PERFORMANCE SPECIFICATIONS
Technology QCM
Range 0.1 to 100 ppmv. Readout capability in ppmw, lb/mmscf, mg/Nm3, and dew point temperature in °C or °F (requires process pressure
as an input)
Online verifications Internal moisture source with NIST-traceable calibration enables on-demand verification of analyzer accuracy and responsiveness
without uninstalling the analyzer. Verification function can be triggered remotely with a voltage signal
Accuracy ±0.03 ppmv or ±10% of reading from 0.1 to 100 ppmv, whichever is greater
Reliability No routine factory calibration required due to highly stable and reliable nature of QCM sensor
QCM response time Near real time. Computer-enhanced response, which may lead to errors, is not required to obtain quick wet-up or dry-down response
Allowable inlet pressure 1.38 to 3.45 Bar (20 to 50 psi) gauge (up to 200 Bar (3000 psi)) with optional pressure reducer; analyzer performance is independent
of process pressure
Sample gas temperature 0 to 100°C (32 to 212°F); analyzer performance is immune to changes in sample gas temperature
Gas flow temperature Approx. 150 sccm. Approx. 1.0 L/min bypass flow for increased speed of response
Outputs Isolated 4-20 mA analog signal, keyboard selectable; 12-bit (0.025%) resolution, RS-232 and RS-485 serial communication ports
(supports Modbus RTU)
Utility requirements 120/240 VAC, 50/60 Hz, 150 Watts Instrument Air: 70 to 100 psi (5 to 7 Bar)
Enclosure IP65
© 2018, by AMETEK, Inc. All rights reserved. Printed in the U.S.A. F-0126 Rev 10 (0918)