Portable and Wide-Range Solid-State Transmission Densitometer For Quality Control in Fi LM Radiography
Portable and Wide-Range Solid-State Transmission Densitometer For Quality Control in Fi LM Radiography
Portable and Wide-Range Solid-State Transmission Densitometer For Quality Control in Fi LM Radiography
Received on: 23.11.09 Review completed on: 02.02.10 Accepted on: 22.02.10 DOI: 10.4103/0971-6203.62135
ABSTRACT
In biology, materials science, radiography quality control or film dosimetry in radiotherapy, a transmission densitometer is
useful for measurements of optical density. The design proposed here is oriented to quality control in radiographic films. The
instrument described here utilizes low-cost solid-state devices and is easy to construct. The use of 1-watt white light-emitting
diode in this densitometer enables low power consumption and a cold light source. Moreover, the instrument does not need a
reference light, which results in decreasing the number of parts and reducing the overall size of the apparatus.
Journal
Journalof
ofMedical
MedicalPhysics,
Physics,Vol.
Vol.35,
35,No.
No.3,
3,151-3
2010
152 Aramburo et al.: Transmission densitometer for quality control
Figure 1: System framework of the transmission densitometer Figure 2: Current source LM234Z and the Log102
logarithmic amplifier to a maximum value in the case of value from the equation (2), where VLOG102 is the voltage
the 4.00 OD suitable for A/D conversion; and in the case of from the logarithmic amplifier.
measurement without film, a minimum voltage that will be
displayed like a 0.00 OD in the LCD. The instrument takes OD = −α + β × VLOG102 (2)
power directly from four AA batteries, which produce 6 volts,
4.5 volts for the microcontroller; and the other voltages Mechanical and thermal noise
required for the operational amplifiers and a negative An important source of noise found in the measurement
voltage supply are obtained from a complementary metal- is the mechanical structure. The photodiode and the LED
oxide-semiconductor (CMOS) switched-capacitor voltage should be aligned, and only up and down movement is
converter (TL7660). The current source for the LED was allowed. The distance between LED and photodiode should
constructed with a linear voltage regulator; it is inexpensive, always be 1.7 mm. Light from the LED should pass through
and due to the feedback and an internal voltage reference of a pinhole of radius 1 mm, and the cup of the LED should
1.25 V, it keeps a constant current in the LED. be removed because the nearest distance between LED and
photodiode is required by the use of little forward current.
Calibration and measurements The thermal noise comes from the increment of the junction
With all the blocks working together, the next step is the temperature in the LED diode junction by the forward
calibration. The reference voltage for the A/D conversion current, causing relative photometric output changes,[4] as
is obtained from a 2.5-volt precision voltage reference well as loss of calibration. The current used to ensure the
integrated circuit. From its output are connected two accuracy of the measurements is only 25 mA, and the time
variable voltage dividers. With high input impedance by which the LED is turned on is about 4 seconds, to avoid
problems of accuracy due to the increment in the junction
voltage followers, the two voltages from the two voltage
temperature. The white LED used is a neutral white LED.
dividers are carried to the analog-to-digital converter of the
microcontroller. The logarithmic amplifier gives a curve of Results and Discussions
response in the range of the OD values measured, which
may be adjusted to the calibrated values of the density The calibration standard for optical density values was an
strip in each step by a linear relation. The constants of AGFA - STRUCTURIX–calibrated step tablet with 14 steps,
this simple linear relation (2) are set by means of variable identification number 6414163, and NIST traceability.
voltage dividers. In the calibration of the instrument, one The uncertainty of this strip tablet is ±0.006 OD, and the
of the potentiometers is used to adjust an offset voltage, measurements made with this instrument in each step of
represented by α, and the other is used to set the constant the tablet give a value with a maximum deviation of ±0.02
β. Finally the microcontroller obtains the optical density OD with respect to the value in the calibration tablet
Table 1: Measurements made with the for optical density measurement of radiographic films was
densitometer in each step of the tablet developed. The calibration system is straightforward and
OD measured with OD values in the Error = xmeasured − xtrue provides stability in measurements because of the circuit
the instrument calibrated step used to maintain the values of the constants of calibration
tablet with a precision voltage reference integrated circuit.
0.00 0.000 0.000 Excellent repeatability was obtained, and the uncertainty in
0.13 0.146 0.016 the measurements was found to be similar to that obtained
0.30 0.296 -0.004 by using commercial densitometers (±0.02 OD). This type
0.59 0.588 -0.002 of densitometer can be used easily in many radio-diagnostic
0.91 0.903 -0.007
and radiotherapy centers.
1.21 1.200 -0.010
1.50 1.500 0.000
1.82 1.810 -0.010
Acknowledgment
2.10 2.090 -0.010
2.40 2.390 -0.010
The authors would like to thank the Dirección de
2.69 2.680 -0.010
Investigación Medellín (DIME) of the National University
2.98 2.980 0.000 of Colombia for their economic support.
3.31 3.300 -0.010
3.58 3.590 0.010 References
3.86 3.880 0.020
4.16 4.180 0.020 1. Brown WN Jr, Birtley WB. A densitometer which records directly in
units of emulsion exposure. Rev Sci Instrum 1951;22:67-72.
2. Karkiewicz LM, Panitz JA, Fowler GL. Low drift optical densitometer.
[Table 1]. The repeatability of the measurements was Rev Sci Instrum 1980;51:1267-8.
achieved by solving the problems related to the different 3. Fu-Ming Tzu, Jung-Hua Chou. Optical density measurement of TFT-
sources of noise. The instrument provides accuracy in LCD by PMT coupled monochrome LED. IEE Electronic Materials
measurements with respect to the values of the calibration and Packaging, 2008. EMAP 2008. International Conference on
Publication Date:22-24-Oct 2008. p. 65 - 8.
standard; and in cases when a deviation was observed, the
4. Gabor M, Gergely N, Zoltan S. A Novel Procedure and Device to
difference with respect to the calibration standard was not Allow Comprehensive Characterization of Power LEDs over a Wide
higher than ±0.02 OD. Range of Temperature. IEE THERMINIC 24-26 Sept 2008 p. 89-92.
Conclusions
Source of Support: National University of Colombia, Conflict of
Interest: None declared.
A simple method using a white-light LED and photodiode