DIAGNOSTIC ULTRASOUND SYSTEM

Product Data
No. MPDUS0079EAB

INTRODUCTION
This system achieves high sensitivity and high quality by
using state-of-the-art digital technology in the T/R section,
which is the core of a diagnostic ultrasound system.
This system features advanced algorithms that run on cir-
cuits that incorporate the latest circuit technology, semicon-
ductor technology, and surface-mount technology.
This system is designed to support a full range of applica-
tions and can be used as a general purpose system or a spe-
cialized system, depending on the installed software.

Full-digital ultrasound beam transmission and


reception
This system employs full-digital transmission and reception
circuits. The high definition ultrasound beams and data pro-
cessing technology available with full-digital systems allow
high sensitivity and image quality to be achieved simultane-
ously.

Enhanced diagnostic capabilities


The spatial resolution, contrast resolution, and temporal res-
olution have been improved through new technologies,
resulting in enhanced diagnostic capabilities.

Transducers supporting a wide range of frequen- Ergonomics


cies The system employs a non-interlace high-definition monitor
Echoes over a wide range of frequencies can be obtained with excellent view ability. This feature decreases operator
using a single transducer, allowing the optimal sensitivity fatigue in long examinations.
and quality to be achieved for each region examined. This The ergonomic design of the system ensures comfortable
function permits a single transducer to be utilized for a wide and efficient examinations for operators, physicians, and
range of applications, greatly improving the throughput and patients.
price-to-performance ratio.
Operability
Intelligent panel and software System operability is optimized for the overall clinical work-
The intelligent panel and software facilitate operation and flow in hospitals.
contribute to a higher throughput.
SYSTEM MATRIX OF CUS-AA450
Unit Model name Remarks
21.5-inch wide LCD monitor, DVD/CD drive, Precision Imaging, D-THI, ApliPure+, Tissue Specific
CUS-AA450
Main unit Optimization, Trapezoid Scan, Quick Scan, Advanced Dynamic FlowTM (ADF), DICOM, Smart 3D, Software
AplioTM a450
full keyboard, Vascularity Index, BEAM, Probe connect holder are included.
<Options for main unit>
USCW-AI900A,
CW kit For cardiovascular examinations, add Continuous Doppler capability to sector and pencil transducers.
USCW-AI900A/EL

USUR-AI900A,
Reference Signal kit This kit is used to display reference signals (ECG waveforms etc.).
USUR-AI900A/EL

For cardiovascular examinations (for regions other than the USA):


UJUR-AI900A
ECG, Respiration, ECG gating, Heart rate
Reference signal cable
UJUR-AI901A For cardiovascular examinations (for the USA): ECG, Respiration, ECG gating, Heart rate

Reference signal sensor unit UJUR-AI902A PCG and Pulse sensor

USSE-AI900A,
Stress Echo kit Adds the cardiac stress examination function to the system.
USSE-AI900A/EL

USWT-AI900A,
2D Wall Motion Tracking kit Adds the cardiac wall motion analysis function to the system.
USWT-AI900A/EL

USEF-AI600A,
Auto EF Measurement kit Software to enable Ejection Fraction to be measured automatically.
USEF-AI600A /EL

Pencil Connector unit UIPC-AA550A This unit is used to add connectors for pencil transducers.

M-TEE hanger kit UAEH-AI900A TEE transducer hanger for the PET-512MA, PTE-512MC, and PET-512MD.

TEE hanger kit UAEH-AI901A TEE hanger for the PET-508MA.

CV kit UACV-AA550A This kit consists of preset data suitable for cardiovascular examinations, a CV sticker, and a startup screen.

This kit is used to add the STC knob on the operating panel.
STC kit UIST-AI900A Use in combination with UIUB-AI900 is not possible.
*Note that this kit can not be instaled if the UIUB-AI900A.

USHI-AA550A,
CHI kit Adds the Contrast Imaging function to the system.
USHI-AA550A/EL

(*1) Not available in the USA. (*3) Some options are required in addition to UIFR-AA550A.
(*2) Only available in the USA.

2
MPDUS0079EAB

Unit Model name Remarks


USMV-AI900A,
4D kit This unit is required for using the 4D transducer or the motor-driven TEE transducer (PET-512MD).
USMV-AI900A/EL

USLM-AI900A, Image processing technology that makes 3D/4D images of fetuses and anatomical structures appear
Luminance kit
USLM-AI900A/EL more realistic. The USMV-AI900A or Smart Sensor 3D kit is needed.

USOB-AI900A, Used for calculation of the volume by extracting the contours for regions with lower brightness in the 3D
Auto Volume Measurement kit
USOB-AI900A/EL Volume image acquired in 4D mode. Enables volume measurement, eg, antral follicle count.

USPF-AI900A,
3D printer format export kit This kit enables output of volume data for Smart 3D, Mecha4D and Smart Sensor 3D to 3D printers.
USPF-AI900A/EL

USEL-AA551A,
Elastography-FLR kit This kit enables Elastography (with FLR measurement) with linear and convex transducers. (*1)
USEL-AA551A/EL

USEL-AA550A,
Elastography kit The kit enables Elastography without FLR measurement. (*2)
USEL-AA550A/EL

USSW-AA550A,
Shear wave kit This kit allows tissue stiffness to be visualized by generating images that show shear wave propagation.
USSW-AA550A/EL

USFN-AA550A, CT/MRI/US volume data is loaded, and a CT/MRI/US planar image and an ultrasound image at the same
Smart Fusion kit
USFN-AA550A/EL position are displayed together. The UIFR-AA550A is required.

This kit is used to generate the magnetic field for acquiring positional information for transducers and
Magnetic Generator kit UIFR-AA550A needles in Smart Fusion, Smart Navigation, and Smart Sensor 3D modes. Sensor securing adaptors and
magnetic sensors are provided (one set for the PVT-375BT or PVT-375SC, and one set for the PVT-475BT).

This magnetic sensor is added for performing Smart Fusion, Smart Navigation, and Smart Sensor 3D using
Sensor kit for Fusion unit UIFR-A501A
multiple transducers. The UIFR-AA550A is required.

USFN-AI901A, Used for the Smart Fusion function. Performs automatic position matching between the ultrasound
Auto Registration kit
USFN-AI901A/EL volume data acquired in a previous exam and the real-time 2D ultrasound image.

(*1) Not available in the USA. (*3) Some options are required in addition to UIFR-AA550A.
(*2) Only available in the USA.

3
Unit Model name Remarks
TM
Position matching device for CT images (CIVCO omniTRAX ), including a magnetic sensor and 12 pieces
Auto Track CT kit UIOT-AI900A of omniTRAX.
* USFN-AA550A is required.
TM
Position matching device for MRI images (CIVCO omniTRAX ), including a magnetic sensor and 12 pieces
Auto Track MR kit UIOT-AI901A of omniTRAX.
* USFN-AA550A is required.

This pole cart allows the magnetic field generator included in the system main unit to be positioned
Fusion Pole Cart UZWT-A500A
independently.

UAFS-001A For PVT-382BT / PVT-482BT. (*3)

UAFS-002A For PVT-350BTP. (*3)

UAFS-003A For PVT-781VT. (*3)

UAFS-004A For PLT-1005BT. (*3)


Mounting kit for fusion sensor
UAFS-005A For PVL-715RST. (*3)

UAFS-006A For PVT-781VTE. (*3)

UAFS-007A For PVT-475BT. One of this kit is included in UIFR-AI900A.

UAFS-010A For PLT-1202BT. (*3)

Smart Navigation Sensor kit UISN-A500A This kit is used to mount a magnetic sensor to omniTRAX purchased by CIVCO.

Superb Micro Vascular Imaging USMI-AI600A,


Visualize Low-velocity blood flow at a high frequency rate.
kit USMI-AI600A/EL

USSS-AI900A, This function is used to display high-precision Smart 3D images by detecting the position of the
Smart Sensor 3D kit
USSS-AI900A/EL transducer based on positional information acquired using the magnetic sensor.

USMP-AI900A,
MicroPure™ kit This kit enables MicroPure, which supports visualization of small structures.
USMP-AI900A/EL

USPV-AI900A,
Panoramic View kit B/W images can be obtained with a wider field of view by moving the transducer in the lateral direction.
USPV-AI900A/EL

USRC-AI900A, This function is used to visualize images with fewer artifacts by cancelling multiple reflections from the
Multi-Reflection Canceller kit
USRC-AI900A/EL body.

Fetal Heart MPI Measurement USFH-AI600A, This kit enables Myocardial Performance Index (MPI). The MPI value can be calculated from the time
kit USFH-AI600A/EL change curve in TDI.

USZS-AI900A,
Measurement Z score kit This kit enables Z-score analysis for the measurement results.
USZS-AI900A/EL

OB kit UAOB-AI900A This kit consists of preset data suitable for OB examinations, an OB sticker, and a startup screen.

(*1) Not available in the USA. (*3) Some options are required in addition to UIFR-AA550A.
(*2) Only available in the USA.

4
MPDUS0079EAB

Unit Model name Remarks


UZRI-AI900A Rack for mounting a B/W printer (For except for UP-D711MD/P95DW-DC)

Rack for mounting a DVD recorder and a color printer on two levels.
UZRI-AI902A
(Not available with UZRI-AI900A)
Mounting kit for peripheral
unit
UZRI-AA550A Rack for mounting a B/W printer in the front or back (For UP-D711/AC or P95DW-DC)

Rack for mounting a color printer or a DVD recorder in addition to a B/W printer in the front. (not available
UZRI-AA551A
with other UZRI options).

Foot switch UZFS-004A Switch used for freezing, printing, and some other operations by foot.

Gel warmer UZGW-008A This unit warms the ultrasound gel to a suitable temperature.

Transducer cable hanger kit UZMK-AI900A Long hanger on which the transducer cable is hooked.

A basket to store transducer connectors is added to the side of the system main unit (one basket is
Transducer Holder kit UZBK-AI900A
provided in the standard configuration)

This kit used to establish connection to the DICOM network via wireless LAN. Complied with the
UIWL-A500A Radio Law of Japan and applicable laws and regulations of USA, EU member states, Iceland, Norway,
Liechtenstein, and Switzerland.
Wireless LAN kit
This kit enables connection to the DICOM network via wireless
UIWL-AI900A
LAN. (For Singapore, Malaysia, Saudi Arabia, Russia, Australia, and Turkey.)

EV/ER Transducer Holder kit UZPH-AI900A This kit is used to add a dedicated holder for endocavitary transducers.

Keyboard kit UIKB-AI900A This kit is used to add a hardware full keyboard for entering the patient ID and comments.

This kit adds a tablet terminal that communicates with the system main unit via wireless LAN to
allow viewing and easy operation of the examination screen. Complied with the Radio Law of Japan
2nd Console kit UITB-AI900A
and applicable laws and regulations of USA, EU member states, Iceland, Norway, Liechtenstein, and
Switzerland. UIWL-A500A is included.

DataBase for External USDB-AI900A, Function for setting up the patient database in the USB HDD in order to perform examinations.
HDD kit USDB-AI900A/EL *UZHI-AI900A is required.

Box with lock for installing the external HDD. HDD itself is not included.
Mounting kit for External HDD UZHI-AI900A
*USDB-AI900A is required.

ECG Cable Hanger kit UZMK-AI901A Hook used to hang the ECG cables on the front of the operating panel.

Kit for adding a USB port to the operating panel.


Panel USB Port kit UIUB-AI900A
Note that this kit cannot be installed if the UIST-AI900A is installed.

A battery kit to provide approximately 30 minutes of operation without being connected to a power out-
Battery unit UEBT-AA550A
let.

Track Ball kit UZTB-AI900A Newly designed (heavier) trackball to improved response to user and therefore workflow.

(*1) Not available in the USA. (*3) Some options are required in addition to UIFR-AA550A.
(*2) Only available in the USA.

5
Unit Model name Remarks

Tilt panel kit UITP-AA550A Kit for enabling tilting of the touch panel.

USHE-AI900A,
Online Help kit Kit for displaying the operation manual on the viewing monitor.
USHE-AI900A/EL

USPA-AI900A, A sequence of operations is registered, and each operation is executed by single switch operation. Not
Protocol Assistant kit
USPA-AI900A/EL necessary when USSE-AI900 is installed.

USRI-AI600A,
Reference Image kit Software to enable images from a previous examination to be displayed at the right of the screen.
USRI-AI600A/EL

This kit provides software for security management of the system.


Security Management kit USSM-AI900A
This kit is already installed in systems.

UZKF-AI900A
(French)

UZKG-AI900A
(German)

UZKI-AI900A
(Italian)

UZKS-AI900A
(Spanish)
This kit is intended for changing the key tops of the full keyboard to support specific languages.
Local Language Key-Top kit
UZKD-AI900A
(Danish)

UZKN-AI900A
(Norwegian)

UZKW-AI900A
(Scandinavian)

UZKR-AI900A
(Russian)

(*1) Not available in the USA. (*3) Some options are required in addition to UIFR-AA550A.
(*2) Only available in the USA.

6
MPDUS0079EAB

BLOCK CHART SYSTEMS

Main unit A
CUS-AA450A

Reference Signal kit CHI kit


USUR-AI900A USHI-AA550A

Reference signal cable


UJUR-AI900A 4D kit
UJUR-AI901A USMV-AI900A

Stress Echo kit Luminance kit


USSE-AI900A USLM-AI900A

Reference signal sensor unit


Auto Volume
UJUR-AI902A
Measurement kit
USOB-AI900A
2D Wall Motion Tracking kit
USWT-AI900A
3D printer format export kit
USPF-AI900A
CW kit
USCW-AI900A
Elastography-FLR kit Elastography kit
Pencil Connector unit USEL-AA551A USEL-AA550A
UIPC-AA550A

Shear wave kit


USSW-AA550A
CV kit
UACV-AA550A

STC kit
UIST-AI900A Note: Not Available with Panel USB Port,
UIUB-AI900A.

ECG Cable Hanger kit


UZMK-AI901A

TEE-Hanger
UAEH-AI900A
UAEH-AI901A

Foot Switch
UZFS-004A

Auto EF Measurement kit


USEF-AI600A

Fetal Heart MPI Measurement kit


USFH-AI600A

A B

7
B C

MicroPure kit Keyboard kit


USMP-AI900A UIKB-AI900A

Panoramic View kit Protocol Assistant kit


* Not necessary when
USPV-AI900A USPA-AI900A USSE-AI900 is installed.

Multi-refrection DataBase
Canceller kit for External HDD kit
USRC-AI900A USDB-AI900A

Mounting kit
Super Micro-vascular for External HDD kit
Imaging kit UZHI-AI900A
USMI-AI600A
Panel USB Port kit
Refference Image kit UIUB-AI900A Note: Not Available with STC kit,
UIST-AI900A.
USRI-AI600A
Online Help kit
Measurement Z score kit USHE-AI900A
USZS-AI900A
Buttery unit
OB kit UEBT-AA550A
UAOB-AI900A

Transducer holder kit


Mounting kit Mounting kit Mounting kit Mounting kit UZBK-AI900A
for a B/W printer for a B/W printer for Peripherals for Peripherals
(except UP-D711MD, (only for UP-D711MD/ (for video recorder, (for B/W printer,
P95D-DC) AC, P95D-DC) color printer etc.) color printer etc.) Tilt panel kit
UITP-AA550A
UZRI-AI900A UZRI-AI901A / UZRI-AA550A UZRI-AI902A UZRI-AA551A
* Not available with * Not available with
UZRI-AI900A. other UZRI options.
Track Ball kit
UZTB-AI900A
Wireless LAN kit Wireless LAN kit
UIWL-A500A UIWL-AI900A

2nd Console kit


UITB-AI900A Note: UIWL-A500A is included.

Gel warmer
UZGW-008A

Transducer cable hanger kit


UZMK-AI900A

EV/ER Transducer Holder kit


UZPH-AI900A

C D

8
MPDUS0079EAB

Magnetic Generator kit


UIFR-AA550A

Fusion Pole Cart


UZWT-A500A

Mounting kit
for fusion sensor
UAFS-001A, UAFS-002A, UAFS-003A,
UAFS-004A, UAFS-005A, UAFS-006A,
UAFS-007A, UAFS-010A

Sensor kit for Fusion unit


UIFR-A501A

(Note)
Smart Fusion kit (Note) Smart Sensor 3D kit
USFN-AA550A USSS-AI900A

Auto Registration kit Luminance kit


USFN-AI901A USLM-AI900A

Auto Track CT kit


UIOT-AI900A

Auto Track MR kit 3D printer format export kit


UIOT-AI901A USPF-AI900A

Note: USFN-AI900A and USSS-AI900A


can be installed simultaneously.

9
TRANSDUCER OPTIONS/OPERATION MODES
Freq. Freq.Range Precision
Transducer name Range 2D ApliPureTM MicroPureTM *1 BEAM M CDI Power
(MHz) (MHz) Imaging

PSI-70BT i10S4 7.0 4.0~9.0     


PST-25BT 5S1 2.5 1.8~4.7     
PST-30BT 5S2 3.0 1.8~5.2     
PST-50BT 6S3 5.0 3.0~8.2     
PVT-350BTP 6CP1 3.5 2.0~4.5      
PVT-375BT 6C1 3.5 1.5~6.0      
PVT-375SC *8 6Cs1 3.5 1.5~6.0      
PVT-382BT 6MC1 3.5 2.0~5.0      
*8
PVT-475BT 8C1 4.0 1.8~6.2      
*8
PVT-482BT 8MC1 4.0 1.8~6.0      
PVT-674BT 10C3 6.0 3.5~9.7      
PVT-675MVL 9CV2 6.0 2.5~7.0      
PVT-675MVS 9CV2 6.0 2.5~7.5      
PVT-681MVL 11CV3 6.0 3.6~11.0      
PVT-712BT 11MC4 7.0 4.3~11.0      
PVT-745BTF 11CI4 7.0 3.0~10.0      
PVT-745BTH 11CI4 7.0 3.0~10.0      
PVT-745BTV 11CI4 7.0 3.3~10.0      
PVT-781VT 11C3 7.0 3.6~10.5      
PVT-781VTE 11C3 7.0 3.6~10.5      
Convex 7.5 4.5~9.0      
PVL-715RST 11CL4
Linear 7.5 4.5~9.0      
PLT-308BTP 6LP3 3.75 2.0~5.5      
PLT-704SBT 11L4 7.5 4.0~10.0       
PLT-705BT 11L3 7.0 3.0~8.0       
PLT-705BTF 11LI4 7.0 3.8~8.4      
PLT-705BTH 11LI4 7.0 3.8~8.4      
PLT-1005BT 14L5 10.0 3.8~10.0        
PLT-1202BT 17LH7 12.0 4.5~17.0      
PLT-1204BT 18L7 12.0 4.5~18.0        
PET-508MA 7S3 5.0 2.5~7.0    
PET-512MA 8S2 5.0 3.0~6.5    
PET-512MC *5 8SM2 5.0 3.0~6.5    
PET-512MD *5 8SM2 5.0 3.0~8.5    
PET-805LA 12LI4 8.0 3.5~9.0      
PC-20M P2 2.0 2.0
PC-50M P5 5.0 5.0

*1: Optional software is required. *6: Optional unit UIPC-AA550A is required.


*2: Depends on the preset. *7: Optional software USCW-AI900A is required.
*3: Not available in the USA. *8: Using Single Crystal technology.
*4: Not available in the USA and Canada.
*5: Optional software USMV-AI900A is required.

10
MPDUS0079EAB

TRANSDUCER OPTIONS/OPERATION MODES


*1
Elasto-*1 SMI /
*1
Shear CHI *5 Volume
Transducer name TDI PWD CWD*7 4D *5
graphy ADF wave*1 2D SMI
*1
ADF MFI VRI color
PSI-70BT      
PST-25BT    
PST-30BT      
PST-50BT    
PVT-350BTP      
PVT-375BT         
PVT-375SC *8         
PVT-382BT      
*8 *4
PVT-475BT          
*8
PVT-482BT      
*4
PVT-674BT      
PVT-675MVL    
PVT-675MVS    
PVT-681MVL        
PVT-712BT  
PVT-745BTF    
PVT-745BTH    
PVT-745BTV  
PVT-781VT       
PVT-781VTE       
Convex     
PVL-715RST
Linear  

PLT-308BTP  
PLT-704SBT     
PLT-705BT     
PLT-705BTF    
PLT-705BTH    
PLT-1005BT         
PLT-1202BT     
PLT-1204BT      
PET-508MA   
PET-512MA   
PET-512MC *5   
PET-512MD *5   
*4
PET-805LA      
PC-20M 
PC-50M 

*1: Optional software is required. *6: Optional unit UIPC-AA550A is required.


*2: Depends on the preset. *7: Optional software USCW-AI900A is required.
*3: Not available in the USA. *8: Using Single Crystal technology.
*4: Not available in the USA and Canada.
*5: Optional software USMV-AI900A is required.

11
TRANSDUCER OPTIONS/OPERATION MODES
*1
*5 STIC Smart Sensor *1 Smart 2D
Transducer name STIC *5 Fusion *1 Remarks
Color 3D 3D*1 Navigation WMT
PSI-70BT 
PST-25BT 
PST-30BT 
PST-50BT 
PVT-350BTP     *4
PVT-375BT    
PVT-375SC     *8
PVT-382BT    
PVT-475BT     *8
PVT-482BT     *8
PVT-674BT 
PVT-675MVL *2 *2 *5
*2 *2
PVT-675MVS   *5
PVT-681MVL *5
PVT-712BT 
PVT-745BTF 
PVT-745BTH 
PVT-745BTV 
PVT-781VT    
PVT-781VTE    
Convex   
PVL-715RST
Linear   
PLT-308BTP  *4
PLT-704SBT 
PLT-705BT 
PLT-705BTF 
PLT-705BTH 
PLT-1005BT    
PLT-1202BT   
PLT-1204BT 
PET-508MA 
PET-512MA  *4
PET-512MC  *5
PET-512MD  *5
PET-805LA
PC-20M *6, *7
PC-50M *6, *7

*1: Optional software is required. *6: Optional unit UIPC-AA550A is required.


*2: Depends on the preset. *7: Optional software USCW-AI900A is required.
*3: Not available in the USA. *8: Using Single Crystal technology.
*4: Not available in the USA and Canada.
*5: Optional software USMV-AI900A is required.

12
MPDUS0079EAB

SPECIFICATIONS • Scan Angle and Scan Width


– Adjustment of the field width (scan width, scan angle) is
System possible.
• Scan methods: Linear scan (some transducers – Adjustment of beam steering (scanning position) is
can perform oblique scanning) possible.
Sector scan – Adjustment of linear beam steering is possible.
Convex scan • PAN/EXPAND
Trapezoid scan Real-time PAN/EXPAND
Curved vector scan – Scale enlargement/reduction using the encoder is
• Monitor: possible.
High-definition 21.5-inch Wide LCD monitor with LED – Movement to the desired section using the trackball is
backlight possible.
– Resolution: 1920 × 1080 (Full HD) – The transmission focus is optimized in steps above.
– Viewing angle: 178 degrees – The specified range of an image can be magnified.
– Contrast ratio: typ. 1500: 1 (Spot Zoom)
– Response time (ms): typ. 14 • Transmission Focus
– Luminance (cd/m2): typ. 300 Transmission conditions: A maximum of 8 steps
– Conformance Standard: DICOM Part 14 • Transmission Frequency
• Presets Multi frequency: 3 frequencies can be selected
– System preset: 1 type from 13 types.
Compatible Peripheral Devices • GAIN
• Black-and-white digital printer – The display brightness for 2D can be changed.
– UP-D711MD/WO: AC (100 V to 240 V, SONY) (Also available when the image is frozen)
– UP-D898MD: AC (100 V to 240 V, SONY) – The display brightness for 2D and M can be changed
– P95DW: AC (100 V to 240 V, MITSUBISHI) simultaneously.
– P95DW-DC: DC (24 V, MITSUBISHI) • STC
• Color digital printer – Software STC
– UP-D25MD: AC (100 V to 240 V, SONY) · Depth direction from the body surface: 8-step slide
– CP30DW: AC (120 V, 220V to 240 V, MITSUBISHI) controls (common for 2D and M)
• DVD video recorder · Lateral direction in the image: 6-step slide controls
– HVO-550MD/FHD: AC (100 V to 240 V, NTSC/PAL, SONY) (common for 2D and M)
• USB flash drive – Hardware STC (UIST-AI900A is required.)
• External HDD 8-step slide controls (common for 2D and M)
• Barcode reader • Acoustic Output
Adjustment is possible to 100%.
2D mode (B mode) • Adjusting the 2D Image Quality
• Viewing Depth – Dynamic range (also available when the image is frozen)
The viewing depth depends on the transducer used. – Time-smoothing (persistence)
– Convex – Gamma (also available when the image is frozen)
· Minimum depth: 2 cm – Frame rate
· Maximum depth: 50 cm – ApliPure
– Linear – Precision
· Minimum depth: 1 cm – Reberb (USRC-AI900A is required.)
· Maximum depth: 14 cm • 2D Map
– Sector – The grayscale pattern can be changed and virtual color
· Minimum depth: 1 cm setting for the 2D image is possible.
· Maximum depth: 28 cm – Settings can also be changed when the image is
• Line density frozen.
– The line density differs depending on the transducer • 2D Quick Scan
used. The gain and STC can be adjusted automatically.
– The line density can be changed. • THI (Tissue Harmonic Imaging)
• Ultrasound Frame Rate THI signal processing methods
– The ultrasound frame rate can be adjusted by using the – Pulse subtractionTM method
following in combination. – Filtering method
· Line density – Differential method
· Parallel signal processing • Display Orientation
– Top/bottom reversal is possible.
– Left/right reversal is possible.

13
• Image Size • Flex-M
The displayed image size can be switched between Any desired plane can be set on the 2D-mode image and
small and large. the M-mode image for the set plane can be reconstructed.
• ApliPure™
– ApliPure Doppler (Spectrum Doppler)
This function reduces ultrasound wave interference with- • Doppler Mode
in tissues, which appear as speckle patterns or speckle – PWD (pulsed-wave Doppler)
noise on 2D images. – HPRF PWD (can be switched to HPRF mode)
– ApliPure+ – CWD (continuous-wave Doppler)
This function can display the boundaries between (USCW-AI900A is required.)
tissues more clearly and reduce speckle noise and acous- – Pencil CWD (pencil-type transducer)
tic shadows. (UIPC-AA550A and USCW-AI900A are required.)
• MicroPure (USMP-AI900A is required.) • Doppler Pulse Repetition Frequency (PRF)
– PWD: 0.3 kHz to 52.1 kHz
– Small structures can be extracted by performing
– CWD: 1.4 kHz to 52.1 kHz
filtering for 2D-mode images.
• Doppler Scan
– Visualization of very small calcifications and other
– 2D/D simultaneous scan
extremely small lesions can be improved.
– D only scan
• Precision Imaging
• Doppler Sampling Volume
– Precision+
The Doppler range gate width can be changed. (Minimum
· Structures in 2D-mode images can be displayed more
0.3 mm)
clearly and the background can be displayed more
• Doppler Sampling Shift
smoothly.
0 cm to the maximum depth
· Saturation in high-intensity regions of tissue structures
• Doppler Cursor Mode
is reduced, allowing the tissue structures to be dis-
Operation for the 2D live image is possible with the
played in a more natural manner.
Doppler sampling volume displayed in it.
• TSO (Tissue Specific Optimization) • Doppler Filter
– Reception focus compensation can be performed. The Doppler filter cutoff can be changed.
– Automatic reception focus compensation can be • Doppler Gain
performed. (Auto TSO) The display brightness for Doppler can be changed.
• BEAM (Biopsy Enhancement Auto Mode) • Doppler Quick Scan
– Display of the needle can be enhanced in the image. The Doppler scale and baseline shift can be adjusted auto-
– The enhancement level can be adjusted. matically.
• Doppler Frequency Analysis and Image Processing
M mode
– Method: FFT
• M Transmission Frequency
– No. of data items: 255 (maximum)
Multi frequency: 5 types (at maximum)
• Indication of Doppler Spectrum Direction
• M Sweep Speed
Reverse display of the velocity spectrum is possible.
The Sweep Speed can be changed in M mode.
• Doppler Baseline Shift (Zero Shift)
• M Gain
– The velocity baseline of Doppler images can be shifted.
M gain can be corrected for 2D gain. – The baseline shift setting can also be adjusted when
• M Image Processing Parameters images that were frozen are displayed.
– M dynamic range (can be changed even after the image • Doppler Audio
has been frozen). Stereo output (blood flow toward and away from the trans-
– M auto gain control ducer)
– M gamma (can be changed even after the image has • Doppler Map
been frozen). The brightness conversion table and the virtual color for
• M Map Doppler images can be set.
M image virtual color setting is possible. The setting can be • Display of Doppler Scale
changed even after the image has been frozen. 2 types (velocity, Doppler shift frequency)
• THI (Tissue Harmonic Imaging) • Doppler Focus
2D mode and THI mode are linked, and M images can be Automatically follows the sample position.
displayed in THI mode. • Doppler Angle Mark
– Pulse subtraction method This mark is displayed for measuring the angle between
– Filtering method the direction of the velocity and the direction of the ultra-
– Differential method sound beam.
• M Mark • Doppler Oblique Scan (PWD Steering)
– The M cursor can be displayed on 2D or C images. – Oblique scans are possible using a specific linear trans-
– The M cursor displayed position can be adjusted. ducer.
– Auto Invert function
14
MPDUS0079EAB

• Doppler Multifrequency • C Multifrequency


The PWD transmission frequency can be changed. The transmission frequency for color Doppler image acqui-
• Doppler Sweep Speed sition can be changed.
The Sweep Speed can be changed in Doppler mode. • C Line density
• Doppler Display Dynamic Range The color Doppler image line density can be changed.
The display dynamic range of the Doppler image can be • C ROI (Region of Interest)
changed. Position, size, and steering adjustment is possible for color
• Doppler Auto Trace (measurement performed after Doppler ROIs.
freezing the image) • C Transmit Focus
Measurement of peak velocity and mean velocity is Automatically follows the color Doppler ROI position.
possible by automatic velocity tracing. • C Filter
The following Doppler waveform trace is possible. Color Doppler low-cut filter can be changed.
– Trace style: Waveform Peak, Mean, Peak + • Variance Curve
Mean – The display of the color variance component can be
– Trace area specified: Forward, Reverse, Full, Auto adjusted.
– Measurement item: Max, Min, Mean, PI, RI, etc. • Color Quick Scan
The following operations are possible when a linear trans-
Color Doppler ducer is used.
Color Doppler Mode · The position of a color ROI and angle of color steering
• Display mode are adjusted automatically.
– CDI mode · When PWD sampling volume is displayed, the Doppler
· Flow velocity gate position, Doppler steering angle, and Doppler
· Flow velocity/variance angle are adjusted automatically.
· Power
– Power Angio mode Color Doppler M mode (MDF Mode)
– TDI mode • Display mode
– TwinViewTM – MCDI mode
Simultaneous dual-screen display with 2D mode is avail- · Velocity display
able. · Velocity/variance display
– SMI (Superb Micro vascular Imaging) mode · Power display
(USMI-AI600A is required.) – M-TDI mode
· Clutter suppressed • M Color Doppler Map (CDI MAP)
· Blood flow enhanced Color Doppler map can be selected for each mode.
– ADF (Dynamic Flow) mode • M Color Doppler Velocity Range Selection (C Scale)
· Direction display The velocity range can be selected.
• C Map • M Color Doppler Baseline (C Baseline)
– C map can be selected for each color Doppler mode. – The zero-velocity line on the M Color Doppler image can
– Changes can also be made when the image is frozen. be shifted.
• C Scale (Switching the Velocity Range) – The baseline shift setting can also be adjusted when
The velocity range can be changed. images that were frozen are displayed or when the
• C Time-Smoothing (Persistence) image in the image memory is played back.
The result of temporal correlation processing between the • Color Reverse Display
previous image and current image can be displayed. – The colors can be reversed.
• C Baseline (Zero Shift) – Changes can also be made when the image is frozen.
– The velocity baseline of color Doppler images can be • Black and White/Color Balance
shifted. – Color weighting for B/W images can be set by compar-
– The baseline shift setting can also be adjusted when ing the M Color Doppler images and B/W images.
images that were frozen or images in the image memory – Changes can also be made when the image is frozen.
are displayed. • Color Gain
• Reverse C Display The display brightness of the M Color Doppler image can
– Coloring is reversed. be changed.
– Changes can also be made when the image is frozen. • M Color Doppler Multi-Frequency
• Black and White/Color Balance Doppler transmission frequency can be selected in M
– By comparing the color Doppler images and B/W imag- Color Doppler image acquisition.
es, color weighting to B/W can be set. • M Color Doppler Filter
– Changes can also be made when the image is frozen. M Color Doppler low-cut filter can be changed.
• C Gain
The display brightness of color Doppler images can be
changed.
15
Reference Signals · The following functions can be used.
(USUR-AI900A and UJUR-AI900A, or USUR-AI900A and UJUR- Volume Color
AI901A are required.) Multi View
• Type Magic Cut
– Electrocardiogram (ECG) VolPureTM
· Lead I is the standard connection. Volume View
· External input is possible. STIC/STIC Color
· DC IN 4D CHI (USHI-AA550A is required.)
The connected device must comply with IEC 60601-1. 4D Biopsy
· Top/bottom inversion is possible. Luminance (USLM-AI900A is required.)
· Lead switching OmniView
· Pacemaker Auto flexible cut line
· INST STL export (USPF-AI900A is required.)
– Phonocardiogram (PCG) (UJUR-AI902A is required.) · The following measurements can be performed.
· PCG microphone: Acceleration type MPR
· Filter: Switching is possible. Multi Auto Volume measurement (USOB-AI900A is
– Pulse (UJUR-AI902A is required.) required.)
Pulse transducer: Air conduction type
– Respiration
Impedance method using the reference signal cable
• Heart Rate 4D OPERATION MODES
The heart mark blinks in synchronization with the heart The optional software USMV-AI900A is required.
beat detected by the ECG.
Transducer Single 4D
The heart rate is displayed. 4D Live STIC STIC Color
name Sweep CHI*1
• Reference Signal Sweep Speed PVT-675MVL    *2
*2
This changes the reference signal sweep speed. PVT-675MVS   *2 *2

Other Diagnostic Function PVT-681MVL   


• CHI (Contrast Harmonic Imaging) (USHI-AA550A is
Transducer Volume 4D
required.) name color Biopsy
The second-harmonic wave signals from the microbubbles PVT-675MVL 
in the contrast medium can be effectively visualized.
PVT-675MVS 
– The following image modes can be selected.
PVT-681MVL  
· PS (Pulse Subtraction) -Low, PS-Low2
· VRI (Vascular Recognition Imaging) Transducer Max sweep
· Fundamental Luminance*1
name range (deg)
· CHI ADF PVT-675MVL  90
· SMI (Superb Micro Vascular Imaging) PVT-675MVS  90
(USMI-AI600 is required.)
PVT-681MVL  150
– The following functions can be selected.
· 2D TwinView
· Micro Flow Imaging (MFI) *1: Optional software is required.
· Image Stabilizer *2: Depends on the preset.
· MI Constant function
• Parametric MFI
– Temporal information can be displayed as a color map • Stress Echo (USUR-AI900A and UJUR-AI900A, or USUR-
superimposed on images acquired by 2D mode (without AI900A and UJUR-AI901A are required.)
CHI starting up), ADF/SMI mode, and CHI mode (contrast Exercise and pharmacological stress echo examinations
image for the period from the start of contrast medium can be performed.
injection to the time when the contrast medium reaches – Data acquisition mode
the target region). This mode is intended for selecting and creating protocols.
– The following functions are available. – Review mode
· Micro Flow Imaging (MFI) This mode provides playback function/data output func-
· Image Stabilizer tion/Wall Motion Scoring (WMS) function.
• Mechanical 4D (USMV-AI900A is required.) • Panoramic View (USPV-AI900A is required.)
Three-dimensional image data (volume data) can be gen- – A continuous image can be acquired by moving the
erated and displayed by using image data acquired for transducer horizontally on the body surface.
three-dimensional image reconstruction. – Measurement using Panoramic View can be performed.

16
MPDUS0079EAB

• Elastography (USEL-AA551A or USEL-AA550A is required.) – ECG Sync Acquisition function can be used.
– Tissue stiffness can be visualized based on the changes – Shear wave measurement can be performed.
in velocity resulting from physical compression and • Smart Sensor 3D (USSS-AI900A, UIFR-AA550A, and UIFR-
decompression of the target region. AI501A are required.)
– FLR measurement can be performed to calculate the Generation of 3D images can be performed based on the
strain within the set ROI. (Not available in the USA.) positional information obtained using the magnetic sensor
• VI (Vascularity Index) attached to the transducer.
The number of pixels in the Power/ SMI image and within – The following functions can be used.
the ROI, the area of the Power/ SMI image and of the ROI, · Volume Color
and the ratio of the number of pixels in the Power/ SMI · Multi View
image to the number of pixels within the ROI can be dis- · Volume View
played for an image acquired in Power/ SMI mode. · Magic Cut
Available on all linear transducers. · CHI (USHI-AA550A is required.)
• Fusion (Smart Fusion) (USFN-AA550A etc. is required.) · Luminance (USLM-AI900A is required.)
Synchronization of ultrasound scanning with CT/MRI · Shear Wave (USSW-AA550A is required.)
image display adjusted according to the examination posi- · OmniView
tion determined using a magnetic sensor attached to the · ECG Sync Construction
transducer can be performed. · STL export (USPF-AI900A is required.)
– The following video modes can be selected. – The following measurements can be performed.
· 2D mode (B mode) · MPR
· Color Doppler · Multi Auto Volume Measurement (USOB-AI900A is
· PWD (pulsed-wave Doppler) required.)
· CWD (continuous-wave Doppler) (USCW-Al900A is • Smart 3D
required.) 3D image can be generated from the 2D image and any
· CHI (Contrast Harmonic Imaging) (USHI-AA550A is input volume shape.
required.) – The following functions can be used.
· Elastography (USEL-AA551A or USEL-AA550A is · Volume Color
required.) · Multi View
– The following image data can be referred to. · Volume View
· CT · Magic Cut
· MR · Omni View
· PET · CHI (USHI-AA550A is required.)
· 3D US (USSS-AI900A is required.) · Shear Wave (USSW-AA550A is required.)
– The following functions can be used. · STL export (USPF-AI900A is required.)
· Blend display • Reference (USRI-AI600A is required.)
· Segment display The ultrasound images of the current examination and the
· Multiplane display (Triple display) previously acquired image of the patient can be displayed
· Multivolume display (Quad display) simultaneously.
· 3D body mark display (Quad display) – The following video modes can be selected.
· Comparative display (Quad display) · 2D mode (B mode)
· Auto Registration (USFN-AI901A is required.) · Color Doppler
· Auto Track (UIOT-AI900A or UIOT-AI901A is required.) – The following image data can be referred to.
– The following measurements can be performed. · US image (still image)
· Basic measurement · Raw data acquired using CUS-AA450
· Cardiac measurement (for ultrasound live image) · CT
• Protocol Assistant (USPA-AI900A is required.) · MR
A series of operations (a protocol) that has been created · MG
for the intended examination can be executed automatically. · PET
Protocols can be created and edited. – The following measurement can be performed.
• Shear wave (USSW-AA550A is required.) Basic measurement
Images representing the speed of propagation of tissue
displacement (shear wave speed) can be visualized (shear
wave scan) by locally displacing tissues by transmitting a
burst wave with high acoustic pressure.
– The following Shear wave display modes are available.
· Speed: shear wave speed display (m/s)
· Elasticity: elasticity display (kPa)
· Propagation: propagation display

17
Display-Related Features · DVD/CD write status
• Display Method · Network use status
– Images on the main unit: 60 Hz non-interlaced display · PACS use status
– Images from external · Used space on HDD
playback devices: 60 Hz non-interlaced display · Saving dynamic/still image
• Monitor Display/Character Display · DICOM printer status/peripheral device status
– ID area · USB flash drive status display
· Patient ID – Multifunction display area
· Patient name Assignment statuses for trackball and surrounding
· Hospital name switches and dials
· Date: Selected from among the formats • Annotation
shown below. – Manual input using the keyboard is possible.
YYYY/MM/DD – Auto annotation (previously specified text) is possible.
MM/DD/YYYY • Pictograms
DD/MM/YYYY Body icons and transducer mark
YYYY: Western calendar year • Biopsy Guide Mark
MM: Month Biopsy guide mark display is possible.
DD: Day • Touch Panel (TCS: Touch Command Screen)
· Time: Selected from among the formats – 12.1-inch LCD monitor: SXGA (1280 x 800)
shown below. – The touch panel tilt angle can be changed by 15°. (UITP-
hh:mm:ss: AM (PM) AA550A is required.)
hh:mm:ss: 24-hour representation • Language
hh: Hour The following languages are supported for the display of
mm: Minute some screens and keyboard entry.
ss: Second
· VIDEO mark, VCR counter
· Age
· Sex
Supported Screen Input by software Input by hardware
· Heart rate display (heart-shaped mark/heart rate)
· Name of the Imaging Preset language display keyboard keyboard*
· Name of the operator English   
· Gestational age
– Acoustic power display area  
English (UK) 
· Acoustic power value (%) (same as"English") (same as"English")
· TI value
German   
– Auto data
Frame rate French   
Acoustic power index = MI indication Italian   
Transducer frequency
Depth Spanish   
Dynamic range Danish   
GAIN
Dutch   
CDI filter
PRF Norwegian   
Doppler filter Swedish   
Doppler angle
Doppler gate size Finnish   
– Thumbnail area Portuguese   
· Image data acquired during the current examination is
Icelandic   
displayed.
· Information from a previous examination of the patient Russian   
currently being examined is displayed. Japanese   
– Information message display area
Chinese   
An operation guide and other messages are displayed.
– Status area : Applicable : Not applicable
The following system statuses can be displayed. *: UIKB-AI900A is required.
· Battery capacity

18
MPDUS0079EAB

Measurement Functions – M-mode measurements


Basic Measurement Functions · LV (left ventricular function) measurements
• 2D-mode measurements · AV (aortic valve) measurements
– Distance · MV (mitral valve) measurements
· Distance – Doppler measurements
· Trace Length · Trans-Aortic valve flow measurement
· Mean-IMT · Trans-Mitral valve flow measurement
– Area · Trans-Pulmonary vein flow measurement
– Angle · Trans-Tricuspid valve flow measurement
· Angle · Trans-Pulmonary valve flow measurement
· Joint · Blood flow waveform auto measurements
– Volume · Coronary measurements
– Stenosis ratio · PISA measurements
· %Stenosis (Distance) – OB (obstetrics) measurements
· %Stenosis (Area) · The data for determining fetal growth based on the
• 4D-mode measurements (USMV-AI900A is required.) measured fetal size is displayed.
– Distance · The list of measured data or a graph of the measured
· Distance value development (fetal growth conditions) is displayed.
· Trace Length · Week function (gestational age)
· Mean-IMT · Measurement data saving is possible.
– Area · Auto NT measurement
– Angle · Fetal heart MPI measurement (USFH-AI600A is
· Angle required.)
· Joint · Anatomy
– Volume · User chart registration
· Volume – Vascular measurement
· Auto Volume Measurement (USOB-AI900A is required.) · CCA (Common Carotid Artery) measurement
– Stenosis ratio · ECA (External Carotid Artery) measurement
· %Stenosis (Distance) · ICA (Internal Carotid Artery) measurement
· %Stenosis (Area) · Vert A (Vertebral Artery) measurement
– Shear Wave measurement · Subclav A (Subclavian Artery) measurement
• M-mode measurements · Auto-IMT measurement
– Slope · IMT-C10 measurement
– Distance – User-registered measurements registration function.
– Time · Measurement items and calculation items based on
– Heart rate the measured values
• PW/CW Doppler measurements · Layout setting on the Worksheet screen
– Velocity · Switch layout setting of the touch panel
– Acceleration · Measurement package DICOM code registration
– Time
– Heart rate Advanced Measurement Functions
– PI • 2D wall motion tracking (USWT-AI900A is required.)
– RI Wall motion can be analyzed by semi-automatically
– S/D extracting the left ventricular myocardium from the image
– Flow volume data acquired by the system and displaying it for the eval-
– Doppler trace uation of myocardial motion.
– Wall motion tracking in 2D dynamic images
Application Measurement Functions – Wall motion information display
• Cardiac measurements – Polar map display
– 2D-mode measurements – Local/whole myocardial wall motion parameter curve
· LV (left ventricular function) measurements display
· LA (left atrial volume) measurements – Parameter setting display of various parameters
· AV (aortic valve) measurements – Analytical data output to a file
· MV (mitral valve) measurements
· PV (pulmonary valve) measurements
· LV MASS measurements
· Auto EF measurements (USEF-AI600A is required.)
19
Measurement of Stored Image Data Other: Single partition
The following measurements can be performed for the – DVD/CD drive
DICOM data (DICOM, with raw data, and without raw data) – USB flash drive
stored in HDD. – Network: DICOM connection
• Basic measurement
• Application measurement Security Function
• Security Control
Report Function This system supports a function for recording the user's
• Worksheet functions authorization and access log in order to protect personal
– The measurement and calculation items can be dis- information.
played for each application measurement. – User authentication
– Data editing is possible (except for some items). – Audit Log
– Display of the following values can be set to ON or OFF. – De-identification (live image/stored image)
Mean value, latest value, maximum value, minimum value • Antivirus
– Trend graphs can be displayed (OB measurement work- Whitelist-based antivirus software is employed. The soft-
sheet). ware permits only executable files registered in the
– Comment entry is possible. whitelist to be executed, preventing execution of malware.
• Report function (On Board Report)
– Reports can be created on the system. Maintenance Function
– The created reports can be printed. • Remote Maintenance (Option)
– The created reports can be output as PDF files. This function makes it possible to remotely control the
– The report template can be edited. above systems for maintenance.

Cine Memory (large-capacity image memory) Image Format to Export


• Memory Capacity – Still: BMP/ JPEG
960 MB – Movie: WMV9/ H.264
• Record/Playback Mode
– Loop playback is possible. Network
– Frame advance playback is possible. • Ethernet: 10BASE-T/100BASE-TX/Gigabit
– Cine playback is possible in Doppler or M mode. Ethernet
– Live images can be recorded. (Clips, Auto Store) • Network client system

Video Recording Wireless Network


The following DVD remote control operations are possible: (UIWL-A500A is required or UIWL-AI900A is required.)
Record, stop, play, fast-forward, rewind, forward search, Wireless network connection is possible with this function.
reverse search, and freeze (pause). • Standard
– IEEE 802.11 b/g/n 2.4 GHz (UIWL-A500A, UIWL-AI900A)
Recording Function – IEEE 802.11 a/n/ac 5 GHz (UIWL-A500A)
• Printers (option) • Security
– Black-and-white printer: USB connection – WPA2-PSK [AES]
– Color printer: USB connection – WPA2-Enterprise [AES] (conformed)
• Video Recording Units (Option) • Frequency
– DVD Video – 2.4 GHz to 2.5 GHz CCK/OFDM modulation (UIWL-A500A,
• Electronic Filing UIWL-AI900A)
– Hard disk drive – 5 GHz OFDM, 802.11 n MCS0-7, 802.11 ac MCS0-9 code
· Built-in HDD (SATA) system (UIWL-A500A)
· External HDD (USB3.0)
* UZHI-AI900A and USDB-AI900A are required. DICOM Function
The HDDs should meet the following specifications: • DICOM data type
Capacity: 500GB to 6 TB – US Image (still image)
Speed: 7200 rpm or higher – US Multi Frame (dynamic image)
Interface: USB 3.0 (bus power) – SC Image (storage in a separate file)
External dimension: 117(W) × 81(D) × 22.5(H) mm or – Enhanced US Volume (Volume data image)
less – Structured Report (measurement result information)
Format: The following windows® formats • Server connection
(read/write compatible) – Storage (Server/Media)
NTFS – MWM (Modality Worklist Management)
FAT32 (up to 2 TB)
20
MPDUS0079EAB

– MPPS (Modality Performed Procedure Step) • 2nd Console (UITB-AI900A is required.)


• Storage function Basic operation of the system can be performed on a tab-
– Storage Commitment let terminal while referring to the image in real time.
– Query/retrieve • Battery Mode (option) (UEBT-AA550A is required.)
• Standard conformity check function The system can be operated in battery mode if the power
– Verification (export/import) supply from the outlet is interrupted due to power failure
• Print function etc.
– DICOM Print
Operating Conditions
Signal I/O Power Supply Requirements
• Transducer Connectors • Line voltage
– Transducer connectors: 4 – Japan 100 VAC ±10%
– Pencil transducer connector: 1 – USA, Canada 120 VAC±10%
(UIPC-AA550A is required.) – Europe 220 to 240 VAC ±10%
• VCR Input/Output Signals – Other 1 110 to 120 VAC ±10%
– Audio output: L, R – Other 2 220 to 240 VAC ±10%
– Audio input: L, R • Line frequency
– DVI signals for TCS 50/60 Hz ± 1 Hz
• External Video Output Signals • Power capacity
– DVI – Japan 1500 VA
• Internal USB – USA, Canada 1440 VA
– 2.0 for printer: 2 ch – Europe 1500 VA
– 2.0 for panel: 1 ch – Other 1 1500 VA
• External USB – Other 2 1500 VA
5 USB ports (2 on the rear of the main unit, 2 on the front
of the main unit (support of USB 3.0), and 1 on the monitor Environmental Conditions
side) • Operating conditions
• Ethernet – Ambient temperature: 10°C to 35°C (20°C to 35°C when a
10BASE-T/100BASE-TX/Gigabit Ethernet: 1 ch 4D transducer is used)
• SATA – Relative humidity: 35% to 80% (no condensation)
– For connecting the built-in HDD: supporting 1 HDD – Atmospheric pressure: 700 hPa to 1060 hPa
– For connecting the built-in SSD: supporting 1 SSD • Storage conditions
– For DVD drive: 1ch – Ambient temperature: -10°C to 50°C
• Footswitch (UZFS-004A is required.) – Relative humidity: 30% to 90% (no condensation)
3-switch footswitch – Atmospheric pressure: 700 hPa to 1060 hPa

DIMENSIONS, MASS, AND POWER CONSUMPTION


External dimensions Mass
mm (in) Power consumption
Unit Model name kg (lb) (approx.)
Width Height Depth (approx.)
1,188 (46.8) 864 (34.0) 359 VA
CUS-AA450 595*1 (23.4) to to to to 95 (209.4)
1,768 (69.6) 930 (36.6) 28 VA*2
Main unit Height adjusted model of 1,248 (49.1) 864 (34.0) 359 VA
CUS-AA450 (Available only in 595*1 (23.4) to to to to 95 (209.4)
USA, Europe and Australia) 1,828 (72.0) 930 (36.6) 28 VA*2
287.7 (11.3)
Sony HVO-550MD/FHD (including
DVD video recorder 212 (8.4) 105.5 (4.2) 3.2 (7.1) 43.2 W
[NTSC/PAL] the projection
section)
Sony UP-D711MD/AC 140 (5.5) 70 (2.8) 125 (4.0) 1.0 (2.2) 72 VA (printing)
Sony UP-D898MD 154 (6.1) 88 (3.5) 240 (9.4) 2.6 (5.7) 190 VA (printing)
B/W digital printer
Mitsubishi P95DW 154 (6.1) 84.5 (3.3) 239 (9.4) 2.6 (5.7) 190 VA (printing)
Mitsubishi P95DW-DC 154 (6.1) 84.5 (3.3) 130 (5.1) 1.6 (3.5) 68 VA (printing)
Mitsubishi CP30DW 212 (8.3) 125 (4.9) 425 (16.7) 7.3 (16.1) 180 VA (printing)
Color digital printer
Sony UP-D25MD 212 (8.3) 98 (3.9) 398 (15.7) 5.5 (12.1) 240 VA (printing)
*1: 533 mm when the provided transducer connector holder is removed.
*2: In Standby mode with internal battery UEBT-AA550A.
21
Safety Classification Conformance Standards
– According to the type of protection against electric • Canada
shock: CLASS I or Internally Powered Equipment – General: CAN/CSA-C22.2 No. 601.1-M90
– According to the degree of protection against electric – Collateral: CAN/CSA-C22.2 No. 60601-1-1-02 (R06)
shock: CAN/CSA-C22.2 No. 60601-1-4-02 (R2006)
EQUIPMENT WITH TYPE-BF APPLIED PARTS (Transducer, – Particular: CAN/CSA-C22.2 No. 60601-2-37-03 + A1 (R2007)
ECG electrodes, PCG sensor, Pulse sensor) • EU and other regions requiring compliance with European
– According to the degree of protection against harmful Directive 93/42/EEC and subsequent amendments
ingress of water: IPX0 (enclosed EQUIPMENT without – General: EN 60601-1: 2006 + A1: 2013
protection against ingress of water) – Collateral: EN 60601-1-2: 2007/AC: 2010
However, the footswitch is IPX8 and the transducers are – Particular: EN 60601-2-37: 2008 + A1: 2015
IPX7 (excluding the connector part). • USA
– According to the degree of safety of application in the – General: UL 60601-1: 2003 R4.06
presence of a FLAMMABLE ANESTHETIC MIXTURE WITH – Collateral: IEC 60601-1-1: 2000
AIR or WITH OXYGEN OR NITROUS OXIDE: IEC 60601-1-2: 2001 + A1: 2004
EQUIPMENT not suitable for use in the presence of a IEC 60601-1-4: 2000
FLAMMABLE ANESTHETIC MIXTURE WITH AIR or WITH – Particular: IEC 60601-2-37: 2001 + A1: 2004 + A2: 2005
OXYGEN OR NITROUS OXIDE • Other regions requiring compliance with IEC 60601-1 Ed. 2
– According to the mode of operation: – General: IEC 60601-1: 1988
CONTINUOUS OPERATION + A1: 1991 + A2: 1995
– Sterilization method – Collateral: IEC 60601-1-1: 2000
· System main unit: Not suitable for sterilization. IEC 60601-1-2: 2001 + A1: 2004
· Transducers: Sterilization methods are specified IEC 60601-1-4: 1996 + A1: 1999
in the relevant operation manuals. – Particular: IEC 60601-2-37: 2001
+ A1: 2004 + A2: 2005
• Other regions requiring compliance with IEC 60601-1 Ed. 3
– General: IEC 60601-1: 2005
– Collateral: IEC 60601-1-2: 2007
– Particular: IEC 60601-2-37: 2007
• Other regions requiring compliance with IEC 60601-1 Ed. 3.1
– General: IEC 60601-1: 2005 + A1: 2012
– Collateral: IEC 60601-1-2: 2007
– Particular: IEC 60601-2-37: 2007 + A1: 2015
* The above standards are applicable to the ultrasound system at the
time of purchase.
These standards continue to remain applicable even if the system
configuration is changed as a result of using options in combina-
tion.
The standards of the ultrasound system are applicable to transduc-
ers.

22
MPDUS0079EAB

MASS
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
System main unit USPV-AI900A Panoramic View kit 0.1 (0.2)
CUS-AA450 Aplio a450 95 (209.4) USRC-AI900A Multi-reflection Canceller kit 0.1 (0.2)
Options/Accessories for main unit USFH-AI600A Fetal Heart MPI Measurement kit 0.1 (0.2)
USCW-AI900A CW kit 0.1 (0.2) USZS-AI900A Measurement Z score kit 0.1 (0.2)
USUR-AI900A Reference Signal kit 0.1 (0.2) UAOB-Al900A OB kit 0.1 (0.2)
UJUR-AI900A Reference signal cable 0.3 (0.7) UZRI-AI900A Mounting kit for peripheral unit 1.1 (2.4)
UJUR-AI901A Reference signal cable 0.1 (0.2) UZRI-AI902A Mounting kit for peripheral unit 3.1 (6.8)
UJUR-AI902A Reference signal sensor unit 0.4 (0.9) UZRI-AA550A Mounting kit for peripheral unit 2.8 (6.2)
USSE-AI900A Stress Echo kit 0.1 (0.2) UZRI-AA551A Mounting kit for peripheral unit 4.5 (9.9)
USWT-AI900A 2D Wall Motion Tracking kit 0.1 (0.2) UZFS-004A Footswitch 0.6 (1.3)
USEF-AI600A Auto EF Measurement kit 0.1 (0.2) UZGW-008A Gel warmer 1 (2.2)
UIPC-AA550A Pencil Connector unit 0.3 (0.7) UZMK-AI900A Transducer cable hanger kit 0.9 (2.0)
UAEH-AI900A M-TEE hanger kit 1.5 (3.3) UZBK-AI900A Transducer Holder kit 0.3 (0.7)
UAEH-AI901A TEE hanger kit 1.2 (2.6) UIWL-A500A Wireless LAN kit 0.5 (1.1)
UACV-AA550A CV kit 1.0 (2.2) UIWL-AI900A Wireless LAN kit 0.5 (1.1)
UIST-AI900A STC kit 0.1 (0.2) UZPH-AI900A EV/ER Transducer Holder kit 1.6 (3.5)
USPF-AI900A 3D printer format export kit 0.1 (0.2) UIKB-AI900A Keyboard kit 1.1 (2.4)
USHI-AA550A CHI kit 0.1 (0.2) UITB-AI900A 2nd Console kit 1.1 (2.4)
USMV-AI900A 4D kit 0.1 (0.2) UZHI-AI900A Mounting kit for External HDD kit 1.0 (2.2)
USLM-AI900A Luminance kit 0.1 (0.2) USDB-AI900A DataBase for External HDD kit 0.1 (0.2)
USOB-AI900A Auto Volume Measurement kit 0.1 (0.2) UZMK-AI901A ECG Cable Hanger kit 0.1 (0.2)
USEL-AA551A Elastography-FLR kit 0.1 (0.2) UIUB-AI900A Panel USB Port kit 0.1 (0.2)
USEL-AA550A Elastography kit 0.1 (0.2) UEBT-AA550A Battery unit 13 (28.7)
USSW-AA550A Shear wave kit 0.1 (0.2) UZTB-Al900A Track Ball kit 0.2 (0.4)
USFN-AA550A Smart Fusion kit 0.1 (0.2) UITP-AA550A Tilt panel kit 1.5 (3.3)
UIFR-AA550A Magnetic Generator kit 15.0 (33.0) USHE-AI900A Online Help kit 0.1 (0.2)
UIFR-A501A Sensor kit for Fusion unit 0.1 (0.2) USPA-AI900A Protocol Assistant kit 0.1 (0.2)
USFN-AI901A Auto Registration kit 0.1 (0.2) USRI-AI600A Reference Imaging kit 0.1 (0.2)
UIOT-AI900A Auto Track CT kit 0.4 (0.9) UZKF-AI900A Local Language Key-Top kit 0.5 (1.1)
UIOT-AI901A Auto Track MR kit 0.6 (1.3) UZKG-AI900A Local Language Key-Top kit 0.5 (1.1)
UZWT-A500A Fusion Pole Cart 26 (57.3) UZKI-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-001A Mounting kit for fusion sensor 0.1 (0.2) UZKS-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-002A Mounting kit for fusion sensor 0.1 (0.2) UZKD-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-003A Mounting kit for fusion sensor 0.1 (0.2) UZKN-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-004A Mounting kit for fusion sensor 0.1 (0.2) UZKW-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-005A Mounting kit for fusion sensor 0.1 (0.2) UZKR-AI900A Local Language Key-Top kit 0.5 (1.1)
UAFS-006A Mounting kit for fusion sensor 0.1 (0.2) UP-D711MD/AC B/W printer 1.0(2.2)
UAFS-007A Mounting kit for fusion sensor 0.1 (0.2) UP-D898MD B/W printer 2.6 (5.7)
UAFS-010A Mounting kit for fusion sensor 0.1 (0.2) P95DW B/W printer 2.6 (5.7)
UISN-A500A Smart Navigation Sensor kit 1.0 (2.2) P95DW-DC B/W printer 1.6 (3.5)
Superb Micro Vascular Imaging UP-D25MD Color printer 5.5 (12.1)
USMI-AI600A 0.1 (0.2)
kit
CP30DW Color printer 7.3 (16.1)
USSS-AI900A Smart Sensor 3D kit 0.1 (0.2)
HVO-550MD/
USMP-AI900A MicroPure kit 0.1 (0.2) DVD video recorder 3.2 (7.1)
FDH

23
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
Transducers PVT-781VTE Endocavitary transducer 0.97 (2.1)
PSI-70BT Phased array transducer 0.65 (1.4) PVL-715RST Endocavitary transducer 1.04 (2.3)
PST-25BT Phased array transducer 0.8 (1.8) PLT-308BTP Linear array biopsy transducer 0.8 (1.8)
PST-30BT Phased array transducer 0.8 (1.8) PLT-704SBT Linear array transducer 0.9 (2.0)
PST-50BT Phased array transducer 0.8 (1.8) PLT-705BT Linear array transducer 0.85 (1.9)
PVT-350BTP Convex array biopsy transducer 0.95 (2.1) PLT-705BTF Linear array transducer 0.78 (1.7)
PVT-375BT Convex array transducer 0.95 (2.1) PLT-705BTH Linear array transducer 0.78 (1.7)
PVT-375SC Convex array transducer 0.95 (2.1) PLT-1005BT Linear array transducer 0.85 (1.9)
PVT-382BT Convex array transducer 0.8 (1.8) PLT-1202BT Linear array transducer 0.75 (1.7)
PVT-475BT Convex array transducer 0.8 (1.8) PLT-1204BT Linear array transducer 0.85 (1.9)
PVT-482BT Convex array transducer 0.75 (1.7) Multi-plane transesophageal
PET-508MA 1.2 (2.6)
PVT-674BT Convex array transducer 0.9 (2.0) transducer
PVT-675MVL Convex array transducer 1.1 (2.4) Multi-plane transesophageal
PET-512MA 1.32 (2.9)
transducer
PVT-675MVS Convex array transducer 1.05 (2.3)
Multi-plane transesophageal
PVT-681MVL Endocavitary transducer 1.15 (2.5) PET-512MC 1.6 (3.5)
transducer
PVT-712BT Convex array transducer 0.8 (1.8)
Multi-plane transesophageal
PVT-745BTF Convex array transducer 0.8 (1.8) PET-512MD 1.5 (3.3)
transducer
PVT-745BTH Convex array transducer 0.8 (1.8) PET-805LA Linear array transducer 1.17 (2.6)
PVT-745BTV Convex array transducer 0.78 (1.7) PC-20M CW Doppler pencil transducer 0.085 (0.2)
PVT-781VT Endocavitary transducer 0.97 (2.1) PC-50M CW Doppler pencil transducer 0.08 (0.2)

1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan

https://global.medical.canon
©Canon Medical Systems Corporation 2018. All rights reserved.
Design and specifications are subject to change without notice.
Model number: CUS-AA450 MPDUS0079EAB V1.3 2018-05 CMSC/Produced in Japan

Canon Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485.
Canon Medical Systems Corporation meets the Environmental Management System standard ISO 14001.

Aplio, Dynamic Flow, ApliPure, MicroPure, Pulse Subtraction, VolPure, TwinView and Made for Life are trademarks of Canon Medical Systems
Corporation.
Windows is a registered trademark of Microsoft Corporation in the United States and / or other countries.
omniTRAX is a trademark of CIVCO Medical Solutions.

24