TECHNICAL MANUAL

MAXI SKY 600


CEILING LIFT SYSTEM

OCTOBER 2005 (rev.2) #001.14155.33


Arjo designs and manufactures quality engineered patient lifting equipment for the
care of the ederly anc disabled in hospitals, nursing homes and home care.
Please take the time to read the entire manual, including the section on "Safety Instructions and Warnings". It
contains important information that will allow you to take advantage of all the characteristics of your ARJO
patient lift.

Arjo Credentials
CE mark (European Union Listing)
Global operations for in-warranty and out-of-warranty service support for Maxi Sky 600.
Each finished Maxi Sky 600 is individually tested in our laboratory.
Lifting capacity is 272 kg (600 lbs). Engineering safety factor exceeds European Standards.
Complete range of patient lift slings and supports designed and manufactured by Arjo using medical
quality materials and fabrics.
FDA Quality System Regulations

APPROVALS:
CSA 601.1
UL 2601.1
EN 60601.1
EN ISO 10535 (Lifting device)
UK Test Report: DUR22107/AC/OO

Note: Arjo is constantly improving its products. For this reason, it may be possible to encounter
modifications of product without revision of this guide.

Thank you for trusting your Arjo Partners.


TABLE OF CONTENTS
INTRODUCTION............................................................................................................................................ 4
How to use this Manual .............................................................................................................................. 4
Equipment Identification ............................................................................................................................. 4
Verifying the Package Contents................................................................................................................. 4
Key to Symbols .......................................................................................................................................... 4
SAFETY INSTRUCTIONS AND WARNINGS............................................................................................... 5
DESCRIPTION OF EQUIPMENT MAXI SKY 600 ..................................................................................... 7
Lift Unit ....................................................................................................................................................... 7
What s included? ........................................................................................................................................ 7
Dimensions................................................................................................................................................. 7
Parts description......................................................................................................................................... 8
Parts description......................................................................................................................................... 8
Maintenance Requirements ....................................................................................................................... 9
Strap Replacement..................................................................................................................................... 9
High Limit Switch Replacement ...............................................................................................................14
Main Circuit Board Replacement .............................................................................................................15
Replacement of Horizontal Motor.............................................................................................................15
Battery Replacement................................................................................................................................16
Connection Diagram, Main Connector .....................................................................................................17
DIAGRAM ....................................................................................................................................................17
MAINTENANCE INSPECTION CHECKLIST..............................................................................................18
ACCESSORIES ...........................................................................................................................................19

WARNING!
All rights reserved. CONFIDENTIAL. The reproduction of this document or the transmittal in any form of any information, contained
herein, without the authority in writing of an officer of the manufacturer is prohibited.
TECHNICAL SPECIFICATIONS AND DESIGN SUBJECT TO CHANGE WITHOUT NOTICE.

Maxi Sky 600 Technical Manual 3


INTRODUCTION

Manual product number and equipment


INTRODUCTION identification number.

How to use this Manual Equipment Identification


DO NOT ATTEMPT TO USE THIS The unit's identification number (specification,
EQUIPMENT WITHOUT UNDER- model, serial number) appears on a silver
STANDING THIS MANUAL. nameplate attached to the end of the lift.
To ensure safe operation, read the entire
manual carefully, especially the section on Verifying the Package Contents
Safety Instructions and Warnings , before Upon receipt of the equipment, verify it against
installing, operating, or servicing this the packing list to ensure it is complete and
equipment. inspect the equipment for possible damage due
If anything is not completely understood, to shipping. If there is any damage, DO NOT
please contact your supplier for more USE the equipment and notify the carrier
details. Failure to comply with warnings in immediately to file a claim. Provide complete
this manual may result in injury. information concerning damage claims or
Keep this manual with the lift and refer to it shipping errors to your supplier. Include all
as required. Contents of this manual are equipment identification numbers and group part
subject to change without prior notice to numbers (if any) as described above along with
users. a full description of damaged parts.

Throughout this manual, the following symbols Key to Symbols


may appear. Pay particular attention to the
information provided under these headings. The following symbols are used on the Maxi Sky
These special annotations are easily recognized 600 lift:
as follows: This symbol indicates the power-on
indicator.
WARNING: This symbol is This symbol indicates the charging
intended to alert the user to indicator light.
hazards or unsafe practices, The symbol indicates the light
which could result in serious "Service".
bodily harm. This symbol represents the emergency
stop.

The following symbols are used on lifter s


CAUTION: This symbol is intended
attachment labels:
to alert the user of the presence of
important operating and
This symbol is required to be displayed
maintenance instructions, which
on regulated products for sale in the
could prevent product damage or
European Market. It indicates that the
possible personal injury.
product complies with applicable
European Directives related to health,
safety, environment and consumer
NOTE: This symbol offers helpful protection.
information concerning certain
operating procedures. Safe Working Load represents the
SWL maximum load the lifter is rated for safe
operation.
You can obtain additional copies of this manual
by contacting your supplier. Include the User

Maxi Sky 600 Technical Manual 4


SAFETY INSTRUCTIONS AND WARNINGS

SAFETY INSTRUCTIONS AND WARNINGS


A. GENERAL
IMPORTANT READ THESE INSTRUCTIONS CAREFULLY OR SERIOUS INJURY MAY RESULT.
KEEP THESE INSTRUCTIONS AND THE KEY PROVIDED WITH THE LIFT AT ALL TIMES.
READ OPERATING AND MAINTENANCE INSTRUCTIONS IN THIS MANUAL BEFORE
INSTALLING, OPERATING, OR SERVICING THIS EQUIPMENT.
Arjo ceiling lifts must be installed by an authorised contractor or installer.
Only a certified technician or installer can remove and reinstall the lift into the track.
Installation must be conformed to accept a 600 lbs (272 kg) load. (for more information, consult
the Installation guide)
YOUR LIFT is for transferring patients only. Do not use the lift for any other purpose.
ALWAYS carry out the daily checklist before using the lift.
Arjo ceiling lifts are specifically designed for Arjo ceiling rail systems, slings and accessories.
Slings and accessories designed by any other manufacturer are prohibited and will prevent
Arjo from being able to guarantee a safe transfer. Use only Arjo Ergofit slings and accessories
to maintain patient safety and product utility.
Arjo ceiling lifts are intended to be used for patients within the specified weight limit indicated
for the lift. Do not attempt to lift more than the weight limit indicated.
Before attempting to transfer, the patient must be assessed by a qualified professional.
Arjo ceiling lifts must be used by a caregiver with proper training to work with the patient to be
transferred. This patient lift should never be used by a patient on their own.
ONLY trained and qualified caregivers should transfer a patient. DO NOT attempt to use the lift
if you have not been properly trained to do so.
ALWAYS be prepared before attempting to transfer a patient.
DO NOT use a sling that is not recommended for the lift.
NEVER use a damaged, torn or frayed sling.
ALWAYS place the sling around the patient according to the instructions enclosed.
FOLLOW the lifting procedures outlined in this manual.
USE all controls and safety features only according to the rules specified in this manual.
Never attempt to force a control or button on the lift.
DO NOT store the charger in a shower, bath or other areas with high humidity.
DO NOT drop the patient lift or batteries. Dropping the lift or batteries may cause internal
damage that is not easily seen. If lift is suspected to be damaged, contact an authorised
technician for servicing.
IMPORTANT: Keep all components of the lift clean and dry, and have electrical and mechanical
safety checkpoints done as instructed in the Maintenance section of this manual.
Replace any precautionary or instruction labels that cannot be easily read.
Avoid violent impact during transportation.

B. SHOCK PREVENTION

DO NOT touch or use a lift with bare conductors or a damaged power cord. Electrically live
equipment can electrocute a patient. If the lift or charger has any exposed or damaged wires,
contact your local dealer immediately.
DO NOT splash or expose electric parts of the device to water or moisture.
CHECK nameplate for voltage and cycle requirements. These requirements differ by country.
Do not attempt to use the lift in an area that has a different voltage and cycle requirement.
DO NOT attempt to expose service or repair the lift, battery or charger. If any unit is
malfunctioning, contact your local dealer.
READ batteries and charger instructions thoroughly before using or storing them.

Maxi Sky 600 Technical Manual 5


SAFETY INSTRUCTIONS AND WARNINGS
C. FIRE AND EXPLOSION PREVENTION

Batteries may explode, leak and cause personal injury if not disposed of properly.
Do not place or store the battery under direct sunlight or near a heat source,
Do not dispose of in fire,
Do not short the battery terminals,
Do not incinerate,
Flush with water if acid comes in contact with skin or eyes.
Batteries must be recycled, disposed of according to local law regulations. When returning
batteries, insulate their terminals with adhesive tape, etc. Otherwise, the residual electricity in
used batteries may cause fire or explosion.

BATTERIES NEED TO BE CHARGED FOR A MINIMUM OF 8 HOURS PRIOR TO USING THIS LIFT
FOR THE FIRST TIME.

D. EQUIPMENT WARNING LABELS


INSPECT all precautionary labels on the equipment. Order and replace all labels that cannot be
easily read.

Maxi Sky 600 Technical Manual 6


DESCRIPTION OF EQUIPMENT
DESCRIPTION OF EQUIPMENT What s included?
MAXI SKY 600 ITEM QTY
Lift Unit Maxi Sky 600 lift cassette 1
The lift unit is a steel frame based system driven Batteries (inside unit) 2
by a gear reduced high torque motor.
Clip on charging station 1
OPERATING FEATURES: Hand control 1
Lifting capacity: 272 kg (600 lbs).
Spreader bar 1
Average weight: 11.5 kg (28 lbs) batteries
included. Power cord 1
Electronic soft-start and soft-stop motor 8-mm Allen key 1
control. User Manual 1
Emergency lowering device.
Emergency stopping device. Dimensions
Current limited for circuit protection in case
of overload. (In millimeters)
Safety device that stops the motor to lift
when batteries are too low.
CEILING
Lifting speed: 6 cm/s (2.3 in/sec.), 3,5
cm/sec. (1.6 in/sec) at 200 kg (440 lbs), and
2.7 cm/sec. (1.2 in/sec) at 272 kg (600 lbs).
Horizontal displacement speeds: 10, 15, 20
and 25 cm/sec. Speed by default 20 cm/sec
(6 in/sec.).
Horizontal axis motor: 24 VDC, 62 watts and
vertical axis motor 24 VDC, 110 W.
Emergency brake (in case of mechanical
failure). Operates on the same principle as a
car safety belt mechanism.
Strap length up to 2.3 m (90 inches) tested
for 1360 kg (2998 lbs.).
(STROKE)

CSA No 601.1, UL No 2601-1 and CE


certifications.
Applicable EMI standards.
ISO 10535.
Safety device that stops the lift motor in case
of abusive use that may cause overheat of
the internal components.

Maxi Sky 600 Technical Manual 7


DESCRIPTION OF EQUIPMENT

Parts description

Parts
description

Notes:
- The yellow "Charger" light flashes while charging and is solid when charge is finished.
- The green "On" light illuminates once the lift is on and ready for use; the green light flashes
when the batteries are low.
- The red Service light illuminates to confirm that the lift is in the programming mode.
- A solid red Service light illuminates in normal mode when the lift is under overheat protection
caused by abusive use.
- The red Service light flashes when servicing is required (contact customer service).
- The Programming mode button (PROG) allows you to modify the settings of the lift.

Maxi Sky 600 Technical Manual 8


MAINTENANCE
Maintenance Requirements 5) Disconnect the batteries. Remove them by
unscrewing the 4 screws using a Torx T-20
The Maxi Sky is equipped with an electronic screwdriver.
monitor. This red light will flash when a
maintenance inspection is necessary to ensure
the durability of the unit and the security of the
patient and user.
Once this light red begins to flash please contact
your distributor/representative in order to perform
the necessary maintenance inspection
requirements.

Strap Replacement
1) Remove the spreader bar.

6) Unplug the hand control from the main circuit


board.

2) Place the Maxi Sky 600 on a solid table with


the trolley facing downwards. Use a piece of
wood on each side of the trolley to stabilize it
and prevent it from moving. Make sure not
to damage the contact blades.

7) Unscrew the 4 plastic screws (A) and the 4


metal taping screws (B) with a Torx-T20
screwdriver.

3) Turn off the lift by gently pulling the red cord.


4) Unsnap and completely remove the panel by
pushing on both ends, this will expose the
batteries.

8) Remove the bottom shell by passing the


battery wires through the assigned holes.

9) Remove the top shell by passing the wheels


of the trolley through the slot.

Maxi Sky 600 Technical Manual 9


MAINTENANCE
a) Turn the drum to get the worm to the
plastic teeth.

b) Turn the transmission shaft to


10) Disconnect the main connector from circuit disengage the worm from the drum.
board. This will open the motor circuit and allows
the motor to turn more easily.

c) Completely unwind the strap.

11) Unlock the transmission with the 8 mm Allen


key.

Maxi Sky 600 Technical Manual 10


MAINTENANCE

12) Align the hole and the strap bolt.

13) Now block the rotation of the drum by inserting


the end of a screwdriver in the small opening
near the drum shaft.
14) Completely unscrew the strap bolt with a 6 mm
Allen key.

20) Grease the drum teeth with food-grade grease


(Refer to "Parts List" section to order it).
a) Engage the transmission by turning the
main shaft with the 8 mm Allen key.

15) Remove the old strap.


16) Prepare for the installation of the new strap.
Place the threaded part (interfacing) facing the
limit plate.
17) Align the loop of the strap with the hole of the
retaining bolt.
21) Reinstall the top part of the shell by placing it
on the trolley. Carefully align the contact
blades.

18) Reinsert the bolt and tighten.


19) Rewind the strap. Make sure it is in the proper
winding direction.

Maxi Sky 600 Technical Manual 11


MAINTENANCE

22) Plug the main connector back into the main 25) Pass the battery wires through their openings.
board.

26) When replacing the shell, make sure lights are


properly aligned.
23) Reinstall the main shell. Pass the strap through
the center opening of the main shell and the red
cord through the opening on the side of the
shell.

27) Remove the bottom of the shell by passing the


battery wires through the assigned holes.

24) Reconnect the hand control to the main circuit


board.

Maxi Sky 600 Technical Manual 12


MAINTENANCE

31) Snap the clips on at each end and near the


strap opening.
28) Pass the hand control wire through retaining
device.

29) Reinstall the batteries. Screw the holding


plates back in place by using a Torx T-20
screwdriver. Reconnect the battery
connectors Match the wire colors (red to
red, black to black).

32) Turn on the lift. For the first use of the unit,
apply a 25 kg (50 lbs) load to ensure the
strap is well wound around the drum. This
will prevent any slipping during use.
30) Replace the top cover. Pass the strap
through the center opening. Pass the hand
control wire through the small slot and align
the up and down buttons.

Maxi Sky 600 Technical Manual 13


MAINTENANCE

High Limit Switch Replacement


1. Repeat steps 1 -10 of the Strap Replacement
section.
2. With an extractor tool #000.00001, remove the
two blue wires of the limit switch from the main
connector.

4. Unscrew the two nuts of the limit switch using a


3/16" or 5-mm socket. Affix the new limit switch
in the same position as the original.

5. Place the plate between the frame and the strap.


Ensure that the wires pass through the round
opening in the frame. Test the system by
pushing the double thickness of the strap
between the strap roller and the plate. If the
installation is correct, you should hear the switch
clicking.
3. Unscrew the two locknuts holding the limit
6. Bolt the limit switch plate back into place.
switch plate using an 8 mm socket tool and an 8
mm (5/16) wrench. 7. Reconnect the two blue wires of the limit switch
plate to the main connector.
8. Follow steps 21 -32 of the Strap Replacement
section.

Maxi Sky 600 Technical Manual 14


MAINTENANCE

Main Circuit Board Replacement


1. Follow steps 1 - 10 of the Strap
Replacement section .
2. Unscrew the 3 holding screws using of an 8
mm socket (5/16).

4. Retract the gears from the shaft of the


motor; a 12 mm wrench could act as a lever
to remove the gear. Make sure the gear
stays greased. Use food grade grease if
needed. Refer to the "Parts list" section to
order the grease.

3. Remove the circuit board and replace it with


the new one.
4. Insert the new circuit board and screw it in.

5. Unscrew the three screws holding the motor


using a 4 mm Allen key.

5. Follow steps 21 through 32 of the Strap


Replacement section.

Replacement of Horizontal
Motor
1. Follow steps 1 through 10 of the Strap
Replacement section.
6. Change the motor.
2. Disconnect the wires from the vertical motor.
You will need an extractor #000.00001 to 7. Screw the new motor back onto its support.
perform this operation. Refer to the Replace the gear using a small plastic
connector diagram on page 23 for positions. hammer. Carefully fit the D shaped hole
with the D shaped shaft. Screw the motor
3. Unscrew the two (2) holding screws located
support back onto the trolley. Align teeth.
in the corners of the motor support using an
8 mm socket tool socket and an 8-mm
wrench (5/16). Remove the motor by
partially sliding it from its axis.

Maxi Sky 600 Technical Manual 15


MAINTENANCE
5. Install the new batteries. Make sure to match
the color codes (red to red, black to black)
6. Screw the support plates back in.
7. Follow steps 29-36 of the Strap
Replacement section.

DO NOT ATTEMPT TO USE A BATTERY


NOT AUTHORIZED BY ARJO.
ARJO batteries are specially designed for ARJO
charging systems. Attempting to use an
unauthorized battery may seriously damage the
8. Reconnect the wires to the main connector lift and/or the charger
in their proper positions. Failure to do so will
reverse motor displacement. Refer to main .
connector diagram (page 23).
9. Repeat steps 21-32 in the Strap
Replacement section.

Battery Replacement
1. Follow steps 1 through 5 of the Strap
Replacement section.

2. Disconnect the batteries.

3. Unscrew the 4 screws using a T-20 Torx


screwdriver.

4. Push out the batteries.

Maxi Sky 600 Technical Manual 16


CONNECTION DIAGRAM MAIN CONNECTORS

Connection Diagram, Main Connector

Please refer to the diagram below for the wire positions.


Position 1: Red and white wire of the positive terminal of the first battery;
Position 2: Red wire of the positive terminal of the second battery;
Position 3: Red wire of the horizontal motor;
Position 4: Blue wire for the high limit switch;
Position 5: Black wire of the negative terminal of the first battery;
Position 6: Blue wire of the horizontal motor;
Position 7: Not used;
Position 8: Blue wire of the high limit switch;
Position 9: Black and white wires of the negative terminal of the second battery;
Position 10: Not used;
Position 11: Red wire of the vertical motor;
Position 12: Blue wire of the vertical motor.

DIAGRAM

12

Maxi Sky 600 Technical Manual 17


MAINTENANCE INSPECTION CHECKLIST
MAINTENANCE INSPECTION CHECKLIST
TO BE INSPECTED BY USER
ITEM INITIALLY BEFORE EVERY EVERY EVERY EVERY 2
EVERY 2 4 YEAR YEARS
USE MONTHS MONTHS
LIFT DEVICE
Inspect for missing hardware or broken panels. X X X
Verify the presence of end stoppers. X
Inspect strap for wear. X X
Recharge batteries. X
Inspect the spreader bar on the strap of the lift X
for damage or cracks.
Inspect wheels in rail for damage, rust or X
cracks. Replace if damaged.
Clean the rail and the clip on charging station X
contacts.
Overall inspection by authorized personnel. X
SLING AND HARDWARE
Check all sling attachments for sign of wear. X
Replace if damaged.
Inspect sling material for wear or deterioration. X
Replace if damaged.
Inspect sling straps for wear. Replace if X
damaged.
Inspect sling for any defects or loose threads in X
the stitched areas. Replace if damaged.
Clean sling as indicated on the tag. When necessary
Verify emergency stop cord. X
Verify emergency lowering device. X

MAINTENANCE BY A CERTIFIED TECHNICIAN


Replace strap. X
Inspect the welding aspect on the frame to X
detect cracks.
Inspect sprockets for wear. X
Inspect connecting joints for proper attachment X
Verify the emergency brake on the drum is X
turning freely.
Verify emergency devices for good functioning. X
Load test with the SWL (Safe Working X
Load)maximum working capacity)
Inspect the track end stoppers X

Maintenance inspection points are suggested by the manufacturer. However in certain cases,
some inspections should be done on a more frequent basis. Continuing to use the lift without
replacing defective parts detected by the maintenance inspection checklist, greatly
compromises caregiver and patient safety.

Maxi Sky 600 Technical Manual 18


ACCESSORIES

ACCESSORIES
CHARGER AND CLIP ON CHARGING RAIL STOPPER
STATION Product
Product #700.15500 #700.11005

2 HOOKS SPREADER HAND CONTROL FOR Maxi


BAR Sky 600
Product #700.05495.33 6-buttons
Product #700.13650.33
2-buttons -
Product #700.13600.33

CHARGER POWER CORD UK 2-METER CHARGER EXTENSION


Product No. Product #402.15505
402.00001 USA (UL)
402.00002 Japan
402.00003 CE (Europe)
402.00004 Australia
402.00005 - UK

FIDO CONTACT STATION


Product number: 700.15400.33

Maxi Sky 600 Technical Manual 19