix.
In the photometric accuracy verification interface, press to perform data loading, just
select certain file with the right suffix to be loaded.
36
3.8.7 Wavelength verification
In the system settings interface, press to enter the wavelength verification interface
(Fig. 3-26).
Fig. 3-26
There are three function shortcuts at the bottom of the screen: Set peaks, Mode, and
Tolerance.
Set peaks: Press to enter the setup, it shows “Edit the number… ”, press , it
shows “Please input wl:”, make sure the input, then input the next value
Mode: Press to select the test method, it shows “Please select mode:”, press or
to select the mode, and there are two modes for selection, Abs, and T%.
Tolerance: Press to enter the setup, it shows “Input tolerance:”, input the tolerance
value and confirm the setting.
In the system settings interface (Fig. 3-20), press to enter the wavelength verification
interface (Fig. 3-26).
37
Step 2: Select the test mode.
Press to enter the mode setup. Press or to select Abs or T% as the test
Press to enter the setting, it shows “Edit the number… ”, press , it shows “Please
input wl:”, input the value and press to confirm the setting. The cursor moves to the
second row automatically and it shows “Edit the number… ”, press , it shows “Please
input wl:”, input the second wavelength value and press to conform the setting.
Several wavelengths can be set. Accordingly input the next value, otherwise, press to
end the setting, and return to the wavelength verification interface.
Press to enter the setup, it shows “Input tolerance:”, input the tolerance value and
confirm the setting.
Press to calibrate blank without anything in the light path. Then put the wavelength
reference material into the sample light path, press to begin the test. The system will
determine the test result (pass or fail) according to the tolerance. At last, all the test date
will be recorded (Fig. 3-27).
Fig. 3-27
38
User can do data saving and loading after completing the wavelength verification. In the
wavelength verification interface, press to save test data, and the file will be saved
using “.wlv” as suffix.
In the wavelength verification interface, press to perform data loading, just select
certain file with the right suffix to be loaded.
3.8.8 Connect to PC
For PC connection, only when PC sends “online” commands in the software interface can it be
effectively executed.
“Controlled by UV Analyst…”will be shown on the screen after successfully connecting to
In the system settings interface, press to switch the beeper status between on and
off.
If the beeper is on, a small icon “ ” will appear on the upper right of the screen. Otherwise,
the small icon will disappear. However, the beeper is on as default setting with each
booting.
39
3.8.10 System baseline calibration
System baseline calibration is performed in the whole wavelength range of the instrument.
In the system settings interface, press to perform the system baseline calibration. If
the beeper is on, the instrument will beep three times after completing the system baseline
calibration. And the wavelength will move to the default position 546.0nm.
In the system settings interface, press , it shows “Delete entire files, are you sure? No”.
Press or to select Yes or No, and press to confirm the setting. User also
Once the file system format is performed, all the files will be cleared. To avoid data lost, please
perform the operation carefully.
In the system settings interface, user can press to perform restore factory default.
If the operation is performed, the parameter settings will be restored to factory default.
However, the saved data won’t be cleared.
40
Chapter 4 Maintenance
4.1 Maintenance
UV-9000 series UV/Vis Spectrophotometer is a precise optical instrument. It was
assembled and debugged carefully before delivery. However, appropriate maintenance will
not only guarantee its reliability and stability, but also prolong its service life. Correct use is
the best maintenance. In addition to previously mentioned installation requirements,
following tips also should be noticed in daily use.
(1) Before switching on the power, make sure that neither sample in the sample
compartment nor any other block in the light path, and the cell holder position is all right,
to avoid error during the self-checking.
(2) Please carefully load the solution into the cuvette, and the height is better no more than
2/3 of the cuvette. Try to avoid the bubble generation, for the bubble on the inner
surface of the cuvette or in the solution will affect the measurement result. Please wipe
off the solution that residue on the outer surface of the cuvette in time. To measure
volatile samples, using with cuvette cover is suggested. Try to avoid contamination to
the cell holder, otherwise, wipe off the residue solution on the cell holder promptly.
(3) Don’t touch both the two optical surfaces of the cuvette with your fingers, for the
fingerprint will absorb the light and furtherly affect the measurement accuracy. Please
handle the cuvette gently, for it is frangible. Clean the cuvette properly. Improper
cleaning or without enough clean also will affect the measurement accuracy, even
cause unstable result.
(4) Whether placing or removing the sample, please close the lid of the sample
compartment in time during the measurement. Please remove the sample from the
sample compartment promptly after completing the measurement, check that there is
no residue in the sample compartment and keep it dry. Any solution sample or residue
left in the sample compartment may cause damage to parts of the instrument such as
filter turning moldy, some components be corroded. Please open and close the lid
gently.
41
(5) To prolong the service life of the lamp, switching off the idle lamp during the
measurement is suggested. Please switch off the instrument and disconnect the plug in
time, to prevent possible damage from thunderstorms.
(6) Be careful in the transport. Don’t place heavy object onto the instrument, to prevent
the light path shift which will furtherly affect the instrument stability and measurement
accuracy.
(7) Don’t disassemble the cover and the inner parts of the instrument without authorization,
especially for the optical parts. Don’t loosen the tightening screws and nuts at will. All
optical surfaces including the light sources can’t be touched by hand or any other
objects. Otherwise, it may affect the normal operation even cause damage.
(8) Keep the instrument surface and the working environment clean. For the surface of the
cover deals with painting process, please don’t clean the cover with organic solutions
such as alcohol, gasoline and ether. If the instrument is not in use, user can cover the
instrument with clean cloth or dust cover to avoid dust accumulation.
(9) A long time not in use should be avoided, and regular boot is suggested to guarantee
the normal operation. In the high temperature and humidity area, user should pay more
attention to keep away from moisture.
The system error may accumulate after transport, moving, and using for a period of
time. When the measurement data differs greatly from the experienced value, or any
above situation occurs, the dark current calibration and wavelength calibration are
suggested to be done.
Danger! Be sure to switch off the power and unplug the socket before replacement!
Following are the steps of replacing the fuse.
Step 1: Power off and unplug the power cord from the instrument.
Step 2: Take out the fuse holder by a 3*75 flat screwdriver with blade, remove the broken
fuse from the working position and replace it with the spare fuse (Fig. 4-1, Fig. 4-2).
42
Fig. 4-1 Fig. 4-2
Step 3: Fit the fuse holder back to the position.
Danger! Be sure to switch off the power and unplug the socket before replacement!
Caution to high temperature! Wait 20 minutes before open the lamp chamber after
power off to avoid scald!
43
Step 5: Disconnect the three leads of the deuterium lamp (Fig. 4-3) from the circuit board,
unscrew the two fixing screws (as shown in Fig. 4-4) and remove the deuterium
from the base of lamp chamber (Fig. 4-5). Then fix the new lamp onto the right
position and connect its three leads to the circuit board.
The deuterium lamp is pre-aligned, there’s no need to re-adjust the position. However, the facula
should focus on the center of the slit (Fig. 4-6).
Do not handle the lamp with bare fingers. Use clean tissue or cloth when handling lamp.
Step 6: Switch on the power, it’s just ok when the deuterium lamp lighting up well.
Step 7: Power off. Reinstall the lamp chamber cover and tighten the screws. Then reinstall
the instrument cover. Be sure to prevent any wires from being pinched in the
process.
Step 8: Reinstall the four screws on the sides of the spectrophotometer.
Fig. 4-5
44
Fig. 4-6
45
Step 6: Switch on the power, observe the entrance facula, it should be focused on the
center of the slit. If the facula is deviated to left or right, then loosen the No.2
screws (as shown in Fig.4-9) and move the lamp seat to left or right until it focus
on the center of the slit. Then fix the screws. If the facula is deviated to up or down,
then loosen the No.1 screws (as shown in Fig.4-9) and move the lamp seat up or
down until the facula focus on the center of the slit. Then fix the No. 1 screws
again.
Step 7: Power off. Reinstall the lamp chamber cover and tighten the screws. Then reinstall
the instrument cover. Be sure to prevent any wires from being pinched in the
process.
Step 8: Reinstall the four screws on the sides of the spectrophotometer.
Fig. 4-9
46
Chapter 5 Troubleshooting
Each UV-9000 series UV/Vis Spectrophotometer is strictly debugged and inspected before
delivery. Commonly, it won’t appear problems in normal storage, transport and use.
However, wrong operation or extreme states, and problems caused by long-term use still
can’t be avoided, such as the damage of electrical and optical units caused by bad storage
and working environment, the damage of vulnerable units or the loosen of the fixing parts
caused by improper transport, the lamp exceeds its lifetime, the wastage of electrical units,
other troubles caused by wrong operation, and so on.
Please carefully refer to the related instructions before operating the instrument. Troubles
and troubleshooting are introduced in following table.
47
No. Trouble Cause Troubleshooting
1) USB cable is not - Make sure the USB cable is
connected. connected well.
2) Communication port is - Make sure no other software is
occupied. occupying the communication port
3) Several workstations run
- Log off or reboot the computer and
at the same time and
ensure that only the instrument
cause the computer
4 Connecting to PC Failure workstation is running
crashed.
- Make sure the USB key has already
4) USB key is not inserted
inserted
5) Software mismatched. - Install and run the right software
6) Communication port of
the computer is - Try again on another computer.
damaged.
1) Wrong position of the cell
- Make sure that the cell holder is in
holder causes block to
right position.
the light path.
2) The pre-warming time is - Pre-warming with enough time, no
not enough. less than 15 min.
3) The tungsten lamp is
- Replace the tungsten lamp with a
exhausted or with bad
The reading is not stable new one.
connection.
5 when adjusting 100% T or
4) Deuterium lamp is - Replace the deuterium lamp with a
0.000 Abs.
exhausted. new one.
- Do dark current calibration and
5) Wavelength error. wavelength calibration, then, try
again.
6) Light path, or the
- Contact the distributor or the factory
amplifier and its power
technical engineer for maintenance.
supply fault.
1) The pre-warming time is - Pre-warming with enough time, no
not enough. less than 15 min.
- Contact the distributor or the factory
2) Unstable voltage.
technical engineer for maintenance.
3) Ambient interference,
- Configure with a steady power
such as unstable power
supply, keep the instrument from
supply, corrosive gas
corrosive gas.
interference.
The sample reading is not
6 - For the sample is unstable, measure
stable.
it as soon as possible. If there is
some bubble in the solution,
4) Unstable sample.
eliminate the bubble or reload the
solution. Measure with a cuvette
cover for volatile sample.
5) The cuvette is
- Make sure that the cuvette is clean
contaminated and it’s too
before measurement.
dirty.
48
No. Trouble Cause Troubleshooting
6) The blank value is much
- The absorbance value of the blank
higher, or the sample
solution or reference solution is better
concentration is too high
below 0.1. Dilute the sample solution
and the absorbance
properly, and the absorbance value is
The sample reading is not reading is out of the
6 better between 0.2 and 0.8.
stable. stable range.
7) The tungsten lamp or
deuterium lamp is
- Change the light source.
exhausted, and the
energy is too weak.
- Calibrate the dark current, then
1) Dark current drift. measure the sample again after
The sample reading is not
7 blank recalibrating.
accurate.
- Make sure that the cuvettes
2) Cuvette matching error
matching well.
1) Loosen connection
- Make sure the connection between
between the instrument
The printer doesn’t work, the instrument and the printer is well.
8 and the printer.
or printing error.
2) The printer model - Choose the factory specified printer
doesn’t match. type.
49
Appendix A
DNA MEASUREMENT
(optional)
50
Appendix B
A.1 Isoabsorbance
When a known interfering component with a known spectrum is present, the error introduced by
this component at the analytical wavelength for the target analyte can be eliminated by selecting a
reference wavelength at which the interfering compound exhibits the same absorbance as it does
at the analytical wavelength. The absorbance at this reference wavelength is subtracted from the
absorbance at the analytical wavelength, as shown in Figure A1.The residual absorbance is the
true absorbance of the analyte.
This technique is less reliable when the spectra of the analyte and of the interferent are highly
similar. Moreover, it can correct for only one interference
The three-point, or Morton-Stubbs correction uses two reference wavelengths, usually those
on either side of the analytical wavelength.
The background interfering absorbance at the analytical wavelength is then estimated using linear
interpolation (see Figure A2).This method represents an improvement over the single-wavelength
reference technique because it corrects for any background absorbance that exhibits a linear
relationship to the wavelength. In many cases, if the wavelength range is narrow, it will be a
reasonable correction for non-linear background absorbances such as that resulting from
scattering of from a complex matrix.
1
Fig A2